BD FACSDiva Option Users Guide
BD FACSDiva Option Users Guide
BD FACSDiva Option Users Guide
BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 USA Tel (877) 232-8995 Fax (408) 954-2347
Canada Tel (888) 259-0187 (905) 542-8028 Fax (905) 542-9391 [email protected] Mexico Tel (52) 5-284-8281 Fax (52) 5-284-8288
2002, Becton, Dickinson and Company. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means: electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without prior written permission from BD Biosciences. The information in this guide is subject to change without notice. BD Biosciences reserves the right to change its products and services at any time to incorporate the latest technological developments. Although this guide has been prepared with every precaution to ensure accuracy, BD Biosciences assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information. BD Biosciences welcomes customer input on corrections and suggestions for improvement. BD, the BD logo, BD CaliBRITE, BD CellQuest, BD FACSComp, BD FACSConvert, BD FACSDiVa, BD FACStation, and BD FACSVantage are trademarks of Becton, Dickinson and Company. PerCP is licensed under US Patent No. 4,876,190; Cy5.5 is licensed under US Patent Nos. 5,268,486; 5,486,616; 5,569,587; 5,569,766; and 5,627,027; APC and PE are licensed under US Patent Nos. 4,520,110; 4,859,582; 5,055,556; European Patent No. 76,695; Canadian Patent No. 1,179,942. Adobe and Acrobat are registered trademarks of Adobe Systems Incorporated. FlowJo is a trademark of Tree Star, Inc. Java is a trademark of Sun Microsystem, Inc. in the US and other countries. Macintosh is a trademark of Apple Computer, Inc., registered in the US and other countries. Microsoft and Windows are registered trademarks of Microsoft Corporation. ModFit LT is a trademark of Verity Software House, Inc. Sentinel SuperPro is a trademark of Rainbow Technologies, Inc. Sybase, Adaptive Server, Adaptive Server Anywhere, and SQL Anywhere are trademarks of Sybase, Inc or its subsidiaries. All other company and product names might be trademarks of the respective companies with which they are associated. This product is for Research Use Only. Not for use in diagnostic or therapeutic procedures.
History
Revision 341756 Rev A 330798 Rev A Date 8/01 1/02 Change Made Production release for software version 1.0. Updated for software version 2.0: enhanced performance, database redesign and data management utility, scalable data display, instrument settings features, Next button, more copy/paste ability, plot display features. Refer to the ReadMe file for details. Updated for software version 2.1: enhanced performance, workspace redesign with separable components, Browser-level folders, functioning Acquisition pointer, Sort Layout redesign, objects duplicated by dragging, drill-down gating, log decade gridlines on plots, view/hide gate boundaries, context-sensitive cursors, histogram smoothing, gate changes downloaded during sorting, automatic acquisition during record/sort, Experiment import/export, Ratio Scaling factor per ratio, Area Scaling factor per laser. Refer to the ReadMe file for details. Instrument features and operation separated from general software information. Instrument procedures updated to reflect version 2.2 of the software. Refer to the BD FACSDiVa Software Users Guide for more information.
330802 Rev A
05/02
334555 Rev A
11/02
Contents
About This Guide Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using Microsoft Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Chapter 1: Features of the BD FACSDiVa Option About the BD FACSDiVa Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BD FACSDiVa Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BD QuadraSort Tube Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BD FACSDiVa Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Chapter 2: Instrument Setup and Optimization Starting Up the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Instrument Optimization and Quality Control . . . . . . . . . . . . . . . . . . . . . . . Preparing the Alignment Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Optimizing Signals from the Primary Laser . . . . . . . . . . . . . . . . . . . . . . Optimizing Signals from the Second-Laser Intercept . . . . . . . . . . . . . . . . Optimizing Signals from the Third-Laser Intercept . . . . . . . . . . . . . . . . . Reusing the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii viii ix ix 11 12 13 14 17 18 21 22 24 24 25 30 37 44 50
iii
Chapter 3: Running Samples Performing Sample Optimization Using Instrument Setup . . . . . . . . . . . . . . . Creating the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjusting the Voltages and Threshold . . . . . . . . . . . . . . . . . . . . . . . . . . Calculating Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Recording and Analyzing Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Acquisition Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . Recording Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Analyzing Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Reusing the Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Saving the Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Chapter 4: Sorting Sorting Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sort Setup Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sort Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sort Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Conflict Resolution with BD FACSDiVa Software . . . . . . . . . . . . . . . . . . . . Yield Mask . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Purity Mask . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Phase Mask . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sort Precision Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General Sorting Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up for Sorting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Main Sorting Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up for Sorting Into Test Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Installing the Sorting Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjusting Sort Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
51 52 54 56 59 61 61 63 64 67 67 69 70 71 77 83 84 85 86 87 88 90 90 92 92 92 95
iv
Calculating the Drop Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Defining the Bead Population . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sorting Beads to Determine the Drop Delay . . . . . . . . . . . . . . . . . . . . . . Sorting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Starting and Monitoring the Sort . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up for Sorting Into a Plate or Slide . . . . . . . . . . . . . . . . . . . . . . . . . . Installing the Sorting Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjusting the Home Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Creating a Custom Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Chapter 5: DNA Analysis Criteria for DNA Experiments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CEN Optimization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Running CEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CTN Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Running CTN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Optimization for Data Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Chapter 6: Calcium Flux Intracellular Calcium Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calcium Flux Optimization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using the Time Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up the Experiment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Optimizing the Calcium Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measuring Calcium Flux . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Analyzing Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
100 100 100 102 104 104 106 106 107 108 110 113 114 115 116 118 122 122 125 127 128 129 129 130 133 135 136
Contents
Chapter 7: Troubleshooting Electronics Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Acquisition Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sorting Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix A: Optical Configurations Six-Color Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Alternate Six-Color Configuration: Five Colors + DNA . . . . . . . . . . . . . Seven-Color Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Eight-Color Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Alternate Eight-Color Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . Configuration Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Index
139 140 142 149 151 152 153 154 155 156 157 159
vi
vii
Conventions
The following tables list conventions used throughout this guide.
Table 1 Notice icons
Icon Notice Type Use Describes important features or instructions Alerts you to potential loss of data or potential damage to an application, system, or device Alerts you to potential personal injury Highlights features or hints that can save time and prevent difficulties
Tip
>
Ctrl-X
viii
Technical Assistance
For technical questions or assistance in solving a problem Read the section of the users guide specific to the operation you are performing. See Chapter 7, Troubleshooting.
If additional assistance is required, contact your local BD Biosciences technical support representative or supplier. When contacting BD Biosciences, have the following information available: product name, part number, and serial number any error messages details of recent instrument performance
For instrument support from within the US, call (877) 232-8995, prompt #2-2. For support from within Canada, call (888) 259-0187. Customers outside the US and Canada, contact your local BD representative or distributor.
ix
1
Features of the BD FACSDiVa Option
The following topics are covered in this chapter: About the BD FACSDiVa Option on page 12 Components on page 13
11
12
Components
The BD FACSDiVa option consists of the following: digital electronics digital oscilloscope BD QuadraSort sorting hardware BD FACSDiVa workstation
13
BD FACSDiVa Module
The digital electronics, oscilloscope, BD FACS AccuDrop monitor, and digital control switch are housed within the BD FACSDiVa module (Figure 1-2). The electronics are adjusted by your field service engineer during installation and do not require any user maintenance.
SAVE/RECALL
MEASURE
ACQUIRE MENU
/RESET
UTILITY
CURSOR
DISPLAY
HARDCOPY
RUN/STOP
VERTICAL
POSITION POSITION
HORIZONTAL
POSITION
TRIGGER
LEVEL
MATH MENU
VOLTS/DIV
VOLTS/DIV
SEC/DIV
SET LEVEL TO 50
FORCE TRIGGER
5V
2mV
5V
2mV
5s
5ns
TRIGGER VIEW
Digital Oscilloscope
The digital oscilloscope displays the digital drop-drive waveform and the digital drop charges (Figure 1-3 on page 15). Use the digital oscilloscope to verify charging of the side streams and the amplitude level. Refer to the documentation provided with the oscilloscope for oscilloscope adjustments.
14
SAVE/RECALL
MEASURE
ACQUIRE MENU
AUTOSET
UTILITY
CURSOR
DISPLAY
HARDCOPY
RUN/STOP
VERTICAL
POSITION POSITION
HORIZONTAL
POSITION
TRIGGER
LEVEL
digital amplitude
Source Type
CURSOR 1 MATH MENU CURSOR 2 CH2 MENU HORIZONTAL MENU HOLDOFF TRIGGER MENU
CH1 Freq
CH1 MENU
drop charge
VOLTS/DIV
VOLTS/DIV
SEC/DIV
SET LEVEL TO 50
FORCE TRIGGER
5V
2mV
5V
2mV
5s
5ns
TRIGGER VIEW
AccuDrop Monitor
The AccuDrop monitor shows an image of the streams illuminated by the AccuDrop excitation source, a diode laser. Use the monitor to accurately set the drop delay and set up the streams for sorting, as described in Calculating the Drop Delay on page 100. Refer to the BD FACS AccuDrop Users Guide for specific information about the AccuDrop option.
Digital Operation
During operation in digital mode, analog settings for gain, threshold, compensation, and the event rate are displayed on the analog oscilloscope. Because threshold and gain are controlled differently depending on the electronic
15
mode, these settings can differ from those displayed by the digital electronics. For an accurate event count, monitor the Acquisition Status frame in BD FACSDiVa software, rather than the analog oscilloscope. When operating in digital mode, the analog oscilloscope reflects adjustments made to the photomultiplier (PMT) voltages in BD FACSDiVa software. These adjustments will also change the data displayed in BD CellQuest (or BD CellQuest Pro) software. However, analog data is subject to any gains set in BD CellQuest software and thus might differ from the data displayed on the BD FACSDiVa workstation. For this reason, it is important to use signals displayed on the analog oscilloscope and in BD CellQuest software for troubleshooting purposes only. CAUTION Because digital data processing is different from analog data processing, do not save BD CellQuest files collected while operating in digital mode. NOTE In digital mode, most controls on the instrument control panel are inactive (see Figure 1-4). Equivalent controls can now be found in BD FACSDiVa software.
BRIGHT
H HOLD
STREAM CONTROLS
ON OFF
INDEX SORT
16
Analog Operation
During operation in analog mode, instrument controls on the BD FACSVantage SE control panel are enabled. Any changes made within BD FACSDiVa software will not be registered on the analog oscilloscope, nor in BD CellQuest software. However, changes made in BD FACSDiVa software will be reflected on the digital oscilloscope.
17
BD FACSDiVa Workstation
The BD FACSDiVa workstation controls the BD FACSVantage SE flow cytometer when operated in digital mode. It consists of a Windows 2000 computer running BD FACSDiVa acquisition and analysis software. Refer to the BD FACSDiVa Software Users Guide for complete instructions on using the software. The BD FACSDiVa workstation is operated independently of the BD FACStation workstation with a separate keyboard and mouse.
18
If you are accustomed to working with analog signals, note that digital data looks very different. See Figure 1-7.
BD Symphony plot
CAUTION Use caution when viewing digital data within BD CellQuest software. Data viewed in BD CellQuest software results from instrument setting adjustments made in BD FACSDiVa software. Digital instrument settings are not updated in BD CellQuest instrument settings files or the resulting FCS files. BD does not recommend saving BD CellQuest files during digital operation; use BD CellQuest signals for troubleshooting purposes only.
19
20
2
Instrument Setup and Optimization
This chapter describes how to start up the BD FACSVantage SE instrument for operation in digital mode, and how to use BD FACSDiVa software to optimize the instrument before acquisition. The following topics are covered in this chapter: Starting Up the Instrument on page 22 Instrument Optimization and Quality Control on page 24 Setting Up the Experiment on page 25 Optimizing Signals from the Primary Laser on page 30 Optimizing Signals from the Second-Laser Intercept on page 37 Optimizing Signals from the Third-Laser Intercept on page 44 Reusing the Experiment on page 50
21
Before beginning this chapter, you should be familiar with the following:
General instrument setup procedures; refer to the BD FACSVantage SE Users Guide for detailed information on instrument setup. BD FACSDiVa workspace components BD FACSDiVa instrument and acquisition controls BD FACSDiVa gating and statistics tools
Refer to the BD FACSDiVa Software Users Guide for information on operating this software.
3 Open the vacuum source and air supply. 4 Verify that the sheath container is full and the waste container is empty.
WARNING The waste container contents might be biohazardous. Expose waste container contents to bleach (10% of total volume) before disposal. Dispose of waste in accordance with local regulations. Use proper precaution and wear suitable protective clothing, eyewear, and gloves.
22
CAUTION Before proceeding with step 5, verify that the BD QuadraSort tube holder is not installed. If you turn on the power with the hardware installed, sorting hardware for the BD CloneCyt Plus option could catch on the BD QuadraSort tube holder. As a result, the motor could be damaged.
5 Turn on the cytometer main power switch. 6 Switch the digital control switch to ON, if necessary.
Turn the switch to ON to operate in digital mode.
7 Turn on the computer main power switch and start up the BD FACSDiVa
workstation; start up the BD FACStation computer, if needed. After logging on to Windows, launch BD FACSDiVa software by doubleclicking the shortcut on the desktop. Verify that the instrument is connected by checking the Instrument frame in the software. The message Instrument Connected appears after the cytometer connects to the workstation (this can take several minutes). If the message Instrument Disconnected remains, refer to the troubleshooting suggestions in the software manual.
8 Switch on the main pressure toggle switch. 9 Remove the sample tube from the sample injection port (SIP). 10 Turn the Fluidics Control knob to Fill for 1020 seconds to remove air
bubbles.
23
When instrument settings and the alignment sample are kept constant, changes in the means and CVs indicate variations in instrument performance over time. Keep track of means and CVs in a quality control (QC) log. QC data should be analyzed for trends over the past 3060 runs. NOTE QC results are affected by laser and fluidics performance. BD Biosciences strongly recommends following the laser and fluidics maintenance procedures in the BD FACSVantage SE Users Guide. CAUTION Do not place heavy objects or lean on the instrument console while performing this procedure. Unnecessary pressure on the instrument during or after instrument optimization could disrupt alignment, and it would have to be performed again.
24
! Tip
To place an Experiment inside a folder, select the folder before creating the Experiment.
25
5 Rename the new Specimen with todays date; rename the first Tube 488 nm.
This Tube will be used to optimize signals from the first laser. Your Experiment should look similar to that shown in the figure at the right.
6 With the 488 nm Tube selected in the Browser, click on the Instr. Settings >
Parameters tab in the Inspector and delete any unnecessary parameters.
! Tip
Save space in the database by listing only appropriate parameters. For example, do not list UV parameters.
NOTE When aligning with beads, you might also need to deselect the checkbox for PerCP-Cy5.5.
26
8 In the Acquisition Controls frame, set the Events to Record to 10,000 evt
and the Events to Display to 500 evt.
! Tip
Decreasing the number of displayed events will increase the data refresh rate.
! Tip
You can later duplicate the Specimen for a subsequent optimization and keep the plots for each day on a separate worksheet.
! Tip
Use the sticky buttons feature to easily create multiple plots. Press the Control key, select a plot tool, and click multiple times on the worksheet. Each click will create a new plot. To unstick the plot tool, select another tool or press the Esc key.
! Tip
Use the Resize tool to resize multiple plots simultaneously. Refer to the software manual for more information.
12 Right-click the 488 nm Tube in the Browser and choose Create Statistics View.
27
Set up the Statistics tab to display the mean and CV for FSC and each fluorescence channel. Set Decimal Places to 1 for the CVs and 0 for the means.
Click OK.
28
14 Resize the Statistics view to fill the remaining 1/3 of the worksheet.
Your worksheet should look similar to the example shown in Figure 2-1.
29
1 2 3 4 5 6 7 8 9 10
X controlmoves nozzle right and left Y controlmoves nozzle away from and toward you Theta lockprevents theta control from moving Z controlmoves nozzle up and down in an arc Theta controlmoves nozzle along an arc so stream moves right and left Alpha controlmoves nozzle along an arc so stream moves away from and toward you Fluorescence channel height adjustment wheelraises and lowers fluorescence objective lens FSC obscuration bar vertical adjustmentmoves FSC obscuration bar up and down Excitation Beam Focus wheelmoves beam focus lens to adjust laser beam focal point on the sample stream Fluorescence Focus control knobmoves objective lens to adjust focal point of the fluorescence image (access through the upper side door)
30
3 Install the alignment sample onto the cytometer; turn the Fluidics Control
knob to Run.
4 Set the sample differential pressure to its standard level for this alignment
procedure.
5 Verify that the green Acquisition pointer is in front of the 488 nm Tube in
the Browser; click once on the pointer to begin acquisition. Alternatively, click the Acquire button in the Acquisition Controls frame; events appear in the plots.
31
NOTE During digital operation, use the analog pulses displayed on the analog oscilloscope for troubleshooting purposes only. If necessary, adjust the threshold gains and log settings in BD CellQuest or BD CellQuest Pro software to change the analog pulse display. These will not affect digital data but can help in alignment.
! Tip
If you dont see any signal, increase the PMT voltage or turn on Log before adjusting the stream controls.
8 Close the FL1 iris incrementally as you continue optimizing the FITC signal.
Continue adjusting the controls and closing the iris until the iris is completely closed.
9 With the iris closed, adjust the Y control and Excitation Beam Focus wheel
for maximum FITC signal.
10 Compare the FITC signal intensity with the iris open and closed.
You should not lose more than half the maximum FITC signal intensity with the iris completely closed.
11 Open the FL1 iris and adjust the beam splitters for maximum fluorescence
intensity. On the appropriate plots, maximize the signal for SSC, PerCP-Cy5.5, and PE.
32
12 Adjust the obscuration bars for minimum FSC and SSC noise, if necessary. 13 Verify the trajectory of the fluid stream.
The stream should remain in the center or front third of the stream aspirator. If necessary, adjust the Alpha and Theta controls to correctly position the fluid stream. After adjusting the controls, repeat steps 6 through 12.
1 Click on the 488 nm Tube in the Browser and display the Parameters tab of
the Instrument Settings Inspector.
2 Select the height (H) checkbox for the FITC parameter. 3 Change the axis on the PerCP-Cy5.5 histogram to FITC-H. 4 Click the Laser tab in the Instrument frame. 5 Adjust Area Scaling for the first laser until the FITC-A intensity is similar to
the FITC-H intensity. See Figure 2-3 on page 34 for an example.
33
Figure 2-3 Primary laser area scaling before (left) and after (right) adjustment
6 Change the FITC-H histogram plot to PerCP-Cy5.5-A. 7 Deselect the checkbox for the FITC-H parameter in the Instrument Settings
Inspector.
34
3 Open the iris; Ctrl-click the Acquisition pointer to begin recording data.
Alternatively, click the Record button in the Acquisition Controls frame.
4 After recording is complete, draw an Interval gate around each peak on the
scatter and fluorescence histograms. Adjust the gate on the FITC histogram, if needed.
! Tip
Ctrl-Click the Auto-Interval Gate tool and create all Interval gates. Then press Escape to unstick the button, and readjust the gates as needed.
35
7 Copy the results into the QC log and print the worksheet for your records.
Keep a record of the primary laser results for future reference. See the following figure for an example.
36
DO5 FL3 OBS2 = triple-laser beam splitter D03 OBS4 SSC DO4 DO2 OBS3 DO1 FL3
OBS2 = half-mirror
NOTE Your optical stage might be configured differently from the examples shown in the figure.
37
2 With the 633 nm Tube selected in the Browser, click on the Instr. Settings >
Parameters tab in the Inspector and select appropriate Area parameters.
! Tip
Save space in the database by listing only appropriate parameters. For example, do not list UV parameters. Change any listed parameter by clicking in the Parameter field and choosing a new parameter from the drop-down menu that appears.
3 Deselect the Log checkboxes for all fluorescent parameters. 4 Add additional vertical pages to the worksheet, as needed.
Click on the worksheet; change the number of pages in the Inspector.
38
6 Right-click on any plot and choose Create Statistics View. 7 Edit the Statistics view to display the mean and CV for each fluorescence
channel. On the Population tab, deselect #Events and %Parent. On the Statistics tab, set Decimal Places to 1 for the CVs and 0 for the means.
2 Click to move the green Acquisition pointer in front of the 633 nm Tube in
the Browser; click Acquire, if needed. Alternatively, click the Next button in the Acquisition Controls frame. After starting acquisition, events appear in the plots.
! Tip
If you do not see any signal, increase the appropriate PMT voltage or change to Log. If you still do not see any signal, verify the Delay setting for the second laser (step 6 on page 40). NOTE During digital operation, use the analog pulses displayed on the analog oscilloscope for troubleshooting purposes only.
3 Adjust the appropriate beam splitter(s) to direct the 633 nm laser signal to
the appropriate PMT(s).
4 Adjust the two rear rotators (rear knobs) to obtain the highest signal
intensity and lowest CV on the fluorescence plots. The knobs are located on the beam steering prism assembly of the laser being optimized.
39
5 If necessary, adjust the two front translators (front knobs) to obtain the
highest signal intensity and lowest CV. Minor adjustment of the translators might be needed after performing laser alignment. The translators do not need adjusting on a daily basis.
! Tip
As a troubleshooting measure, estimate the laser delay setting from the analog oscilloscope using the Special Setup feature in BD CellQuest software. Draw an Interval gate around each peak on the fluorescence histograms. Use the Population Hierarchy to rename each population defined by the Interval gates. For example, change P1 to APC and P2 to APC-Cy7. Select the Laser tab in the Instrument frame. Adjust the second laser Delay setting in increments of 1 to obtain the highest mean channel for the fluorescent populations (Figure 2-7 on page 41). Change the Window Extension to zero to capture pulses within the narrowest time window.
40
Adjust the second laser Delay setting in increments of 0.1 to obtain the highest mean channel for the fluorescent populations.
Figure 2-7 Second laser delay before (left) and after (right) adjustment
Reset the Window Extension to the appropriate setting (typically 2). A larger Window Extension allows more flexibility for capturing pulses.
1 Select the height (H) checkbox for APC-Cy7 in the Parameters tab of the
Instrument Settings Inspector.
3 In the Laser tab of the Instrument frame, adjust Area Scaling for the second
laser until the APC-Cy7-A intensity is similar to the APC-Cy7-H intensity (Figure 2-8 on page 42).
41
Figure 2-8 Second laser area scaling before (left) and after (right) adjustment
4 (Optional) Delete the APC-Cy7-H histogram. 5 Deselect the checkbox for the APC-Cy7-H parameter in the Instrument
Settings Inspector.
42
3 Ctrl-click the Acquisition pointer to record data for the 633 nm Tube.
Alternatively, click the Record button in the Acquisition Controls frame.
43
! Tip
Save space in the database by listing only appropriate parameters. For example, do not list fluorescence parameters for the primary laser. Change any listed parameter by clicking in the Parameter field and choosing a new parameter from the drop-down menu that appears.
3 Deselect the Log checkbox for all fluorescent parameters. 4 Add additional vertical pages to the worksheet, as needed.
Click on the worksheet; change the number of pages in the Inspector.
44
6 Right-click on any plot and choose Create Statistics View. 7 Edit the Statistics view to display the mean and CV for each UV channel.
On the Population tab, deselect #Events and %Parent. On the Statistics tab, set Decimal Places to 1 for the CVs and 0 for the means.
2 Click to move the green Acquisition pointer in front of the UV Tube in the
Browser; click Acquire, if needed. Alternatively, click the Next button in the Acquisition Controls frame. After starting acquisition, events appear in the plots.
! Tip
If you do not see any signal, increase the appropriate PMT voltage or change to Log. If you still do not see any signal, verify the Delay setting for the third laser (see step 6 on page 46). NOTE During digital operation, use the analog pulses displayed on the analog oscilloscope for troubleshooting purposes only.
45
4 Adjust the two rear rotators (rear knobs) to obtain the highest signal
intensity and lowest CV on the UV plots. The knobs are located on the beam steering prism assembly of the laser being optimized.
5 If necessary, adjust the two front translators (front knobs) to obtain the
highest signal intensity and lowest CV. Minor adjustment of the translators might be needed after performing laser alignment. The translators do not need adjusting on a daily basis.
! Tip
As a troubleshooting measure, estimate the laser delay setting from the analog oscilloscope using the Special Setup feature in BD CellQuest software. Draw an Interval gate around each peak on the UV histograms. Use the Population Hierarchy to rename each population defined by the Interval gates. For example, change P1 to UV1 and P2 to UV2. Select the Laser tab in the Instrument frame. Adjust the third laser Delay setting in increments of 1 to obtain the highest mean channel for the UV populations (Figure 2-10 on page 47). Change the Window Extension to zero to capture pulses within the narrowest time window. Adjust the third laser Delay setting in increments of 0.1 to obtain the highest mean channel for the UV populations (Figure 2-10 on page 47).
46
Figure 2-10 Third-laser delay before (left) and after (right) adjustment
Reset the Window Extension to the appropriate setting (typically 2). A larger Window Extension allows more flexibility for capturing pulses.
1 Select the height (H) checkbox for the UV2 parameter in the Instrument
Settings Inspector.
3 In the Laser tab of the Instrument frame, adjust Area Scaling for the third
laser until the UV2-A intensity is similar to the UV2-H intensity (Figure 2-11 on page 48).
47
Figure 2-11 Third-laser area scaling before (left) and after (right) adjustment
4 (Optional) Delete the UV2-H histogram. 5 Deselect the checkbox for the UV2-H parameter in the Instrument Settings
Inspector.
48
49
1 Open the instrument optimization Experiment. 2 Create a new worksheet. 3 Right-click the Specimen and choose Duplicate without Data.
The three optimization Tubes appear under the new Specimen; empty plots and Statistics views appear on the new worksheet.
4 Rename the Specimen and the worksheet. 5 Double-click the 488 nm Tube to locate the plots. 6 Continue optimization.
50
3
Running Samples
This chapter describes how to use BD FACSDiVa software to record and analyze sample data. If this is your first time using BD FACSDiVa software, BD recommends that you first practice the steps in this chapter using BD CaliBRITE beads. This exercise will familiarize you with digital data and help you establish target channel values. The following topics are covered in this chapter: Performing Sample Optimization Using Instrument Setup on page 52 Recording and Analyzing Data on page 61
51
1 Create an Experiment, specify parameters, and add compensation Tubes. 2 Adjust the FSC and SSC voltages and the FSC threshold. 3 Gate the population of interest. 4 Adjust fluorescence detectors. 5 Calculate compensation.
52
This section describes how to perform sample optimization using the Instrument Setup feature. Instrument Setup can be used to automatically calculate compensation settings. For more information about this feature, refer to the BD FACSDiVa Software Users Guide. If you are performing compensation manually, not all steps will apply. This section presents an example of sample optimization using a lysed, washed, whole-blood sample (LWB) stained with the following mouse anti-human antibodies: Unstained control Mouse IgG1 FITC/Mouse IgG1 PE/ Mouse IgG1 PerCP-Cy5.5/Mouse IgG1 APC/ Mouse IgG1 APC-Cy7 CD8 FITC/Mouse IgG1 PE/ Mouse IgG1 PerCP-Cy5.5/Mouse IgG1 APC/ Mouse IgG1 APC-Cy7 Mouse IgG1 FITC/CD8 PE/ Mouse IgG1 PerCP-Cy5.5/Mouse IgG1 APC/ Mouse IgG1 APC-Cy7 Mouse IgG1 FITC/Mouse IgG1 PE/ CD8 PerCP-Cy5.5/Mouse IgG1 APC/Mouse IgG1 APC-Cy7 Mouse IgG1 FITC/Mouse IgG1 PE/ Mouse IgG1 PerCP-Cy5.5/CD8 APC/Mouse IgG1 APC-Cy7 Mouse IgG1 FITC/Mouse IgG1 PE/ Mouse IgG1 PerCP-Cy5.5/Mouse IgG1 APC/ CD8 APC-Cy7
FITC-stained control
PE-stained control
PerCP-Cy5.5stained control
APC-stained control
APC-Cy7stained control
53
3 (Optional) Press Ctrl-N to add a new folder to the Browser; rename the
folder with your name. Alternatively, you can name the folder Sample Optimization or you can create a Sample Optimization folder inside another folder.
! Tip
To place an Experiment inside a folder, select the folder before creating the Experiment.
54
55
compensation values. This is especially noticeable in tandem conjugates due to lot-to-lot variation. Refer to the BD FACSDiVa Software Users Guide for more information about this feature.
1 Install the unstained control tube on the cytometer. 2 Verify that the green Acquisition pointer is in front of the Unstained
Control Tube in the Browser; click once on the pointer to begin acquisition. Alternatively, click the Acquire button in the Acquisition Controls frame.
3 Adjust the FSC and SSC voltages to appropriately display the scatter
properties of the LWB sample (Figure 3-2). Select the Unstained Control Tube in the Browser. Click the Instr. Settings tab in the Inspector, and then click the Parameters tab. Click the up and down arrows or drag the sliders to adjust the values.
56
4 Click the Threshold tab and adjust the FSC Threshold, if needed.
Set the threshold to remove most of the debris without cutting off the lymphocyte population (Figure 3-2 on page 56).
6 Select all fluorescence histograms on the Unstained Control worksheet. 7 In the Plot Inspector, select the Show Grid checkbox (Figure 3-3).
checkbox selected
In a four-log display, values are displayed from 26262,143. Thus, the first log decade ranges from 26262. Gridlines are used to delineate log decades on plots.
57
8 Optimize the voltages to place the negative population for each fluorescent
parameter within the first log decade (Figure 3-4). Refer to your instrument manual if you need assistance optimizing the fluorescent signal.
9 Click Record; when all events have been recorded, remove the unstained
control tube from the cytometer. NOTE Do not change the PMT voltages after the first compensation Tube has been recorded. In order to calculate compensation, all Tubes must be recorded with the same PMT voltage settings. If you need to adjust the PMT voltage for a subsequent compensation Tube, you will need to record all compensation Tubes again.
58
Calculating Compensation
Before you can calculate compensation, you will need to record data for each single-stained control.
1 Install the first stained control tube onto the cytometer. 2 Click the Next button; click Acquire, if needed.
Alternatively, move the Acquisition pointer to the next Tube. If the Start acquisition on pointer change User Preference is enabled (default option), acquisition starts automatically when the pointer is moved.
3 Adjust the P1 gate to encircle the lymphocytes. 4 Ctrl-click the Acquisition pointer to record data. 5 When recording is finished, install the next stained control tube onto the
cytometer.
6 Repeat steps 2 through 5 until data for all stained control tubes has been
recorded. Now that data has been recorded, you need to create regions around the fluorescence-positive populations on the histogram for each stained control Tube.
59
! Tip
Use the Auto-Interval tool to quickly create an Interval gate around the population. To keep the tool selected so you dont have to click the tool for each remaining Tube, hold down the Control key the first time you click the tool. The tool remains selected until you press Esc or select another tool.
9 Double-click the next Stained Control Tube in the Browser to locate the
corresponding plots on the worksheet.
! Tip
To keep track of compensation Setups, include the Experiment name, date, or both in the Setup name. The named Setup is automatically linked to the Experiments instrument settings.
60
1 Create a new Specimen; rename the Specimen LWB. 2 Create two Tubes under the LWB Specimen; rename the Tubes
appropriately. For example, T/B/NK_001 and T/B/NK_002.
61
4 Use the Experiment Layout dialog box to define labels and to specify the
number of events to record for each Tube. Parameter labels are defined in the Experiment Layout view. Labels will appear on the plot axes and in all statistics views. Choose Experiment > Experiment Layout. On the Labels tab, enter appropriate labels for the Tube. For example, enter CD4 in the FITC field; use the Tab key to move to the next field.
On the Acquisition tab, enter 10,000 events for Tubes 001 and 002. Notice that the Acq. tab in the Inspector updates automatically.
Click OK.
62
! Tip
plots.
Ctrl-click the Plot tool to keep the tool selected until you create all
Recording Data
1 Install the first sample tube onto the cytometer. 2 Move the Acquisition pointer to the first Tube. 3 While data is being acquired, draw a region around the lymphocytes; set
the other plots to show data from the Lymphocyte population.
4 Ctrl-click the Acquisition pointer to record data. 5 When all events have been recorded, remove the tube from the cytometer. 6 Install the next sample tube onto the cytometer; move the pointer to the
corresponding Tube in the Browser.
8 Repeat steps 5 through 7 until data has been recorded for all tubes. 9 (Optional) Print the Experiment-level instrument settings.
Right-click the instrument settings icon and choose Print.
63
Analyzing Data
This section describes how to set up plots, gates, and a statistics view to analyze the recorded data. By the end of this section, your analysis should look similar to that shown in Figure 3-6 on page 66.
1 Create a new Acquisition Template; rename the template T/B/NK Analysis. 2 Select the first Tube under the LWB Specimen and create the following plots
on the Analysis template: FSC vs SSC CD3 PerCP-Cy5.5 vs CD16+56 PE CD3 PerCP-Cy5.5 vs CD19 APC CD3 PerCP-Cy5.5 vs CD8 APC-Cy7 CD3 PerCP-Cy5.5 vs CD4 FITC
3 Create a Population Hierarchy and a statistics view and move them below
the plots on the worksheet.
5 Select all plots except the FSC vs SSC plot and specify to show only the
Lymphocyte population.
6 Select all plots and click the Title tab in the Plot Inspector; select the
checkboxes to display the Tube and Population names in the plot titles.
64
7 Edit the statistics view to show only the Lymphocyte population and to
display the mean for all fluorochromes.
65
66
1 Move the Acquisition pointer to the next Tube under the LWB Specimen. 2 View the data on the Acquisition Template; make adjustments to gates, as
needed. NOTE Adjustments will also apply to the next Tube that is viewed on the template. If you dont want to alter the template, save the analysis as described in the next section and make adjustments on the Tubes worksheet.
1 Expand the T/B/NK Analysis Template in the Browser. 2 Right-click on the Analysis object and choose Copy.
67
) to switch to the
4 Create a new worksheet for the destination Tube; rename the worksheet
appropriately.
5 Collapse the Tube in the Browser; right-click the Tube icon and choose
Paste. The elements on the template are copied to the new worksheet. You can view the analysis by double-clicking the Tube in the Browser.
! Tip
Apply the analysis to multiple Tubes by selecting more than one Tube before you choose Paste. Enable the Tube-specific worksheet user preference to automatically create a new worksheet for the pasted Analysis objects.
68
4
Sorting
You can program BD FACSDiVa software to sort a specified number of particles from multiple populations into a variety of sorting devices, including tubes, plates, and slides. Specialized BD QuadraSort hardware, included with the option, provides the ability to sort into four tubes simultaneously. Up to four defined populations can be sorted into each tube, allowing up to 16 populations to be sorted at one time. Any subpopulation can be used for sorting, including populations defined by Quadrant gates, Interval gates, or derived gates. A single sort population can be defined by up to eight gates. The following topics are covered in this chapter: Sorting Controls on page 70 Conflict Resolution with BD FACSDiVa Software on page 84 General Sorting Overview on page 90 Setting Up for Sorting Into Test Tubes on page 92 Calculating the Drop Delay on page 100 Sorting on page 104 Setting Up for Sorting Into a Plate or Slide on page 106
69
Sorting Controls
There are four sets of sorting controls provided with the BD FACSDiVa option. Controls on the BD FACSVantage SE control panelonly shaded controls are active in Digital mode (Refer to the BD FACSVantage SE Users Guide for a full description.)
H HOLD STREAM CONTROLS
V HOLD
ON OFF
INDEX SORT
Drop Drive and side stream controls in the Sort Setup frame (See Sort Setup Controls on page 71.) Sorting instructions, controls, and counters in the Sort Layout window (See Sort Layout on page 77.) Sort Precision modes, Sort Layouts, and sorting device commands in the Sort menu Choose commands in the Sort menu for the following. Sort Precisionopens a dialog box where you can choose or define a Sort Precision mode for handling sorting conflicts; see Conflict Resolution with BD FACSDiVa Software on page 84. New Sort Layoutopens the default 2-Tube Sort Layout where other sorting instructions can be specified; see Sort Layout on page 77.
70
Open Sort Layoutopens an existing Sort Layout. A Sort Layout must be selected in the Browser for this menu command to be enabled. Alternatively, double-click any Sort Layout to open it. Home Deviceopens a dialog box containing commands to move the tray support arm, either manually or to the home position; see Adjusting the Home Location on page 108. Custom Devicesopens a dialog box where custom devices can be defined; see Creating a Custom Device on page 110. Sort Reportdisplays a report showing the sort settings, acquisition counters, and Sort Layout information from the current sort. See Sort Report on page 83.
The Sort Setup frame contains drop-drive controls in the Breakoff tab and stream charging controls in the Streams tab. The values entered in each tab apply globally to BD FACSDiVa software and are not saved with Experiments or Tubes. The default values at startup are the last set of values used by the software.
Chapter 4: Sorting
71
Sort setup values for different applications can be saved and recalled using the Sort Setup option in the Instrument menu.
Default settings are provided for high-, medium-, and low-pressure sorts; additional custom settings can be defined and saved. See Saving and Recalling Sort Setup Values on page 76.
Adjusting Settings
Within each tab, make adjustments using the software controls or your keyboard. To adjust a setting, click in the field containing the value you want to change, or press the indicated function key (eg, F2 for Amplitude). NOTE To use the function keys, the Sorting frame must be active (highlighted).
Do one of the following to change any value: Select the value in the field and enter a new value. Click the slider button next to the arrow keys to access a slider control. Click the pointer in the slider bar and drag it to a new value.
slider control
Use the mouse to click the up and down arrows or press the arrow keys on your keyboard to increase or decrease the values in small increments.
72
Hold down the Control key while clicking the arrows or pressing the keys to increase the increments of the values.
Drop Drive on
NOTE The drop drive should be off when an experiment does not involve sorting. Test Sortgenerates test side streams based on the Drop sequence specified in the Breakoff tab when the button is clicked. The drop drive must be on to enable the Test Sort button.
Test Sort on
! Tip
Attenuationdecreases the amplitude of the drop drive oscillation when the button is clicked. At startup, attenuation defaults to off. As a general rule, turn on attenuation when sorting below 30 psi.
Attenuation on
Attenuation off
Chapter 4: Sorting
73
Breakoff values are used for the following: Frequencyadjusts the frequency of the drop drive from 1.0102.0 KHz; determines the number of drops formed per second. Amplitudeadjusts the amplitude or intensity of the drop drive from 0.080.0 V. Phaseadjusts the phase between drop generation and charging of the droplets from 0360 degrees. The selected value is sent to both the dropcharging electrode and the drop strobe. Drop delaysets the amount of time between when an event is measured and the breakoff point from 10140 drops. The drop delay value determines which drop will be deflected. Drop sequencedetermines the sequence of test pulses. Changes to the numerical field are not saved after restarting the application.
74
Far Left, Leftadjust the Far Left and Left streams by changing the amount of charge applied to sorted droplets, from 0100%. The Far Left stream is used for sorting onto a plate or slide and for four-way sorting; the Left stream is used for two- and four-way sorting. Right, Far Rightadjust the Far Right and Right streams by changing the amount of charge applied to sorted droplets, from 0100%. The Far Right stream is used for four-way sorting; the Right stream is used for two- and four-way sorting. 2nd, 3rd, 4th dropapply a correction factor for the drop charge as a percentage of the previous drop, from 100 to 100%. NOTE BD recommends that the amount of charge applied to the far left and far right streams not exceed 80%. When the charge is set to 100%, there is no remaining charge available for stream shaping via the 2nd, 3rd, and 4th drop correction factors.
Chapter 4: Sorting
75
Choose Instrument > Sort Setup > Recall to switch between predefined value sets. Choose a named setup from the drop-down menu, and then click OK.
Choose Instrument > Sort Setup > Delete to delete a predefined value set. Choose a named setup from the drop-down menu, and then click OK.
76
Sort Layout
A Sort Layout is a floating window containing all sorting instructions and controls. The Sort Layout designates which device will be used to collect sorted particles and which particles will be sorted into each sort location. Up to four sort counters can be displayed in the window to give ongoing status during a sort. Only one Sort Layout can be open at a time, but you can create several layouts for a single Tube, as long as each Sort Layout has a different name. Sort Layouts can also be added to Acquisition Templates. Sort Layouts are available for up to ten default collection devices; additional custom devices can be defined. (See Creating a Custom Device on page 110.)
Examples of Sort Layouts for different devices are shown in the following figures; instructions for setting up a Sort Layout can be found in Setting Up the Experiment on page 104.
Chapter 4: Sorting
77
collection device
Figure 4-2 Sort Layout for tubes (top) and for 48-well plate (bottom)
78
2 In the Sort Layout window, choose the type of device from the Device
menu. Default sorting devices are listed along with any defined custom devices. The Sort Layout window changes depending on the selected device: the number of rows and columns in the window matches the number of tubes, wells, or spots in the collection device.
5 Select the field(s) corresponding to the tube(s), well(s), or spot(s) where the
population will be sorted and choose a defined population from the Add menu. After you click in a sort location field, a menu appears where you can choose to add, delete, or clear all populations in the field (Figure 4-4 on page 80).
Chapter 4: Sorting
79
After you add a population, the population and the number of target events are written to the corresponding sort location field.
! Tip
Select a row or column header to select all fields in that row or column. After adding a population, it will be written to all selected fields at once.
For a four-tube layout, conflicts for the Far Left tube are sorted to the left; conflicts for the Far Right tube are sorted to the right. No other populations can be sorted into the center-most tubes.
80
Sortstarts sorting events for the current acquisition Tube. All counters reset to zero when this button is clicked. Events are sorted until the requested number of sorted events has been reached. Click the Sort button again to stop sorting before reaching the requested number of events; the counters stop at the number of sorted events. If you click Sort to restart sorting, the counters reset to zero.
Pausestops sorting, but not acquisition; sort counters freeze when the Pause button is clicked. Click the Pause button again to continue sorting and to continue incrementing the sort counters.
Chapter 4: Sorting
81
Using Counters
Counters provide ongoing status during sorting; the fields cannot be edited. To display fewer counters in the Sort Layout window, click the View Counters button and choose a menu option. The corresponding counter is hidden. (Only counters with a checkmark next to the name are displayed.)
NOTE
Counters display the following information: Sort Ratenumber of events/second that met the sort criteria and were sorted Conflict Countnumber of events that met the sort criteria but were not sorted because of conflicts Conflict Ratenumber of conflicts/second Efficiencynumber of sorted events/(sort conflicts + sorted events) x 100
82
Monitoring a Sort
During sorting, each sort location field displays the number of actual sorted events. When a target number is specified, the field displays the actual number of events along with the number of target events. A progress bar appears behind the Sort Rate counter field showing the progress of the sort.
Sort Report
Choose Sort > Sort Report to view a report of the current Sort Layout. This menu item is enabled only if a Sort Layout is open and the instrument is not sorting. The Sort Report can be printed or exported. NOTE After closing a Sort Layout, all counter information is lost. Thus, you should print a Sort Report immediately after sorting. A Sort Report contains the following: Header informationTube name, Sort Layout name, type of collection device, and the date and time of printing Sort settingsSort Setup values (information in the Breakoff and Streams tabs), precision mode, and masks definition Acquisition countersThreshold count, processed events count, electronic conflicts count, and elapsed time
Chapter 4: Sorting
83
Sort counterscounter values per sort destination, or total sort count if sorting sequentially Sort LayoutPopulation(s), sort count, and target event count for each sort location field
The Sort Report window contains a File menu where you can choose to print or export the report. Exported comma-separated values (CSV) files can be opened with a spreadsheet application such as Microsoft Excel.
84
Yield Mask
The Yield Mask setting defines how close to the edge of the drop, in 1/32-drop increments, a particle of interest can be located before sorting an additional drop. Half of each Yield Mask setting defines an equal area at each end of the drop. For example, when the Yield Mask is set to 16 and an event is within 8/32 from the beginning of a drop, the previous (leading) drop will be sorted. If an event is within 8/32 from the end of a drop, the following (trailing) drop will be sorted. See Figure 4-5.
trailing drop: sorted drop being interrogated leading drop: not sorted
Yield Mask
If the Yield Mask were set to 8 for the same target particle, the target particle would fall outside of the Yield Mask; thus no additional drops would be sorted. See Figure 4-6.
trailing drop: not sorted drop being interrogated leading drop: not sorted
Yield Mask
When the Yield Mask is set to zero, only one drop (the drop containing the target particle) will be deflected; when the mask is set to 32, two drops will always be deflected. Yield Masks between 032 will sort either one or two drops. When more than one drop is deflected in the same direction, residual charge from the first drop will degrade the quality of the side stream. Thus, when four-way sorting or sorting into small wells where precise deflection is required, a Yield Mask of zero is recommended.
Chapter 4: Sorting
85
NOTE Yield Masks cannot be used in conjunction with Phase Masks. Thus, when the Yield Mask is greater than zero, the Phase Mask automatically reverts to zero.
Purity Mask
The Purity Mask setting defines how close, in 1/32-drop increments, a contaminating drop can be located before ignoring the drop being interrogated. For example, when the Purity Mask is set to 16, the drop being interrogated will not be sorted if a non-target particle falls within the first or last 8/32 of the leading or trailing drop. In the following example, a non-target particle falls within the first 8/32, so the interrogated drop will not be sorted (Figure 4-7).
trailing drop drop being interrogated leading drop
Purity Mask
(not sorted)
Purity Mask
non-target particle
If the Purity Mask were set to 8 for the same target particle, the non-target particle would fall outside of the Purity Mask, so the interrogated drop would be sorted. See Figure 4-8. NOTE With any Purity mask greater than zero, the drop being interrogated must be free of contaminating particles or the drop will not be sorted.
trailing drop drop being interrogated leading drop
Purity Mask
(sorted)
Purity Mask
non-target particle
86
Phase Mask
Particles near the drop edge can affect the breakoff and alter the trajectory of the deflected drop. The Phase Mask restricts drop deflection when an event is too close to the edge of a drop or when there are events close to the edge of adjacent drops. A Phase Mask is used to improve counting accuracy and side-stream quality at the expense of yield. For example, when the Phase Mask is set to 16, the drop being interrogated will be sorted only if the target particle falls outside the Phase Mask (Figure 4-9).
trailing drop (drop sorted) leading drop
Phase Mask
Decreasing the Phase Mask to 8 allows more drops to be sorted. However, because the target particle is closer to the edge of the drop, there is more variability in drop trajectory (Figure 4-10).
trailing drop (drop sorted) leading drop
Phase Mask
! Tip
Chapter 4: Sorting
87
NOTE Phase Masks cannot be used in conjunction with Yield Masks. Thus, when the Phase Mask is greater than zero, the Yield Mask automatically reverts to zero.
Precision Mode: Yield Mask: Purity Mask: Phase Mask: Single Cell:
Purity 32 32 0
Yield 32 0 0
Single Cell 0 32 16
Initial 32 0 0
Fine Tune 0 0 0
"
"
"
"
In Purity mode, the Yield Mask is set to the maximum to obtain the greatest number of particles; because the Purity Mask is also at the maximum, only drops with a target particle will be sorted. Sorting in Purity mode results in a sorted sample that is highly pure, at the expense of recovery and yield. In Yield mode, only the Yield Mask is used at its maximum value; thus recovery and yield are optimized at the expense of purity.
88
In Single Cell mode, the Purity Mask is set to the maximum, so only drops containing a target particle will be sorted. The Phase mask is set at half the maximum, so only particles centered within the sorted drop are deflected. Drop trajectory and count accuracy are optimized at the expense of yield. This mode is recommended for single-cell sorting or situations where precise counting is required. NOTE Select the Single Cell checkbox to obtain the highest quality side streams and the most accurate counts. When the checkbox is selected, drops containing two target events (acceptable with a Purity Mask) are discarded. The Yield Mask is disabled when Single Cell is selected.
In Initial mode, only the Yield Mask is used at its maximum value; thus recovery and yield are optimized at the expense of purity. NOTE Initial mode is equivalent to the Yield mode. It is named differently as a reminder to use this as the initial mode when setting the drop delay using the AccuDrop option.
In Fine Tune mode, all Masks are set to zero for deflecting the maximum number of drops. This mode is used to fine-tune the drop-delay value using the AccuDrop option.
The current sort mode is duplicated and the Mask fields are enabled.
Chapter 4: Sorting
89
2 (Optional) Change the name of the mode in the Precision Mode field. 3 Enter values for the Yield, Purity, and Phase Masks. 4 Click to select the Single Cell checkbox, if needed. 5 Click Close.
The new mode is added to the Precision Mode drop-down menu. NOTE To delete a mode, choose it from the drop-down menu and then click Delete.
1 Install all sorting hardware, including the appropriate nozzle tip for the size
of the cells to be sorted. For specific sorting hardware, see Installing the Two-Way Sorting Hardware on page 93, Installing the BD QuadraSort Tube Holder on page 93, or Installing the Sorting Hardware on page 107. WARNING Sorting hardware could be contaminated with biohazardous material. Follow universal precautions when handling instrument hardware. As a general guideline, the nozzle tip should be six to ten times the particle diameter. Perform daily instrument optimization and quality control each time the nozzle tip is changed.
90
NOTE Because the drop-delay value cannot be <10, the BD FACSDiVa option cannot sort with nozzle tips >100 m. Use one of the following nozzle tips with the BD FACSDiVa option.
Nozzle Tip Size (m) 50 60 70 80 90 100a BD Catalog No. 343592 343588 343593 343589 343591 343594
! Tip
When sorting, perform sample optimization with the drop drive on and the frequency at an appropriate level.
Chapter 4: Sorting
91
1 Adjust the (drop drive) Frequency for the shortest droplet breakoff
distance.
2 Adjust the Amplitude to optimize the last connected drop. 3 Adjust the Phase to obtain single side streams. 4 Use the AccuDrop option to determine the drop delay.
92
3 Install the pegs at the appropriate positions in the support bracket; slide the
collection tube holder onto the pegs. If you are using the 12 x 75-mm tube holder, ports for cooling water face toward the front of the instrument. If you are using the 15-mL tube holder, ports for cooling water face toward the rear of the instrument.
Chapter 4: Sorting
93
NOTE When used with 15-mL tubes, the BD QuadraSort tube holder is designed to hold only polypropylene tubes (BD Catalog No. 352096), not polystyrene tubes. BD recommends placing the 15-mL tubes into the two center tube holders only, and using the outer tube holders for smaller tubes.
1 Verify that the Fluidics Control knob is set to Off and that the deflection
plates are not charged. WARNING To prevent shock, do not touch the deflection plates when the red warning light appears on the control panel.
3 Install the pegs into the fifth and eighth holes from the top of the support
bracket; slide the BD QuadraSort tube holder onto the pegs (Figure 4-12). NOTE Optimize the position of the tube holder and the angle of the tubes for your system.
94
standard pressure
high pressure
Figure 4-13 Preliminary Breakoff settings for standard (left) and high (right) pressure
2 Click the buttons to turn on the Drop Drive and Test Sort; for standard
pressure (1012 psi), turn on Attenuation as well.
standard pressure only
WARNING To prevent shock, do not touch the nozzle when the drop drive is on or drops are being charged. In digital mode, the drop drive is on when the Drop Drive button shows a drop pattern in the Sort Setup frame; drops are being charged when the Test Sort button shows a drop pattern or when the Sort button has been clicked.
Chapter 4: Sorting
95
When sorting with the 70-m nozzle tip at 35 psi, the Frequency should be in the range of 5577 kHz. For other sorting pressures, see the following table.
Sheath Pressure (psi) 5060 3050 2030 Frequency (kHz) 6599 5577 4055
Optimize these values for your sorting application. In general, a smaller nozzle opening requires a higher Frequency at a given sheath pressure. CAUTION The standard nozzle holder is not intended to be used at sheath pressures >20 psi. For higher sorting pressures, verify that the high-speed sort head (provided with the BD TurboSort Plus option) is installed.
4 Adjust the Amplitude to optimize the breakoff point (minimize stream noise).
! Tip
Breakoff values can be adjusted using the keyboard. Press the indicated Function key (ie, press F2 for Amplitude), and use the up and down arrow keys to adjust the value. Hold the Control key down while pressing the arrow keys to adjust the values in larger increments.
96
two-way sort
four-way sort
Figure 4-14 Preliminary Streams settings for a two-way (left) and four-way (right) sort
digital amplitude
VERTICAL
POSITION
HORIZONTAL
POSITION
TRIGGER
LEVEL
AUTOSET button
Source Type
CURSOR 1 MATH MENU CURSOR 2 CH2 MENU HORIZONTAL MENU HOLDOFF TRIGGER MENU
CH1 Freq
CH1 MENU
drop charge
VOLTS/DIV
VOLTS/DIV
SEC/DIV
SET LEVEL TO 50
FORCE TRIGGER
5V
2mV
5V
2mV
5s
5ns
TRIGGER VIEW
SEC/DIV knob
Chapter 4: Sorting
97
7 Turn the Plate Voltage knob on the instrument control panel counterclockwise to its minimum setting.
8 Turn on the deflection plates, using the push button on the instrument
control panel. WARNING To prevent shock, do not touch the deflection plates when the red warning light appears on the control panel. The plates remain energized even when the camera door is open.
9 Slowly turn up the plate voltage until the side streams are visible and
deflecting away from the waste aspirator. WARNING Do not allow the streams to touch the deflection plates because this could result in arcing (sparking). You might need to adjust the Frequency, Amplitude, Phase, or Streams settings to optimize the angle of the streams. Once optimized, the plate voltage can be increased.
10 Click the Streams tab and adjust the 2nd, 3rd, and 4th drop settings to
tighten the center stream and fine-tune the side streams.
! Tip
Generally, settings of 20, 10, and 5 are good starting values for the 2nd, 3rd, and 4th drops, respectively. If the 2nd drop must be set to zero to obtain a narrow center stream, the Frequency setting probably needs adjustment.
11 Adjust the Plate Voltage knob and Stream deflection percentages to direct
the streams into the tubes.
12 Press F12 to turn off Test Sort. 13 Calculate the drop delay.
See Calculating the Drop Delay on page 100.
98
NOTE After closing a Sort Layout, all counter information is lost. Thus, you should print a Sort Report immediately after sorting.
Chapter 4: Sorting
99
1 Create a new Experiment and rename it AccuDrop. 2 Rename the first Tube AccuDrop Beads. 3 With the AccuDrop Beads Tube selected in the Browser, click on the Instr.
Settings > Parameters tab in the Inspector and delete all parameters except FSC and SSC.
1 Move the emission filter away from the camera. 2 In the Sort Setup frame, click the button to turn on Test Sort ( 3 While viewing the streams on the AccuDrop monitor, adjust the
micrometer dial to obtain even illumination of the center and side streams. When illuminated evenly, the streams appear to sparkle (Figure 4-15 on page 101). ).
100
4 Turn off Test Sort. 5 Turn off the stream lamps to better view the streams. 6 Install a sample tube filled with a dilute suspension of AccuDrop beads
(1 drop of beads in 0.5 mL sheath fluid); click the Acquisition pointer to start acquisition.
7 Click the Parameters tab in the Inspector and adjust the FSC voltage to
place the bead population at channel 125,000.
8 Draw an Interval gate that encloses the entire histogram (Figure 4-16).
Set the endpoints of the interval at 0 and 262 x 103. CAUTION For an accurate setting, ensure that the histogram region encompasses the entire bead population, including doublets.
Chapter 4: Sorting
101
2 Adjust the micrometer dial to obtain the brightest bead spot on the center
stream.
3 Move the emission filter in front of the camera. 4 Adjust the Sample Differential knob to achieve a bead event rate of
approximately 4,000 events/second.
5 Click Sort in the Sort Layout window. 6 Optimize the drop delay.
In the Breakoff tab in the Sort Setup frame, adjust the Drop Delay setting until most of the beads are in the left stream and the least are in the center stream (Figure 4-17), using a 1-drop increment. Before you adjust the drop delay, the beads will appear as a bright spot on the center stream and a faint spot on the left side stream. Adjust the drop delay until the spot on the left stream is as bright as possible. This will yield the most accurate drop delay.
left stream left stream
center stream
center stream
Figure 4-17 Viewing beads on AccuDrop monitorbefore (left) and after (right) adjustment
102
7 Click Sort to stop sorting. 8 In the Sort Layout, change the Precision mode to Fine Tune.
10 Move the emission filter away from the camera; remove the tube from the
cytometer.
Chapter 4: Sorting
103
Sorting
Before beginning the sort, do the following: Perform sample optimization with the Drop Drive on and the Frequency at an appropriate level. Use gating tools and subsetting methods to define the population(s) of interest. Examples of gating analysis can be found in Recording and Analyzing Data on page 61 and in the BD FACSDiVa Software Users Guide.
104
Choose the collection device from the Device menu. Choose the Precision mode from the Precision menu. Enter the number of Target Events by choosing a value from the dropdown menu or entering a number in the field. Select the sort location field(s) to be sorted into. Select multiple fields by dragging the mouse; select a row or column by clicking the row or column header. Add the required population(s) to each sort location field. If the Add menu doesnt appear after selecting the sort location field(s), right-click the selected fields to see the menu. Enter the number of Target Events and the population(s) for the remaining sort location fields, if necessary.
NOTE To change the number of target events in a sort location field, change the value in the Target Events field before adding the population. Once a population has been added, you cannot change the number of sorted events except by deleting, and then adding the sort population. See Editing a Sort Layout on page 81.
Chapter 4: Sorting
105
2 Install the sample tube on the cytometer and close the camera door. 3 Turn the Fluidics Control knob to Run. 4 Verify that the green Acquisition pointer is indicating the appropriate Tube
in the Browser; click Sort.
106
2 Remove the center stream aspirator and replace it with the right and center
stream aspirator supplied with the BD CloneCyt Plus option.
3 Install the metal ground shield and the plastic tray shield. 4 Install the tray support on the support arm using the metal thumbscrew.
NOTE Install only the shorter tray support provided with the BD FACSDiVa option. Previous versions of the tray support do not allow sufficient clearance between the plate and the sort chamber.
tray support
Chapter 4: Sorting
107
2 Install the collection device on the tray support. 3 Choose Sort > Home Device.
The Device Setup dialog box appears (Figure 4-19 on page 109). If a Sort Layout is currently open, the corresponding collection device will be selected in the list of devices. Otherwise, select the appropriate collection device in the list.
108
4 Double-click the Test Sort button to deposit a drop on the Home location. 5 Carefully remove the collection device from the tray support and note
where the drop was deposited.
6 Wipe the collection device dry, and place it back on the tray support. 7 Adjust the Home location, if necessary.
Click the appropriate Arrow buttons to move the tray support as needed. Large arrows move the tray by 5 steps; small arrows move the tray by 1 step.
8 Repeat steps 4 through 7 until the drop is centered appropriately. 9 Click Set Home, and then Close. 10 Proceed with Sorting on page 104.
Chapter 4: Sorting
109
1 Choose Sort > Custom Devices. 2 Click the Add button in the Custom Devices dialog box.
A new device is added to the list of custom devices. By default, devices are named Custom Device_00x, where x is the next consecutively numbered device (Figure 4-20).
3 Select the text in the Name field and enter a new name. 4 Enter the number of sort location Rows and Columns.
A device can have up to 60 rows and 25 columns.
110
5 Use the Arrow keys to move to the Home location; click Set Home.
See Adjusting the Home Location on page 108. There are no default values for custom devices, so more initial adjustment with the Arrow keys is required.
6 Use the same procedure to move to the Farthest location; click Set Farthest.
The Farthest sort location is the well or spot on the lower-right corner of the collection device.
3 Click Delete.
The device is deleted from the custom device list, but is retained within any Sort Layouts where it was used.
Chapter 4: Sorting
111
112
5
DNA Analysis
The following topics are covered in this chapter: Criteria for DNA Experiments on page 114 CEN Optimization on page 115 CTN Resolution on page 122 Optimization for Data Recording on page 125
113
Obtaining good resolution for the DNA signal depends on proper sample preparation and instrument optimization of the optics and fluidics. The resolution of a flow cytometer can be assessed by measuring the CV of a reference particle: the lower the CV, the better the resolution. Linearity is critical for DNA experiments. To verify the linearity of DNA data, the pulse-area signal is used to measure the amount of DNA fluorescence detected from cells and nuclei. For example, the G2+M peak should be located at twice the mean channel of the G0/G1 peak (Figure 5-1).
Doublet discrimination, or the ability to resolve singlets from aggregates, is also important for DNA experiments. Aggregated cells or nuclei are detected as events that have two or more times the amount of singlet fluorescence. For cell-cycle analysis, it is important to resolve singlets from aggregates because doublets of G0/G1 cells have the same amount of DNA fluorescence as singlet G2+M cells. Therefore, these doublets accumulate in the same fluorescence area channel as singlet G2+M cells (Figure 5-2 on page 115).
114
NOTE
Before beginning this chapter, do the following. Prepare biological standards for instrument quality control using the BD DNA QC Particles kit (Catalog No. 349523). Prepare one tube each of chicken erythrocyte nuclei (CEN) and calf thymocyte nuclei (CTN) sample according to the kit instructions. The CEN sample is used for instrument optimization and to check instrument resolution (CV) and linearity. The CTN sample is used to verify the systems ability to resolve singlets from aggregates.
CEN Optimization
Use the following procedure to set up BD FACSDiVa software for a DNA experiment that uses propidium iodide (PI) as the DNA-staining dye (eg, the BD DNA QC Particles kit). If you are using another sample type, modify the steps accordingly.
115
! Tip
To place the Experiment inside an existing folder, select the folder before creating the Experiment.
3 Rename the new Specimen DNA QC Kit and rename the first Tube CEN.
This Tube will be used to optimize signals from the first laser. Your Experiment should look similar to that shown in the figure at the right.
4 Click the Parameters tab in the Inspector and make the following changes:
Delete all parameters except FSC, SSC, and PI. Select the Height and Width checkboxes for PI. Verify that the Log checkbox is deselected for all parameters.
116
5 Click the Threshold tab and change the threshold parameter to PI; verify
that the threshold value is set to 5,000.
7 Create a Statistics view and display the mean and CV for PI-A and PI-H.
Right-click the CEN Tube and choose Create Statistics View. Click the Edit Statistics button in the Inspector. On the Population tab, deselect # Events and %Parent. On the Statistics tab, select the Mean and CV for PI-A and PI-H (Figure 5-3 on page 118). Set Decimal Places to 1 for the CVs.
117
8 In the Acquisition Controls frame, set the Number to Record to 10,000 evt
and the Events to Display to 500 evt.
! Tip
Decreasing the number of displayed events will increase the data refresh rate.
Running CEN
1 Install the CEN sample tube on the cytometer; turn the Fluidics Control
knob to Run.
2 Verify that the green Acquisition pointer is in front of the CEN Tube in the
Browser; click once on the pointer to start acquisition. Events appear in the plots.
118
4 Adjust the FSC and SSC voltages to place the CEN on scale in the FSC vs
SSC dot plot.
6 While viewing the PI-A vs PI-H plot, adjust the Y control and the
Excitation Beam Focus wheel to maximize the signals.
119
8 Draw an Interval gate around the first two peaks on the PI-A histogram;
name the populations Singlets and Doublets (Figure 5-6).
CV <3%
9 Verify that the FL1/FL2 iris is open and check the CV of the Singlet
population. If the CV is 3%, continue with step 10. If the CV is >3%, skip to the following section; then return to step 10 to finish this section.
10 Click Record to save the data. 11 Check the linearity and print the worksheet.
Note the means of the Singlet and Doublet populations. Divide the mean of the Doublets by the mean of the Singlets. The Doublet/Singlet ratio should be 2.00 +/-0.05. If you cannot achieve a ratio between 1.952.05, contact BD Customer Support. Copy the means, CVs, and the calculated linearity result into the QC log.
120
1 Adjust the event rate to approximately 1,000 events/second. 2 Adjust the instrument optimization controls to obtain the lowest possible
CV for the Singlet population. Ensure that the Display is set to 100500 events. Close the FL1/FL2 iris. While viewing the PI-A vs PI-H plot, adjust the Y control, Excitation Beam Focus wheel, X control, Fluorescence focus control knob, Fluorescence channel height adjustment wheel, and FL1/FL2 beam splitter, as necessary.
NOTE PI-stained nuclei are used in this exercise, so the fluorescence signal will be generated by the first laser and collected in the FL2 channel. If you are setting up for an experiment using a different laser and detector channel, peak the signals from that laser using the appropriate rotators, translators, and beam splitters.
3 Decrease the event rate to approximately 200 events/second. 4 Open the FL1/FL2 iris and check the CV of the Singlet population.
If the CV is within an acceptable range, go to step 10 on page 120. If the CV is not within an acceptable range, repeat steps 1 through 4 in this section.
121
CTN Resolution
Singlets can be distinguished from aggregates based on size. With BD FACSDiVa software, aggregates can be resolved from singlets on an Area vs Height plot in conjunction with an Area vs Width plot. On the Area vs Height plot singlets can be distinguished from doublets by their height on the y-axis; singlets have slightly more height. On the Area vs Width plot, singlets are distinguished from doublets by the Width measurement; singlets have a smaller Width measurement. Discriminating the singlets from the aggregates enhances the accuracy of cellcycle analysis.
Running CTN
1 Install the CTN sample tube onto the cytometer. 2 Adjust the event rate to approximately 500 events/second. 3 Click the Next button in the Acquisition Controls frame;
change the name of the new Tube to CTN. The Next button duplicates the CEN Tube and Analysis objects. The new plots and Statistics view appear below the previous objects on the worksheet. Acquisition starts automatically and events appear in the plots. Notice the lack of resolution between the singlets and doublets (Figure 5-7).
doublets singlets
122
123
8 (Optional) Use the Zoom-In tool to magnify the area showing the singlets
and doublets on the PI-A vs PI-W plot. Magnify the area from the left of the singlets to the right of the doublets, including only the area of interest. The zoomed-in area should be long and narrow.
9 (Optional) Draw a gate around the singlet CTN. 10 Decrease the event rate to approximately 200 events/second. 11 Click Record to save the data; print the worksheet. 12 Remove the CTN Tube from the cytometer; put the instrument in Standby.
124
1 Install the sample tube and optimize the FSC and SSC signals. 2 Optimize the PI voltage to place the singlets close to channel 50 x 103. 3 Verify the doublet discrimination by zooming in on the PI-A vs PI-W plot.
If the singlets are not resolved, repeat step 6 on page 119. After optimizing the instrument settings, record data for each sample tube. Export data files for analysis in a third-party application such as ModFit LT.
125
126
6
Calcium Flux
The following topics are covered in this chapter: Intracellular Calcium Concentration on page 128 Calcium Flux Optimization on page 129 Measuring Calcium Flux on page 135
127
128
! Tip
To allow enough time for Ca++ flux response and resolution, enter a large value for the Events to record before recording events. You cannot enter a specific time in which to record events or assign a time resolution. Do not restart data recording during a calcium flux experiment.
129
! Tip
Add Violet to the UV1 Parameter name and Blue to the UV2 Parameter name to help keep track of UV parameters.
CAUTION For accurate data results, the instrument optics setup must match the current Instrument Configuration.
! Tip
To place the Experiment inside an existing folder, select the folder before creating the Experiment.
130
4 Click the Parameters tab in the Inspector and make the following changes:
Delete all parameters except FSC, SSC, UV1 Violet, and UV2 Blue. Verify that the Log checkbox is deselected for all parameters.
5 Click the Ratio tab and click the Add button; choose UV1Violet-A for the
Numerator and UV2 Blue-A for the Denominator.
131
8 Create a Statistics view and display the mean for the UV parameters and
the ratio. Right-click the Ca 1 Tube and choose Create Statistics View. Select the Statistics view, and click the Edit Statistics View button in the Inspector. Add the required statistics; delete #Events and %Parent from the Population tab.
NOTE Only the specified number of events is displayed in plots during acquisition and recording. After data recording is complete, all recorded events will be displayed.
132
2 Verify that the green Acquisition pointer is in front of the Ca 1 Tube in the
Browser; click once on the pointer to start acquisition. Events appear in the plots.
4 While viewing the FSC vs SSC plot (Figure 6-3), make the following
adjustments: Adjust the FSC and SSC voltages to place the sample on scale in the FSC vs SSC dot plot. Adjust the FSC threshold to remove debris without cutting into the population of interest. Draw a gate around the lymphocytes; use the Population Hierarchy view to rename the population Lymphocytes.
133
5 Format the remaining two dot plots to show the Lymphocyte population.
Select the two plots, right-click inside one of the plots and choose Show Populations > Lymphocytes.
134
1 Change the Events to Display to 50,000 events. 2 Verify that the unstimulated sample is still running; adjust the event rate to
approximately 200 events/second. The event rate is displayed in the Acquisition Status frame.
3 Click Record. 4 When approximately 10,000 events have been recorded, turn the Fluidics
Control knob to Standby and remove the unstimulated sample tube from the cytometer. CAUTION Do not stop recording or acquisition or the data display will reset to zero.
5 Add the stimulus to the tube and mix thoroughly. 6 Reinstall the tube on the cytometer; turn the Fluidics Control knob to Run.
After a few seconds, the Ca++ concentration begins to increase on the Time vs Ratio plot (Figure 6-4).
unstimulated sample
stimulus added
135
7 Click the Acquisition pointer when the cells are no longer reacting to the
stimulus.
8 Remove the tube from the cytometer. 9 Clean the fluidics system with 10% bleach for 5 minutes, and then with
deionized water for 5 minutes. NOTE Make sure to remove any remaining stimulus that would activate cells in subsequent samples.
10 To run another sample, install the next sample tube onto the cytometer. 11 Click the Next button; rename the new Tube as appropriate. 12 Repeat steps 3 through 10.
Analyzing Data
NOTE You can export the data for analysis in a third-party application, such as FlowJo.
136
2 Use the Statistics view to display the increase in the Ratio mean over time.
137
138
7
Troubleshooting
The tips in this section are provided to help you troubleshoot issues that might arise when using the BD FACSDiVa option. For instrument-specific troubleshooting, refer to the BD FACSVantage SE Users Guide; for softwarespecific troubleshooting, refer to the BD FACSDiVa Software Users Guide. If additional assistance is required, contact your local BD Biosciences technical support representative. See Technical Assistance on page ix. Troubleshooting suggestions in this chapter are grouped under the following headings: Electronics Troubleshooting on page 140 Acquisition Troubleshooting on page 142 Sorting Troubleshooting on page 149
139
Electronics Troubleshooting
Observation Instrument Disconnected in Instrument frame Possible Causes Power switched off on BD FACSDiVa module Communication failure between workstation and instrument Recommended Solutions Check the digital oscilloscope. If the screen is blank, switch on the Digital control switch. Quit the software and then restart it. If restarting does not work, reset the BD FACSDiVa electronics by switching off the power switch on the back of the BD FACSDiVa module, and then switching the power back on. Restart the computer. Unplug and then plug in the cable connectors and make sure they are secure. Enter the correct IP address. Call BD Biosciences for assistance. Wait two minutes. If the message remains, restart the computer. Decrease the event rate or verify the threshold. Delete Analysis objects, decrease the Display value, or delete parameters from the Instrument Settings Inspector.
Ethernet cable disconnected between workstation and instrument IP address changed Upgrading firmware in Instrument frame Master DAQ Overflow in Instrument frame Firmware loading incomplete Event rate too high Too many Analysis objects on worksheet or too many events displayed
140
NOTE If this occurs during sorting, turn off the deflection plates before resetting the electronics.
Chapter 7: Troubleshooting
141
Acquisition Troubleshooting
Observation No events in plots after clicking Acquire Possible Causes Acquisition pointer not set to current Tube Not in digital mode Viewing plots for a different Tube Incorrect population(s) in plot Recommended Solutions Click to move the Acquisition pointer in front of the appropriate Tube. Switch the Digital control switch to On. Double-click the current Tube in the Browser to display the plots for that Tube. Right-click the plot and choose Show Populations. Verify that the appropriate populations are displayed. Format the plot to display all events. Assign a color to the population displayed in the plot. Verify the population drawing order. Current Instrument Configuration different from optical bench Verify that the current Instrument Configuration corresponds to the optical bench setup. See Instrument Optimization and Quality Control on page 24. Add sample to tube or install new sample tube. Mix sample to suspend cells. Replace the sample tube. Set the threshold to the correct parameter for your application.
No sample in tube Sample not mixed properly Sample tube cracked Threshold not set to correct parameter (usually FSC)
142
Chapter 7: Troubleshooting
143
144
Sample not adequately mixed Sample too dilute Erratic event rate Sample aggregates Sample tube cracked Sample tube O-ring worn
Chapter 7: Troubleshooting
145
Dead cells or debris in sample Sample contaminated High CVs Instrument not aligned Event rate too high Poor sample preparation Old or contaminated quality control (QC) particles Window Extension too low
146
Window Extension set incorrectly Increasing threshold results in decreased Area signal Window Extension too small
CAUTION Increasing the Window Extension too much results in more electronic aborts or high CVs.
Chapter 7: Troubleshooting
147
148
Sorting Troubleshooting
Observation Function keys not responding for Sort Setup values Defined population not listed in Add menu Unusual pattern on digital oscilloscope while setting Phase Sort button disabled Possible Causes Sort Setup frame inactive Recommended Solutions Select the frame title bar to make it active and then press the required function key. Redefine the population using another gate type. Turn on Test Sort, and then push the AUTOSET button on the oscilloscope control panel. Click to move the Acquisition pointer in front of the appropriate Tube. Double-click the appropriate Tube in the Browser to view worksheet objects for that Tube. Decrease the event rate. Verify the gating hierarchy.
Acquisition pointer not set to current Tube Viewing Sort Layout for another Tube Event rate too high for Drop Drive frequency Sorting parent and child populations into two different tubes Purity Mask too high
Decrease the Purity Mask. Use the arrow buttons to choose the appropriate stream. See Using Sorting Controls on page 81. Decrease the Drop Drive amplitude. Decrease the event rate.
Chapter 7: Troubleshooting
149
150
This appendix also includes a blank configuration worksheet that can be photocopied and filled in for any custom configurations. See Configuration Worksheet on page 157.
151
Six-Color Configuration
712/21 610 SP 405/20 UV1 (FL4) 505 SP 485/22 SSC UV2 (FL5) APC (FL6) 660/20 PerCP-Cy5-5 (FL3)
152
712/12 610 SP 424/22 Hoechst 33258 (FL4) 780/60 APC-Cy7 (FL6) 710 LP 660/20 SSC APC (FL7) PerCP-Cy5-5 (FL3)
153
Seven-Color Configuration
712/21 610 SP 405/20 UV1 (FL4) 505 SP 485/22 SSC UV2 (FL5) APC (FL7) 780/60 APC-Cy7 (FL6) 710 LP 660/20 PerCP-Cy5-5 (FL3)
154
Eight-Color Configuration
FITC (FL1) 575/26 PE (FL2) 530/30 560 SP PE-Cy5 (FL8) 710/20 740 LP 750 LP or 780/60 PE-Cy7 (FL3) 780/60 APC-Cy7 (FL6) 710 LP 660/20 UV2 (FL5) SSC APC (FL7)
155
FITC (FL1) 575/26 PE (FL2) 530/30 560 SP PE-Tx Red (FL8) 610/20 640 LP 740 LP or 780/60 PE-Cy7 (FL3) 780/60 or 740 LP APC-Cy7 (FL6) 710 LP 660/20 Hoechst Red (FL5) APC (FL7)
* For more information, see Isolation and functional properties of murine hematopoietic stem cells that are replicating in vivo. Goodell MA, Brose K, Paradis G, et al. J Exp Med; 1996:183;1797-1806.
156
Configuration Worksheet
Configuration Name: ________________________________
FL ____ =
FL ____ =
FL ____ =
FL ____ =
FL ____ =
FL ____ =
FL ____ = SSC
FL ____ =
157
158
Index
A
aborts electronic 147 See also conflicts, sort. 80 AccuDrop determining drop delay 102 Experiment 100 monitor 15, 100 acquisition events to record 62 starting 31 troubleshooting 142 Acquisition Templates creating 61 previewing data 61, 67 adjusting area scaling 33, 41, 47, 123 Home location 71, 108 laser delay 40, 46 sort settings 72, 95 streams 31, 75 threshold 56 voltages 56 Window Extension 40, 46 alignment sample, preparing 24 Alpha control 30 amplitude about 74 optimizing 96 analog data (vs digital) 19 operation 17 oscilloscope 15, 32 analysis calcium flux data 136 data 61 DNA data 125 immunophenotyping 64 primary-laser results 35 reusing 67 saving 67 second-laser results 43 third-laser results 49 applications calcium flux 128 DNA 114 sorting 90 area scaling adjusting 41, 47 DNA experiment 123 primary laser 33 troubleshooting 143, 148 assistance, technical ix Attenuation 73
159
B
Backspace key, troubleshooting ix beads AccuDrop 100 alignment 24 sorting for drop delay 102 biohazardous hardware 90 waste 22 breakoff about 74 adjusting settings 95 adjusting with keyboard 72, 96 controls 74 optimal distance 96
C
calcium flux about 128 data analysis 136 Experiment 130 measuring 135 optimization 129 optimizing sample 133 Time parameter 129 calculating compensation 59, 60 drop delay 100 calf thymocyte nuclei (CTN) preparing 115 resolution 122 running 122 CellQuest Pro, viewing 18 chicken erythrocyte nuclei (CEN) Experiment 116 optimization 115 preparing 115 running 118 coefficient of variation (CV), high 146
compensation calculating 59, 60 gating data 59 Tubes, creating 55 components AccuDrop monitor 15 digital oscilloscope 14 FACSDiVa module 14 QuadraSort 17 computer starting up 23 workstation 18 configurations alternate eight-color 156 eight-color 155 five colors + DNA 153 instrument defaults 151 seven-color 154 six-color 152 variants 37 worksheet 157 conflicts, sort about 84 counting 82 printing 83 saving 80 troubleshooting 149 control switch about 15 analog operation 17 digital operation 15 controls Breakoff controls 74 compensation 55 inactive 16 instrument 16, 30, 70 single-stained 53 sort setup 71 sorting 70, 81 Streams controls 75
160
conventions viii counters, sort 82, 149 creating Acquisition Templates 61 Analysis objects 64 compensation Tubes 55 custom devices 71, 110 Sort Layouts 70, 79, 104 Sort Precision modes 70, 89 custom devices creating 71, 110 deleting 111 customer support ix cytometer See instrument.
DNA experiments, about 114 five colors + configuration 153 QC Particles kit 115 setting up Experiment 116 verifying linearity 114 doublet discrimination 114, 120, 124 drop charge 75 conflicts 84 correction factor 75 delay 74, 100 drive 73 sequence 74 starting values 98
D
data analyzing 61, 64 digital vs analog 19 gating 59, 64 recording 61, 63 delay drop 100 laser 40, 46 Delete key, troubleshooting ix deleting custom devices 111 sort populations 81 sort setup values 76 digital control switch 15 data (vs analog) 19 oscilloscope 14 digital mode active controls 70 inactive controls 16 operation 15 starting instrument in 22
E
editing Sort Layouts 81 eight-color configuration 155, 156 electronics aborts 147 about 12 control switch 15 troubleshooting 140 error messages instrument disconnected 140 instrument not responding 141 Master DAQ overflow 140 Unable to move stage 150 upgrading firmware 140 events not showing in plots 142, 144 rate, troubleshooting 144, 145 troubleshooting 147 Excitation Beam Focus wheel 30
Index
161
Experiments AccuDrop optimization 100 calcium flux 130 CEN optimization 116 immunophenotyping 61 instrument optimization 25 reusing 50 sample optimization 54 exporting Sort Reports 84
H
hardware biohazardous 90 plate sorting, installing 107 QuadraSort, installing 93 two-way sorting, installing 93 Home Device 71 Home location adjusting 71, 108 custom devices 110
F
FACSDiVa option about 12 components 13 workstation 18 FACStation 18 Far Left stream 75 Far Right stream 75 Fine Tune mode 89 fluorescence channel height adjustment wheel 30 focus control knob 30 signal, troubleshooting 143 frequency about 74 optimal ranges 96 function keys, troubleshooting 72, 149
I
immunophenotyping analysis 64 Experiment 61 Initial mode 89 Inspector, troubleshooting 148 installing nozzle tips 90, 92 plate-sorting hardware 107 QuadraSort hardware 93 two-way sorting hardware 93 instrument controls 16, 30, 70 default configurations 151 disconnect error 140 not responding 141 optimization 24 starting up 22 intracellular calcium concentration 128
G
gating compensation Tubes data 64 59
K
keyboard shortcuts 72 troubleshooting
162
L
labels parameter 62 lasers delay 40, 46 primary, optimizing signals 30, 32 QC results 35, 43, 49 second, optimizing signals 37, 39 third, optimizing signals 44, 45 Layout, Sort See Sort Layouts. Left stream 75 linearity, DNA experiments 114 low Area signal 143
O
obscuration bar 30 on/off switch about 15 analog operation 17 digital operation 15 optical configurations 151 optimization AccuDrop 100 CEN 115 CV of singlet population 121 for calcium samples 129, 133 for DNA samples 114, 125 for LWB samples 53 instrument 24, 25 primary-laser signal 30, 32 sample 52 second-laser signal 37, 39 third-laser signal 44, 45 oscilloscope analog 15, 32 digital 14 troubleshooting 149
M
main sorting adjustments 92 Masks about 84 default Precision Modes 88 Phase 87 Purity 86 Yield 85 Master DAQ overflow error 140 measuring calcium flux 135 modes defining 89 Sort Precision 84, 88 monitoring sorts 83, 106 monitors AccuDrop 15, 100 computer 18
P
parameters labels 62 scatter, distorted 146 Time 129 pausing sorting 81, 106 phase about 74 optimizing 97 Phase Masks about 87 Yield Masks, using with plates, sorting into 106
N
nozzle tips installing 90, 92 optimal frequency with 96 sizes and catalog numbers 91
86, 88
Index
163
plots excessive debris 146 no events in 142 unexpected events in 144 populations AccuDrop bead, defining 100 sorting 79, 91, 104 troubleshooting 147, 149 Precision Modes 84, 88 See also Sort Precision modes. primary laser optimizing signals 30, 32 results 35 verifying area scaling 33 printing Sort Reports 84, 99 Purity Masks 86 Purity mode 88
S
sample alignment 24 tubes, QuadraSort 93, 94 sample optimization about 52 Experiment 54 LWB example 53 samples, running 63 Save Conflicts 80 saving analyses 67 sort conflicts 80 sort setup values 76, 99 scaling, area DNA experiment 123 primary laser 33 second laser 41 third laser 47 troubleshooting 143, 148 scatter parameters, distorted 146 second laser adjusting delay 40 optimizing signals 37, 39 results 43 verifying area scaling 41 seven-color configuration 154 sheath filling tank 22 pressures, optimal frequency with 96 shortcuts, keyboard 72 signals low Area 143 no fluorescent 143 primary laser, optimizing 30, 32 second laser, optimizing 37, 39 third laser, optimizing 44, 45 troubleshooting 148 Single Cell mode 89 single-stained controls 53
Q
QuadraSort about 17 installing hardware 93 tube placement 94 quality control 24
R
recording calcium flux data 135 compensation Tubes 58, 59 data 61, 63 DNA data 125 primary-laser results 35 second-laser results 43 third-laser results 49 reports, sort 83 results, troubleshooting 143, 150 reusing analyses 67 Right stream 75
164
singlet population defining 120 optimizing CV 121 six-color configuration 152 slides, sorting into 106 software about 12 starting up 23 Sort Layouts about 77 creating 70, 79, 104 custom 71, 110 editing 81 entering populations 79 examples 77 Sort Precision modes about 84 creating 70, 89 defaults 88 Fine Tune 89 Initial 89 Purity 88 Single Cell 89 Yield 88 sort rate, troubleshooting 149 Sort Reports displaying 71, 83 exporting 84 printing 84, 99 Sort Setup controls about 71 Breakoff controls 74 button functions 73 Streams controls 75 sorting about 69, 90 adjusting settings 72, 95 beads for drop delay 102 collection devices 77 conflicts 80, 82, 84
controls 70, 81 counters 82 into plates 106 into slides 106 into tubes 92 main adjustments 92 monitoring 83, 106 pausing 81, 106 populations 79, 91, 104 report 83 saving settings 76, 99 setting up for 90 Sort button 81 starting 81, 104 stopping 81 test mode 73 troubleshooting 149 starting acquisition 31 computers 23 instrument 22 sample flow 31 software 23 sorting 81, 104 stopping sorting 81, 104 streams adjusting 31, 70, 75 Streams tab 75 troubleshooting 150 turning on 73
T
tanks, servicing 22 Target Events 79 technical assistance ix Test Sort 73 test tubes, sorting into 92 Theta control 30
Index
165
third laser adjusting delay 46 optimizing signals 44, 45 results 49 verifying area scaling 47 threshold adjusting 56 troubleshooting 147 Time parameter 129 troubleshooting acquisition 142 CVs 146 electronic aborts 147 electronics 140 event rate 144, 145 Inspector 148 keyboard keys ix, 72, 149 low Area signal 143 oscilloscope 149 plots 142, 144, 146 populations 147, 149 results 150 scatter parameters 146 signals 143, 148 sort conflict rate 149 sort counters 149 sort rate 149 sorting 149 streams 150 tubes 150 Window Extension 147 Tubes compensation 55 tubes installing two-tube holder 93 sorting into 92 troubleshooting 150 used with QuadraSort 93, 94
U
upgrading firmware error 140
V
viewing CellQuest Pro voltages, PMT adjusting 56 18
W
waste tank, emptying 22 Window Extension adjusting 40, 46 troubleshooting 147 worksheet, configuration 157 workspace, view options 18 workstation, about 18
X
X control 30
Y
Y control 30 Yield Masks about 85 Phase Masks, using with 86, 88 Yield mode 88
Z
Z control 30 Z distance 31
166