Electrophysiological Recording Techniques: Goal of Physiological Recording
Electrophysiological Recording Techniques: Goal of Physiological Recording
Electrophysiological Recording Techniques: Goal of Physiological Recording
Goal of Physiological Recording To detect the communication signals between neurons in real time (s to hours) Current clamp measure membrane potential, PSPs, action potentials, resting membrane potential Voltage clamp measure membrane current, PSCs, voltage-ligand activated conductances
Example of an electrical circuit with various parts. Current always flows in a complete circuit.
Ohm's Law
For electrophysiology, perhaps the most important law of electricity is Ohm's law. The potential difference between two points linked by a current path with a conductance G and a current I is:
Instruments used to measure potentials must have a very high input resistance Rin.
capacitance
B Now, if we apply a pulse of current to the circuit, the current first charges up the capacitance, then changes the voltage
Response of an RC parallel circuit to a step of current
The voltage V(t) approaches steady state along an exponential time course:
The steady-state value Vinf (also called the infinite-time or equilibrium value) does not depend on the capacitance; it is simply determined by the current I and the membrane resistance R:
This is just Ohm's law, of course; but when the membrane capacitance is in the circuit, the voltage is not reached immediately. Instead, it is approached with the time constant t, given by
Thus, the charging time constant increases when either the membrane capacitance or the resistance increases. Consequently, large cells, such as Xenopus oocytes that are frequently used for expression of genes encoding ion-channel proteins, have a long charging phase.
Electrodes
In circuits, we use wires In biology, nature uses liquids Electrodes are used to transform current flow from electrons to ions
Cl-
Ag
AgCl coating
Electrode reaction: Ag + ClAgCl + electron (e-) The silver/silver chloride electrode is reversible but exhaustible
Platinum Electrode
Getting a recording
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. Find a health cell Fill a pipette, place in holder and apply positive pressure Put pipette in the bath Get the test pulse running and make sure it works Zero the offset Bridget balance the pipette Position the pipette above the cell Verify positive pressure and advance into the slice Push pipette tip onto cell surface, a small dimple would appear, and then release pressure Apply slight suction and a negative holding potential (-60 mV) If everything is right, you will get a gigaohm seal
Intracellular solutions
Standard intracellular (patch) solution: 135 K-Gluconate or K-MeSO4, 10 HEPES, 7 NaCl, 2 Na2-ATP, 2 MgCl2 (pH 7.2 with KOH) Osmolarity - ~10% lower than extracellular (higher is better for lower Rs) Could also add/replace: - GTP (0.3 mM) - Phosphocreatine (10 mM) - Fluorescence dyes or biocytin - Cs instead of Gluconate/MeSO4 to block K+ currents - High Cl to increase inhibitory responses at resting membrane potentials (eg. replace K-Gluconate with KCl)
10
See http://www.olympusmicro.com/primer/techniques/dic/dicconfiguration.html
Steps to gigaseal
11
Surface cleaning is critical for successful recording Use of a cleaning pipette to remove tissue overlying a cell in tissue slices.
12
Record Voltage
Positive potential - a positive voltage at the headstage input with respect to system ground Transmembrane potential, Vm - Vinside relative to Voutside Depolarizing potential - is a positive shift in Vm
13
Record Current
Positive current - flows out of amplifier into the electrode - flows out of the pipette tip into the cell Inward current - flows from the outside surface of membrane to the inside surface Positive and Negative currents and voltages are always based on the headstages perspective
14
15
In a patch recording, currents through the seal also flow through the measuring circuit, increasing the noise on the measured current.
16
Technical issues
1. Loosing the seal on break-in (leaky cell; depolarized membrane potential, such as negative (downward) shift, jump in holding current, Series resistance (VC/CC, filtering, voltage-drop/error, compensation), Noise (60 Hz line frequency, grounding, high-frequency noise single channel), Wash-in/wash-out (perforated patch recording: Amphotericin or Gramicidin), Offsets (junction potentials, Ag/AgCl electrodes), Recording issues (amplifier gain, Analog/Digital boards, saturation, sample rates), Space clamp.
2. 3. 4. 5. 6. 7.
Expensive microscope & camera required, better manipulators (?) Best for cells near the surface Typically lower series resistance Can record from multiple cells Can record from multiple locations on the same cell Can identify cells before recording (morphology, fluorescence)
Only need dissecting microscope & no camera Can patch deep cells (also in vivo) Typically higher series resistance Difficult to record from multiple cells
Impossible (?) to record from multiple locations on the same cell Identification only possible after experiment (but can use electrical cues)
17
Data analysis
Phases of the Action Potential
Firing threshold is the point at which the number of activated Na+ channels > inactivated Na+ channels
18
19