HP Agilent 8590 Series Service Manual
HP Agilent 8590 Series Service Manual
HP Agilent 8590 Series Service Manual
Manufacturing Part Number: 08590-90316 Supersedes: 08590-90300 Printed in USA April 2001 Copyright 1992-1998, 2001 Agilent Technologies, Inc.
The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and tness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Certication
Agilent Technologies certies that this product met its published specications at the time of shipment from the factory. Agilent Technologies further certies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
Regulatory Information
The specications and characteristics chapter in this manual contain regulatory information.
Warranty
This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country. Agilent Technologies warrants that its software and rmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or rmware will be uninterrupted or error-free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modication or misuse, operation outside of the environmental specications for the product, or improper site preparation or maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedies
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Ofce.
Documentation Options
Option 041 or 043: Programmer's Guide 8590 E-Series and L-Series Spectrum Analyzers, and 8591C Cable TV Analyzer Programmer's Guide Describes analyzer operation via a remote controller (computer) for Parallel/RS-232 or Parallel/GPIB interface. Option 910: Additional User's Documentation Provides an additional copy of the user's guides, the calibration guide, and the quick reference guide. Option 915: Service Guide and Component-Level Information Describes troubleshooting and repair of the analyzer. Option 915 consists of two manuals: 8590 Series Analyzers Assembly-Level Repair Service Guide Describes adjustment and assembly level repair of the analyzer. 8590 Series Analyzers Component-Level Repair Service Guide Provides information for component-level repair of the analyzer.
This represents a key physically located on the instrument front panel. This indicates a softkey, a key whose label is determined by the instrument's rmware. Softkeys are located on the display bezel. This indicates text displayed on the instrument screen.
Screen Text
Contents
1. Introduction The 8590 Series Analyzers Assembly-Level Repair Service Guide . . . . . . . . . . . . . . . . . . . . . Service Guide Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Analyzer Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Firmware Revision Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Firmware Upgrade Kit Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Reliability Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjustment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Test Equipment You Will Need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . When A 3335A Source Is Not Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . If There Are Abnormal Indications During Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Periodically Verifying Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard-Value Replacement Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . If You Replace or Repair an Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14 15 16 18 18 19 21 23 23 23 24 24 24 25
2. Making Adjustments Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 1a. Display (with four access holes in top of shield) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 1b. Display (with eight access holes in top of shield) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 2. Sampler Match . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 3. 10 MHz Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 4. 10 MHz Precision Frequency Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 5. Crystal and LC Bandwidth Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 6. Cal Attenuator Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 7. Log and Linear Amplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 8. CAL FREQ Adjustment Routine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 9. CAL AMPTD Adjustment Routine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 10. CAL YTF Adjustment Routine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 11. CAL MXR Adjustment Routine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 12. Second Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 13. Third Converter and Second IF Bandpass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 14. Comb Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 15. Frequency Response of the 8590L, 8591C, and 8591E . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q . . . . . . . . . . . 106 17. Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 18. Modulator Offset and Gain for Option 010 or 011 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 19. Entering External ALC Correction Constants for Option 010 or 011 . . . . . . . . . . . . . . 129 20. Checking the External ALC for Option 010 or 011 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131 21. Correcting for External ALC Error for Option 010 or 011 . . . . . . . . . . . . . . . . . . . . . . . 135 22. First LO Distribution Amplier for Option 009 or 010 . . . . . . . . . . . . . . . . . . . . . . . . . . 139 23. BITG Power Level for Option 010 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142 24. Tracking Oscillator for Option 010 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145 25. Checking the Absolute Amplitude Accuracy for Option 050 . . . . . . . . . . . . . . . . . . . . . . 151 26. Correcting for Absolute Amplitude Accuracy for Option 050 . . . . . . . . . . . . . . . . . . . . . 157 27. Checking the Absolute Amplitude Accuracy for Option 051 . . . . . . . . . . . . . . . . . . . . . . 160 28. Correcting for Absolute Amplitude Accuracy for Option 051 . . . . . . . . . . . . . . . . . . . . . 171
Contents
2a. Making Adjustments: If 3335A Source Not Available Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .178 6a. Cal Attenuator Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .179 7a. Log and Linear Amplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .186 15a. Frequency Response of the 8590L, 8591C, and 8591E . . . . . . . . . . . . . . . . . . . . . . . . . .191 3. Backing Up and Reloading Correction Constants Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200 Backing Up Analyzer Correction Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .201 Analyzer Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .204 Reloading the Correction Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .207 Instrument Calibration after Reloading the Correction Constants . . . . . . . . . . . . . . . . . . . .209 4. Troubleshooting the Analyzer Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .216 Problems at Instrument Power-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .219 Troubleshooting an Inoperative Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .221 Troubleshooting the A2 Display Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .227 Isolating an RF, LO, IF, or Video Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .232 Using the Internal Service-Diagnostic Routines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .238 Troubleshooting the A16 Processor/Video Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . .249 5. Troubleshooting the IF Section Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .254 Service Equipment You Will Need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .254 After an Analyzer Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .255 IF Section Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .255 IF Power-Level Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .256 Troubleshooting the A15 Motherboard Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .266 6. Troubleshooting the RF Section Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .282 Service Equipment You Will Need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .283 After an Analyzer Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .283 6a. 8590L, 8591C, and 8591E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284 6b. 8592L/94L, 8594Q, and 8593E/94E/95E/96E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .287 7. Replacing Major Assemblies Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .294 Service Equipment You Will Need . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .294 After an Analyzer Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .295 Removal and Replacement Procedures in this Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . .295 Instrument Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .296 A1 Front-Frame Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .298 A1A1 Keyboard/Front-Panel Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .302 A2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303 RF Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308 A3 Front-End Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311
10
Contents
A7 Analog Interface Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A8 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A15 Motherboard and IF Extrusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A16 Processor/Video Board and A16A1 Memory Board Assembly . . . . . . . . . . . . . . . . . . . . A16 Processor/Video Board Firmware ROMs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A16A1BT1 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A17 Memory Card Reader Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A25 Counter-Lock Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B1 Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Rear-Frame Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8. If You Have A Problem Contacting Agilent Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Check the Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calling Agilent Technologies Sales and Service Ofces . . . . . . . . . . . . . . . . . . . . . . . . . . . . Returning Your Analyzer for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9. Assembly Descriptions and Block Diagrams Assembly Descriptions and Block Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RF and LO Section for 8590L, 8591E, and 8591C Spectrum Analyzers . . . . . . . . . . . . . . . RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IF Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10. Analyzer Options and Block Diagrams Analyzer Options and Block Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Option Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Troubleshooting Analyzer Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010 . . . . . . . . . . . . . . . . . . . . . . First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Option 190 DVB-C Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Option 190 DVB-C Service Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
318 322 324 330 335 337 339 342 344 346
11. Major Assembly and Cable Locations Major Assembly and Cable Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 454 12. Replaceable Parts Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard-Value Replacement Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Component-Level Information Packets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11
Contents
14. Analyzer Messages Interpreting Analyzer Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .614 15. Service Equipment and Tools Service Equipment and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .660 16. Safety and Regulatory Information Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .668 Safety Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .668 Instrument Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .669 General Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .670 Protection from Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .672
12
Introduction
The 8590 Series Analyzers Assembly-Level Repair Service Guide provides the information needed to adjust and repair the 8590 E-Series and L-Series spectrum analyzers, the 8591C cable TV analyzer and 8594Q QAM analyzer to the assembly level.
13
14
Chapter 1
Chapter 1
15
Chapter 10, Analyzer Options and Block Diagrams, contains information on the available analyzer options. Chapter 11, Major Assembly and Cable Locations, contains gures identifying all major assemblies and cables. Chapter 12, Replaceable Parts, contains information needed to order assemblies and component-level information for the analyzer. Chapter 13, Softkey Descriptions, contains a description of the calibration, service, and diagnostic softkeys. Chapter 14, Analyzer Messages, contains a description of the analyzer messages that are displayed when there is a problem with the analyzer. Chapter 15, Service Equipment and Tools, contains information about equipment and tools used to service the analyzer.
Analyzer Description
The 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers provide measurement capabilities over the RF and Microwave frequency ranges. The frequency ranges of 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers are described below. All 8590 E-Series spectrum analyzers, 8591C cable TV analyzers and 8594Q QAM analyzers are equipped with a memory card reader. The memory card reader allows the analyzer to use downloadable software programs. The functional personality of the analyzer can also be altered using the memory card reader. This feature provides the user with one analyzer that can perform the function of other types of microwave test equipment. Further information about the 8590 E-Series and L-Series spectrum analyzers, and the available options, is provided in Chapter 10, Analyzer Options and Block Diagrams, of this service guide and the 8590 E-Series and L-Series Spectrum Analyzers User's Guide. Further information about the 8591C cable TV analyzer and 8594Q QAM analyzer is provided in the spectrum analyzer reference for your instrument. Further information about the 8594Q QAM analyzer is provided in 8594Q QAM Analyzer Spectrum Analyzer Reference User's Guide.
16
Chapter 1
Table 1-1
8590 E-Series and L-Series Spectrum Analyzer, 8591C Cable TV Analyzer and 8594Q QAM Analyzer Frequency Ranges
Analyzer Model 8590L 8591E 8591C 8592L 8592L (Option 026) 8593E 8593E (Option 026) 8594E 8594L 8594Q 8595E 8596E Frequency Range 9 kHz to 1.8 GHz 9 kHz to 1.8 GHz 1 MHz to 1.8 GHz 9 kHz to 22.0 GHz 9 kHz to 26.5 GHz 9 kHz to 22.0 GHz 9 kHz to 26.5 GHz 9 kHz to 2.9 GHz 9 kHz to 2.9 GHz 9 kHz to 2.9 GHz 9 kHz to 6.5 GHz 9 kHz to 12.8 GHz
Chapter 1
17
18
Chapter 1
Reliability Considerations
Analyzer Input Protection
The analyzer input circuitry can be damaged by power levels that exceed the maximum safe input-level specications. Table 1-2 through Table 1-5 provides the input specications. To prevent input damage, these specied levels for your analyzer must not be exceeded. The analyzer input can also be damaged by large transients. If it is likely that your analyzer will be exposed to potentially damaging transients, take whatever precautions are necessary to protect its input circuitry. The analyzer input can easily be protected by disconnecting it from the signal source whenever it is likely that large transients will be present. When it is impractical to disconnect the analyzer, a transient-limiting device should be used. The 11947A Transient Limiter is an analyzer accessory that protects the input circuitry from transients and accidental overloads. Contact your local Agilent Technologies sales or service ofce for more information about the 11947A.
CAUTION
Transients are often produced during electromagnetic interference (EMI) conducted emissions testing. One type of device, the line impedance stabilization network (LISN), can produce large transients when its switch position or voltage input is changed.
Chapter 1
19
Table 1-2
8590L and 8591E Spectrum Analyzer and 8591C Cable TV Analyzer &Maximum Safe Input Level
INPUT 50 (Input Attenuation 10 dB) INPUT 75 (Input Attenuation 10 dB) +75 dBmV (0.4 W) +75 dBmV (0.4 W) 100 Vdc
Table 1-3
+30 dBm (1 W, 7.1 V rms), input attenuation 10 dB +50 dBm (100 W) for <10 s pulse width and <1% duty cycle, ssfs input attenuation 30 dB 0 Vdc
dc
Table 1-4
8594E and 8594L Spectrum Analyzer and 8594Q Analyzer Maximum Safe Input Level
INPUT 50 (Input Attenuation 10 dB)
+30 dBm (1 W, 7.1 V rms) +50 dBm (100 W) for <10 s pulse width and <1% duty cycle, input attenuation 30 dB 0 V (dc coupled) 50 V (ac coupled)
Table 1-5
+30 dBm (1 W, 7.1 V rms) +50 dBm (100 W) for <10 s pulse width and <1% duty cycle, input attenuation 30 dB 0 V (dc coupled) 50 V (ac coupled)
20
Chapter 1
Adjustment Procedures
The following table lists the analyzer adjustments in Chapter 2 and Chapter 2a and the models that they apply to. Select the model of the analyzer being adjusted and note the adjustments marked in the analyzer model column to determine which adjustments should be performed. Table 1-6 8590 E-Series and L-Series Spectrum Analyzer, 8591C Cable TV and 8594Q QAM Analyzer Adjustment Procedures
Adjustment for Spectrum Analyzer Model: 8594Q 8591C 8591E 8594E 8593E 8596E 8595E 21 8590L 8592L 8594L Adjustment Procedure Name
Display Sampler Match 10 MHz Reference 10 MHz Precision Frequency Reference Crystal and LC Bandwidth Filter Cal Attenuator Errora Cal Attenuator Errorb Log and Linear Ampliera Log and Linear Amplierb CAL FREQ Adjustment Routine CAL AMPTD Adjustment Routine CAL YTF Adjustment Routine CAL MXR Adjustment Routine Second Converter Third Converter and Second IF Bandpass Comb Generator Frequency Response of the 8590L, 8591C, and 8591Ea
Chapter 1
Table 1-6
8590 E-Series and L-Series Spectrum Analyzer, 8591C Cable TV and 8594Q QAM Analyzer Adjustment Procedures
Adjustment for Spectrum Analyzer Model: 8594Q ` 8591C 8591E 8594E 8593E 8596E 8595E 8590L 8592L 8594L
15a.
Frequency Response of the 8590L, 8591C, and 8591Eb Frequency Response of the 8592L, 8593E, 8594E, 8594L, 8594Q,8595E, 8596E Time and Date Modulator Offset and Gain for Option 010 or 011 Entering External ALC Correction Constants for Option 010 or 011 Checking the External ALC for Option 010 or 011 Correcting the External ALC Error for Option 010 or 011 First LO Distribution Amplier for Option 009 or 010 BITG Power Level for Option 010 Tracking Oscillator for Option 010 Checking the Absolute Amplitude Accuracy for Option 050 Correcting for Absolute Amplitude Accuracy for Option 050 Checking the Absolute Amplitude Accuracy for Option 051 Correcting for Absolute Amplitude Accuracy for Option 051
16.
17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28.
a. If a 3335A source is not available, use the alternative adjustment with the same number found in Chapter 2a , Making Adjustments: If 3335A Source Not Available. b. If a 3335A source is not available, substitute this adjustment for the one with the same number found in Chapter 2 , Making Adjustments.
22
Chapter 1
Introduction Safety
Safety
Familiarize yourself with the safety symbols marked on the analyzer, and read the general safety instructions and the symbol denitions given in Chapter 16 , Safety and Regulatory Information, before you begin the procedures in this chapter.
Chapter 1
23
24
Chapter 1
Table 1-7
Frequency Response Residual Responses Frequency Response Residual Responses Comb Generator Frequency Accuracy Frequency Response Residual Responses Frequency Response Residual Responses Frequency Response Residual Responses Noise Sidebands System Related Sidebands Frequency Response Other Input Related Spurious Responses Spurious Response Gain Compression Displayed Average Noise Residual Responses Frequency Response Residual Responses
A3A5 Input Attenuator CAL FREQ CAL AMPTD CAL YTF Frequency Response
Chapter 1
25
Table 1-7
Noise Sidebands System Related Sidebands Frequency Response Other Input Related Spurious Responses Spurious Response Gain Compression Displayed Average Noise Residual Responses Frequency Readout & Marker Count Accy Noise Sidebands System Related Sidebands Residual FM Frequency Span Readout Accy Frequency Response Other Input Related Spurious Responses Spurious Response Residual Responses Noise Sidebands System Related Sidebands Frequency Response Other Input Related Spurious Responses Spurious Response Displayed Average Noise Residual Responses None
First LO Distribution Amplier for Option 009 or 010 CAL FREQ CAL AMPTD CAL YTF
A3A8 YIG-Tuned Filter CAL FREQ or A3A8 Switched CAL AMPTD YIG-Tuned Filter CAL YTF CAL MXR Frequency Response A3A9 Bandpass Filter CAL FREQ CAL AMPTD A3A10 Directional Coupler CAL FREQ CAL AMPTD CAL YTF
System Related Sidebands Frequency Response Other Input Related Spurious Responses Spurious Responses Residual Responses Comb Generator Frequency Accuracy None System Related Sidebands Frequency Response Other Input Related Spurious Responses Spurious Responses Residual Responses
A3A12 3 dB Attenuator CAL FREQ CAL AMPTD A3A13 Isolator CAL FREQ CAL AMPTD CAL YTF
26
Chapter 1
Table 1-7
A3A14 LO Distribution First LO Amplier (LODA) Distribution Amplier for Option 009 or 010 only CAL FREQ CAL AMPTD CAL YTF A3A15 Tracking Generator First LO Distribution Amplier for Option 009 or 010 CAL FREQ CAL AMPTD CAL YTF BITG Power Level for Option 010 A3 Input Attenuator A4 First Converter CAL FREQ CAL AMPTD Frequency Response CAL FREQ CAL AMPTD
Frequency Readout and Marker Count Accy Noise Sidebands System Related Sidebands Residual FM Frequency Span Readout Accuracy Frequency Response Other Input Related Spurious Responses Spurious Response Residual Responses Frequency Readout and Marker Count Accy Noise Sidebands System Related Sidebands Residual FM Frequency Span Readout Accuracy Frequency Response Other Input Related Spurious Responses Spurious Response Residual Responses Frequency Response Noise Sidebands System Related Sidebands Frequency Response Spurious Response Gain Compression Displayed Average Noise Residual Responses Noise Sidebands Residual FM System Related Sidebands Frequency Response Spurious Response Gain Compression Displayed Average Noise Residual Responses Noise Sidebands Residual FM System Related Sidebands Frequency Span Readout Accuracy Frequency Response Spurious Response Displayed Average Noise Residual Responses
A5 Second Converter
A6 YIG-Tuned Oscillator
Chapter 1
27
Table 1-7
Frequency Readout and Marker Count Accy Noise Sidebands Residual FM System Related Sidebands Frequency Span Readout Accuracy Sweeptime Accuracy Scale Fidelity Reference Level Accuracy Absolute Amplitude Calibration and Resolution Bandwidth Switching Uncertainties Frequency Response Absolute Amplitude, Vernier, and Power Sweep Accuracy Tracking Generator Level Flatness Harmonic Spurious Outputs Non-Harmonic Spurious Outputs Tracking Generator Feedthrough
Modulator Offset and Gain for Option 010 or 011 Entering External ALC Correction Constants for Option 010 or 011 Checking the External ALC for Option 010 or 011 Correcting the External ALC Error for Option 010 or 011 None
A8 Power Supply
System Related Sidebands Spurious Responses Residual Responses Noise Sidebands System Related Sidebands Absolute Amplitude Calibration and Resolution Bandwidth Switching Uncertainties Other Input Related Spurious Spurious Responses Displayed Average Noise Residual Responses Frequency Response Absolute Amplitude, Vernier and Power Sweep Accuracy Tracking Generator Level Flatness
A9 Third Converter
Third Converter &Second IF Bandpass CAL FREQ CAL AMPTD BITG Power Level for Option 010 only
A10 LODA Control (Opt. 009) or Tracking Generator Control (Opt. 010)
First LO Distribution Amplier for Option 009 or 010 only CAL FREQ CAL AMPTD CAL YTF
28
Chapter 1
Table 1-7
Resolution Bandwidth Accuracy Absolute Amplitude Calibration and Bandwidth Switching Uncertainties Scale Fidelity Displayed Average Noise Resolution Bandwidth Accuracy Absolute Amplitude Calibration and Bandwidth Switching Uncertainties Scale Fidelity Reference Level Accuracy
A12 Amplitude Control CAL FREQ CAL AMPTD A13 Bandwidth Crystal and IC Bandwidth Filter CAL FREQ CAL AMPTD Log and Linear Amplier CAL FREQ CAL AMPTD CAL FREQ CAL AMPTD 10 MHz Reference CAL FREQ CAL AMPTD CAL YTF (8592L, 8593E, 8595E, 8596E only) CAL MXR (8592L, 8593E, 8595E, 8596E only) Cal Attenuator Error Frequency Response Time and Date
These adjustments are not necessary if valid correction constants are recovered from the defective A16 processor/video assembly. See the A16 processor/video board assembly replacement procedure in this service guide.
Chapter 1
29
30
Chapter 1
Making Adjustments
The procedures in this chapter adjust the analyzer electrical performance to the specications described in Chapter 2 of the calibration guide for your instrument. Most adjustments require access to the interior of the analyzer. If a 3335A source is not available, use the alternative adjustments with the same number found in Chapter 2a.
31
32
Chapter 2
Making Adjustments 1a. Display (with four access holes in top of shield)
Procedure
Display Position 1. Press the following analyzer keys to adjust the horizontal position.
PRESET CAL, More 1 of 4 CRT HORZ POSITION
2. Rotate the knob until the display is centered horizontally. 3. Press the following analyzer softkey to adjust the vertical position.
CRT VERT POSITION
4. Rotate the knob until the display is centered vertically. 5. Press the following analyzer keys to store the horizontal and vertical position values into nonvolatile memory.
CAL CAL STORE
Chapter 2
33
Making Adjustments 1a. Display (with four access holes in top of shield)
Fine Focus 1. Adjust the front panel INTENSITY control for a comfortable viewing intensity. 2. Use an adjustment tool to access the ne focus adjustment. See Figure 2-1. Adjust as necessary for a focused display. If one of the end-stops of the ne focus adjustment is reached, proceed with the Coarse Focus Adjustment section. Figure 2-1 Fine Focus Adjustment Location
34
Chapter 2
Making Adjustments 1a. Display (with four access holes in top of shield)
Coarse Focus The Coarse Focus adjustment point is located at the rear of the display assembly, therefore, it is required to place the display in a service position to perform this procedure. 1. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. 2. Place the display in a service position.
WARNING
The analyzer contains potentially hazardous voltages. Refer to the safety symbols provided on the analyzer, and in the general safety instructions in this guide, before operating the unit with the cover removed. Ensure that safety instructions are strictly followed. Failure to do so can result in severe or fatal injury. 3. Turn the analyzer LINE switch to ON. 4. Center the ne focus adjustment. Refer to Figure 2-1 for the adjustment location. 5. Adjust the coarse focus adjustment for the best possible focus. See Figure 2-2.
Figure 2-2
Chapter 2
35
Making Adjustments 1b. Display (with eight access holes in top of shield)
All display adjustments are described in this adjustment procedure. However, the FOCUS adjustment is the only adjustment that can be performed without removing the instrument cover. For all other adjustments, it is strongly recommended that you return the instrument to your local sales and service ofce. There are two different types of displays. If your display has eight holes in the top of the shield, continue with this procedure. If it has four holes, go to the preceding procedure. If you are going to make adjustments to your display, only make the adjustments that would correct the problem that the display is experiencing. The following is a list of adjustments described in this procedure:
Focus Horizontal Hold Vertical Size Brightness Centering Vertical Hold Vertical Linearity Contrast
The horizontal and vertical display positions should rst be adjusted using front-panel softkeys. These positions are then stored in nonvolatile memory. The Focus adjustment is located on the left side of the instrument and can be accessed with the instrument cover on or off. All other adjustments must be made with the instrument cover removed.
Equipment Required
Display adjustment tool, 0.075 inch hex (part number 8710-1010)
36
Chapter 2
Making Adjustments 1b. Display (with eight access holes in top of shield)
Procedures
Before Performing any Adjustment
Before performing any display adjustment, press the following analyzer keys.
CONFIG, More 1 of 3, More 2 of 3 DEFAULT SYNC
CAL, More 1 of 4, More 2 of 4
Focus
The Focus potentiometer is located on the left side of the instrument and can be accessed with the instrument cover on or off. Use the specied adjustment tool to make this adjustment. 1. Adjust the front-panel INTENSITY control for a comfortable viewing intensity. 2. Adjust as necessary for a focused display. See Figure 2-1. Figure 2-3 Focus Adjustment Location
Chapter 2
37
Making Adjustments 1b. Display (with eight access holes in top of shield)
Centering
Perform this adjustment only if the displayed image is not horizontally centered on the display. The Centering potentiometer is accessible through the top of the display's cover. The instrument cover must be removed for access to the display's cover. Refer to Figure 2-4 for the location of this adjustment. Use the specied adjustment tool to make this adjustment. 1. Press CAL, More 1 of 4, CRT HORZ POSITION. 2. Set the CRT horizontal position setting to 12. 3. Set the Centering potentiometer so the displayed image is positioned approximately at the center of the display. Figure 2-4 Display Adjustment Locations
38
Chapter 2
Making Adjustments 1b. Display (with eight access holes in top of shield)
Horizontal Hold
Perform this adjustment only if the displayed image is rolling horizontally across the display. The Horizontal Hold potentiometer is accessible through the top of the display's cover. The instrument cover must be removed for access to the display's cover. Refer to Figure 2-4 for the location of this adjustment. Use the specied adjustment tool to make this adjustment. 1. If the display is unreadable, rst adjust the Horizontal Hold potentiometer for a stable display. 2. Press CAL, More 1 of 4, CRT HORZ POSITION. 3. Set the CRT horizontal position setting to 12. 4. Find the clockwise position of the Horizontal Hold potentiometer where the display begins to roll horizontally. 5. Find the counter-clockwise position of the Horizontal Hold potentiometer where the display begins to roll horizontally. 6. Set the adjustment to a position approximately midway between the clockwise and counter clockwise position where the display does not roll horizontally.
Vertical Hold
Perform this adjustment only if the displayed image is rolling vertically on the display. The Vertical Hold potentiometer is accessible through the top of the display's cover. The instrument cover must be removed for access to the display's cover. Refer to Figure 2-4 for the location of this adjustment. Use the specied adjustment tool to make this adjustment. 1. If the display is unreadable, rst adjust the Vertical Hold potentiometer for a stable display. 2. Press CONFIG, More 1 of 3, More 2 of 3. 3. Press SYNC NRM NTSC (NTSC). 4. Adjust the Vertical Hold potentiometer so that the display image is not rolling vertically.
Chapter 2
39
Making Adjustments 1b. Display (with eight access holes in top of shield)
5. Press SYNC NRM PAL (PAL). 6. Adjust the Vertical Hold potentiometer so that the display image is not rolling vertically. 7. Repeat step 1 to step 5 until no adjustment is necessary for either the NTSC or the PAL mode. 8. Press DEFAULT SYNC. This is the third softkey from the top.
Vertical Size
Perform this adjustment only if the displayed image is too tall or too short for the display. The Vertical Size potentiometer is accessible through the top of the display's cover. The instrument cover must be removed for access to the display's cover. Refer to Figure 2-4 for the location of this adjustment. Use the specied adjustment tool to make this adjustment. 1. Press CAL, More 1 of 4, CRT VERT POSITION. 2. Set the CRT vertical position setting so that softkey labels are adjacent to the keys at the right to which they are associated. 3. Adjust the Vertical Size potentiometer so that the uppermost softkey label is adjacent to the uppermost softkey and the lowermost softkey label is adjacent to the lowermost softkey. 4. Repeat step 2 and step 3 if necessary.
Vertical Linearity
Perform this adjustment only if the vertical space between the horizontal graticules are not equal on the display. The Vertical Linearity potentiometer is accessible through the top of the display's cover. The instrument cover must be removed for access to the display's cover. Refer to Figure 2-4 for the location of this adjustment. Use the specied adjustment tool to make this adjustment. 1. Adjust the Vertical Linearity potentiometer so that the vertical space between each of horizontal graticules are approximately equal.
40
Chapter 2
Making Adjustments 1b. Display (with eight access holes in top of shield)
Chapter 2
41
2. Sampler Match
This adjustment applies to: All 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers The match between the sampling oscillator and the sampler is optimized by rst setting the sampling-oscillator frequency for midrange, then adjusting the sampler-match adjustment for maximum dc volts as read on a digital multimeter.
Equipment Required
Digital multimeter DMM test leads
Procedure
1. Turn the analyzer LINE switch to OFF. Remove the instrument over assembly. 2. Turn the analyzer LINE switch to ON. 3. Set the center frequency of the analyzer to 194 MHz. If you are adjusting an 8590L, or 8591E spectrum analyzer, or an 8591C cable TV analyzer, set the center frequency to 265 MHz. 4. Press the following analyzer keys.
SPAN, 10, MHz
5. Connect the digital multimeter (DMM) from chassis ground to A25TP1. Refer to Figure 2-5. 6. Adjust A25C107 for maximum voltage as read on the DMM. This voltage must be 1.5 V 1.0 V. 7. Measure the voltage at A25TP2. It should be +1.5 V 1.0 V. If it is not, readjust A25C107 until a compromise is established between the two test points, such that the voltage specications of step 6 and step 7 are met.
42
Chapter 2
Figure 2-5
Chapter 2
43
3. 10 MHz Reference
This adjustment applies to: All 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers with Option 704 8594Q QAM analyzers with Option 704 The internal 10 MHz time base is adjusted for frequency accuracy. This procedure does not adjust for long-term drift or aging rate. Only short-term accuracy is adjusted. A frequency counter is connected to the CAL OUT, which is locked to the 10 MHz reference. This yields better effective resolution. The time base is adjusted for a frequency of 300 MHz as read by the frequency counter.
Equipment Required
Microwave frequency counter Frequency standard Cable, 122 cm (48 in) (two required) Figure 2-6 10 MHz Reference Adjustment Setup
44
Chapter 2
Procedure
Note that to properly adjust the time base, a frequency standard with a better time base accuracy than that of the analyzer is required. 1. Connect the equipment as shown in Figure 2-6. 2. Set the microwave frequency counter controls as follows: 50 /1 M ...........................................50 Impedance 10 Hz500 MHz/500 MHz18 GHz ...10 Hz500 MHz SAMPLE RATE .......................................... Midrange FREQUENCY STANDARD ................... EXTERNAL 3. Press the following analyzer keys.
PRESET FREQUENCY, 2001, Hz CAL, More 1 of 4, More 2 of 4, More 3 of 4 SERVICE CAL, CAL TIMEBASE
4. A number will be displayed in the active function block of the analyzer display. This is the setting of the DAC (0 to 255) which controls the frequency of the internal time base. Use the knob or keyboard to change the DAC setting until the frequency counter reads 300 MHz 75 Hz (0.25 ppm). 5. Once the time base has been adjusted for minimum deviation from 300 MHz, press CAL, CAL STORE. The new DAC number is now stored in nonvolatile memory.
Chapter 2
45
Equipment Required
Frequency standard Frequency counter Cable, BNC, 122 cm (48 in) (two required) Figure 2-7 Precision Frequency Reference Setup
46
Chapter 2
Procedure
1. Place the analyzer on its side as shown in Figure 2-7 and set the LINE switch of the analyzer to ON. 2. Allow the analyzer to remain powered ON and undisturbed for at least 24 hours, so that both the temperature and frequency of the OCXO can stabilize. 3. Connect the frequency standard to the frequency counter rear-panel TIMEBASE IN/OUT connector. Refer to Figure 2-7. 4. Disconnect the jumper between the 10 MHz REF OUTPUT and EXT REF IN jacks on the analyzer rear panel. Connect a BNC cable between the 10 MHz REF OUTPUT jack and INPUT A on the frequency counter. 5. Set the frequency counter controls as follows: FUNCTION/DATA ....................................... FREQ A INPUT A: x10 ATTN ........................................................ OFF AC ............................................. OFF (DC coupled) 50 Z .................... OFF (1 M input impedance) AUTO TRIG ........................................................... ON 100 kHz FILTER A .............................................. OFF INT/EXT switch (rear panel) .............................. EXT 6. On the frequency counter select a 1 second gate time and a 10 MHz offset of the displayed frequency by pressing the following frequency counter keys.
GATE TIME, 1, GATE TIME MATH, SELECT/ENTER CHS/EEX, 10, CHS/EEX, 6, SELECT/ENTER SELECT/ENTER
The frequency counter should now display the difference between the frequency of the INPUT A signal and 10.0 MHz with a displayed resolution of 10 MHz (0.010 Hz). 7. Locate the FREQ ADJ control on the OCXO. See Figure 2-8. Remove the dust cap screw. 8. Use a nonconductive adjustment tool to adjust the FREQ ADJ control on the OCXO for a frequency counter indication of 0.00 Hz.
Chapter 2
47
9. Select a 10 second gate time by pressing the following frequency counter keys.
GATE TIME, 10, GATE TIME
The frequency counter should now display the difference between the frequency of the INPUT A signal and 10.0 MHz with a resolution of 0.001 Hz (1 MHz). 10.Wait at least 2 gate periods for the frequency counter to settle, and then adjust the FREQ ADJ control on the OCXO for a stable frequency counter indication of 0.000 0.010 Hz. 11.Replace the dust cap screw on the OCXO. Figure 2-8 Oven Reference Adjustment Location
48
Chapter 2
Equipment Required
Crystal shorts (set of three) Cable, BNC, 23 cm (9 in)
Procedure
1. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. 2. Turn the analyzer LINE switch to ON, then press the following analyzer keys.
PRESET CAL, More 1 of 4 CORRECT ON OFF (OFF)
Chapter 2
49
Crystal Alignment
3. Connect the CAL OUT to the INPUT 50 with the BNC cable. 75 input: Connect the CAL OUT to the INPUT 75 with the 75 BNC cable. Figure 2-9 Crystal Symmetry and Centering
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 4. Press the following analyzer keys.
FREQUENCY, 300, MHz SPAN, 10, MHz PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON)
50
Chapter 2
5. Press AMPLITUDE, then use the knob to place the signal at the sixth graticule line from the bottom. 6. Press BW, 30, kHz on the analyzer.
CAUTION
Shorting the crystal test points to ground may permanently damage the bandwidth board assembly. If you make your own shorts, it is advisable to insulate the bare wires and connectors. 7. Connect the crystal shorts (through the access holes on the assembly cover) across the following pairs of test points. A13TP1 and A13TP2 A11TP1 and A11TP2 A11TP4 and A11TP5 8. Adjust A13C54 CTR for minimum signal amplitude. Then adjust A13C38 SYM and A13C54 CTR for a centered and symmetrical bandpass response as shown in Figure 2-9. 9. Remove the crystal short from A13TP1 and A13TP2 and connect it across A13TP4 and A13TP5. 10.Adjust A13C25 CTR for minimum signal amplitude. Then adjust A13C15 SYM and A13C25 CTR for a centered and symmetrical bandpass response. 11.Remove the crystal short from A11TP4 and A11TP5. Connect the short across A13TP1 and A13TP2. 12.Adjust A11C54 CTR for minimum signal amplitude. Then adjust A11C38 SYM and A11C54 CTR for a centered and symmetrical bandpass response. 13.Remove the crystal short from A11TP1 and A11TP2. Connect the short across A11TP4 and A11TP5. 14.Adjust A11C25 CTR for minimum signal amplitude. Then adjust A11C15 SYM and A11C25 CTR for a centered and symmetrical bandpass response.
Chapter 2
51
15.Remove the crystal shorts and press the following analyzer keys.
PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) SPAN, 50, kHz MKR FCTN, MK TRACK ON OFF (OFF) BW, 3, kHz PEAK SEARCH, MARKER DELTA BW, 30, kHz PEAK SEARCH
16.Verify that the MARKER frequency does not exceed 3 kHz. If the signal shift is out of tolerance, repeat step 3 to step 16. 17.Press the following analyzer keys.
MKR MARKER 1 ON OFF (OFF)
52
Chapter 2
LC Alignment
18.Press the following analyzer keys.
BW, 100, kHz SPAN, 5, MHz
19.Widen all but one of the LC lter poles by shorting A11TP10 and A11TP11, A11TP12 and A11TP13, and A13TP10 and A13TP11 using the crystal shorts.
CAUTION
Use a tool with a nonmetallic body to make the LC dip adjustment. Shorting components to ground may result in permanent damage to the bandwidth board assembly. 20.Center the signal on the analyzer display by pressing the following analyzer keys.
PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON)
21.Adjust A13C47 LC dip for a minimum signal amplitude. 22.Move the short from A13TP10 and A13TP11 to A13TP12 and A13TP13, then adjust A13C17 LC dip for a minimum signal amplitude. 23.Move the short from A11TP10 and A11TP11 to A13TP10 and A13TP11, then adjust A11C17 LC dip for a minimum signal amplitude. 24.Move the short from A11TP12 and A11TP13 to A11TP10 and A11TP11, then adjust A11C47 LC dip for a minimum signal amplitude.
LC Centering Note that the center frequency of the 100 kHz bandwidth is referenced to the 30 kHz bandwidth. During this procedure it is advisable to switch to the 30 kHz bandwidth occasionally and recenter it using PEAK SEARCH, MARKER CF. 25.Short A11TP10 and A11TP11, A11TP12 and A11TP13, and A13TP10 and A13TP11 using the crystal shorts used in the crystal alignment section. Press the following analyzer keys.
BW, 30, kHz SPAN, 200, kHz MKR FCTN, MK TRACK ON OFF (OFF) BW, 100, kHz
Chapter 2
53
26.Adjust A13C45 LC CTR for maximum signal at center-screen. 27.Move the short from A11TP10 and A11TP11 to A11TP12 and A11TP13, then adjust A11C23 LC CTR for maximum signal at center-screen. 28.Move the short from A13TP12 and A13TP13 to A11TP10 and A11TP11, then adjust A13C45 LC CTR for maximum signal at center-screen. 29.Move the short from A13TP10 and A13TP11 to A13TP12 and A13TP13, then adjust A13C23 LC CTR for maximum signal at center-screen. 30.Disconnect all the shorts from A11 and A13 bandwidth board assemblies.
32.Press AMPLITUDE and adjust the signal level one division below the top graticule using the knob. 33.Press the following analyzer keys.
PEAK SEARCH, MARKER MKR FCTN, MK TRACK ON OFF (ON) BW, 100, kHz
It may be necessary to occasionally recenter the 30 kHz bandwidth. 34.Adjust A11R26 LC and A13R26 LC equally for a MARKER amplitude of 0 dB. Each potentiometer should be adjusted to accomplish one-half of the necessary increase in signal amplitude. If A11R26 or A13R26 reaches its limit, recenter both potentiometers and repeat step 31 to step 35. 35.Press MKR FCTN, MK TRACK ON OFF (OFF) on the analyzer.
54
Chapter 2
Final LC Centering
36.Press the following analyzer keys.
BW, 30, kHz SPAN, 100, kHz PEAK SEARCH, MARKER CF BW, 100, kHz
37.Make nal adjustments by adjusting A11C23, A11C45, A13C23, and A13C45 in succession to peak the amplitude of the marker at center-screen. 38.Repeat step 36 and step 37 until the 30 kHz and 100 kHz bandwidths are centered in relation to each other. 39.Press the following analyzer keys.
BW, 30, kHz PEAK SEARCH MARKER BW, 100, kHz PEAK SEARCH
40.Verify that the MARKER frequency does not exceed 10 kHz. If the signal shift is out of tolerance, repeat step 26 to step 39.
42.Adjust A11R31 XTL and A13R31 XTL equally for a MARKER amplitude reading of 0 dB. Each potentiometer should be adjusted to accomplish one-half of the necessary increase in signal amplitude. 43.If A11R31 or A13R31 reaches its limit, recenter both potentiometers and repeat step 41 and step 42.
Chapter 2
55
47.Refer to the BW-AMP column of the analyzer display to locate the XTAL and LC bandwidth amplitude-correction numbers of the analyzer. All LC and XTAL bandwidth readings should be between 0.8 dB to +0.5 dB. Table 2-1 describes and shows an example of the XTAL and LC bandwidth amplitude-correction numbers that will be displayed on the analyzer. Perform the following steps that apply to your analyzer. If the difference between the bandwidth amplitude-correction numbers of the 30 kHz XTAL and 3 MHz LC is greater than 0.8 dB, pad A11R8 or A13R8 and repeat the Final BW Amplitude Check. Refer to the Component-Level Information Package for the location of A11R8 and A13R8. If the 30 kHz XTAL bandwidth amplitude correction is greater than the 3 MHz LC bandwidth correction, increase the value of A11R8 or A13R8. If the 3 MHz LC bandwidth amplitude correction is greater than the 30 kHz XTAL bandwidth correction, decrease the value of A11R8 or A13R8. If just the 100 kHz LC amplitude is out of range, repeat step 23 to step 38 and step 44 to step 47. If the 1 kHz XTAL amplitude is out of the above range, repeat step 44 to step 46.
56
Chapter 2
Table 2-1
LC:
Chapter 2
57
Equipment Required
Synthesizer/level generator Cable, BNC, 120 cm (48 in) Adapter, Type N (m) to BNC (f)
58
Chapter 2
Procedure
The accuracy of the amplitude control attenuator is critical to the proper calibration of the instrument; therefore, this procedure must be carefully and accurately performed. 1. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. 2. Turn the analyzer LINE switch to ON. 3. Connect the 50 output of the synthesizer/level generator to the analyzer INPUT 50 . See Figure 2-10. 75 input: Connect the 75 output of the synthesizer/level generator to the analyzer INPUT 75 . Figure 2-10 Cal Attenuator Error Correction Setup
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 4. Set the frequency of the synthesizer/level generator to 25 MHz and the output to 19 dBm. 5. To turn the amplitude attenuator correction constants off, press the following analyzer keys.
PRESET CAL, More 1 of 4 CORRECT ON OFF (OFF)
Chapter 2
59
1 dB Step Check
6. To measure the 1 dB step correction, press the following analyzer keys.
FREQUENCY, 25, MHz SPAN, 10, MHz
8. The MKR amplitude reading should be 0.0 dB 0.01 dB. If it is not, repeat step 7. 9. Set the analyzer by pressing AMPLITUDE, 17, dBm. 10.Set the amplitude of the synthesizer/level generator to 18 dBm. 11.Press MKR on the analyzer. 12.Record the MKR amplitude reading in Table 2-2. This is the 1 dB attenuator step error of the A12 amplitude control assembly.
60
Chapter 2
2 dB Check
13.Press the following analyzer keys.
AMPLITUDE, 16, dBm.
14.Set the amplitude of the synthesizer/level generator to 17 dBm. 15.Press the following analyzer keys.
PEAK SEARCH MARKER DELTA, MARKER DELTA MKR FCTN, MK TRACK ON OFF (ON)
16.MKR amplitude reading should be 0.0 0.01. If it is not, repeat step 15. 17.Set the analyzer by pressing AMPLITUDE, 14, dBm. 18.Set the amplitude of the synthesizer/level generator to 15 dBm. 19.Press MKR on the analyzer. 20.Record the MKR amplitude reading in Table 2-2. This is the 2 dB attenuator step error of the A12 amplitude control assembly.
Table 2-2
Attenuator OFF 19 17 18 19
Attenuator ON 18 15 14 11
Chapter 2
61
Note that the frequency of 2001 Hz is necessary to access the SERVICE CAL routines. 23.When the analyzer prompts you with the message ENTER CAL ATTEN ERROR, enter the data from Table 2-2 and terminate with the reverse sign using the dBm or dBm keys. For example, if the error recorded in Table 2-2 is 0.07, enter 0.07 and terminate with dBm. 24.When the analyzer prompts you to enter the 16 dB step, enter 0 dBm. At the completion of entering the 16 dB error, the analyzer will reset. 25.To conrm that the correct data is stored, access the cal attenuator corrections by pressing the following analyzer keys.
CAL, More 1 of 4, More 2 of 4 SERVICE DIAG DISPLAY CAL DATA
Note that the cal-attenuator correction data are the rst ve corrections located in the ERR column.
62
Chapter 2
Equipment Required
Synthesizer/level generator Digital multimeter (DMM) IF test board Cable, BNC, 120 cm (48 in) Cable, BNC (f) to dual banana plug Test cable
Chapter 2
63
Procedure
1. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. 2. Remove the rst IF bandwidth lter assembly, A11. Install the IF test board into the A11 slot. Turn the analyzer LINE switch to ON. 3. Set the DMM to read dc volts. 4. Press the following analyzer keys.
PRESET CAL, More 1 of 4 CORRECT ON OFF (OFF), More 2 of 4 SERVICE DIAG STP GAIN ZERO SPAN, 0, Hz BW, 10, kHz VID BW AUTO MAN, 300, Hz AMPLITUDE, 10, dBm SCALE LOG LIN (LIN)
5. Set the synthesizer as follows: FREQUENCY ............................................. 21.4 MHz MANUAL TUNE ................................................... ON AMPTD INCR .............................................. 0.01 dBm AMPLITUDE .................................................. 6 dBm 6. Connect equipment as shown in Figure 2-11. Connect the output of the synthesizer to J2 of the IF test board. Connect the DMM to AUX VIDEO OUT (located on the rear panel).
64
Chapter 2
Figure 2-11
Chapter 2
65
Figure 2-12
12.Set the synthesizer amplitude 60 dB below that recorded in step 8 by pressing AMPLITUDE and then pressing (step-down key) six times. Adjust A14R10 OFFSET for the DMM reading of 250 mV 1 mV. 13.Repeat step 10 and step 11 until no further adjustment is necessary. 14.Set the synthesizer amplitude 30 dB below that recorded in step 8 and adjust the A14R23 SLOPE for a DMM reading of 625 mV 1 mV. 15.Set the synthesizer amplitude to the level recorded in step 8 and adjust the A14R69 30 dB for a DMM reading of 1000 mV 1 mV. 16.Repeat step 13 and step 14 until no further adjustment is necessary. 17.Set the synthesizer amplitude 10 dB below that recorded in step 8 and adjust the A14R23 SLOPE for a DMM reading of 875 mV 1 mV. 18.Set the synthesizer amplitude to the level recorded in step 8 and adjust the A14R39 10 dB for a DMM reading of 1000 mV 1 mV. 19.Repeat step 16 and step 17 until no further adjustment is necessary. 20.Repeat step 10 to step 18 until the limits in Table 2-3 are met.
66
Chapter 2
Table 2-3
22.Set the synthesizer amplitude to the level recorded in step 8 and adjust A14R34 LIN for a DMM reading of 1000 mV 1 mV. 23.Make the adjustments indicated in Table 2-9. Table 2-4
Adjust A14R34 A14R33 A14R30 A14R27 N/A
Chapter 2
67
Equipment Required
Cable, BNC, 20 cm (9 in) Adapter, Type N (m) to BNC (f)
68
Chapter 2
Procedure
Interrupting this routine may result in corrupt data being stored in RAM. If this occurs, rerun the CAL FREQ adjustment routine. 1. Connect the CAL OUT to the INPUT 50 with the BNC cable. See Figure 2-13. 75 input: Connect the CAL OUT to the INPUT 75 with the 75 BNC cable. Figure 2-13 CAL FREQ Adjustment Routine Setup
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 2. Press the following analyzer keys.
CAL, CAL FREQ
The CAL FREQ adjustment routine will take a few minutes to run. The internal adjustment data will be stored in working RAM. To store this data in nonvolatile memory, press CAL STORE. If the CAL AMPTD adjustment routine has not been done, the following error message will be displayed: 'CAL: DATA NOT STORED' and 'CAL AMP NEEDED'. Proceed with the CAL AMPTD adjustment routine and then press CAL STORE when done.
Chapter 2
69
Equipment
Cable, BNC, 20 cm (9 in) Adapter, Type N (m) to BNC (f)
70
Chapter 2
Procedure
It is recommended to complete the CAL FREQ Adjustment Routine prior to performing the CAL AMPTD Adjustment Routine. 1. Connect the CAL OUT to the INPUT 50 using a BNC cable. See Figure 2-14. 75 input: Connect the CAL OUT to the INPUT 75 with the 75 BNC cable. Figure 2-14 CAL AMPTD Adjustment Routine Setup
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 2. Press the following analyzer keys.
CAL, CAL AMPTD
The CAL AMPTD routine takes approximately 5 to 7 minutes to run. The internal adjustment data will be stored in working RAM. To store this data in nonvolatile memory, press CAL STORE.
Chapter 2
71
Equipment Required
Cable, Type N (m) to SMA (m) (part number 8120-5148)
72
Chapter 2
Procedure
1. Perform the CAL FREQ routine as indicated in the CAL FREQ Adjustment Routine in this chapter. 2. Connect the 100 MHz COMB OUT to INPUT 50 using the YTF CAL cable. Refer to Figure 2-15. Model 8595E only: Connect the CAL OUT to the INPUT 50 using a BNC cable. Figure 2-15 CAL YTF Adjustment Setup
The CAL YTF routine will take a few minutes to run. The message CAL: DONE will be displayed when the routine has nished. The internal adjustment data will be stored in working RAM. If the message LOST COMB SIGNAL is displayed, perform the Mixer Bias DAC Initialization procedure. 4. Press CAL STORE to store the YTF correction data in nonvolatile memory.
Chapter 2
73
2. Record the displayed Optimum Bias DAC values for each frequency band. Table 2-5 Acceptable Mixer Bias DAC Correction Values
Band 1 2 3 4 Optimum Bias _________________ _________________ _________________ _________________ Acceptable Bias Range 800 to 1900 0 to 200 1000 to 1800 400 to 750
3. If the recorded Optimum Bias values are within the acceptable ranges indicated in Table 2-5, RF section troubleshooting is necessary. If the recorded Optimum Bias values are not within the acceptable ranges indicated in Table 2-5, complete the rest of this procedure to initialize the mixer bias DAC values. 4. Press the following keys to allow entry of the default mixer bias DAC values.
PRESET FREQUENCY, 2001, Hz DISPLAY CHANGE TITLE
5. Enter CALMXRDATA 1600,1; as a title entry to set the mixer bias value to 1600 for band 1. 6. Press the following keys to store the mixer bias value for band 1 in nonvolatile memory.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL EXECUTE TITLE
74
Chapter 2
7. Press the following keys to allow entry of the default mixer bias DAC value for band 2.
DISPLAY CHANGE TITLE CLEAR
8. Enter CALMXRDATA 20,2; as a title entry to set the mixer bias value to 20 for band 2. Press the following keys to store the mixer bias value for band 2 in nonvolatile memory.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL EXECUTE TITLE
9. Press the following keys to allow entry of the default mixer bias DAC value for band 3.
DISPLAY CHANGE TITLE CLEAR
10.Enter CALMXRDATA 1300,3; as a title entry to set the mixer bias value to 1300 for band 3. 11.Press the following keys to store the mixer bias value for band 3 in nonvolatile memory.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL EXECUTE TITLE
12.Press the following keys to allow entry of the default mixer bias DAC value for band 4.
DISPLAY CHANGE TITLE CLEAR
13.Enter CALMXRDATA 520,4; as a title entry to set the mixer bias value to 520 for band 4.
Chapter 2
75
14.Press the following keys to store the mixer bias value for band 4 in nonvolatile memory.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL EXECUTE TITLE
15.Press the following keys to check the default mixer bias DAC values.
PRESET CAL, More 1 of 4, More 2 of 4 SERVICE DIAG DISPLAY CAL DATA
Conrm that the Optimum Bias DAC values displayed for band 1 through band 4 are 1600, 20, 1300, and 520. 16.Repeat the CAL YTF adjustment procedure. 17.Perform the CAL MXR adjustment procedure.
76
Chapter 2
Equipment Required
Cable, Type N (m) to SMA (m)
Chapter 2
77
Procedure
1. Perform the CAL YTF routine as indicated in the CAL YTF Adjustment Routine in this chapter. 2. Connect the 100 MHz COMB OUT to INPUT 50 using the CAL YTF cable (Type N to SMA cable). Refer to Figure 2-16. Model 8595E only: Connect the CAL OUT to the INPUT 50 using a BNC cable. Figure 2-16 CAL MXR Adjustment Setup
Note that the frequency of 2001 Hz is necessary to access the SERVICE CAL routines. The CAL MXR routine will take a few minutes to run. The message CAL: DONE will be displayed when the routine has nished. The internal adjustment data will be stored in working RAM. 4. Press CAL STORE to store the correction data in nonvolatile memory.
78
Chapter 2
Equipment Required
Microwave spectrum analyzer Tuning tool Cable, SMB (f) to BNC (m) Adapter, SMB (f) to SMB (f) Adapter SMB (m) to SMB (m) Adapter, N (m) to BNC (f)
Chapter 2
79
Procedure
Tripler Adjustment
1. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. Refer to the Instrument Cover removal procedure in Chapter 3. 2. Turn the analyzer LINE switch to ON. 3. Press INSTRUMENT PRESET on the microwave spectrum analyzer and set the controls as follows: CENTER FREQUENCY ............................... 1.8 GHz FREQUENCY SPAN ..................................... 10 MHz REFERENCE LEVEL .................................. 20 dBm dB/DIV ......................................................... 2 dB/DIV 4. Connect the microwave spectrum analyzer to the 2ND LO 1.8 GHz port on the A5 Second Converter using SMB adapters and the test cable. See Figure 2-17. Figure 2-17 Tripler Adjustment Setup
5. Use the tuning tool to adjust A5C4 FREQUENCY TRIPLER (see Figure 2-18) for maximum signal amplitude as observed on the analyzer. Readjust the reference level of the microwave spectrum analyzer as necessary to keep the signal on screen. The 1.8 GHz signal should be 19 dBm to 24 dBm.
80
Chapter 2
Note that when adjusting with the tuning tool, the signal amplitude will change as the lock nuts are tightened. Optimum signal amplitude should be maintained while tightening these lock nuts. Figure 2-18 Second Converter Adjustment Location
Chapter 2
81
8. With nothing connected to the analyzer INPUT 50 (Option 001: INPUT 75 ), press the following keys.
PRESET PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) SPAN, 10, MHz
Wait for AUTO ZOOM message to disappear. 9. Set the analyzer as follows:
MKR FCTN, MK TRACK ON OFF (OFF) SPAN, 0, Hz
10.Loosen the lock nut on A5C1, A5C2, and A5C3. Carefully turn tuning screws clockwise until they reach the bottom of the cavity. Do not force the tuning screws down. 11.Turn A5C1 counterclockwise for a peaked signal on the 8566A/B. Lightly tighten the lock nut. 12.Turn A5C2 counterclockwise for a minimum signal on the analyzer. Lightly tighten the lock nut. 13.Turn A5C3 counterclockwise for peak signal on the analyzer. Lightly tighten the lock nut. 14.Repeat step 11 to step 13 until no further improvement is noticeable. Tighten the lock nuts without changing the amplitude on the display.
82
Chapter 2
Figure 2-20
16.Set the microwave spectrum analyzer controls as follows: CENTER FREQUENCY .......................... 321.4 MHz REFERENCE LEVEL .................................. 40 dBm SPAN .............................................................. 20 MHz dB/DIV ......................................................... 1 dB/DIV 17.Adjust A5 2ND MIXER MATCH (refer to Figure 2-18) for maximum amplitude as displayed on the microwave spectrum analyzer. Adjust the reference level as necessary to keep the signal near top-screen.
19.On the microwave spectrum analyzer, press trace B and MAX HOLD. Wait for at least one complete sweep. Then press the following keys to measure the 3 dB bandwidth.
TRACE B VIEW AB PEAK SEARCH MARKER MODE,
20.Rotate the knob counterclockwise so that the marker delta amplitude reads 3 dB.
Chapter 2
83
21.Press MARKER MODE, and rotate the knob clockwise so that the marker delta amplitude reads 0 dB and is on the right-hand side of the response. Read the marker delta frequency; it should be 15 MHz 2.5 MHz. See Figure 2-21. 22.The maximum ripple should be less than 1 dB for a bandwidth of 6 MHz. Figure 2-21 Second Converter Bandpass Ripple Measurement
84
Chapter 2
Equipment Required
Synthesized sweeper Microwave spectrum analyzer Measuring receiver (used as a power meter) Power meter Low power sensor with a 50 MHz reference attenuator Power sensor, 300 MHz Power splitter Attenuator, 10 dB, Type N (m to f), dc-12.4 GHz Low-pass lter, 300 MHz IF test board Cable, Type N, 152 cm (60 in) Cable, BNC, 120 cm (48 in) Test cable, SMB (f) to BNC (m) (two required) Adapter, APC 3.5 (f) to Type N (f) Adapter, Type N (f) to BNC (m) (two required) Adapter, Type N (m) to BNC (f)
Chapter 2
85
Procedure
Second IF Bandpass Filter
1. Press INSTRUMENT PRESET on the microwave spectrum analyzer, then set the controls as follows: CENTER FREQUENCY ............................ 21.4 MHz FREQUENCY SPAN ..................................... 50 MHz REFERENCE LEVEL .................................. 30 dBm dB/DIV ......................................................... 1 dB/DIV 2. Set the synthesized sweeper controls as follows: CW ............................................................. 321.4 MHz POWER LEVEL ........................................... 26 dBm 3. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. 4. Remove the rst IF bandwidth lter assembly, A11. 5. Install the IF test board into the A11 slot. 6. Remove the W9 cable from A9J4, 321.4 MHz IF INPUT. 7. Connect the synthesized sweeper output to A9J4. Refer to Figure 2-22.
86
Chapter 2
Figure 2-22
8. Press the analyzer LINE switch to ON. 9. Press the following analyzer keys.
PRESET SPAN, 0, Hz
10.Connect the microwave spectrum analyzer RF INPUT to J1 of the IF test board. Refer to Figure 2-22. 11.Adjust A9C44, A9C46, and A9C47 for maximum signal amplitude as observed on the microwave spectrum analyzer. Adjust the reference level of the microwave spectrum analyzer, as necessary, to display the signal below the top graticule. 12.Remove the test board from the A11 slot and install the A11 bandwidth lter assembly. 13.Reconnect W9 to A9J4, 321.4 MHz INPUT. If you are adjusting an 8590L Option 713, continue with the 600 MHz Adjustment. If you are adjusting an 8591C 8591E, 8593E, 8594E, 8594Q, 8595E, or 8596E, continue with the LPF, Attenuator and Adapter Insertion Loss Characterization.
Chapter 2
87
Do not attempt to calibrate the power sensor without the reference attenuator or damage to the power sensor will occur. 20.Zero and calibrate the power meter and 300 MHz power sensor, as described in the power meter operation manual. 21.Connect the equipment as shown in Figure 2-23. Connect the 300 MHz power sensor directly to the Power Splitter (bypass the LPF, Attenuator and Adapters).
88
Chapter 2
Figure 2-23
LPF Characterization
22.Press INSTRUMENT PRESET on the synthesized sweeper. Set the controls as follows: CW ................................................................ 300 MHz POWER LEVEL ........................................... 15 dBm 23.Allow the power sensors to settle, then on the measuring receiver, press RATIO mode. Power indication should be 0 dB. 24.On the power meter, press the dB REF mode key. Power indication should be 0 dB. 25.Connect the LPF, Attenuator and adapters as shown in Figure 2-23. 26.Record the measuring receiver reading in dB. This is the relative error due to mismatch. Mismatch Error____________________dB 27.Record the power meter reading in dB. This is the relative uncorrected insertion loss of the LPF, attenuator, and adapters. Uncorrected Insertion Loss____________________dB
Chapter 2
89
28.Subtract the Mismatch Error (step 22) from the Uncorrected Insertion Loss (step 22). This is the corrected insertion loss. Corrected Insertion Loss____________________dB For example, if the Mismatch Error is +0.3 dB and the uncorrected Insertion Loss is 10.2 dB, subtracting the mismatch error to the insertion loss gives a corrected reading of 10.5 dB.
30.On the power meter, press the dBm mode key. Record the power meter reading in dBm. Power Meter Reading____________________dBm 31.Subtract the Corrected Insertion Loss (step 23) from the power meter reading (step 25) and record as the CAL OUT power. The CAL OUT should be 20 dBm 0.4 dB. CAL OUT Power = Power Meter Reading Corrected Insert Loss
90
Chapter 2
For example, if the Corrected Insertion Loss is 10.0 dBm, and the measuring receiver reading is 30 dBm, then 30 dBm (10.0) dBm = 20 dBm. CAL OUT Power____________________dBm 32.Adjust A9R19 CAL OUT ADJ accordingly if the CAL OUT amplitude is not 20 dBm 0.4 dB as calculated in step 26.
Chapter 2
91
Equipment Required
Oscilloscope Frequency counter Measuring receiver (used as a power meter) Power sensor, 300 MHz Attenuator, 20 dB Cable, SMA, 90 cm (36 in.) Cable, BMC, 120 cm (48 in.) Adapter, Type N (m) to SMA (f) Adapter, Type N (f) to Type N (f) Adapter, Type N (m) to BNC (f)
92
Chapter 2
Figure 2-25
Procedure
1. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. 2. Remove the front-end assembly. Refer to the Front-End Assembly removal procedure. 3. Remove the A3A1 comb generator assembly from the front-end assembly and place in a service position, leaving W13 connected to the A7 Analog Interface. 4. Remove W24 from the A3A1J1 comb generator output connector. Connect the SMA cable to A3A1J1. Refer to Figure 2-25. 5. Turn the analyzer LINE switch to ON. 6. Press the following analyzer keys.
AUX CTRL, COMB GEN ON OFF (ON)
Chapter 2
93
Frequency
7. Connect the other end of the SMA cable to the 20 dB attenuator. Connect the output of the 20 dB attenuator to the 10 Hz to 500 MHz input of the frequency counter using adapters and the BNC cable. 8. Adjust A3A1C5 FREQ for 100.000000 MHz 0.00004 MHz. Refer to Figure 2-25 for adjustment location. 9. Disconnect the BNC cable from the frequency counter and 20 dB attenuator, then connect the BNC cable to CH1 of the oscilloscope. Set the oscilloscope by activating AUTOSCALE, located under the setup key. 10.Adjust both A3A1C15 OUTPUT MATCH and A3A1C3 OSC PEAK for maximum peak-to-peak voltage on the oscilloscope display. 11.Reconnect the BNC cable through the 20 dB attenuator to the frequency counter input. The comb generator frequency must be 100.000000 MHz 0.00004 MHz. 12.Repeat step 8 to step 11 until the frequency is within specication. 13.Perform the following steps only if the comb generator frequency cannot be brought within tolerance with the output power peaked. Set the LINE switch to OFF and remove the power cord. Remove the A3A1 comb generator cover plate. Change the selected value of A3A1L3 to obtain an output frequency of 100.000500 MHz 0.004 MHz with A3A1C5 FREQ centered. Increasing the value of A3A1L3 increases the output frequency of the comb generator. The frequency of the oscillator decreases about 500 Hz when the cover plate is installed. Refer to Figure 2-26 for component location. Figure 2-26 Comb Generator Component Location
94
Chapter 2
Each time the value of A3A1L3 is changed, reconnect the power cord, set the LINE switch to ON, and adjust A3A1C3 OSC PEAK for maximum signal. The output frequency changes when A3A1C3 OSC PEAK is adjusted. Reinstall the comb generator cover plate and repeat step 8 to step 12.
Output Power
14.Zero and calibrate the power sensor/measuring receiver combination in log mode (power reads out in dBm). Enter the power sensor 0.1 GHz cal factor into the measuring receiver. 15.Connect the power sensor to the output of the 20 dB attenuator using an adapter. 16.Adjust A3A1C15 OUTPUT MATCH for maximum power output. The measuring receiver should measure +6 dBm 0.8 dB. 17.If the output power of the comb generator is out of tolerance, perform the following: Set the LINE switch to OFF, remove the power cord, and remove the A3A1 comb generator cover plate. Change the selected value of A3A1R6 to obtain an output power reading of +6.0 dBm 0.8 dB. Increasing the value of A3A1R6 decreases the output power of the comb generator, while decreasing the value increases the output power. Refer to Figure 2-26 for component location. Each time the value of A3A1R6 is changed, reconnect the power cord, set the LINE switch to ON, and adjust A3A1C16 OUTPUT MATCH for maximum power out. 18.Reinstall the comb generator assembly cover plate and all the screws. Connect a BNC cable from the 20 dB attenuator to the frequency counter input using adapters. 19.Adjust A3A1C5 FREQ for a frequency counter reading of 100.000000 MHz 0.000010 MHz (tolerance of 10 Hz). 20.Set the LINE switch to OFF and reconnect W24 to A3A1J1. 21.Reinstall A3A1 comb generator assembly into RF Section. Refer to the A3A1 comb generator assembly replacement procedure. 22.Reinstall the front-end assembly in the analyzer. Refer to the Front-End Assembly replacement procedure.
Chapter 2
95
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
Equipment Required
Synthesized sweeper Measuring receiver (used as a power meter) Frequency synthesizer Power sensor, 1 MHz to 1.8 GHz Power splitter Adapter, Type N (f) to APC 3.5 (m) Adapter, Type N (m) to Type N (m) Cable, BNC, 122 cm (48 in.) Cable, Type N, 183 cm (72 in.)
96
Chapter 2
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 5. Adjust the synthesized sweeper POWER LEVEL for a 0 dBm reading on the measuring receiver. 6. Record the power meter reading in Column 4 of Table 2-6, taking into account the cal factors of both power sensors.
Chapter 2
97
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
7. On the synthesized sweeper, press CW and STEP UP, to step through the remaining frequencies listed in Table 2-6. At each new frequency repeat step 5 and step 6, and enter each power sensor cal factor into the respective power meter.
Procedure
1. Zero and calibrate the measuring receiver and 1 MHz to 1.8 GHz power sensor in log mode as described in the measuring receiver operation manual. 2. Connect the equipment as shown in Figure 2-28. Figure 2-28 Frequency Response Setup
3. Press INSTRUMENT PRESET on the synthesized sweeper. Set the synthesized sweeper controls as follows: CW ................................................................ 300 MHz FREQ STEP ................................................... 37 MHz POWER LEVEL ............................................. 9 dBm
98
Chapter 2
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
5. Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 14 dBm 0.1 dB. 6. Press RATIO mode on the measuring receiver. 7. Press the following analyzer keys.
FREQUENCY, 41, MHz
8. Set the synthesized sweeper CW to 41 MHz. 9. Adjust the synthesized sweeper POWER LEVEL for an analyzer MKR-TRK amplitude reading of 14 dBm 0.1 dB. 10.Record the power ratio here and in Column 2 of Table 2-6 for 41 MHz. Measuring Receiver Reading at 41 MHz____________________dB 11.Press the following analyzer keys.
FREQUENCY, 78, MHz
12.Set the synthesized sweeper CW to 78 MHz. 13.Adjust the synthesized sweeper POWER LEVEL for an analyzer MKR-TRK amplitude reading of 14 dBm 0.1 dB. 14.Record the power ratio displayed on the measuring receiver in Column 2 of Table 2-6 for 78 MHz.
Chapter 2
99
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
15.On the analyzer, press FREQUENCY and (step up), to step through the remaining frequencies listed in Column 1 of Table 2-6. At each new frequency repeat step 13 to step 15, entering the power sensor Cal Factor into the measuring receiver as indicated in Column 3 of Table 2-6. 16.On the synthesized sweeper, press CW and STEP UP.
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur.
100
Chapter 2
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
18.Set the frequency synthesizer controls as follows: FREQUENCY ................................................ 41 MHz AMPLITUDE ................................................ 15 dBm AMPTD INCR ................................................. 0.05 dB 19.Press the following analyzer keys.
MKR, MARKER 1 ON OFF (OFF) SPAN, 12, MHz FREQUENCY, 41, MHz PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) BW, 10, kHz SPAN, 100, kHz
Wait for AUTO ZOOM message to disappear. 20.Adjust the frequency synthesizer AMPLITUDE until the MKR-TRK reads 14 dBm. This corresponds to the amplitude at 41 MHz recorded in step 10. Record the frequency synthesizer amplitude here. AMPLITUDE setting (41 MHz)____________________dBm 21.Set the frequency synthesizer FREQUENCY to 4 MHz. 22.AUTO ZOOM on the 4 MHz signal by pressing the following analyzer keys.
FREQUENCY, 4, MHz MKR, MKR FCTN, MK TRACK ON OFF (OFF) FREQUENCY, 4, MHz SPAN, 12, MHz PEAK SEARCH, NEXT PK RIGHT MKR FCTN, MK TRACK ON OFF (ON) SPAN, 100, kHz
Chapter 2
101
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
23.Adjust the frequency synthesizer AMPLITUDE for a MKR amplitude reading of 14.00 dBm.05 dB. Record the frequency synthesizer AMPLITUDE setting here. AMPLITUDE setting (4 MHz)____________________dBm 24.Subtract the frequency synthesizer AMPLITUDE setting (4 MHz) recorded in step 23 from the frequency synthesizer AMPLITUDE setting (41 MHz) recorded in step 20. Record the result as the Amplitude Relative to 41 MHz here. 4 MHz Amplitude Relative to 41 MHz____________________dBm 25.Add the result from step 24 to the reading from step 10 and enter that result in Column 2 of Table 2-6 (Option 001 only: Column 5) as the 4 MHz error (relative to 300 MHz). 75 input only: Starting with the error at 41 MHz, add Column 3 (System Error) to Column 2 (Error Relative to 300 MHz) and record the result in Column 5 (Corrected Error Relative to 300 MHz).
75 input only: AMPLITUDE, More 1 of 2, AMPTD UNITS, dBm 27.To access the atness correction menu, press the following analyzer keys.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL FLATNESS DATA
Perform the next step only if all the atness correction data must be replaced in memory due to the repair or replacement of the A16 processor/video assembly.
102
Chapter 2
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
28.To initialize the area of memory where the atness correction data is stored, press the following keys.
INIT FLT PRESET FREQUENCY, 2001, Hz
29.To enter atness corrections, press EDIT FLATNESS. 30.The frequency of the rst data point, 4.00 MHz, will be displayed in the active function block of the analyzer display. 31.Use the data keys on the analyzer to enter the amplitude value for 4 MHz from Column 2 of Table 2-6 (75 input: Column 5), Frequency Response Errors. Terminate the entry with the dB key. When entering negative amplitude values, precede the numeric entry with the and dB keys or the dB key. The BK SP (backspace) key may be used to correct any entry if the terminator, dB or dB key has not been pressed. Re-enter the data if the terminator has been pressed. 32.Press (step-up key) and enter the data from Table 2-6, Column 2 (75 input: Column 5) for the next data point as described in step 30. 33.Repeat step 30 for the remaining atness correction data points listed in Table 2-6. At each point, verify that the frequency listed in the active function block corresponds to the frequency at which the data was taken. If these two frequencies do not correspond, press (step up) or (step down) until the proper frequency is displayed in the active function block. If some data is incorrect after entering all of the data from Table 2-6, select the incorrect data point using (step up) or (step down) and re-enter the proper data. 34.After all corrections have been input, press the STORE FLATNESS softkey to store the correction data in nonvolatile memory. The instrument will automatically preset and display CAL: DONE in the active function block of the analyzer.
Chapter 2
103
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
Table 2-6
Column 1 Frequency (MHz)
4 41 78 115 152 189 226 263 300 337 374 411 448 485 522 559 596 633 670 707 744 781 818 855 892
_______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________
N/A 0.03 0.1 0.1 0.1 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0
N/A _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________
_______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________
104
Chapter 2
Making Adjustments 15. Frequency Response of the 8590L, 8591C, and 8591E
Table 2-6
Column 1 Frequency (MHz)
929 966 1003 1040 1077 1114 1151 1188 1225 1262 1299 1336 1373 1410 1447 1484 1521 1558 1595 1632 1669 1706 1743 1780 1817
Chapter 2
105
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Analyzer Model 8592L 8592L (Option 026) (Option 027) 8593E 8593E (Option 026) (Option 027) 8594E 8594L 8594Q 8595E 8596E
The frequency response (atness) of the analyzer is measured with corrections off. The source amplitude is adjusted for a marker amplitude reading which is dependent upon the harmonic number of the band being measured. This is necessary since the source and power meter combination does not have sufcient dynamic range to measure the uncorrected frequency response. The difference between the target marker amplitudes is taken into account when calculating the uncorrected atness. The atness data is then entered into the analyzer using the SERVICE CAL functions. The error corrections are stored in battery backed RAM on the A16 processor/video assembly.
106
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Equipment Required
Synthesized sweeper Measuring receiver (used as a power meter) Power sensor, 50 MHz to 26.5 GHz Power sensor, 1 MHz to 2.9 GHz Power splitter Adapter, Type N (m) to APC 3.5 (m) Adapter, Type N (f) to APC 3.5 (m) Adapter, APC 3.5 (f) to APC 3.5 (f) Cable, APC 3.5, 91 cm (36 in.)
Procedure
Measuring Uncorrected Flatness
1. Zero and calibrate the measuring receiver and 1 MHz to 2.9 GHz power sensor in log mode as described in the measuring receiver operation manual. 2. Connect the equipment as shown in Figure 2-30. Figure 2-30 Frequency Response Setup
Chapter 2
107
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
3. Press INSTRUMENT PRESET on the synthesized sweeper. Set the synthesized sweeper controls as follows: CW ................................................................ 300 MHz FREQ STEP ................................................... 72 MHz POWER LEVEL ............................................. 3 dBm 4. To enter band 0, press the following analyzer keys.
PRESET CAL, More 1 of 4 CORRECT ON OFF (OFF) FREQUENCY, BAND LOCK, 0 - 2.9 Gz BAND 0
(Skip these key presses for the 8594E, 8594L and 8594Q analyzers.) 5. Set the analyzer controls by pressing the following keys.
FREQUENCY, 300, MHz CF STEP AUTO MAN (MAN), 72, MHz SPAN, 10, MHz AMPLITUDE, 0, dBm SCALE LOG LIN (LOG), 10, dB BW, 1, MHz PEAK SEARCH MKR FCTN, MKR TRACK ON OFF (ON)
6. Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 9 dBm 0.1 dB on the analyzer. 7. Press RATIO on the measuring receiver.
Band 0
8. Set the synthesized sweeper CW to 12 MHz. 9. Press the following analyzer keys.
FREQUENCY, 12, MHz
10.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 9 dBm 0.1 dB on the analyzer. 11.Record the power ratio displayed on the measuring receiver in Column 2 of Table 2-8. 12.On the synthesized sweeper, press CW and (step-up key).
108
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
14.Step through the remaining frequencies listed in Table 2-8. At each new frequency, repeat step 10 to step 13 and enter the appropriate power sensor cal factor into the measuring receiver as listed in Column 3 of Table 2-8. Adjustment for the 8594E and 8594L spectrum analyzer and 8594Q QAM analyzer is complete. Continue with Entering Flatness Correction Data at the end of this adjustment. Continue with the next step for all other analyzers.
Band 1
15.Replace the 1 MHz to 2.9 GHz power sensor with the 50 MHz to 26.5 GHz power sensor. 16.Calibrate the measuring receiver and 50 MHz to 26.5 GHz power sensor combination in log mode. 17.Set the synthesized sweeper CW to 300 MHz. 18.Press the following analyzer keys.
FREQUENCY, 300, MHz
19.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 9 dBm 0.1 dB on the analyzer. 20.Press RATIO on the measuring receiver. 21.To enter band 1, press the following analyzer keys.
FREQUENCY, BAND LOCK, 2.75 - 6.4 BAND 1
23.Set the synthesized sweeper CW to 2.75 GHz and FREQ STEP to 234.3 MHz. 24.Press the following analyzer keys.
PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) AMPLITUDE, PRESEL PEAK
Chapter 2
109
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
25.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 9 dBm 0.1 dB on the analyzer. 26.Record the power ratio displayed on the measuring receiver in Column 2 of Table 2-9. 27.On the synthesized sweeper, press CW and STEP UP. 28.On the analyzer, press the following keys.
FREQUENCY,
29.Step through the remaining frequencies listed in Table 2-9. At each new frequency, repeat step 24 to step 28 and enter the appropriate power sensor cal factor into the measuring receiver as listed in Column 3 of Table 2-9.Adjustment for the 8595E spectrum analyzer is complete. Continue with Entering Flatness Correction Data at the end of the adjustment. Continue with the next step for all other analyzers.
Band 2
30.To enter band 2, press the following analyzer keys.
FREQUENCY, BAND LOCK, 6.0-12.8 BAND 2
32.Set the synthesized sweeper CW to 6.2 GHz and FREQ STEP to 184 MHz. 33.Press the following analyzer keys.
PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) AMPLITUDE, PRESEL PEAK
34.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 15 dBm 0.1 dB on the analyzer. 35.Add 6 dB to the power ratio displayed on the measuring receiver and record the result in Table 2-11. For example, if the power ratio displayed on the measuring receiver is +2.8 dB, enter 8.8 dB (Error = 2.8 dB + 6.0 dB = 8.8 dB). The 6 dB correction comes from the difference between the MKR-TRK amplitudes set in step 6 (9 dBm) and step 34 (15 dBm).
110
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
36.On the synthesized sweeper, press CW and STEP UP. 37.Press the following analyzer keys.
FREQUENCY,
38.Step through the remaining frequencies listed in Table 2-10. At each new frequency, repeat step 33 to step 37 and enter the appropriate power sensor cal factor into the measuring receiver as listed in Column 3 of Table 2-10. Adjustment for the 8596E spectrum analyzer is complete. Continue with Entering Flatness Correction Data at the end of this adjustment. Continue with the next step for all other analyzers.
Band 3
39.To enter band 3, press the following analyzer keys.
FREQUENCY, BAND LOCK, 12.4-19. BAND 3
41.Set the synthesized sweeper CW to 12.45 GHz and FREQ STEP to 230 MHz. 42.Press the following analyzer keys.
PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) AMPLITUDE, PRESEL PEAK
43.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 20 dBm 0.1 dB on the analyzer. 44.Add 11 dB to the power ratio displayed on the measuring receiver and record the result in Table 2-11. For example, if the power ratio displayed on the measuring receiver is 6.7 dB, enter 17.7 dB (Error = 6.7 dB + 11.0 dB = 17.7 dB). The 11 dB correction comes from the difference between the MKR-TRK amplitudes set in step 6 (9 dBm) and step 43 (20 dBm). 45.On the synthesized sweeper, press CW and STEP UP.
Chapter 2
111
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
47.Step through the remaining frequencies listed in Table 2-10. At each new frequency, repeat step 42 to step 46 and enter the appropriate power sensor cal factor into the measuring receiver as listed in Column 3 of Table 2-11.
Band 4
48.To enter band 4, press the following analyzer keys.
MKR, MARKERS OFF SPAN, BAND LOCK, 19.1-22 BAND 4
Option 026 only: 19.1-26.5 BAND 4 49.Set the controls on the analyzer by pressing the following keys.
FREQUENCY, 19.15, GHz CF STEP AUTO MAN (MAN), 150, MHz
50.Set the synthesized sweeper CW to 19.15 GHz and FREQ STEP to 150 MHz (Option 026: 148 MHz). 51.Press the following analyzer keys.
PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) AMPLITUDE, PRESEL PEAK
52.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK amplitude reading of 34 dBm 0.1 dB on the analyzer. 53.Add 25 dB to the power ratio displayed on the measuring receiver and record the result in Table 2-12 (Option 026: Table 2-13). For example, if the power ratio displayed on the measuring receiver is 3.4 dB, enter 21.6 dB (Error = 3.4 dB + 25.0 dB = 21.6 dB). The 25 dB correction comes from the difference between the MKR-TRK amplitudes set in step 6 (9 dBm) and step 52 (34 dBm). 54.On the synthesized sweeper, press CW and STEP UP.
112
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
56.Step through the remaining frequencies listed in Table 2-12 (Option 026: Table 2-13). At each new frequency, repeat step 51 to step 55 and enter the appropriate power sensor cal factor into the measuring receiver as listed in Column 2 of Table 2-12 (Option 026: Table 2-13). Adjustment for the 8592L and 8593E spectrum analyzers is complete. Continue with the next step.
58.To access the atness correction menu, press the following analyzer keys.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL FLATNESS DATA CAUTION
The next step will erase all current atness correction. This step is to be performed only if the A16 processor/video assembly has been repaired or replace. 59.Initialize the area of memory where the atness correction data is stored, by pressing the following keys.
INIT FLT 22 GHz For Option 026, 027 only: Use INIT FLT 26.5 GHz PRESET FREQUENCY, 2001, Hz CAL, More 1 of 4, More 2 of 4 SERVICE CAL FLATNESS DATA
60.Enter the atness corrections, by pressing EDIT FLATNESS. 61.The frequency of the rst data point, 12.00 MHz, will be displayed in the active function block of the analyzer display.
Chapter 2
113
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
62.Use the data keys on the analyzer to enter the amplitude value for 12 MHz from Column 2 of Table 2-8, Frequency Response Errors. Terminate the entry with the dB key. When entering negative amplitude values, precede the numeric entry with the and dB keys or the dB key. Note that the BK SP (backspace) key may be used to correct any entry if the terminator, dB or dB key has not been pressed. Re-enter the data if the terminator has been pressed. 63.Press (step-up key) and enter the data from Column 2 of the Frequency Response Errors tables for the next data point as described in step 61. 64.Repeat step 62 for the remaining atness correction data points listed in the Frequency Response Errors tables. At each point, verify that the frequency listed in the active function block corresponds to the frequency at which the data was taken. If these two frequencies do not correspond, press (step up) or (step down) until the proper frequency is displayed in the active function block. If some data is incorrect after entering all of the data from the Frequency Response Errors tables, select the incorrect data point using (step up) or (step down) and re-enter the proper data. 65.After all corrections have been input, press STORE FLATNESS to store the correction data in nonvolatile memory. The instrument will automatically preset and display CAL: DONE in the active function block of the analyzer.
114
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 0 0.012 0.084 0.156 0.228 0.300 0.372 0.444 0.516 0.588 0.660 0.732 0.804 0.876 0.948 1.020 1.092 1.164 1.236 1.308 1.380 1.452 1.524 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 2.0 2.0 2.0 2.0 2.0 2.0 2.0
Chapter 2
115
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 0 1.596 1.668 1.740 1.812 1.884 1.956 2.028 2.100 2.172 2.244 2.316 2.388 2.460 2.532 2.604 2.676 2.748 2.820 2.892 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ Frequency Response Errors Band 1 6.100 6.284 6.468 6.652 6.836 7.020 7.204 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 6.0 6.0 6.0 7.0 7.0 7.0 7.0 2.0 2.0 2.0 2.0 2.0 2.0 2.0 2.0 2.0 2.0 2.0 2.0 2.0 3.0 3.0 3.0 3.0 3.0 3.0
116
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 1 7.388 7.572 7.756 7.940 8.124 8.308 8.492 8.676 8.860 9.044 9.228 9.412 9.596 9.780 9.964 10.148 10.332 10.516 10.700 10.884 11.068 11.252 11.436 11.620 11.804 11.988 12.172 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 7.0 8.0 8.0 8.0 8.0 8.0 8.0 9.0 9.0 9.0 9.0 9.0 10.0 10.0 10.0 10.0 10.0 11.0 11.0 11.0 11.0 11.0 11.0 12.0 12.0 12.0 12.0
Chapter 2
117
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 1 12.356 12.540 12.724 _____________________ _____________________ _____________________ Frequency Response Errors Band 2 6.100 6.284 6.468 6.652 6.836 7.020 7.204 7.388 7.572 7.756 7.940 8.124 8.308 8.492 8.676 8.860 9.044 9.228 9.412 9.596 9.780 9.964 10.148 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 6.0 6.0 6.0 7.0 7.0 7.0 7.0 7.0 8.0 8.0 8.0 8.0 8.0 8.0 9.0 9.0 9.0 9.0 9.0 10.0 10.0 10.0 10.0 12.0 13.0 13.0
118
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 2 10.332 10.516 10.700 10.884 11.068 11.252 11.436 11.620 11.804 11.988 12.172 12.356 12.540 12.724 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ Frequency Response Errors Band 3 12.450 12.680 12.910 13.140 13.370 13.600 13.830 14.060 14.290 14.520 14.750 14.980 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 12.0 13.0 13.0 13.0 13.0 14.0 14.0 14.0 14.0 15.0 15.0 15.0 10.0 11.0 11.0 11.0 11.0 11.0 11.0 12.0 12.0 12.0 12.0 12.0 13.0 13.0
Chapter 2
119
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 3 15.210 15.440 15.670 15.900 16.130 16.360 16.590 16.820 17.050 17.280 17.510 17.740 17.970 18.200 18.430 18.660 18.890 19.120 19.350 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ Frequency Response Errors Band 4 19.150 19.300 19.450 19.600 19.750 19.900 20.050 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 19.0 19.0 19.0 20.0 20.0 20.0 20.0 15.0 15.0 16.0 16.0 16.0 16.0 17.0 17.0 17.0 17.0 18.0 18.0 18.0 18.0 18.0 19.0 19.0 19.0 19.0
120
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 4 20.200 20.350 20.500 20.650 20.800 20.950 21.100 21.250 21.400 21.550 21.700 21.850 22.000 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 20.0 20.0 20.0 21.0 21.0 21.0 21.0 21.0 21.0 22.0 22.0 22.0 22.0
Frequency Response Errors Band 4 (Option 026) 19.100 19.248 19.396 19.544 19.692 19.840 19.988 20.136 20.284 20.432 20.580 20.728 20.876 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 19.0 19.0 19.0 20.0 20.0 20.0 20.0 20.0 20.0 20.0 21.0 21.0 21.0
Chapter 2
121
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 4 (Option 026) 21.024 21.172 21.320 21.468 21.616 21.764 21.912 22.060 22.208 22.356 22.504 22.652 22.800 22.948 23.096 23.244 23.392 23.540 23.688 23.836 23.984 24.132 24.280 24.428 24.576 24.724 24.872 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 21.0 21.0 21.0 21.0 22.0 22.0 22.0 22.0 22.0 22.0 23.0 23.0 23.0 23.0 23.0 23.0 23.0 24.0 24.0 24.0 24.0 24.0 24.0 24.0 25.0 25.0 25.0
122
Chapter 2
Making Adjustments 16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
Table 2-8
Frequency Response Errors Band 4 (Option 026) 25.020 25.168 25.316 25.464 25.612 25.760 25.908 26.056 26.204 26.352 26.500 _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ _____________________ 25.0 25.0 25.0 25.0 26.0 26.0 26.0 26.0 26.0 26.5 26.5
Chapter 2
123
Procedure
1. To turn the time and date ON or OFF, press the following analyzer keys.
CONFIG TIMEDATE TIMEDATE ON OFF (ON) or (OFF) as desired
The time and date will be displayed in the top-left corner with ON underlined. 2. The time and date may be displayed as month, day, and year (MDY) or as day, month, and year (DMY). To change the display, press the following analyzer keys.
CONFIG TIMEDATE DATEMODE MDY DMY (MDY) or (DMY) as desired
The active function block of the analyzer will display YYMMDD (year, month, and day). Use the data keys on the analyzer to enter the correct date as YYMMDD. Terminate the entry with one of the ENTER data keys.
124
Chapter 2
The active function block of the analyzer will display HHMMSS (hours, minutes, and seconds). Use the data keys on the analyzer to enter the correct time as HHMMSS. Terminate the entry with an ENTER data key.
Chapter 2
125
Making Adjustments 18. Modulator Offset and Gain for Option 010 or 011
Equipment Required
Digital multimeter (DMM) DMM test leads Adapter, Type N (m) to BNC (f) (two required) Cable, BNC, 50
Procedure
1. Switch the analyzer on and let it warm up 30 minutes. 2. Connect the BNC cable to the analyzer CAL OUT to the RF INPUT connector.
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 3. Press PRESET, then perform the CAL AMPTD and CAL FREQ adjustment routines. Refer to CAL AMPTD Adjustment Routine and CAL FREQ Adjustment Routine procedures. It is normal for FREQ UNCAL to appear on the display. This will disappear after the CAL routines are nished.
126
Chapter 2
Making Adjustments 18. Modulator Offset and Gain for Option 010 or 011
Note that if the display shows CAL SIGNAL NOT FOUND, the routines default settings are not sufcient to nd the signal. Press FREQ, 37, Hz before performing the CAL routines. This causes the routines to bypass the CAL signal check. 4. Press CAL STORE on the analyzer.
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 5. Connect the analyzer RF OUT to the RF INPUT using a BNC cable and adapters. 6. Perform the tracking generator self-cal routine by pressing the following analyzer keys.
CAL, More 1 of 4, More 2 of 4, CAL TRK GEN
Wait for the completion of the routine (less than 1 minute). 7. Press CAL STORE. 8. Press PRESET, then wait for preset routine to nish. Set the analyzer by pressing the following keys.
FREQUENCY, 300, MHz SPAN, 0, Hz
75 only: Press AMPLITUDE, More 1 of 2, AMPTD UNITS, dBm. Set the analyzer reference level to +12 dBm. 9. Press the following analyzer keys.
AUX CTRL, TRACK GENSRC PWR ON OFF (ON) CAL, More 1 of 4, More 2 of 4 SERVICE DIAG, More 1, More 2, More 3, More 4, More 5 ALC TEST
This breaks the ALC loop, causing TG UNLVL to appear on the screen; this is normal. 10.Connect the ground () lead of the DMM to A7TP1. Connect the positive (+) lead to A7TP2. 11.Adjust A7A1R43 and A7A1R49 fully clockwise. 12.Connect the MOD_TP test point, located on the A7A1 tracking generator control assembly, to A7TP2 by pressing DISPLAY, CHANGE TITLE, then type in the following: DET TGMOD;
Chapter 2
127
Making Adjustments 18. Modulator Offset and Gain for Option 010 or 011
13.Press the following analyzer keys to execute the screen title command.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL EXECUTE TITLE
Note that TGMOD should appear in the upper-left screen annotation below REF. 14.Press the following analyzer keys.
AMPLITUDE, SCALE LOG LIN (LOG), 1, dB AUX CTRL, TRACK GEN, SRC PWR ON OFF (ON)
15.Adjust the front-panel knob for 0.0 V 0.01 V on the DMM, then press the following keys.
TRACE,CLEAR WRITE A, More 1 of 3, DETECTOR SAMPL PK (PK) MKR, MARKER
16.Adjust A7A1R49 so that the MKR reading drops approximately 0.1 to 0.15 dB. 17.Press the following analyzer keys.
MKR, MARKER , MARKER CAL, More 1 of 4, More 2 of 4, SERVICE CAL, EXECUTE TITLE AMPLITUDE, SCALE LOG LIN (LOG) 10 dB AUX CTRL, TRACK GEN, SRC PWR ON OFF (ON)
18.Adjust the front-panel knob for 1.0 V 0.01 V measured with the DMM, then press the following analyzer keys.
TRACE, CLEAR WRITE A, More 1 OF 4, DETECTOR SAMPL PK (PK)
19.Adjust A7A1R43 so that the MARKER DELTA reading is 33 dB 0.5 dB. 20.Press PRESET, then wait for the preset routine to nish. 21.Perform the tracking generator self-cal routine by pressing the following analyzer keys.
CAL, More 1 of 4, More 2 of 4, CAL TRK GEN
Wait for the completion of the routine (less than 1 minute). 22.Press CAL STORE.
128
Chapter 2
Making Adjustments 19. Entering External ALC Correction Constants for Option 010 or 011
19. Entering External ALC Correction Constants for Option 010 or 011
This adjustment applies to: 8590L spectrum analyzers, Option 010 or 011 8591E spectrum analyzers, Option 010 or 011 8591C cable TV analyzers, Option 011 External automatic level control (ALC) operation is corrected by entering the correction constants (CALTGX slope and offset) into analyzer memory using the EXECUTE TITLE function. Correction constants that pertain to each tracking generator are developed at the factory. The two constants are recorded on a label which is located on the A7A1 tracking generator control board assembly.
Equipment Required
No equipment is required for this procedure.
Procedure
1. Record the CAL TGX slope and offset numbers (located on the A7A1 assembly) below. Slope =____________________ Offset =____________________ 2. Press the following analyzer keys: DISPLAY, CHANGE TITLE, More 1 of 2. 3. Enter the CALTGX slope and offset correction constants in the format CALTGX slope, offset;. For example, a slope of 0.3079 and an offset of 3095 would be entered as: CALTGX 0.3079,3095;
Chapter 2
129
Making Adjustments 19. Entering External ALC Correction Constants for Option 010 or 011
Verify that the XSLOP and XOFST values are the same as you entered in step 3. 5. Perform the Checking the External ALC for Option 010 or 011" procedure.
130
Chapter 2
Making Adjustments 20. Checking the External ALC for Option 010 or 011
Equipment Required
Digital multimeter (DMM) Power meter Power sensor, 1 MHz to 350 MHz Cable, BNC (m) to dual banana plug Cable, BNC, 122 cm (48 in) Cable, BNC, 23 cm (9 in) Adapter, BNC Tee Adapter, Type N (m) to BNC (f) (two required)
Chapter 2
131
Making Adjustments 20. Checking the External ALC for Option 010 or 011
Procedure
1. Switch the analyzer on and let it warm up 30 minutes. 2. Connect the analyzer CAL OUT to the RF INPUT.
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 3. Press PRESET, then perform the CAL FREQ and CAL AMPTD adjustment routines. Refer to CAL FREQ Adjustment Routine and CAL AMPTD Adjustment Routine procedures. It is normal for FREQ UNCAL to appear on the display. This will disappear after the CAL routines are nished. Note that if the display shows CAL SIGNAL NOT FOUND, the routines default settings are not sufcient to nd the signal. Press FREQ, 37, Hz before performing the CAL routines. This causes the routines to bypass the CAL signal check. 4. Press CAL STORE on the analyzer. 5. Connect a BNC cable from the RF INPUT to the RF OUTPUT. 6. Press PRESET, then wait for preset routine to nish. 7. Perform the tracking generator self-cal routine by pressing the following analyzer keys.
CAL, More 1 of 4, More 2 of 4, CAL TRK GEN
Wait for the completion of the routine (less than 1 minute). 8. Press CAL STORE. 9. Zero and calibrate the power meter and power sensor as described in the power meter operation manual.
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur.
132
Chapter 2
Making Adjustments 20. Checking the External ALC for Option 010 or 011
Figure 2-31
10.Connect the equipment as shown in Figure 2-31. For Option 011 only: Use the 75 power sensor with an adapter, instead of the 50 power sensor. 75 input only: Press AMPLITUDE, More 1 of 2, AMPTD UNITS, dBm. 11.Set the DMM to measure dc volts. 12.Set the analyzer by pressing the following keys.
FREQUENCY, 300, MHz SPAN, 0, Hz AUX CTRL, TRACK GEN, SRC PWR ON OFF (ON), 2, dBm.
13.Ensure that the power meter reads 2 2 dB, then press RANGE HOLD on the power meter. 14.On the analyzer, press More 1 OF 2, ALC MTR INT XTAL (MTR). 15.Adjust the SRC PWR for 0 dBV by turning the front-panel knob on the analyzer. Note that the display will read the power dBm. 16.Note the voltage on the DMM. This is the error at source power 0 dBV. Record the automatic leveling control (ALC) voltage in the space provided. ALC voltage at 0 dBV # = ______________volts dc
Chapter 2
133
Making Adjustments 20. Checking the External ALC for Option 010 or 011
17.Convert the ALC voltage recorded in the previous step to dBv using the following equation. ALC Voltage in dBv = ______________20 log (V) The calculated ALC voltage, in dBV, should be 0.75 dBV and +0.75 dBV. 18.Adjust the SRC PWR to 34 dBV using the front-panel knob. (Display reads 34 dBm.) 19.Note the voltage on the DMM. Wait until the reading stabilizes before recording it in the space provided. This is the error at a source power of 34 dBV. ALC voltage at -34 dBV = ______________volts dc 20.Convert the ALC voltage recorded in the previous step using the following equation.
134
Chapter 2
Making Adjustments 21. Correcting for External ALC Error for Option 010 or 011
21. Correcting for External ALC Error for Option 010 or 011
This adjustment applies to: 8590L spectrum analyzers, Option 010 or 011 8591E spectrum analyzers, Option 010 or 011 8591C cable TV analyzers, Option 011 External automatic level control (ALC) operation is corrected by developing two correction constants (CALTGX Slope and Offset). This procedure is only necessary if the Checking the External ALC for Option 010 or 011" procedure does not pass, or if the factory correction constants are not available.
Equipment Required
Digital multimeter (DMM) Power meter Power sensor, 1 MHz to 350 MHz Cable, BNC (m) to dual banana plug Adapter, BNC Tee Adapter, Type N (m) to BNC (f) (two required) Cable, BNC, 122 cm (48 in) Cable, BNC, 32 cm (9 in)
Chapter 2
135
Making Adjustments 21. Correcting for External ALC Error for Option 010 or 011
Procedure
1. Connect a BNC cable from the RF INPUT to the RF OUTPUT.
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 2. Press PRESET, then wait for preset routine to nish. 3. Perform the tracking generator self-cal routine by pressing the following analyzer keys.
CAL, More 1 of 4, More 2 of 4, CAL TRK GEN
Wait for the completion of the routine (less than 1 minute). 4. Press CAL STORE. 5. Zero and calibrate the power meter and power sensor as described in the power meter operation manual.
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. External ALC Error Correction Adjustment
Figure 2-32
136
Chapter 2
Making Adjustments 21. Correcting for External ALC Error for Option 010 or 011
6. Connect the equipment as shown in Figure 2-32. For Option 011 only: Use the 75 power sensor with an adapter, instead of the 50 power sensor. Option 011 only: Press AMPLITUDE, More 1 of 2, AMPTD UNITS, dBm. 7. Set the DMM to measure dc volts.
CAUTION
The default values must be entered rst for the correction procedure to be successful. 8. Load the default correction constants for Slope and Offset into nonvolatile memory by performing the following steps. a. Press the following analyzer keys.
DISPLAY, CHANGE TITLE
b. Enter the following screen title. CALTGX 0.3079,3095; c. Press the following analyzer keys.
CAL, More 1 OF 4, More 2 OF 4, SERVICE CAL, EXECUTE TITLE CAL, More 1 OF 4, More 2 OF 4, SERVICE DIAG, NEXT PAGE
Verify that the XSLOP and XOFST values are the same as you entered in the previous step. 9. Set the analyzer by pressing the following keys.
FREQUENCY, 300, MHz SPAN, 0, Hz AUX CTRL, TRACK GEN, SRC PWR ON OFF (ON), 2, dBm.
10.Ensure that the power meter reads 2 2 dBm. Press RANGE HOLD on the 436A Power Meter. 11.On the analyzer, press More 1 OF 2, ALC MTR INT XTAL (until MTR is underlined). 12.Adjust the SRC PWR for 0 dBV as read on the analyzer display. Note that the display will read the power in dBm. 13.Note the voltage on the DMM. Wait until the reading stabilizes before recording it in the space provided. This is the ALC voltage at a source power of 0 dBV (VHigh). VHigh = ______________volts dc
Chapter 2
137
Making Adjustments 21. Correcting for External ALC Error for Option 010 or 011
14.Adjust the SRC PWR for 34 dBV as read on the analyzer display. 15.Note the voltage on the DMM. Wait until the reading stabilizes before recording it in the space provided. This is the ALC voltage at a source power of 34 dBV (VLow). VLow = ______________volts dc 16.Calculate the actual slope from the two voltage measurements using the following equation.
138
Chapter 2
Making Adjustments 22. First LO Distribution Amplier for Option 009 or 010
Equipment Required
Measuring receiver Digital multimeter Power sensor DMM test leads Adapter, dual banana plug
Procedure
1. Set the analyzer LINE switch to off, then disconnect the line cord. Remove the cover assembly, then reconnect the line cord. 2. Remove the 50 termination from the analyzer rear-panel LO OUTPUT connector. 3. Connect the positive DMM test lead to A10TP5, GB (gate bias). Connect the negative DMM test lead to A10TP2, AGND (analog ground). See Figure 2-33.
Chapter 2
139
Making Adjustments 22. First LO Distribution Amplier for Option 009 or 010
Figure 2-33
4. Set the digital voltmeter controls as follows: FUNCTION .............................................. DC VOLTS RANGE ................................................................ 10 V RESOLUTION .................................................... 1 mV 5. Set the analyzer LINE switch to on. 6. Adjust A10R29 (GATE) for a digital voltmeter reading within 5 mV of the GATE (gate bias) voltage printed on the RF section label. 7. Zero and calibrate the measuring receiver and power sensor in LOG mode. (Power levels read in dBm.) Enter the power sensor's 5 GHz cal factor into the measuring receiver. 8. Connect the power sensor to the analyzer LO OUTPUT. 9. On the analyzer, press PRESET, SPAN, ZERO SPAN, FREQUENCY, 300, MHz. 10.Connect the positive DMM test lead to A10TP4, LOS (LO sense). 11.Note the SENS (LO sense) voltage printed on the RF section label. Adjust A10R25, LO AMP (LO power), until the DMM reads equal to the SENS voltage printed on the RF section label.
140
Chapter 2
Making Adjustments 22. First LO Distribution Amplier for Option 009 or 010
12.Check that the measuring receiver power level reads greater than +12 dBm. 13.Disconnect the power sensor from LO OUTPUT, then reconnect the 50 termination to LO OUTPUT. 14.Disconnect the DMM leads from A10TP4 and A10TP2.
Chapter 2
141
Equipment Required
Measuring receiver Power sensor, 1 MHz to 350 MHz Cable, Type N, 62 cm (24 in.)
142
Chapter 2
Procedure
1. Set the analyzer LINE switch to off. Disconnect the line cord. Remove the cover assembly, then reconnect the line cord. 2. Set the analyzer AOFST by pressing the following keys.
PRESET FREQUENCY, 2001, Hz CAL, More 1 of 4, More 2 of 4, DEFAULT CAL DATA CAL, More 1 of 4, More 2 of 4 SERV DIAG, DISPLAY CAL DATA, NEXT PAGE
Verify that AOFST=0 under the tracking generator readouts. 3. Connect the cable between the RF OUT 50 and INPUT 50 connectors on the analyzer. 4. Press PRESET on the analyzer and set the controls as follows: CENTER FREQ ........................................... 300 MHz SPAN .................................................................... 0 Hz 5. On the analyzer, press the following keys.
BW, 10, kHz AUX CTRL, TRACK GEN, SRC PWR ON OFF (ON), 10, dBm
6. On the analyzer, press TRACKING PEAK. Wait for the PEAKING message to disappear. 7. Zero and calibrate the measuring-receiver/power-sensor combination in log mode (power levels readout in dBm). Enter the power sensor 300 MHz cal factor into the measuring receiver. 8. Disconnect the cable from the RF OUT 50 connector, then connect the power sensor to the RF OUT 50 connector. See Figure 2-34. Figure 2-34 BITG Power Level Adjustment Setup
Chapter 2
143
9. On the analyzer, press SRC PWR ON OFF (ON), 0, dBm, SGL SWP. Note that some analyzers may have sealing compound over A3A15R13 (10 dB ADJ) and A3A15R18 (0 dB ADJ) adjustments. Remove this compound before making these adjustments. 10.Adjust 10 dB ADJ (A3A15R13) for a 0 dBm 0.05 dB reading on the measuring receiver. Refer to Figure 2-35 for adjustment location. Figure 2-35 BITG Power Level Adjustment Locations
11.Set the SRC PWR level to 10 dBm. Note the power displayed on the measuring receiver. If the power level is -9.77 dBm to -10.23 dBm, then the adjustment is complete. If the power level is not within the range, then continue with step 11. Power at 10 dBm Setting ______________dBm If the power level noted in step 10 was outside the range of 10 dBm 0.23 dB, perform the following: a. With the SRC PWR level set to 10 dBm, adjust 10 dB ADJ (A3A15R13) for a 10 dBm 0.1 dB reading on the measuring receiver. Refer to Figure 2-35 for adjustment location. b. Set the SRC PWR level to 0 dBm. Adjust 0 dB ADJ (A3A15R18) for a 0 dBm 0.2 dB reading on the measuring receiver. Refer to Figure 2-35 for adjustment location. c. Repeat this step until the output power level is within the tolerances indicated at both the 10 dBm and 0 dBm SRC PWR level settings. Adjust 10 dB ADJ only with the SRC POWER level set to 10 dBm, and adjust 0 dB ADJ only with the SRC PWR level set to 0 dBm.
144
Chapter 2
Equipment Required
Microwave frequency counter Termination, 50 termination Alignment tool, non-metallic Cable, BNC, 122 cm (48 in.) (two required) Adapter, Type N (f) to APC-3.5 (f) Adapter, Type N (m) to BNC (f)
Chapter 2
145
Procedure
Frequency Tracking Range Check
1. Connect a cable between the RF OUT 50 and INPUT 50 connectors on the analyzer. Figure 2-36 Frequency Tracking Range Setup
2. Remove the rear-panel jumper that is between the 10 MHz REF OUTPUT and EXT REF IN jacks. Connect the frequency counter FREQ STD OUT connector to the analyzer EXT REF IN connector as shown in Figure 2-36. 3. Press PRESET on the analyzer, then set the controls as follows: CENTER FREQ ............................................ 500 MHz SPAN .....................................................................0 Hz 4. On the analyzer, press the following key.
BW, 10, kHz AUX CTRL, TRACK GEN, SRC PWR ON OFF (ON), 5, dBm
5. On the analyzer press TRACKING PEAK. Wait for the PEAKING message to disappear.
146
Chapter 2
6. Set the microwave frequency counter controls as follows: SAMPLE RATE ........................................... Midrange 10 Hz-500 MHz SWITCH ...........500 MHz - 26.5 GHz 500 MHz-26.5 GHz SWITCH .....500 MHz - 26.5 GHz RESOLUTION ......................................................1 Hz 7. Connect the RF OUT 50 connector to the microwave frequency counter input as shown in Figure 2-36. 8. Wait for the microwave frequency counter to gate two or three times, then record the microwave frequency counter reading below as the peaked frequency. Peaked Frequency: ______________MHz 9. On the analyzer, press MAN TRK ADJUST, 4095, ENTER. Wait for the microwave frequency counter to gate two or three times, then record the microwave frequency counter reading below as the minimum frequency. Minimum Frequency: ______________MHz 10.On the analyzer, press MAN TRK ADJUST, 0, ENTER. Wait for the microwave frequency counter to gate two or three times, then record the microwave frequency counter reading below as the maximum frequency. Maximum Frequency: ______________MHz 11.If the absolute value of the difference between either the minimum or maximum frequency and the peaked frequency is less than 5 kHz, proceed with the adjustment procedure below. If the differences are greater than 5 kHz, no adjustment is necessary. 12.Disconnect the cable from the EXT REF IN connector, then replace the rear-panel jumper.
Chapter 2
147
Figure 2-37
15.Set the analyzer LINE switch to on. Press AUX CTRL, TRACK GEN, SRC PWR ON OFF (ON). Allow the analyzer to warm up for at least 5 minutes. Set the controls as follows: CENTER FREQ ............................................ 300 MHz SPAN .....................................................................0 Hz 16.Set the microwave frequency counter controls as follows: SAMPLE RATE ......................................... Fully CCW 10 Hz-500 MHz SWITCH ................. 10 Hz-500 MHz 500 MHz-26.5 GHz SWITCH ............ 10 Hz-500 MHz 50 - 1 M SWITCH ..........................................50 17.Remove the screw, located on the front of the tracking generator, used to seal the tracking oscillator adjustment. 18.On the analyzer, press AUX CTRL, TRACK GEN, MAN TRK ADJUST, 0, Hz. 19.Record the microwave frequency counter reading in Table 2-9 as F1. 20.On the analyzer, press MAN TRK ADJUST, 4095, Hz. 21.Record the microwave frequency counter reading in Table 2-9 as F2.
148
Chapter 2
22.Calculate Fcenter as shown below, and record it in Table 2-9. Fcenter = (F1 + F2)/2
Chapter 2
149
23.Set SRC TRACK ADJ to 350. This sets the tracking oscillator near the center of its frequency range. (The relationship between the SRC TRACK ADJ DAC number and the output frequency is nonlinear.) Adjust SRC TRACK ADJ until the microwave frequency counter reads Fcenter 100 Hz. 24.Record the value of SRC TRACK ADJ in Table 2-9.
CAUTION
A3A15C3 (TRK OSC CTR) is rated for a maximum of 10 adjustment cycles. Due to this limitation, adjust TRK OSC CTR only when absolutely necessary. 25.Adjust A3A15C3 (TRK OSC CTR) until the microwave frequency counter reads 300 MHz 500 Hz. 26.Repeat step 17 to step 24 at least once more until no further adjustment of A3A15C3 is necessary. 27.Set the analyzer LINE switch to off, then replace the screw removed in step 17. 28.Reinstall the A3 RF Section assembly into the analyzer. 29.Replace the rear-panel jumper between the 10 MHz REF OUTPUT and EXT REF IN connectors.
Table 2-9
N 1 2 3 4 5 6
150
Chapter 2
Making Adjustments 25. Checking the Absolute Amplitude Accuracy for Option 050
Equipment required
Synthesized sweeper Measurement receiver Power splitter Power sensor, 818 MHz to 948 MHz Adapter, Type N (f) to APC 3.5 (m) Adapter, APC 3.5 (f) to APC 3.5 (f) Adapter, Type N (m) to Type N (m) Cable, Type N, 183 cm (72 in)
Chapter 2
151
Making Adjustments 25. Checking the Absolute Amplitude Accuracy for Option 050
Procedure
1. Zero and calibrate the measuring receiver and power sensor in log mode as described in the measuring receiver operation manual. Note that the absolute amplitude accuracy test should only be performed if the ambient temperature is between 20 C and 30 C. 2. Connect the equipment as shown in Figure 2-38. Connect the power splitter to the analyzer using an adapter. Figure 2-38 Absolute Amplitude Accuracy Verication
3. Press instrument preset on the synthesized sweeper, then set the controls as follows: CW ................................................................. 818 MHz POWER LEVEL ................................................... 2 dBm
152
Chapter 2
Making Adjustments 25. Checking the Absolute Amplitude Accuracy for Option 050
4. Press PRESET on the analyzer and wait for the preset to nish, then press the following analyzer keys.
FREQUENCY, 818, MHz SPAN, 400, kHz BW, 100, kHz VID BW AUTO MAN, 30, kHz AMPLITUDE, 4, dBm ATTEN AUTO MAN, 10, dB PEAK SEARCH
Log Fidelity
5. Set the power sensor cal factor for 818 MHz on the measuring receiver. 6. On the synthesized sweeper, press POWER LEVEL and adjust the output amplitude so that the analyzer marker amplitude reads 9 dBm 0.05 dB. 7. Record the measuring receiver power reading in Table 2-10. 8. Adjust the output amplitude of the sweeper for analyzer marker amplitude readings of 14 dBm and 19 dBm. 9. Record the measuring receiver power readings in Table 2-10. The readings should be within the limits shown. Table 2-10 Log Fidelity
Measuring Receiver Min(dBm) 9.7 14.7 19.7 Reading (dBm) _________________ _________________ _________________ Max(dBm) 8.3 13.3 21.3
Chapter 2
153
Making Adjustments 25. Checking the Absolute Amplitude Accuracy for Option 050
13.On the synthesized sweeper, press POWER LEVEL and adjust the output amplitude so the analyzer marker amplitude reads 9 dBm 0.05 dB. 14.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-11. 15.Repeat step 10 to step 14 for frequencies of 836 MHz, 881 MHz, and 948 MHz. Record the results in Table 2-11. The results should be within the limits shown. Table 2-11 Frequency Response Attenuator 10 dB
Measuring Receiver Min (dBm) 9.7 9.7 9.7 9.7 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) 8.3 8.3 8.3 8.3
17.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-12. 18.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 19.On the analyzer, press PEAK SEARCH. 20.On the sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +1 dBm 0.05 dB. 21.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-12. 22.Repeat step 17 to step 21 for frequencies of 881 MHz, 836 MHz, and 818 MHz. Record the results in Table 2-12. The results should be within the limits shown.
154
Chapter 2
Making Adjustments 25. Checking the Absolute Amplitude Accuracy for Option 050
Table 2-12
24.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-13. 25.On the analyzer, press PEAK SEARCH. 26.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 27.On the synthesized sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +5 dBm 0.05 dB. 28.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-13. 29.Repeat step 24 to step 28 for frequencies of 836 MHz, 881 MHz, and 948 MHz. Record the results in Table 2-13. Table 2-13 Frequency Response Attenuator 30 dB
Measuring Receiver Min (dBm) +4.3 +4.3 +4.3 +4.3 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) +5.7 +5.7 +5.7 +5.7
Chapter 2
155
Making Adjustments 25. Checking the Absolute Amplitude Accuracy for Option 050
31.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-14. 32.On the synthesized sweeper, press CW, then set the frequency to the same measurement frequency as the analyzer is set in the previous step. 33.On the analyzer, press PEAK SEARCH. 34.On the synthesized sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +5 dBm 0.05 dB. 35.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-14. 36.Repeat step 31 to step 35 for frequencies of 881 MHz, 836 MHz, and 818 MHz. Record the results in Table 2-14. Table 2-14 Frequency Response Attenuator 40 dB
Measuring Receiver Min (dBm) +4.0 +4.0 +4.0 +4.0 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) +6.0 +6.0 +6.0 +6.0
156
Chapter 2
Making Adjustments 26. Correcting for Absolute Amplitude Accuracy for Option 050
The next step disables the protection for the factory correction constants. 2. Press the following analyzer keys.
FREQUENCY, 2001, Hz CAL, More 1 of 4, More 2 of 4 Service Cal Flatness Data EDIT FLATNESS
3. Press (step up key) until the ERROR readout for 804 MHz is displayed (781 MHz for an 8591E spectrum analyzer or an 8591C cable TV analyzer).
Chapter 2
157
Making Adjustments 26. Correcting for Absolute Amplitude Accuracy for Option 050
4. Read the amplitude error from the display and record it in Table 2-15 for the 8593E, 8594E, 8595E, and 8596E spectrum analyzers. Use Table 2-16 for 8591E spectrum analyzers or 8591C cable TV analyzers.
Table 2-15
Table 2-16
Repeat step 3 and step 4 for the next frequency indicated in Table 2-15 or Table 2-16 until all four points have been recorded. 5. Subtract the Average Error, calculated in step 1, from each Amplitude ERROR from Table 2-15 or Table 2-16, then record column 3 as the Corrected Flatness ERROR. Do not subtract more than 0.5 dB from the Amplitude ERROR. If the Average Error is more than 0.5 dB, it may not be possible to make corrections during the adjustment. 6. Using (step down key), set the frequency displayed on the analyzer to 804 MHz (781 MHz for 8591E spectrum analyzers or 8591C cable TV analyzers).
158
Chapter 2
Making Adjustments 26. Correcting for Absolute Amplitude Accuracy for Option 050
7. Enter the Corrected Flatness ERROR from Table 2-15 or Table 2-16 for the frequency displayed using the DATA keys. Terminate the entry with the +dBm. After the new data is entered, the analyzer will automatically jump to the next frequency correction point. Enter the next Corrected Flatness ERROR from the table. To conrm the entries, press (step down key) to the frequency of interest. Re-check the displayed error against the Corrected Flatness ERROR from column 3. 8. Perform the frequency response verication test from the calibration guide for your instrument. Only checking Band 0 is necessary.
Chapter 2
159
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Equipment required
Synthesized sweeper Measurement receiver Power splitter Power sensor, 810 MHz to 1501 MHz Adapter, Type N (f) to APC 3.5 (m) Adapter, APC 3.5 (f) to APC 3.5 (f) Adapter, Type N (m) to Type N (m) Cable, Type N, 183 cm (72 in)
160
Chapter 2
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Figure 2-39
Procedure
1. Zero and calibrate the measuring receiver and power sensor in log mode as described in the measuring receiver operation manual. Note that the absolute amplitude accuracy test should only be performed if the ambient temperature is between 20 C and 30 C. 2. Connect the equipment as shown in Figure 2-39. Connect the power splitter to the analyzer using an adapter. 3. Press instrument preset on the synthesized sweeper, then set the controls as follows: CW ................................................................. 810 MHz POWER LEVEL ....................................................2 dBm 4. Press PRESET on the analyzer and wait for the preset to nish, then press the following analyzer keys.
FREQUENCY, 810, MHz SPAN, 400, kHz BW, 100, kHz VID BW AUTO MAN, 30, kHz AMPLITUDE, 4, dBm ATTEN AUTO MAN, 10, dB PEAK SEARCH
Chapter 2
161
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Log Fidelity
5. Set the power sensor cal factor for 810 MHz on the measuring receiver. 6. On the synthesized sweeper, press POWER LEVEL and adjust the output amplitude so that the analyzer marker amplitude reads 9 dBm 0.05 dB. 7. Record the measuring receiver power reading in Table 2-17. 8. Adjust the output amplitude of the sweeper for analyzer marker amplitude readings of 14 dBm and 19 dBm. 9. Record the measuring receiver power readings in Table 2-17. The readings should be within the limits shown. Table 2-17 Log Fidelity
Measuring Receiver Min (dBm) 9.7 14.7 19.7 Reading (dBm) _________________ _________________ _________________ Max (dBm) 8.3 13.3 21.3
162
Chapter 2
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Range 810 MHz to 956 MHz Frequency Response Input Attenuator 10 dB 10. Set the frequency of the analyzer to the rst measurement frequency shown in Table 2-18. 11.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 12.On the analyzer, press PEAK SEARCH. 13.On the synthesized sweeper, press POWER LEVEL and adjust the output amplitude so the analyzer marker amplitude reads 9 dBm 0.05 dB. 14.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-18. 15.Repeat step 10 to step 14 for frequencies of 826 MHz, 940 MHz, and 956 MHz. Record the results in Table 2-11. The results should be within the limits shown. Table 2-18 Frequency Response Attenuator 10 dB
Measuring Receiver Min (dBm) 9.6 9.6 9.6 9.6 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) 8.4 8.4 8.4 8.4
Chapter 2
163
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Response Input Attenuator 20 dB 16.On the analyzer, press the following keys.
AMPLITUDE, ATTEN AUTO MAN, 20, dB AMPLITUDE, 6, +dBm
17.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-19. 18.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 19.On the analyzer, press PEAK SEARCH. 20.On the sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +1 dBm 0.05 dB. 21.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-12. 22.Repeat step 17 to step 21 for frequencies of 948 MHz, 826 MHz, and 810 MHz. Record the results in Table 2-19. The results should be within the limits shown. Table 2-19 Frequency Response Attenuator 20 dB
Measuring Receiver Min (dBm) +0.4 +0.4 +0.4 +0.4 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) +1.6 +1.6 +1.6 +1.6
164
Chapter 2
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Response Input Attenuator 30 dB 23.On the analyzer, press the following keys.
AMPLITUDE, ATTEN AUTO MAN, 30, dB AMPLITUDE, 10, +dBm
24.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-20. 25.On the analyzer, press PEAK SEARCH. 26.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 27.On the synthesized sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +5 dBm 0.05 dB. 28.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-20. 29.Repeat step 24 to step 28 for frequencies of 826 MHz, 940 MHz, and 956 MHz. Record the results in Table 2-20. Table 2-20 Frequency Response Attenuator 30 dB
Measuring Receiver
Synthesized Sweeper Frequency (MHz) Min (dBm) 810 826 940 956 +4.4 +4.4 +4.4 +4.4
Chapter 2
165
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Response Input Attenuator 40 dB 30.On the analyzer, press the following keys.
AMPLITUDE, ATTEN AUTO MAN, 40, dB AMPLITUDE, 10, +dBm
31.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-21. 32.On the synthesized sweeper, press CW, then set the frequency to the same measurement frequency as the analyzer is set in the previous step. 33.On the analyzer, press PEAK SEARCH. 34.On the synthesized sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +5 dBm 0.05 dB. 35.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-21. 36.Repeat step 31 to step 35 for frequencies of 940 MHz, 826 MHz, and 810 MHz. Record the results in Table 2-21. Table 2-21 Frequency Response Attenuator 40 dB
Measuring Receiver Min (dBm) +4.0 +4.0 +4.0 +4.0 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) +6.0 +6.0 +6.0 +6.0
166
Chapter 2
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Range 1429 MHz to 1501 MHz Frequency Response input Attenuator 10 dB 37.Set the frequency of the analyzer to the rst measurement frequency shown in Table 2-22. 38.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 39.On the analyzer, press PEAK SEARCH. 40.On the synthesized sweeper, press POWER LEVEL and adjust the output amplitude so the analyzer marker amplitude reads 9 dBm 0.05 dB. 41.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-22. 42.Repeat step 37 to step 41 for frequencies of 1453 MHz, 1477 MHz, and 1501 MHz. Record the results in Table 2-22. The results should be within the limits shown. Table 2-22 Frequency Response Attenuator 10 dB
Measuring Receiver
Synthesized Sweeper Frequency (MHz) Min (dBm) 1429 1453 1477 1501 9.6 9.6 9.6 9.6
Chapter 2
167
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Response Input Attenuator 20 dB 43.On the analyzer, press the following keys.
AMPLITUDE, ATTEN AUTO MAN, 20, dB AMPLITUDE, 6, +dBm
44.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-23. 45.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 46.On the analyzer, press PEAK SEARCH. 47.On the sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +1 dBm 0.05 dB. 48.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-23. 49.Repeat step 44 to step 48 for frequencies of 1477 MHz, 1453 MHz, and 1428 MHz. Record the results in Table 2-23. The results should be within the limits shown. Table 2-23 Frequency Response Attenuator 20 dB
Measuring Receiver Min (dBm) +0.4 +0.4 +0.4 +0.4 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) +1.6 +1.6 +1.6 +1.6
168
Chapter 2
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Response Input Attenuator 30 dB 50.On the analyzer, press the following keys.
AMPLITUDE, ATTEN AUTO MAN, 30, dB AMPLITUDE, 10, +dBm
51.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-24. 52.On the analyzer, press PEAK SEARCH. 53.On the synthesized sweeper, press CW and set the frequency to the same measurement frequency as the analyzer is set in the previous step. 54.On the synthesized sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +5 dBm 0.05 dB. 55.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-24. 56.Repeat step 51 to step 55 for frequencies of 1453 MHz, 1477 MHz, and 1501 MHz. Record the results in Table 2-24. Table 2-24 Frequency Response Attenuator 30 dB
Measuring Receiver Min (dBm) +4.4 +4.4 +4.4 +4.4 Reading (dBm) _________________ _________________ _________________ _________________ Max (dBm) +5.6 +5.6 +5.6 +5.6
Chapter 2
169
Making Adjustments 27. Checking the Absolute Amplitude Accuracy for Option 051
Frequency Response Input Attenuator 40 dB 57.On the analyzer, press the following keys.
AMPLITUDE, ATTEN AUTO MAN, 40, dB AMPLITUDE, 10, +dBm
58.Set the FREQUENCY of the analyzer to the measurement frequency shown in Table 2-25. 59.On the synthesized sweeper, press CW, then set the frequency to the same measurement frequency as the analyzer is set in the previous step. 60.On the analyzer, press PEAK SEARCH. 61.On the synthesized sweeper, press POWER LEVEL and adjust the amplitude so the analyzer marker amplitude reads +5 dBm 0.05 dB. 62.Set the power sensor cal factor (for frequency being measured) on the measuring receiver, then record the measuring receiver power reading in Table 2-25. 63.Repeat step 58 to step 62 for frequencies of 1477 MHz, 1453 MHz, and 1429 MHz. Record the results in Table 2-25.
Table 2-25
170
Chapter 2
Making Adjustments 28. Correcting for Absolute Amplitude Accuracy for Option 051
The next step disables the protection for the factory correction constants. 2. Press the following analyzer keys.
FREQUENCY, 2001, Hz CAL, More 1 of 4, More 2 of 4 Service Cal Flatness Data EDIT FLATNESS
Chapter 2
171
Making Adjustments 28. Correcting for Absolute Amplitude Accuracy for Option 051
3. Press (step up key) until the ERROR readout for 804 MHz is displayed (781 MHz for an 8591E spectrum analyzer or an 8591C cable TV analyzer). 4. Read the amplitude error from the display and record it in Table 2-26 for the 8593E, 8594E, 8595E, and 8596E spectrum analyzers. Use Table 2-27 for 8591E spectrum analyzers or 8591C cable TV analyzers. Table 2-26 Frequency Response Errors: 8593E, 8594E, 8595E, 8596E
Amplitude ERROR _________________ _________________ _________________ _________________ Corrected Flatness ERROR _________________ _________________ _________________ _________________
Frequency
Table 2-27
Frequency
Repeat step 3 to step 4 for the next frequency indicated in Table 2-26 or Table 2-27 until all four points have been recorded. 5. Subtract the Average Error, calculated in step 1, from each Amplitude ERROR from Table 2-26 or Table 2-27, then record column 3 as the Corrected Flatness ERROR. Do not subtract more than 0.5 dB from the Amplitude ERROR. If the Average Error is more than 0.5 dB, it may not be possible to make corrections during the adjustment. 6. Using (step down key), set the frequency displayed on the analyzer to 804 MHz (781 MHz for 8591E spectrum analyzers or 8591C cable TV analyzers).
172
Chapter 2
Making Adjustments 28. Correcting for Absolute Amplitude Accuracy for Option 051
7. Enter the Corrected Flatness ERROR from Table 2-26 or Table 2-27 for the frequency displayed using the DATA keys. Terminate the entry with the +dBm. After the new data is entered, the analyzer will automatically jump to the next frequency correction point. Enter the next Corrected Flatness ERROR from the table. To conrm the entries, press (step down key) to the frequency of interest. Re-check the displayed error against the Corrected Flatness ERROR from column 3. 8. Press More, EXIT after all corrections are complete.
Frequency Range 1429 MHz to 1501 MHz 9. Looking at Table 2-22, Frequency Response Attenuator 10 dB, calculate the error from the nominal value of 9 dB for each frequency. For example, if the measuring receiver reading at 1429 MHz is 9.2 dB, then the error is 0.2 dB. Find the average error by adding the four errors for each frequency and then dividing the sum by four. Average Error__________________dB
CAUTION
The next step disables the protection for the factory correction constants. 10.Press the following analyzer keys.
PRESET (wait for completion) FREQUENCY, 2001, Hz CAL, More 1 of 4, More 2 of 4 Service Cal Flatness Data EDIT FLATNESS
11.Press (step up key) until the ERROR readout for 1.380 GHz is displayed (1.410 GHz for an 8591E spectrum analyzer or an 8591C cable TV analyzer).
Chapter 2
173
Making Adjustments 28. Correcting for Absolute Amplitude Accuracy for Option 051
12.Read the amplitude error from the display and record it in Table 2-28 for the 8593E, 8594E, 8595E, and 8596E spectrum analyzers. Use Table 2-29 for 8591E spectrum analyzers or 8591C cable TV analyzers.
Table 2-28
Frequency
Table 2-29
Frequency
Repeat step 10 and step 11 for the next frequency indicated in Table 2-28 or Table 2-29 until all four points have been recorded. 13.Subtract the Average Error, calculated in step 9, from each Amplitude ERROR from Table 2-28 or Table 2-29, then record column 3 as the Corrected Flatness ERROR. Do not subtract more than 0.5 dB from the Amplitude ERROR. If the Average Error is more than 0.5 dB, it may not be possible to make corrections during the adjustment. 14.Using (step down key), set the frequency displayed on the analyzer to 1.380 GHz (1.410 GHz for 8591E spectrum analyzers or 8591C cable TV analyzers).
174
Chapter 2
Making Adjustments 28. Correcting for Absolute Amplitude Accuracy for Option 051
15.Enter the Corrected Flatness ERROR from Table 2-28 or Table 2-29 for the frequency displayed using the DATA keys. Terminate the entry with the +dBm. After the new data is entered, the analyzer will automatically jump to the next frequency correction point. Enter the next Corrected Flatness ERROR from the table. To conrm the entries, press (step down key) to the frequency of interest. Re-check the displayed error against the Corrected Flatness ERROR from column 3. 16.Press More, EXIT after all corrections are complete. 17.Perform the frequency response verication test from the calibration guide for your instrument. Only checking Band 0 is necessary.
Chapter 2
175
Making Adjustments 28. Correcting for Absolute Amplitude Accuracy for Option 051
176
Chapter 2
2a
177
178
Chapter 2a
Making Adjustments: If 3335A Source Not Available 6a. Cal Attenuator Error
Equipment Required
Synthesized Signal generator 8663A 1 dB step attenuator 8494G Attenuator/switch driver (if programmable step attenuators are used) 11713A Cable, Type-N 152 cm (60 in) Cable, BNC, 122 cm (48 in) Attenuator interconnect kit Adapter, Type N (m) to BNC (f)
Chapter 2a
179
Making Adjustments: If 3335A Source Not Available 6a. Cal Attenuator Error
Figure 2a-1
180
Chapter 2a
Making Adjustments: If 3335A Source Not Available 6a. Cal Attenuator Error
Procedure
The accuracy of the amplitude control attenuator is critical to the proper calibration of the instrument; therefore, this procedure must be carefully and accurately performed. 1. Connect the equipment as shown in Figure 2a-1. The 75 RF input requires an external pad as shown in Figure 2a-1. 2. Set the frequency of the synthesized signal generator to 200 MHz and the amplitude to -10dBm. 3. Set the external 1 dB step attenuator to 9 dB. 4. To turn the cal attenuator corrections off, press the following analyzer keys.
PRESET CAL, More 1 of 4 CORRECT ON OFF (OFF)
1 dB Step Check 5. To measure the 1 dB step correction, press the following analyzer keys.
FREQUENCY, 200, MHz SPAN, 10, MHz AMPLITUDE, 18, dBm SCALE LOG LIN (LIN), More 1 of 2 AMPTD UNITS, dBm PEAK SEARCH MKR FCTN, MK TRACK ON OFF (ON) SPAN, 50, kHz BW, 3, kHz VID BW AUTO MAN, 300, Hz
75 input: AMPLITUDE, More 1 of 2, AMPTD UNITS, dBm 6. Press the following analyzer keys.
PEAK SEARCH, MARKER DELTA MKR FCTN, MK TRACK ON OFF (ON)
7. The MKR amplitude reading should be 0.0 dB 0.01 dB. If it is not, repeat step 5.
Chapter 2a
181
Making Adjustments: If 3335A Source Not Available 6a. Cal Attenuator Error
8. Set the analyzers reference level by pressing AMPLITUDE, 17, dBm. 9. Set the 1 dB step attenuator to 8 dB. 10.The 1 dB attenuator step error is the MKR reading minus the error in the 1 dB step attenuator when set to 9 dB and 8 dB in Table 2a-1. 1 dB atten step error = Analyzer MKR - (9dB atten error - 8 dB atten error) 11.Enter the 1 dB attenuator step error in Table 2a-2.
2 dB Check
12.To measure the 2 dB step correction, press the following analyzer keys.
AMPLITUDE, 16, dBm
13.Set the 1 dB step attenuator to 7 dB. 14.Press the following analyzer keys.
PEAK SEARCH, MARKER DELTA MKR FCTN, MK TRACK ON OFF (ON)
15.The MKR amplitude reading should be 0.0 dB 0.01dB. If it is not, repeat step 5. 16.Set the analyzers reference level by pressing AMPLITUDE, 14, dBm. 17.Set the 1 dB step attenuator to 5 dB. 18.The 2 dB attenuator step error is the MKR reading minus the error in the 1 dB step attenuator when set to 7 dB and 5 dB in Table 2a-1. 2 dB attenuator step error = Analyzer MKR - (7dB attenuator error - 5 dB attenuator error) 19.Enter the 2 dB attenuator step error in Table 2a-2.
182
Chapter 2a
Making Adjustments: If 3335A Source Not Available 6a. Cal Attenuator Error
4 dB Step Check
20.To measure the 4 dB step correction, press the following analyzer keys.
AMPLITUDE, 17, dBm
21.Set the 1 dB step attenuator to 8 dB. 22.Press the following analyzer keys.
PEAK SEARCH, MARKER DELTA MKR FCTN, MK TRACK ON OFF (ON)
23.The MKR amplitude reading should be 0.0 dB 0.01dB. If it is not, repeat step 5. 24.Set the analyzers reference level by pressing AMPLITUDE, 13, dBm. 25.Set the 1 dB step attenuator to 4 dB. 26.The 4 dB attenuator step error is the MKR reading minus the error in the 1 dB step attenuator when set to 8 dB and 4 dB in Table 2a-1. 4 dB attenuator step error = Analyzer MKR - (8dB attenuator error - 4 dB attenuator error) 27.Enter the 4 dB attenuator step error in Table 2a-2.
8 dB Step Check
28.To measure the 8 dB step correction, press the following analyzer keys.
AMPLITUDE, 18, dBm
29.Set the 1 dB step attenuator to 9 dB. 30.Press the following analyzer keys.
PEAK SEARCH, MARKER DELTA MKR FCTN, MK TRACK ON OFF (ON)
31.The MKR amplitude reading should be 0.0 dB 0.01dB. If it is not, repeat step 5. 32.Set the analyzers reference level by pressing AMPLITUDE, 10, dBm. 33.Set the 1 dB step attenuator to 1dB.
Chapter 2a
183
Making Adjustments: If 3335A Source Not Available 6a. Cal Attenuator Error
34.The 8 dB attenuator step error is the MKR reading minus the error in the 1 dB step attenuator when set to 9 dB and 1 dB in Table 2a-1. 8 dB atten step error = Analyzer MKR - (9dB atten error - 1 dB atten error) 35.Enter the 8 dB attenuator step error in Table 2a-2.
Table 2a-2
Note that the frequency of 2001 Hz is necessary to access the SERVICE CAL routines. 37.When the analyzer prompts you with the message ENTER CAL ATTEN ERROR, enter the data from Table 2a-2 and terminate with the reverse sign using the dBm or dBm keys. For example, if the error recorded in Table 2a-2 is 0.07, enter 0.07 and terminate with dBm. 38.When the analyzer prompts you to enter the 16 dB step, enter 0 dBm. At the completion of entering the 16 dB error, the analyzer will reset.
184
Chapter 2a
Making Adjustments: If 3335A Source Not Available 6a. Cal Attenuator Error
39.To conrm that the correct data is stored, access the cal attenuator corrections by pressing the following analyzer keys.
CAL, More 1 of 4, More 2 of 4 SERVICE DIAG DISPLAY CAL DATA
Note that the cal-attenuator corrections are the rst ve corrections located in the ERR column.
Chapter 2a
185
Making Adjustments: If 3335A Source Not Available 7a. Log and Linear Amplier
Equipment Required
Synthesized signal generator Digital multimeter (DMM) IF test board Cable, BNC, 120 cm (48 in) Cable, BNC (f) to dual banana plug Test cable BNC (m) to SMB (f) Adapter tip FN (m) to BNC (f) 10 dB step attenuator (8496 A/G) Attenuator/switch driver (if programmable step attenuators are used) Attenuator interconnect kit
186
Chapter 2a
Making Adjustments: If 3335A Source Not Available 7a. Log and Linear Amplier
Procedure
1. Turn the analyzer LINE switch to OFF. Remove the instrument cover assembly. 2. Remove the rst IF bandwidth lter assembly, A11. Install the IF test board into the A11 slot. Turn the analyzer LINE switch to ON. 3. Set the DMM to read dc volts. 4. Press the following analyzer keys.
PRESET CAL, More 1 of 4 CORRECT ON OFF (OFF), More 2 of 4 SERVICE DIAG STP GAIN ZERO SPAN, 0, Hz BW, 10, kHz VID BW AUTO MAN, 300, Hz AMPLITUDE, 10, dBm SCALE LOG LIN (LIN)
5. Set the synthesized signal generator as follows: FREQUENCY ............................................. 21.4 MHz AMPLITUDE .................................................. 6 dBm 6. Connect equipment as shown in Figure 2a-2.
Chapter 2a
187
Making Adjustments: If 3335A Source Not Available 7a. Log and Linear Amplier
Figure 2a-2
188
Chapter 2a
Making Adjustments: If 3335A Source Not Available 7a. Log and Linear Amplier
Figure 2a-3
11.Set the 10 dB step attenuator to 60 dB. Adjust A14R10 OFFSET for the DMM reading of 250 mV 2 mV. 12.Repeat step 10 and step 11 until no further adjustment is necessary. 13.Set the 10 dB step attenuator to 30 dB. Adjust the A14R23 SLOPE for a DMM reading of 625 mV 2 mV. 14.Set the 10 dB step attenuator back to 0 dB. Adjust the A14R69 30 dB for a DMM reading of 1000 mV 1 mV. 15.Repeat step 13 and step 14 until no further adjustment is necessary. 16.Set the 10 dB step attenuator to 10 dB. Adjust the A14R23 SLOPE for a DMM reading of 875 mV 2 mV. 17.Set the 10 dB step attenuator back to 0 dB. Adjust the A14R39 10 dB for a DMM reading of 1000 mV 2 mV. 18.Repeat step 16 and step 17 until no further adjustment is necessary. 19.Repeat step 10 to step 18 until the limits in Table 2a-3 are met.
Chapter 2a
189
Making Adjustments: If 3335A Source Not Available 7a. Log and Linear Amplier
Table 2a-3
Synthesized Sweeper Level Reference from step 8 Reference 10 dB Reference 20 dB Reference 30 dB Reference 40 dB Reference 50 dB Reference 60 dB Reference 70 dB
21.Set the synthesized sweeper amplitude to the level recorded in step 8 and adjust A14R34 LIN for a DMM reading of 1000 mV 1 mV. 22.Make the adjustments indicated in Table 2a-4.
Table 2a-4
Adjust
190
Chapter 2a
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
Equipment Required
Synthesized signal generator Measuring receiver (used as a power meter) Frequency synthesizer Power sensor, 1 MHz to 1.8 GHz Power splitter Adapter, Type N (f) to APC 3.5 (m) Adapter, Type N (m) to Type N (m) Cable, BNC, 122 cm (48 in.) Cable, Type N, 183 cm (72 in.)
Chapter 2a
191
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 5. Adjust the synthesized signal generator POWER LEVEL for a 0 dBm reading on the measuring receiver.
192
Chapter 2a
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
6. Record the power meter reading in Column 4 of Table 2a-5, taking into account the cal factors of both power sensors. 7. On the synthesized signal generator, press CW and STEP UP, to step through the remaining frequencies listed in Table 2a-5. At each new frequency repeat step 5 and step 6, and enter each power sensor cal factor into the respective power meter.
Procedure
1. Zero and calibrate the measuring receiver and 1 MHz to 1.8 GHz power sensor in log mode as described in the measuring receiver operation manual. 2. Connect the equipment as shown in Figure 2a-5. Figure 2a-5 Frequency Response Setup
CAUTION
Use only 75 cables, connectors, or adapters on instruments equipped with 75 inputs or damage to the input connectors will occur. 3. Press INSTRUMENT PRESET on the synthesized signal generator. Set the synthesized signal generator controls as follows: CW ................................................................ 300 MHz FREQ STEP ................................................... 37 MHz POWER LEVEL ............................................. 9 dBm
Chapter 2a
193
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
5. Adjust the synthesized signal generator POWER LEVEL for a MKR-TRK amplitude reading of 14 dBm 0.1 dB. 6. Press RATIO mode on the measuring receiver. 7. Press the following analyzer keys.
FREQUENCY, 4, MHz
8. Set the synthesized signal generator CW to 4 MHz. 9. Adjust the synthesized signal generator POWER LEVEL for an analyzer MKR-TRK amplitude reading of 14 dBm 0.1 dB. 10.Record the power ratio in Column 2 of Table 2a-5 for 4 MHz. 11.Press the following analyzer keys.
FREQUENCY, 41, MHz
12.Set the synthesized signal generator CW to 41 MHz. 13.Adjust the synthesized signal generator POWER LEVEL for an analyzer MKR-TRK amplitude reading of 14 dBm 0.1 dB. 14.Record the power ratio displayed on the measuring receiver in Column 2 of Table 2a-5 for 41 MHz. 15.On the analyzer, press FREQUENCY and (step up), to step through the remaining frequencies listed in Column 1 of Table 2a-5. At each new frequency repeat step 13 to step 15, entering the power sensor Cal Factor into the measuring receiver as indicated in Column 3 of Table 2a-5.
194
Chapter 2a
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
75 input: AMPLITUDE, More 1 of 2, AMPTD UNITS, dBm 18.To access the atness correction menu, press the following analyzer keys.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL FLATNESS DATA
Perform the next step only if all the atness correction data must be replaced in memory due to the repair or replacement of the A16 processor/video assembly. 19.To initialize the area of memory where the atness correction data is stored, press the following keys.
INIT FLT PRESET FREQUENCY, 2001, Hz
20.To enter atness corrections, press EDIT FLATNESS. 21.The frequency of the rst data point, 4.00 MHz, will be displayed in the active function block of the analyzer display. 22.Use the data keys on the analyzer to enter the amplitude value for 4 MHz from Column 2 of Table 2a-5 (75 input: Column 5), Frequency Response Errors. Terminate the entry with the dB key. When entering negative amplitude values, precede the numeric entry with the and dB keys or the dB key. The BK SP (backspace) key may be used to correct any entry if the terminator, dB or dB key has not been pressed. Re-enter the data if the terminator has been pressed.
Chapter 2a
195
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
23.Press (step up) and enter the data from Table 2a-5, Column 2 (75 input: Column 5) for the next data point as described in step 21. 24.Repeat step 22 for the remaining atness correction data points listed in Table 2a-5. At each point, verify that the frequency listed in the active function block corresponds to the frequency at which the data was taken. If these two frequencies do not correspond, press (step up) or (step down) until the proper frequency is displayed in the active function block. If some data is incorrect after entering all of the data from Table 2a-5, select the incorrect data point using (step up) or (step down) and re-enter the proper data. 25.After all corrections have been input, press the STORE FLATNESS softkey to store the correction data in nonvolatile memory. The instrument will automatically preset and display CAL: DONE in the active function block of the analyzer.
196
Chapter 2a
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
Table 2a-5
Column 1 Frequency (MHz)
4 41 78 115 152 189 226 263 300 337 374 411 448 485 522 559 596 633 670 707 744 781 818 855 892
_______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________
N/A 0.03 0.1 0.1 0.1 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 0.3 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0
N/A _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________
_______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________ _______________
Chapter 2a
197
Making Adjustments: If 3335A Source Not Available 15a. Frequency Response of the 8590L, 8591C, and 8591E
Table 2a-5
Column 1 Frequency (MHz)
929 966 1003 1040 1077 1114 1151 1188 1225 1262 1299 1336 1373 1410 1447 1484 1521 1558 1595 1632 1669 1706 1743 1780 1817
198
Chapter 2a
199
Commands within parenthesis after a softkey, for example (LOG), are used throughout this chapter to indicate the part of a softkey which should be underlined when the key is pressed. Refer to Chapter 4 for information that is useful when rst starting to troubleshoot an analyzer failure.
The analyzer contains potentially hazardous voltages. Refer to the safety symbols on the analyzer and the general safety considerations in this guide before operating the unit with the cover removed. Failure to heed the safety precautions can result in severe or fatal injury.
200
Chapter 3
Chapter 3
201
4. Press VERIFY TIMEBASE, then record the number that is displayed in the active-function block in Table 3-2. 5. Press the following keys.
SERVICE CAL, FLATNESS DATA, EDIT FLATNESS
6. The signal trace represents the frequency-response (atness) correction-constant data. The active-function block displays the frequency response error. 7. Record the frequency-response error in the appropriate table for your analyzer.Table 3-3 is for the 8590L and 8591E spectrum analyzers and 8591C cable TV analyzers. Table 3-5 through Table 3-12 are for all other 8590 E-Series and L-Series spectrum analyzers. 8. Press , then record the next frequency-response error in the appropriate table. 9. Repeat the previous step until all frequency-response errors are recorded. Use to view previous data points. 10.Press EXIT when all frequency-response errors have been recorded.
202
Chapter 3
2. Look at the rst ve entries in the CA ATT ERR column; they are the amplitude errors for the 1 dB, 2 dB, 4 dB, 8 dB, and 16 dB step-attenuators. 3. Record the amplitude errors (correction constants) for the ve step-attenuators in Table 3-14. Step 4 is for analyzers equipped with Option 010 or 011 only. Skip this step for all other analyzers. 4. Record the CALTGX slope and offset correction constants in Table 3-15. The correction constants are printed on a label that is located on the A7A1 Tracking Generator Control Board assembly. File the completed copy of the Correction Constant Backup-Data Record for future reference.
Chapter 3
203
Analyzer Initialization
This procedure is used to restore the factory/service correction constants to the processor/video board assembly, and to initialize the analyzer settings after a non-volatile memory loss. The loss of non-volatile memory may be caused by the following conditions. Installation of a new A16 Processor/Video board assembly Dead B1 battery
204
Chapter 3
If the analyzer is an 8591C, the analyzer's startup routine will identify it as an 8591E. Use this procedure to change the identity back to an 8591C. Press the following keys.
PRESET (wait until preset is complete) DISPLAY, Change Title
Use the softkeys to type in the following remote command. Don't forget to include the semicolon (;). FACTSET 11023,1;
NOTE
Chapter 3
205
206
Chapter 3
Type the value from the Table (corr backup) using the DATA Keys, then press ENTER.
INT FLAT For Option 026: Press INIT FLT 26.5 GHz
EDIT FLATNESS
2. Enter each correction constant listed in the Correction Constant Backup-Data Record, then terminate the entry with the +dBm or dBm key, as appropriate. Each entry is displayed briey before the data-entry routine steps to the next correction data point. Use the and keys to edit previously entered correction data. 3. When all atness-correction constants are entered, press STORE FLATNESS, More, EXIT.
Chapter 3
207
REF LVL OFFSET is displayed in the active-function block above the prompt ENTER CAL ATTEN ERROR 1. 2. At the prompt, enter the ve step-attenuator correction constants (resolution to 0.01 dB) listed in the Correction Constant Backup-Data Record, then terminate each entry with either +dBm of dBm, as appropriate. Typically 1 and 2 dB step errors are negative. The 4 and 8 dB steps are positive. The 16 dB step is always 0 dB. Each entry is displayed to the left of the graticule as an amplitude offset, but only with 0.1 dB resolution. A PRESET occurs after the 16 dB step-attenuator error is entered.
208
Chapter 3
Backing Up and Reloading Correction Constants Instrument Calibration after Reloading the Correction Constants
Instrument rmware expects the cal output signal to be 300 MHz 2 MHz. Sometimes the instrument default data is not able to tune the cal signal within this range and a cal signal not found message may appear on screen. Perform a cal output bypass check by pressing Frequency, -37, Hz, Cal, Cal Freq. This will bypass the cal check and start by calibrating the sweep ramp.
Chapter 3
209
Backing Up and Reloading Correction Constants Instrument Calibration after Reloading the Correction Constants
Table 3-1
Serial No.:_______________________
Date:___________________________
Table 3-2
Timebase
_________________________________
Table 3-3
Frequency (MHz) 4 41 78 115 152 189 226 263 300 337 374 411 448
210
Chapter 3
Backing Up and Reloading Correction Constants Instrument Calibration after Reloading the Correction Constants
Table 3-4
Serial No.:_______________________
Date:___________________________
Table 3-5
Frequency-Response Correction Constants for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, or 8596E Band 0
Error (dB) Frequency (GHz) Error (dB) Frequency (GHz) Error (dB) Frequency (GHz) Error (dB)
Frequency (GHz)
0.012 0.084 0.156 0.228 0.300 0.372 0.444 0.516 0.588 0.660 0.732
_______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______
0.804 0.876 0.948 1.020 1.092 1.164 1.236 1.308 1.380 1.452 1.524
_______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______
1.596 1.668 1.740 1.812 1.884 1.956 2.028 2.100 2.172 2.244 2.316
_______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______
Table 3-6
Frequency (GHz)
6.235 6.5084
_______ _______
Chapter 3
211
Backing Up and Reloading Correction Constants Instrument Calibration after Reloading the Correction Constants
Table 3-7
Serial No.:_______________________
Date:___________________________
Table 3-8
Frequency (GHz)
6.100 6.284 6.468 6.652 6.836 7.020 7.204 7.388 7.572 7.756
_______ _______ _______ _______ _______ _______ _______ _______ _______ _______
7.940 8.124 8.308 8.492 8.676 8.860 9.044 9.228 9.412 9.596
_______ _______ _______ _______ _______ _______ _______ _______ _______ _______
9.780 9.964 10.148 10.332 10.516 10.700 10.884 11.068 11.252 11.436
_______ _______ _______ _______ _______ _______ _______ _______ _______ _______
Table 3-9
Frequency (GHz) 12.450 12.680 12.910 13.140 13.370 13.600 13.830 14.060
212
Chapter 3
Backing Up and Reloading Correction Constants Instrument Calibration after Reloading the Correction Constants
Table 3-10
Serial No.:_______________________
Date:___________________________
Table 3-11
Frequency (GHz)
Table 3-12
Frequency (GHz) 19.100 19.248 19.396 19.544 19.692 19.840 19.988 20.136 20.284 20.432 20.580 20.728 20.876
Chapter 3
213
Backing Up and Reloading Correction Constants Instrument Calibration after Reloading the Correction Constants
Table 3-13
Serial No.:_______________________
Date:___________________________
Table 3-14
1 dB 2dB
_______ _______
Table 3-15
Slope
_______________________
Offset
_______________________
214
Chapter 3
This chapter provides information that is useful when starting to troubleshoot an analyzer failure. It provides procedures for troubleshooting common failures and isolating problems in the analyzer.
215
Additional troubleshooting details for specic assemblies are available in Chapter 5 and Chapter 6 of this service guide. Assembly descriptions are located in Chapter 9. Component-level information for the 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, 8594Q QAM analyzers is provided in the 8590 Series Analyzers Component-Level Repair Service Guide binder. Refer to Chapter 12 for a list of available component-level service information.
The analyzer contains potentially hazardous voltages. Refer to the safety symbols on the analyzer and the general safety considerations at the beginning of this service guide before operating the unit with the cover removed. Failure to heed the safety precautions can result in severe or fatal injury.
216
Chapter 4
Replacement assemblies
Some of the 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzer, and 8594Q QAM analyzer assemblies are not repairable to the component level. The following assemblies are not repairable to the component level and must be replaced as an assembly. A2 display A3 input attenuator (8590L, 8591E, and 8591C only) A3A2 RF switch (8592L, 8593E, 8595E, and 8596E only) A3A3 2.9 GHz low-pass lter (8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E only) A3A4 second converter (8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E only) A3A5 input attenuator (8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E only) A3A6 dual mixer (8592L, 8593E, 8595E, and 8596E only) A3A6 low band mixer (8594E, 8594L, and 8594Q only) A3A7 YTO (8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E only) A3A8 YTF (8592L and 8593E only) A3A8 switched YTF (8595E, and 8596E only) A3A9 321.4 MHz bandpass lter (8592L, 8593E, 8594E, 8594L, 8594Q,8595E, and 8596E only) A3A14 rst LO distribution amplier (Option 009, 010 only) A3A15 tracking generator (Option 010 only) A4 rst converter (8590L, 8591E, and 8591C only) A6 YTO (8590L, 8591E, and 8591C only) A8 power supply A22 10 MHz precision frequency reference
Chapter 4
217
A25A1 sampler A26 TV tuner module (Option 107) Qualied Service Center Only A130 narrow bandwidth (Option 030 only) Refer to Chapter 12 when ordering replacement assemblies.
218
Chapter 4
Immediately unplug the analyzer from the ac power line if the unit shows any of the following symptoms. Smoke, arcing, or unusual noise from inside the unit. No response of any kind when unit is plugged into ac power mains and turned on. The analyzer ac power fuse blows. A circuit breaker or fuse on the main ac power line opens. These potentially serious faults must be corrected before proceeding. Refer to Troubleshooting an Inoperative Analyzer. Note that analyzers equipped with the AM/FM speaker (Options 102, 103, or 110) normally emit noise from the speaker at power-up. This is not a problem, adjust the volume control as desired. Table 4-1 lists symptoms that can occur when the instrument is rst powered on. Refer to this table for an overview of symptoms at power-up and their possible cause.
Chapter 4
219
Table 4-1
A8 LED 15 V
Line LED
ON OFF OFF ON ON ON ON ON ON ON ON ON ON ON ON ON
ON ON OFF M ON M ON M OFF ON M ON ON ON ON ON
M = LED or assembly is momentarily on, then goes off. X = Status of LED or assembly does not matter. * DS13 remains on after PRESET is pressed.
220
B1 Fan ON ON
Chapter 4
For continued protection against re hazard, replace fuse only with same type and ratings, (type 5A/250V). Failure to use the proper fuse specied for the analyzer can cause substantial instrument damage. The use of other fuses or materials is prohibited.
The analyzer contains potentially hazardous voltages. Refer to the safety symbols provided on the analyzer, and in the general safety instructions in this guide, before operating the unit with the cover removed. Ensure that safety instructions are strictly followed. Failure to do so can result in severe or fatal injury.
Chapter 4
221
Figure 4-1
3. Turn the analyzer LINE switch to ON. 4. Locate the power supply test point using Figure 4-1. Check the supply voltages as indicated in Table 4-2. Be sure the voltage readings are within the limits shown in Table 4-2. 5. Repeat step 4 for each power supply listed in Table 4-2. Table 4-2 Power Supply Tolerances
Power Supply +5.1 V +12.0 V +15.0 V 15.0 V Test Point A16TP403 A16TP404 A16TP401 A16TP402 Specication +5.0 to +5.25 Vdc +11.1 to +12.8 Vdc +14.7 to +15.25 Vdc 14.7 to 15.25 Vdc
The A8 power supply assembly is a switching power-supply and does not operate normally without a load on the dc power-supply outputs. Do not attempt to operate the power supply out of the instrument. Damage to the power supply may occur.
222
Chapter 4
The A8 power supply may be hot if the instrument has been operating without the fan running. Allow the instrument to cool down before troubleshooting. The B1 Fan receives +12 Vdc from the A8 power supply after the power supply receives the start-up signal, low-power-on (LPWRON) from the front-panel LINE switch. First check the +12 Vdc supply to the fan at the Rear Frame. 1. Disconnect the line-power cord from the analyzer. 2. Look through the fan grill. If the cables routed near the fan are jammed in the fan, remove the fan from the rear frame, reroute the cables, and remount the fan. 3. Remove the fan from the rear frame and disconnect its wiring connector from A15J19. 4. Reconnect line power to the analyzer and turn it on. 5. Refer to A15 Motherboard Assembly in Chapter 5 for the pin identication of the fan-supply connector, A15J19, and the A8 power-supply connector, A8J1. 6. Check that pin 2 on A15J19 is connected to digital common-ground (DCOM). If the ground connection is open, suspect the A15 Motherboard. 7. Check pin 1 on A15J19 for +12 Vdc. If the +12 Vdc is present, check for a defective B1 fan assembly by connecting the fan to an external DC supply.
Chapter 4
223
If the +12 Vdc is absent, do the following: Disconnect the analyzer from the line power. Remove the A8 power supply. Check the continuity of the A15 motherboard from pin 1 on A15J19 to pin 8 on A15J13. If the continuity of the motherboard is correct, suspect a bad connection on the D sub-connector (between pin 8 on A15J13 and pin 8 on A8J1) or a defective power supply.
224
Chapter 4
Figure 4-2
Chapter 4
225
226
Chapter 4
Chapter 4
227
WARNING
High voltage is present within the A2 display and remains for some time after it has been disconnected. Be careful while troubleshooting the display outside of its enclosure. If an A2 display assembly failure is suspected, the display must be removed from its enclosure for further troubleshooting. Refer to the A2 display assembly removal procedure in Chapter 3 before checking the continuity of the W2 and W3 wire harnesses. If both W2 and W3 are good, suspect the A2 display. The A2 display is not eld-repairable, and must be replaced as an assembly. Refer to Chapter 10 for part number information. After replacement, adjust the new display, refer to step 2.
The A1R1 potentiometer leads are easily broken. Do not twist the W3 wire harness where it attaches to A1R1. 1. Disconnect W3 from the A1R1 intensity control potentiometer on the front frame. 2. Check the continuity and resistance range (0 to 100 k) of the inner potentiometer on A1R1. 3. Reconnect W3 to A1R1.
Do not short any two of the A16J8 pins together. Damage to the output buffer ampliers located on the processor/video assembly may occur. Disconnect the W2 wire harness.
228
Chapter 4
Refer to Figure 4-4 for the location of the test points and check the following signals at A16J8 pins, using an oscilloscope. t Check for a 60 Hz TTL signal with a pulse width of approximately 0.2 ms and a repetition rate of approximately 18 ms at pin 2, vertical sync signal. t Check for an approximately 2.5 V pk-pk video signal at pin 3, video signal. t Check for a 15.75 kHz TTL signal with a pulse width of approximately 10 s and a repetition rate of approximately 64 s at pin 5, horizontal sync signal. If the video, horizontal, and vertical signals are missing, but the +12 V supply voltage is present, continue with step 3. 3. Check the +12 Vdc power supplied to the A2 display. There is a separate 12 V supply for the display assembly. This supply can be checked at the output connector, A8J6, located on the right side of the A8 power supply. Refer to Figure 4-3. If there is a failure, check the continuity of the cable assembly W51 supplying the display, and the integrity of the power supply using the procedures in this chapter. Figure 4-4 A16J8 Display Signal Output Pins
Chapter 4
229
4. Verify the High-Power-Up signal from the A8 power supply by checking for the +5 V TTL WRUP signal at pin 54 on the A16J1. The A16J1 connector is on the trace side of the A16 processor/video assembly. Refer to Figure 4-5 for the numbering order of A16J1 connector-pins accessible from the component side of the A16 processor/video assembly. Note that the A8 power supply assembly sends the high-power-up (WRUP) signal to the A16 processor/video assembly after it receives the low-power-on (LPWRON) signal and the +5 V supply has stabilized. If WRUP is low, the display remains blank and the A16 failure LEDs (DS1 through DS16) will not come on during power-up. If the WRUP signal is missing, trace the signal back through the A15 motherboard to the A8 power supply assembly, pin 37 on A8J1. Refer to A15 Motherboard Assembly in Chapter 5 for the connector-pin designation of the WRUP signal path. Note that if your analyzer is an 8590 E-Series, 8591C, or 8594Q QAM analyzer, the WRUP signal can be measured on pin 55 of each card-cage slot on the A15 motherboard. Refer to Figure 4-6 for the numbering order of the connector pins on A15J3 through A15J6. If the WRUP signal is present, yet one or more of the A16 processor/video assembly outputs is incorrect, suspect a defective A16 processor/video assembly. Before replacing the A16 processor/video assembly, refer to Chapter 3 in this service guide. If all the A16 display outputs are correct, suspect a defective A2 display.
230
Chapter 4
Figure 4-5
Figure 4-6
Chapter 4
231
232
Chapter 4
The DAC value is unique for each analyzer. The CAL AMPTD self-calibration routine adjusts this value based on the amplitude of the digitized video signal on the processor/video board assembly. The calibration routine makes this adjustment while calibrating the reference level at top screen. After running the CAL AMPTD calibration routine, the REF LVL CAL DAC has a typical value of 165 with a range of 130 to 185. The higher the DAC value, the greater the output from the A9 third converter assembly. 4. Check the signal path from the analyzer input to the output of the A14 log amplier assembly. Refer to IF Power-Level Measurement located in Chapter 5 of this guide. 5. Measure the 321.4 MHz input to the A9 third converter assembly by connecting the CAL OUT to the INPUT 50 with a BNC cable. 75 input only: Connect the CAL OUT to the INPUT 75 with the 75 BNC cable. Then, disconnect the W9 cable assembly from the third converter and measure at the cable output. The 321.4 MHz input typically measures between 36 and 39 dBm with zero input attenuation. If the REF LVL CAL DAC value is 255 and the input to the A9 third converter assembly is above the test limit, suspect a low gain problem in the IF section. Continue with the troubleshooting procedures in IF Power-Level Measurement. If the REF LVL CAL DAC value is 0 and the input to the A9 third converter assembly is nominal, suspect a high gain problem in the IF section. Continue with the troubleshooting procedures in IF Power-Level Measurement. If the REF LVL CAL DAC value is 255 and the input to the A9 third converter is below the test limit, suspect a low gain problem in the RF/LO sections. Continue with the next step. If the REF LVL CAL DAC value is 0 and the input to the A9 third converter is above the test limit, suspect a high gain problem in the RF section. Continue with the next step. 6. Set up the analyzer as shown in Instrument Settings for RF Power-Level Measurement in this chapter. 7. Isolate the failure to an RF or LO assembly by performing the RF Power-Level Measurements procedure in Chapter 6 of this guide. Once the problem has been isolated to one of the four sections, use standard troubleshooting methods to locate the source of the failure. The following failure symptoms are examples to help isolate a failure to either the RF, LO, IF, or Video sections.
Chapter 4
233
3. Check the signal path from the analyzer input to the output of the A14 log amplier assembly. Refer to IF Power-Level Measurement located in Chapter 5 of this service guide.
234
Chapter 4
The analyzer may also fail the displayed average noise performance verication test. 1. Check the REF LVL CAL DAC value using DISPLAY CAL DATA as described in Using the Internal Service-Diagnostic Routines. This type of problem causes the DAC value to be close to the 255 maximum. 2. Measure the 321.4 MHz input to the A9 third converter assembly by connecting the CAL OUT to the INPUT 50 with a BNC cable. 75 input only: Connect the CAL OUT to the INPUT 75 with the 75 BNC cable. Then, disconnect the W9 cable assembly from the third converter and measure at the cable output. The 321.4 MHz input typically measures between 36 and 39 dBm with zero input attenuation. 3. Check the signal path from the analyzer input to the output of the A14 log amplier assembly. Refer to IF Power-Level Measurement located in Chapter 5 of this service guide.
Chapter 4
235
236
Chapter 4
Chapter 4
237
Note that whenever an analyzer setting has been changed while in a service-diagnostic routine, you must return to the service-diagnostic routine softkey menu by pressing the key sequence listed above.
238
Chapter 4
2. Verify that the digital correction values are within their limits. Refer to Chapter 14 for typical examples of the calibration data displays. If default values are being used, all digital corrections should be zero. If corrections are being used, make sure that all digital corrections are present. Typical values should be below 1.00. 3. Return to the diagnostics menu by pressing the following analyzer keys.
PRESET FREQUENCY, BAND LOCK, 0-2.9 GHz BAND 0 (except 8590L,
Chapter 4
239
240
Chapter 4
Chapter 4
241
3. Verify the main coil driver by pressing MAIN COIL DRIVE. If the signal trace is similar to Figure 4-9, the main coil driver is functioning properly. If the signal is not similar, suspect the A7 analog interface assembly. Figure 4-9 Main Coil Driver Typical Display
242
Chapter 4
3. Verify the FM coil driver by pressing FM COIL DRIVE. If the signal trace is similar to Figure 4-10, the FM coil driver is functioning properly. If the signal is not similar, suspect the A7 analog interface board assembly. For analyzers equipped with an A25 counterlock assembly only: If all other service-diagnostic routines operate properly, disconnect the A25 counterlock assembly, perform the FREQ CAL calibration routine, then operate the analyzer as an unlock instrument. If the analyzer is corrected, suspect the A25 counterlock assembly. Figure 4-10 FM Coil Driver with 10 MHz Span
Chapter 4
243
3. Verify the FM coil driver by pressing FM COIL DRIVE. If the signal trace is similar to Figure 4-11 the FM coil driver is functioning properly. If the signal is not similar continue with the next service-diagnostic routine. Figure 4-11 FM Coil Driver with >10 MHz Span
244
Chapter 4
If the signal trace is similar to Figure 4-12 the sweep ramp is functioning properly.
NOTE
Before a successful frequency calibration is performed, the ramp will not be corner to corner, but will have some overshoot or undershoot. If the signal is not similar suspect the A7 analog interface board assembly. Refer to Chapter 6 for further information about troubleshooting the A7 board assembly. 2. Change the analyzer sweep time.
NOTE
It is recommended that all sweep times be checked due to the A7 analog interface board assembly operating differently for the different RC combinations. This can result in the analyzer appearing to operate properly in one sweep time and not in another. Typical Sweep Ramp Display
Figure 4-12
Chapter 4
245
2. Verify the 10 V reference detector by pressing 10V REF DETECTOR. If the signal trace is at the top of the display, the 10 V reference detector is functioning properly. See Figure 4-13. If the signal is not at the top of the display, the problem is isolated to the A7 analog interface board assembly. Refer to Chapter 6 for further troubleshooting. Figure 4-13 10 V Reference Detector
246
Chapter 4
Chapter 4
247
Note that any band, other than band 0, is acceptable. 2. Verify the YTF driver by pressing:
CAL, More 1 of 4, More 2 of 4, SERV DIAG, More 1, More 2, More 3, More 4, More 5, YTF DRIVER
Bands 2, 3, and 4 will have a similar display at different positions on the display. If the signal trace is similar to Figure 4-15, the YTF driver is functioning properly. If the signal is not similar, suspect the A7 analog interface board assembly. Refer to Chapter 6 for further information about troubleshooting the A7 board assembly. Figure 4-15 YTF Driver
248
Chapter 4
Chapter 4
249
5. Switch between POS PEAK and SAMPLE detector modes. If the signal appears when in the SAMPLE detector, suspect something in the positive-peak detector circuit.
If the analyzer is equipped with Option 021 (IB remote interface), connect it to a controller. Execute the following remote program line to erase user memory and clear the problem. SEND 7;UNL MTA LISTEN 18 CMD 12
While measuring pins on the microprocessor chip, exerting too much pressure on a pin may cause one of the data lines to open. This may result in factory correction constants to be erased.
250
Chapter 4
5. Check A16U12 for the conditions described in Table 4-3 while in the free-run mode. Refer to the component-level information for the A16 processor/video board assembly. Figure 4-16 A16U12 Pin Location
Table 4-3
6. Check for +5 V at pin 1 of ROM's U6, U7, U23, and U24. The Free-Run Check is now complete. 7. Turn the analyzer power off. 8. Remove the clip lead from A16TP1 pin 7 and ground. 9. Replace A16U25 (16 pin DIP jumper pack). 10.Turn the analyzer power on.
Chapter 4
251
252
Chapter 4
253
The analyzer contains potentially hazardous voltages. Refer to the safety symbols on the analyzer and the general safety considerations in the front of this guide before operating the unit with the cover removed. Failure to heed the safety precautions can result in severe or fatal injury.
254
Chapter 5
IF Section Information
This section provides the control details for the assemblies in the IF section. The control-line outputs are valid when the instrument settings provided with each table are used. For many IF functions, the A7 analog interface assembly converts the digital control signals from the A16 processor/video assembly to analog control signals. Some analyzer functions are performed directly by the A16 assembly. The A16 assembly also makes amplitude error corrections to improve instrument performance. The analyzer-setting changes and error-correction functions performed on the A16 assembly are a combined mathematical offset of the digitized video signal. For more information about the A15 motherboard assembly and the tracing of specic signals in the IF section, refer to A15 motherboard assembly in this chapter.
Chapter 5
255
IF Power-Level Measurement
The following measurement procedures are used for troubleshooting along the 21.4 MHz IF signal path from the A9 third converter assembly, through the IF section, to output of the A14 log amplier assembly. To calibrate the reference level of the analyzer, the CAL AMPTD self-calibration routine adjusts the gain of the 21.4 MHz IF variable amplier on the A9 assembly and mathematically offsets the digitized video signal on the A16 processor/video assembly. Due to component variations it is unlikely that any two analyzers will have the same 21.4 MHz IF signal level for the same instrument settings. Furthermore, a defective analyzer may produce misleading IF signal levels if the CAL AMPTD routine has been run. An IF test board is used in the following procedures to simplify troubleshooting. The test board is used to isolate the RF section from the IF section and allows the testing of individual IF assemblies. Refer to Chapter 15, Service Equipment and Tools for the part number of the IF test board.
256
Chapter 5
Chapter 5
257
258
Chapter 5
Chapter 5
259
To check bandwidth control lines for the A11/A13 bandwidth lter assemblies
Table 5-1 provides nominal bandwidth control voltages sent to the A11 and A13 bandwidth lter assemblies from the A7 analog interface assembly. A calibrated analyzer produces control voltages similar to the values in Table 5-1. (The values in Table 5-1 were measured while default calibration data was in use.) Use the difference in control voltage between bandwidths from the table as a guide for normal bandwidth operation. Use an extender board to measure the bandwidth control voltages at the motherboard connector for the A7 assembly or the A11/A13 lter assemblies. Refer to Troubleshooting the A15 Motherboard Assembly for the location of the control lines for each assembly. Table 5-1 Nominal Resolution Bandwidth Control Line Voltages
Resolution Bandwidth 3 dB 5 MHz 3 MHz 1 MHz 0.3 MHz 0.1 MHz 30 kHz 10 kHz 3 kHz 1 kHz 0.3 kHz 6 dB EMI 9 kHz 120 kHz +14.0 +14.0 +14.0 +14.0 +14.0 1.0 1.0 1.0 1.0 1.0 9.90 9.90 9.90 9.90 9.90 +9.25 +8.96 +8.51 +7.50 +6.00 +4.4 +6.4 +8.5 +9.4 +9.7 +9.7 +9.7 +9.7 +9.7 +9.7 BW5 BW6 BW7
1.0 +14.0
+8.81 9.90
+9.7 +9.6
260
Chapter 5
Chapter 5
261
Table 5-2
0 1 2 3 4 5 6 7 8 9 10 20 30 40 50 60 70 80 90 100 110 120 130 131 132 133 134 135 136 137 138 139
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 10 20 30 40 40 40 40 40 40 40 40 40 40 40 40 40 40
* For instruments with 75 Input Impedance, change the display units to dBmv. The gain of the 10 dB step-gain stage is 15 dB when enabled and 5 dB when disabled. Turn CORRECT ON OFF off to remove the amplitude error corrections generated by CAL AMPTD.
262
Chapter 5
Be sure that the input attenuator setting remains at 10 dB. Table 5-3 10 dB Step-Gain Control Lines
Reference Level (dBm) 10 dB Step Gains IFG1 (10 dB Step) A12J2-14 L H L H L H IFG2 (20-1 dB Step) A12J2-13 L L H H H H IFG3 (20-2 dB Step) A12J2-12 L L L L H H
For each reference level in Table 5-3, the Calibration Attenuator control lines are set to 10 dB of attenuation as shown in the rst entry of Table 5-4. The control line settings from Table 5-4 repeat, starting with each reference level in Table 5-3 to produce reference-level changes in 1 dB increments.
Chapter 5
263
Table 5-4
Reference Level (dBm)
264
Chapter 5
To check linear gain control lines for the A14 log amplier
Table 5-5 provides the control line voltages from the A7 analog interface assembly to the A14 log amplier assembly. When the analyzer is in linear mode, the seven A14 log amplier stages are biased to operate as linear ampliers. Four of the seven stages are also used to provide 40 dB of gain in 10 dB increments. The linear gains are enabled for the reference levels indicated in Table 5-2 and Table 5-5. Two of the four stages operate as one 20 dB amplier and are controlled by the same control line, IFG6. Refer to Troubleshooting the A15 Motherboard Assembly in this chapter when tracing control lines in the IF section. When enabled, each control line has a 7.6 Vdc output. This voltage is supplied by the 8 VT temperature-compensated power supply located on the A14 assembly. The entries in Table 5-5 are valid when the instrument is set up as follows:
PRESET SCALE LOG LIN, LIN AMPTD UNITS, dBm CAL More 1 of 3 CORRECT ON OFF (OFF)
Table 5-5
Reference Level (dBm) 50 60 70 80 90
Chapter 5
265
266
Chapter 5
Figure 5-1
All connector pins on A9J20 are tied to ground (ACOM). The A15J1 connector for the A7 analog interface assembly is installed in a position that is reversed when compared to card-cage connectors A15J3, A15J4, A15J5, and A15J6; therefore, the pin-numbering order is also reversed for A15J1 when compared to the card-cage connectors.
Chapter 5
267
Figure 5-2
268
Chapter 5
Figure 5-2
The A15 motherboard pin designation for the 21.4 MHz IF signal path is provided in Table 5-7. All the connector pins with the same number are connected in parallel for the card-cage connectors A15J3, A15J4, A15J5, and A15J6.
Chapter 5
269
Figure 5-3
270
Chapter 5
Figure 5-4
Chapter 5
271
Table 5-6
Mnemonic 21.4 MHz IF
+10V
+10 V Supply
ACOM ADC_SYNC
ADR0-ADR4 ANA_TEST
AUX_IF
Auxiliary IF
AUX_IF_BP
272
Chapter 5
Table 5-6
Mnemonic AUX_VIDEO
CRD_ANLG_1
CRD_ANLG_2
EXT_HSWP
FAN WRUP
Chapter 5
273
Table 5-6
Mnemonic HSWP
IFA1 IFA2 IFA3 IFA4 IFA5 IFG1 IFG2 IFG3 IFG4 IFG5 IFG6 INTERBUS
IF Attenuation 1 IF Attenuation 2 IF Attenuation 3 IF Attenuation 4 IF Attenuation 5 IF Gain 1 IF Gain 2 IF Gain 3 IF Gain 4 IF Gain 5 IF Gain 6 Interbus
IOB0-IOB15
LBIO LINE_TRIG
274
Chapter 5
Table 5-6
Mnemonic LOG_LIN LPWRON
LTIO REF_CAL
SWEEP_RAMP
Sweep Ramp
VIDEO_IF
Video IF
VTO_TUNE
Chapter 5
275
Table 5-7
Mnemonic
A7
A8
A9
A11
A12
A15J13
A15J10
A15J11
A15J12
A15J1
A15J7
A15J8
A15J9
21.4 MHz IF
22, 23
20
22, 23 b 29 b
44 b
29
27, 57 27 b 8, 18 27 b 27b 8, 38
+5 V
19, 49
10, 20
15 V
4, 34
4, 5, 6, 23, 24, 25 c
6, 16 b
25 b
8 VT ACOM
23 2, 32 7, 26, 27 1-8, 10, 12, 14, 16, 18, 20 1-21, 24-26, 28, 33- 44 2-5, 7, 9, 11-15, 17, 19 5- 10, 15- 19 1- 21, 24-26, 28, 33- 44
276
A15J2 9, 39 30 c 11 c 41 c 12 c
Chapter 5
Table 5-7
Mnemonic
A7
A8
A9
A11
A12
A15J13
A15J10
A15J11
A15J12
A15J1
A15J7
A15J8
A15J9
ADR3 ADR4 ANA_TEST AUX_IF AUX_IF_BP AUX_VIDEO b BW5 BW6 BW7 COUNT_IF b CRD_ANLG_1 CRD_ANLG_2 DCOM
37 8 31 22 b, c 22 b, c
24 c 54 c 25 c
30 31 32
30 31 32
14-17, 33-36
8c 50 37 c 54 58 c 58 c 29 c 4 3
Chapter 5
A15J2 42 c 13 c 31 c 7c 3c 5c
277
Table 5-7
Mnemonic
A7
A8
A9
A11
A12
A15J13
A15J10
A15J11
A15J12
A15J1
A15J7
A15J8
A15J9
IFA3 IFA4 IFA5 IFG1 IFG2 IFG3 IFG4 IFG5 IFG6 INTERBUS IOB0 IOB1 IOB2 IOB3 IOB4 IOB5 IOB6 IOB7 IOB8 IOB9 IOB10 IOB11 IOB12 IOB13
59 c 30 c 60 c 55 c 26 c 56 c 27 c 57 c 28 c
2 1 11 14 13 12 38 39 40
278
A15J2
Chapter 5
Table 5-7
Mnemonic
A7
A8
A9
A11
A12
A15J13
A15J10
A15J11
A15J12
A15J1
A15J7
A15J8
A15J9
IOB14 IOB15 LBIO LINE_TRIG LOG_LIN LPWRON LTIO REF_CAL SWEEP_RAMP b VIDEO_IF VTO_TUNE
47 18 5 18 c 52 c 19 31
53 c 51 c
a. 8590 E-Series, 8591C and 8594Q only. b. Refer to the gure, A15 Connectors with Additional Associated Circuitry in this section. c. Pin numbers indicate assembly where signal or voltage originates.
Chapter 5
A15J2 52 23 10 c 60 59 c 24 c
279
280
Chapter 5
281
Troubleshooting the RF Section Before You Start Section 6a, 8590L, 8591C, and 8591E, describes troubleshooting
information for the 8590L and 8591E spectrum analyzers and 8591C cable TV analyzers.
Section 6b, 8592L/94L, 8594Q.and 8593E/94E/95E/96E, describes troubleshooting information for the 8592L, 8593E, 8594E, 8594L, 8595E, and 8596E spectrum analyzers and 8594Q QAM analyzer.
Refer to Chapter 4, Troubleshooting the Analyzer, for troubleshooting procedures that are useful when rst starting to troubleshoot an analyzer failure.
WARNING
The analyzer contains potentially hazardous voltages. Refer to the safety symbols on the analyzer and the general safety considerations in this manual before operating the unit with the cover removed. Failure to heed the safety precautions can result in severe or fatal injury.
282
Chapter 6
Chapter 6
283
Table 6-1
Power Levels at Measurement Points Measurement Frequency 50 A B C D E F G 300 MHz 2.1214 GHz 2.1214 GHz 321.4 MHz 2.4214 GHz 2.4214 GHz 1.8 GHz 29 to 31 39 to 43 39 to 44 42 to 46 +6 to +12 10 to 20 17 to 23 Power Level Range * (dBm) 75 35 to 37 46 to 49 46 to 50 51 to 55 Same as 50 Same as 50 Same as 50
Measurement Point
* A frequency-selective measuring device, such as another analyzer, is recommended for making these measurements. Broadband measuring devices, such as power meters, give erroneous results due to the presence of other, higher-amplitude signals.
284
Chapter 6
A7J2 Connector Pin Designation (Top-Side View) W13 Wire Color Brown Black Orange N/C Green Control line for 20 dB step on A3 Input Attenuator Signal Description +8 VF power supply for A5 Second Converter Analog ground for A5 Second Converter +15 VF power supply for A3 Attenuator
6 7 8
Chapter 6
285
H = +15 V (A high output indicates that the attenuator step is in the signal path.) L = +0.8 V These control voltages are valid only if the A3 Input Attenuator is connected to A7J2.
286
Chapter 6
Chapter 6
287
Troubleshooting the RF Section 6b. 8592L/94L, 8594Q, and 8593E/94E/95E/96E Table 6-4 Measurement Point Power Levels at Measurement Points Measurement Frequency Band 0 92L, 93E, 94E, 94L, 94Q, 95E, 96E A Same as input frequency B Same as input frequency C 321.4 MHz 30 to 33 dBm 14 to 19 dBm 10 to 16 dB below B D 3.9214 GHz 8 to 12 dB below A E 321.4 MHz 5 to 11 dB below D 1 to 6 dB below C 1 to 6 dB below C 1 to 6 dB below C 1 to 6 dB below C 14 to 19 dBm 18 to 26 dB below B 14 to 20 dBm 21 to 31 dB below B 15 to 23 dBm 25 to 35 dB below B Power Level Range * Band 1 92L, 93E, 95E, 96E Band 2 92L, 93E, 96E Band 3 92L, 93E Band 4 92L,93E
* A frequency-selective measuring device, such as another analyzer, is recommended for making these measurements. Broadband measuring devices, such as power meters, will give erroneous results due to the presence of other, higher-amplitude signals. The power-level range is relative to the actual measurement taken at the measurement point indicated.
288
Chapter 6
A7J2 Connector Pin Designation for Option 026 for 8595E A7J2 Pin Number 1 2 3 4 5 6 7 8 9 10 Brown Red N/C N/C Green Blue N/C N/C N/C N/C W13 Wire Color Signal Description Normally Open Normally Closed ACOM +15 VF POS_MXR_BIAS PIN_Switch COMB +15 VF 15 VF
Chapter 6
289
Troubleshooting the RF Section 6b. 8592L/94L, 8594Q, and 8593E/94E/95E/96E Table 6-7 A7J4 2nd Converter Drive Pin Designation for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, 8596E A7J4 Pin Number 1 2 3 4 5 6 7 8 9 10 Table 6-8 Signal Description 2nd Mixer Bias +5 A ACOM +10 VF PIN_SW KEY N/C N/C N/C N/C
A7J301 YTF Driver Pin Designation for 8592L, 8593E, 8595E, 8596E A7J301 Pin Number 1 2 3 4 Brown Orange Green Violet W34 Wire Color Signal Description YIG_FLT+ YIG_FLT15 VF +15 VF
Table 6-9
A10J1 Tracking Generator Control Pin Designation for 8593E, 8594E, 8594L, 8594Q, 8595E, 8596E A10J1 Pin Number 1 2 3 4 5 6 7 8 9 10 Signal Description +15 VF ACOM 15 VF OSC_PWR TUNE ALC_MON KEY PWR_LVL +5 VF1 ALC_EXT
290
Chapter 6
H = 10 V (with the attenuator connected at A7J5) H = 0 V (A oating output if the attenuator is disconnected.) L = 15 V (A low at pin 2, 9, 8, or 7 indicates that the attenuator step is in the signal path.)
Chapter 6
291
Troubleshooting the RF Section 6b. 8592L/94L, 8594Q, and 8593E/94E/95E/96E Table 6-11 Input Attenuator Control Output at A7J5 for 8594E, 8594L, 8594Q, 8595E, 8596E 10 dB Step Attenuator Setting (dB) 0 10 20 30 40 50 60 70 Blocking Capacitor In Blocking Capacitor Out Pin 2 H L H L H L H L Pin 1 L H L H L H L H Pin 9 H H H H L L L L Pin 4 L L L L H H H H Pin 8 H H L L H H L L Pin 5 L L H H L L H H Pin 7 L H Pin 6 H L 40 dB Step 20 dB Step Blocking Capacitor
H = 10 V (with the attenuator connected at A7J5) H = 0 V (A oating output if the attenuator is disconnected.) L = 15 V (A low at pin 2, 9, or 8 indicates that the attenuator step is in the signal path. A low at pin 7 indicates the blocking capacitor is the signal path.)
292
Chapter 6
293
The words right and left are used throughout these procedures to indicate the sides of the analyzer as normally viewed from the front of the instrument. Numbers in parentheses, for example (1), indicate numerical callouts on the gures.
The analyzer contains potentially hazardous voltages. Refer to the safety symbols on the analyzer and the general safety considerations in this guide before operating the unit with the cover removed. Failure to heed the safety precautions can result in severe or fatal injury.
294
Chapter 7
Chapter 7
295
Instrument Cover
This procedure applies to: 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers
Removal
1. Disconnect the analyzer from ac power.
CAUTION
To prevent damage to the front frame, use a soft cloth or towel between the work surface and the front frame. 2. Carefully place the analyzer on the work surface with the front frame facing down. 3. Remove the four screws and washers attaching the instrument cover to the rear frame. 4. Unscrew, but do not remove, the four rear-feet screws, using a 4 mm hex wrench. 5. Pull the instrument cover off towards the rear of the instrument.
296
Chapter 7
Replacement
1. Disconnect the analyzer from ac power.
CAUTION
To prevent damage when replacing the instrument cover, remember the following: Place a soft cloth or towel between the work surface and the front frame. Ensure that cables do not bind between the instrument cover and its internal assemblies. 2. Carefully place the analyzer on the work surface with the front frame facing down. 3. Replace the instrument cover assembly by matching the seam on the cover with the bottom of the instrument. 4. Fit the leading edge of the cover completely into the slot on the back of the front-frame assembly. The cover should t snugly against the EMI gasket in the slot. 5. Tighten the four rear-feet screws with a 4-mm hex wrench. 6. Replace the four screws and washers attaching the instrument cover assembly to the rear frame.
Chapter 7
297
A1 Front-Frame Assembly
This procedure applies to: 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers
CAUTION
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Place the analyzer with the bottom side facing up on the work surface. 3. If the instrument came equipped with the A17 Memory Card Reader, remove it. Refer to the A17 Memory Card Reader Assembly. removal procedure. 4. On 8592L, 8593E, and 8596E spectrum analyzers, remove the cable tie (2) that secures the A3A11 and A3A12 assemblies to the W10 semi-rigid cable. See Figure 7-1. 5. On 8592L, 8593E, and 8596E spectrum analyzers, disconnect the W37 semi-rigid cable (3) from the 100 MHz COMB OUT connector on the front-frame assembly. 6. Remove the two screws (1) that secure the front frame to the main chassis. 7. If the instrument came equipped with a tracking generator, disconnect the tracking generator output semi-rigid cable (5) from the RF OUT connector on the front-frame assembly. 8. Disconnect the W10 semi-rigid cable (4) from the INPUT connector. 9. Disconnect W3 (6) from A1R1, the front-panel INTEN control. It may be necessary to remove the speaker bracket from LS1 as well as A22 10 MHz reference in order to get to W3. 10.If the instrument came equipped with an A102 AM/FM Speaker and TV Synch Trigger (Option 102) or A103 Quasi Peak Detector (Option 103), disconnect W102 from the VOL control wiring connector (8) at A1R1.
298
Chapter 7
11.Disconnect W1 (7) from A16J9. 12.Place the analyzer with the front frame forward and the top side facing up on the work surface. 13.Disconnect W7 from A9J2. 14.Remove the ve screws (9) that secure the front frame to the main chassis sides. There are three screws on the right and two screws on the left.
CAUTION
To prevent damage to the semi-rigid cables, make sure they are completely disconnected from the front panel connectors before removing the front frame. 15.Separate the front frame from the main chassis and disconnect the W4 ribbon cable from the A1A1 Keyboard assembly. 16.Remove the front-frame assembly.
Chapter 7
299
Figure 7-1
A1 Front-Frame Replacement
300
Chapter 7
Replacement
1. Connect W7 to A9J2 and route the cable down between the A2 Display assembly and the Front-End assembly. 2. Connect the W4 ribbon cable to the A1A1 Keyboard assembly. 3. Align the W10 semi-rigid cable (4) with the INPUT connector while replacing the front frame on the main chassis. Make sure that the cables attached to the front frame are not pinched between adjacent assemblies. See Figure 7-1. 4. Replace the ve screws (9) that secure the front frame to the main chassis sides. There are three screws on the right and two screws on the left. 5. Place the analyzer with the front frame forward and the bottom side facing up on the work surface. 6. Connect W1 (7) to A16J9. 7. Connect W3 (6) to the connector from A1R1, the front-frame INTEN control. 8. If the instrument came equipped with an A102 AM/FM Speaker and TV Synch Trigger (Option 102), connect W102 to the VOL control wiring connector (8) at A1R1. 9. If the instrument came equipped with a tracking generator, connect the tracking generator output semi-rigid cable (5) to the RF OUT connector on the front-frame assembly to 10 inch-pounds. 10.Connect the W10 semi-rigid cable (4) at the INPUT connector to 10 inch-pounds. 11.On 8592L, 8593E, and 8596E spectrum analyzers, connect the W37 semi-rigid cable (3) to the 100 MHz COMB OUT on the front-frame assembly. 12.On 8592L, 8593E, and 8596E spectrum analyzers, replace the cable tie (2) around W10 and the A3A11 and A3A12 assemblies. Tighten the cable tie to secure the A3A11 and A3A12 assemblies to the W10 semi-rigid cable. 13.Replace the two screws (1) that secure the front frame to the main chassis. 14.If the instrument came with an A17 Memory Card Reader, replace it. Refer to the A17 Memory Card Reader Assembly. replacement procedure. 15.Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
Chapter 7
301
Removal
1. Remove the front frame from the analyzer. Refer to the A1 Front-Frame Assembly. removal procedure. 2. Place the front frame face down on the work surface. 3. Refer to Foldout 12-5 at the end of Chapter 12 to remove the keyboard assembly. 4. Note that the front-panel menu key actuators are part of the display bezel assembly and are not replaceable. If the menu keys are damaged, replace the bezel assembly.
Replacement
1. Refer to Foldout 12-5 at the end of Chapter 12 to replace the keyboard assembly. 2. Install the front frame to the analyzer. Refer to the A1 Front-Frame Assembly. replacement procedure.
302
Chapter 7
A2 Display
This procedure applies to: 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers
CAUTION
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Remove the A17 Memory Card Reader. Refer to the A17 Memory Card Reader Assembly. removal procedure. 3. Remove the A1 Front-Frame assembly. Refer to the A1 Front-Frame Assembly. removal procedure. 4. Place the instrument on the work surface with the bottom side facing up. 5. Remove the three screws (1) that secure the A2 Display to the chassis. See Figure 7-2. 6. Disconnect the W2 wire harness (2) from A16J8 on the A16 Processor/Video Board assembly. 7. Place the instrument on its right side. 8. Disconnect the W51 wire harness from the power supply and, on 8592L and 8593E spectrum analyzers, from the A7 Analog Control assembly.
CAUTION
The plastic locking clip on the W3 connector (3) is fragile. Protect the connector by wrapping it with a piece of tape before pushing it down into the slot (4) on the instrument chassis. 9. Push the W2 wire harness (2) and the W3 wire harness (3) down into the slot (4) on the instrument chassis. 10.Hold the display in position and remove the two screws (5) that secure the display to the left side of the instrument chassis.
CAUTION
A corner of the A8 Power Supply partially blocks the slot (4) in the instrument chassis. Make sure that the display's wiring harness and connectors do not jam between the power supply and the chassis.
Chapter 7
303
11.Slowly remove the display assembly from the front of the instrument. Continue to route W2 and W3 through the slot (4) as the display is removed. Figure 7-2 A2 Display Replacement
12.Remove the screws and the ber washer (6) that secure the A2 Display to its enclosure (7). See Figure 7-4. 13.Slowly pull the display out of its enclosure while routing W2 and W3 through the opening in the enclosure (8).
304
Chapter 7
WARNING
A high-voltage potential may remain within the A2 Display assembly for some time after it has been removed from the instrument. Do not attempt to remove the post-accelerator lead from the cathode-ray tube. 14.Disconnect W2 (9) and W3 (10) from the display.
Figure 7-3
CRT Assembly
Chapter 7
305
Figure 7-4
Replacement
1. Connect W2 and W51 (8) and W3 (9) to the A2 Display. See Figure 7-4.
CAUTION
The plastic locking clip on the W3 connector (9) is fragile. Protect the connector by wrapping it with a piece of tape before pushing it through the opening (7) on the enclosure. 2. Place the display close to the front of the enclosure and route W2, W51, and W3 out through the opening (7). 3. Slowly push the display into the enclosure while pulling W2, W51, and W3 through the opening (7).
306
Chapter 7
4. Place the instrument on the work surface so that it is resting on its right side. 5. Position the display assembly part of the way into the front of the instrument. 6. Begin to push W2 and W3 out through the slot (4) in the instrument chassis. See Figure 7-2.
CAUTION
A corner of the A8 Power Supply partially blocks the slot (4) in the instrument chassis. Make sure that the display's wiring harness and connectors do not jam between the power supply and the chassis. 7. Replace the display in the instrument while carefully pulling W2 and W3 through the slot (4) in the instrument chassis. 8. Hold the display as far in as possible and replace the two screws (5) that secure the display to the left side of the instrument chassis. 9. Connect the W51 wire harness to the power supply and, on 8592L and 8593E spectrum analyzers, to the A7 Analog Control assembly. See Figure 7-8. 10.Place the instrument on the work surface with the bottom side facing up. 11.Remove the tape protecting the plastic clip on W3 (3). 12.Connect W2 (2) to A16J8 on the A16 Processor/Video Board assembly. 13.Replace the three screws (1) that secure the display to the chassis. 14.Replace the A1 Front-Frame assembly. Refer to the A1 Front-Frame Assembly. replacement procedure. 15.Replace the A17 Memory Card Reader. Refer to the A17 Memory Card Reader Assembly. replacement procedure. 16.Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
Chapter 7
307
RF Assembly
This procedure applies to: 8590L and 8591E spectrum analyzers 8591C cable TV analyzers
CAUTION
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Place the analyzer with the bottom side facing up. 3. Disconnect the W10 semi-rigid cable (2) from the INPUT connector. See Figure 7-5. 4. If the instrument came equipped with a tracking generator, disconnect W31 (8) from the RF OUTPUT. 5. On 8591E spectrum analyzers and 8591C cable TV analyzers, disconnect W20 (7) from counter lock assembly A25A1J1. 6. Remove the four screws (3) securing the RF assembly to the chassis. 7. Place the analyzer with the top side facing up. 8. Clip the tie wraps (9) which secure W8 and W9 to the other cables. 9. Disconnect the W9 cable (4) from A9J4, the 321.4 MHz INPUT. 10.Disconnect the W8 cable (5) from A9J5, the 600 MHz OUT. 11.Remove the four screws (10) securing the A7 Analog board to the chassis. Note that one of these is a pozidrive screw which attaches to the RF assembly. 12.Remove the A7 Analog Interface assembly. Refer to the A7 Analog Interface Assembly. removal procedure. 13.If the instrument came equipped with a tracking generator assembly, disconnect the SMA cables connected to the A7A1 Tracking Generator Control assembly. 14.Remove the bracket (11) which secures the RF assembly to the IF section.
CAUTION
To prevent cable damage, make sure that W10 and W31 are completely disconnected before removing the RF assembly.
308
Chapter 7
To disassemble or assemble the RF assembly, refer to the Foldouts at the end of Chapter 12. Chapter 11 provides cable location information for the RF assembly and the tracking generator assembly. RF Assembly Replacement
Figure 7-5
Chapter 7
309
Replacement
1. Place the analyzer with the top side facing up. 2. Align the W10 semi-rigid cable (2) with the INPUT connector, and W31, if equipped with a tracking generator, with the OUTPUT connector as the Rf assembly is placed into the analyzer. Make sure that none of the cables are pinched while installing the RF assembly. 3. Replace the bracket (11) which secures the RF assembly to the IF section. 4. Replace the A7 Analog Interface assembly (1). Refer to the A7 Analog Interface Assembly. replacement procedure. 5. Connect W9 (4) to A9J4, and W8 (5) to A9J5. 6. Replace the tie wraps (9) which secure W8 and W9 to the other cables. 7. Place the analyzer with the bottom side facing up. 8. Replace the four screws (3) that secure the RF assembly to the analyzer. 9. Tighten the W10 semi-rigid cable (2) on the INPUT connector to 10 inch-pounds. 10.If the instrument came equipped with a tracking generator assembly, tighten the W31 semi-rigid cable (8) on the OUTPUT connector to 10 inch-pounds. 11.Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
310
Chapter 7
A3 Front-End Assembly
This procedure applies to: 8592L spectrum analyzers 8593E spectrum analyzers 8594E spectrum analyzers 8594L spectrum analyzers 8594Q QAM analyzers 8595E spectrum analyzers 8596E spectrum analyzers
General Instructions
CAUTION
The A3 Front-End assembly has static-sensitive components. Read Chapter 15 before proceeding. Hold the A3A3 Low-Pass Filter, A3A9 Bandpass Filter, and FL1 Low-Pass Filter rmly in place when loosening or tightening either of their semi-rigid cables. The cables can be easily damaged if the lters are allowed to twist. The bracket assembly is made up of four separate brackets. Whenever the brackets are called out in a replacement procedure, they are designated by the letters A through D, as illustrated in Figure 7-6. To disassemble the A3 Front-End assembly, refer to the foldouts at the end of Chapter 12. Chapter 11 provides cable location information for the various option and model mixes. The torque specication for all of the SMA connections within the A3 Front-End assembly is 10 inch-pounds. Be sure that the assemblies are securely mounted during initial loosening or nal torquing of the semi-rigid cables.
Chapter 7
311
NOTE
When replacing the A3A8 YTF, closely examine the replacement YTF. It may have its wiring label installed in one of two ways. Either orientation is correct as long as A3A8J2 is connected to the SMA adapter and proper polarity of the tuning wires is observed. Use the following schedule to connect the wires to the replacement YTF: Connect the violet wire to one HTR terminal. Connect the green wire to the other HTR terminal. Connect the orange wire to the Tune terminal. Connect the brown wire to the + Tune terminal.
Figure 7-6
312
Chapter 7
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Remove the front-frame assembly. Refer to the A1 Front-Frame Assembly. removal procedure. 3. Place the instrument so that its left side is resting on the work surface. 4. Disconnect the W20 coaxial cable (1) from AT1, the 10 dB pad. See Figure 7-7. 5. Push W20 through the slot (2) on the instrument chassis. 6. Remove the four screws (3) that secure the A3 Front-End assembly to the instrument chassis. Figure 7-7 A3 Front-End Replacement, Bottom View
Chapter 7
313
7. Place the instrument on the work surface with the top side facing up. 8. Remove the A7 Analog Interface assembly. Refer to the A7 Analog Interface Assembly. removal procedure. 9. Note how the cables attached to the A9 Third Converter assembly are dressed. (This information will be used later during the A3 Front-End Assembly. replacement procedure.
NOTE
Some card-cage options have a different cable dress which include the use of tie-wraps. The tie-wraps are required to hold the wires in place during installation of the instrument cover. 10.Disconnect W8 (1) from A9J5 and W9 (2) from A9J4. See Figure 7-8. 11.If the instrument came equipped with a tracking generator (Option 010), disconnect W39 from A10J1 and W38 from A10J2. 12.If the instrument came equipped with an LO Output on the rear panel (Option 009), disconnect W42 (3) from the LO OUTPUT connector on the rear-frame assembly. 13.Remove the three screws and two washers (4) from the front-end-to-IF support bracket, then remove the bracket. 14.Tuck the cables and wire harness that are attached to the front-end assembly down into the space (5) between the D bracket and the A3A9 lter. This reduces the possibility that wires will be damaged while the front-end assembly is removed. 15.Carefully lift the A3 Front-End assembly out of the instrument.
314
Chapter 7
Figure 7-8
Chapter 7
315
Replacement
1. Tuck all cables and wire harnesses on the A3 Front-End assembly (except for cables W20 and W24) down into the space (5) between the D bracket and the A3A9 lter. This reduces the possibility of damaging wires while replacing the front-end assembly. See Figure 7-8. 2. Place the instrument on the work surface with the top side facing up. 3. Set the front-end assembly on top of the A2 Display assembly and route the W20 coaxial cable into the slot on the instrument chassis that is just below the lower-right rear corner of the A2 Display assembly. 4. Position the front-end assembly over the front-end section of the instrument chassis.
CAUTION
Protect the center pin of the W10 semi-rigid cable and make sure that cables W20 and W24, as well as all other wiring, are not pinched between the front-end assembly and the instrument. 5. Carefully lower the front-end assembly into the instrument chassis. 6. Replace the front-end-to-IF support bracket, and install the three screws and two washers (4) on the bracket without tightening. 7. Connect W8 (1) to A9J5 and W9 (2) to A9J4. See Figure 7-8. 8. If the instrument came equipped with an LO Output on the rear panel (Option 009), connect W42 (3) to the LO OUTPUT connector on the rear-frame assembly. 9. Place the instrument so that its left side is resting on the work surface. 10.Install and tighten the four screws (3) that secure the front-end assembly to the chassis. See Figure 7-7. 11.Route W20 through the slot (2) in the instrument chassis and under the ribbon cable. 12.Connect W20 (1) to the AT1 assembly and tighten it to 10 inch-pounds. 13.Place the instrument on the work surface with the top side facing up. 14.Tighten the three screws (4) on the front-end-to-IF support bracket. See Figure 7-8. 15.Replace the A7 Analog Board assembly. Refer to the A7 Analog Interface Assembly. replacement procedure.
316
Chapter 7
16.Dress all cables as noted during the removal procedure. 17.For card-cage options with additional wiring, replace the tie-wraps and dress the wiring so that there will be no interference with the instrument cover assembly. 18.Replace the front-frame assembly. Refer to the A1 Front-Frame Assembly. replacement procedure. 19.Replace the instrument cover assembly. Refer to the Instrument Cover. replacement.
Chapter 7
317
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Remove the four screws (1), which secure the A7 Analog Interface assembly to the right side of the instrument chassis. See Figure 7-9. 3. Note how the cables and wiring attached to the A7 assembly are dressed. (This information will be used later during the replacement procedure.) 4. Pull the A7 assembly partially out of its slot and disconnect W12, W13, W34, W35, W36, and W51 from the A7 assembly (2). 5. If you have an A7A1 Tracking Generator Control assembly on your 8590L, 8591C, or 8591E, you will need to remove the SMA cables attached to it as you are removing the A7 assembly. 6. Remove the A7 assembly and place it in a static-safe container.
318
Chapter 7
Figure 7-9
Chapter 7
319
Replacement
CAUTION
Serious instrument damage will result if any wire connector is not installed correctly. A connector can be installed backwards or in the wrong position. Inspect each connector and make sure it is not damaged or missing a key plug. Also inspect the A7 assembly for bent connector pins. Incorrect connector installation is most likely to occur with the connectors on W12, W35 and W36. See Figure 7-10. End View of Connectors for W12, W35, and W36
Figure 7-10
1. Lower the A7 assembly part way into the A7 slot and connect W12, W13, W34, W35, W36, and W51 (3) to the A7 assembly. 2. If your 8590L, 8591C, or 8591E has an A7A1 Tracking Generator Control assembly, you will need to reconnect the SMA cables to it as you lower the A7 assembly into the instrument.
NOTE
To ensure proper installation of the A7 assembly, perform the following: a. Arrange all wiring so that the wires do not press against the A7 assembly. b. Align the two tabs on the bottom of the A7 assembly with the slots on the instrument chassis. c. Avoid bending the pins on the A15 motherboard connector while positioning the A7 assembly. d. Align the A7 Analog Interface assembly connector with the A15 motherboard connector carefully.
320
Chapter 7
3. Install the A7 assembly. Slide the A7 assembly toward the back of the A7 slot so that the front-end-to-chassis spacer, mounted on the leading edge of the A7 assembly, will drop smoothly into place. 4. Replace the four screws (1) that secure the A7 assembly to the right side of the chassis. 5. Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
Chapter 7
321
A8 Power Supply
This procedure applies to: 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Remove the instrument rear frame. Refer to the Rear-Frame Assembly. removal procedure. 3. Remove W51 (4) from the A8 Power Supply. 4. Remove the four TORX screws and washers on the left side of the chassis adjacent to the A8 Power Supply. 5. Grasp the power supply pull-tab (1), pull straight up, and remove the A8 assembly from the instrument chassis. See Figure 7-11.
Replacement
1. Position the 37-pin connector (2) on the power-supply so it is able mate with the A15 motherboard connector. 2. Lower the A8 assembly into the instrument chassis. 3. With the connectors properly aligned, push down on top of the A8 assembly close to the pull tab (3). The assembly is correctly installed when the four mounting holes on the left side of the chassis are aligned with the tapped holes on the A8 assembly. 4. Replace the four screws and washers on the left side of the chassis. 5. Replace the rear frame. Refer to the Rear-Frame Assembly. replacement procedure. 6. Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
322
Chapter 7
Figure 7-11
Chapter 7
323
324
Chapter 7
CAUTION
Use the two-prong board-puller tool to remove the IF board assemblies that do not have attached covers. Traces on the printed circuit boards can be damaged if tools with sharp edges are used to remove the IF board assemblies. A15 Motherboard Replacement
Figure 7-12
11.Remove the ve IF board assemblies (8) from the IF extrusion and place them in a static-safe container. 12.Remove the ve screws (9) that secure the A15 motherboard to the instrument chassis.
Chapter 7
325
CAUTION
Do not twist the motherboard sideways while removing it from the instrument. This avoids damaging the connector on the A16 Processor/Video Board assembly that is mated with the connector on the back side of the A15 motherboard. 13.Pull the motherboard straight up and out of the instrument chassis. If the motherboard is difcult to remove, a slight rocking action, from front to rear, will help it disconnect from the connector on the processor/video board assembly. 14.If the instrument is equipped with a card-cage: Remove the four screws (10) that secure the card-cage bracket to the sides of the IF extrusion. See Figure 7-13. Turn the motherboard assembly upside down on the work surface and remove the two screws (11) that secure the card-cage bracket to the motherboard. Remove the bracket. 15.Remove the 40 screws (12) that secure the IF extrusion assembly to the motherboard.
NOTE
If you are replacing a defective A15 motherboard, continue at the Replacement of the A15 Motherboard. procedure. 16.Remove the motherboard and place it in a static-safe container.
326
Chapter 7
Figure 7-13
Chapter 7
327
NOTE
The IF extrusion assembly contains one at extrusion end-section. Make sure that the extrusion assembly is positioned with the at extrusion end-section next to the A8 Power Supply connector (18). See Figure 7-13. 19.Turn the A15 motherboard upside down and align it with the IF extrusion assembly. 20.Replace the 40 screws (12) that secure the IF extrusion assembly to the motherboard. 21.If the instrument came equipped with a card-cage: a. Replace the card-cage bracket and align it with the two holes on the motherboard. b. Replace the two screws (11) that secure the card-cage bracket to the motherboard without tightening. See Figure 7-13. c. Turn the motherboard assembly right side up and replace the four screws (10) that secure the card-cage bracket to the sides of the IF extrusion assembly. d. Tighten the two screws (11) on the bottom-side of the motherboard.
CAUTION
Do not twist the motherboard sideways while inserting its connector into the A16 board assembly connector. This prevents damage to either connector. 22.Lower the motherboard assembly into the instrument chassis and align it with the connector on the processor/video board assembly. 23.If it is difcult to insert the motherboard connector into the processor/video connector, a slight rocking action, from front to rear, will help it drop into place. 24.Make sure that the holes in the motherboard assembly line up with the holes in the instrument chassis.
NOTE
If the holes do not line up, loosen the screws that secure the A16 Processor/Video Board assembly to the instrument chassis. Refer to the A16 Processor/Video Board and A16A1 Memory Board Assembly. replacement procedure. 25.Replace the ve screws (9) that secure the A15 motherboard to the instrument chassis. 26.Replace the ve IF board assemblies (8) in the IF extrusion. 27.Place each IF cover over its corresponding IF board assembly.
328
Chapter 7
28.Replace the 40 screws (7) that secure the IF board assemblies to the IF extrusion. 29.Connect the B1 Fan wire harness (6) to A15J19 at the rear of the motherboard. 30.Connect the W19 coaxial cable (5) on A15J18 at the rear of the motherboard. 31.Replace the two screws and at washers (4) on the front-end-to-IF support bracket. See Figure 7-12. 32.Replace the rear frame. Refer to the Rear-Frame Assembly. replacement procedure. 33.Replace the A7 Analog Board assembly. Refer to the A7 Analog Interface Assembly. replacement procedure. 34.Replace W7, W8, W9, and W21 (1) on the A9 Third Converter. 35.If the instrument is equipped with card-cage options: Replace the option assemblies in the card-cage slot that they were removed from. Connect all option cables and wire harnesses as noted in the removal procedure. Replace all cable ties as noted in the removal procedure. Position all wiring and cables so that they do not interfere with the installation of the instrument cover. 36.Replace the A8 Power Supply. Refer to the A8 Power Supply. replacement procedure. 37.Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
Chapter 7
329
Replacing Major Assemblies A16 Processor/Video Board and A16A1 Memory Board Assembly
Before Replacing the A16 Processor/Video Board Assembly or the A16A1 Memory Board Assembly
Frequency response and cal attenuator correction constants (for the A12 amplitude control board assembly) are stored in battery-backed RAM on the A16A1 memory board assembly. If replacing the A16A1 memory board, record these correction constants using the procedures in Chapter 3 prior to removing the A16A1 memory board assembly. If not replacing the A16A1 memory board, remove the board by prying the board off the standoffs and place in a static-safe bag.
NOTE
If you are unable to record the current correction constants, or they are invalid, new correction constants must be generated after replacing the A16A1 memory board. Refer to the 10 MHz Reference, Frequency Response, and Cal Attenuator Error Correction adjustment procedures in Chapter 2 of this manual. If valid correction constants can be removed from analyzer memory before replacing the A16A1 memory board, record the correction values using the procedures in Chapter 3.
CAUTION
330
Chapter 7
Replacing Major Assemblies A16 Processor/Video Board and A16A1 Memory Board Assembly
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Place the analyzer on the work surface with the bottom side facing up. 3. If the instrument is equipped with Option 041 or Option 043: Remove the ve screws securing the IO connector to the IO board. Remove the four screws (4) that secure the A40 or A41 IO board assembly. Remove the A40 or A41 I0 board assembly (5). 4. Detach the instrument rear frame from the chassis. Pull the rear frame back so that the A16 Processor/Video Board assembly is completely exposed. 5. Disconnect the W14 ribbon cable (1) from A16J2 and A25J4. See Figure 7-14. 6. Remove the four screws (2) that secure the A25 Counter-Lock assembly. 7. Remove the A25 assembly (3). 8. If the instrument is equipped with Option 041 or Option 043. 9. Remove the seven screws (6) that secure the A16 assembly.
CAUTION
The A16 assembly may be damaged if it is placed on a conductive surface. Use a static-safe workstation. Ensure that the assembly is not placed on any conductive material. 10.Remove the A16A1 Memory board assembly and place in a static-safe container. 11.Remove the A16 assembly and place it in a static-safe container.
Chapter 7
331
Replacing Major Assemblies A16 Processor/Video Board and A16A1 Memory Board Assembly
Figure 7-14
332
Chapter 7
Replacing Major Assemblies A16 Processor/Video Board and A16A1 Memory Board Assembly
Replacement
CAUTION
To prevent damage to connectors A16J1 and A15J2, make sure that they are mated correctly before pushing the A16 assembly into place. Avoid twisting the A16 assembly from side to side during installation. 1. Replace the A16 processor/video board assembly. 2. Place two screws in the mounting holes by J5 and J7. Align the A16 board as far toward the rear panel as it will go. 3. Replace the remaining ve screws (6) that secure the A16 assembly. See Figure 7-14. 4. If the instrument is equipped with Option 041 or 043: Replace the A40 or A41 IO board assembly (5). Replace the four screws (4) that secure the A40 or A41 IO board assembly. 5. Replace the A25 counter-lock assembly (3). 6. Replace the four screws (2) that secure the A25 assembly. 7. Connect the W14 ribbon cable (1) to A16J2 and A25J4.
CAUTION
To prevent cable and wire damage, ensure all wiring is routed to one side of the SMB connector, A15J18, and through the notch located on the rear edge of the A15 motherboard. 8. Replace the rear-frame assembly. 9. Install the IO connector through the rear panel into the IO board and replace the ve screws securing the IO connector to the IO board. 10.Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
Chapter 7
333
Replacing Major Assemblies A16 Processor/Video Board and A16A1 Memory Board Assembly
334
Chapter 7
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Remove the A25 Counter-Lock assembly. Refer to the A25 Counter-Lock Assembly. removal procedure. 3. Note the location of the four rmware ROMs: A16U6, A16U7, A16U23, and A16U24. Refer to Figure 7-15. 4. Carefully pry the ROMs upward using a small-blade screwdriver. Pull the ROMs from their sockets evenly so that the IC pins (1) at the rear of each ROM are not bent.
Replacement
CAUTION
To prevent damage to the ROMs, make sure that the ROMs are oriented correctly with the sockets on the A16 assembly. Match the notch (2) at the end of each ROM with the notch at the end of its socket. Align each ROM IC pin with the opening of its pin receptacle on the socket. 1. Position each ROM on the A16 assembly as illustrated in Figure 7-15. 2. Install each ROM by carefully pushing down on both ends of the ROM with equal pressure. 3. Replace the A25 Counter-Lock assembly. Refer to the A25 Counter-Lock Assembly. replacement procedure. 4. Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure. 5. Continue at the Adjustments Required after ROM Replacement. procedure.
Chapter 7
335
Figure 7-15
A FREQ UNCAL message may appear on screen but will disappear after the successful completion of the self-calibration routines. 4. Perform the following adjustments from Chapter 2. a. CAL FREQ Adjustment Routine b. CAL AMPTD Adjustment Routine c. CAL YTF Adjustment Routine (8592L, 8593E, 8595E, and 8596E only)
336
Chapter 7
A16A1BT1 Battery
This procedure applies to: 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers
CAUTION WARNING
Use ESD precautions when performing this replacement procedure. Battery A16A1BT1 contains lithium iodide. There is a danger of explosion if the battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Do not incinerate or puncture this battery. Discard used batteries according to the manufacture's instructions.
Removal/Replacement
CAUTION
The A16A1 memory board assembly may be damaged if it is placed on a conductive surface. Use a static-safe workstation. Ensure that the memory board assembly is not placed on any conductive material. 1. Remove the A16A1 memory board assembly by lifting it up from the A16 processor/video board.
CAUTION
Placing the bottom side of the A16A1 assembly on a conductive work surface can result in loss of correction data stored in RAM. Protect the A16A1 assembly from contact with any conductive surface. 2. Remove the nylon insulator from the bottom of the A16A1 board assembly. 3. Locate the battery leads on the A16A1 assembly, desolder them, and remove the old battery.
Chapter 7
337
NOTE
A16A1C1 maintains the voltage required to back up RAM for a maximum of 8 hours. A16A1C1 is the large round capacitor next to A16A1BT1. 4. Replace and solder the leads of the new battery, ensuring proper polarity as silkscreened on the A16A1 board assembly. 5. Record the battery-replacement date on the battery label located on the analyzer rear frame. 6. Replace the A16A1 memory board assembly onto the A16 processor/video board. Refer to the A16 Processor/Video Board and A16A1 Memory Board Assembly. replacement procedure. 7. Connect the power cord to the analyzer and turn the power on for about ve minutes. This allows the supercap to fully charge and insures the battery switching circuitry is in the correct mode.
338
Chapter 7
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Place the analyzer on the work surface with the bottom side facing up. 3. Disconnect the W14 ribbon cable (1) from the A17 Memory Card Reader assembly. See Figure 7-16. 4. Remove the jumper (5) between the A17 Memory Card Reader and LS1. See Figure 7-16. 5. Remove the two screws (2) that secure the A17 assembly to the chassis.
CAUTION
When removing the A17 assembly, be sure to slide the assembly straight backward to prevent damage to the two alignment tabs on the front of the A17 support bracket. 6. Remove the A17 assembly (3), gently sliding the assembly towards the rear of the analyzer and up. 7. Remove the four screws (4) that secure the A17 assembly to the memory-card base and remove the A17 assembly.
Chapter 7
339
Replacement
1. Place the analyzer on the worksurface with the bottom side facing up. 2. Place the A17 Memory Card Reader assembly on the memory-card base. 3. Replace the four screws (4) that secure the A17 assembly to the memory-card base.
CAUTION
The two alignment tabs on the memory-card base beneath the A17 Memory Card Reader assembly can be easily damaged. When replacing the A17 Memory Card Reader assembly, carefully mate the two alignment tabs with the memory-card bezel on the front-frame assembly. 4. Replace the A17 assembly (3), gently sliding the assembly along the chassis toward the memory-card reader bezel on the front-frame assembly. 5. Replace the two screws (2) that secure the A17 assembly to the chassis. 6. Replace the jumper (5) between the A17 Memory Card Reader and LS1. 7. Connect the W14 ribbon cable (1) to the A17 assembly. 8. Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
340
Chapter 7
Figure 7-16
Chapter 7
341
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Place the analyzer on the worksurface with the bottom side facing up. 3. Disconnect W20 (1) from the counter-lock assembly. See Figure 7-17. 4. Disconnect the cables from A25J3 and A25J5 (2). 5. Disconnect W4 (4) from the counter-lock assembly. 6. Remove the four mounting screws (5). 7. As you remove the counter-lock assembly, disconnect the cables from A25J1 and A25J2 (3). 8. Place the counter-lock in a static-safe container.
Replacement
1. As you replace the counter-lock assembly, reconnect the cables to A25J1 and A25J2 (3). See Figure 7-17. 2. Replace the four mounting screws (5). 3. Connect W20 (1) to the counter-lock assembly. 4. Connect the cables to A25J3 and A25J5 (2). 5. Connect W4 (4) to the counter-lock assembly. 6. Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure.
342
Chapter 7
Figure 7-17
Chapter 7
343
B1 Fan
This procedure applies to: 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers
Removal
1. Disconnect the analyzer from ac power. 2. Place the instrument on the worksurface with the back end facing out. 3. Hold the B1 Fan assembly in position and loosen the four screws (1) that secure the fan assembly to the rear-frame assembly. See Figure 7-18. 4. Pull the fan 2 inches away from the instrument, reach into the opening in the rear frame, and disconnect the fan wiring from A15J19 (2). 5. Remove the four screws, washers, and spacers from the fan (1). 6. Remove the fan grill (3).
344
Chapter 7
Figure 7-18
B1 Fan Replacement
Replacement
1. Position the fan grill on the B1 Fan (3). See Figure 7-18. The fan label must be visible through the fan grill to make sure that the fan provides the correct air ow. 2. Position the four spacers on the fan and replace the screws and washers (1). 3. Hold the fan 2 inches away from the instrument and connect the fan wiring to A15J19 (2). 4. Position the fan assembly on the instrument and tighten the four screws (1) that secure it to the rear frame.
Chapter 7
345
Rear-Frame Assembly
This procedure applies to: 8590 E-Series and L-Series spectrum analyzers 8591C cable TV analyzers 8594Q QAM analyzers
CAUTION
Removal
1. Remove the instrument cover assembly. Refer to the Instrument Cover. removal procedure. 2. Place the analyzer with the top side facing up. 3. Remove the six screws (three on each side) that secure the rear frame to the main chassis. 4. Remove the W6 jumper from the 10 MHz REF OUTPUT and the EXT REF IN connectors. 5. Pull the rear frame straight back away from the main chassis.
NOTE
Different models and option mixes will cause the conguration of the cables and wires to the rear frame to vary widely. Therefore, you will need to note these connections for reference during reassembly. See Figure 7-19. 6. Disconnect the fan assembly from J19 on the A15 IF Motherboard assembly. 7. Disconnect the wires and cables from the rear-frame assembly.
Replacement
1. Connect the wires and cables to the rear-frame assembly as noted during the removal procedure. See Figure 7-18. 2. Connect the fan assembly to J19 on the A15 IF Motherboard assembly.
CAUTION
Wiring can be pinched between the rear frame and the edge of the A15 motherboard. Make sure that all wiring is routed to one side of the SMB connector, A15J18, and through the notch located on the rear edge of the motherboard.
346
Chapter 7
3. Place the rear frame on the main chassis. 4. Replace the W6 jumper between the 10 MHz REF OUTPUT and the EXT REF IN connectors. 5. Replace the six screws (three on each side) that secure the rear frame to the main chassis. 6. Replace the instrument cover assembly. Refer to the Instrument Cover. replacement procedure. Figure 7-19 Rear Frame Replacement
Chapter 7
347
348
Chapter 7
349
350
Chapter 8
Chapter 8
351
If You Have A Problem Calling Agilent Technologies Sales and Service Ofces
352
Chapter 8
If You Have A Problem Calling Agilent Technologies Sales and Service Ofces
A serial-number label is attached to the rear of the analyzer. This label has two instrument identication entries: the rst provides the instrument's serial number and the second provides the identication number for each option built into the instrument. The serial number is divided into three parts. The rst four digits are the serial-number prex; the letter indicates the country of origin; the last ve digits are the sufx. The serial-number label contains a three-digit option number for each option built into the analyzer. Each option number is entered sequentially below the serial number. See Table 8-1. Figure 8-1 Typical Serial-Number Label
The serial-number prex is a code that identies the date of the last major design change that is built into the analyzer. The letter identies the country where the instrument was manufactured. The ve-digit sufx is a sequential number and is different for each instrument.
Chapter 8
353
If You Have A Problem Calling Agilent Technologies Sales and Service Ofces
354
Chapter 8
Analyzer damage can result from using packaging materials other than those specied. Never use styrene pellets in any shape as packaging materials. They do not adequately cushion the instrument or prevent it from shifting in the carton. Styrene pellets cause equipment damage by generating static electricity and by lodging in the analyzer fan. 2. Use the original packaging materials or a strong shipping container that is made of double-walled, corrugated cardboard with 159 kg (350 lb) bursting strength. The carton must be both large enough and strong enough to accommodate the analyzer and allow at least 3 to 4 inches on all sides of the analyzer for packing material. 3. Surround the instrument with at least 3 to 4 inches of packing material, or enough to prevent the instrument from moving in the carton. If packing foam is not available, the best alternative is SD-240 Air Cap from Sealed Air Corporation (Commerce, CA 90001). Air Cap looks like a plastic sheet covered with 1-1/4 inch air-lled bubbles. Use the pink Air Cap to reduce static electricity. Wrap the instrument several times in the material to both protect the instrument and prevent it from moving in the carton. 4. Seal the shipping container securely with strong nylon adhesive tape. 5. Mark the shipping container FRAGILE, HANDLE WITH CARE to ensure careful handling. 6. Retain copies of all shipping papers.
Chapter 8
355
356
Chapter 8
357
Assembly Descriptions and Block Diagrams Assembly Descriptions and Block Diagrams
358
Chapter 9
Assembly Descriptions and Block Diagrams RF and LO Section for 8590L, 8591E, and 8591C Spectrum Analyzers
J1 Input Connector
The analyzer RF section receives input signals via a 50 input, or 75 input. To protect the attenuator and input mixer from damage, refer to Analyzer Input Protection in Chapter 1 for maximum input level requirements.
A3 Input Attenuator
The A3 input attenuator has a 0 to 60 dB range that provides input attenuation in 10 dB steps. Pressing PRESET or analyzer power-on selects 10 dB attenuation. If POWERON LAST is selected, the analyzer is reset to the settings in use when the analyzer was turned off.
A4 First Converter
The A4 rst converter assembly converts incoming signals to a 2.1214 GHz rst IF by mixing the 2.1214 GHz to 3.9214 GHz LO signal from the A6 YTO assembly with the input signal. The rst IF signal is the summed signal from the mixing process. For analyzers that are not equipped with a tracking generator, Option 010 or Option 011, the J4 output for the tracking generator is terminated.
Chapter 9
359
Assembly Descriptions and Block Diagrams RF and LO Section for 8590L, 8591E, and 8591C Spectrum Analyzers
A5 Second Converter
The 2.1214 GHz rst IF signal is downconverted by the A5 second converter to a 321.4 MHz second IF. This is accomplished by mixing the rst IF signal with the 1.8 GHz second LO signal. The second LO signal is produced by multiplying the 600 MHz signal from the A9 Third Converter by three. The 2.1212 GHz bandpass lter is a high Q lter that eliminates unwanted mixing products. Its performance is due to the resonant characteristics of three precisely machined cavities in an aluminum block. The 321.4 MHz matching lter is an adjustable bandpass lter that prevents unwanted mixing products from passing on to the A9 third converter.
A9 Third Converter
The A9 third converter performs ve functions: 1. Produces the 300 MHz CAL OUT signal. The CAL OUT signal is rich in harmonics; these are required for the CAL FREQ self-calibration routine. 2. Provides the 600 MHz reference to the multiplier on the A5 second converter assembly. 3. Sends a 300 MHz signal to the A25 counterlock assembly. The A25 assembly locks the 300 MHz signal to the 10 MHz frequency reference. Note that this step applies to all 8591C, 8591E and 8590L analyzers. 4. Converts the 321.4 MHz IF signal to the nal 21.4 MHz IF signal. The 321.4 MHz IF signal is rst amplied and ltered by the 321.4 MHz bandpass lter. The 600 MHz reference is divided to produce the 300 MHz third LO. The ltered 321.4 MHz IF signal and the 300 MHz third LO mix at the third mixer, a double-balanced mixer, to produce the 21.4 MHz IF. The 21.4 MHz IF passes through a 21.4 MHz bandpass lter, a buffer amplier with a nominal 19 dB of gain, and a variable IF calibration amplier. The variable amplier has a gain range of 15 to +2 dB.
360
Chapter 9
Assembly Descriptions and Block Diagrams RF and LO Section for 8590L, 8591E, and 8591C Spectrum Analyzers
5. Changes the amplitude of the 21.4 MHz IF signal to calibrate the analyzer during the CAL AMPTD self-calibration routine. The amplitude of the detected 21.4 MHz IF signal is measured on the A16 processor/video assembly. The A7 analog interface assembly adjusts the gain of the IF calibration amplier based on the A16 amplitude measurement. The CAL AMPTD adjustment routine adjusts the output of the A9 IF calibration amplier for a displayed amplitude of 20 dBm at 300 MHz. The analyzer is set for a reference level of 20 dBm and a center frequency of 300 MHz. (The calibration routine uses the CAL OUT signal.)
A25 Counterlock
The A25 counterlock assembly performs four main functions: Phase-locks the 600 MHz SAW oscillator on the A9 third converter to the 10 MHz reference. Counts the rst LO frequency. Provides discriminator output, DISCRIM, to the A7 analog interface assembly. Counts the 21.4 MHz IF. Refer to Figure 9-5, at the end of this chapter for details about the operation of the A25 counterlock assembly. On the A9 third converter, the output of the 600 MHz oscillator is divided by two and the resulting 300 MHz signal is routed to the A25 counterlock assembly. The A25 counterlock assembly divides the 300 MHz signal by 40 to generate a 7.5 MHz reference for the sampling oscillator, stabilizer, and frequency counter. The 7.5 MHz reference and the 10 MHz reference are divided further and compared in a phase-frequency detector. The output of the phase-frequency detector, VTO_TUNE, is fed back to A9 to tune the 600 MHz oscillator.
Chapter 9
361
Assembly Descriptions and Block Diagrams RF and LO Section for 8590L, 8591E, and 8591C Spectrum Analyzers
The sampling oscillator provides a 279 to 298 MHz driving signal to the A25A1 sampler. The rst LO signal is also applied to A25A1 sampler. The rst LO signal is mixed with a harmonic of the sampling oscillator signal to generate the sampler IF. This IF is divided by 10 and fed to one input of the frequency counter. The equation used to produce the rst LO frequency is:
The stabilizer mixes the divided sampler IF (nominally 8.25 MHz) with the 7.5 MHz, and feeds the difference signal into the discriminator. The discriminator output, DISCRIM, is fed back to the A7 analog interface assembly to tune the A3A7 YTO precisely to the center frequency. When the frequency count marker is active, the analyzer pauses at the marked frequency and counts the rst LO as described previously. The 21.4 MHz IF is then divided by four, to a nominal value of 5.35 MHz, fed to another frequency counter input, and counted. With the second LO, the third LO, the counted rst LO, and the counted nal IF all referenced to the 10 MHz reference, the actual input signal frequency is calculated.
A25A1 Sampler
The A25A1 sampler mixes the rst LO with a harmonic of the sampling oscillator to produce the 76 MHz to 89 MHz sampler IF output. In the phase-locking process, the sampler IF is counted and the YTO tune DACs on the A7 analog interface assembly are adjusted until the actual sampler IF frequency is equal to the desired sampler IF. In YTO spans of 10 MHz or less, the sampler IF is also applied to a discriminator. The output of the discriminator (DISCRIM) ne-tunes the YTO precisely to the center frequency.
362
Chapter 9
Assembly Descriptions and Block Diagrams RF and LO Section for 8590L, 8591E, and 8591C Spectrum Analyzers
A7 Analog Interface
The A7 analog interface assembly receives digital control input on the I/O bus control lines from the A16 processor/video assembly, and produces analog control signals for most of the analyzer functions. This section describes several of the control functions illustrated on Figure 9-1. The A7 analog interface assembly provides control voltages for the Sweep Generator and Span Dividers: The SWEEP RAMP output is 0 to +10 V. The RAMP output is 10 V to +10 V for any span. Two SPAN signals, MAIN SPAN and FM SPAN, are generated by attenuating the RAMP signal. The MAIN SPAN and FM SPAN signals are fed to the YTO drivers to sweep the YTO frequency. The amplitude of the SPAN signals depends on the span setting selected. If the LO span is greater than 10 MHz, then the MAIN SPAN signal is used. If the LO span is less than or equal to 10 MHz, then the FM SPAN signal is used.
Chapter 9
363
Assembly Descriptions and Block Diagrams RF and LO Section for 8590L, 8591E, and 8591C Spectrum Analyzers
The A7 analog interface assembly controls the center frequency of the YTO: The YTO tune DACs in the A7 YTO driver circuitry determine the center frequency of the YTO. For LO spans greater than 10 MHz, the MAIN SPAN signal is summed with the output of the YTO tune DAC in the YTO main coil driver. For LO spans less than or equal to 10 MHz, the FM SPAN signal is summed with the A25 DISCRIM signal in the FM coil driver. The DISCRIM signal is used to lock the YTO precisely to the center frequency. The control voltages produced using either SPAN signal are then converted to currents that drive the A6 YTO.
364
Chapter 9
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
The 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E analyzer RF and LO section includes the following assemblies: A3A1 comb generator (except 8595E) A3A2 RF switch (8592L and 8593E only) A3A3 2.9 GHz LPF A3A4 second converter A3A5 input attenuator A3A6 dual-band mixer (low-band mixer for 8594E, 8594L, and 8594Q) A3A8 YTF (YIG-tuned Filter) (except 8594E, 8594L, and 8594Q) A3A8 SYTF (Switched YIG-tuned lter) (8595E and 8596E only) A3A9 321.4 MHz low-pass lter A3A11 step recovery diode A3A12 3 dB pad A3A15 tracking generator (Option 010) A7 analog interface A9 third converter FL1 3.9214 GHz low-pass lter For details about the operation of the 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E, RF section, refer to Figure 9-3 and Figure 9-4, located in the back of this chapter. The RF section converts all input signals to a xed 21.4 MHz IF. The microcircuits in the RF section are controlled by signals from the A7 analog interface assembly. The A7 analog interface assembly also includes circuitry for controlling the LO and IF sections. Five frequency bands cover the 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E spectrum analyzers input range.
Chapter 9
365
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
Band 0 1 2 3 4 5
Frequency Range 9 kHz to 2.9 GHz 2.75 to 6.5 GHz 6.0 to 12.8 GHz 12.4 to 19.4 GHz 19.1 to 22 (except Option 026 and 027) 19.1 to 26.5
Analyzer Model 8592L, 93E, 94E, 94L, 94Q, 95E, 96E 8592L, 93E, 95E, 96E 8592L, 93E, 96E 8592L, 93E 8592L, 93E 8592L, 93E (Option 026 and 027)
Band 0 (low band) uses triple conversion to produce the nal 21.4 MHz IF. The A3A6 dual band mixer up-converts the RF input to a 3.9214 GHz rst IF. The A3A4 second converter down-converts the 3.9214 GHz rst IF to a 321.4 MHz second IF. The A9 third converter down-converts the second IF to the nal 21.4 MHz IF. Bands 1 through 4 (high bands) use double conversion. The A3A6 dual-band mixer down-converts the RF input to a 321.4 MHz rst IF. Although this rst IF passes through the A3A4 second converter, it bypasses the second mixer. The second and nal conversion occurs in the A9 third converter where the rst IF is down-converted to produce the nal 21.4 MHz IF.
LO Section for the 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
The LO Section includes the following assemblies: A3A7 YTO (YIG-Tuned Oscillator) A3A10 directional coupler A3A13 isolator A7 analog interface A9 third converter A22 precision frequency reference A25 counterlock (not available for the 8592L) AT1 10 dB pad
Refer to Figure 9-5 at the end of this chapter for details about the operation of the LO section.
366
Chapter 9
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
The LO section provides a 3.0 to 6.8214 GHz rst LO, a 600 MHz second LO, a 300 MHz third LO, a 300 MHz CAL OUT signal, and an IF frequency counter. The counter and the local oscillators use one of two 10 MHz frequency references: the standard room temperature crystal oscillator (RTXO) or the A22 precision frequency reference, oven-controlled crystal oscillator (OCXO). The 10 MHz reference phase-locks the 600 MHz oscillator on the A9 third converter. The 600 MHz signal drives the second converter and is divided to produce the 300 MHz third LO and CAL OUT signals. A 300 MHz signal that is sent to the A25 counterlock assembly is divided down further to produce a 7.5 MHz reference signal. This 7.5 MHz reference is used by the stabilizer, the sampling oscillator, and frequency counter. The phase-frequency detector is located on the A25 counterlock assembly. The rst LO output of the A3A7 YTO is fed through the A3A13 isolator to the A3A10 directional coupler. The main output of the coupler is sent to the A3A6 dual-band mixer (low-band mixer for the 8594E and 8594Q). The coupled output is fed through the AT1 10 dB pad to the A25A1 sampler. The sampler mixes the rst LO with a harmonic of the sampling oscillator to generate a 60 to 100 MHz sampler IF. This value is then divided by 10, producing a 6 to 10 MHz output. During retrace, the YTO is locked to the selected rst LO frequency. The divided sampler IF is then counted in the frequency counter. The YTO tune DACs on the A7 analog interface assembly are adjusted until the counted frequency is equal to the desired frequency. In LO spans less than or equal to 10 MHz, the divided sampler IF is also fed to a stabilizer, which generates the DISCRIM (discriminator) signal. DISCRIM tunes the YTO precisely to the desired rst LO frequency. When a trigger signal occurs, a sweep ramp is applied to the main coil drivers while in LO spans greater than 10 MHz, or to the FM coil drivers while in LO spans less than or equal to 10 MHz. During a main coil sweep, the sampling oscillator is disconnected from the A25A1 sampler. In frequency count mode, the rst LO sweeps up to the marked signal and pauses. The divided sampler IF is counted to calculate the actual rst LO frequency. Then the prescaled 21.4 MHz IF, nominally 5.35 MHz, is counted and the actual input frequency calculated.
Chapter 9
367
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
368
Chapter 9
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
The buffered 21.4 MHz IF signal is amplied in the IF calibration amplier (IF Cal Amp). The gain of the calibration amplier is controlled by a DAC on the A7 analog interface assembly via the REF_LVL_CAL control line. During the CAL AMPTD routine, the gain of the calibration amplier is adjusted so that the reference level at top-screen is calibrated. The amplitude reference for the routine is provided by the 20 dBm CAL OUT signal with 10 dB of input attenuation in band 0. Refer to IF Power-Level Measurement in Chapter 5 for more information about the 21.4 MHz output from the A9 third converter assembly.
Chapter 9
369
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
The stabilizer mixes the divided sampler IF (nominally 8.25 MHz) with the 7.5 MHz, and feeds the difference signal into the discriminator. The discriminator output, DISCRIM, is fed back to the A7 analog interface assembly to tune the A3A7 YTO precisely to the center frequency. When the frequency count marker is active, the analyzer pauses at the marked frequency and counts the rst LO as described previously. The 21.4 MHz IF is then divided by four, to a nominal value of 5.35 MHz, fed to another frequency counter input, and counted. With the second LO, the third LO, the counted rst LO, and the counted nal IF all referenced to the 10 MHz reference, the actual input signal frequency is calculated.
370
Chapter 9
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
Tracking Oscillator
The tracking oscillator enables the ne adjustment of the tracking generator output frequency to compensate for the frequency inaccuracies of the analyzer 21.4 MHz IF. The tracking oscillator determines the residual FM and frequency drift of the tracking generator. The 184.28 MHz output frequency is obtained by doubling the output of a crystal oscillator operating at 92.14 MHz.
Upconverter
The upconverter mixes the tracking oscillator output with the buffered 600 MHz reference from the A9 third converter assembly. The upconverter also contains a lter to pass only the 784.28 MHz upper sideband.
Pentupler
The pentupler multiplies the 784.28 MHz signal by ve to generate 3.9214 GHz, the analyzer rst IF in low band. A dual cavity bandpass lter centered at 3.9214 GHz eliminates all unwanted multiples of 784.28 MHz.
Modulator
The output of the pentupler is passed through a modulator to adjust the power level into the output mixer. The modulator is controlled by an ALC circuit on the bias board, which is fed by a detector in the output amplier. If the detected output power is too high, the ALC will drive the modulator to decrease the input level into the output mixer, resulting in a decrease in output power.
Coupler
The rst LO signal from the A3A14 rst LO distribution amplier assembly is coupled off, then buffered to drive the output mixer. The main line of the coupler is fed to the LO OUTPUT connector on the rear panel. The loss through the coupler main line is less than 2.5 dB.
Output Mixer
The 3.9214 GHz signal from the modulator is fed into the RF port of the output mixer. The LO port of the output mixer is driven by the buffered rst LO signal from the coupler. The output of the mixer is then amplied.
Chapter 9
371
Assembly Descriptions and Block Diagrams RF and LO Section for 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E Analyzers
Output Amplier
The output amplier lters the signal emerging from the output mixer and then amplies it into a usable range. The amplier also contains a detector for leveling the output.
Bias Board
The bias board contains the ALC circuitry for the tracking generator and distributes dc power from the A10 tracking generator control assembly to the rest of the tracking generator. The ALC inputs come from the A10 tracking generator control assembly (for controlling the power level), the EXT ALC INPUT line, and the detector in the output amplier. The ALC loop drives the modulator.
372
Chapter 9
IF Section
This section applies to all 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers, and describes the operation of the IF section assemblies and the A16 processor/video assembly. The related operation of the A7 analog interface assembly and the A8 Power Supply are also described. Figure 9-1 illustrates the assembly descriptions in this section. The A15 motherboard pin designation of related assemblies along the IF signal path are detailed on Figure 9-1. Refer to Chapter 5 for more information about the connection of analyzer assemblies to the motherboard.
Chapter 9
373
374
Chapter 9
Chapter 9
375
Final processing of the detected 21.4 MHz IF signal before the video signal is converted by the ADC for further digital processing by the central processing unit (CPU). t Video bandwidths from 30 Hz to 3 MHz are available in a 1, 3, 10 sequence. t The ADC input MUX selects the positive-peak detector, or bypasses the positive-peak detector, and selects the sample detector. In sample mode, the video signal passes directly to the ADC from the video bandwidth circuitry. t The MUX can also select the processed video signal from an assembly in the card cage. Mathematical offset of the digitized video signal for greater reference-level resolution and analyzer calibration accuracy. Digital control of analyzer assemblies directly over the IO bus. Analog control of analyzer assemblies via the A7 analog interface assembly. Nonvolatile RAM memory-storage of DLP software, analyzer calibration data, and error correction data. Refer to Chapter 13 for more information about analyzer calibration and error correction. Processing and integration of trace and text information for output to the A2 display assembly. The digitized video signal is merged by the CPU with other trace information. The trace information is then combined with text information for input to the display drive circuitry. Generation of the A2 display drive signals. The digital display input is converted back into analog voltages by the A16 display drive-circuitry and sent to the A2 assembly. The display signal is also sent to MONITOR OUTPUT on the rear panel.
376
Chapter 9
A7 Analog Interface
The A7 analog interface assembly converts the digital commands from the A16 assembly to analog control signals for the following assemblies shown on Figure 9-1: DAC control of the A11/A13 bandwidth lter assemblies. Refer to Replacing Major Assemblies in Chapter 7 for more information about bandwidth switching. t LC to crystal mode switching. BW5 controls switching between crystal and LC bandwidth lter modes. t Bandwidth control. The A7 bandwidth control DACs supply two control lines to drive the PIN diodes on the A11 and A13 assemblies. Companding DACs are used because their nonlinear output compensates for the nonlinear resistance-versus-current of the PIN diodes they control. BW6 control line. Controls the crystal bandwidths from 30 kHz to 1 kHz. More DAC current produces a narrower bandwidth in crystal mode. BW7 control line. Controls the LC bandwidths from 5 MHz to 100 kHz. In LC mode, more DAC current produces a wider bandwidth. Bandwidth error is corrected by the CAL AMPTD self-calibration routine. Refer to Chapter 13 for a description of CAL AMPTD. A12 amplitude control assembly. When the reference level is changed, the A7 assembly switches the calibration attenuators and the step gains on the A12 assembly to change the displayed signal position. Refer to Replacing Major Assemblies in Chapter 7 for more information about switching for the calibration attenuator, the step gains, and the log/linear switch. t Calibration attenuators. A TTL high on selected A7 control lines activates a combination of attenuator steps. The 1, 2, 4, and 8 dB step attenuators are combined to provide attenuation in 1 dB increments. Currently, the 16 dB attenuator is not used. t 10 dB step gains. Temperature-compensated control voltages activate a combination of the three A12 step-gain stages. Step gains are produced in 10 dB increments over a 50 dB range. Step gain errors are corrected by the CAL AMPTD self-calibration routine. Refer to Chapter 13 for a description of CAL AMPTD.
Chapter 9
377
A14 log amplier assembly. The A7 assembly controls two functions on the A14 assembly: t Log/linear mode switching. A temperature-compensated control bias voltage switches the seven A14 amplier stages to function as either linear ampliers or logging ampliers. t 10 dB linear step gains. In linear mode, three temperature-compensated control lines bias four of the seven linear ampliers to provide step gains in 10 dB increments over a 40 dB range. Two ampliers are switched by one control line to provide the 20 dB step-gain stage. Linear step gain errors are corrected by the CAL AMPTD self-calibration routine. Refer to Chapter 13 for a description of CAL AMPTD.
A8 Power Supply
The A8 Power Supply is a switching power supply that plugs into a connector on the A15 motherboard. It is not repairable to the component level. When the line switch is on, it produces the low-power-on (LPWRON) signal that activates the power supply. The power supply then provides the following: +12 V for the B1 fan. The +5 V, +15 V, 15 V, and +12 V supply voltages. A line-trigger signal (LINE_TRIG) at the power-line frequency. The high-power-on (WRUP) signal used to coordinate the start-up of the A16 assembly. +12 V for the A2 display. 24 V for the YTF on the 8592L and 8593E.
378
Chapter 9
Refer to the A130 Narrow Bandwidth Assembly section of Foldout 11-1. The narrow bandwidth assembly is inserted into one of the four card-cage slots. The narrow bandwidth assembly receives the 21.4 MHz AUX_IF from the A15 motherboard. The AUX_IF signal is available to all the card-cage assemblies. The 21.4 MHz signal is downconverted to a 8.56 kHz IF. This low IF frequency was chosen to facilitate the necessary Q for the bandwidth lters and to eliminate the need for bulky crystal circuits. The 8.56 kHz IF is buffered, put through two poles of active ltering, then the signal goes through 0 to 40 dB of step gain. After the step gain, there are two more active lter poles before the signal is put through a detector. If the analyzer is in log mode the signal is amplied logarithmically. Otherwise, the signal is amplied with linear circuits. The detected video signal is then routed back to the A15 motherboard as VIDEO_IF. This signal then goes to the A16 processor video assembly and converted to digital. Refer to the A14 Log Amplier/Detector section of the IF/Control block diagram. Note that jumper W1 must be removed when Option 130 is installed in the analyzer. The video signal takes a different path when Option 130 is installed and bandwidths greater than 300 Hz (wide bandwidths) are selected. After the signal is detected on the A14 log amplier assembly, the video is routed to A14J2, DAISY CHAIN VIDEO OUTPUT, then W130 sends the video signal to the A130 narrow bandwidth assembly at A130J3. This video signal goes to the output multiplexer located on the A130 narrow bandwidth assembly. In the wide bandwidths, this signal is selected and is then routed to back to the A15 motherboard as VIDEO_IF. The A130 narrow bandwidth assembly also has a DAISY CHAIN VIDEO OUTPUT (A130 J2). However, this output is not currently in use.
Chapter 9
379
380
Chapter 9
382
Chapter 9
384
Chapter 9
386
Chapter 9
388
Chapter 9
390
Chapter 9
392
Chapter 9
10
393
Analyzer Options and Block Diagrams Analyzer Options and Block Diagrams
394
Chapter 10
Analyzer Options and Block Diagrams Analyzer Options and Block Diagrams
Table 10-1
Option Table
8594Q 8591C 8591E 8593E 8594E 8595E 8596E 8590L 8592L 8594L 395 Option
001 003 004 009 010 011 012 015 016 041 043 026 027 040 042 050 051 101 102 103 105 107 110 111 112 119 130 140
75 Ohm Input Memory Card Reader Precision Freq Reference LO/Sweep+Tune Out Tracking Generator TG/75 Ohm Output DECT Source Soft Tan Carrying Case Soft Yellow Operating/Carrying Case IB and Parallel Interface RS-232 and Parallel Interface 26.5 GHz/APC Conn 26.5 GHz/N Conn Impact Cover Soft Carry Case Improved Amplitude Accy Improved Amplitude Accy Fast ADC TV SYNC/AM-FM Demod Quasi-Peak Detector Gated Video TV Receiver/Video Tester CT2 Demodulator Group Delay DECT Demodulator Noise Figure Narrow Res Bandwidth Narrow BW/Freq Ref
Chapter 10
Analyzer Options and Block Diagrams Analyzer Options and Block Diagrams
Table 10-1
151 161 301 E04 H80 H81 701 704 711 713 908 909 910 915
Digital Demod NADC ROMs Fast ADC/TV SYNC/AM-FM Demod DIRTS PAL Secam Disp NTSC TV Display Delete ADC/TV Sync Delete Prec Freq Ref 50/75 Matching Pad Delete Counter-Lock Rack Mount Without Handles Rack Mount With Handles Extra Manual Service Guides
396
Chapter 10
Option Descriptions
75 Input Impedance (Option 001)
For 8590L and 8591E only. This option provides a 75 input impedance instead of the standard 50 impedance. analyzers with this option use cables, circuit boards, and front panels that are different from the standard units.
Chapter 10
397
8593E, 8594E, 8595E, and 8596E: Option 010 is also available, with factory installation (Option K10) after the purchase of your analyzer. For 8590L and 8591E only. Option 010 (100 kHz to 1.8 GHz) or 011 (1 MHz to 1.8 GHz) provides a built-in tracking generator. Option 010 has the standard 50 output impedance. Option 011 has 75 output impedance. 8590L: Options 010 and 011 are also available as retrot kits (Option R10 or R11) after the purchase of your analyzer, or as kits part number 5062-6463 and 5062-6479 respectively. 8591E: Options 010 and 011 are also available as retrot kits (Option R10 or R11) after the purchase of your analyzer, or as kits part numbers 5062-6460 and 5062-6480 respectively.
398
Chapter 10
Option 041 allows the analyzer to copy its screen to a printer or plotter, and includes a separate connector that accepts programming commands from an external keyboard. Option 041 is also available as a retrot kit (Option R41) after the purchase of your analyzer, or as a kit part number 08590-60380.
NOTE
Option 041 can be converted to an Option 043 by ordering the RS-232 and parallel interface connector assembly part number 08590-60369.
Option 043 can be converted to an Option 041 by ordering the IB and parallel interface connector assembly part number 08590-60368.
Chapter 10
399
NOTE
The extended frequency specications only apply when using the Type-N to SMA adaptor that is supplied with the option.
400
Chapter 10
Chapter 10
401
AM/FM Demodulator with Speaker and TV Sync Trigger Circuitry (Option 102)
For 8591E, 8593E, 8594E, 8595E, and 8596E only. Option 102 enables you to use amplitude or frequency demodulation and to listen to a demodulated signal. Option 102 also allows you to TV trigger on the selected line of a TV video picture frame if both Option 101 and 102 are installed. The sweep triggering works with interlaced or noninterlaced displays for the NTSC, PAL, and SECAM formats. Also see Option 301 below. Option 102 is also available as a retrot kit (Option R02) after the purchase of your analyzer, or as a kit part number 5062-6457.
402
Chapter 10
Chapter 10
403
404
Chapter 10
Option 160 implements a coherent downconversion, and calculates the minimum RMS error vector magnitude of a timeslot. Carrier frequency error and I-Q origin offset are also extracted from sampled data. Option 151 with Option 160 and the 85718B NADC-TDMA measurements personality provide a complete NADC transmitter RF measurement solution including modulation metrics. Option 151 with Option 160 and the 85720C PDC measurements personality provide a complete PDC transmitter RF measurement solution including modulation metrics. Option 151 with Option 160 and the 85725B CDMA measurements personality provide a CDMA transmitter measurement solution with fast DSP-assisted measurements. Option 151 with Option 160 and the 85726B PHS measurements personality provide a complete PHS transmitter RF measurement solution including modulation metrics.
Chapter 10
405
TV Sync Trigger Capability/Fast Time-Domain Sweeps and AM/FM Demodulator (Option 301)
For 8591E, 8593E, 8594E, 8595E, and 8596E only. Option 301 is a combined option made of Options 101 and 102, allowing sweep times down to 20 s in zero span, enabling use of amplitude or frequency demodulation, and allowing a demodulated signal to be listened to. See Options 101 and 102 above for more detailed descriptions.
406
Chapter 10
Chapter 10
407
Modulator drive voltage Drain voltage (.5 to 2.0V) Gate 1 bias voltage (5.4 V) (0.5 to 2.0 V) Drain voltage (.5 to 2.0V) Gate 2 bias voltage (5.4 V) (0.5 to 2.0 V) Control for 30 dB attenuator step Switched +15 VF Control for 20 dB attenuator step +15 VF for Amplier/Detector Control for 10 dB attenuator step +15 VF for Attenuator
408
Chapter 10
The tracking generator bias voltages from A7A1 tracking generator control board assembly to the A24A2 tracking converter are described in the table below.
Table 10-3
a. These bias voltages should remain constant even when the tracking generator source power changes, as long as the ALC loop is level. If the A24A3 amplier/detector is suspect, you can perform some simple checks using a DVM. Measure the +15 V line to ground using an ohmmeter. You should measure approximately 200 . Check the detector diode using the diode checker on the DVM. Connect the black lead to the center pin of the Video Output and the red lead ground. You should measure approximately 3 k. Connect the red lead to the center pin of the Video Output and the black lead to ground. you should measure approximately 15 k. If this measures approximately 8 k, the detector diode inside is too leaky and will cause a failure.
Chapter 10
409
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
Wait for the PEAKING message to disappear. Step the SRC POWER setting in 1 dB increments and note at which power levels the output is unlevelled. The output should be unlevelled only when the power level is greater than +1 dBm.
410
Chapter 10
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
3. Check maximum power available from the tracking generator. Connect the RF OUT 50 connector to the INPUT 50 connector. Press the following keys.
PRESET AMPLITUDE, 20, +dBm LOG dB/DIV, 5, dB AUX CTRL TRACK GEN SRC PWR ON OFF (ON underlined) MORE 1 OF 2 ALC INT EXT (EXT underlined)
The available power should always be greater than +1 dBm. If the output is unlevelled only at specic frequencies, a power hole will usually be visible at those frequencies. 4. Check the LO OUTPUT power level as follows: a. Set the analyzer to zero span at a 0 Hz center frequency. b. Zero and calibrate a power-meter/power-sensor combination. Set the power meter to readout power in dBm. Enter the power sensor's 4 GHz cal factor into the power meter. c. Connect the power sensor to the LO OUTPUT connector on the analyzer rear panel. d. Record the power meter reading. The power level should be greater than +12.5 dBm. e. Increase the analyzer center frequency setting by 100 MHz. The LO OUTPUT frequency will be 3.9214 GHz greater than the center frequency setting. f. Enter the appropriate power sensor cal factor into the power meter. g. Repeat step a to step f until the center frequency setting is 2.9 GHz. 5. If the LO OUTPUT power-level check fails, note the center frequency setting at which the power level was out of tolerance. If the LO OUTPUT power level check passes, proceed to step 7.
Chapter 10
411
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
6. Place the A3 front end assembly in its service position. Place the A3A15 tracking generator assembly in its service position. Disconnect W41 from A3A15J4 (LO IN). Connect the power sensor to the free end of W41. Repeat the LO OUTPUT power level check above, noting the center frequency settings at which the power level is out of tolerance. The power level for this check should be +16.5 dBm 2 dB. If the power level is within tolerance at W41, but out of tolerance at the LO OUTPUT (rear panel), and the center frequency setting of the out-of-tolerance power levels is close to the frequencies at which the output is unlevelled, suspect A3A15. If the power level at W41 is also out of tolerance, suspect either the A3A14 LODA assembly, A3A7 YTO assembly, or W41. Refer to the LO Section information in Chapter 9. 7. If the output is unlevelled only at certain power level settings or certain frequencies, monitor A10J1 pin 8 with a DVM. Connect the negative DVM lead to A7JTP1. Vary the SRC POWER setting or center frequency setting, as appropriate, and plot the voltage variation versus power level or frequency. A discontinuity in the plot near the frequency or power level at which the output is unlevelled indicates a problem on the A10 tracking generator control assembly.
412
Chapter 10
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
Chapter 10
413
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
414
Chapter 10
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
Chapter 10
415
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
The 1st LO typically has a higher harmonic content than the tracking generator output. For the purposes of this check, it is the variation in harmonic content versus frequency which is important. If the variation of the harmonic level of the 1st LO versus frequency tracks the harmonic level variation of the tracking generator output, repeat step 1 while measuring the 1st LO signal at the end of W41 nearest A3A15. If there is little variation in the 1st LO harmonic level between the LO OUTPUT connector and W41, and the relative variation in harmonic level tracks with the tracking generator output harmonic level, suspect either the A3A14 First LO Distribution Amplier assembly or the A3A7 YTO assembly. If the harmonic level variation of the 1st LO versus frequency does not track the harmonic level variation of the tracking generator output, suspect A3A15. 2. If sidebands are present at the same frequency offset at every output frequency, use another analyzer to check the spectral purity of the 1st LO and the 600 MHz drive signals. When checking the 1st LO, the 8594A must be set to zero span. The 1st LO frequency will be 3.9214 GHz greater than the center frequency setting. A 1 MHz sideband on the 1st LO will appear as a 1 MHz sideband on the output signal. To verify that the 600 MHz drive or 1st LO signal is responsible for the sidebands, substitute a clean signal for the 600 MHz drive or 1st LO signal. If the sidebands on the output disappear when using the clean signal, the substituted signal was responsible for the sidebands.
416
Chapter 10
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
NOTE
The 600 MHz drive signal should be 8 dBm 3.5 dB. The 1st LO signal should be +16 dBm 2 dB.
c. The amplitude of the positive-going ramp displayed on the oscilloscope should be approximately 7.8 V. d. Change the SRC PWR SWEEP setting to any value between 1 and 11 dB. The ramp amplitude displayed on the oscilloscope should be 780 mV per dB of the SRC PWR SWEEP setting.
NOTE
Although the source power sweep may be set to a 12.75 dB sweep width, the power sweep function is only warranted to have a 11 dB sweep width. 2. Perform the Vernier Accuracy performance test. See the calibration guide for your instrument. If this test fails, refer to If vernier accuracy is out of tolerance. earlier in this section.
Chapter 10
417
Analyzer Options and Block Diagrams Tracking Generator 8593E, 8594E, 8595E, 8596E Option 010
The following step requires adjustment of A3A15C3. The lifetime of A3A15C3 is rated for less than 10 cycles. Do not adjust A3A15C3 unless it is absolutely necessary. 7. If all of the checks above are acceptable, the tracking oscillator might not be functioning. Set up the 8594A as indicated in the Tracking Oscillator adjustment procedure in Chapter 2, using an analyzer, such as an 8566A/B, in place of the frequency counter. Try to adjust A3A15C3 until a signal is displayed on the 8566A/B. If adjusting A3A15C3 does not result in the tracking generator beginning to function, the A3A15 Tracking Generator assembly is suspect.
418
Chapter 10
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
The following troubleshooting information is aimed at isolating 1st LO amplitude-related faults to either the A3A14 LODA assembly, A10 Tracking Generator Control assembly, or the A3A7 YTO assembly. The A3A7 YTO and A3A14 LODA are not eld repairable. The A10 Tracking Generator Control assembly is eld repairable. However, if the repair time exceeds 30 minutes, it may be more cost-effective to order a replacement assembly rather than to continue component-level repair. 1. Place the A3 Front End assembly in its service position. Place the A10 Tracking Generator Control assembly on an extender board. Remove W33, the semi-rigid coax cable between A3A14 and A3A6. Do not reconnect W20 to AT1. 2. Connect a DVM's negative lead to A10TP2, AGND. Connect the positive lead to A10TP6 (PIN). 3. Set the analyzer by pressing the following keys.
FREQUENCY CENTER FREQ, 300, MHz SPAN, 20, MHz TRIG, CONT SGL (SGL underlined)
4. Measure the LO power at A3A14J2 and the free end of W20. Refer to Figure 9-5 for acceptable power level ranges. 5. If both LO power levels are lower than acceptable, the voltage on A10TP6 (PIN) should be above 0 V. If both LO power levels are higher than acceptable, this voltage should be more negative than 10 V. 6. If the voltage measured in step 5 is as described, the LODA drive circuitry is acceptable. Check the A3A7 YTO output power level. Refer to Foldout 12-2 for acceptable power level range. 7. If the voltage measured in step 5 is not as described, either the LODA drive circuitry or the LODA itself is malfunctioning. Check that the operational amplier's output is consistent with its inputs. 8. Set the analyzer LINE switch to off, disconnect W38 from A10J2, then set the analyzer LINE switch to on. 9. Connect a jumper between A10J2 and A10TP2. This connects ground A2, a oating ground, to AGND. Connect another jumper between A10TP4, LOS (LO sense) and +10 VR (refer to the A10 schematic).
Chapter 10
419
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
10.The voltage measured on the DVM should be greater than +14 Vdc. 11.Move the jumper from +10 VR to 10 VR (refer to the A10 schematic). The voltage measured should be more negative than 13 Vdc. 12.If the voltages do not meet the limits described in step 10 and step 11, troubleshoot the A10 Tracking Generator Control assembly. 13.Connect the positive DVM lead to A10J2 pin 1. 14.The measured voltage should be approximately +5 Vdc. If the voltage is not +5 Vdc, troubleshoot the A10 Tracking Generator Control assembly. 15.Connect the positive DVM lead to A10TP5, GB (gate bias). The voltage should measure within 5% of the GATE voltage listed on the RF section label. 16.If the voltage is not within this range, rotate A10R29 (GATE) through its range while monitoring the DVM. 17.If the voltage varies between 0 Vdc and 2 Vdc, adjust A10R29 (GATE) for a DVM reading within 5% of the GATE voltage listed on the RF section label. If the voltage does not vary outside the range of 0 Vdc and 2 Vdc, troubleshoot the A10 assembly. 18.Set the analyzer LINE switch to off, reconnect W38 to A10J2, then set the analyzer LINE switch to on. 19.If the DVM reading changes more than 50 mV, the A3A14 LODA is probably defective.
420
Chapter 10
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
The 8594Q Option 190 DVB-C QAM analyzer is a complex instrument to troubleshoot. It is recommended that the following troubleshooting procedures be performed by an approved Agilent Technologies service center only. This information is divided into two sections. Troubleshooting Tree Option 190 Service Menus
Troubleshooting Tree
The following troubleshooting trees are provided as a tool to use to isolate the most common problems in an 8594Q QAM analyzer. More complex interactive problems will require more detailed troubleshooting knowledge to be gained with experience and help from division engineers. Use the following guidelines to use the 8594Q Option 190 DVB-C QAM analyzer troubleshooting trees use the following guidelines. 1. Start with TREE 1. 2. Read and evaluate each key question in the tree you are working with before moving on to the next tree. 3. Follow through the ow of the trees, sequentially, starting with TREE 1, for every new problem. There are ve troubleshooting trees included in this section. Tree 1: Spectrum Analyzer Platform vs Option 190 Tree 2: Internal Preamplier Tree 3: Option 190 Tree 4: EVM Problems Tree 5: Average Power Test Problems
Chapter 10
421
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 1: Spectrum Analyzer Platform vs Option 190 IS THE PROBLEM A SPECTRUM ANALYZER PLATFORM OR OPTION 190 PROBLEM? Comments Isolate the problem. Note that the base spectrum analyzer platform is the same as an 8594E spectrum analyzer. Problems/Reasons If the failure is in a spectrum analyzer test, it is most likely a problem in the base spectrum analyzer platform. Diagnostic/Fix Action Check the list of Option 190 failures that can look like spectrum analyzer problems. If none match, follow the typical 8594Q analyzer troubleshooting procedure documented in the 8590 Series Analyzers Assembly-Level Repair Service Guide. Check the list of spectrum analyzer platform problems that can look like Option 190 problems. If none match, go to TREE 3.
If the failure is in an Option 190 specic test under the DVB Analyzer Menu or at 859XE Option 190 Recalibration Software, it is most likely to be an Option 190 problem in the card cage.
422
Chapter 10
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 1: Spectrum Analyzer Platform vs Option 190 IS THE PROBLEM A SPECTRUM ANALYZER PLATFORM OR OPTION 190 PROBLEM? (Continued) Comments Spectrum Analyzer Platform Problems Sometimes an Option 190 problem can look like a spectrum analyzer failure. High spurious in noise oor, or showing up in spectrum analyzer residual FM test. Problems/Reasons One or more power supply voltages can be shorted to ground. Diagnostic/Fix Action Detach and remove the Option 190 card cage boards, one at a time, in the following order: DSP, DEMOD, and RF card. After each card is removed, power up the instrument and look for normal spectrum analyzer behavior. The short is probably on the last card you removed. The Option 190 cards have many high-speed clock signals, and high amplitude oscillators. Sometimes these signals are coupled into the main spectrum analyzer path, and seen as spurs. 1. Check the Option 190 80-pin ribbon cable for good seating in its connector. 2. Check the Option 190 gray coaxial cables for good seating in its connector. 3. Check for proper routing of cables. Refer to the Service Guide. No signal on screen. The Option 190 RF card has an electronic switch that ips between the input signal to the spectrum analyzer path, or the Option 190 RF card input. This switch may be broken, or frozen so that the input signal never reaches the spectrum analyzer path. Remove the RF card. Reconnect the 321.4 MHz path as in an 8594E spectrum analyzer. Check for a trace on screen.
Chapter 10
423
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 1: Spectrum Analyzer Platform vs Option 190 IS THE PROBLEM A SPECTRUM ANALYZER PLATFORM OR OPTION 190 PROBLEM? (Continued) Comments Option 190 Problem Sometimes a spectrum analyzer platform problem can look like an Option 190 problem. Problems/Reasons Center frequency tuning problem. Diagnostic/Fix Action If the analyzer tuning/sweep circuits are broken, the Option 190 EVM test will fail. 1. Check the 10 MHz reference. 2. Perform the spectrum analyzer frequency and amplitude calibration. If it fails, follow the typical 8594Q troubleshooting procedure documented in the 8590 Series Analyzers Assembly-Level Repair Service Guide. Signal level is too high or too low at the 321.4 MHz tap point into the Option 190 cards. If the QAM analyzer has a problem in the down-conversion chain, the RF board will not receive the correct power level. Probe the input to the RF card. The power level should be between 33 and 25 dBm.
424
Chapter 10
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 2: Internal Preamplier IS THE PROBLEM IN THE 8594Q QAM ANALYZER INTERNAL PREAMPLIFIER? Comments Internal Preamplier Problems Problems/Reasons No signal on screen Diagnostic/Fix Action Refer to Set up to use the Preamplier in the Making Measurements chapter of the user's guide for your QAM analyzer. 1. Go to the Preamplier Control menu, then select PREAMP ON OFF several times to turn the preamplier on and off. You should hear a loud click as the coaxial switch puts the preamplier in and out of the spectrum analyzer path. 2. If you do not hear the click, the problem is either in the coaxial switch, or in the control board mounted on top of the IF. Check the wiring harness to and from the control board. Replace the control board, and try to turn the internal preamplier on and off again. 3. Measure the signal at the input to the RF card. If there is no signal, the problem is in the preamplier, the preamplier control, or in the analyzer front end. Signal compressed at the top of the display. or Displayed Average Noise Level (DANL) appears to be too high. or Signal level much larger than true power level of the input signal. Refer to Set up to use the Preamplier the Making Measurements chapter of the user's guide for your QAM analyzer. 1. Go to the Preamplier Setup menu, then select PREAMP GAIN, then enter 0 dB. If the signal jumps 20-26 dB in amplitude, the internal preamplier is on. Try to turn it off, using PREAMP ON OFF softkey. 2. If it does not turn off, the problem is in the coaxial switch, or the preamplier control board located on top of the IF.
Chapter 10
425
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 3: Option 190 HOW DO I CHECK THE BASIC CONFIGURATION? Comments Problems/Reasons The Option 190 consists of three card-cage option cards, DSP rmware, host analyzer rmware, and the DLP. All revisions must be up to date. Diagnostic/Fix Action Access the option listing under Show Options. Conrm that the option software revisions are current, and that no conguration error message shows. The following is a list of the initial revisions. Firmware: 961216 DLP: 961218 Option 190 DSP: 961206 If you get a conguration error, refer to the Error Messages section of the 8594Q Option 190 DVB-C QAM Analyzer User's Guide. WHICH OPTION 190 TEST FAILED? Comments Problems/Reasons The Option 190 Recalibration routine or DVB Demod Test exercises both the Option 190 hardware, and the base spectrum analyzer platform hardware in different tests. Diagnostic/Fix Action If the Option 190 Recalibration routine or DVB Demod Test fails the EVM test limit, the problem is most likely in the Option 190 hardware/software or the DLP. Go to TREE 4. If the Option 190 Recalibration routine or DVB Demod Test fails the average power limit, the problem is most likely in the basic spectrum analyzer platform hardware, the preamplier, or the DLP. Go to TREE 5.
IS THE ANALYZER RECEIVING A GOOD INPUT SIGNAL? The test signal comes from the 8770 arbitrary waveform synthesizer. It is very sensitive to improperly connected or terminated cables. View the spectrum on a spectrum analyzer screen. It is centered at 27.6 MHz. Refer to the Average Channel Measurement located in the Making Measurements chapter of the 8594Q Option 190 DVB-C QAM Analyzer User's Guide, for an example of the correct signal shape. The bandwidth depends on the exact test being performed, but the signal will be of the same shape as the example.
426
Chapter 10
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 4: EVM Problems IS THE FAILURE IN A SINGLE CHANNEL BANDWIDTH? Comments Problems/Reasons If the problem is concentrated in a single-channel bandwidth setting, the problem is probably in the RF option board. Diagnostic/Fix Action Replace the suspected RF board assembly with a known good RF board. If the analyzer passes the EVM test, save the bad board and the matching test data for division analysis.
IS THE FAILURE IN A SINGLE SYMBOL RATE? Comments Problems/Reasons If the problem is concentrated in a single-channel bandwidth setting, the problem is likely in the DEMOD board assembly. Diagnostic/Fix Action Replace the suspect DEMOD board assembly with a known good DEMOD board. If the analyzer passes the EVM test, save the bad board and the matching test data for division analysis.
Chapter 10
427
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 4: EVM Problems IS THE FAILURE IN A SINGLE CHANNEL BANDWIDTH? Comments Problems/Reasons Diagnostic/Fix Action
IS THE FAILURE WITH THE PREAMPLIFIER ON ONLY? The test conditions with the preamplier ON are designed to test the functionality of the preamplier and to stress the analyzer sensitivity to low signals. If the problem could be in the preamplier, the extra noise in the RF option board path, or it could be in the spectrum analyzer platform DANL. Evaluate the preamplier. Refer to Set up to use the preamplier in the Making Measurements chapter of the user's guide for your QAM analyzer. 1. Remove the 8770 test signal, then apply the CAL signal to the RF input. 2. Do the Measuring the Preamplier Gain procedure to nd out what suspect preamplier gain truly is. It should be 23 3 dB. If it is not, the problem is in the preamplier. 3. Test the preamplier control by turning the preamplier on and off, using PREAMP ON OFF. Evaluate the spectrum analyzer platform DANL. If the DANL is out of specication, the problem is in the spectrum analyzer down-conversion path. Swap the RF option card with a known good card. Rerun the test. If the problem is resolved, save the bad board and the matching test data for return to Agilent.
428
Chapter 10
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 4: EVM Problems IS THE FAILURE IN A SINGLE CHANNEL BANDWIDTH? Comments Connector Problems Care must be taken on inserting the cables into the board to keep the male connectors well aligned with the female connectors. Inspect the cable male connector before EVERY insertion. Correct any tilt or off-center of the center pin, if necessary. Problems/Reasons The MCX connectors used on the gray cables on the option cards are delicate. Diagnostic/Fix Action Inspect both the female connector on the board, and the male connector on the cable. If the female connector on the board is not severely crushed, it can often be cleared using a dental pick. If the male connector pin appears to be off center, recenter using ne tweezers. Replace the board only if the connector center hole is crushed beyond repair.
Chapter 10
429
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
Problems/Reasons
The problem is in the main demodulation path used by all EVM test modes.
Diagnostic/Fix Action
1. Check the adjustment of the 71.025 MHz oscillator on the RF option board. The nominal setting at room temperature after warmup and instrument preset (PRESET) of the instrument should be 71.025 MHz. If it has drifted off frequency, use a tweaker to adjust C402 to bring the frequency to 71.025 MHz. Watch the signal. Is it stable, or does it drift again? If it drifts again, Q400 is probably bad. Q400 is a specially selected 2N3904 transistor. Check with the division to obtain an approved replacement. Rerun the EVM test to check if the problem is resolved. If not, proceed to step 2. 2. Replace the RF board with a known good board. Does the instrument now pass the EVM test? If so, save the bad board and the matching test data for division analysis. If not, continue to step 3. 3. Replace the DEMOD option board with a known good board. Does the instrument now pass the EVM test? If so, save the bad board and the matching test data for division analysis. If not, continue to step 4. 4. Replace the DSP option board with a known good board. Does the instrument now pass the EVM test? If so, save the bad board and the matching test data for division analysis. If not, after swapping to a full set of known good boards, set the instrument aside for further analysis.
430
Chapter 10
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 5: Average Power Test Problems IS THE TEST SIGNAL AT THE PROPER LEVEL? Comments Test Signal Level Problems/Reasons The test signal comes from the 8770 arbitrary waveform synthesizer. Diagnostic/Fix Action View the spectrum on a spectrum analyzer screen. It is centered at 27.6 MHz. Refer to the Average Channel Measurement located in the Making Measurements chapter of the 8594Q Option 190 DVB-C QAM Analyzer User's Guide for an example of the correct signal shape. The bandwidth depends on the exact test being performed, but the signal will be of the same shape as the example. If the signal appears to be normal, check the output power through a power meter. Is the test signal at the right power level? It should be approximately 4 dBm 0 (8770A attenuator setting) pad with the preamplier OFF, and 4 dBm 50 (8770A attenuator setting) pad when the preamplier is on. Preamplier OFF If the problem appears with the preamplier OFF (PREAMP ON OFF, OFF is underlined), the root cause is likely to be a spectrum analyzer platform path problem or DSP problem. The DLP uses only the 8594E spectrum analyzer hardware and the DSP card to execute the average power measurement. Replace the DSP option board with a known good board. Rerun the test to see if it passes. If so, save the bad board and the matching test data for division analysis. If not, after swapping the DSP does not work, perform an analyzer self calibration. Troubleshoot as a spectrum analyzer platform problem if it does not pass.
Chapter 10
431
Analyzer Options and Block Diagrams First LO Distribution Amplier (LODA) 8593E, 8594E, 8594Q, 8595E, 8596E Option 009
TREE 5: Average Power Test Problems IS THE TEST SIGNAL AT THE PROPER LEVEL? Comments Preamplifer ON Problems/Reasons If the problem appears with the preamplier ON (PREAMP ON OFF, ON is underlined) the root cause is likely to be the preamplier or the preamplier control. Diagnostic/Fix Action 1. Evaluate the preamplier. Refer to Set up to use the preamplier in the Making Measurements chapter of the user's guide for your QAM analyzer. a. Remove the 8770 test signal, then apply the CAL signal to the RF input. b. Do the Measuring the Preamplier Gain procedure to nd out what suspect preamplier gain truly is. It should be 23 3 dB. If it is not, the problem is in the preamplier. c. Test the preamplier control by turning the preamplier on and off, using PREAMP ON OFF. 2. If you cannot control the preamplier, the problem is in the preamplier control board. Swap the preamplier control board with a known good board. 3. If the problem persists, replace the DSP with a known good board. Rerun the test to see if it passes. If so, save the bad board and the matching test data for division analysis. 4. If swapping the DSP does not work, perform an analyzer self calibration. Troubleshoot as a spectrum analyzer platform problem if it does not pass.
432
Chapter 10
Analyzer Options and Block Diagrams Option 190 DVB-C Connection Diagram
Chapter 10
433
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
It is important to realize that all the keys, with one exception (RCVR CAL ON OFF,) under Service Menus intentionally retain their values through presets and power cycles like all other Setup keys. This is done so board swapping may be done during troubleshooting and the previous state can be retained. After troubleshooting is completed, the DEFAULT SETUP key should be press to clear any conditions set via the Service Menus.
Passcode
To enable the Service Menus you must enter a passcode. Perform the following procedure to access the Service Menus. 1. Press the following keys.
PRESET DVB-C ANALYZER Setup Analyzer Setup MAX MIXER LEVEL , 1, 2, 3, dBm
434
Chapter 10
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
Menu 1 of 3
DIAGS ON OFF enables the display of the diagnostics data on
demod-based measurements along with the measurement results. This data is shown as a small highlighted block on the screen. The exact meaning of each of the data items is detailed in the Diagnostic Data section.
IGNORE ON OFF enables a mode in which demod-based measurements will ignore any QAM acquisition or measurement failure and show the measurement anyway. This is useful to observe the constellation under this condition since it can yield valuable clues about the particular fault (spinning or random constellation). Note that this only ignores acquisition or measurement unlocks; it will not ignore a 240 MHz PLL unlock or level set failure. KEYS ON OFF enables the following diagnostic and optional control keys located elsewhere in the menus.
Service is located in the lq Constln menu. AGC HOLD RUN and DNCVRTR ATTEN are shown during a continuous (live) demod-based measurement. DNCVRTR ATTEN is used to manually enter a value of Option 190 downconverter attenuation independent of the value set by the level-set algorithm, and display in the diagnostic data. Useful information can be gleaned by entering values different than that set by the level-setting algorithm, then observing the effects on the constellation, or other measurement. Note that values are entered blind into the keypad, terminated by
dB; there is no visible active function area on the screen. Note also
that the diagnostic data does not change to reect the new value of downconverter attenuation. Also, only even values of attenuation are utilized, due to only 2 dB steps are available on the board. AGC HOLD RUN (hold underlined) will freeze the AGC loop on the Option 190 demodulator input. This allows subtle effects to be observed without being obscured by the action of the AGC loop. VARIANCE SD 90% is located in the Modulation Accuracy menu. When set to SD (SD is underlined), this will make averaged measurements display the standard deviation instead of the 90% condence value of measurement variance. METRIC ON OFF located in the SER, C/N Margin menu, when set to ON (on is underlined) will make the SER measurement display the EVM, or MER, that accompanies the measurement in a manner identical to the channel response measurements.
Chapter 10
435
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus Dncvtr Test leads to a set of menus that allows the setup of various
Option 190 downconverter functions in a manner convenient to troubleshooting, for example, no measurements are run. The softkeys located under Dncvtr Test are documented in the Downconverter Test section.
Menu 2 of 3
RCVR CAL ON OFF allows you to enable or disable use of the Option 190
calibration by demod-based measurements. The Option 190 calibration is a special factory calibration that allows us to compensate for the linear distortion of the analyzer. This improves the channel response measurement results as well as increases the ability of the demodulator to lock to a signal. The calibration values are kept in the analyzer CMOS RAM memory and automatically uploaded by the analyzer rmware to the Option 190 board set at power-up. If the cal is missing or corrupt, RCVR CAL ON OFF will remain in the OFF position (OFF is underlined) and cannot be toggled ON. This key is the one exception to the Setup keys: it does NOT retain its value through preset or power cycles. The reason being the mode entry routine will automatically turn RCVR CAL ON OFF to OFF (OFF is underlined) if it detects that the Option 190 cal is missing or corrupted (and display a message). The mode entry routine will also turn RCVR CAL ON OFF to ON (ON is underlined) if the CAL is OK. Pressing DEFAULT SETUP key will set RCVR CAL ON OFF to ON (if there is a valid CAL). The state of RCVR CAL ON OFF is also shown in the Diagnostic Data section.
Dsp Evm Corr key leads to a menu which allows you to individually
enable or disable DSP based evm correction for each of the 3 bandwidths, 2,4 or 8 MHz. For example, the 8 MHz CH ON OFF key when set to ON enables DSP based corrections on 8 MHz channels. As described in Modulation Accuracy (metrics) above, the 2 &4 MHz bandwidths normally have this enabled, but not the 8 MHz bandwidth.
EVM CORR ON OFF enables DLP based evm corrections on channels that
do not have DSP based evm corrections enabled. The diagnostics data (see the Diagnostic Data section) displays if DSP or DLP based evm corrections are on for the current analyzer state.
CNTL A 1 0 allows you to toggle the preamp bypass switch independently of PREAMP ON OFF key. Selecting 1 puts the switch in preamp mode, 0 puts the switch in bypass mode. This is useful for troubleshooting the preamp drive circuits. SHOW REVDATES key displays all three software revision dates:
the 8594Q QAM analyzer rmware (F/W:), Option 190 DSP rmware (DSP:) and DLP.
436
Chapter 10
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
Menu 3 of 3
COUPLING ON OFF (when set to OFF) allows you to disable the automatic selection of channel bandwidth when a new channel is entered. This may be used if you want to ignore the FTZ `D' channel layout while using `D' channels. RETUNE ON OFF (when set to OFF) allows you to disable the automatic
re-setting of the 71.1 MHz oscillator that is done when a demod-based measurement signal acquisition fails. This can be set to off to investigate signal acquisition for a xed 71.1 MHz tuning state (lower, centered or upper).
ACQ MER LIMIT allows entry of values from 0 to 50 dB (the default is 20 dB). This value is used during a demod-based measurement signal-acquisition routine. If the measured value of MER is less than this limit and the MER LMT ON OFF key in the Demod Setup menu is ON, the acquisition routine will abort with a MER TOO LOW message. MEAS MER LIMIT allows entry of values from 0 to 50 dB (the default is 20 dB). This value is used during a demod-based measurement. If the measured value of MER is less than this limit and the MER LMT ON OFF key in the Demod Setup menu is ON, the measurement will abort with a MER TOO LOW message. NOTE
Setting MER LMT ON OFF in the Demod Setup menu to OFF will ignore both ACQ and MEAS MER limits. This is essentially a shortcut to setting both ACQ and MEAS MER limits to 0 dB to allow acquisition and measurement of very noisy signals.
IGNOR BW ON OFF is used to disable the automatic channel bandwidth
verication that is done at the start of single signal demod based measurements. If this key is set to ON, the signal's (approximate) measured occupied bandwidth is not compared against the setting of the CHAN BW SELECT key.
Chapter 10
437
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
Diagnostic Data
When DIAGS ON OFF in the Service Menus is set to ON, demod-based measurements will include the display of extra diagnostic data in a small highlighted block on the screen. The format is: XX dB YY,Z ABCD where: XX is the amount of dB set on the Option 190 downconverter assembly after the automatic level setting algorithm has run. This will be an even number from 0 to 60 dB. YY is the nal AGC (signal) reading from the Option 190 demodulator input stage in hexadecimal after the automatic level setting algorithm has run. This will be 80 (128 decimal) or slightly more than this. The target for the algorithm is 80 to 82. Z is a number that indicates the sequence of action taken by the algorithm. There are 6 possible values: 0: the initial AGC reading indicated no adjustment in was needed. This should rarely happen since we intentionally set the gain equation for the initial amount of attenuation to always make it excessive to guarantee under-drive at the demodulator input. 1: The initial AGC reading indicated excess attenuation of less than 12 dB. The attenuation was reduced by this amount and a subsequent AGC reading indicated the signal was within the target. This is the most common case. 2: The initial AGC reading indicated excess attenuation of less than 12 dB. The attenuation was reduced by this amount and a subsequent AGC reading indicated adjustment was still required (but less than 12 dB). The attenuation was reduced again by the required amount and a nal AGC reading was taken. This is the next most common case. 3: The initial AGC reading indicated excess attenuation of 12 dB (or more). The attenuation was reduced by 12 dB and a subsequent AGC reading indicated the signal was within target. This will rarely happen. 4: The initial AGC reading indicated excess attenuation of 12 dB (or more). the attenuation was reduced by 12 dB and a subsequent AGC reading indicated adjustment was still required (but less than 12dB). The attenuation was reduced again by the required amount and a nal AGC reading was taken. This is rare, but possible (if the downconverter has more loss than usual). Note that this case doesn't check the nal reading and reduce the attenuation a third time if the target is not reached. So the level may not be optimum.
438
Chapter 10
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
5: The second AGC reading in case 2 or 4 above indicated adjustment was required and the amount of excess attenuation was 12 dB or more (for example, the cumulative amount of attenuation decrease would be 24 dB or more). This can happen if the signal path has excess loss. This case generates the Downconverter Level Set Failed error message. A is a number which represents the current Option 190 71.1 MHz oscillator state: 1: lower 2: centered 3: upper B is the character R if RCVR CAL is ON, r if it is OFF. C is the character D if DSP based evm corrections are ON, d if they are not. D is the character E if DLP based evm corrections are ON, e if they are not. A typical readout of a 20 dBm, 8 MHz channel might be: 46 dB 80,2 2RdE
Chapter 10
439
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
Downconverter Test
Dncvrtr Test in the Service Menus leads to a set of menus used to troubleshoot Option 190 RF Downconverter boards. Note that pressing it also sets up the analyzer for 0 span and routes the signal path to Option 190 board set instead of the spectrum analyzer. Thus, if a signal was present on the analyzer input, the screen will show only the analyzer noise oor after pressing Dncvrtr Test.
The Switch Control key leads to a menu which will allow you to select one of the 5 possible paths: 1. Spectrum Analyzer will route the signal back to the spectrum analyzer. If you press this, the screen will show the signal (in 0 span) if one is present on the analyzer input. 2. Demod 8 Mz SAW will route the signal through the Option 190 demod, 8 MHz SAW lter path. 3. Demod 4 Mz SAW will route the signal through the Option 190 demod, 4 MHz SAW lter path. 4. Demod 2 Mz SAW will route the signal through the Option 190 demod, 2 MHz SAW lter path. 5. Demod Bypass will route the signal through the Option 190 demod, bypass (no SAW lter) path.
DNCVRTR ATTEN allows entries from 0 to 60 dB to set the amount of
attenuation. Note that only even integer values are actually utilized and the value is set to 30 dB when entering the downconverter test menu.
PLL Control leads to a menu which will allow you to select the 240 MHz
PLL oscillator frequency (based on the channel bandwidth): 1. PLL 8 Mz Channel will set it to 243.4 MHz. 2. PLL 4 Mz Channel will set it to 246.85 MHz. 3. PLL 2 Mz Channel will set it to 248.571 MHz.
Custom Freq in the PLL Control menu leads to another menu which will allow you to set the 240 MHz oscillator to a custom frequency by specifying the `N' and `R' words which are sent as 19 individual (serial, MSB 1st) bits to the National semi LMX2325 PLL chip. The highest 11 bits of the `N' word is used to load the `B' register, the next 7 bits are used to load the `A' register, the lowest order bit always a 0 for `N' words. The highest 3 bits of the `R' word are not used, the next bit is a 1 to specify the 32/33 prescalar, the next 14 bits are used to load the `R' register, the lowest order bit always a 1 for `R' words. The frequency is given by:
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
For the 8 MHz channels, R=50, B=38 and A=1 which yields 243.4 MHz. For the 4 MHz channels, R=35, B=27 and A=0 which yields 246.85 MHz. For the 2 MHz channels, R=49, B=38 and A=2 which yields 246.571 MHz. The two separate keys, DECIMAL PLL N and DECIMAL PLL R, are used to enter the 2 values and contain the default values for the 8 MHz channel: 9720 (2602 hex) for the N word and 32869 (8065 hex) for the R word.
SHOW AGC VALUES will take a single AGC (signal) reading at the input
of the Option 190 demodulator input and display it both as a hex value and as its equivalent `gain' (for example, the amount of downconverter attenuation to subtract). With a signal at the input, by entering a value into DNCVRTR ATTEN and reading the AGC value returned by {SHOW AGC VALUES}}, you can accomplish manually what the DLP automatic level setting algorithm does. Note that if you want to duplicate the value of attenuation shown during a measurement and the measurement set the analyzer attenuator (AT) to 0 dB, you'll also have to (manually) set AT to 0 dB. The AGC HOLD RUN key allows you to freeze the AGC circuit when toggled to HOLD.
Chapter 10
441
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
IQ Constellation
If SHOW KEYS ON OFF in the Service menus is set to ON, a Service key will be present in the IQ Constellation menu. All the keys on this Service menu are duplicates (for convenience) of keys under Service menus discussed in the menu sections, except Osc Control. Osc Control leads to a menu which allows control of both the 240 MHz PLL and 71.1 MHz oscillators. PLL oscillator by 200 KHz when a demod based measurement is started and the is underlined or by +200 KHz when the + is underlined. This key is also a left-over from prototype days when we did not have control of the 71.1 MHz oscillator.
PLL Custom is also a left-over from prototype days and is a duplicate of the Custom Freq menu key described in the Downconverter Test section. TUNE 71.1 OSC shows the current state of the 71.1 MHz oscillator and allows you to enter a new state (tune number). Note that even though the automatic retune algorithm will only select the lower (1), centered (2) or upper (3) states, you can enter values here from 0 to 7. The 0 state is twice as low as the 1 state, the 4 and 5 states are don't cares', the 6 is twice as high as the 3 state and the 7 state is 3 times as high as the 3 state. It turned out that only the lower, centered, and upper states were required by the automatic retune algorithm to compensate for the temperature drift of the 71.1 oscillator. RETUNE ON OFF is a duplicate of the same key in the Service menus PLL OFS - 0 + is a 3-way toggle key. This key will offset the 240 MHz
numbers used during the last full signal acquisition sequence. A 2 means only tune state 2 was required. A 21 means that states 2 then 1 were tried. A 213 means that 2 then 1 then 3 were tried.
442
Chapter 10
FOLDOUT 10-1. 8590L/91C, 91E TRACKING GENERATOR OPTION 010 AND 011, OVERALL BLOCK DIAGRAM
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
444
Chapter 10
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
446
Chapter 10
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
448
Chapter 10
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
450
Chapter 10
SPECTRUM ANALZYER
A3A18 INPUT 50 W A3A17 0-70 dB ATTN 10 dB STEP AT 3-6.8 GHz _ G=+23 +3dB NF= 7 dB 3.6 GHz
RF
321.4 MHz
IF LO
RF
RBW
RF
IF LO
IF LO
TO BASEBAND DETECTOR
300 MHz
A190A1 RF
70 MHz
A190A2 DEMOD
4x SYMBOL RATE
RF
RF
IF LO
IF LO
2 dB STEP 4
I
A D C
FIR
CARRIER LO SYNCH
SYMBOL CLOCK
DMA
A190A3 DSP
8594DY HOST/PROCESSOR CARD CAGE BUS HOST INTERFACE DSP SAMPLE RAM RAM EEPROM
sq15
FOLDOUT 10-5. 8594Q OPTION 190 DVB-C QAM ANALYZER HARDWARE BLOCK DIAGRAM.
Analyzer Options and Block Diagrams Option 190 DVB-C Service Menus
452
Chapter 10
11
453
Major Assembly and Cable Locations Major Assembly and Cable Locations
454
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-2
Chapter 11
455
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-3
Instrument Top View 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E
456
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-4
Chapter 11
457
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-5
Card-Cage (Options 102, 103) 8590 E-Series, 8591C Cable TV Analyzers, and 8594Q QAM Analyzers
458
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-6
Figure 11-6
Card-Cage (Options 009, 010) 8593E, 8594E, 8594Q, 8595E, and 8596E
Chapter 11
459
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-7
460
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-8
Chapter 11
461
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-9
462
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-10
Instrument Bottom View with A25 Counter-Lock Removed 8590 E-Series and L-Series Spectrum Analyzers, 8591C Cable TV Analyzer, and 8594Q QAM Analyzers
Chapter 11
463
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-11
Instrument Bottom View 8590 E-Series and L-Series Spectrum Analyzers, 8591C Cable TV Analyzers, and 8594Q QAM Analyzers
464
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-12
Figure 11-13
Chapter 11
465
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-14
Figure 11-15
466
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-16
Chapter 11
467
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-17
468
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-18
A24 Tracking Generator 8591C (Option 011) and 8591E (Options 010 and 011)
Chapter 11
469
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-19
A3 Front-End Assembly, Top and Left-Side Views 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E
470
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-20
Chapter 11
471
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-21
472
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-22
Figure 11-22
Chapter 11
473
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-23
474
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-24
A3 Front-End Assembly, Rear View with A3A5 Removed 8592L and 8593E (Options 009 and 010)
Chapter 11
475
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-25
A3 Front-End Assembly, Rear View with A3A5 Removed 8594E (Options 009 and 010)
476
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-26
A3 Front-End Assembly, Rear View with A3A5 Removed 8595E and 8596E (Options 009 and 010)
Chapter 11
477
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-27
478
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-28
Chapter 11
479
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-29
A3 Front-End Assembly, Right-Side View 8595E and 8596E (Options 009 and 010)
480
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-30
A3 Front-End Assembly, Rear View with A3A5 Removed 8593E (Options 009 and 010)
Chapter 11
481
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-31
A3 Front-End Assembly, Rear View with A3A5 Removed 8594E (Options 009 and 010)
482
Chapter 11
Major Assembly and Cable Locations Major Assembly and Cable Locations
Figure 11-32
A3 Front-End Assembly, Rear View with A3A5 Removed 8595E and 8596E (Options 009 and 010)
Chapter 11
483
Major Assembly and Cable Locations Major Assembly and Cable Locations
484
Chapter 11
12
Replaceable Parts
This chapter contains information for identifying and ordering replacement assemblies for the 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers. Each assembly applies to all 8590 E-Series and L-Series spectrum analyzers, except where specied with the applicable analyzer models.
485
Replaceable Parts
Major assembly and cable location information is given in Chapter 11. The following tables and gures are also included in this chapter: Table 12-1 list reference designations, page 490 begins a list of abbreviations, and Table 12-2 lists value multipliers used in the parts lists. Table 12-2 lists standard value replacement capacitors. Table 12-3 lists standard value replacement resistors, 0.125 W. Table 12-4 lists standard value replacement resistors, 0.5 W. Table 12-5 lists all major assemblies, and all major mechanical and electrical parts that are not part of a major assembly. Table 12-6 lists the cable assemblies. Table 12-7 lists available component-level information packets. Figure 12-1 through Figure 12-7 give the analyzer parts identication information.
NOTE
The parts lists, schematics, and component-location diagrams for the 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers board assemblies are available separately in the 8590 Series Analyzers Component-Level Repair Service Guide.
486
Chapter 12
Ordering Information
Parts can be ordered by addressing the order to the nearest Agilent Technologies ofce. Customers within the USA can also use either the direct mail-order system, or the direct phone-order system described below. The direct phone-order system has a toll-free phone number available.
Chapter 12
487
Regular Orders
The toll-free phone number, (800) 227-8164, is available Monday through Friday, 6 AM to 5 PM (Pacic time). Regular orders have a four-day delivery time.
Hotline Orders
Hotline service is available 24 hours a day, 365 days a year, for emergency parts ordering. The toll-free phone number, (800) 227-8164, is available Monday through Friday, 6 AM to 5 PM (Pacic time). To cover the cost of freight and special handing, there is an additional hotline charge on each order (three line items maximum per order). Hotline orders are normally delivered the next business day after they are ordered.
488
Chapter 12
Table 12-1
Reference Designations
REFERENCE DESIGNATIONS
A AT
F FL HY
RT S T
B BT C
TB TC TP
CP CR
K L M MP
V VR
DC DL
Cable, Wire, Jumper Socket Crystal Unit (Piezoelectric, Quartz) Tuned Cavity, Tuned Circuit
DS
X Y
Chapter 12
489
Abbreviations
A ADJ ANSI ASSY AWG BCD BD BE-CU BNC BRG BRS BSC BTN C CD CER CHAM CHAR CMOS CNDCT CONT CONV CPRSN CCP CUP-PT CW D DA DAP-GL Across Flats, Acrylic, Air (Dry Method), Ampere Adjust, Adjustment American National Standards Institute (formerly USASI-ASA) Assembly American Wire Gage Binary Coded Decimal Board, Bundle Beryllium Copper Type of Connector Bearing, Boring Brass Basic Button Capacitance, Capacitor, Center Tapped, Cermet, Cold, Compression Cadmium, Card, Cord Ceramic Chamfer Character, Characteristic, Charcoal Complementary Metal Oxide Semiconductor Conducting, Conductive, Conductivity, Conductor
Contact, Continuous, Control, Controller
Converter Compression Carbon Composition Plastic Cup Point Clockwise, Continuous Wave Deep, Depletion,Depth, Diameter, Direct Current Darlington Diallyl Phthalate Glass
490
Chapter 12
FT JFET DBL DCDR DEG D-HOLE DIA DIP DIP-SLDR D-MODE DO DP DP3T DPDT DWL E-R EXT F FC FDTHRU FEM FIL-HD FL FLAT-PT FR FREQ FXD GEN GND GP
Current Gain Bandwidth Product (Transition Frequency), Feet, Foot Junction Field Effect Transistor Double Decoder Degree D-Shaped Hole Diameter Dual In-Line Package Dip Solder Depletion Mode Package Type Designation Deep, Depth, Dia-metric Pitch, Dip Double Pole Three Throw Double Pole Double Throw Dowell E-Ring Extended, Extension, External, Extinguish Fahrenheit, Farad, Female, Film(Resistor), Fixed, Flange, Frequency Carbon Film/Composition, Edgeof Cutoff Frequency, Face Feedthrough Female Fillister Head Flash, Flat, Fluid Flat Point Front Frequency Fixed General, Generator Ground General Purpose, Group
Chapter 12
491
H HEX HDW HLCL HP IC ID IF IN INCL INT J-FET K KNRLD KVDC LED LG LIN LK LKG LUM M MA MACH MAX MC MET MHZ MINTR MIT MLD MM
Henry, High Hexadecimal, Hexagon, Hexagonal Hardware Helical High Pass Collector Current, Integrated Circuit Identication,Inside Diameter Forward Current, Intermediate Frequency Inch Including Integral, Intensity, Internal Junction Field Effect Transistor Kelvin, Key, Kilo, Potassium Knurled Kilovolts, Direct Current Light EmittingDiode Length, Long Linear, Linearity Link, Lock Leakage, Locking Luminous Male, Maximum, Mega, Mil, Milli, Mode Milliampere Machined Maximum Molded CarbonComposition Metal, Metallized Megahertz Miniature Miter Mold, Molded Magnetized Material, Millimeter
492
Chapter 12
MOM MTG MTLC MW N N-CHAN NH NM NO NOM NPN NS NUM NYL OA OD OP AMP OPT PA PAN-HD PAR PB PC PCB P-CHAN PD PF PKG PLSTC PNL PNP POLYC
Momentary Mounting Metallic Milliwatt Nano, None N-Channel Nanohenry Nanometer, Nonmetallic Normally Open, Number Nominal Negative Positive Negative (Transistor) Nanosecond, Non-Shorting, Nose Numeric Nylon (Polyamide) Over-All Outside Diameter Operational Amplier Optical, Option, Optional Picoampere, PowerAmplier Pan Head Parallel, Parity Lead (Metal),Pushbutton Printed Circuit Printed Circuit Board P-Channel Pad, Power Dissipation Picofarad, Power Factor Package Plastic Panel Positive Negative Positive (Transistor) Polycarbonate
Chapter 12
493
POLYE POT POZI PREC PRP PSTN PT PW Q R REF RES RF RGD RND RR RVT SAWR SEG SGL SI SL SLT SMA SMC SMB SPCG SPDT SPST SQ SST STL
Polyester Potentiometer Pozidriv Recess Precision Purple, Purpose Piston Part, Point, Pulse Time Pulse Width Figure of Merit Range, Red, Resistance, Resistor, Right, Ring Reference Resistance, Resistor Radio Frequency Rigid Round Rear Rivet, Riveted Surface Acoustic Wave Resonator Segment Single Silicon, Square Inch Slide, Slow Slot, Slotted Subminiature, A Type (ThreadedConnector) Subminiature, C-Type (ThreadedConnector) Subminiature, B Type (Slip-onConnector) Spacing Single Pole Double Throw Single Pole Single Throw Square Stainless Steel Steel
494
Chapter 12
SUBMIN SZ T TA TC THD THK TO TPG TR-HD TRMR TRN TRSN UCD UF UH UL UNHDND V VAC VAR VDC W W/SW WW X YIG ZNR
Subminiature Size Teeth, Temperature, Thickness, Time, Timed, Tooth, Typical Ambient Temperature, Tantalum Temperature Coefcient Thread, Threaded Thick Package Type Designation Tapping Truss Head Trimmer Turn, Turns Torsion Microcandela Microfarad Microhenry Microliter, Underwriters' Laboratories, Inc. Unhardened Variable, Violet, Volt, Voltage Vacuum, VoltsAlternating Current Variable VoltsDirect Current Watt, Wattage, White, Wide, Width With Switch Wire Wound By (Used withDimensions), Reactance Yttrium-Iron-Garnet Zener
Chapter 12
495
MULTIPLIERS Abbreviation T G M k da d c Prex tera giga mega kilo deka deci centi Multiple 1012 109 106 103 10 101 102 Abbreviation m n p f a Prex milli micro nano pico femto atto Multiple 103 106 109 1012 1015 1018
496
Chapter 12
Chapter 12
497
Table 12-3
Value ()
Part Number 0757-0346 0757-0378 0757-0379 0698-3427 0698-3428 0757-0382 0757-0294 0698-3429 0698-3430 0698-3431 0698-3432 0698-3433 0757-0180 0698-3434 0698-3435 0757-0316 0698-4037 0757-0394 0757-0395 0757-0276 0757-0397 0757-0398 0757-0399 0757-0400 0757-0401 0757-0402 0757-0403 0698-3437 0698-3438 0757-0405 0698-3439 0698-3440
CD
Value ()
Part Number 0698-3447 0698-0082 0757-0416 0757-0417 0757-0418 0757-0419 0757-0420 0757-0421 0757-0422 0757-0280 0757-0424 0757-0274 0757-0317 0757-1094 0757-0428 0757-0278 0698-0083 0698-0084 0698-3150 0698-0085 0698-3151 0757-0279 0698-3152 0698-3153 0698-3154 0698-3155 0757-0438 0757-0200 0757-0290 0757-0439 0757-0440 0757-0441
CD
10.0 11.0 12.1 13.3 14.7 16.2 17.8 19.6 21.5 23.7 26.1 28.7 31.6 34.8 38.3 42.2 46.4 51.1 56.2 61.9 68.1 75.0 82.5 90.9 100 110 121 133 147 162 178 196
2 0 1 0 1 6 9 2 5 6 7 8 2 9 0 6 0 0 1 7 3 4 5 9 0 1 2 2 3 4 4 7
422 464 511 562 619 681 750 825 909 1.0K 1.1K 1.21K 1.33K 1.47K 1.62K 1.78K 1.96K 2.15K 2.37K 2.61K 2.87K 3.16K 3.48K 3.83K 4.22K 4.64K 5.11K 5.62K 6.19K 6.81K 7.50K 8.25K
4 7 7 8 9 0 3 4 5 3 7 5 7 9 1 9 8 9 6 0 7 0 8 9 0 1 3 7 5 4 7 8
498
Chapter 12
Table 12-3
Value () 215 237 261 287 316 348 383 17.8K 19.6K 21.5K 23.7K 26.1K 28.7K 31.6K 34.8K 38.3K 42.2K 46.4K 51.1K 56.2K 61.9K 68.1K 75.0K 82.5K 90.9K
Part Number 0698-3441 0698-3442 0698-3132 0698-3443 0698-3444 0698-3445 0698-3446 0698-3136 0698-3157 0757-0199 0698-3158 0698-3159 0698-3449 0698-3160 0757-0123 0698-3161 0698-3450 0698-3162 0757-0458 0757-0459 0757-0460 0757-0461 0757-0462 0757-0463 0757-0464
CD 8 9 4 0 1 2 3 8 3 3 4 5 6 8 3 9 9 0 7 8 1 2 3 4 5
Value () 9.09K 10.0K 11.0K 12.1K 13.3K 14.7K 16.2K 100K 110K 121K 133K 147K 162K 178K 196K 215K 237K 261K 287K 316K 348K 383K 422K 464K
Part Number 0757-0288 0757-0442 0757-0443 0757-0444 0757-0289 0698-3156 0757-0447 0757-0465 0757-0466 0757-0467 0698-3451 0698-3452 0757-0470 0698-3243 0698-3453 0698-3454 0698-3266 0698-3455 0698-3456 0698-3457 0698-3458 0698-3459 0698-3460 0698-3260
CD 1 9 0 1 2 2 4 6 7 8 0 1 3 8 2 3 5 4 5 6 7 8 1 9
Chapter 12
499
Table 12-4
Value ()
Part Number 0757-0984 0575-0985 0757-0986 0757-0001 0698-3388 0757-0989 0698-3389 0698-3390 0698-3391 0698-3392 0757-0003 0698-3393 0698-3394 0698-3395 0698-3396 0698-3397 0698-3398 0757-1000 0757-1001 0757-1002 0757-0794 0757-0795 0757-0796 0757-0797 0757-0198 0757-0798 0757-0799 0698-3399 0698-3400 0757-0802 0698-3334 0757-1060
CD
Value ()
Part Number 0698-3404 0698-3405 0698-0090 0757-0814 0757-0815 0757-0158 0757-0816 0757-0817 0757-0818 0757-0819 0757-0159 0757-0820 0757-0821 0698-3406 0757-1078 0757-0873 0698-0089 0698-3407 0698-3408 0698-3409 0698-0024 0698-3101 0698-3410 0698-3411 0698-3412 0698-3346 0698-3348 0757-0833 0757-0834 0757-0196 0757-0835 0757-0836
CD
10.0 11.0 12.1 13.3 14.7 16.2 17.8 19.6 21.5 23.7 26.1 28.7 31.6 34.8 38.3 42.2 46.4 51.1 56.2 61.9 68.1 75.0 82.5 90.0 100 110 121 133 147 162 178 196
4 5 6 6 2 9 3 6 7 8 8 9 0 1 2 3 4 7 8 9 4 5 6 7 2 8 9 5 9 5 8 9
383 422 464 511 562 619 681 750 825 909 1.00K 1.10K 1.21K 1.33K 1.47K 1.62K 1.78K 1.96K 2.15K 2.37K 2.61K 2.87K 3.16K 3.48K 3.83K 4.22K 4.64K 5.11K 5.62K 6.19K 6.81K 7.50K
3 4 7 9 0 4 1 2 3 4 5 7 8 5 9 0 4 6 7 8 7 7 1 2 3 2 4 2 3 0 4 5
500
Chapter 12
Table 12-4
Value () 215 237 261 287 316 348 16.2K 17.8K 19.6K 21.5K 23.7K 26.1K 28.7K 31.6K 34.8K 38.3K 42.2K 46.4K 51.1K 56.2K 61.9K 68.1K 75.0K 82.5K 90.9K 100K 110K 121K 133K 147K
Part Number 0698-3401 0698-3102 0757-1090 0757-1092 0698-3402 0698-3403 0757-0844 0698-0025 0698-3415 0698-3416 0698-3417 0698-3418 0698-3103 0698-3419 0698-3420 0698-3421 0698-3422 0698-3423 0757-0853 0757-0854 0757-0309 0757-0855 0757-0856 0757-0857 0757-0858 0757-0367 0757-0859 0757-0860 0757-0310 0698-3175
CD 0 8 5 7 1 2 5 8 6 7 8 9 9 0 3 4 5 6 6 7 7 8 9 0 1 7 2 5 0 5
Value () 8.25K 9.09K 10.0K 12.1K 13.3K 14.7K 162K 178K 196K 215K 237K 261K 287K 316K 348K 383K 422K 464K 511K 562K 619K 681K 750K 825K 909K 1M 1.1M 1.21M 1.33M 1.47M
Part Number 0757-0837 0757-0838 0757-0839 0757-0841 0698-3413 0698-3414 0757-0130 0757-0129 0757-0063 0757-0127 0698-3424 0757-0064 0757-0154 0698-3425 0757-0195 0757-0133 0757-0134 0698-3426 0757-0135 0757-0868 0757-0136 0757-0869 0757-0137 0757-0870 0757-0138 0757-0059 0757-0139 0757-0871 0757-0194 0698-3464
CD 6 7 8 2 4 5 2 9 0 7 7 1 0 8 9 5 6 9 7 3 8 4 9 7 0 4 1 8 8 5
Chapter 12
501
Table 12-5
Reference Designator A1A1
A1A2
Rotary Pulse Generator All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers Old Style New Style 5
0960-0745 1990-1525
A1R1
Volume-Intensity Dual Pot Assembly All 8590 E-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers All 8590 L-Series spectrum analyzers 1 08590-60133 08590-60041
A2
Display All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers 7 08590-60377
A3
Input Attenuator 8590L and 8591E 8591C Front End (Includes A3A1 through A3A15) 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E See A3A5 2 2 0955-0453 0955-0700
A3A1
A3A1A1
Comb Generator Board (part of A3A1) 8592L, 8593E, and 8596E 4 08569-60026
A3A2
Microwave Switch 8592L and 8593E 8592L and 8593E, (Option 026 and Option 027) 3 4 33314-60011 33314-60012
A3A3
Low-Pass Filter 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 3 0955-0420
502
Chapter 12
Table 12-5
Reference Designator A3A4
A3A5
Input Attenuator 8592L and 8593E 8592L and 8593E, (Options 026 and 027) 8594E, 8594L, 8594Q, and 8595E 8596E 4 3 1 2 33321-60039 33321-60038 33321-60036 33321-60037
A3A6
Dual Mixer 8592L, 8593E, 8595E, and 8596E Low Band Mixer 8594E, 8594L, and 8594Q 5086-7748 2 5086-7952
A3A7
Yig-Tuned Oscillator (YTO) 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 3 5086-7903
A3A8
Yig-Tuned Filter (YTF) 8592L and 8593E 8592L and 8593E, (Options 026 and 027) Switched Yig-Tuned Filter (SYTF) 8595E 8596E 2 5 5086-7803 5086-7880 8 9 0955-0277 5086-7850
A3A9
Bandpass Filter Assembly 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 0 9135-0252
A3A10
A3A11
A3A12
Chapter 12
503
Table 12-5
Reference Designator A3A13
A3A14
A3A15
Tracking Generator 8593E, 8594E, 8595E, and 8596E Tracking Generator Exchange 8593E, 8594E, 8595E, and 8596E 3 5086-6917 5 5086-7917
A3A16
A3A17
A3A18
A4
A5
A5A1
A6
Yig-Tuned Oscillator (YTO), 2-4 GHz 8590L 8591C and 8591E 0955-0974 0955-0974
A6A1
504
Chapter 12
Table 12-5
Reference Designator A7
A7A1
A8
Power Supply All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers Power Supply Exchange All 8590 E-Series and L-Series spectrum analyzers and 8591C cable TV analyzers 08590-69005 8 5062-8229
A9
Third Converter 8590L and 8591E (50 input) 8590L, 8591C, and 8591E (75 input) All 8590 E-Series and 8592L spectrum analyzers, and 8594Q QAM analyzers 9 6 8 08591-60081 08590-60344 08593-60062
A10
LO Distribution Amplier (LODA) Control for Option 009 8593E, 8594E, 8595E, and 8596E Tracking Generator Control for Option 010 8593E, 8594E, 8595E, and 8596E
5062-8232
5063-0635
A11
Bandwidth Board All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers 08590-60398
A12
Amplitude Control Board All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers 7 08590-60105
Chapter 12
505
Table 12-5
Reference Designator A13
A14
Log Amplier Board All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers 08590-60386
A15
Motherboard 8590L, 8592L, and 8594L 8591C, 8591E, 8593E, 8594E, 8594Q, 8595E, and 8596E 3 6 08590-60218 08591-60062
A16
Processor Board All 8590 E-Series and L-Series spectrum analyzers and 8591C cable TV analyzers 8594Q 4 08590-60416 08594-60054
A16A1
2X Memory Board All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzer, and 8594Q QAM analyzers 08590-60399
A16A1BT1
Battery All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzer, and 8594Q QAM analyzers 1420-0394
A17
Memory Card All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers 9 08590-60396
A22
10 MHz Reference All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers and 8594Q QAM analyzers equipped with Option 704 8 08591-60048
10 MHz Precision Frequency Reference 8591E, 8593E, 8594E, 8594Q, 8595E and 8596E (Option 004), 8591C, and 8594Q 4 1813-0644
506
Chapter 12
Table 12-5
Reference Designator A24
N/A
A24A1
A24A2
A24A3
A24A4
A24FL1
A25
Counter Lock Board 8590L, 8591C, 8591E, 8592L, 8593E, 8595E, and 8596E 8594E, 8594L and 8594Q 08591-60098 08594-60088
A25A1
Sampler Assembly 8590L, 8591C, 8591E, 8592L, 8593E, 8595E, and 8596E 8594E, 8594L, and 8594Q 5086-7806 5086-7956
A26
A40
IB and Parallel Interface Board All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers and 8594Q QAM analyzers equipped with Option 041 0 08590-60366
A40A1
IB and Parallel Interface Connector All 8590 E-Series and L-Series spectrum analyzers and 8591C cable TV analyzers and 8594Q QAM analyzers (Option 041) 0 08590-60368
Chapter 12
507
Table 12-5
Reference Designator A41
A41A1
RS-232 and Parallel Interface Connector All 8590 E-Series and L-Series spectrum analyzers and 8591C cable TV analyzers and 8594Q QAM analyzers (Option 043) 1 08590-60369
A41/A43
Dual I/O Board 8591C, 8591E, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E (Option 041 or Option 043) 08590-60366
A101
Fast ADC Board 8591C, or 8591E, 8593E, 8594E, 8595E, and 8596E (Option 101) 4 5062-7079
A102
Demod/TV-Sync Board 8591C, or 8591E, 8593E, 8594E, 8595E, and 8596E (Option 102) 6 5062-1982
A103
Demod/QPD Board 8591E, 8593E, 8594E, 8595E, and 8596E (Option 103) 7 5062-1983
A105
Gated Video Board 8591C, or 8591E, 8593E, 8594E, 8595E, and 8596E (Option 105) 2 5062-7754
A110
CELLULAR DEMOD/TV 8591E, 8593E, 8594E, 8595E, and 8596E (Option 110) 1 5063-0234
A111
A119
Noise Card Board 8591C, 8591E, 8593E, 8594E, 8595E, and 8596E (Option 119) 3 5062-8258
A130
Narrow Bandwidth Board 8591C, 8591E, 8593E, 8594E, 8595E, and 8596E (Option 130) 7 08590-60352
508
Chapter 12
Table 12-5
Reference Designator A151
A160
DSP Assembly 8591E, 8593E, 8594E, 8595E, and 8596E (Option 160) without rmware 8591E, 8593E, 8594E, 8595E, and 8596E (Option 160) with rmware 5062-8245 08591-60101
A163
DSP Assembly (GSM/DCS Formats) 8591E, 8593E, 8594E, 8595E, and 8596E (Option 163) with rmware 5062-0577
A190A1
A190A2
A190A3
AT1
Isolator (2 to 4 GHz) 8591C only 10 dB Attenuator 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 0960-0084 0955-0122
AT2
50 Ohm Termination 8590L and 8591E 8593E, 8594E, 8595E, and 8596E 9 3 0960-0053 1250-2289
B1
Fan (includes cable assembly) All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers 2 5063-0243
F1
Fuse 5.0 A 250 V F All 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers 3 2110-0709
Chapter 12
509
Table 12-5
Reference Designator FL1
J1
Input 50 8590L, 8591E, 8594E, 8594L, and 8594Q 8592L, 8593E, 8595E, and 8596E 8592L and 8593E Option 026 8592L and 8593E Option 027 Input 75 (MLA) 8591C 8590L, and 8591E (Option 001) 0955-1005 0955-1005 6 1250-2191 1250-2180 08673-60040 5063-0230
J3
RF OUT 50 8590L and 8591E Option 010 8593E, 8594E, 8594Q, 8595E, and 8596E Option 010 TG-RF OUT 50 8593E Option 026 RF OUT 75 (MLA) 8590L, 8591C, and 8591E Option 011 0955-1005 5061-5311 1250-2191 1250-2191
510
Chapter 12
Table 12-6
Reference Designator W1
Replaceable Cables
Description CABLE ASSY WITH LINE SWITCH, A1 TO A16J9 All 8590 Series analyzers 2 5062-4809 CD Part Number
W2
CABLE ASSY, DISPLAY VIDEO, A2 TO A16J8 All 8590 -Series analyzers 7 5063-0602
W3
CABLE ASSY, INTENSITY, WITHOUT POTENTIOMETER, A1R1 TO A2 All 8590 Series analyzers 4 08590-60037
W4
CABLE ASSY, KEYBOARD, A1A1J1 TO A16J10 All 8590 Series analyzers 6 08592-60037
W5
CABLE ASSY, PROBE POWER, WITH FILTER, WITHOUT CONNECTOR, A1A1 TO A1 All 8590 Series analyzers 3 8120-8152
W6
JUMPER, 10 MHz REF OUTPUT TO EXT REF IN All 8590 Series analyzers 5 1250-1499
W7
CABLE ASSY, CAL OUTPUT, A9J2 TO A1 8590L, 8591C, and 8591E (75 ) 8590L, 8592L, 8591E, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E (50 ) 3 6 08590-60028 8120-5052
W8
CABLE ASSY, 600 MHz DRVR, A9J5 TO A3A4J4 8592L, 8593E, 8594E, 8594Q, 8595E, and 8596E CABLE ASSY, 600 MHz DRVR, A9J5 TO A5J5 8590L, 8591C, and 8591E 8 8120-5020 8 8120-5020
W9
CABLE ASSY, 2ND IF, A5J2 TO A9J4 8590L, 8591C, and 8591E CABLE ASSY, 2ND IF, A3A9J2 TO A9J4 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 1 8120-5106 1 8120-5023
Chapter 12
511
Table 12-6
Reference Designator W10
W11
CABLE ASSY, ATTENUATOR OUTPUT, A3 TO A4J1 8590L and 8591E 8591C CABLE ASSY, ATTENUATOR OUTPUT, A3A5J2 TO A3A2J1 8592L and 8593E 8594E, 8594L, and 8594Q 8595E, and 8596E 7 3 4 08592-20066 08594-20002 08595-20002 5 08590-20110 08591-20057
W12
CABLE ASSY, YTO DRIVER, A7J3 TO A3A7 8590L, 8591C, and 8591E CABLE ASSY, YTO DRIVER, A7J3 TO A3A7 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 4 08592-60019 2 08590-60035
512
Chapter 12
Table 12-6
Reference Designator W13
W14
CABLE ASSY, COUNTER LOCK/MEMORY CARD CONTROL, A16J2 TO A17J1/A25J4 All 8590 Series analyzers except 8590L with Option 713 0 8120-5535
W15
CABLE ASSY, PRECISION FREQUENCY REFERENCE POWER, A25J3 TO A22 All 8590 E-Series and L-Series spectrum analyzers, and 8594Q QAM analyzers 8591C, 8594Q, or 8590 Series analyzers (Option 004) 5 6 5063-0204 5063-0205
W17
CABLE ASSY, 10 MHz OUTPUT, A22J1 TO REAR PANEL 8590 E-Series and L-Series spectrum analyzers and 8591C 7 08591-60013
W18
CABLE ASSY, EXT REF IN 8590 E-Series and L-Series spectrum analyzers and 8591C 3 08591-60009
W19
CABLE ASSY, COUNT IF, A25J5 TO A15J8 8590 E-Series and L-Series spectrum analyzers, and 8594Q QAM analyzers 6 08591-60012
W20
CABLE ASSY, SAMPLER LO DRIVE A3A10 TO AT1 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 5 8120-5068
Chapter 12
513
Table 12-6
Reference Designator W21
W22
CABLE ASSY, FIRST CONVERTER OUTPUT, A4J3 TO FL1 8590L and 8591E 8591C CABLE ASSY, FIRST MIXER HI BAND, A3A6J5 TO A3A4J3 8592L, 8593E, 8595E, and 8596E 3 8120-5067 2 2 08590-20203 08590-20112
W23
CABLE ASSY, LPF TO SECOND CONVERTER, FL1 TO A5J1 8590L and 8591E CABLE ASSY, AT1 TO SECOND CONVERTER, AT1 TO A5J1 8591C CABLE ASSY, 2ND CONV. 321.4 MHz OUT, A3A4J2 TO A3A9J1 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 2 8120-5058 08590-20113 3 08590-20204
W24
CABLE ASSY, YTO OUTPUT, A6 TO A4J2 8590L, 8591C, and 8591E CABLE ASSY, COMB GENERATOR OUTPUT, A3A1J1 TO A3A11 8592L, 8593E, 8596E 1 08592-60058 1 08590-20111
514
Chapter 12
Table 12-6
Reference Designator W25
W26
CABLE ASSY, TG TRIPLER OUTPUT, A24A1J2 TO A24A2J1 8590L, 8591C, and 8591E CABLE ASSY, 2.9 GHZ LPF IN, A3A2 TO A3A3J1 8592L and 8593E 8595E and 8596E 8 5 08592-20067 08595-20003 4 08590-20180
W27
CABLE ASSY, TG LO INPUT, A4J5 TO A24A2J2 8590L and 8591E (Option 010) CABLE ASSY, 2.9 GHZ LPF OUT, A3A3J2 TO A3A6J 8592L, 8593E, 8595E, and 8596E 8593E, 8595E, and 8596E (Option 010) 8594E, 8594L, and 8594Q 5 2 4 08592-20048 08593-20010 08594-20003 9 08590-20177
W28
CABLE ASSY, CONVERTER OUTPUT, A24A2J3 TO FL1 8590L AND 8591E (Option 010) CABLE ASSY, YTF OUT, A3A8J2 TO A3A6J3 8592L AND 8593E CABLE ASSY, SYTF OUT, A3A2J2 TO A3A6J3 8595E AND 8596E 5 1250-1788 5 1250-1788 0 08590-20178
Chapter 12
515
Table 12-6
Reference Designator W29
W30
CABLE ASSY, LPF OUTPUT, FL1J2 TO A3A4J1 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E CABLE ASSY, AMP-DET/ATTN, A24A3 TO A24A4 8591E Options 010 and 011 8 08590-20179 4 08593-20012
W31
CABLE ASSY, ATTN/TG OUT, A24A4 TO J3 8591E Options 010 and 011 CABLE ASSY, YTO OUTPUT, A3A7J2 TO A3A13J1 8592L, 8593E, 8595E, and 8596E 8593E, 8594E, 8595E, and 8596E (Option 009, 010) CABLE ASSY, YTO OUTPUT, A3A7J2 TO A3A13J1 8594E, 8594L, and 8594Q 9 08594-20008 2 4 08592-20053 08593-20004 7 5062-0761
W32
CABLE ASSY, ISOLATOR OUTPUT, A3A13J2 TO A3A10 8592L 8593E, 8595E, and 8596E 8594E, 8594L, and 8594Q CABLE ASSY, AMPLIFIER-DETECTOR OUTPUT, A24A3 to A7A1J2 8590L, 8591E, Option 010 and Option 011 8591C, Option 011 8120-5105 8120-5105 5 1 0 08592-20064 08592-20052 08594-20009
516
Chapter 12
Table 12-6
Reference Designator W33
08592-20065
08593-20003
W34
CABLE ASSY, YTF DRIVER, A7J301 TO A3A8 8592L, 8593E, 8595E, and 8596E 3 08592-60018
W35
CABLE ASSY, 2ND CONV DRIVER, A7J4 TO A3A4J5 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 3 5062-6434
W36
CABLE ASSY, ATTENUATOR DRIVER, A7J5 TO A3A5 8592L, 8593E, 8594E, 8594L, 8594Q, 8595E, and 8596E 8 5062-4813
W37
CABLE ASSY, COMB GENERATOR OUTPUT, A3A12 TO J2 8592L, 8593E, AND 8596E 2 08592-20061
W38
CABLE ASSY, LODA DRIVE, A10J2 TO A3A14 8593E, 8594E, 8595E, and 8596E (Option 010) 9 08593-60009
W39
CABLE ASSY, TRACKING GENERATOR CONTROL, A10J1 TO A3A15J 8593E, 8594E,8595E, and 8596E (Option 010) 5 5062-7715
W40
CABLE ASSY, 600 MHz TRACKING GENERATOR DRIVE, A3A15J8 TO A9J7 8593E, 8594E, 8595E, and 8596E (Option 010)
8120-5020
Chapter 12
517
Table 12-6
Reference Designator W41
W42
CABLE ASSY, FIRST LO OUTPUT, A3A15J3 TO REAR PANEL 8593E, 8594E, 8595E, and 8596E (Options 009 and 010) 0 5062-7710
W43
CABLE ASSY, TG OUT, A3A16 TO J3 8593E, 8594E, 8595E, and 8596E (Option 010) 08593-20009
W47
CABLE ASSY, EXT ALC INPUT, REAR PANEL TO A3A15J6 8592L, 8593E, 8594E, 8595E, and 8596E (Option 010) 6 08560-60001
W48
CABLE ASSY, SWEEP + TUNE OUTPUT, A7J6 TO REAR PANEL 8592L, 8593E, 8594E, 8595E, and 8596E (Option 009) 6 8120-5401
W49
CABLE ASSY, TRACKING GENERATOR OUTPUT, A3A15J2 TO A3A16 8593E, 8594E, 8595E, and 8596E (Option 010) 8 08593-20008
W50
CABLE ASSY, TG ATTENUATOR DRIVE, A10J3 TO A3A16 8593E, 8594E, 8595E, and 8596E (Option 010) 5 8120-5530
W101
CABLE ASSY, TV TRIGGER OUTPUT, A102J2 TO REAR PANEL 8591E, 8593E, 8594E, 8595E, and 8596E (Option 101) 9 08591-60024
W102
CABLE ASSY, VOLUME CONTROL, A1R1 TO A102 8591C, or 8591E, 8593E, 8594E, 8595E, and 8596E (Option 102) 4 08590-60136
W104
CABLE ASSY, HEADPHONE, A102 TO REAR PANEL 8591C, 8591E, 8593E, 8594E, 8595E, and 8596E (Options 102 and 103) 3 08590-60135
518
Chapter 12
Table 12-6
Reference Designator W130
W131
CABLE ASSY, A130 TO A15J200 8591C, 8591E, 8593E, 8594E, 8595E, and 8596E (Option 130) 0 8120-5030
W132
W133
W134
CABLE ASSY, TV MONITOR OUT, A107J8 TO REAR PANEL Option 107 8 08591-60080
W135
CABLE ASSY, TUNER CONTROL, A107 TO A26 AND A16 TO A107P11 Option 107 1 5062-3446
W136
CABLE ASSY, TV IN, FRONT PANEL TO A26 8591C (Option 107) CABLE ASSY, TV IN, REAR PANEL TO A26 8590 E-Series (Option 107) 0 5062-3445 1 08591-60075
W137
W138
CABLE ASSY, GATE TRIGGER IN, REAR PANEL TO A16J501 8590 E-Series (Option 105) factory installed time-gated analyzers only 2 08591-60076
W139
CABLE ASSY, GATE OUT, REAR PANEL TO A16J500 8590 E-Series (Option 105) factory installed time-gated analyzers only 3 08591-60077
W190
CABLE ASSY, GREY, A190A1 PORT A TO A190A2 PORT A 8594Q (Option 190) 8120-6990
Chapter 12
519
Table 12-6
Reference Designator W191
W192
CABLE ASSY, VIOLET, A190A1 PORT C TO A190A2 PORT C 8594Q (Option 190) 8120-6990
W193
W194
CABLE ASSY, GREEN, A190A2 OUT TO IF ASSEMBLY J4 8594Q (Option 190) 8120-8334
W195
W196
W197
W198
W199
CABLE ASSY, PREAMP CONTROL, A3A17 TO A26J1 8594Q (Option 190) 08594-60065
W200
W201
W202
CABLE ASSY, A3A18 PORT 4 TO A26 INPUT 8594Q (Option 190) 08594-60065
W203
CABLE ASSY, A3A18 PORT 1 TO A26 OUTPUT 8594Q (Option 190) 08594-60065
520
Chapter 12
Ordering CLIPs
For ordering convenience, current CLIPs for a specic instrument are combined into Component-Level Information binders. The current set of CLIPs contains information supporting the instrument assemblies manufactured at the time this manual was printed, plus a packet containing general CLIP information. A complete set of CLIPs can be obtained by ordering the 8590 Series Analyzers Component-Level Repair Service Guide. Updated or replacement CLIPs may be ordered through your local Agilent Technologies Sales or Service ofce using the CLIP part number provided in Table 12-2.
NOTE
Chapter 12
521
Table 12-7
8590 E-Series and L-Series Spectrum Analyzer Board Assembly, Serial Number, and CLIP Part Number Cross-Reference
Assembly Instrument Serial Prex All serials All serials All serials All serials Board Assembly Part Number 08590-60201a 08569-60026a 08590-60117a 08590-60155a CLIP Part Number 08590-90204 08590-90132 08590-90117 08590-90117
A1A1 Keyboard A3A1A1 Comb Generator A5 Second Converter A5A1 Tripler A7 Analog Interface, RF 8590L 8591E 8591C 8592L 8593E 8593E 8594E/8594L 8594E 8595E/8596E 8595E/8596E A7A1 Tracking Generator Control A9 Third Converter Standard: 8590L
Serial Prex 3407A Serial Prex 3205A Serial Prex 3325A Serial Prex 3407A Serial Prex 3251A Serial Prex 3205A Serial Prex 3251A Serial Prex 3205A Serial Prex 3304A Serial Prex 3205A All serials
08590-60197a 08590-60197a 08591-60072a 08592-60097a 08592-60097a 08592-60073 08594-60030a 08594-60008 08594-60022a 08594-60008 08590-60196a
08590-90205 08590-90205 08591-90121 08592-90080 08592-90080 08592-90077 08594-90035 08594-90027 08595-90031 08595-90026 08590- 90157
08591-60081a
08591-90120
522
Chapter 12
Table 12-7
8590 E-Series and L-Series Spectrum Analyzer Board Assembly, Serial Number, and CLIP Part Number Cross-Reference
Assembly Instrument Serial Prex Serial Prex 3325A Serial Prex 3407A Serial Prex 3346A Serial Prex 3205A Serial Prex 3407A Serial Prex 3407A Serial Prex 3346A Serial Prex 3205A Board Assembly Part Number 08590-60344a 08591-60081a 08593-60061 08593-60021 08593-60062a 08593-60062a 08593-60061 08593-60021 08590-90146 08590-90146 08593-90054 08593-90054 CLIP Part Number 08590-90223 08591-90120
8591C 8591E 8591E 8591E 8592L 8593E, 8594E, 8594L, 8595E, and 8596E
Option 001: 8590L 8591E 8591E A10 Tracking Generator Control (Options E02 and 010) A10 LODA Control (Option 009) A11/A13 Bandwidth Filter A12 Amplitude Control A14 Log Amp All serials All serials Serial Prex 3520A 08590-60363a 08590-60105a 08590-60373 08590-90231 08590-90121 5961-0483 All serials 5062-8232a 08590-90266 Serial Prex 3407A Serial Prex 3407A Serial Prex 3205A All serials 08590-60344a 08590-60344a 08590-60193 5063-0635 08590-90223 08590-90223 08590-90146 5961-0480
Chapter 12
523
Table 12-7
8590 E-Series and L-Series Spectrum Analyzer Board Assembly, Serial Number, and CLIP Part Number Cross-Reference
Assembly Instrument Serial Prex Board Assembly Part Number CLIP Part Number
A15 Motherboard 8590L, 8592L, and 8594L Serial Prex 3308A Serial Prex 3205A 8591C, 8591E, 8593E, 8594E, 8595E, and 8596E Serial Prex 3308A Serial Prex 3205A A16 Processor/Video Serial Prex 3523A Serial Prex 3501A Serial Prex 3322A Serial Prex 3303A Serial Prex 3205A A16A1 Memory Assembly A17 Memory Card Reader (Option 003) A20 IB I/O (Options 021 and EO2) A21 RS-232 I/O (Option 023) A22 10 MHz Reference A25 Counter Lock 8590L and 8592L Serial Prex 3407A 08591-60073a 08591-90119 All serials 08591-60048a 08591-90086 All serials 08590-60109a 08590-90126 All serials 08590-60108a 08590-90125 Serial Prex 3523A All serials 08590-60218a 08590-60213 08591-60062a 08590-90217 08590-90206 08591-90103
524
Chapter 12
Table 12-7
8590 E-Series and L-Series Spectrum Analyzer Board Assembly, Serial Number, and CLIP Part Number Cross-Reference
Assembly Instrument Serial Prex Serial Prex 3249A Serial Prex 3324A Serial Prex 3205A Board Assembly Part Number 08591-60073a 08591-60061 08591-60053 08590-60366 5062-7079a 08591-90085 08590- 90257 5958-7174 CLIP Part Number 08591-90119
All others
A41/43 Dual I/O A101 Fast ADC (Options 101 and 301) A102 AM/FM TV Synch (Options 102 and 301) A103 Demodulator/ Quasi Peak Det (Option 103) A105 Gate Card (Option 105) A110 CT2 Board (Option 110) A111 Event Counter (Option 111)
All serials
5062-1982a
5958-7055
All serials
5062-1983a
5960-2504
All serials
5062-7754a
5960-2568
All serials
5063-0234a
5961-0481
All serials
5063-0214a
5961-0487
a. These CLIPS are included in the current component-level repair service guide.
Chapter 12
525
Figure 12-1
Item 1 2 3 4 5 6 7 8 9 10 11 12
Part Number 5041-8992 5041-8912 0515-1114 1460-2164 5021-6343 5021-6344 5021-8667 5001-8728 0515-1367 0515-1133 0515-0382 5022-1122
CD 8 2 2 8 7 8 2 4 0 5 4 2
Description HANDLE TRIM CAP SCREW-MACH M40.7 10MM-LG PAN-HD SPRING-CPR RING GEAR SOCKET GEAR HANDLE PLATE BACKUP PLATE SCREW-MACH M40.7 8MM-LG FLH-HD SCREW-MACH M50.8 16MM-LG SCREW-MACH M4 12MM-LG PAN-HD (8591C) HANDLE BUTTON (8591C)
526
Chapter 12
Figure 12-2
Item 1 2 3 4 5 6 7 8 9
Part Number 08590-60378 08590-00070 0515-1114 5041-8907 0900-0024 2190-0587 0515-1218 3050-0893 0515-1069 08590-40005
CD 3 2 5 8 3 7 9 6 2
Description COVER ASSEMBLY COVER SCREW-MACH M40.7 10MM-LG PAN-HD REAR FOOT O-RING .145-IN-ID .07-IN-XSECT-DIA SIL WASHER-LK HLCL 5.0 MM 5.1-MM-ID SCREW-SKT-HD-CAP M5.00.8 40 MM-LG WASHER-FLAT SCREW M4.010MM PAN-HD REAR FOOT SPACER
Chapter 12
527
Figure 12-3
Item
Part Number 08590-60379 08594-60058 08590-00062 5002-0690 08594-00013 0380-1900 3050-0105 0515-1038 3160-0309 6960-0076 08590-60135 0590-1251 6960-0150 6960-0082 08590-80012 7120-1232 5040-0345 1250-0118 08590-00065
CD
Description REAR FRAME ASSEMBLY (E-Series, L-Series, and 8591C only) REAR FRAME ASSEMBLY (8594Q only) REAR FRAME, DUAL I/O REAR FRAME, SINGLE I/O REAR FRAME (8594Q only) SPACER-ROUND .688L.166ID WASHER-FLAT .125ID SCREW-MACHINE M3.0 SEMPNPD FAN GRILL HOLE PLUG .312D MINI-JACK CONNECTOR AND NUT (OPTION 102) HEX NUT 15/32-32 HOLE PLUG 12.7DMM HOLE PLUG .922D (NOT ON OPTION 021 OR 023) HOLE COVER (NOT ON OPTION 021 OR 023) BATTERY LABEL INSULATOR CONNECTOR CONNECTOR RF FEMALE BNC PLATE (BLANK)
2 3 4 5 6 7 8 9 10 11 12 13 14 15
7 6 9 5 2 3 6 3 0 7 5 7 3
528
Chapter 12
Figure 12-4
Item 1 2 3 4 5 6 7 8 9 10 11 12
Part Number 5021-9907 5021-9908 0515-0372 5001-8782 5002-0671 5001-8769 1400-0510 2360-0526 0515-0372 5001-8767 5021-9395 5022-0931
CD 5 6 2 0 2 3 8 1 2 1 5 9
Description EXTRUSION, END PLATE ENCLOSURE EXTRUSION, CIRCUIT ENCLOSURE, TAPPED SCREW-MACH M3.0 X 8MM COVER-AMPLITUDE CONTROL COVER-BANDWIDTH FILTER COVER-LOG AMPLIFIER CABLE CLAMP SCREW-MACH 440 .250 PNPD SCREW-MACH M3.0 X 8MM CARD CAGE EXTRUSION, CIRCUIT ENCLOSURE EXTRUSION, CIRCUIT ENCLOSURE
Chapter 12
529
Figure 12-5
Item 1 2 3 4 5 6
CD 0 6 3 9 0 7 9 CHASSIS WASHER-FLAT.125ID
Description
SCREW-MACH M3.0 40 SEMPNPD, 8592L, 93E, 94E, 95E, AND 96E SCREW-MACH M3.0 45 SEMPNPD, 8590L, 8591C, AND 8591E SCREW-SQBU M3.0 6MM TORX-HD SCREW-MACH M4.0 8 PCFLPDS SCREW-ST M4.0 12 PNPD
530
Chapter 12
Figure 12-6
Item 1 2 3 4 5
CD 2 3 2 9 1
Description SCREW-MACH M3.0 8 CWPNPDS FRONT END TO IF SUPPORT BRACKET SCREW-MACH M3.0 8 CWPNPDS WASHER-FLAT M4.0ID DISPLAY SHIELD
Chapter 12
531
Figure 12-7
Item 1 2 3 4 5 6 7 8 9 10
Part Number 0515-0372 0535-0031 0380-1363 0515-1453 0515-0808 0515-1110 0515-0898 5002-0635 0515-0372
CD 2 2 6 0 9 8 7 8 2
Description SCREW-MACH M3.0 8 CWPNPDS NUT-MACH WLKWR M3.0 (STANDARD ONLY) STANDOFF (STANDARD ONLY) SCREW-MACH M3.0 12 SEMPNPD SCREW-ST M4.0 12 PNPD SCREW-MACH M3.0 12 PCPNPDS SCREW-MACH M4.0 6 PCPNPDS OCXO MOUNTING BRACKET (OPTION 004 or 8591C) SCREW-MACH M3.0 8 CWPNPDS WASHER
532
Chapter 12
Figure 12-8
Item 1 2 3 4 5
CD 8 0 4 3 0
Description IMPACT COVER ASSEMBLY GASKET, 3.2 FT. CATCH LATCH SCREW-MACH M2.5 6MM-LG PAN-HD DRAW CATCH ASSEMBLY
Chapter 12
533
Foldout 12-1 Parts Identication 8592L, 8593E, 8593E, 8596E A3 Front-End Assembly
Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Part Number 0515-0372 2360-0123 5022-0179 0515-0382 1400-0024 0515-1462 0520-0174 0515-1373 0515-1946 2190-0654 0535-0069 0515-0973 0535-0070 5002-0700 CD 2 4 7 4 9 3 3 5 8 5 6 9 9 8 Description SCREW-MACH 3.0 8MM PAN-HD SCREW-MACH 6-23 .625 PAN-HD WASHER, EYO SHOULDER SCREW-MACH 4.0 12MM TORX CABLE CLAMP SCREW-MACH 3.0 20MM PAN-HD SCREW-MACH 2-56 .25 PAN-HD SCREW-MACH 2.5 16MM PAN-HD SCREW-MACH 3.0 6MM PAN-HD LOCK WASHER, 2.0MM HEX NUT-MACH 2.0MM SCREW-MACH 2.0 20MM PAN-HD HEX NUT-MACH 2.5MM RF BRACKET - 4 PIECES
534
Chapter 12
Foldout 12-4 Parts Identication 8590L, 8591C, and 8591E Tracking Generator Assembly
Item 1 2 3 4 5 Part Number 08590-00037 5002-0621 0515-0433 0515-1227 0520-0174 CD 8 2 6 8 3 Description TG BRACKET TG ATTENUATOR BRACKET SCREW-MACH 4.0 8MM TORX SCREW-MACH 3.0 6MM TORX SCREW-MACH 2-56 .25 PAN-HD
Foldout 12-5 Parts Identication 8590 C-Series, E-Series and L-Series Front Frame Assembly
Item 1 2 3 4 5 6 7 8 Part Number 0515-1622 5041-8914 5041-8962 1000-0846 1000-0897 0370-3069 0370-3079 0535-0082 2190-0016 08590-00060 08590-00061 08591-00015 08591-00016 08591-00033 08592-00024 08592-00025 08593-00010 08593-00009 08594-00002 CD 7 4 2 5 6 2 4 3 3 Description SCREW-SKT-HD-CAP M40.7 8 MM-LG DISPLAY BEZEL ASSEMBLY DISPLAY BEZEL ASSEMBLY, (Option 103) DISPLAY FILTER DISPLAY FILTER (OPTION 103) RPG TUNING KNOB INTENSITY KNOB, 8590 L-SERIES ONLY NUT-HEX-3/832 WASHER-LK INTL T 3/8 IN .377-IN-ID 859OL FRONT PANEL-DRESS 8590L FRONT PANEL-DRESS (OPTION 001) 8591E FRONT PANEL-DRESS 8591E FRONT PANEL-DRESS (OPTION 001) 8591C FRONT PANEL-DRESS 8592L FRONT PANEL-DRESS 8592L FRONT PANEL-DRESS (OPTION 026, 027) 8593E FRONT PANEL-DRESS 8593E FRONT PANEL-DRESS (OPTION 026, 027) 8594E FRONT PANEL-DRESS
3 4 5
0 7 1
Chapter 12
535
Foldout 12-5 Parts Identication 8590 C-Series, E-Series and L-Series Front Frame Assembly (Continued)
Item Part Number 08594-60017 08594-60053 08595-00002 08596-00001 5060-0467 0590-1251 1250-1666 0515-2145 5062-4806 2190-0067 2950-0216 5041-8917 5041-8936 5022-0200 5022-1114 8160-0520 0535-0082 2190-0016 2950-0043 1250-2180 5063-0230 08673-60040 23 24 25 26 27 28 0515-0430 0515-2145 5041-8989 5041-8984 5041-8924 5041-8925 0515-1934 08590-60133 08590-60041 3050-0103 5021-9320 5041-1682 CD 1 2 2 6 6 8 1 9 4 7 7 5 7 3 3 8 3 7 9 3 1 3 8 6 7 4 1 0 4 6 Description 8594L FRONT PANEL-DRESS 8594Q FRONT PANEL-DRESS (OPTION 190) 8595E FRONT PANEL-DRESS 8596E FRONT PANEL-DRESS PROBE POWER CONNECTOR (M) NUT-SPCLY 15/32-32-THD (SECURES W7) ADAPTER SMA F TO SMA F SCREW-MACH M3.0 8MM-LG PAN-HD BUMPER KIT (SET OF 4) WASHER NUT-HEX 1/4 36 MEMORY CARD BEZEL MEMORY CARD BEZEL (SOLID) FRONT FRAME CASTING FRONT FRAME CASTING (8591C) GASKET-EMI, 3 FT. NUT-HEX 4.0MM LOCK WASHER .377ID HEX NUT 3/8 32 ADAPTER,(F) SMA TO (F) TYPE N (INPUT 50) ADAPTER, (F) SMA TO (F) TYPE N (OPTION 027) ADAPTER, (M) APC 3.5 TO (F) APC 3.5 (OPTION 026) SCREW M3.06MM TORX SCREW M3.08MM TORX (OPTION 026, 027) 8590 L-SERIES RUBBER KEYPAD 8590 E-SERIES RUBBER KEYPAD VOLUME KNOB, 8590 E-SERIES ONLY INTENSITY KNOB, 8590 E-SERIES ONLY SCREW M2.56MM TORX 8590 E-SERIES VOLUME-INTENSITY DUAL-POT ASSEMBLY 8590 L-SERIES INTENSITY POT ASSEMBLY WASHER-FLAT .250ID12 BRACKET - RF INPUT CONNECTOR (OPTION 026) KEYCAP "LINE"
9 10 11 12 13 14 15 16 16 17 17 18 19 20 21 22
29 30 31
536
Chapter 12
Chapter 12
537
538
Chapter 12
540
Chapter 12
542
Chapter 12
544
Chapter 12
546
Chapter 12
548
Chapter 12
550
Chapter 12
552
Chapter 12
13
Softkey Descriptions
Refer to this chapter for an explanation of the instrument passcodes, the two types of instrument correction data, and the service-related softkeys that are available after pressing CAL.
553
The calibration, service, and diagnostic softkey functions are listed alphabetically. A list of service-related softkeys follows each softkey description, when appropriate. Refer to the description of each related softkey to understand how the softkey functions interrelate.
Correction-constant data can be lost if the SERVICE CAL softkeys are used improperly. Refer to the appropriate softkey description for instructions on softkey usage.
554
Chapter 13
Figure 13-1
Calibration, Service, and Diagnostic Softkey Tree: 8590 E-Series and L-Series, and 8594Q QAM Analyzers
Chapter 13
555
Figure 13-2
556
Chapter 13
37 Hz Passcode
Use the 37 Hz passcode to perform the following functions. Replace the current calibration-factors with the default calibration data provided by DEFAULT CAL DATA. Check the accuracy of the 10 MHz frequency reference using VERIFY TIMEBASE. Refer to the 10 MHz reference accuracy verication test in the calibration guide for your instrument. Bypass the CAL OUT set-up check within individual self-calibration routines.
2001 Hz Passcode
Use the 2001 Hz service passcode to perform the following functions. Set the start frequency, stop frequency, and step size of the atness correction points using INIT FLT. Edit the atness-correction constants using EDIT FLATNESS. Edit the A12 amplitude control step-attenuator correction constants using SET ATTN ERROR. Set the correction factor for the 10 MHz frequency reference (standard timebase). Modify the displayed power units that appear when the instrument is rst turned on using STOR PWR ON UNITS. Bypass the CAL OUT signal check within individual self-calibration routines.
Entering a Passcode
The passcode must be entered before pressing CAL. Press the following keys to enter a passcode.
FREQUENCY 37, HZ or 2001, HZ CAL
When SRVC appears in the lower-left corner of the display, the passcode has been accepted.
Chapter 13
557
558
Chapter 13
The frequency of the 10 MHz standard timebase, a room temperature crystal oscillator (RTXO), is controlled by a DAC located on the A25 counterlock assembly. Refer to the 10 MHz reference procedure in Chapter 2 to adjust the timebase DAC using CAL TIMEBASE.Analyzers equipped with a A24 Tracking Generator (Options 010 and 011) use two correction constants to improve the performance of the external automatic level control (ALC). Refer to 19. Entering External ALC Correction Constants for Option 010 or 011 and 21. Correcting for External ALC Error for Option 010 or 011 in Chapter 2 for details about the CALTGX slope and offset correction constants. The initial service-calibration correction constants are stored by the factory in nonvolatile memory on the A16 assembly. After shipment, new correction constants must be manually generated whenever an adjustment or repair affects frequency response or A12 amplitude control step attenuation. Refer to After an analyzer repair in Chapter 4 for the specic assemblies that affect frequency response. Whenever the A16 processor/video assembly is replaced, new correction constants must also be generated, unless valid correction constants were saved prior to the instrument failure. Note that the manual generation of the new service-calibration correction constants is a lengthy procedure. Avoid having to generate new correction constants manually by periodically recording the current correction constants. Refer to Chapter 3 for information about backing up correction constants. Refer to the softkey descriptions in this section for specic information about the use and modication of the correction-constant data.
Chapter 13
559
560
Chapter 13
level control in the tracking generator. Refer to the modular offset and gain adjustment procedures in Chapter 2 for information on the use of ALC TEST.
Chapter 13
561
Display Label RF Atten 3rd Conv DAC 21.4 Gain Cal Atten Lin Gain Nbw S Gain*
Description RF Attenuator Variable Gain Step Gain Calibration Attenuator Linear Gain Nbw Step Gain
AUXB Auxiliary B
DESCRIPTION Note that this function is available for 8591C cable TV analyzers and 8591E spectrum analyzers only. Displays the voltage level at an unused input to the Test Point MUX circuitry located on the A7 analog interface assembly. The AUX B input to the A16 processor/video assembly oats to a high positive voltage. When AUX B is selected, a horizontal line above the top graticule line, is displayed.
562
Chapter 13
self-calibration, service-diagnostics, and service-calibration functions. A passcode is required for access to specic softkeys. RELATED SOFTKEYS Refer to Figure 13-1 and Figure 13-2 at the beginning of this chapter for the softkeys that are available after pressing CAL.
Chapter 13
563
564
Chapter 13
While CAL: 3dB BW is displayed, the following calibrations are performed: t The 3 dB and 6 dB EMI resolution bandwidths are measured. t DACs on the A7 analog interface assembly adjust the bandwidth of the LC and crystal lters on the A11 and A13 Bandwidth Filter assemblies. t The DAC bandwidth correction factors are stored on the A16 processor/video assembly. While CAL: ATTEN is displayed, the following calibrations are performed: t The amplitude error of the 10 dB step gains on the A12 amplitude control assembly and the 10 dB linear gains on the A14 Log Amplier/Detector assembly are corrected. The 10 dB step gain on the A12 assembly is the amplitude reference. Step-gain and linear-gain errors are corrected with digital offsets of the video signal on the A16 assembly. The video offsets are stored as correction factors. t Amplitude errors for the A3 Input Attenuator are corrected. The 10 dB step attenuator is the amplitude reference. Input attenuator errors are corrected with digital offsets of the video signal on the A16 assembly. The attenuator offsets are stored as correction factors.
Chapter 13
565
While CAL: LOG AMP is displayed, the following calibrations are performed: t The log delity is measured in 1 dB steps. t Errors are corrected with digital offsets of the video signal on the A16 assembly. t The video offsets are stored as correction factors. If a failure occurs during this calibration routine, an error message is displayed. Refer to Chapter 14 for the description of displayed error messages. Be sure to press CAL STORE after running the CAL AMPTD routine. RELATED SOFTKEYS DISPLAY CAL DATA
CAL FREQ CAL FREQ & AMPTD CAL FETCH CAL STORE CORRECT ON OFF
factors from the area of memory that retains data when the analyzer is turned off and places it in working (volatile) RAM memory.
CAUTION
Pressing CAL FETCH clears the correction-data-error ag and can permit erroneous data to be stored. Use CAL FETCH only for troubleshooting. RELATED SOFTKEYS CAL AMPTD
CAL FREQ CAL FREQ &TD CAL STORE
566
Chapter 13
Note that it is normal for the FREQ UNCAL error message to appear briey during CAL FREQ. During the CAL FREQ self-calibration routine, messages are sequentially displayed, indicating a specic calibration activity. Each function performed by CAL FREQ is described below. Before the calibration routine starts, an instrument setup check is performed. The CAL OUT signal must be within 300 MHz 50 MHz and greater than or equal to 45 dBm to pass the setup check. If the CAL OUT signal is not present, the routine stops. If the DEFAULT CAL DATA correction factors are in use, a frequency offset may occur that prevents the CAL OUT signal from being found. If desired, the instrument setup check can be bypassed. Enter the 37 Hz passcode before pressing CAL and CAL FREQ. Refer to the CAL SIGNAL NOT FOUND error message description in Chapter 14 for more information. While CAL: SWEEP is displayed: t The sweep ramp is calibrated. t The values are stored as correction factors. While CAL: FREQ is displayed: t Harmonics of the CAL OUT signal are used to adjust the start and stop end-points for the A6 YTO. t The end-point values are stored as correction factors.
Chapter 13
567
While CAL: SPAN is displayed: t The main-coil sweep sensitivity and span attenuator are adjusted for LO spans greater than 10 MHz. t The adjustment routine checks for three displayed signals: 0 Hz, 300 MHz, and 600 MHz. The test passes if all three signals are found within a 750 MHz span. FM DAC Error Check: t The analyzer changes the center frequency setting to move the displayed signal four divisions. t If the signal moves within 1.25 divisions of the expected display position, the FM coil drive is within tolerance. Instruments equipped with Option 102, AM/FM Speaker and TV Sync Trigger, display the message CAL: FM GAIN + OFFSET. While CAL: FM GAIN + OFFSET is displayed: t The FM offset DAC is adjusted to position the signal at center-screen. This calibrates the center screen-position so that it represents an unmodulated signal. t The top to bottom screen deviation of a demodulated FM signal, referenced to center screen, is calibrated for a specic frequency deviation. A demodulated signal with a 100 kHz frequency deviation is simulated using the CAL OUT signal and a center frequency step of 100 kHz. The center frequency is stepped up and the FM gain is adjusted to position the signal at top screen. The center frequency is stepped down and the FM gain is adjusted to position the signal at bottom-screen.
568
Chapter 13
If a failure occurs during this calibration routine, an error message is displayed. Refer to Chapter 14 for descriptions of displayed error messages. Be sure to press CAL STORE after running the CAL FREQ routine. RELATED SOFTKEYS CAL AMPTD
CAL FREQ & AMPTD CAL FETCH CAL STORE DISPLAY CAL DATA FM GAIN FM OFFSET
Chapter 13
569
amplitude self-calibration routines. Connect CAL OUT to the analyzer input before initiating CAL FREQ & AMPTD. Be sure to press CAL STORE after running the CAL FREQ & AMPTD routine. RELATED SOFTKEYS CAL AMPTD
CAL FREQ CAL FETCH CAL STORE
instrument for the cable part number. The CAL MXR routine adjusts the bias-current DAC setting for the optimum displayed-signal amplitude, using the 100 MHz COMB OUT signal. The routine adjusts the bias current at several frequencies in each band; an optimum bias setting for that band is then determined. During the routine, TRACE A displays the frequency spectrum, while TRACE B displays the signal amplitude versus mixer-bias DAC setting. The 2001 passcode must be entered to activate the CAL MXR calibration routine. Enter the passcode before pressing the CAL key.
570
Chapter 13
Note that the CAL YTF routine must be performed before running the CAL MXR routine. New frequency response correction constants must also be generated whenever the CAL MXR routine is used. Refer to the CAL MXR adjustment routine in Chapter 2 for instructions on the use of the CAL YTF and CAL MXR routines. Once the CAL MXR routine is nished, the optimum bias-current value for each frequency band is displayed. The displayed values are the mixer bias DAC setting multiplied by 16. The DAC values are automatically stored as correction constants in nonvolatile memory on the A16 processor/video assembly. RELATED SOFTKEYS CAL YTF
EDIT FLATNESS
CAUTION
If the correction factors are not stored, they will be lost when the analyzer is turned off. RELATED SOFTKEYS CAL AMPTD
CAL FREQ CAL FREQ & AMPTD CAL FETCH CAL TIMEBASE CRT VERT POSITION CRT HORZ POSITION CAL TRK GEN
Chapter 13
571
572
Chapter 13
self-calibration routine. Be sure to perform the CAL AMPTD and CAL FREQ self-calibration routines before using CAL TRK GEN. Connect the RF OUT from the tracking generator to the RF INPUT using the CAL cable. Refer to Table 1-4 for the part number of the 50 CAL cable or 75 CAL cable. When the CAL TRK GEN routine is complete, press CAL STORE to store the tracking generator correction factors in nonvolatile memory. The correction factors cannot be viewed.
CAUTION
If the correction factors are not stored, they will be lost when the analyzer is turned off. RELATED SOFTKEYS CAL AMPTD
CAL FREQ CAL STORE
Chapter 13
573
574
Chapter 13
Chapter 13
575
If the error message POS-PK FAIL is displayed, the mean of the data from the positive-peak detector is less than the mean from the sample detector data. 2. Sample-Detector Check and error message SAMPLE FAIL. The test checks switching for the sample detector. The peak-to-peak amplitude of the noise oor for the sample detector should be wider than the noise oor for the positive-peak detector.The test detects the wider noise oor by statistically comparing the peak-to-peak amplitude of the noise oor for the sample detector to the noise oor for the positive-peak detector. If the error message SAMPLE FAIL is displayed, the standard deviation of the data from the sample detector is less than the standard deviation of the data from the positive-peak detector. 3. Video-Bandwidth Check and error message VID-BW FAIL. The test checks video-bandwidth switching by stepping the analyzer from the widest video bandwidth setting to the narrowest. The test detects switching activity by comparing the peak-to-peak amplitude of the noise oor for each video bandwidth. If the error message VID-BW FAIL is displayed, the peak-to-peak amplitude did not decrease when a narrower video bandwidth was switched on. 4. Resolution-Bandwidth Noise Check and error message RES-BW NOISE FAIL. The test makes a relative comparison of noise-oor amplitude for each resolution bandwidth. The test compares the noise-oor amplitude of each resolution bandwidth to that of the next, narrower resolution bandwidth. If the error message RES-BW NOISE FAIL is displayed, a decrease in the noise oor amplitude did not occur when the test switched to the narrower bandwidth. The test displays the narrower bandwidth as the test failure and displays all bandwidths that fail.
576
Chapter 13
5. A12 Step-Gain Check and error message STEP GAIN/ATTEN FAIL. The test checks the switching of the step gains on the A12 amplitude control assembly. The test does not check step-gain accuracy. The test steps the reference level from 60 dBm to +30 dBm, in 10 dB increments, with the input attenuator set to 60 dB. The test detects step-gain switching activity by comparing the noise-oor level for each 10 dB step gain. If the error message STEP GAIN/ATTEN FAIL is displayed, the displayed noise level did not decrease when the reference level was changed. 6. 3 dB Resolution-Bandwidth Check and error message RES-BW SHAPE FAIL. The test checks the 3 dB resolution bandwidth of the 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz, and 3 MHz bandwidth lters. If the error message RES-BW SHAPE FAIL is displayed, the bandwidths that are displayed with the error message are not within 20%. Resolution bandwidth accuracy is a characteristic, not a specication. Check the failed bandwidths manually using the 3 dB POINTS softkey.
Chapter 13
577
578
Chapter 13
The CORRECT ON OFF function does not affect the following DAC adjustments. Resolution bandwidth corrections. Reference level amplitude correction. YTO frequency and span corrections. 10 MHz reference (standard timebase) DAC correction. Refer to the CAL AMPTD description in this chapter and Chapter 4 when the displayed signal exhibits symptoms of either low or high gain. RELATED SOFTKEYS CAL AMPTD
CAL FREQ CAL STORE DEFAULT CAL DATA
Chapter 13
579
DACS DACs
DESCRIPTION Use DACS to change the DAC numbers of the span, YTO coarse-tune, YTO ne-tune, and YTO FM tune DACs located on the A7 analog interface assembly. The following terminator keys are used to select the desired DAC.
YTO Adjustment DAC Selection Span DAC YTO coarse-tune DAC YTO ne-tune DAC YTO FM tune DAC Terminator Key
GHz MHz kHz Hz
580
Chapter 13
Each DAC may be set to any integer between 0 and 4095 using the A1A2 knob on the front panel. Use the and keys to make large changes in the DAC settings. The key increments the DAC setting in an even binary progression of 0, 2, 4, 8, 16,, 4096. The key decrements the DAC setting in an odd binary progression of 4095, 2047, 1023, 511,, 1. Refer to Chapter 6 for additional LO troubleshooting information. RELATED SOFTKEYS COARSE TUNE DAC
FINE TUNE DAC X FINE TUNE DAC SERVICE DIAG MAIN SPAN
Chapter 13
581
582
Chapter 13
RELATED SOFTKEYS
DISPLAY CAL DATA CAL FREQ CAL AMPTD CAL TIMEBASE CORRECT ON OFF
Chapter 13
583
584
Chapter 13
The following text describes each column of correction data illustrated in Figure 13-6, Figure 13-7, and Figure 13-8. The sections of Figure 13-6, Figure 13-7, and Figure 13-8 that contain numbers retain data that is the same for all analyzers; the data is stored in memory at the factory and cannot be changed. TUNING. These miscellaneous frequency correction values correct for variations in instrument hardware performance. BW-AMP. The bandwidth amplitude corrections are mathematical offsets of the digitized video signal on the A16 processor/video assembly. Refer to the CAL AMPTD softkey description for more information. LC-XTAL. The A7 analog interface assembly uses these DAC values to adjust the bandwidths of the 21.4 MHz crystal and LC bandwidth lters on the A11 and A13 Bandwidth Filter assemblies. The CAL AMPTD self-calibration routine produces DAC values that adjust each bandwidth to within 20%. The 9 kHz and 120 kHz entries are for the 6 dB EMI resolution bandwidths. All other entries are for the 3 dB resolution bandwidths. Refer to the CAL AMPTD softkey description for more information. RFATN. The input attenuator corrections are mathematical offsets of the digitized video signal on the A16 processor/video assembly.
Chapter 13
585
The 10 dB attenuator step is the amplitude reference used by the CAL AMPTD self-calibration routine while calibrating the input attenuator. Refer to the CAL AMPTD softkey description for more information about the calibration of the input attenuators. Refer to To check control of the A3 input attenuator in Chapter 6 to identify the attenuators that are in use for a given input attenuation setting. SGAIN. The step-gain and linear-gain corrections are mathematical offsets of the digitized video signal on the A16 processor/video assembly. The rst six entries in the SGAIN column are for the 10 dB step gains on the A12 amplitude control assembly. The CAL AMPTD self-calibration routine uses the 10 dB step gain as an amplitude reference while calibrating the step gains. The last four entries in the SGAIN column are for the 10 dB linear gains on the A14 Log Amplier assembly. Refer to the CAL AMPTD softkey description for more information about the calibration of the step-gain and linear-gain stages. Refer to Chapter 5 to identify the gain stages that are in use for a given reference-level setting. GAIN. The reference-level-vernier error corrections are DAC values that calibrate the reference level at top-screen during the CAL AMPTD self-calibration routine. The rst ve entries are xed constants. The second ve entries provide a coarse gain-adjustment of the A9 Third Converter assembly. Only one DAC value is produced by the calibration routine and is duplicated for all ve entries. The A7 analog interface assembly uses the DAC value to adjust the gain of the A9 assembly. The DAC adjustment range is from 0 to 255. The higher the DAC value, the greater the gain on the A9 assembly. Refer to the CAL AMPTD softkey description for more information. ERR. The ERR column, next to GAIN, is the video offset ne tuning of the reference level calibration.
586
Chapter 13
CAL ATT ERR. These error corrections are mathematical offsets of the digitized video signal on the A16 processor/video assembly. The rst ve entries in the CA ATT ERR column are correction constants that correct amplitude errors for the calibration attenuators on the A12 amplitude control assembly. Currently, the entry for the 16 dB attenuator is not used. The calibration attenuators on the A12 amplitude control assembly are the amplitude reference for the CAL AMPTD self-calibration routine. The ve constants are stored in nonvolatile memory initially at the factory. The constants can be retrieved from, and restored to, nonvolatile memory using the procedures for backing up analyzer correction constants in Chapter 3 . When required, new constants are produced using the Cal Attenuator adjustment procedures in Chapter 2 . Refer to the SET ATTN softkey description for more information about the calibration attenuators on the A12 assembly. The second ve entries in the ERR column are one correction factor that provides a ne adjustment of the reference-level during the CAL AMPTD self-calibration routine. The calibration routine produces one video offset and stores it for all ve entries. The video offset is used to make a ne reference-level adjustment after a coarse reference-level adjustment is done using the DAC correction values from RL-VENR. RELATED SOFTKEYS CAL FREQ
CAL AMPTD DEFAULT CAL DATA CAL STORE CAL FETCH SET ATTN ERROR CAL TIMEBASE
Chapter 13
587
DROOP Droop
DESCRIPTION Droop disables the reset of the peak detector on the A16 processor/video assembly after each analog-to digital conversion. When an impulse signal is applied, the decrease, or droop, in the peak-detector output is visible. The peak detector will charge up to the peak value of the input signal and then its amplitude will decrease over time. Refer to Chapter 4 for additional troubleshooting information about the peak detector. RELATED SOFTKEYS CONF TEST
588
Chapter 13
To safeguard the current atness-correction constants, refer to Chapter 3 for information about backing up analyzer correction constants before using this function. The atness-correction constants can be viewed or modied using EDIT FLATNESS. To view the atness-correction constants, do not enter the passcode before proceeding to EDIT FLATNESS. To edit the atness data, enter the 2001 Hz passcode, press CAL and proceed to EDIT FLATNESS. 1. After pressing EDIT FLATNESS, the STORE FLATNESS softkey appears, and the atness data is ready for editing. 2. Use the A1A2 knob, the key, or the key to move along the frequency range of the analyzer. 3. Enter the amplitude offset for the desired frequency point and terminate the entry with dBm or +dBm.
CAUTION
Avoid pressing the INIT FLT softkey when pressing the STORE FLATNESS softkey. All the atness-correction constants will be lost if INIT FLT is pressed accidentally. 4. Store the atness constants by pressing STORE FLATNESS; the new correction constants are stored and an instrument preset is performed. Exit the routine at any time by pressing EXIT: no changes are made to the existing correction constants and an instrument preset is performed. Refer to the frequency response adjustment procedures in Chapter 2 for complete instructions related to the atness-correction constants. RELATED SOFTKEYS INIT FLT
STORE FLATNESS FLATNESS DATA EXIT
Chapter 13
589
EXIT Exit
DESCRIPTION Use EXIT to withdraw from a softkey function. No changes are made within the function and an instrument preset is performed.
590
Chapter 13
CAUTION
Avoid pressing the INIT FLT softkey when pressing the STORE FLATNESS softkey. All the atness-correction constants will be lost if INIT FLT is pressed accidentally. RELATED SOFTKEYS EDIT FLATNESS
STORE FLATNESS INIT FLT
Chapter 13
591
FM GAIN FM Gain
DESCRIPTION Note that this function is available for 8593E, 8594E, 8594Q, 8595E, and 8596E spectrum analyzers equipped with Option 102, AM/FM Speaker and TV Sync Trigger, only. This softkey duplicates the functions of the FM GAIN key that is accessed using AUX CTRL. FM GAIN adjusts the top-to-bottom screen deviation range of a demodulated FM signal, referenced to center-screen. The CAL FREQ self-calibration routine calibrates the FM screen deviation and modulation offset. Use FM GAIN to do a functional check of the demodulation circuitry. The CAL OUT signal can be used in place of a FM modulated signal source. 1. Connect the CAL OUT signal to the analyzer input and make the following instrument settings.
PRESET FREQUENCY.................................................300 M
Hz
CF STEP AUTO MAN(MAN).........................500 kHz SPAN................................................................. 0 Hz BW...................................................................5 MH
z
AMPLITUDE...................................................20 dB
When FM GAIN is rst enabled, it has a 100 kHz deviation from center-screen. 3. Set the maximum deviation from center screen by entering 500 kHz using the data keys, the A1A2 knob, or the and keys.
592
Chapter 13
If FM GAIN is functioning correctly, the displayed signal is deected from center screen to bottom screen as the center frequency is stepped up 500 kHz from the original center frequency setting. 5. Press the key twice. The displayed signal moves from bottom-screen to top-screen as the center frequency is stepped down 500 kHz from the original center frequency. RELATED SOFTKEYS DEMOD
DEMOD AM FM FM OFFSET
Chapter 13
593
FM OFFSET FM Offset
DESCRIPTION Note that this function is available for 8593E, 8594E, 8594Q, 8595E, and 8596E spectrum analyzers equipped with Option 102, AM/FM Speaker and TV Sync Trigger, only. Use FM OFFSET to adjust the horizontal trace for center-screen with no modulation on the carrier. This function is useful for adjusting the carrier offset when FM GAIN is set for a modulated signal with a small frequency deviation. The CAL FREQ self-calibration routine calibrates an initial center-screen offset. Use FM OFFSET to do a functional check of the demodulation circuitry. The CAL OUT signal can be used in place of an unmodulated carrier signal. 1. Connect the CAL OUT signal to the analyzer input and make the following instrument settings.
PRESET FREQUENCY..............................................300 MHz SPAN............................................................0 Hz BW................................................................5 MHz AMPLITUDE ...............................................20 dBm
594
Chapter 13
3. Adjust the position of the horizontal trace to center-screen using the A1A2 knob. 4. To calculate the actual frequency offset in kHz, multiply the displayed value by 300. RELATED SOFTKEYS DEMOD
DEMOD AM FM FM GAIN
FM SPAN FM Span
DESCRIPTION Displays the FM_SPAN signal from the Span Dividers on the A7 analog interface assembly. Perform the following steps to observe the FM-SPAN signal for FM spans (LO spans less than or equal to 10 MHz). 1. Activate SPAN before pressing CAL to select the FM SPAN function. 2. Press FM SPAN to display a negative-going ramp. 3. Adjust the span setting while observing the displayed ramp. The slope of the ramp increases as the span is increased. Due to quantization errors, the display appears at for LO spans less that 500 kHz. For the main-coil spans (LO spans greater than 10 MHz) the display is a at line. Refer to Chapter 6 additional LO troubleshooting information. RELATED SOFTKEYS MAIN SPAN
Chapter 13
595
596
Chapter 13
1 2 3 4 5 6 7 8 9
Calculated coarse-tune DAC (0-4095) Sampler harmonic number Calculated sampler IF in Hz Wide-band discriminator sensitivity factor Actual frequency error in Hz on last count Actual measured sampler IF in Hz Actual coarse tune DAC number (0-4095) Main coil sensitivity in number of bits per Hz Actual ne tune DAC number (0-4095) Spans >10 MHz, changes from sweep to sweep Spans 10 MHz, relatively stable from sweep to sweep
10 11 12
Actual extra-ne tune DAC number (0-4095) Indicates which DAC is being adjusted in the iteration process; coarse, ne, or extra-ne. Last iterations of the extra-ne tun DAC. Spans > 10 MHz, readout is in frequency Spans 10 MHz, readout is DAC number Extra-ne tune DAC iterations to get to correct frequency (not present for spans 10 MHz) Fine-tune DAC iterations to get to correct frequency Counter timebase Coarse-tune DAC iterations to get to correct frequency Drift calculation of marker count, a large difference between the two LO counts indicates drift (only present when MKR count is enabled) Discriminator sensitivity factor Sampling oscillator frequency in Hz
13 14 15 16 17
18 19
Refer to Chapter 6 for additional LO troubleshooting information. RELATED SOFTKEYS DISPLAY CAL DATA
Chapter 13
597
598
Chapter 13
additional information about analyzer recovery and initialization after a A16 Memory Loss. Selects the identity of the analyzer. There is one rmware set for the 8590 E-Series and L-Series spectrum analyzers, 8591C cable TV analyzers, and 8594Q QAM analyzers. By selecting the appropriate identication number (IDNUM), the rmware congures the analyzer to the proper frequency range and analyzer functions. The following table shows the relationship between the IDNUM and the analyzer model number.
IDNUM 5 6 8595E 8596E Analyzer
To safeguard the current atness-correction constants, refer to Chapter 3 for more information about backing up analyzer correction constants before using this function.
DESCRIPTION
Note that this function is available for 8590L, 8591C, 8591E, 8594E, 8594Q, 8595E, and 8596E spectrum analyzers only. Prepares the analyzer memory for the entry of new atness correction constants. The 2001 Hz passcode is required and must be entered before pressing CAL.
INIT FLT performs the following functions.
Sets up the start frequency, stop frequency, and step size of the frequency-response-correction points. Initializes A16 memory by entering a default value of 0 dB into memory for all frequency points.
CAUTION
The current atness-correction constants are lost when INIT FLT is pressed. RELATED SOFTKEYS EDIT FLATNESS
STORE FLATNESS
Chapter 13
599
Sets up the start frequency, stop frequency, and step size of the frequency-response-correction points up to 22 GHz. Initializes A16 memory by entering a default value of 0 dB into memory for all frequency points.
CAUTION
The current atness-correction constants are lost when INIT FLT, INIT FLT 22 GHz, or INIT FLT 26.5 GHz is pressed. RELATED SOFTKEYS EDIT FLATNESS
INIT FLT 26.5 GHz STORE FLATNESS
Sets up the start frequency, stop frequency, and step size of the frequency-response-correction points up to 26.5 GHz. Initializes A16 memory by entering a default value of 0 dB into memory for all frequency points.
CAUTION
The current atness-correction constants are lost when INIT FLT 22 GHz or INIT FLT 26.5 GHz is pressed.
600
Chapter 13
Chapter 13
601
602
Chapter 13
(20 dB). Note that the reference level offset is not changed. The marker readout and reference level readout must be divided by 10 to obtain the correct amplitude readout.
Chapter 13
603
604
Chapter 13
Chapter 13
605
Enter the new correction constants into A16 RAM with the following procedure. 1. Press the following keys.
PRESET FREQUENCY, 2001, Hz SET ATTN ERROR
2. The data-entry instruction for the 1 dB step, ENTER CAL ATTEN ERROR 1, is displayed in the active-function block, just below the active entry for the reference-level-offset value, REF LVL OFFSET. 3. Enter the characterization data for the 1 dB, 2 dB, 4 dB, and 8 dB steps, with a 0.01 dB resolution for each entry. 4. Terminate each entry with either the +dBm or dBm key. 5. Enter the original factory value, or a default value of 0 dBm, for the 16 dB step. After each entry the characterized value is displayed, with 0.1 dB resolution, to the left of the graticule, followed by the offset-reference-level indicator, OFFSET. If desired, use DISPLAY CAL DATA to review the new attenuator-characterization data with 0.01 dB resolution. Refer to DISPLAY CAL DATA for an explanation of the displayed calibration data. RELATED SOFTKEYS CAL AMPTD
DISPLAY CAL DATA
606
Chapter 13
Note that selecting INPUT Z (50) ensures that the analyzer will make amplitude calculations based on a 50 system. Select INPUT Z (75) for instruments equipped with 75 input. Changing INPUT Z does not affect the input impedance of the analyzer. To return the analyzer to the standard power-on settings, use the procedures in Analyzer Initialization in Chapter 3 of this service guide.
Chapter 13
607
Avoid pressing the INIT FLT softkey when pressing the STORE FLATNESS softkey. All the atness-correction constants will be lost if INIT FLT is pressed accidentally. RELATED SOFTKEYS EDIT FLATNESS
608
Chapter 13
Chapter 13
609
610
Chapter 13
Chapter 13
611
612
Chapter 13
14
Analyzer Messages
Refer to this chapter for information about hardware-error, user-error, and informational messages that are displayed when a problem with the operation of the analyzer occurs.
613
614
Chapter 14
Related Assemblies
A6 YTO, A9 third converter, A25 counter lock.
Troubleshooting Hints
If this message appears constantly, perform the CAL FREQ routine to try to eliminate this message.
Related Assemblies
A7 analog interface, A8 power supply, A15 motherboard, A16 processor/video assembly, assemblies installed in the card cage.
Troubleshooting Hints
The +15 V and -15 V supplies from the A8 power supply can cause this error message. If the power-supply LEDs for both power supplies are on, check that the supplies are within tolerance using the test points on the A16 processor/video assembly.A voltage greater than +10 Vdc on pins 1,3,5,7, or 31 of A16J1 can cause this error message. Refer to Figure 14-2 for the numbering order of the A16J1 connector pins that are accessible from the component side of the A16 assembly. If a voltage greater than +10 V is present on any one of the A16J1 pins indicated, use Table 14-2 to locate the source of the error message.
Chapter 14
615
Table 14-1
31
Figure 14-1
616
Chapter 14
Before removing the A16 processor/video assembly, check all assemblies that are identied in Table 14-1. 1. Turn the instrument power off and remove each assembly, one at a time. 2. Turn the instrument on. If the ADC ground fault remains, the instrument preset at power-up displays the error message. If the error message goes away, the last assembly removed is the cause of the problem. If the error message remains after the removal of all related assemblies, suspect the A16 processor/video assembly. An ADC ground failure can also be checked manually, using the service diagnostic. Press CAL, More 1 of 4, More 2 of 4, Service Diag, More 1, GND REF DETECTOR.
Chapter 14
617
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
If the instrument does not sweep, and all other instrument functions are normal, the probable cause of the error message is a defective A16 processor/video assembly.
Related Assemblies
A8 power supply, A15 motherboard, A16 processor/video assembly, assemblies installed in the card cage.
Troubleshooting Hints
The +15 V and 15 V supplies from the A8 power supply can cause this problem. If the power-supply LEDs for both supplies are on, check that the supplies are within tolerance using the test points on the A16 processor/video assembly. A voltage greater than +10 Vdc on pins 1, 3, 5, 7, or 31 of A16J1 can cause this error message. Refer to Figure 14-2 for the numbering order of the A16J1 connector pins that are accessible from the component side of the A16 assembly. If a voltage greater than +10 V is present on any one of the A16J1 pins indicated, use Figure 14-2 to locate the source of the error message.
618
Chapter 14
Table 14-2
A16J1 Pin Number 1 3 5 7
31
Figure 14-2
Chapter 14
619
Before removing the A16 processor/video assembly, check all assemblies that are identied in Table 14-2. 1. Turn the instrument power off and remove each assembly, one at a time. 2. Turn the instrument on. If the ADC 2 V fault remains, the instrument preset at power-up displays the error message. If the error message goes away, the last assembly removed is the cause of the problem. If the error message remains after the removal of all related assemblies, suspect the A16 processor/video assembly. Check the ADC 2 V reference voltage on the A16 processor/video assembly. Refer to 8590 Series Analyzers Component-Level Repair Service Guide for the location of the 2 V reference circuit in the analog-to-digital function block. A 2 V reference failure can also be checked manually, using the service diagnostics. Press CAL, More 1 of 4, More 2 of 4, Service Diag, More 1, 2v REF DETECTOR.
620
Chapter 14
CAL:_ _ _ (M)
During the self-calibration routine, messages may appear on the display to indicate how the calibration routines are progressing. For example, sweep, freq, span, MC delay, FM coil, and atten can appear on the analyzer display. Please note that it is normal for the LOCK OFF and REF UNLOCK error messages to appear briey during CAL FREQ.
Troubleshooting Hints
If the rmware detects a problem during the self-calibration routine, the routine will stop and display an error message (H). Refer to the appropriate error-message description in this section.
Troubleshooting Hints
Reset the preamplier gain to 0 dB using the following routines. 1. Press AMPLITUDE, More 1 of 3, EXTERNAL PREAMPG and set the preamplier gain to 0 dB. 2. Perform the CAL AMPTD routine. 3. Press CAL STORE to store the new calibration factors and the preamplier setting of 0 dB. Please note that pressing PRESET does not initialize the preamp-gain setting to 0 dB. If this message is still displayed after resetting the preamp gain, it is likely there is a failure in RAM on the A16 processor/video assembly.
Chapter 14
621
Related Assemblies
A16 processor/video assembly
Troubleshooting Hints
Perform the CAL FREQ and CAL AMPTD. For 8592L, 8593E, 8595E, and 8596E spectrum analyzers, also perform the CAL YTF routine. If the error message reappears after pressing CAL STORE, the CAL AMPTD routine was not successfully completed. Pressing CAL FETCH also clears the correction-data error ag and can permit erroneous data to be stored. Use CAL FETCH only for troubleshooting.
Related Assemblies
8590L and 8591E spectrum analyzers only: A6 YTO, A7 analog interface. All other 8590 E-Series and L-Series spectrum analyzers: A3A7 YTO, A7 analog interface.
Troubleshooting Hints
Perform the CAL FREQ routine again.
622
Chapter 14
Related Assemblies
8590L and 8591E spectrum analyzers only: A3 input attenuator, A4 rst converter, A5 second converter, A6 YTO, A7 analog interface, A9 third converter, A11 bandwidth lter, A12 amplitude control, A13 bandwidth lter, A14 log amplier. All other 8590 E-Series and L-Series spectrum analyzers: A3 front end, A7 analog interface, A9 third converter, A11 bandwidth lter, A12 amplitude control, A13 bandwidth lter, A14 log amplier.
Troubleshooting Hints
The CAL AMPTD routine checks the amplitude of the video signal as it adjusts the A7 reference-level-calibration DAC. Normally, the video-signal amplitude should move within tolerance as the A7 DAC changes the gain of the IF calibration amplier on the A9 third converter assembly. Figure 9-1, IF/Control Overall Block Diagram, and Figure 9-2, Figure 9-3, or Figure 9-4, RF Section Overall Block Diagram for more information. Check the signal amplitude along the signal path from the RF input, through the IF section, to the output of the A14 log amplier. Refer to Isolating an RF, LO, IF, or Video Problem in Chapter 4 for further troubleshooting information.
Related Assemblies
A9 third converter.
Troubleshooting Hints
Ensure that the CAL OUT connector is connected to the analyzer input, perform the CAL FREQ &TD routine, and then perform the CAL YTF routine again.
Chapter 14
623
Related Assemblies
8590L and 8591E spectrum analyzers only: A6 YTO, A7 analog interface. All other 8590 E-Series and L-Series spectrum analyzers: A3A7 YTO, A7 analog interface.
Troubleshooting Hints
If this message appears, press FREQUENCY, 37, Hz, CAL, More 1 of 4, More 2 of 4, DEFAULT CAL DATA, and perform the CAL FREQ routine again.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the CAL AMPTD.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the CAL AMPTD.
624
Chapter 14
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the CAL AMPTD.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the CAL AMPTD.
Related Assemblies
A130 narrow resolution bandwidth card.
Troubleshooting Hints
Perform the CAL AMPTD.
Troubleshooting Hints
Before performing any function that requires a service passcode, be sure you understand the consequences of the function. Improper use of functions that require a passcode can delete factory correction constants that are difcult to replace. Refer to the passcode information in Chapter 13 for information about the proper use of the passcodes.
Chapter 14
625
Related Assemblies
A7 analog interface, A11 bandwidth lter, A13 bandwidth lter, A14 log amplier/detector.
Troubleshooting Hints
The CAL AMPTD routine will stop at the resolution bandwidth that failed. If the A11/A13 bandwidth lter assemblies require realignment, use the crystal and LC bandwidth lter adjustment procedures in Chapter 2. If a defective bandwidth lter board is suspected, refer to IF Power-Level Measurement in Chapter 5.
Related Assemblies
A7 analog interface, A9 third converter.
Troubleshooting Hints
If one of the following conditions are present, the error message is displayed. The CAL OUT signal is missing. Ensure that the CAL OUT is connected to the input connector using the CAL OUT cable supplied with the analyzer. Refer to Table 1-4 for the part number of the CAL OUT cable. If the calibration cable is connected properly, manually check the CAL OUT signal. The CAL OUT signal is not within 50 MHz of 300 MHz, or it has an amplitude less than or equal to 45 dBm. The test limits are large enough to indicate an obvious failure for either the CAL OUT signal or the analyzer. The DEFAULT CAL DATA is in use and the frequency of the CAL OUT signal appears out of tolerance.
626
Chapter 14
The default data introduces frequency offsets that can make an accurate CAL OUT signal appear to be outside the test limits for the CAL OUT check. If desired, the instrument CAL OUT check can be bypassed. Enter the 37 Hz passcode before pressing CAL and the desired self-calibration softkey. Refer to the description of DEFAULT CAL DATA in Chapter 13 for additional information. The analyzer has extremely low gain. Refer to Chapter 4. Be sure to perform all internal self-calibration routines after resolving the calibration signal problem. Refer to the calibration guide for your instrument.
NOTE
Instrument rmware expects the cal output signal to be 300 MHz 2 MHz. Sometimes the instrument default data is not able to tune the cal signal within this range and a cal signal not found message may appear on screen. Perform a cal output bypass check by pressing Frequency, -37, Hz, Cal, Cal Freq. This will bypass the cal check and start by calibrating the sweep ramp.
Related Assemblies
8590L and 8591E spectrum analyzers only: A6 YTO, A7 analog interface, A9 third converter, A16 processor/video assembly. All other 8590 E-Series and L-Series spectrum analyzers: A3A7 YTO, A7 analog interface, A9 third converter, A16 processor/video assembly.
Troubleshooting Hints
Spurious signals can cause this routine to fail. Refer to the MAIN SPAN softkey description for further information in Chapter 13.
Chapter 14
627
Troubleshooting Hints
Interruption of the CAL AMPTD routine can cause the routine to terminate and produce this message. Refer to Chapter 13 for more information about DEFAULT CAL DATA.
Related Assemblies
8592L and 8593E model analyzers only: A3A2 microwave switch, A3A8 YIG-tuned lter, A7 analog interface assembly. All other 8590 E-Series spectrum analyzers and 8591C cable TV analyzers: (except 8594E and 8594Q) A3A8 switched YIG-tuned lter, A7 analog interface assembly.
Troubleshooting Hints
If this message appears, ensure that the CAL OUT connector (for the 8595E) or 100 MHz COMB OUT connector (for the 8592L, 8593E, or 8596E) is connected to the analyzer input, perform the CAL FREQ &TD routine, and then perform the CAL YTF routine again.
Related Assemblies
8590L and 8591E spectrum analyzers only: A6 YTO, A7 analog interface assembly. All other 8590 E-Series and L-Series spectrum analyzers: A3A7 YTO, A7 analog interface assembly.
Troubleshooting Hints
If this message appears, press FREQUENCY, 37, Hz, CAL, More 1 of 4, More 2 of 4, DEFAULT CAL DATA, and perform the CAL FREQ routine again.
628
Chapter 14
Cannot engage phase lock with current CAL FREQ data (U)
Indicates that the CAL FREQ routine needs to be performed before phase locking can be turned on.
Related Assemblies
A9 third converter assembly.
Troubleshooting Hints
If this message appears, press FREQUENCY, 37, Hz, CAL, More 1 of 4, More 2 of 4, DEFAULT CAL DATA, and perform the CAL FREQ routine again.
Related Assemblies
A9 third converter, A16 processor/video assembly.
Troubleshooting Hints
Access the detector by pressing TRACE, More 1 of 3, DETECTOR SMP PK. Toggle between SAMPLE and PEAK detector while looking for a difference in trace amplitude. If there appears to be a drastic difference in amplitude, the problem lies in the PEAK detector section of the A16 processor/video assembly.
Related Assemblies
A16 processor/video assembly, A105 gated video board assembly.
Troubleshooting Hints
Ensure there is a trigger pulse connected to the GATE TRIGGER INPUT connector on the rear panel of analyzer and that the GATE OUTPUT is connected the EXT TRIG INPUT connector.
Chapter 14
629
Related Assemblies
8590L and 8591E spectrum analyzers only: A3A8 YIG-tuned lter, A7 analog interface assembly. All other 8590 E-Series and L-Series spectrum analyzers: (except 8594D) A3A8 switched YIG-tuned lter, A7 analog interface assembly.
Troubleshooting Hints
Ensure that the 100 MHz COMB OUT connector (for an 8592L, 8593E, or 8596E) or the CAL OUT connector (for an 8595E) is connected to the analyzer input with a low-loss, short cable before the CAL YTF routine is performed.
Related Assemblies
A3A1 comb generator, A3A11 step-recovery diode, A3A12 3dB attenuator, J2 comb-generator output connector.
Troubleshooting Hints
Make sure that the 100 MHz COMB OUT is connected to the analyzer input using the comb generator cable (SMA to Type N) supplied with the analyzer. (A comb generator cable (SMA to SMA) is supplied with Option 026 instruments.) Perform the following steps to verify the comb-generator output manually. 1. Press the following keys.
PRESET FREQUENCY, 100, MHz SPAN, 500, MHz
630
Chapter 14
2. Look for the 100 MHz comb-generator signal to the right of the local oscillator signal. When COMB GEN ON OFF is OFF, this signal has an amplitude of approximately 8 dBm and the 100 MHz harmonics are less than 20 dBc. Press the following keys.
AUX CTRL COMB GEN ON OFF (On)
3. When the comb-generator output amplier is enabled, the amplitude of the 100 MHz signal and the 100 MHz harmonics increase. The 100 MHz signal amplitude increases to approximately +12 dBm. t If the signal amplitude does not increase, check that the comb-generator control line from the A7 analog interface assembly is switching correctly. Refer to Figure 9-2, Figure 9-3, or Figure 9-4, RF Section Overall Block Diagram for your analyzer, for the comb-generator control voltage values. t If the A7 control line is correct, suspect the A3A11 step-recovery diode or the A3A1 comb generator. Refer to Figure 9-2, Figure 9-3, or Figure 9-4, RF Section Overall Block Diagram for your analyzer, for more information.
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
Refer to the 8590 E-Series and L-Series Spectrum Analyzers, and 8591C Cable TV Analyzer Programmer's Guide for a description of legal commands.
Troubleshooting Hints
The condence test routine identies the analyzer function that failed the test. Refer to the CONF TEST description in Chapter 13 for more information.
Chapter 14
631
The four-digit failure code (DF40) is interpreted in Table 14-3. The three fail codes within the 10-digit section (0625C1F72D) are interpreted in the The 10-Digit Failure Code description.
632
Chapter 14
Table 14-3
Digits From Example D
Refer to the 10-digit failure code description for more information. LED DS12 remains on after pressing PRESET.
Chapter 14
633
Each failure code from the example is converted from hexadecimal to binary and interpreted in Table 14-4, Table 14-5, and Table 14-6. The I/O Address Bus Failure Code. The rst two digits of the 10-digit code identify the failed address lines on the I/O address bus. Convert the two digits to binary and interpret them with Table 14-4. Table 14-4 I/O Bus Address Failure-Code Interpretation
Digits From Example Equivalent Binary Value 8 4 2 0 1 8 6 4 2 1 A16 Circuit Tested Not Used Not Used Not Used Error at ADR 4 Error at ADR 3 Error at ADR 2 Error at ADR 1 Error at ADR 0 U18 U18 U18 U18 U18 A16 Reference Designator
634
Chapter 14
The I/O Data-Bus Failure Code. The next four digits identify failed data lines on the I/O data bus. Convert the four digits to binary and interpret them with Table 14-5. Table 14-5 I/O Data Bus Failure Code Interpretation
Digits From Example Equivalent Binary Value 8 4 2 2 1 8 5 4 2 C 1 8 4 2 1 8 4 2 1 1 A16 Circuit Tested Error at IOB 15 Error at IOB 14 Error at IOB 13 Error at IOB 12 Error at IOB 11 Error at IOB 10 Error at IOB 9 Error at IOB 8 Error at IOB 7 Error at IOB 6 Error at IOB 5 Error at IOB 4 Error at IOB 3 Error at IOB 2 Error at IOB 1 Error at IOB 0 A16 Reference Designator U3 U3 U3 U3 U3 U3 U3 U3 U2 U2 U2 U2 U2 U2 U2 U2
Chapter 14
635
The Video RAM Failure Code. The last four digits identify the failure address for the video RAM on the A16 assembly. A video RAM failure code indicates an internal failure on the A16 assembly; the section of the video RAM circuitry that is tested is not connected to other instrument assemblies. Convert the four digits to binary and interpret them with Table 14-6. Table 14-6 A16 Video Ram Address Failure Code Interpretation
Digits From Example Equivalent Binary Value A16 Circuit Tested A16 Reference Designator
8 4 2 1 8
Error at MAD 15 * Error at MAD 14 Error at MAD 13 Error at MAD 12 Error at MAD 11 Error at MAD 10 Error at MAD 9 Error at MAD 8 Error at MAD 7 Error at MAD 6 Error at MAD 5 Error at MAD 4 Error at MAD 3 Error at MAD 2 Error at MAD 1 Error at MAD 0
U306 U305 U306 U305 U306 U305 U306 U305 U306 U305 U306 U305 U306 U305 U306 U305
4 2 1 8 4
2 1
8 4 2
* The acronym MAD stands for Multiplexed Address and Data. There are no separate address and data lines within the A16 video RAM.
636
Chapter 14
Related Assemblies
A7 analog interface, A16 processor/video assembly, A17 memory card, A25 counter lock, assemblies installed in the card cage.
Troubleshooting Hints
Refer to Figure 9-1, IF/Control Overall Block Diagram, and Figure 9-2, Figure 9-3, or Figure 9-4, RF Section Overall Block Diagram for your analyzer, in Chapter 9 for an overview of the assemblies attached to the I/O bus. Refer also to Troubleshooting the A15 Motherboard Assembly in Chapter 5 for the location of specic I/O signal traces on related assemblies. If a failure code for an I/O bus address or data line occurs, remove the related assemblies one at a time until the failure message disappears. Be sure to turn the analyzer power off while removing and installing each assembly. t If the failure message disappears, suspect the assembly that was removed last. t If the failure message remains after all related assemblies have been removed, suspect the A16 assembly. If a A16 video RAM fail code occurs, the A16 assembly is defective.
Chapter 14
637
Related Assemblies
8590L and 8591E spectrum analyzers only: A6 YTO, A7 analog interface, A25 counter lock. All other 8590 E-Series and L-Series spectrum analyzers: A3A7 YTO, A7 analog interface, A25 counter lock.
Troubleshooting Hints
This error message may occur when using correction data from DEFAULT CAL DATA. Refer to the DEFAULT CAL DATA description in Chapter 13 for more information. If DEFAULT CAL DATA is in use, the instrument self- calibration routines must be run. Refer to the calibration guide for your instrument. The error message may occur if incomplete frequency calibration factors are used. Interrupting the CAL FREQ self-calibration routine produces incomplete frequency calibration data. Performing the CAL FREQ routine may eliminate the problem. Be sure to press CAL STORE to store the correction factors produced by CAL FREQ. The error message may occur if either the reference oscillator or the sampling oscillator on the A25 assembly is unlocked. Refer to Isolating an RF, LO, IF, or Video Problem in Chapter 4 for further information.
638
Chapter 14
Related Assemblies
A105 gated video board assembly.
Troubleshooting Hints
If your analyzer is equipped with an Option 105, ensure that nothing is connected to the GATE TRIGGER INPUT connector when the CAL AMPTD or CAL FREQ &TD routines are performed.
Chapter 14
639
640
Chapter 14
Related Assemblies
A16 processor/video assembly, A20 IB interface board assembly.
Troubleshooting Hints
Check that there is only one controller (the analyzer) connected to the printer or plotter.
Related Assemblies
A16 processor/video assembly, A20 IB interface board assembly.
Troubleshooting Hints
This is usually caused by trying to print or plot when a controller is on the interface bus with the analyzer. To use the analyzer print or plot functions, you must disconnect any other controllers on the IB. If you are using programming commands to print or plot, you can use an BASIC command instead of disconnecting the controller. Ensure the analyzer is congured properly for PRINT or PLOT, B/W or Paintjet, and the printer or plotter address is set correctly set. See the 8590 E-Series and L-Series Spectrum Analyzers, and 8591C Cable TV Analyzer Programmer's Guide for more information.
Chapter 14
641
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
The key name may have conicted with an analyzer programming command. To avoid this problem, use an underscore as the second character in the key name, or avoid beginning the key name with the following pairs of letters: LB, OA, OL, TA, TB, TR, MA, MF, TS, OT, and DR.
642
Chapter 14
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
Reset all instruments attached to the bus.
Related Assemblies
All internal assemblies.
Troubleshooting Hints
Remove instrument from bus. Perform an instrument preset.
Chapter 14
643
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
Use DISPOSE USER MEM and then load the user-dened item or downloadable program into analyzer memory. Press the following analyzer keys.
CONFIG, More 1 of 3 DISPOSE USER MEM
If the above keys are not accessible perform the following steps. 1. Press the following analyzer keys.
DISPLAY Change Title
Use the knob to select the letters DISPOSE ALL;, be sure to include the semicolon (;). 2. Press the following analyzer keys.
CAL, More 1 of 4, More 2 of 4 SERVICE CAL EXECUTE TITLE
If the analyzer still appears to be locked up, refer to Isolating an RF, LO, IF, or Video Problem in Chapter 4 .
644
Chapter 14
Related Assemblies
A7 analog interface board.
Troubleshooting Hints
Concentrate on the RF control sections of the A7 analog interface board.
Chapter 14
645
LO UNLVL (U)(H)
Indicates that the analyzer local oscillator distribution amplier is not functioning properly. This message may appear on the 8592L, 8593E, 8594E, 8595E, and 8596E with Options 009/010, Option 009, or Option 010.
Related Assemblies
A10 LO distribution amplier (LODA) control/tracking generator control.
Troubleshooting Hints
Check the LOsense and gate bias adjustments on A10. Bias voltages are marked on a label on the top of the high frequency RF. Marker Count Reduce SPAN (U) Indicates the resolution bandwidth to span ratio is too small to use the marker count function. Check the span and resolution bandwidth settings.
646
Chapter 14
Related Assemblies
A3A1 comb generator, A3A11 step-recovery diode, A3A12 3dB attenuator, J2 connector.
Troubleshooting Hints
Make sure that the 100 MHz COMB OUT is connected to the analyzer input using the comb generator cable (SMA to Type N) supplied with the analyzer. (A comb generator cable (SMA to SMA) is supplied with Option 026 instruments.) Perform the following steps to verify the comb-generator output manually. 1. Press the following keys.
PRESET FREQUENCY, 100, MHz SPAN, 500, MHz
2. Look for the 100 MHz comb-generator signal to the right of the local oscillator signal. When COMB GEN ON OFF is OFF, this signal has an amplitude of approximately 8 dBm and the 100 MHz harmonics are less than 20 dBc. 3. Press the following keys.
AUX CTRL COMB GEN ON OFF (On)
Chapter 14
647
4. When the comb-generator output amplier is enabled, the amplitude of the 100 MHz signal and the 100 MHz harmonics increase. The 100 MHz signal amplitude increases to approximately +12 dBm. t If the signal amplitude does not increase, check that the comb-generator control line from the A7 analog interface assembly is switching correctly. Refer to Foldout 11-2, 11-3, or 11-4, RF Section Overall Block Diagram for your analyzer, for more information about the comb-generator control voltage values. t If the A7 control line is correct, suspect the A3A11 step-recovery diode or the A3A1 comb generator. Refer to Foldout 11-2, 11-3, or 11-4, RF Section Overall Block Diagram for your analyzer, for more information.
Related Assemblies
A22 precision frequency reference.
Troubleshooting Hints
This is a timed message that comes on whenever the instrument is turned on; the temperature of the A22 precision frequency reference oven-controlled crystal oscillator (OCXO) is not measured. The instrument rmware displays the message only when it senses that the A22 assembly is connected to its power supply through W15, the OCXO power cable. If the message does not appear, check the W15 cable.
648
Chapter 14
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
This test performs a functional check of switching for the positive-peak detector on the A16 processor/video assembly. Refer to the CONF TEST description in Chapter 13 for more information about the condence test routine.
Related Assemblies
A22 precision frequency reference, standard 10 MHz reference, A25 counter lock.
Troubleshooting Hints
Refer to Foldout 11-2, 11-3, or 11-4, RF Section Overall Block Diagram for your analyzer, for more information. 1. Check the rear-panel cable connections for the 10 MHz reference. When using either the standard 10 MHz timebase or the precision frequency reference, make sure that the 10 MHz REF OUT on the rear panel is connected to the EXT REF IN. The W6 jumper is supplied with the analyzer for this purpose. When using an external 10 MHz reference, make sure it is connected to the EXT REF IN on the rear panel.
Chapter 14
649
2. If the rear panel connection is correct, check the power output of the 10 MHz frequency reference. The A22 standard 10 MHz timebase (RTXO) has an typical output of 0 dBm 6 dB. A22 precision frequency reference (OCXO), has a typical output of 0 dBm 3 dB. An external 10 MHz reference should have an output between 2 dBm and +10 dBm. 3. If the output of the 10 MHz reference is correct, check the continuity of W18, the EXT REF IN cable, and W17, the 10 MHz OUT cable. 4. Check the continuity of the VTO_TUNE control line. 5. Check the continuity of the W14 ribbon cable and the W21 coaxial cable. 6. Check the continuity of the A25 assembly connections to the A9 assembly that pass through the A15 motherboard. 7. If the error message is still present, suspect the A25 counter lock assembly.
650
Chapter 14
Related Assemblies A11 bandwidth lter, A13 bandwidth lter. Troubleshooting Hints Refer to the CONF TEST description in Chapter 13 for more information about the condence test routine.
A high noise oor can be caused by a defective bandwidth control line. Look for a change in noise oor level and bandwidth when manually switching between bandwidths. If no change occurs, check the bandwidth control lines from the A7 analog interface assembly. Refer to Foldout 11-1, IF/Control Overall Block Diagram, and bandwidth control line information in Chapter 5 for further information. A high noise oor can also be caused by loose screws in the IF assembly section. Make sure the IF assembly cover screws are properly tightened.
Chapter 14
651
2. Use DISPLAY CAL DATA to view the current bandwidth control factors. 3. Perform the CAL AMPTD self-calibration routine. Use DISPLAY CAL DATA to monitor changes in the bandwidth control factors each time the CAL AMPTD routine is performed. 4. Repeat the CONF TEST routine. If the error message is still present, look for an unstable bandwidth-lter assembly or bandwidth control line. Refer to bandwidth control line information in Chapter 5 when troubleshooting the bandwidth control lines. The bandpass shape of the failed resolution bandwidths may have drifted since the last time the self-calibration routines were run. 1. Look at the bandpass shape of each failed bandwidth. If realignment is needed, refer to the crystal and LC bandwidth lter adjustment procedure in Chapter 2 . 2. Run the CAL AMPTD self-calibration routine and store the data using CAL STORE. 3. Repeat the CONF TEST routine. If the error message is still present, one of the bandwidth-lter assemblies may be defective. Refer to the CONF TEST description in Chapter 13 for more information about the condence test routine.
Related Assemblies
A7 analog interface assembly, A3A8 (switched) YIG-tuned lter.
Troubleshooting Hints
Ensure that the signal source is connected to the RF input.
Related Assemblies
A7 analog interface assembly, A3A8 (switched) YIG-tuned lter.
Troubleshooting Hints
Ensure that the signal source is connected to the RF input.
652
Chapter 14
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
This test performs a functional check of switching for the sample detector on the A16 processor/video assembly. Refer to the CONF TEST description in Chapter 13 for more information about the condence test routine.
Chapter 14
653
Related Assemblies
8591E and 8590L spectrum analyzers and 8591C cable TV analyzers only: A6 YTO, A25 counter lock board assembly. All other 8590 E-Series and L-Series spectrum analyzers and 8591C cable TV analyzers: A3A7 YTO, A25 counter lock board assembly.
Troubleshooting Hints
Check that the external frequency reference is correctly connected to the EXT REF INPUT. Observe the frequency diagnostics. Look for numbers that are out of tolerance.
SRQ _ _ _ (M)
The specied service request is active. Service requests are a form of informational message and are explained in Appendix A of the 8590 E-Series and L-Series Spectrum Analyzers User's Guide.
Related Assemblies
A12 amplitude control.
Troubleshooting Hints
The condence test performs a functional check of the A12 assembly step gains; it does not check the function of the A12 step attenuators. The test passes if it detects any increase in noise oor level for each 10 dB change in reference level. There is no manual adjustment for the 10 dB step gains. The CAL AMPTD self-calibration routine corrects for the 10 dB step gains and input attenuator errors. Refer to the CAL AMPTD description in Chapter 13 for more information.
654
Chapter 14
If one of the 10 dB step gains appears to be faulty, check the step-gain control lines from the A7 analog interface assembly. Refer to Troubleshooting the A15 Motherboard Assembly in Chapter 5 for the location of the step gain control lines. If the control lines function correctly, suspect a A12 amplitude control assembly failure. Refer to the IF section gain control information in Chapter 5 for further information about the operation of the 10 dB step gains. Refer to the CONF TEST description in Chapter 13 for more information about the condence test routine.
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
Use DISPOSE USER MEM to clear analyzer memory. Perform instrument calibrations. If the message is still displayed, it may indicate a hardware failure.
Chapter 14
655
Description .
During the CAL TRK GEN self-calibration routine, a signal response above the rst division from bottom screen is not present. The primary purpose of the error message is to indicate that the CAL cable is not connected from the tracking generator RF OUT to the RF INPUT.
Related Assemblies
A7A1 tracking control, A24 FL1, A24A1 tripler, A24A2 tracking converter, A24A3 amplier detector, A24A4 output attenuator.
Troubleshooting Hints
If a low signal response is present within the rst division from bottom screen, check the cable connection from the RF OUT to the RF INPUT. A cable with excessive loss can cause a low-level signal. Use the CAL cable that is supplied with the instrument. Be sure to use the same CAL cable when performing all the self-calibration routines. Refer to Table 1-4 for the part number of the CAL cable. If no signal response is present, check the output of the tracking generator at the RF OUT connector. Refer to Figure 9-5, Tracking Generator Block Diagram, in Chapter 9 for more information.
TG UNLVL (U)(H)
This message can indicate the following: that the source power is set higher or lower than the analyzer can provide, that the frequency span extends beyond the specied frequency range of the tracking generator, or that the calibration data for the tracking generator is incorrect. See the 8590 E-Series and L-Series Spectrum Analyzers User's Guide for more information.
Related Assemblies
8590L and 8591E spectrum analyzers only: A7A1 tracking generator control assembly, A24 tracking generator control assembly.
Troubleshooting Hints
Perform the tracking generator control routine.
656
Chapter 14
USING DEFAULTS: _ _ _
The analyzer has determined that the contents of certain RAM locations are not within the expected values. The number following the colon corresponds to the location that caused the default condition. This number was used during the initial design and is not applicable to troubleshooting.
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
This message may indicate that the battery used to power the RAM is dead. Refer to restoring analyzer memory.
Chapter 14
657
Related Assemblies
A14 log amplier/detector, A16 processor/video assembly.
Troubleshooting Hints
The CONF TEST routine performs a functional check of switching for the video bandwidth circuitry on the A16 processor/video assembly. Refer to CONF TEST description in Chapter 13 for more information about the condence test routine. Problems with the low-pass lter on the A14 log amplier/detector assembly can cause the video bandwidths to appear to be incorrect.
Related Assemblies
A16 processor/video assembly, A105 gated video board assembly.
Troubleshooting Hints
Before using the time-gating functions, ensure there is a trigger pulse connected to the GATE TRIGGER INPUT connector on the rear panel of analyzer and that the GATE OUTPUT is connected the EXT TRIG INPUT connector. If the connections are correct and, after an instrument preset, the error message is still displayed, the internal triggering circuitry on the A16 processor/video assembly may not be functioning correctly. If you do not want to use the time-gating functions, press PRESET.
658
Chapter 14
15
659
These techniques for a static-safe workstation should not be used when working on circuitry that has a voltage potential greater than 500 volts. For more information about preventing ESD damage, contact the Electrical Overstress/Electrostatic Discharge (EOS/ESD) Association, Inc. The ESD standards developed by this agency are sanctioned by the American National Standards Institute (ANSI).
660
Chapter 15
Table 15-1
Static-Safe Accessories
Description Set includes: 3M static control mat 0.6 m 1.2 m (2 ft 4 ft) and 4.6 m (15 ft) ground wire. (The wrist-strap and wrist-strap cord are not included. They must be ordered separately.) Wrist-strap cord 1.5 m (5 ft) Wrist-strap, color black, stainless steel, without cord, has four adjustable links and a 7 mm post-type connection. ESD heel-strap (reusable 6 to 12 months).
Chapter 15
661
662
Chapter 15
Table 15-2
Item 1 2 3 4 5 6
Board puller, two prongs to lift PC boards Extender board, pin and socket, 20 contacts (two required) Extender board, pin and socket, 60 contacts Extender board, 22 pin edge connector, 44 contacts Injector board Crystal bandwidth shorts (set of three) Components Needed to Build One Crystal Short
6a 6b 6c 6d 6e 7 8 9 10 11 12 13 14 15
0.01 F capacitor 90.0 resistor Square, single-connector terminal (two required) Two-terminal connector body 3/16 inch dia. heat shrink tubing, 1 1/8 inches long Combination wrench, 5/16 inch, with slotted box end (two required) Combination wrench, 1/4 inch Open end wrench, 15/64 inch, open end Torque wrench, break-away, 10 inch-pounds, 5/16 inch, open end Cable puller, pry-bar style Alignment tool, metal tip, plastic body Alignment tool, nonmetallic tip, ber body Alignment tool TORX hand driver with required T8 and T10 bits, included in TORX driver kit with multiple bits (See tool-tip illustration)
0160-4832 0757-0400 1251-4182 1251-0689 0890-0029 08555-20097 8720-0014 8710-0946 40-60271 5021-6773 8710-0630 8710-0033 A,R R R R A,T,R A,T A,T A,T 8710-1426 A,R
The part number provided is a non-part number. This tool can be ordered from: Assembly Systems Inc. 16595 Englewood Avenue Los Gatos, California 95032 (408) 395-5313
If you order a similar tool from your local supplier, it is important that the outside dimension of the wrench be no wider than 0.518 inches. This allows the wrench to be used on semirigid cable connectors in conned areas.
Chapter 15
663
Figure 15-1
664
Chapter 15
Table 15-3
Hex (Allen) wrench, 3mm Hex (Allen) wrench, 4mm Hex (Allen) wrench, no. 4 Hex (Allen) wrench, no. 6 Nut driver, 7mm Nut driver, 3/8 inch Nut driver, 5/16 inch Nut driver, 7/16 inch Nut driver, 9/16 inch, drilled out, end covered with heatshrink tubing to protect front/rear-panel surface Phillips Screwdriver, small no. 0 Pozidriv screwdriver, small no. 1 Pozidriv screwdriver, large no. 2 Long-nose pliers Wire cutters Wire strippers * A = Adjustment, R = Replacement Procedure
Chapter 15
665
666
Chapter 15
16
667
Safety Considerations
Before servicing the analyzer, familiarize yourself with the safety markings on the analyzer and the safety instructions in this manual. This analyzer has been manufactured and tested according to international safety standards. To ensure safe operation of the instrument and the personal safety of the user and service personnel, the cautions and warnings in this manual must be heeded.
Safety Symbols
The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument.
WARNING
Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met. The caution sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
CAUTION
668
Chapter 16
Instrument Markings
The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
The CE mark is a registered trademark of the European Community. (If accompanied by a year, it is when the design was proven.) The CSA mark is a registered trademark of the Canadian Standards Association.
Chapter 16
669
Failure to ground the analyzer properly can result in personal injury, as well as analyzer damage. Before turning on the analyzer, connect a three-wire power cable with a standard IEC 320-C13 (CEE 22-V) inlet plug to the analyzer power receptacle. The power cable outlet plug must be inserted into a power-line outlet socket that has a protective earth-contact. DO NOT defeat the earth-grounding protection by using an extension cable, power cable, or autotransformer without a protective ground conductor. If you are using an autotransformer, make sure its common terminal is connected to the protective ground conductor of its power-source outlet socket.
WARNING
This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor, inside or outside the instrument, is likely to make the instrument dangerous. Intentional interruption is prohibited. The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from all voltage sources while it is being opened. These servicing instructions are for use by qualied personnel only. To avoid electrical shock, do not perform any service unless you are qualied to do so. There is a danger of explosion if the battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Discard used batteries according to manufacture's instructions.
WARNING
WARNING
WARNING
670
Chapter 16
WARNING
For continued protection against re hazard replace line fuse only with same type and rating (F 5A/250V). The use of other fuses or material is prohibited. The power cord is connected to internal capacitors that may remain live for 10 seconds after disconnecting the plug from its power supply. The detachable power cord is the instrument disconnect device. It disconnects the mains circuits from the mains supply before other parts of the instrument. The front panel switch is only a standby switch and is not a LINE switch (disconnect device). Before switching on this instrument, make sure that the line voltage selector switch is set to the voltage of the power supply and the correct fuse is installed.
WARNING
WARNING
CAUTION
Chapter 16
671
These techniques for a static-safe workstation should not be used when working on circuitry with a voltage potential greater than 500 volts. Example of a Static-Safe Workstation
Figure 16-1
672
Chapter 16
Always handle printed circuit board assemblies by the edges. This will reduce the possibility of ESD damage to components and prevent contamination of exposed plating.
Chapter 16
673
674
Chapter 16