Designed with safety in mind and in conformance with the international safety standard IEC 61010-2-020. Conformance to this standard will become increasingly important in the future.
It is designed so that no fragments will be ejected from the centrifuge even if the rotor is broken. The lid lock is interlocked, and a twin lock system of 2 right and left hooks is employed. The interlock is protected against shocks and employs an electrical double protection circuit.
Complete with various sensors for higher safety. Stops when an abnormality is detected.
Comes with automatic rotor identification function (patented). Rotor type is automatically identified and overspeed detection is enabled when a rotor is simply placed on the drive shaft. Comes with imbalance detection, abnormally high temperature detection, abnormality detection of spin sensor and temperature sensor, and a circuit breaker.
Speed
Time
Sample temperature is kept at 4C at the maximum speed of any rotor (at ambient temperature of 25C). Designed considering friendliness to the global environment and recycling.
Employs CFC-free HFC134a for the refrigerant. On the main unit, only a small amount of plastic is used to reduce the load on the environment.
A thermostatic refrigerated centrifuge (-20 to 60C) is available (optional). A safety cabinet with an anti-biohazard HEPA filter (Class I) can be set (optional).
RCF
Sealing cap
Code No. S17111 (optional)
ST-720
7
ST-410
7
ST-480
7
86 (W) x 128 (D) x 52 (H) 86 (W) x 130 (D) x 62 (H) 86 (W) x 130 (D) x 89 (H) 86 (W) x 128 (D) x 90 (H)
055-6370 055-6380
Angle Rotors
RA-288M3 RA-410M3 RA-500M3 AT-360 15.9~16.5 x 97~120 17 x 121 30 x 117 26.4~29.2 x 98~121 15~17.2 x 96~130 8,500 8,500 8,500 8,500 3,500 9,690 9,530 9,690 9,690 2,270 055-1280
1 Conical tubes 175 mL, 200 mL, and 225 mL can be centrifuged by using supporters. The 175 mL conical tube is made by Falcon (code No. 2076) and needs a cushion (code No. 2090). The 200 mL conical tube is made by Nalge Nunc (code No. 376813) and needs a cushion adapter (code No. 377585). The 225 mL conical tube is made by Falcon (code No. 2075) and needs a cushion (code No. 2090). 2 sealing cap cannot be used on this tube rack. sealing cap (code no. S17111) can be used on a 12 x 15 mL conical tubes with code No. S13130. 3 sealing adapter against biohazard for 50 mL conical tube (code no. 055-5760) and for 15 mL conical tube (code no. 055-5750) can be used. 4 15 mL glass tubes can be used if an adapter (code no. 055-7400) is added and need a rubber cushion (code no. 024-1400) sold separately. 5 15 mL conical tubes can be used if an adapter (code no. 055-1280) is added. 6 If micro plates are used, it is possible to stack them in double or more on each bucket. However, be sure to test the strength of those plates prior to centrifugation. The micro plate at the bottom may be crushed. 7 Contact your dealer in case you use two or more swinging bucket rotors and/or swinging bucket rotors for micro plates. 8 5/7mL narrow tube made by Sekisui (InsepackII), Termo (VenojectII), Nipro (Neotube) and Becton Dickinson (Vacutenar) can be used. Tube for flow cytometry can be used. 9 The CPT cell preparation tube is made by Becton Dickinson (Vacutenar 8mL for CPT cell separation).
Specifications
Model 5910
Maximum Speed Maximum RCF Maximum Capacity Control system Speed setting Speed indication RCF setting RCF indication Integrator setting and indication 8500 rpm 9690 xg 1020 mL Micro-processor control (Brushless motor) Speed, RCF, Time, Acceleration & Deceleration, 5 channel memories From 200 to 8500 rpm, 10 rpm increments Digital display, From 0 to 8900 rpm, 10 rpm increments From 10 to 9690 xg, 10 xg increments Digital display, From 0 to 10620 xg, 10 xg increments Digital display, From 1.00 to 9.99 x 10 9 g sec from 1 sec. to 100 sec., in 1 sec. increment setting from 100 sec. to 990 sec., in 10 sec increments setting from 1 min. to 100 min., in 1 min. increment setting from 100 min. to 990 min., in 10 min. increments setting Digital display, 1C increments setting and indication from 20C to 40C Single phase AC115V10%, 50/60Hz, 15A Single phase AC220V10%, 230V10%, 50/60Hz, 10A 115V, 12A 220/230V, 7.5A 115V, 910W 3.3MJ/h 220/230V, 1.2kW 4.3MJ/h 50(W) x 65(D) x 85(H) cm1, 110kg Manufactured and tested in accordance with IEC 61010-2-020
Temperature setting and indication Power requirements Rated voltage and Rated current Power consumption and Heat output Dimensions, weight Test standard
(1) Those of projections are not included. All specifications and external appearance subject to change without notice. * On Model 5910 only use KUBOTA made rotor, adapter and tube rack. Do not attempt use rotors, adapters or tube racks made by other manufacturers. * The rotor mentioned in the catalogue is usable on other KUBOTA centrifuge models. For details, refer to the catalogues relevant to each model. * Follow instructions given in the relevant Instruction & Service Manual for operating centrifuges.
Safety Precautions
For your safe use of this equipment, be sure to read the instruction manual carefully before you start its operation.
Parts may be adversely affected by heat, humidity and specimens or deteriorated by aging resulting in functional failures or may even lose their safety characteristic to cause accidents. To prevent occurrences of functional failures or accidents, turn off the power switch, pull out the power cable plug from the plug socket and be sure to contact your nearest Kubota Dealer, or an office of Kubota Corporation when there is no Kubota Dealer in your neighbourhood.
The rotor or bucket has been found to be damaged or corroded. Always discontinue A burnt smell comes out of the equipment. use when You receive a minor electrification when you touch the IMPORTANT phenomena listed equipment body by a naked hand. DISCONTINUE at right appear. Some other abnormality or failure has been found occurring. USE
http://www.centrifuge.jp
29-9 Hongo 3-chome, Bunkyo-ku, Tokyo 113-0033, Japan Telephone: (03) 3815-1331 Facsimile: (03) 3814-2574 E-mail: [email protected]
Note: This catalogue is not for distribution in the USA, Canada and Mexico as products shown are not for sale in these countries.
Copyright 2006 KUBOTA CORPORATION G1, Printed in Japan