Principles of CT and CT Technology : P CT CTT - Goldman 115
Principles of CT and CT Technology : P CT CTT - Goldman 115
Principles of CT and CT Technology : P CT CTT - Goldman 115
Lee W. Goldman
Department of Radiation Therapy and Medical Physics, Hartford Hospital, Hartford, Connecticut
14 2 10 5 14 for X3
and
FIGURE 4. ART. (A) ART algorithm for 4-voxel ‘‘patient.’’ (B)
Attenuation measurements. (C) Starting estimate is constructed 6 2 10 5 2 4 for X4 :
by dividing measurements from first view equally along their
ray paths. (D–F) This estimate is iteratively adjusted to match Next, these adjustments are divided equally along the
measurements for each consecutive view, stopping when rays, adding 2 to u1 and u3 (total adjustment of 14) and
transmission measurements predicted by current estimate
match all actual measurements to within some preset tolerance. subtracting 2 from u2 and u4 (total adjustment of 24),
yielding the second estimate (Fig. 4D). The process con-
tinues in this manner, with the actual values for each view
being compared in turn with the values predicted by the
of the voxel attenuation values along each ray. In this case, latest estimate. For the third view (measurement 5), the
there are 6 equations: second estimate predicted an attenuation value of 10 (7 1
3); the actual value was 12. Thus, the 12 total adjustment is
X1 5 u1 1 u3 ; divided equally along the ray for X59, subtracting 1 each
from u1 and u4, to generate the third estimate (Fig. 4E). The
X2 5 u2 1 u4 ; required adjustment for the fourth view (a total of 22) is
divided along the ray (subtracting 21 from u2 and u3) to
X3 5 u1 1 u2 ; generate the fourth and last estimate (Fig. 4F). Because the
attenuation values predicted by this last estimate match all
X4 5 u3 1 u4 ; actual measurements, the image is accepted as the true re-
constructed image (and one can see that it matches the pa-
X5 5 u1 1 u4 ; renthetical values in Fig. 4B).
ART images were typically available within 20 s of scan
and completion. In practice, however, ART was sensitive to
quantum noise and could result in poor image quality if
X6 5 u2 1 u3 : transmitted x-ray intensities were low. As in conventional
radiography, noise (graininess) in CT images arises from a
limited number of x-ray photons contributing to each mea-
In principle, these equations may be solved by use of si- surement (because a limited patient radiation dose is used).
multaneous equations to calculate the 4 unknowns (u1, u2, This noise, known as quantum mottle, appears as unavoidable
APPENDIX For the ray shown in Figure 3, wi is the voxel width W and
To understand how each measurement can be reduced to is equal for all voxels. However, in the more general case
a sum of the attenuation values in the voxels along the path for other angles, a ray may pass through a voxel at an angle