Alignment Face and Rim Method
Alignment Face and Rim Method
Alignment
Alignment procedures must be followed to ANGULAR ALIGNMENT
prevent unintended contact of rotating A unit is in angular alignment when indicator A
parts. Follow coupling manufacturer’s (Angular indicator) does not vary by more that .002 in.
coupling installation and operation (.05 mm) as measured at four points 90° apart.
procedures.
Vertical Correction (Top-to-Bottom)
SET UP 1. Zero indicator A at top dead center (12 o’clock) of
1. Mount two dial indicators on one of the coupling coupling half Y.
halves (X) so they contact the other coupling half
(Y) (Fig. III-1). 2. Rotate indicators to bottom dead center
(6 o’clock). Observe needle and record reading.
2. Check setting of indicators by rotating coupling
half X to ensure indicators stay in contact with 3. Negative Reading - The coupling halves are
coupling half Y but do not bottom out. Adjust further apart at the bottom than at the top. Correct
indicators accordingly. by either raising the driver feet at the shaft end
(add shims) or lowering the driver feet at the other
end (remove shims), (Fig. III-2).
Positive Reading - The coupling halves are
closer at the bottom than at the top. Correct by
either lowering the driver feet at the shaft end
(remove shims) or raising the driver feet at the
other end (add shims).
2. Take indicator measurements with driver feet 2. Rotate indicators through top dead center to the
hold-down bolts tightened. Loosen hold down right side, 180° from the start (3 o’clock). Observe
bolts prior to making alignment corrections. needle and record reading.
3. Take care not to damage indicators when moving 3. Negative Reading - The coupling halves are
driver during alignment corrections. further apart on the right side than the left. Correct
by either sliding the shaft end of the driver to the
left or the other end to the right.
Fig. III-4
Vertical Correction (Top-to-Bottom) 2. Rotate indicators through top dead center to the
right side, 180° from the start (3 o’clock). Observe
1. Zero indicators A and P at top dead center the needle, measure and record the reading.
(12 o’clock) of coupling half Y.
3. Make corrections as outlined previously.
2. Rotate indicator to bottom dead center (6 o’clock).
Observe the needles and record the readings. 4. Recheck both vertical and horizontal readings to
ensure adjustment of one did not disturb the
3. Make corrections as outlined previously. other. Correct as necessary.