Mp55 A Diferencijalni Presostat
Mp55 A Diferencijalni Presostat
Mp55 A Diferencijalni Presostat
PME industrial
measurement
electronics linked
to a field bus
MP55 module
A0563-5.6 en
PME-MP55
Contents
Page
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1 List of components and accessories supplied . . . . . . . . . . . . . . . .
1.2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7
7
7
12
13
4 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1 Functional overview of the MP55 . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Supply voltage and remote contact I/Os . . . . . . . . . . . . . . . . . . . . .
4.2.1 External supply voltage for the control I/Os . . . . . . . . . . . . . .
14
14
15
16
4.3 Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.4 CAN-interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.5 Synchronisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
17
19
20
21
21
24
25
26
30
38
38
38
39
41
44
53
55
57
9 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
60
10 Keyword index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
63
A0563-5.6 en
HBM
PME-MP55
Safety instructions
Use in accordance with the regulations
The MP55 module and its connected transducers are to be used exclusively
for measurement tasks and directly related control tasks. Use for any additional purpose shall be deemed to be not in accordance with the regulations.
In the interests of safety, the instrument should only be operated as described
in the User Manual. It is also essential to observe the appropriate legal and
safety regulations for the application concerned during use. The same applies
to the use of accessories.
The device must not be connected directly to the mains supply. The
maximum permissible supply voltage is 18...30 V.
General dangers of failing to follow the safety instructions
The MP55 module corresponds to the state of the art and is safe to operate.
The instrument can give rise to further dangers if it is inappropriately installed
and operated by untrained personnel.
Everyone involved with the installation, commissioning, maintenance or repair
of the instrument must have read and understood the User Manual and in particular the technical safety instructions.
Conditions on site
Protect the device from direct contact with water (IP20).
Maintenance and cleaning
The MP55 module is maintenance-free. Please note the following points when
cleaning the housing:
- Before cleaning, disconnect the devices from the power supply.
- Clean the housing with a soft, slightly damp (not wet!) cloth. You should
never use solvent, since this could damage the labelling on the front panel
and the display.
- When cleaning, ensure that no liquid gets into the device or connections.
HBM
A0563-5.5 en
PME-MP55
Residual dangers
The scope of supply and list of components provided with the MP55 cover
only part of the scope of measurement technology. In addition, equipment
planners, installers and operators should plan, implement and respond to the
safety engineering considerations of measurement technology in such a way
as to minimise residual dangers. Prevailing regulations must be complied with
at all times. There must be reference to the residual dangers connected with
measurement technology.
Any risk of residual dangers when working with the MP55 is pointed out in this
introduction by means of the following symbols:
Symbol:
WARNING
Meaning: Dangerous situation
Warns of a potentially dangerous situation in which failure to comply with
safety requirements can lead to death or serious physical injury.
CAUTION
Symbol:
Meaning: Possible dangerous situation
Warns of a possibly dangerous situation in which failure to comply with safety
requirements can cause damage to property or lead to some form of physical
injury.
Symbol:
NOTE
Means that important information about the product or its handling is being
given.
Symbol:
Meaning: CE mark
The CE mark enables the manufacturer to guarantee that the product complies with the requirements of the relevant EC directives (the declaration of
conformity is available at http://www.hbm.com/support/dokumentation).
A0563-5.5 en
HBM
PME-MP55
Working safely
Error messages should only be acknowledged if the cause of the error is removed and no further danger exists.
The instrument complies with the safety requirements of DIN EN 61010,
Part 1 (VDE 0411, Part 1).
To ensure adequate immunity from interference, use only Greenline shielded
ducting (place the shield of the transducer cable onto the connector housing).
The MP55 module must be operated with an extra-low safe voltage (supply
voltage 18 to 30 V DC).
Conversions and modifications
The MP55 module must not be modified from the design or safety engineering
point of view except with our express agreement. Any modification shall exclude all liability on our part for any damage resulting therefrom.
In particular, any repair or soldering work on motherboards is prohibited.
When exchanging any modules, only original HBM parts must be used.
Qualified personnel
This instrument is only to be installed and used by qualified personnel strictly
in accordance with the technical data and with the safety rules and regulations
which follow. It is also essential to comply with the appropriate legal and
safety regulations for the application concerned during use. The same applies
to the use of accessories.
Qualified personnel means persons entrusted with the installation, assembly,
commissioning and operation of the product who possess the appropriate
qualifications for their function.
Maintenance and repair work on an open device with the power on must only
be carried out by trained personnel who are aware of the dangers involved.
HBM
A0563-5.5 en
PME-MP55
Introduction
1.2 Introduction
The MP55 module from the PME product line is a carrier-frequency amplifier,
and is ideal for connecting the widest possible technological varieties of force,
pressure, torque and displacement transducers as well as load cells. The
MP55 module is set up, and its parameters are assigned, using either the keyboard and display or the PME Assistent. The PME Assistent provides an easy
operator interface under MS-Windows and this is used for assigning parameters to modules (as described in the PME Assistant online help).
Carrier-frequency
transducer excitation
A
DC
MP55
Fig. 1.1:
A0563-5.5 en
DC
24 V Supply voltage
Analogue output
Control I/Os
CANopen interface
HBM
PME-MP55
NOTE
The DIP switches must be set up/changed before the PME is fitted.
Various settings are defined with DIP switches and can be read out via the
display (see chapter 5.3). These are the settings for
bridge excitation voltage, effective range, bridge type, analogue output,
master/slave, terminating bus impedance, edge steepness
To set up the DIP switches proceed as shown in Fig. 2.1.
1
Unscrew the
cover
S10
S11
S12
Fig. 2.1:
HBM
S1/S2
S5
A0563-5.5 en
PME-MP55
Factory settings:
ON
S2
ON
Lower
board:
Upper
board:
S1 and S2
S1
1 2 3 4 5 6 1 2 3 45 6
ON
ON
S5
3
2
1
S12
1
2
3
4
5
6
ON
S11
ON
S10
1
2
3
4
5
6
1
2
3
4
5
6
ON
Analogue output
1 2 3
ON
S10
1 2 3456
Master/
Slave
ON
S11
12 34 56
Analogue
output
S12
1 23456
Bridge excitation
voltage
Input
range
A0563-5.5 en
ON
Bridge
type
ON
S2
ON
S1
1 2 34 5 6 123 4 56
Amplifier type
HBM
10
PME-MP55
Factory setting:
S2 S1
S11
S12
S5
Bridge type1)
ON
Input range*) 2)
ON
ON
S1
Full bridge
1 23456
S12
ON
1 23456
S12
S1
1 23456
1 23456
ON
S12
S1
1 23456
S1
1 23456
1 23456
ON
500 mV/V
Switch pos. III
S11
1 23456
ON
ON
LVDT
ON
50 mV/V
Switch pos. II
S11
1 23456
ON
ON
Half bridge
S1
3 mV/V
Switch pos. I
1 23456
ON
S1
1 23456
S11
1 23456
S11
1 23456
1 23456
ON
ON
2.5 V
ON
ON
S2
5V
Analogue output4)
S2
123 4 5 6
ON
"10 V
S11
1 23456
123
S11
"20 mA
S11
ON
1 23456
mA
ON
1234 56
Fig. 2.2:
1)
2)
3)
4)
*)
ON
S2
1V
4...20 mA
S11
1 23456
S5
ON
1 23456
ON
S11
S5
123
1 23456
Setting up an amplifier
Check this in the display under the TRANSDUCER group, parameter Trans.type; see page 25
Check this in the display under the TRANSDUCER group, parameter Input; see page 25
Check this in the display under the TRANSDUCER group, parameter Excitation; see page 25
Check this in the display under the ANALOG OUTPUT group, parameter ModeUa, see page 25
mV/V values by reference to 5 VUB (see table Tab. 1.1 on the following page)
HBM
A0563-5.5 en
11
PME-MP55
ON
S10
Master
Master/Slave
1 23456
S10
ON
Slave
S10
1 23456
Fig. 2.3:
Fig. 2.4:
OFF
Switch position II
50
500
2.5
100
1000
15
250
2500
Bridge type
Bridge excitation
voltage
Input range
Full bridge
5V
3 mV/V
Half bridge
2.5 V
100 mV/V
Half bridge
1V
15 mV/V
Half bridge
1V
250 mV/V
2.5 V
1000 mV/V
A0563-5.5 en
HBM
12
3
PME-MP55
Fig. 3.1:
Fig. 3.2:
Removal
CAUTION
The support rail must be on protection circuit potential
HBM
A0563-5.5 en
13
PME-MP55
1st
Colour coding
on pin 1
2nd
3rd
58mm
Recommended distance from the ribbon cable jack-connector
Fig. 4.1:
Several MP55 modules can be connected via one ribbon cable. This cable
serves as the local link for supply voltage and synchronisation between modules. No more than eight modules should be interconnected over one ribbon
cable.
A0563-5.5 en
HBM
14
4
PME-MP55
Connections
Warning
Comply with the safety instructions before putting the instrument into
service.
HBM
A0563-5.5 en
15
PME-MP55
Warning
24 V
Terminal plug 2
SYN
Labeling
Terminal plug 1
Terminal plug 3
Terminal plug 4
Labeling
Labeling
0V
AnalogueOUT
24 V
OUT1
IN3
"10 V
IN2
"20 mA
4...20 mA IN1
IN = digital input
IN4
OUT4
OUT2
OUT3
Fig. 4.2:
The 4 terminal plugs are coded so that they can be inserted in the 4 jacks
without any confusion. Jacks are fitted with coded lateral guides and terminal
plugs are fitted with coded pins.
A0563-5.5 en
HBM
16
PME-MP55
PLC
max. 0.5 A
Relay
max. 0.5 A
OUT1
24 V*
0 V*
24 V
0 V*
Terminal plug 4
Fig. 4.3:
Connection to a PLC
HBM
A0563-5.5 en
17
PME-MP55
4.3 Transducers
The following transducer types can be connected to the MP55 module:
Transducer
connection jack
15
5
15
6
13
12
Cable shielding
Hsg.
Measuring
signal (+)
Bridge excitation
voltage (-)
Bridge excitation
voltage (+)
2
1
Wiring colours:
Fig. 4.4:
A0563-5.5 en
Bridge excitation
voltage (+)
Sensor circuit (+)
13
12
ye
Cable shielding
Hsg.
LVDT transducers
8
5
Measuring signal
(+)
Bridge excitation
voltage (-)
13
Bridge excitation
voltage (+)
Measuring signal
(-)
Cable shielding
Sensor circuit (+)
12
Cable shielding
bu
gy
Potentiometric transducers
Bridge excitation
voltage (-)
gn
bk
Hsg.
15
Hsg.
13
12
wh= white; bk= black; bu= blue; re= red; ye= yellow; gn= green;
gy= grey
HBM
18
PME-MP55
8
5
Four-wire connection:
Half bridge
wh
15
6
13
15
bu
Hsg.
gn
13
gy
12
ye
bk
ye
12
Hsg.
Fig. 4.5:
wh= white; bk= black; bu= blue; re= red; ye= yellow; gn= green; gy= grey
NOTE
Use standard HBM cable for the transducer connection. When using
other shielded, low-capacitance measuring cable, connect the transducer cable shielding to the connector housing in accordance with the
HBM Greenline concept (publication S1578). This ensures EMC protection.
1)
For cable lengths in excess of 50 m, one resistor with half the value of the bridge resistance (RB/2) must
be switched on at the transducer in place of each of the feedback bridges. If the transducers are calibrated in six-core circuit, the resistors must be switched on directly in the sensor circuit.
HBM
A0563-5.5 en
19
PME-MP55
4.4 CAN-interface
The CAN-bus is connected via terminal plug 1. A maximum of 32 CAN users
can be connected to a bus segment (in accordance with the CANopen specification). The CAN-bus needs a terminating impedance of 120 in the first and
last bus users. The bus line can have a maximum of two terminating resistors.
The MP55 module has a built-in terminating resistor which is enabled by
toggle switch S14 (see page 11).
Low
High
1st device
Fig. 4.6:
CAN connection
in accordance
with Fig. 4.6
First device in
the bus line
connect the terminating
resistor here (toggle
Do not connect the terminaswitch)
ting resistor
Fig. 4.7:
Last device in
the bus line
connect the terminating
resistor here (toggle
switch)
NOTE
If the first or last device in the bus line is not a PME module, one 120 W
resistor must be connected for each of these outside devices.
A0563-5.5 en
HBM
20
PME-MP55
4.5 Synchronisation
Synchronisation is advisable when
D the transducer cables are connected in parallel to several devices
D unshielded channels lie close together
Synchronisation prevents carrier-frequency differences giving rise to beat interference.
S10
S10
ON
ON
S10
Slave
Master
1 2 3 4 56
Fig. 4.8:
1 23 4 5 6
Setting up master/slave
HBM
A0563-5.5 en
21
PME-MP55
-18.0024
kg
Status field
Value
unit
The keys
are pressure-sensitive:
Hold key down - the values scroll (the harder you press, the faster they scroll)
Press key briefly - go to next value
Function of the
buttons:
SET
A0563-5.5 en
Select
parameter/group
+
-
HBM
22
PME-MP55
Measuring mode
Password status
inactive (factory setting):
Password status
enabled
SET 2 sec
SET 2 sec
PASSWORD
DIALOGUE
Set-up mode
Input password:
SET
Password
SET
Input mode
Confirm
SET
Example:
Change
settings:
+
SET
Select
group
Input numerical
value
Select table
value
TRANSDUCER
CONDITIONING
Select first
parameter
unit
Nom. mV/V
0.000
SET
+
SET
Back to
measuring
mode:
kN
Filter
_ _ Hz
2.000
0.5 Hz
Confirm
SET
saves
directly
HBM
Go to input mode
>0<
YES
NO
SET
+/-
SET
Save ?
No
A0563-5.5 en
23
PME-MP55
During measurement you can press
play:
1. Display mode
2. The status of input and output
3. Error types (ERROR)
Status field
-18.0024
kg
and
Value
unit
Display mode
No character
Gross signal
>T<
Net signal
Maximum peak value signal
Minimum peak value signal
Peak/peak signal
+
-
mV/V
Input signal
V or mA
Outp
Inpt
not set
e.g. StoreMax
Error messages
Status field
Error occurred
Standstill status occurred
Shunt resistor on
A0563-5.5 en
HBM
24
PME-MP55
5.2 Commissioning
D Set up the DIP switches in accordance with chapter 2 (pages 10 and 11).
Example:
Transducer type and rated data
Bridge type
Bridge excitation
voltage
Input range
Full bridge
5V
3 mV/V
Half bridge
2.5 V
100 mV/V
Half bridge
1V
15 mV/V
Half bridge
1V
250 mV/V
Half bridge
2.5 V
1000 mV/V
D Connect the power supply cable and the transducer to the module, as
described in chapters 4.2 and 4.3.
CAUTION
Be sure to follow the safety instructions!
D Switch on the power supply.
The instrument carries out a function test (approx. 15 sec) and if
functioning correctly, switches to measuring mode. During the function
test, the remote contacts stay at 0 V.
NOTE
If the error message HardwOvf is displayed at this point, please refer to
chapter 8 Error messages for more details.
A green LED also tells you whether the MP55 is ready to begin measuring.
If the LED shows yellow or red, please refer to chapter 8 Error messages for
more details.
NOTE
When connecting transducers in parallel, please take into account the
resulting total resistance. If required, reduce the excitation voltage.
HBM
A0563-5.5 en
SET
SET
25
Groups
DIALOG
DATA SET
DISPLAY
TRANSDUCER
TRANSD.CALIBRAT
CONDITIONING
ANALOGOUTPUT
LIMIT VAL.
1...4
PEAK STORE
IN/OUT
CAN-BUS
ADDITION FUNCTION
Password
Recall ?
DecPoint
Unit
P1Meas.?
>0< kN1)
SourceUa
Operatn.
Operatn.
Output1
Baudrate
AmplType
PassStat
Save ?
Step
Transd.Type
P1 mV/V
>0<set ?
Mode UA
Source
InputMin
ModeOut1
Address
PrgVers
>0<save
kN1)
SwtchDir
InputMax
Output2
Profil
>0<Rf kN1)
ModeOut2
Output
MotionDsp
Output3
OutR. ms
MTime ms
ModeOut3
PDO-Frmt
MAmp kN1)
MAINGR
HW Synchr
Language
Up
I.DataS
InputRng
P2Meas.?
>T< kN1)
Zero V
Value kN1)
I.Displ.
ZeromV/V
P2 mV/V
>T<set?
EndV kN1)
Hyst kN1)
Down
I.Transd
Zero kN1)
P2 kN1)
>T<save
EndV V
On Del ms
I.Calibr
Nom.mV/V
MAINGR
filtre
MAINGR
Off Del ms
Output4
I.Condit
NVal kN1)
FiltChar
MAINGR
ModeOut4
Keyboard
I.Analog
Zero.Adj
MAINGR
Zeroing
I.LimVal
NomV.Adj
Tare
HW-Vers.
I.PStore
Shunt
PkMomMax
MAINGR
I.I/O
ShuntPol
PkHldMax
I.CAN
MAINGR
PkMomMin
Overview of parameters
MAINGR
MAINGR
Excitatn
P1
kN1)
Zero
ClearPkV
kN/s1)
MAINGR
I.AddFnc
PkHldMin
MAINGR
ParaCo1
ParaCo2
InpFunc
MAINGR
SET
back to group
A0563-5.6 en HBM
2sec
Meas. value
PassStat
active
SET
Yes
PASSWORD
Groups
No
DIALOG
SET
____
+/-
Password
DATA SET
SET
SET
Password
A0563-5.6 en HBM
SET
SET
____
+/-
Recall ?
PassStat
SET
Language
SET
active
inactiv
Deutsch
English
Francais
SET
DataS1
DataS2
DecPoint
+/-
DataS3
DataS4
FactSet
No
I.Displ.
Save ?
SET
back
MAINGRP
SET
DataS1
DataS2
DataS3
DataS4
No
SET
back
MAINGRP
SET
I.Calibr
I.Condit
SET
I.Analog
free
locked
+
-
I.I/O
Confirm input:
+/- =
+/-
or
SET
press
SET
2sec
26
back
MAINGRP
0005
0010
0020
0050
0100
0200
0500
1000
I.AddFnc
0001
0002
Select parameter
I.PStore
I.CAN
+/-
Parameter values
+/-
I.LimVal
SET
.000
.00
.0
.
Step
+/-
I.Transd
-
.00000
.0000
+/-
SET
+/-
I.DataS
SET
Define
password
Input
password
Continue
with DIALOG
Display
Transducer
TRANSDUCER
+
-
TRANSD.-CALIBRAT
SET
SET
BrdgeTyp
FullBrdg
+/-
fixed
w.switches
SET
fixed
w.switches
SET
SET
fixed
w.switches
0.0000
+/-
Zero kN1)
Nom.mV/V
NVal
kN1)
Shunt
SET
SET
SET
SET
0.00
2.0000
10.00
Off
On
+/-
+/-
+/-
kg
T
kT
TON
lb
oz
N
kN
Bar
mBar
Pa
Pas
hPas
kPas
psi
m
mm
cm
m
inch
Nm
kNm
FTLB
INLB
m/m
M/S
M/SS
p/0
p/00
ppm
S
MP
MN
g
SET
nein
no
ja
yes
P1 mV/V
SET
SET
SET
+Unbala
-Unbala
>0<
0.0000
kN1)
SET
no
yes
SET
kN1)
>0<save
+/-
0.00
SET
0.0000
>T<set?
SET
no
yes
SET
0.00
StoreMax
StoreMin
Store PP
CAN
+/-
+/-
+/-
Mode Ua
10V
SET
fixed
w.switches
Zero kN1)
SET
0.00
+/-
SET
0.00
+/-
SET
10.00
+/-
SET
10.00
+/-
>T<
P2Meas.?
SET
SET
>T<save
kN1)
SET
filtre
save
volatil
SET
+/-
Zero V
SET
EndV kN1)
+/-
0.0000
FiltChar
+/-
SET
Bessel
Butterw
EndV V
0.05 Hz
SET
back
MAINGRP
+/-
SET
0.0000
+/-
P2
+/-
+/-
P2 mV/V
kN1)
nein
no
ja
yes
1)
GrosValu
NetValue
SET
+/-
save
volatil
SET
SourceUa
+/-
+/-
+/-
back
MAINGRP
SET
>0<set ?
+/-
ShuntPol
LimVal.1
ANALOG OUTPUT
SET
P1Meas.?
P1
CONDITIONING
SET
Unit
27
Groups
Display
back
MAINGRP
+/-
0.1 Hz
0.2 Hz
0.5 Hz
1 Hz
2 Hz
5 Hz
10 Hz
20 Hz
50 Hz
100 Hz
200 Hz
500 Hz
SET
back
MAINGRP
A0563-5.6 en HBM
Groups
Analogue output
LIMIT VAL.1
LIMIT VAL.2...4
A0563-5.6 en HBM
+
PEAK STORE
IN/OUT
SET
SET
Operatn.
Source
SET
SET
On
Off
+/-
GrosValu
NetValue
StoreMax
StoreMin
Store PP
SwtchDir
SET
Higher
Lower
SET
On
Off
InputMin
SET
GrosValu
NetValue
InputMax
SET
+/-
+/-
LimVal1
LimVal2
SET
Operatn.
+/-
As
appropriate
for 1...4
+/-
GrosValu
NetValue
CAN-Bus
Output1
SET
ModeOut1
SET
Zeroing
SET
LimVal3
LimVal4
Motion
Error
inactiv
Act. On
Act.Off
+/-
+/-
+/-
Tare
Value kN1)
SET
0.00
+/-
ClearPkV
-
Hyst
kN1)
SET
0.00
Off Delms
SET
SET
0000.0
0000.0
No
Yes
+
+/-
PkMomMax
+/-
Input 1
Input 2
kN/s
On Del ms
SET
SET
0.00000
PkHldMax
+/-
+/-
+/-
Input 3
Input 4
NoInput
PkMomMin
SET
+/-
back
MAINGRP
PkHldMin
ParaCo1
SET
back
MAINGRP
ParaCo2
InpFunc
1)
SET
back
MAINGRP
free
locked
+/-
28
SET
29
Groups
In/Out
CAN-Bus
SET
Baudrate
10 kB
20 kB
SET
+/-
50 kB
125 kB
250 kB
500 kB
800 kB
1000 kB
Address
SET
SET
no Prof
000
ADDITIONFUNCTION
MEAS MODE
SET
SET
AmplType
Save ?
SET
Yes
No
+/-
PrgVers
>0< Rf kN1)
+/-
SET
0.00000
+/-
Meas. value
Profil
Output
+/-
StoreMax
StoreMin
Store PP
GrosDsp
NetDsp
SET
MTime ms
SET
GrosValu
NetValue
SET
MotionDsp
MAmp kN1)
+/-
HW Synchr
Slave
OutpR ms
PDO-Frmt
SET
SET
000
float
integer
SET
SET
SET
+/-
Keyboard
+/-
SET
Off
On
+/-
0000.0
0000.0
+/-
+/-
fixed
w.switches
slow
fast
medium
Preset with
DIP switch S10
+/-
SNo
HW-Vers.
back
MAINGRP
1)
SET
back
MAINGRP
A0563-5.6 en HBM
30
6
PME-MP55
Parameter
Meaning
DIALOG
Password
PassStat
I.DataS to
I.AddFnc
Recall ?
You can load either the factory settings or one of the four
parameter sets that have been stored.
Save ?
DATA SET
TRANSDUCER
Zero mV/V
Nom.mV/V (
1)
mV/V
2 mV/V)
HBM
A0563-5.6 en
31
PME-MP55
Group
Parameter
TRANSDUCER
Meaning
Information on scaling
Input characteristics:
The range of values for scale factors is limited. Scaling is
dependent on the chosen resolution. In the case of set-ups
that lead to overshooting the respective limits, Scaling error
is reported (see page 58).
Maximum display resolution: 999 999 digits at 6.67 % of the input range
Minimum display resolution: 10 Digits at 100 % of the
input range
TRANSD.CALIBRAT
Shunt
ShuntPol
P1Meas.?
P1 mV/V
P1 kN1)
p2
P1Meas.? Yes
p1
0,0457 mV/V
P1
Enter 0 kg
(physic. Unit is assigned)
mV/V
P2Meas.? Yes
P2
7,873 mV/V
Enter 4 kg
CONDITIONING
Container
Steps in the
weighing procedure
1)
Action
Display
Gross
Net
Attach platform
(35 kg)
> 0<
Input 35 kg
before 35 kg
after 0 kg
before 35 kg
after 0 kg
Attach container
(8 kg)
> T<
Input 8 kg
before 8 kg
after 8 kg
before 8 kg
after 0 kg
A0563-5.6 en
HBM
32
PME-MP55
Group
Parameter
Meaning
CONDITIONING
>0<kN1)
Enter zero value. Zeroing effects both, the gross value and
the net value.
>0< set ?
>0< save
>T< kN1)
>T< set ?
>T<save
Filter
FiltChar
0.05 Hz
0.1 Hz
0.2 Hz
0.5 Hz
1 Hz
2 Hz
5 Hz
10 Hz
20 Hz
50 Hz
100 Hz
200 Hz
500 Hz
Step response
The diagram shows a linear amplitude response which falls away steeply above the
cut-off frequency. There is an overshoot of
approx. 10 %.
Time
HBM
A0563-5.6 en
33
PME-MP55
Group
Parameter
Meaning
ANALOGOUTPUT
SourceUa
The gross or net value and the peak value can be chosen as
the source for the analogue signal.
Mode Ua
Zero kN1)
Zero V
EndV kN1)
EndV V
EndV V
Zero V
0
Zero
kN
EndV
kN
Physical unit
Information on scaling
Output characteristics:
The scale factor for the analogue output is derived from the
input and output characteristics. If the span that has been set
up corresponds to the measurement range in mV/V, the
minimum output voltage that can be set up is 0.17 V. If the
settings cause the respective limits to be exceeded,
Analogue scaling error is reported (see page 58).
Scale range min. for analogue output: 0.17 V at 100 % of amplifier input
range
Scale range max. for analogue output: 10 V at 3.67 % of amplifier input
range
1)
A0563-5.6 en
HBM
34
PME-MP55
Group
Parameter
Meaning
LIMIT VAL.
1...4
Source
You can choose from the following as the source for the limit
value signal: Gross, Net, Peak value Max/Min / Peak-to-peak
SwtchDir
Value
Hyst
Value
Over limit
Below limit
Hyst
OFF
OFF
Hyst
Value
Limit2, ON
24 V
0V
Limit1 ON
24 V
0V
Limit2 ON
PEAK
STORE*)
On Del ms
Off Del
ms
InputMin/Max
ClearPkV
kG/s
HBM
A0563-5.6 en
35
PME-MP55
Inputs/Outputs (I/Os)
Terminal plug 3: This is equipped with 4 inputs which you can use to control
the functions of the PME.
Terminal plug 4: This has 4 outputs available.
Group
Parameter
Meaning
IN/OUT
Output1...4
Mode
Out1...4
Function
Input value 0 V
Input value 24 V
Taring
Zero
balance
PkMomMax
PkMomMin
PkHldMax
PkHldMin
ParaCo1
ParaCo2
A0563-5.6 en
Parameter set
ParaCo2
ParaCo1
HBM
36
Group
Parameter
Meaning
Measuring signal
Trend of stored
value
Amplitude
IN/OUT
PME-MP55
Function
Operating
mode
Run
Hold
Peak value (Store1)
t
Run
Hold
Current value
Amplitude
Measuring signal
t
Function
Operating
mode
CAN-Bus
HBM
Run
Hold
Current value
Run
Run
Baudrate
10 kB, 20 kB, 50 kB, 125 kB, 250 kB, 500 kB, 800 kB, 1000 kB
Address
Profile
Output
You may choose which signal is output over the CAN bus:
Gross, Net or Peak value Max/Min.
OutR. ms
A0563-5.6 en
37
PME-MP55
Group
Parameter
Additional
function
>0< Rf
Meaning
Zero reference
Example:
"20 mm
Engine seating
MotionDsp
MTime ms
MAmp kg
Standstill time;
Standstill is reported if Amplitude MAmp is not exceeded in
standstill time t.
Signal
MAmp
MTi
me
24 V
Warning
A0563-5.6 en
0V
Time
(Standstill time)
Standstill
HBM
38
7
PME-MP55
7.1 Introduction
The MP55 module has a built-in CAN-interface which may be used not only
for data transmission but also for assigning parameters to the module. The
baud rate is selectable; the maximum possible is 1 MBaud. The interface protocol is based on the CANopen standard.
Receive PDO:
CAN identifier
1st data byte
HBM
0
1
Module address (0 = all)
A0563-5.6 en
39
PME-MP55
To exit from Operational status use the message Enter_Pre_Operational_State:
CAN identifier
1st data byte
2nd data byte
0
128
Module address (0 = all)
Reading a parameter:
Send value (PC or PLC to MP55)
CAN identifier
1st data byte
2nd + 3rd data byte
4th data byte
5th-8th data byte
A0563-5.6 en
HBM
40
PME-MP55
HBM
A0563-5.6 en
41
PME-MP55
Subindex
Name
Data type
Attr.
1000
1001
0
0
Device type
Error register
Unsigned32
Unsigned8
ro
ro
1003
1003
0
1..7
Unsigned8
Unsigned32
rw
ro
1005
1008
0
0
Unsigned32
Vis-String
rw
ro
1009
Vis-String
ro
100A
Vis-String
ro
100B
100C
100D
100E
100F
1010
0
0
0
0
0
0..2
Unsigned32
Unsigned16
Unsigned8
Unsigned32
Unsigned32
Unsigned32
ro
rw
rw
rw
ro
rw
65766173Hex
1011
0..2
Unsigned32
rw
64616F6CHex
1012
1014
0..2
0
Unsigned32
Unsigned32
rw
rw
1200
0..2
ro
1400
0..2
1401
0..2
SDOParame
ter
1st Receive PDO parameter PDOComm
Par
2nd Receive PDO parameter PDOComm
Par
A0563-5.6 en
Values
rw
rw
HBM
42
PME-MP55
Index
(hex)
Subindex
Name
Data type
Attr.
1600
1601
1800
1801
1A00
1A01
0..2
0..2
0..2
0..2
0..2
0..2
PDOMapping
PDOMapping
PDOCommPar
PDOCommPar
PDOMapping
PDOMapping
rw
rw
rw
rw
rw
rw
Values
Data structures:
PDO CommPar:
Index
Subindex
0020
0
1
2
3
4
Name
Number of entries
CAN identifier for PDO
Transmission type
Off-time
Priority group
Data type
unsigned 8
unsigned32
unsigned8
unsigned16
unsigned8
Value
31 (MSB)
0
1
0
1
0
1
X
30
29
28..0
Meaning
PDO valid
PDO invalid
RTR allowed
RTR not allowed
11 bit ID
29 bit ID
CAN-ID
PDO mapping:
Index
Subindex
0021
0
1
2
...
Name
Data type
unsigned8
unsigned32
unsigned32
unsigned32
HBM
A0563-5.6 en
43
PME-MP55
SDO parameter:
Index
Subindex
0022
0
1
2
3
Name
Number of entries
COB-ID client->server
COB-ID server->client
Node ID (optional)
Data type
unsigned8
unsigned32
unsigned32
unsigned8
0
1000
8100
FF00
Meaning
No error
Fatal error
Communication
Device-specific
Meaning
0
1
2
3
4
5
6
7
8
10
11
12
13
14
21
22
No error
Transmission error
System error
Unknown command
Wrong number of parameters
Wrong parameter value
Filter frequency error
Amplifier overflow
Command cannot be executed
Wrong channel selection
Measuring error
Triggering error
Range error
Taring error
Filter frequency warning
Tare status warning
A0563-5.6 en
HBM
44
PME-MP55
Subindex
Name
Measured values:
Gross measured value
Net measured value
maximum
minimum
Peak-to-peak
Measured value in
mV/V
Analogue output value
V
Measured value status
Format
Attr.
integer32
integer32
integer32
integer32
integer32
integer32
ro
ro
ro
ro
ro
ro
Decimal places
integer32
ro
Decimal places
unsigned8
ro
2000
2001
2002
2003
2004
2005
1
1
1
1
1
1
2006
2010
2011
Measured value
status_2
unsigned32
ro
2020*
I/O status
unsigned8
ro
HBM
Values
A0563-5.6 en
45
PME-MP55
Index
(hex)
Subindex
2080
2081
2082
2083
Format
Attr.
Edit mode
unsigned8
ro
Restart executed
unsigned8
0
0
Serial number
Exit from edit mode
vis.string
unsigned8
2101
Dialog:
Dialog language
unsigned16
2103
2104
0
1
2110
2111
2112
1
1
A0563-5.6 en
Name
Password
Enable keyboard and
menu
Parameter sets
Activate parameter set
integer16
unsigned16
unsigned16
unsigned16
unsigned16
Values
1: Edit mode on
0: Edit mode off
rw 1: Restart executed
0: Write = Delete
ro 12 char.
wo Value display after writing
with alloc. value
rw 1500 German
1501 English
rw
rw 0:
Input enabled
1:
Input disabled
Bit 0: Password input
Bit 1: Dialog
Bit 2: Parameter set
Bit 3: Display
Bit 4: Transducer
Bit 5: Conditioning
Bit 6: Analogue output
Bit 7: Limit values
Bit 8: Peak values
Bit 9: I/Os
Bit 10: CAN
Bit 11: Additional functions
Bit 15: Keyboard lock
rw 6600: Factory set-up
6601: Parameter set 1
6602: Parameter set 2
6603: Parameter set 3
6604: Parameter set 4
rw See above
ro See above
HBM
46
Index
(hex)
2120
2121
2122
HBM
PME-MP55
Subindex
Name
Format
Attr.
Values
1
1
Display adaptation
Decimal point position
Step
unsigned16
unsigned16
rw
rw
0..5
110:
111:
112:
113:
114:
115:
116:
117:
118:
119:
Transducers
Physical unit
unsigned16
rw
1603:
1604:
1605:
1606:
1607:
1608:
1609:
1610:
1611:
1612:
1613:
1614:
1615:
1616:
1617:
1618:
1619:
1620:
1621:
1622:
1623:
1624:
1625:
1626:
1627:
1628:
1629:
1630:
1631:
1632:
1633:
1634:
1635:
1636:
1637:
1
2
5
10
20
50
100
200
500
1000
g
kg
T
kT
TON
lb
oz
N
kN
bar
mbar
Pa
Pas
hPas
kPas
psi
m
mm
cm
m
inch
Nm
kNm
FTLB
INLB
m/m
m/s
m/s2
percent
perthou
ppm
S
MPas
MN
Blank
A0563-5.6 en
47
PME-MP55
Index
(hex)
Subindex
2130
2131
Name
Format
Attr.
Transducer type
unsigned16
ro
Excitation
unsigned16
ro
Values
2132
Range
unsigned16
ro
2133
Shunt
unsigned16
rw
2134
unsigned16
rw
2140
2141
1
1
integer32
integer32
rw
rw
2142
integer32
rw
Value in mV/V
2143
integer32
rw
Value e.g. in kN
2150
integer32
rw
Value in mV/V
2151
integer32
rw
Value in mV/V
2160
integer32
rw
Value e.g. in kN
2161
Shunt mismatch
direction
Transducer null mV/V
Transducer null phys.
unit
Transducer sensitivity
mV/V
Transducer nominal
value phys. unit
Input characteristics at
1st point mVV
Input characteristics at
2nd point mVV
Input characteristics at
1st point in phys. unit
Input characteristics at
2nd point in phys. unit
integer32
rw
Value e.g. in kN
A0563-5.6 en
HBM
48
Index
(hex)
PME-MP55
Subindex
Name
Format
Attr.
integer32
integer32
unsigned16
rw
rw
rw
Values
2180
2181
2182
1
1
1
Conditioning
Tare value
Zero balance value
Memory mode for taring
2183
unsigned16
rw
2185
2190
1
1
Zero reference
Filter frequency
integer32
unsigned16
rw
rw
2191
Filter characteristics
unsigned16
rw
21A0
unsigned32
rw
21A1
integer32
rw
21A2
unsigned16
rw
1:
0:
21C0
Standstill monitoring by
time window
Standstill monitoring by
amplitude
Activate motion count
indication
Analogue output
Mode of analogue output
(voltage/current)
unsigned16
ro
21C1
unsigned16
rw
21D0
integer32
rw
21D1
integer32
rw
Value e.g. in kN
21D2
integer32
rw
Value in V
21D3
290: "10 V
291: "20 mA
292: 4..20 mA
214: Gross
215: Net
204: Max
205: Min
218: Peak-to-peak
Value e.g. in kN
integer32
rw
Value in V
HBM
6611:
6610:
6611:
6610:
volatile
permanent
volatile
permanent
908: 0.05 Hz
914: 0.1 Hz
917: 0.2 Hz
921: 0.5 Hz
927: 1 Hz
931: 2 Hz
935: 5 Hz
941: 10 Hz
945: 20 Hz
949: 50 Hz
955: 100 Hz
958: 200 Hz
962: 500 Hz
141: Butterworth
142: Bessel
ms
on
off
A0563-5.6 en
49
PME-MP55
Index
(hex)
Subindex
Name
Format
Attr.
unsigned16
rw
2210
Limit switches
Enable limit value 1
2211
unsigned16
rw
2212
unsigned16
rw
2214
2215
2216
1
1
1
integer32
integer32
integer32
rw
rw
rw
2217
2218
2220
1
1
1
Starting delay LV 1
Cut-off delay LV 1
Switching level for limit
value 1
Hysteresis for limit value 1
Status of limit value 1
Enable limit value 2
integer32
unsigned8
unsigned16
rw
ro
rw
2221
unsigned16
rw
2222
unsigned16
rw
2224
2225
2226
1
1
1
integer32
integer32
integer32
rw
rw
rw
2227
2228
2230
1
1
1
Starting delay LV 2
Cut-off delay LV 2
Switching level for limit
value 2
Hysteresis for limit value 2
Status of limit value 2
Enable limit value 3
integer32
unsigned8
unsigned16
rw
ro
rw
2231
unsigned16
rw
2232
unsigned16
rw
A0563-5.6 en
Values
1: yes
0: no
214: Gross
215: Net
204: Min
205: Max
218: Peak-to-peak
130: Above limit
131: Below limit
ms
ms
1: yes
0: no
214: Gross
215: Net
204: Min
205: Max
218: Peak-to-peak
130: Above limit
131: Below limit
ms
ms
1: yes
0: no
214: Gross
215: Net
204: Min
205: Max
218: Peak-to-peak
130: Above limit
131: Below limit
HBM
50
PME-MP55
Index
(hex)
Subindex
Name
Format
Attr
.
2234
2235
2236
1
1
1
integer32
integer32
integer32
rw
rw
rw
2237
2238
2240
1
1
1
Starting delay LV 3
Cut-off delay LV 3
Switching level for limit value
3
Hysteresis for limit value 3
Status of limit value 3
Enable limit value 4
integer32
unsigned8
unsigned16
rw
ro
rw
2241
unsigned16
rw
2242
unsigned16
rw
2244
2245
2246
1
1
1
integer32
integer32
integer32
rw
rw
rw
2247
2248
1
1
Starting delay LV 4
Cut-off delay LV 4
Switching level for limit value
4
Hysteresis for limit value 4
Status of limit value 4
integer32
unsigned8
rw
ro
2260
Peak values
Input signal Min store
unsigned16
rw
2261
unsigned16
rw
2262
integer32
rw
2263
unsigned16
rw
1: enabled
2: disabled
2271
Additional functions
Hardware synchronisation
unsigned16
ro
2272
Sensitivity of keyboard
unsigned16
rw
6700:
6701:
7601:
7602:
7603:
HBM
Values
ms
ms
1: yes
0: no
214: Gross
215: Net
204: Min
205: Max
218: Peak-to-peak
130: Above limit
131: Below limit
ms
ms
214: Gross
215: Net
214: Gross
215: Net
Display / s
Master
Slave
low
medium
high
A0563-5.6 en
51
PME-MP55
Index
(hex)
Subindex
Name
Format
Attr.
Values
unsigned16
rw
200:
221:
222:
223:
224:
230:
231:
2310
Digital I/Os
Function of output 1
2311
Mode Outp. 1
unsigned16
rw
2312
2313
2314
2315
2316
2317
2320
1
1
1
1
1
1
1
Function of output 2
Mode Outp. 2
Function of output 3
Mode Outp. 3
Function of output 4
Mode Outp. 4
Remote function Taring
unsigned16
unsigned16
unsigned16
unsigned16
unsigned16
unsigned16
unsigned16
rw
rw
rw
rw
rw
rw
rw
2322
unsigned16
rw
2323
unsigned16
rw
see above
2324
unsigned16
rw
See above
2325
unsigned16
rw
See above
2326
=2327
1
1
unsigned16
unsigned16
rw
rw
See above
See above
2328
unsigned16
rw
See above
2330
Remote function
Max/Current value
Remote function
Min/Current value
Remote function Hold
Max value
Remote function Hold
Min value
Remote function Zeroing
Remote function Select
parameter set 1
Remote function Select
parameter set 2
Enable remote contacts
135: normal
136: inverse
See above
See above
See above
See above
See above
See above
100: no input
101: Input 1
102: Input 2
103: Input 3
104: Input 4
See above
unsigned16
rw
5: free
4: locked
A0563-5.6 en
No function
Limit value 1
Limit value 2
Limit value 3
Limit value 4
Error / Warning
Standstill
HBM
52
Index
(hex)
PME-MP55
Subindex
Name
Format
Attr.
unsigned16
rw
2400
CAN-interface
CAN baudrate
2410
PDO contents
unsigned16
rw
2411
2412
1
1
integer32
unsigned16
rw
rw
2600
2610
2620
1
1
1
Functions
SetZero
Tare
Clear Max store
unsigned8
unsigned8
unsigned8
wo
wo
wo
2621
unsigned8
wo
2622
2623
2630
1
1
1
unsigned8
unsigned8
unsigned8
rw
rw
rw
HBM
Values
1409: 10 kBaud
1411: 20 kBaud
1413: 50 kBaud
1427: 100 kBaud
1417: 125 kBaud
1419: 250 kBaud
1421: 500 kBaud
1423: 800 kBaud
1424: 1000 kBaud
214: Gross
215: Net
204: Max
205: Min
218: Peak-to-peak
0.1 ms
1253: Integer32
1257: Floating
1: Zeroing
1: Taring
1: Constant delete
2: 1x delete
1: Constant delete
2: 1x delete
1: Hold
1: Hold
Bit 0: Zeroing
Bit 1: Taring
Bit 4: Clear Max.
Bit 5: Clear Min.
Bit 6: Hold Max.
Bit 7: Hold Min.
A0563-5.6 en
53
PME-MP55
Subindex
Name
Format Attr.
1
1
1
1
1
1
1
Measured values:
Gross value
Net value
Maximum
Minimum
Peak-to-peak
Value in mV/V
Analogue output value
float
float
float
float
float
float
float
ro
ro
ro
ro
ro
ro
ro
3140
3141
1
1
Transducers
Transducer null mV/V
Transducer null physical unit
float
float
rw
rw
3142
3143
1
1
float
float
rw
rw
3150
3151
3160
1
1
1
float
float
float
rw
rw
rw
3161
float
rw
float
float
float
float
rw
rw
rw
rw
float
float
rw
rw
float
float
rw
rw
3000
3001
3002
3003
3004
3005
3006
3180
3181
3185
31A1
1
1
1
1
31D0
31D1
1
1
31D2
31D3
1
1
A0563-5.6 en
Conditioning
Tare value
Zero balance value
Zero reference
Standstill monitoring by amplitude
Analogue output
Zero point of analogue output phys. unit
Final value of analogue output phys.
unit
Zero point of analogue output V
Final value of analogue output V
Values
Value in mV/V
Value e.g. in
kN
Value in mV/V
Value e.g. in
kN
HBM
54
Index
(hex)
3216
3217
3226
3227
3236
3237
3246
3247
3262
HBM
PME-MP55
Subindex
Name
Format Attr.
1
1
1
1
1
1
1
1
Limit switches
Switching level for limit value 1
Hysteresis for limit value 1
Switching level for limit value 2
Hysteresis for limit value 2
Switching level for limit value 3
Hysteresis for limit value 3
Switching level for limit value 4
Hysteresis for limit value 4
float
float
float
float
float
float
float
float
rw
rw
rw
rw
rw
rw
rw
rw
Peak values
Envelope curve function discharge
float
rw
Values
Display value/s
A0563-5.6 en
55
PME-MP55
7.7 Examples
Example 1:
Read a net measured value as a float value via SDO transfer from an amplifier with module address 3.
Protocol on the amplifier:
Identifier
0603
CAN
identifier
1st
byte
40
Read
2nd byte
3rd byte
4th byte
5th byte
01
Index low
byte
30
Index
high byte
01
Subindex
6th
7th
byte
byte
X
X
dont care
8th
byte
X
1st byte
2nd byte
3rd byte
4th byte
5th byte
0583
CAN
identifier
43
Read
acknowl
edge
01
Index low
byte
30
Index
high byte
01
Subinde
x
m0
Low byte
6th
7th
byte
byte
m1
m2
Measured
value as a
float
8th
byte
m3
High
byte
Example 2:
Setting up the filter frequency as 200 Hz.
Protocol on the amplifier:
Identifier
1st byte
0603
CAN
identifier
2B
Write 2
bytes
4th
5th byte
6th
byte
byte
90
21
01
BB
03
Index
Index
Subin Low byte High byte
low byte high byte dex
958 = (03BB Hex)
7th
8th
byte
byte
X
X
dont care
1st byte
0583
60
CAN
Write
identifier acknowle
dge
A0563-5.6 en
2nd byte
3rd byte
4th byte
5th byte
90
Index low
byte
21
Index
high byte
01
Subindex
6th
7th
byte
byte
X
X
dont care
8th
byte
X
HBM
56
PME-MP55
Example 3:
The intention is to set up the tare value as 23.250 kg (transfer as a long value,
i.e. 23.250 =23250).
Assumed settings: unit kg; decimal places: 3
Protocol on the amplifier:
Identifier
1st byte
0603
CAN
identifier
23
Write 4
bytes
2nd byte
3rd byte
4th
5th byte
6th byte 7th
byte
byte
80
21
01
D2
5A
00
Index
Index
Subin Low byte
low byte high byte dex 23.250 kg=23500(=5AD2Hex)
8th
byte
00
High
byte
1st byte
0583
60
CAN
Write
identifier acknowle
dge
HBM
2nd byte
3rd byte
4th byte
5th byte
80
Index low
byte
21
Index
high byte
01
Subindex
6th
7th
byte
byte
X
X
dont care
8th
byte
X
A0563-5.6 en
57
PME-MP55
ADC+Ovf
ADC-Ovf
Grs+Ovfl
Grs-Ovfl
Scal.Err
UrCalErr
HrdwOvfl
ADC+Ovf
ADC-Ovf
Net+Ovfl
Net-Ovfl
Scal.Err
UrCalErr
PkMaxOvf
UrCalErr
PkMinOvf
UrCalErr
!
Gross
-------
! >T<
Net
-------
!
Max. peak value signal
-------
When enabled
!
Min. peak value signal
-------
PkPk Ovf
UrCalErr
HrdwOvfl
ADC+Ovf
ADC-Ovf
UrCalErr
HrdwOvfl
ADC+Ovf
ADC-Ovf
AnlgOvfl
AScalErr
!
Peak/peak signal
-------
mV/V
Input signal
-------
UrCalErr
A0563-5.6 en
HBM
58
PME-MP55
The current errors are displayed consecutively (see also page 23). Press
until you come to the ERROR display mode.
2)
3)
4)
Error message
Cause
Remedy
Hrdware1)
Connecting transducers
See pin assignment on page17
Adapt amplifier under
TRANSDUCER group
Connect sensor circuits
AD Conv
(ADC+Ovfl, ADC-Ovfl)
AnlgOutp
(AnlgOvfl)
StoreMin
(PkMinOvf)
StoreMax
(PkMaxOvf)
Net
(Net+Ovf; Net-Ovf)
Gross
(Grs+Ovf; Grs-Ovf)
NomVal over
Adapt range
Transducer
Connecting transducers
Connect sensor circuits
Scaling4)
(Scal.Err)
AnlgScal
(AScalErr)
ISyncErr
No internal synchronisation
(UrCalErr)
CAN Tx
(HrdwOvfl)2)
1)
Error messages without brackets: errors that are displayed consecutively in the ERROR display mode.
Error messages in brackets: errors that are displayed in the appropriate display mode (e.g. Gross, Net,
Analogue output signal).
On CAN bus "1 000 000 is output
See page 31
HBM
A0563-5.6 en
59
PME-MP55
Operating status:
LED
colour
Status
Meaning
Measuring mode
Bus mode
CAN Operational (PDO transfer possible)
Green
Steady
light
Ready to take
measurements
Green
Flashing
Yellow
Steady
light
Ready to take
measurements
LED
colour
Status
Red
Flashing
Meaning
Remedy
Measuring mode
Bus mode
Measured value
overflow
Adapt range
Restart
LCD error
Transducer resistance
too low
Red
Steady
light
A0563-5.6 en
Reduce excitation
voltage
CAN bus not ready for
communication
Wait
Connect
transducer, poss.
restart
Send PME to the
manufacturer
(HBM)
HBM
60
9
PME-MP55
Specifications
Type
MP55
Accuracy class
Supply voltage
VDC
VDC
W
0.12)
24; Potential separation from measuring
system, test voltage 350 Veff3))
18...30
9 maximum
kHz
Vrms
4.8"1 %
5 / 2.5 / 1
mH
m
Amplifier
Carrier frequency
Bridge excitation voltage UB ("5 %)
Transducers that can be connected
SG half and full bridge
Inductive half and full bridge, LVDTs
Permitted cable length between transducer and amplifier
Maximum permitted common mode
voltage
Common mode rejection
0...500 Hz
0...4800 Hz
Maximum differential voltage
Linearity deviation (typical)
Noise voltage1)
0...10 Hz
0...500 Hz
Measurement frequency range, adjustable (-1 dB)
Max. display resolution
Min. display resolution
Input sensitivities (measuring ranges
can be adjustet by DIP switch)
at UB=5 V
at UB=2.5 V
at UB =1 V
Low pass filter
Effect of operating voltage when specified range is changed (in relation to final
value)
on zero point
on sensitivity
dB
dB
mV
%
V/VPP
V/VPP
120
72
"30
0.025
Measuring range [mV/V]
3
50
500
0.2
3
30
1.5
25
250
Hz
0.05...500
999 999 digits at 6.67 % of amplifier input
range
10 digits at 100 % of amplifier input range
mV/V
mV/V
mV/V
Low
Medium
High
0.15...3
2.5...50
25...500
0.3...6
5...100
50...1000
0.75...15 12.5...250 125...2500
Adjustable in steps of 0.05 to 500 Hz
(Bessel and Butterworth filter characteristics)
%
%
1)
2)
0.25 with irradiation per EN61326 in the range of 700 MHz up to 1 GHz
3)
HBM
A0563-5.6 en
61
PME-MP55
3 mV/V
V/V
V/V
%
1
10
0.05
50 mV/V
500 mV/V
10
20
0.05
100
100
0.05
V/V
V
kOhm
Ohm
mA
Ohm
kOhm
10
10
10
"20; 4...20
500
100
mVSS
mV
<< 3
mV
A
%
3
6
0.05
%
%
ms
4
Gross, net, peak values
0...100
0.0033
1
ms
ms
ms
Unit/
sec
2
Positive, negative, peak-to-peak
1
2
2
0 to 999999
By reference to UB=5 V
A0563-5.6 en
HBM
62
Control outputs
Number
Rated voltage, external power supply
Permitted supply voltage range
Output current, max.
Short-circuit current, typically
Short-circuit period
Test voltage, typical
Control inputs
Input voltage range, LOW
Input voltage range, HIGH
Input current, typically, HIGH value =
24 V
Test voltage typical
Interface
Sampling rate, approx.
Protocol
Hardware bus link
baud rate
Maximum length of cable
Parameter memory (EEPROM)
PME-MP55
V
V
mA
4
24
18...30
0.5
0.8
unlimited
500
4
0...5
10...30
12
VDC
500
V
V
A
A
VDC
kBit/s
m
Display
Type
Keyboard
Rated temperature range
Service temperature range
Storage temperature range
Protection class
Dimensions, overall (W x H x D)
Weight, approx.
HBM
C [F]
C [F]
C [F]
mm
55 x 146 x 162
470
A0563-5.6 en
63
PME-MP55
10
Keyword index
A
Additional functions, 50
Address, 36
Amplifier settings, 8
Analogue output, 9 , 10 , 14 , 15 , 33 , 48
, 53
Current value, 36
Cyclical data transmission, 38
D
Dialog, 45
Digital I/Os, 51
Digital input, 15
Digital output, 15
Baudrate, 36
DIP switches, 8
Bessel, 32
Discharge rate, 34
Display adaptation, 46
Bridge type, 9 , 10
Display mode, 23
Butterworth, 32
Display resolution, 31
CAN bus, 14 , 19
connection, 14
Error acknowledge, 40
Error messages, 23
Error message, 24 , 58
F
Factory setting, 10
Factory settings, 9
Feedback bridges, 18
filter, 32
Fitting, 12
Four-wire connection, 18
Function, 52
Function test, 24
G
Grenzwert, 34
Control I/Os, 7 , 14
Control inputs, 14 , 16 , 35
Control outputs, 14 , 16
A0563-5.6 en
HBM
64
PME-MP55
H
hysteresis, 34
I
Input range, 9 , 10
Input remotes, 35
Inputs, 35
Interface, connection, 19
L
LED, 59
Level, 34
Limit switches, 49 , 54
Limit value level, 34
M
Master/Slave, 9 , 11 , 20
Measured value, 44
Measured values, 53
Measuring mode, 22
Motion count indication, 37
O
Object directory, 41 , 44
Operating direction, 34
Output characteristics, 33
Output ports, 35
Output rate, 36
Output remotes, 35
P
Parameters, 25
Description, 30
Read, Writen, 39
Recall, Save, 30
set up, 26
Password, 22 , 30
HBM
Peak value, 34 , 35 , 36 , 58
Peak values, 50 , 54
Peak-value store, 34
Pin assignment for terminal plugs, 15
PLC, 39
PLC connection, 16
Power failure, 15
Power supply, 14 , 15
Profile, 36
R
Recall ?, 30
Remotes, 35
Removal, 12
Ribbon cable, 20
S
Save ?, 30
Scale factor, 31 , 33
Scale range, 33
Scaling, 31 , 33
Set up of parameters, 26
Set-up mode, 22
Setting up, 24
Setting up , 21
Setting up an amplifier, 10
Setting zero, 32
Standstill, 23
Standstill time, 37
Starting delay, 34
Supply voltage, 15
Synchronisation, 14 , 20 , 59
T
Tare, 31 , 32
Terminal plug, 14 , 15
A0563-5.6 en
65
PME-MP55
Power supply, CAN bus, Synchronisation, Remote contacts, Control outputs, 14
Z
Zero balance, 31 , 32
Zero offset, 31
Terminating resistor, 19
Zero reference, 37
Transducer connection, 14
Strain gauge full and half bridges, Inductive full and half bridges, Potentiometric, Piezoelectric, LVDT, 17
Zero value, 32
Transducer excitation, 14
Transducers, 46 , 53
A0563-5.6 en
HBM
66
PME-MP55
HBM
A0563-5.6 en
PME-MP55
67
A0563-5.6 en
HBM
Modifications reserved.
All details describe our products in general form only.They are
not to be understood as express warranty and do not constitute
any liability whatsoever.
A0563-5.6 en