Unit 3 - Mammography Equipment

Download as pdf or txt
Download as pdf or txt
You are on page 1of 74

Screen-Film

Mammography Equipment
Unit 3
Bonnie A.Barnes,BA,R.T.(R)(M)(CT)(f)
Xuan Ho, Ph.D., R.T. (R)

Equipment

Dedicated Mammography Equipment


Digital Mammography
FDA/MQSA Requirements
Quality Control of Equipment

Dedicated Mammography
Equipment
C-arm X-ray tube stand
Mammography tube

X-ray tube stand

The range of vertical movement should


accommodate very tall patients, as well as
those who need to be seated.
There should be no obstacles under the
receptor tray that would interfere with the
patient should she need to be seated for
the exam.

C-arm X-ray tube stand


There should be at least 180 degree rotation of the Carm.
The point at which the C-arm connects to the tower
(the collar) should be isocentrically designed so that
when manipulating from the craniocaudal to the
oblique position no adjustments in height is needed.
Almost all current units require the C-arm to be
adjusted when changing to the oblique position.

Dedicated Mammography
Equipment
The single most important part of
dedicated equipment in mammography
is that the tube is designed and
constructed specifically for imaging the
soft tissue of the breast.

Rationales for Dedication


Small x-ray attenuation differences between
normal and cancerous tissues in the breast
require the x-ray equipment dedicated for use
to optimize breast cancer detection.
Attenuation differences between these minute
tissues is highest a very low x-ray energies
(10 to 15 keV) and is poor at higher energies
(>35 keV).
Low x-ray energies provide best differential
attenuation between the tissues; high
absorption results in high tissue doses and
long exposure time.

Dedicated Mammography
Equipment
Specialized components are

target, window,
filter, focal spot,
Source-to-image distance,
object-to-image receptor distance,
Generator, automatic exposure control,
grids, beam limiting devices,
compression devices, and
processor, and
magnification.

Dedicated Mammography
Equipment

Mammography Tube
Stationary Versus Rotating Anodes
An advantage of the rotating anode is a higher tube
loading capacity.
Higher milliamperage allows for shorter exposure times.
The disadvantage is that there is off-focus radiation
produced with this system.
Stationary tubes produce less off-focus radiation, which
creates less overall haze of the image, but have lower
tube rating capabilities.

Anode design
Anode construction:
Molybdenum
rhodium & tungsten

Combination of anode angle and tube tilt


SID of 65 cm requires the effective
anode angle to be at least 20 to avoid
field cutoff for the 24 x 30 cm field area

Mammography Tube
Target materials consist of three main types.
Molybdenum
Specialized Tungsten
Rhodium
Molybdenum
the best material to be used in mammography,
allows production of low energy spectrums of radiation
low kVp (26-30 kVp)

Tungsten and Rhodium are used for higher beam


needs, in dense breast tissues.

Filament design
Mammographic x-ray tubes typically have
dual filaments in a focusing cup that produces
0.3 and 0.1 mm nominal focal spot sizes

Minimize geometric blurring and maintain


spatial resolution necessary for detection of
microcalcifications

Focal spot
Important Factors

Standard: is used in obtaining routine mammographic views, where


images of the entire breast are needed.

0.30 to 0.60mm

Magnified: is used in imaging of a specific area of the breast using


magnification, spot compression or coning.
0.10-0.20mm

Actual: is the area of the target bombarded by electrons. The size is


dependent on the target angle, and the heat capacity of the anode. As
the target angle decreases from 90 degrees, the size of the actual focal
spot increases.

Effective : is a downward projection of the actual focal spot and is


measured by diameter (mm). The size of the effective output and
necessary field size. Two focal spot sizes are used in mammography in
order to accommodate views of the entire breast and special views
using spot compression, cone-down and magnification.

Source-to-Image distance
This is the entire distance of the X-ray beam
from the focal spot (source) to the image
receptor (film-screen). The greater the distance,
the less geometric blur occurs.
This affects the focal spot size, which in turn
affects the size of the object being imaged.
Typically the larger the SID, the larger the field
size.
With larger distance of the source to the image
(65 cm), more beam penetrability will be needed,
which then increases the heel effect.

Object to Image receptor


Distance
It is the distance of the object (the breast) to the image
receptor.
An increase of the object to IR distance will increase
magnification of an area of breast tissue. It will increase
the resolution and contrast of the breast image.
With an increase in distance, this will also increase the
amount of radiation dose to the patient, and the possibility
of motion artifact.
It is very important to explain to the patient that she will
need to have a longer breath hold during exposure.

Dedicated Mammography
Equipment
Filtration
Tube filtration
Exit window filtration
Heel effect
Effects on dose
Effects on contrast

Filtration
Materials that are placed in the path of the X-ray
beam in order to absorb those X-rays with energies
above and below the desired spectrum.
Tube filtration types:

Molybdenum (Mo) (best used for lower kVp)


Rhodium (Rh)
Yttrium
Aluminum (used for above 30 kVp)

Target/Filter Combinations
Mo target with 0.03 mm Mo filter
(Mo/Mo)
Rh target with 0.025 mm Rh filter
(Rh/Rh)
Mo target with Rh filter
Never use Rh target with Mo filter!

Filtration
Exit window
Beryllium instead of glass in a dedicated
mammography tube housing.
Glass acts as a filter when dealing with this
soft end of the X-ray spectrum, and it filters
out photons that would not provide contrast.

Heel effect
Due to the angle of the anode, some of the X-rays it just
produced will be reabsorbed by the anode material. This is
called the heel effect.
The steeper the anode angle (the closer to 0 degrees) the
more heat units the tube can withstand but the more
pronounced the heel effect will be.
The greater the heel effect the lower the intensity of the Xray beam at the anode (nipple) side. The lower the X-ray
intensity at the nipple side of the film the less field
coverage there is available for the 24 dimension of a
24x30 cm image receptor.

Filtration
Effects on dose
It is important to use optimal filtration.
Filtration lowers the amount of radiation dose
the patient receives.
Filtration shortens the exposure time, giving the
patient only the radiation dose needed to obtain
a good quality radiograph.
With excessive filtration, a built-in microswitch
should prohibit an exposure from being made.

Filtration
Effects on contrast
With more filtration, image contrast
is diminished, less scatter radiation
is produced.
The beam gains penetrability, good
quality image is resulted.

Generator

Types
Ripple factor

Generators
Single phase unit: uses a slower alternating current
to produce radiation. This allows for longer
exposures to achieve a quality image, giving the
patient a larger dose of radiation.
Three phase unit: uses a faster 6-pulse current to
achieve the same quality image. It uses less time,
which gives the patient less radiation dose, and less
chance for motion artifact.
High-frequency/Constant Potential units: does not
use alternating or pulse currents. It uses a constant
current creating no ripple effect. Radiation dose to
the patient is even less, and the same quality image
is achieved.

Generators
Ripple Factor
occurs in single and three-phase generators.
Fluctuations of the current cause the ripple
factor.
A near-constant (minimal ripple) current
generated by three-phase rectified or high
frequency generator is necessary to produce
optimum image quality with the lowest
radiation dose to patients.

Generator - Tube Capacity


(mA output)
The amount of heat a dedicated mammography tube
can handle with:

a molybdenum target
a beryllium window
added molybdenum filtration
a large focal spot of 100mA
a small focal spot of 25 mA is:

250,000hu (heat units)


The anode cooling rate (Max) is: 25,000 hu/min.

Dedicated Mammography
Equipment
Automatic exposure control (AEC)
Purpose
Types

Dedicated Mammography
Equipment
Automatic exposure control (AEC)
Radiation that is transmitted through the
breast is converted into an electronic
signal, which terminates the exposure
when the predetermined level of
radiation has been reached.

Automatic exposure control


(AEC)
Photomultipliers
a series of photocells rather than a
single pickup.
A high-technology microprocessor
chip allows the phototimer to average
the multiple photocell readings.

Automatic exposure control


(AEC)
Solid State
allows the new phototimers to
remember and to repeat kVp
settings.
are able to respond to varying kVp
settings.

Grids
Reciprocating
These grids are best.
The movement helps clean up scatter radiation.
The ratios of these grids is usually 4:1 or 5:1, with
some units having higher grid ratios.
The higher is the grid ratio, the better the clean-up of
scatter radiation, giving better contrast.
Dose to the patient increases in this case.
Materials used for this type of grid can be of carbon
fiber, pressed cardboard, or wood. These materials
allow to keep bucky factors low, between 2 and 2.5.

Grids
Stationary:
These grids are focused and must be centered to the
central ray in order to minimize the grid lines.
Offsetting the grid will increase the shadowing of the
lead lines on the radiograph.
The grid ratio for these grids is typically with a 3 -5:1.
Since these grids are so thin, aluminum interspace
material is used to provide some rigidity.

Beam limiting devices

The purpose to having beam limiting


devices is to allow the proper collimation
of the X-ray beam on the image
receptor.
This allows for only the primary beam to
be used, instead of scatter radiation.

Beam limiting devices


Three types are currently used:
fixed apertures,
a set of interchangeable cones of various sizes,
an internal set of collimating blades.

Three-sided collimation is also used to


collimate the sides of the bucky that are
lateral, medial and anterior to the breast. The
side closest to the patients chest wall, is not
collimated.

Compression devices

Purpose
Compression testing
Types
Paddle shapes, sizes and purposes

Compression devices
Purpose
Vigorous compression of the breast has an important
role in screen-film and digital mammography.
It creates decreased motion unsharpness, geometric
unsharpness, increased contrast, separation of
superimposed breast tissues, reduced radiation dose,
more uniform film density and more useful assessment
of the apparent density of masses.
The only negative aspect of compression is the varying
degree of discomfort that each patient will experience.
Every woman will have her own level of tolerance for
compression.

Compression devices
Compression testing
It is very important to test the compression
of the compression paddle at least every
six months.
The degrees of compression of any
mammography unit may be:
maximum 40 lbs. of pressure
minimum is 25 lbs. of pressure.

Compression testing

Sscale used to measure degree of pressure

Compression devices
Types
Manual
This type of compression is used by the
technologist using her hand to manually
compress the breast, using the compression
paddle.
It may be necessary to use this type of
compression for very anxious women, or if
doing an invasive procedure requiring slow
and controlled compression.

Compression devices
Motorized
Foot-controlled by the technologist
Allows the technologist to have two free
hands to position the breast.
The RT still is in control of how fast and how
much compression is used, but it is done so to
allow the technologist maneuverability.
This type of compression is used for most
mammography exams.

Compression device
Paddle shapes, sizes and purposes
Shapes
There is a paddle used for 18x24cm and 24x30cm
cassettes, and there are paddles used for spot
compression, which are much smaller.
There are also paddles that are fenestrated which are
used for wire-localization procedures.

Compression devices
The purpose of having the proper shaped
paddle, is to allow for proper compression
of the entire breast.
In situations where spot compression or
magnification are necessary, the smaller
paddle is used to compress a small area
of the breast.
This is to try to smooth out tissue that was
not well compressed, or to magnify only a
small area of the breast for better
visualization.

Dedicated Mammography
Equipment

Magnification

Purpose
Focal spot size
Air gap technique
Effect of dose
Magnification factor

Magnification
Used to better visualize smaller areas
of breast tissue.
Calcifications are not visualized well
on standard views.
They can only be diagnosed, whether
benign or suspicious for malignancy, if
they are magnified.

Magnification
Improved resolution occurs with a smaller
focal spot size.
Most mammographic units have magnification
factors of 1.5x to 2.0x.
An air gap technique is used between the
breast and the image receptor to achieve
magnification.
The larger the magnification factor the better
the resolution, giving more radiation to the
patient.

Dedicated Mammography
Equipment
The further
away the
image
receptor is to
the tube, the
larger the
magnification
is obtained.

Digital Mammography
Theory
Approaches of digital mammography
Benefits

Digital Mammography
The theory behind digital mammography is
that it is to increase the incidence of detecting
more breast cancers, that would not normally
be detected as easily in film-screen systems.
It allows for faster image production, with less
radiation to the patient.
This new system is more costly, but in the
long run will save the facility money in storage
and film processing costs.

Digital Mammography
Direct digital systems use amorphous selenium
photodetector screens. These screens have high X-ray
absorption efficiency, high intrinsic resolution, low noise
and a well-established manufacturing process.
Indirect digital systems use cesium iodide doped with
thallium to absorb the X-rays. These systems generate
light scintillation that then is detected by an array of
photodiodes.
Indirect systems have resolution degradation caused by
light spread in the scintillator, and poor quantum efficiency
with thin scintillators.
Some systems use a charged-couple-device (CCDs) as an
alternative light collection array and readout method.

Digital Mammography
In producing the image, spatial resolution plays a big
part.
Spatial resolution is determined by the pixel size.
With fewer microns per pixel, the spatial resolution of
the detector is improved.
The amount of contrast information available in the
image is defined by the bit depth per pixel.
The digital image is formed as a 2-dimensional matrix
of square picture elements of a fixed size, typically
0.04mm to 0.1mm on a side. It is a sampled
representation of the pattern of X-ray transmission
through the breast.

Digital Mammography
Noise to Radiation Dose Ratio
A lower the dose, the noisier is the image.
A higher dose means less noise in the image.
Radiation dose in digital systems may be the same
or less than that used in film-screen mammography.
The reduction in dose is the direct result of greater
sensitivity and the lower overall system noise of the
charge-coupled devices found in digital systems.

Digital Mammography
With digital systems image enhancement is much greater
than that with film-screen systems.
Digital mammographys excellent contrast resolution ,
especially in dense breast tissues, will eventually make
way for more accurate breast cancer detection.
In film-screen systems there are blind areas where the film
is very dark or very light, and lesions are more likely to be
overlooked.
Because there is a linear response of digital detectors to
incident radiation, there is more contrast between areas of
similar X-ray absorption. In addition, image contrast can be
manipulated on the computer monitor for improved
conspicuity.

Digital Mammography
Another benefit to digital mammography systems is
the amount of time it takes to produce an image. It
takes approximately four seconds to image the breast
in one projection.
The technologist is able to look at her positioning right
away, and let the patient go when the exam is
finished.
If any additional imaging is needed, enhancement of
the image is done instead. The patient does not have
to be called back to the facility to have more X-rays
taken. This alleviates any additional anxiety patients
may have.

Computer-Aided Detection
This computer digitizes mammography film. It
aids the radiologist in detecting breast cancer.
It is like having a second radiologist looking
at the patients mammogram.
The computer can pick up to 20% more breast
cancers than with standard/previous methods
of diagnoses.
This system is less costly than going to an all
digital mammography system, but does not
alleviate additional imaging procedures from
being performed.

Digital Mammography
R2 unit is a CAD
(Computer aided
detection)
system used to
digitize
mammograms.

Telemammography
This is being used more and more, as
facilities become all digital. Radiologists are
able to be off-site, and look at a womans
mammogram from a remote location.
The benefit of this is that other physicians
within the network may review a patients
mammogram. It saves time and money.
Immediate reading of mammograms can be
performed in mobile units.
The downside to this is that the patient does
not have a physician to talk to in person, at
the time of her exam.

FDA/MQSA Requirements
Mammography Equipment
Dedicated
Gantry assembly motion
Image receptors
Compression
Beam limitation and light fields

FDA/MQSA Requirements
Mammography equipment
Dedicated: MQSA requires that all
radiographic equipment used for
mammography shall be specifically
designed for mammography and
shall be certified.

FDA/MQSA Requirements
Gantry assembly motion
The assembly shall be capable of being fixed
in any position where it is designed to operate.
Once fixed in any such position, it shall not
undergo unintended motion.
The mechanism ensuring compliance shall not
fail in the event of power interruption.
The gantry shall have visual indicators of the
gantry angle, and must be able to rotate 180
degrees.

FDA/MQSA Requirements
Image receptors
Systems using screen-film image receptors shall
provide, at a minimum, for operation with image
receptors of 18x24cm and 24x30cm.
Systems using screen-film image receptors shall
be equipped with moving grids matched to all
image receptor sizes provided.
Systems used for magnification procedures shall
be capable of operation with the grid removed from
between the source and image receptor.

FDA/MQSA Requirements
Compression
All mammography systems shall incorporate a
compression device.
Each system shall provide: an initial power-driven
compression activated by hands-free controls
operable from both sides of the patient; and fine
adjustment compression controls operable from
both sides of the patient.

FDA/MQSA Requirements
Beam limitation and light fields
All systems shall have beam-limiting
devices that allows the entire chest wall
edge of the X-ray field to extend to the
chest wall edge of the image receptor and
provide means to assure that the X-ray field
does not extend beyond any edge of the
image receptor by more than 2% of the
SID.

FDA/MQSA Requirements
Beam limitation and light fields
Illumination requirements: The facility shall
make special lights for film illumination, i.e.,
hot-lights, capable of producing light levels
greater than that provided by the view box,
available to the interpreting physicians.
Exposure interlock systems: should forbid
exposure when in the open position. Any
interlocks on the machine should be checked
by attempting exposure while they are in the
open position.

FDA/MQSA Requirements
Source-to-image receptor distance
The source-to-image receptor distance
(SID) may be fixed or it may be variable. In
either case the size of the focal spot must
match the SID of the unit to provide good
resolution as well as to allow the use of
acceptable technical factors when dealing
with the inverse square law.

FDA/MQSA Requirements
Dose limitations
MQSA requires that the average glandular dose
delivered during a single cranio-caudal view of
an FDA-accepted phantom simulating a
standard breast shall not exceed 3.0 milligray
(mGy) (0.3 rad) per exposure.
The dose shall be determined with technique
factors and conditions used clinically for a
standard breast.

FDA/MQSA Requirements
Infection control
Facilities shall establish and comply with a system
specifying procedures to be followed by the facility
for cleaning and disinfecting mammography
equipment after contact with blood or other
potentially infectious materials. This system shall
specify the methods for documenting facility
compliance with the infection control procedures
established and shall:
Comply with all applicable Federal, State, and local
regulations pertaining to infection control; and
comply with the manufacturers recommended
procedures for the cleaning and disinfection of the
mammography equipment used in the facility.

FDA/MQSA Requirements

Foaming cleanser is
used for cleaning
hands in between
patients.

FDA/MQSA Requirements

Quality Assurance of Equipment


Each mammography facility shall have a
quality assurance program in place.
The required maintenance of log books,
phantom images, clinical image monitoring,
clinical image interpretation, a physicist
survey, and all medical records.
For the sake of not missing a lesion, it is our
responsibility to adhere to and uphold the final
regulations mandated by the FDA/MQSA.

You might also like