C&i Systems
C&i Systems
C&i Systems
Presentation Outline:
1.Some definitions & basics of Pressure, Flow & Temp. measurement
Measurement: Pressure
Outline:
Some Definitions
Pressure Units
Manometers
Pressure Switches
Measurement: Pressure
Terminology
Measurement: Pressure
Repeatability refers to the ability of a pressure sensor to provide
the same output with successive applications of the same pressur
e.
Hysteresis is a sensor's ability to give the same output at a given
pressure while increasing and decreasing the pressure.
Measurement: Pressure
Pressure : Definitions
Absolute pressure
Atmospheric pressure
Differential pressure
Gauge pressure
Measurement: Pressure
Zero Reference , Gauge, Absolute, Atmospheric
Pressure
Any pressure above atmosphere is called gauge pressure
Any pressure below atmosphere is a vacuum (negative gauge pressure)
Absolute pressure (psia) is measured from a perfect vacuum
Measurement: Pressure
Units
The SI unit for pressure is the Pascal (Pa);1Pa= 1 Nm-2
Non-SI unit pound (Lb) per square inch (psi) and bar are commonly used
Pressure is sometimes expressed in grams-force/cm 2or as kgf/cm2 (KSC)
1 atm=1.03 ksc=14.696 psi=760mmHg=10000 mmWC
=101325 Pa
Standard pressure:Pressure of normal (standard) atmosphere is defined as
standard pressure
Measurement: Pressure
Pressure Measuring devices
Manometers
using water ,mercury and other liquids of known density
For measuring low pressures.
Mechanical/Elastic Pressure Sensors
Electrical Pressure Transducers
For measuring pressure of all ranges for telemetering purposes.
Manometer:
May be used for gauge, differential, and absolute measurements with a suitable reference.
Useful mainly for lower pressure work because the height of the column of mercury will
otherwise become very high.
Measurement: Pressure
Elastic Pressure Sensors
The basic pressure sensing elements:
A: C-shaped Bourdon tube , B: Helical Bourdon tube , C: flat diaphragm
D: Convoluted diaphragm, E: Capsule , F: Set of bellows
Measurement: Pressure
Electrical Pressure Sensors
Potentiometer Sensor
2. Inductive
3. Capacitive
4. Piezoelectric
5. Strain Gauge
1.
Measurement: Pressure
Pressure Switches
Applications
Alarm (Status)
Shutdown (Hi/Lo Limits)
Control (ON/OFF)
A switch is an instrument that automatically senses some process
variable (such as pressure) and provides an on/off signal relative to
some reference point.
Set Point
Sensing
Element
Conditioning
Circuit
Bourdon Tube
Bellows
Diaphragm
Strain Gauge
Mechanical
Switch
Transistor
Measurement: Pressure
High Pressure In High Temperature
* When high process temperatures are present, various methods
of isolating the pressure instrument from the process are used.
* These include siphons, chemical seals with capillary tubing for
remote mounting, and purging.
Measurement: Pressure
Pressure Snubbers
To filter out pressure spikes, or to average out pressure pulses, snubbers are
installed between the process and the instrument
Instrument indicates avg pr.
Snubber
Before use
After use
Measurement: Pressure
Chemical Seal or diaphragm Protector
Chemical seals are used when media can falsify the
pressure measurements due to high temperature, high
viscosity or their property to crystallise
Measurement: Pressure
Siphon
A siphon is a coiled tube. This coil provides a large cooling surface and
the trap created prevents the condensate from draining away.
A siphon is required for hot condensing. fluids, such as steam, to assure
a liquid trap.
It is used to prevent live steam from entering and damaging the device.
It is used to protect the instrument from hydraulic or thermal shocks.
The two most common forms of siphon tube are the 'U' and Pigtail
types.
Measurement: Flow
Types of flow meters:
1.
2.
3.
4.
5.
Measurement: Flow
Orifice Flow-meters
An orifice plate is a restriction with an opening smaller than the pipe diameter which
is inserted in the pipe; the typical orifice plate has a concentric, sharp edged opening.
Because of the smaller area the fluid velocity increases, causing a corresponding
decrease in pressure.
The flow rate can be calculated from the measured pressure drop across the orifice
plate, P1-P3.
The orifice plate is the most commonly used flow sensor, but it creates a rather large
non-recoverable pressure due to the turbulence around the plate, leading to high
energy consumption.
Measurement: Flow
Venturi Tube
Measurement: Flow
Pitot Tubes
Pitot tubes were invented by Henri Pitot in 1732 to measure the
flowing velocity of fluids. Basically a differential pressure (dp) flow
meter, a pitot tube measures two pressures: the static and the total
impact pressure.
Pitot tubes are used to measure air flow in pipes, ducts, stacks, and
liquid flow in pipes, open channels.
While accuracy and rangeability are relatively low, pitot tubes are
simple, reliable, inexpensive, and suited for a variety of
environmental conditions, including extremely high temperatures
and a wide range of pressures.
Measurement: Flow
Pitot Tubes
A single-port pitot tube can measure the flow velocity at only a single point
in the cross-section of a flowing stream.
The probe must be inserted to a point in the flowing stream where the flow
velocity is the average of the velocities across the cross-section, and its
impact port must face directly into the fluid flow.
Measurement: Flow
Pitot Tubes
The point velocity of approach (VP) can be calculated by taking the square root
of the difference between the total impact pressure (PT) and the static pressure (P)
and multiplying that by the C/D ratio, where C is a dimensional constant and D is
density:
The pitot tube measures the static and dynamic (or impact) pressures of the fluid
at one point in the pipe.
The flow rate can be determined from the difference between the static and
dynamic pressures which is the velocity head of the fluid flow.
Measurement: Flow
Vortex Flow-meters
Measurement: Flow
Vortex Flow-meters
This is detected by a sensor, such as capacitive sensor and
fed to the electronic processor as a primary, digitized, linear
signal.
Capacitive sensors with integrated temperature measurement
can directly register the mass flow of saturated steam as well.
Measurement: Flow
Ultrasonic flow-meters
Swimming against the flow requires more power and more time
than swimming with the flow. Ultrasonic flow measurement is
based on this elementary transit time difference effect.
Measurement: Flow
Ultrasonic flow-meters
By using the absolute transit times both the averaged fluid velocity and
the speed of sound can be calculated.
This time difference is a measure for the averaged velocity of the fluid
along the path of the ultrasonic beam
Measurement: Flow
Ultrasonic flow-meters
Advantages:
Non-invasive measurement
Measurement: Flow
Coriolis Mass Flow-meters
Measurement: Flow
Coriolis Mass Flow-meters
The processor analyzes this information and uses it to
compute the rate of mass flow.
Advantage
This principle is used in a huge range of industry sectors,
including pharmaceuticals, chemicals and petrochemicals, oil
and gas, food etc.
Measurement: Flow
Major issues for selecting flow-meters
Accuracy
Repeatability
Linearity
Reliability
Range/Span
Dynamics(Response time)
Safety
Maintenance
Cost
Measurement: Temp.
Measurement Devices
Thermocouples
Resistance Thermometers
Thermistors
Bimetallic Thermometers
Acoustic Pyrometers
Local Instruments
Measurement: Temp.
Thermocouple
IT IS BASED ON SEEBECK EFFECT WHICH SAYS THAT WHEN
HEAT IS APPLIED TO A JUNCTION OF TWO DISSIMILAR METALS
AN EMF IS GENERATED WHICH CAN BE MEASURED AT THE
OTHER JUNCTION
T/C Connection
COMPENSATING CABLE
HOT JUNCTION
TO DDC CARDS
TERMINAL END
CJC BOX
Measurement: Temp.
Thermocouple
Types of T/C:E,J,K,T,R,S,B
K (Chromel & Alumel; Ni-Cr &Ni-Al) Type: mostly used in power plant for low
temp. application )
R (Platinum & Platinum-Rhodium) Type: Used for high temp. application. Highly
resistant to oxidation & corrosion
Advantages: Disadvantages: - Low Cost
- Sensitivity low & low voltage output
- No moving parts, less likely to be broken. susceptible to noise
-Wide temperature range.
- Accuracy not better than 0.5 C
-Reasonably short response time.
- Requires a known temperature
- Reasonable repeatability and accuracy.
reference
Measurement: Temp.
RESISTANCE THERMOMETER (RTD)
THE RESISTANCE OF A CONDUCTOR CHANGES WHEN ITS TEMPERATURE IS CHANGED .THIS
PROPERTY IS UTILISED TO MEASURE THE TEMPERATURE.
Rt = Ro (1+dT)
Measurement: Temp.
THERMISTORS
THERMISTORS ARE GENERALLY COMPOSED OF SEMICONDUCTOR
MATERIALS.THEY HAVE A NEGATIVE COEFFICIENT OF TEMPERATURE SO
RESISTANCE DECREASES WITH INCREASE IN TEMP.
Measurement: Temp.
ACOUSTIC PYROMETER
For measurement of temperatures across large spaces of known distance in a noisy, dirty and
corrosive environment such as a coal-fired utility boiler, or a chemical recovery boiler.
LOCAL INDICATION
BIMETALLIC THERMOMETER
b)
c)
d)
b)
c)
ii)
Gas analyzers
iii)
Smoke monitors
SCALING
CORROSION
CARRY OVER
EMBRITTLEMENT
b)
c)
d)
PARAMETER
DM WATER
a) Conductivity
b) Cation Conductivity
Condensate
pump
discharge
(CEP)
a) Conductivity
b) Cation Conductivity
Economizer
Inlet
Boiler water
a) Conductivity
b) Cation Conductivity
c) pH
S/cm
S/cm
d) Na+
ppb
ppb
c) Hydrazine
a) Conductivity
b) pH
UNIT
S/cm
ppb
S/cm
LIMIT
<0.3
<5
<0.3
9.0-9.2
<5ppb
<10
<5
<0.3
10-20
100
9.1-9.4
c) Silica
ppb
100
a) Conductivity
b) Cation Conductivity
S/cm
<5
<0.3
CALIBRATION
REPAIR
TESTING
CALIBRATION:
Pressure switch , Transmitter , Gauge
Flow Transmitter
Level Switch
b)
c)
Test benches with standard power supply sockets (e.g. 24VDC, 48VDC,
220VDC, 110VAC, 230VAC etc.) in each bench depending on requirement.
d)
2.
3.
PULVERISERS/FEEDERS ON/OFF
4.
5.
6.
CONTROL
Group Control
Sub-Group Control
Sub-Loop Control
Drive Level Control
Programmable Logic Control(PLC)
Group Control : Start and Stoppage of a Group of equipment is
accomplished by Group Control(GC).
Ex. :CEP GC, Equipment Cooling GC etc.
Bump less Transfer : The arrangement where the transfer from auto to manual
mode does not affect the process.
Runback Mode
BLI signal as feed forward signal for boiler firing rate control
Load deviation as feed orward signal for boiler firing rate control
Unit master demand is limited by unit capability , TSE margins and unit max/min load
set points.
Unit target load is derived from unit master after the limitations.
Unit target load is used as feed forward signal to the boiler firing rate control.
Turbine control utilises the unit load as turbine load set point after adapting the same
by steam generation delay.
DATA ACQUISITION
DATA PROCESSING
DATA REPRESENTATION
Analog inputs:
1. Thermocouple Input ( mV )
K-Type T/C ( Cr-Al ) : For temp < 600 Deg C& used in Flue Gas path
after FSH outlet.
R-Type T/C ( Pt-Pt-Rh ) : For temp > 600 Deg C used in PSH & FSH
region of FG path.
2. RTD Input ( Resistance )
Pt-100 RTD : For Brg. Temp measurement.
Cu-53 RTD : For HT motor & Generator Stator winding temp.
measurement.
MAGTAPE UNIT
TREND RECORDER
PRINTERS
If all the pages are full (normally no. of alarm pages & alarm per
page is predefined) and any new alarm comes , then oldest alarm
will disappear from the alarm page as FIFO basis
CRT Displays
1. Alarm CRT display
2. Utility CRT display
TIME ACTIVATED
EVENT ACTIVATED
DEMAND LOGS
PRINOUTS
SOE PRINTOUT
SOLID-STATE CONTROLLERS
COMPUTERISED CONTROLS
CONTROL SYSTEM
PROVISION OF LVS
OLCS - THE SEQUENCE CONTROL, INTERLOCK OF ALL THE PLANT SYSTEMS WHICH ARE
NOT COVERED IN THE SG-C&I AND TG-C&I. THIS INCLUDES MAJOR AUXILIARIES LIKE FD/ID/PA
FANS, AIR-PREHEATER, BFP/CEP/CWP/ BCWP , DMCWP/CLCWP AND ELECTRICAL BREAKERS.
CLCS
FW FLOW (DRUM LEVEL), FURNACE DRAFT, COMBUSTION CONTROL (FUEL FLOW AND AIR
FLOW), PA HDR PRESSURE CONTROL, DEAERATOR/HOTWELL/HEATER LEVEL CONTROLS
ETC.
IN CONTROL STRATEGY
NO NEED OF
CHANGING TECHNOLOGY
Indian:
Total installed capacity only 1362 MW in 1947
Per Capita consumption 631 units (2005-06) only with installed capacity of 1,77,000 MW
GDP growth of 8%, power growth required 10%
To add 1,00,000MW capacity by 2017
Liberalizations of the sector
5. Sealing & Cable dressing in MAX DNA panels during unit Shutdown
6. Disabling various ports for removable drives of MAX DNA work
stations for system reliability
7. Installation of ON Line printers of MAX DNA system for daily LOGs.
Daily LOGs are essential for analysis of different plant parameters by
O&E dept.