Manual Sanwa YX360TRF (Analog)

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Black

Attachment of Hand Strap


1 Loosen the screws fixing the rear
case and remove it.
2 Hand strap is attached to connecting point.
3 Put back the rear case where it
was and fix it with the screws.





Storage of Test Leads


When placing the test leads in the
storage space, roll it 3 times, then
put in the test pin side first as shown
below.

Fig. 2

DCA 50

X1

DCV 0.1

X 0.01

DCA 0.25

X 0.001

DCA 25m

X 0.1

DCA 2.5m

X 0.01

DCV 50

X1

ACV 50

X1

1 X 1k

X 1k

1 X 100

X 100

1 X 10

X 10

1X1

X1

DCV 250

X1

DCV 2.5

X 0.01

DCV 0.25

X 0.001

ACV 250

X1

Range

Multiplied

1 X 100k

X 100k







Range
DCV 10
DCV 1000
ACV 750
ACV10
C (F)
DCV 25
DCV 5
150mA at X 1
15mA at X 10
1.5mA at X 100
150 A at X 1k
1.5 A at X 100k
LV
hFE
ACV 10
ACV 50
ACV 250
ACV 750

Multiplied
X1
X 100
X 100
X1
X1
X1
X1
X 10
X1
X 0.1
X 10
X 0.1
X1
X1
X1
14dB added
28dB added
40dB added

Use of Cover (example for the body cover)


When this tester is out of use:
Attach the cover to the panel face for safekeeping.
When measuring:
Attach it either to the rear case side
or use it as a stand as shown below.

The specifications described in this manual are subject to


change without notice.
Application
This instrument is portable multitester designated for measurement of weak current circuits.

APPLICATION
Note: The definition of installation category,i.e.
CAT II : Local level, appliances, portable equipment etc.,
with smaller transient overvoltages than installation category III.
CAT III : Distribution level, fixed installation, with smaller
transient overvoltages than installation category IV.

D-1034-236 YX360 TRF


*1 Not including the resistance of fuse.
*2 The maximum value when the pointer was moved by
charged current in the capacitor.
7
0 ~ 150mA at X1 range
0 ~ 15mA at X10 range
LI
0 ~ 1.5mA at X100 range Approximate Value
0 ~ 150 A at X1K range
0 ~ 1.5 A at X100K range
Use the optional probe
HV (DC
DC25kV
high volt)
hFE
1000 atX10 range

Drop shock proof

A taut-band structure is adopted in the meter part.


The meter part is designed to withstand shock.

Circuit protection

The circuit is protected by fuse even when voltage of up


to AC 230V is impressed on each range for 5 seconds.

Internal battery

R6 (IEC) or UM-3 1.5V X 2

Internal fuse

F500mAH/250V 5.2 X 20mm Fast acting fuse

Operating temperature
and humidity range

5 ~ 31C, 80%RH max.


31< ~ 40C, 80 ~ 50%RH (decreasing linearly)

Storage temperature/
Humidity

-10 ~ 50C 70%RH max. No condensation

Withstand voltage

6kV AC (1min.) between input terminal and case

2.5m/25m/0.25

Dimensions and weight


Accessories

159.5 X 129 X 41.5 mm/ approx. 320g


Instruction manual 1, Hand strap 1,

2k/20k/200k/2M
(X1/X10/X100/X1K)
200M
(X100K)
10F
-10dB ~ +22dB
(for 10VAC)
~ +62dB

dB
C

HV probe, HV-10T hFE probe, HFE-6T Test lead for repair, TL-6IT
1
DCACV

Items

50
DCV
(NULL)

5/25

ACV ~

10/50/250/750
250/1000

DCVCV

SCALE READING

0.25/2.5/10/50
0.1

Specification

General Specifications
(Temperature : 232C humidity 75% RH max. No condensation)

BODY COVER, TEST LEADS, HAND STRAP


02-0807 2040 2040

Function

Full scale value

HFE-6T probe
HV-10T probe
Current across test pins

5% of arc

Center value 201


Max. value 2k1
Release voltage 3V

Approximate Value

*2

Approximate Value

Input impedance 9k1/V

3% of arc
3% against
full scale
3% against
full scale
4% against
full scale
Accuracy
5% against
full scale
3% against
full scale
3% against
full scale
5% against
full scale

*1 Voltage drop 0.25V


*1 Voltage drop 0.1V
Input impedance 9k1/V
30Hz ~ 100kHz within
3% f.s. (AC10V range)
Input impedance 40k1/V
Input impedance 9k1/V
Input impedance 20k1/V
Remarks

Measurement Range and Accuracy

SPECIFICATIONS

Dempa Bldg., 4-4 Sotokanda 2-Chome


Chiyoda-Ku,Tokyo,Japan

Fig. 1

11. Never use the meter on the electric circuits that exceeds 3kVA.
12. Pay special attention when measuring the voltage of AC 33 Vrms
(46.7V peak) or DC 70V or more to avoid injury.
13. Never apply input signals exceeding the maximum rating input
value.
14. Never use the meter for measuring the line connected with equipment (i.e. motors) that generates induced or surge voltage since it
may exceed the maximum allowable voltage.
15. Never use the meter if the meter or test leads are damaged or broken.
16. Never use an uncased meter.

SANWA ELECTRIC
INSTRUMENT CO.,LTD.
INSTRUCTION MANUAL

MULTITESTER

YX360TRF

To ensure that the meter is used safely, follow all safety and operating instructions.

WARNING
Precautions for Safety Measurement
Symbols
The following cautionary signs appear on the multitester and
in this manual.
Disobedience to instructions with this sign may lead to trouble
with the tester and accidents such as electrical shock.
This sign cautions that high voltage is applied to parts marked
with it.

NAMES OF COMPONENTS
Thank you for purchasing a SANWA tester Model YX360TRF. You are
kindly requested to thoroughly read this manual before use for safety.
Especially, SAFETY INFORMATION and MEASURING PROCEDURE are important. Keep this manual together with the tester so as
no to lose it.

The following are precautions to prevent accidents such as electrical


shocks.
Be sure to read them before using the tester.

INTRODUCTION

17. Be sure to use a fuse of the specified rating or type. Never use a
substitute of the fuse or make a short circuit of the fuse.
18. Always keep your fingers behind the finger guards on the probe
when making measurements.
19. Be sure to disconnect the test pins from the circuit when changing
the function or range.
10. Before starting measurement, make sure that the function and
range are properly set in accordance with the measurement.
11. Never use the meter with wet hands or in a damp environment.
12. Never use test leads other than the specified type.
13. Never open the case except when replacing batteries or fuses. Do
not attempt any alteration of original specifications.
14. To ensure safety and maintain accuracy, calibrate and check the
meter at least once a year.
15. Indoor use only.

SAFETY INFORMATION
MEASURING PROCEDURE

WARNING
Confirm the range of use before measurement.
Preparation for Measurement
1 Adjustment of meter zero position:
Turn the zero position adjuster so that the pointer may align
right to the zero position.
2 Range selection:
Select a proper range for the item to be measured and set
the range selector knob accordingly.

NOTE
When determining a measuring range, select a one higher
voltage than the value to be measured as well as where the
pointer of a meter moves to a considerable extent.
However, select the maximum range and measure in case
the extent of value to be measured can not be predicted.
Measuring DCV
1 Set the range selector knob to
an appropriate DCV range.
2 Apply the black test pin to the
negative potential, and the red
test pin to the positive potential of the circuit.
3 Read the movement of the
pointer by V and A scale.

Measuring ACV
1 Turn the range selector knob to an appropriate ACV range.
2 Apply the test leads to the circuit to be measured.
3 Read the movement of the pointer by V and A scale. (Use
AC 10V scale for 10V range only.)
Since this instrument provides
the mean value system for its AC
voltage measurement circuit, AC
waveform other than sine wave
may cause an error.
Errors occur under such frequencies other than those in the speciFig. 4
fication table.
Measuring DCA

1 Turn the range selector knob to an appropriate DCA range.


2 Take out the circuit to be measured and apply the black test
pin to the negative potential,
and the red test pin to the
positive potential of the circuit.
3 Read the movement of the
pointer by V and A scale.
Fig. 5

Cut off.

Measuring 1
WARNING
Do not measure the resistance in a circuit where
a voltage is present.
1
2

3
4

Fig. 6

Note: How to replace battery.


1 Loosen the screws fixing the rear
case and remove it.
2 Replace R6 (UM-3) to fresh dry
batterise.
3 Put back the rear case where it
was, and fix it with the screws.
Note: Be sure to use the same rated
Fig. 7
fuse. In case a fuse other than
the same rate (see SPECIFICATIONS) is used, an error in
indication occurs and/or circuit protection is made unable.

MAINTENANCE

USAGE OF OPTIONAL PROBES

10

Usage of High Voltage Probe (HV-10T)


Up to DC 25kV of CRT anode voltage can be measured by
connecting optional HV-10T probe.

WARNING
Keep the hand (finger) away from high voltage
power supply. Electric shock may occur due to discharge.
Measurement should be limited only to micro current circuits.
1 Turn the range selector knob and set it to HV PROBE (DC
2.5V range).
2 Connect the jack of the black lead of the probe to the black
test pin, and the jack of the red lead to the red test pin.
3 Apply the probe clip to the earth side and the measuring pin
to the point to be measured.
4 Read out measured value on 0 ~ 250 of V scale in kV unit
after multiplying it by 0.1.

1 Loosen the screws fixing the


rear case and remove it.
2 Pull the fuse out of holder on the
Fig. 11
circuit board and replace it.(Fig. 11)
3 Put back the rear case where it was and tighten the screws.
4 Check and see whether or not indications of respective
ranges are normal (check other parts for any failures).
Storage and Other Precautions
1 Avoid giving the meter any excessive shock or vibration by
loading it on the motorbike, for instance.
2 Keep off dust and moisture from the meter.
3 Do not leave the meter for a long time in places of a high
temperature (higher than 55C), a high humidity (higher than
80%), and dew condensation.
4 The meter cover is treated with antistatic coating. Do not
wipe it hard or clean it with volatile solvent. Use a soft
brush to remove dust.

13

Measuring Capacity (C)


1 Set the range selector knob to C(F) .
2 Measure the capacitance by applying
the test pin to the capacitor to be measured after the 01 adjustment is made
in the same manner as in the resistFig. 8
ance measurement.
3 The pointer moves full scale by the charge current to the
capacitor. However, the pointer gradually starts returning to its
original position. Read the indicated maximum value on C(F)
scale.
Note: Be sure to short circuit the both ends of the capacitor for
discharge prior to the initial measurement or in such case
after the measurement is made.
Note: Pay due attention to the polarity (+ and -) of the capacitor.
(Connect + side of the capacitor to - side of the meter.)
Measuring AF Output (dB)
dB (decibel) is measured in the same way as the ACV
measurement, but by reading the dB scale instead.
For measurement on the 10V range, the dB scale (- 10dB ~
+22dB) is read directly, but, when measured on the 50V
range, 14dB is added. On the 250V range, 28dB is added to
the reading on the scale, and on the 1000V range, 40dB
added.
Thus, the maximum dB readable is 22 + 40 = 62 (dB) measured on the 1000V range.
Note: Cut direct current with a capacitor of 0.1F or more when
measuring such signal as having direct current.

11

AFTER - SALES SERVICE

How to Replace the Fuse


If voltage over 100V is applied
to DCA and 1 ranges, the fuse
may blow out to protect the
circuit.

C
black
B

C
red

Turn the range selector knob to an appropriate 1 range.


Short the red and black test pins and turn the 01 adjuster so
that the pointer may align exactly to 01. (If the pointer fails
to swing up to 01 even when the 01 adjuster is turned
clockwise fully, replace the internal battery with a fresh one.)
Apply the test pin to measur
resistance.
Read the movement of the pointer by 1 scale.

Note: The polarity of + and - is


reverse to that of the test
leads when measurement is
done in 1 range.

WARNING
Connect the meter in series with the load.
Fig. 3

Measuring DCV (NULL)


1 Set the range selector knob to
an appropriate DCV (NULL)
range.
2 Turn the 01
adjuster so
that the pointer may align
exactly to 0 by DCV scale.
3 Apply the black test pin to the
negative potential side, and
the red test pin to the positive
potential side of the circuit.
4 Read the movement of the pointer by DCV scale.

MEASURING PROCEDURE
Measuring of Iceo (Leak Current) for Transistor
1 Adjust 01 by setting the range selector
NPN
knob to a proper range from X1 ~ X1k.
2 For NPN transistor, apply a black test E
lead to the collector and a red one to
red
the emitter.
For PNP transistor, the red one to the colPNP
lector and the black one to the emitter.
3 Determine the leak current by ICEO
E
scale indicated on the scale plate. (Unit
black
in A, mA)

Fig. 9

Measuring of Diode (including LED)


1 Adjust 01 by setting the range selector knob to a proper
range from X1 (150mA) ~ X100k (1.5A).
2 Apply the black test lead to anode side and the red one to
cathode side when measuring IF (forward current). Apply
the black test lead to cathode side and the red one to anode
side when measuring IR (reverse current).
3 Read the indicated value by LI
black
red
I
scale. (The pointer moves to a
considerable extent for IF, and
A
K
little extent for IR)
I
4 The value indicated on LV scale
during the measurement is the
forward voltage of diode.
Fig. 10

Usage of hFE PROBE (HFE-6T)


1 Set the range selector knob to X10 range ( hFE PROBE).
2 Short circuit both the red and black test pins to adjust 01.
3 Connect the black test pin to the probe jack when a transistor to be measured is NPN, and the red pin to the probe
jack for PNP transistor.
4 Connect the black clip of the probe to the transistor base
and the red clip to the collector.
5 Connect the remaining test lead to the emitter and measure
hFE.
6 Read the indicated value of the meter on hFE scale.

12

Warranty and Provision


Under Sanwa's general warranty policy, each instrument is warranted to be free from defects in workmanship or material
under normal use for the period of one (1) year from the date of
purchase.
This warranty policy is valid within the country of purchase only,
and applied only to the product purchased from Sanwa authorized agent or distributor.
This warranty does not apply to fuses, disposables batteries, or
any product or parts, which have been subject to one of the following causes:
1. A failure due to improper handling or use that deviates
the instruction manual.
2. A failure due to inadequate repair or modification by people other than Sanwa service personnel.
3. A failure due to causes not attributable to this product
such as fire, flood and other natural disaster.
4. Non-operation due to a discharged battery.
5. A failure or damage due to transportation, relocation or
dropping after the purchase.
Repair
Please contact Sanwa authorized agent/distributor/service
provider, listed in our website, in your country with your
information.
SANWA web site
http://www.sanwa-meter.co.jp
E-mail: [email protected]

14

15

D-1034-236 YX360 TRF

Black

Attachment of Hand Strap


1 Loosen the screws fixing the rear
case and remove it.
2 Hand strap is attached to connecting point.
3 Put back the rear case where it
was and fix it with the screws.




X1

ACV 50

Storage of Test Leads


When placing the test leads in the
storage space, roll it 3 times, then
put in the test pin side first as shown
below.

X 0.01

DCV 0.1
DCA 50

Fig. 2

X1
X1

DCV 50

X 0.01

DCA 2.5m

X 0.1

DCA 25m

X 0.001

DCA 0.25

X1

ACV 250

X 0.001

DCV 0.25

X 0.01

DCV 2.5

X1

DCV 250

X1

1X1

X 10

1 X 10






X 100

1 X 100

X 1k

1 X 1k

X 100k

1 X 100k

Multiplied

Range

6
Range
DCV 10
DCV 1000
ACV 750
ACV10
C (F)
DCV 25
DCV 5
150mA at X 1
15mA at X 10
1.5mA at X 100
150 A at X 1k
1.5 A at X 100k
LV
hFE
ACV 10
ACV 50
ACV 250
ACV 750

Multiplied
X1
X 100
X 100
X1
X1
X1
X1
X 10
X1
X 0.1
X 10
X 0.1
X1
X1
X1
14dB added
28dB added
40dB added

Use of Cover (example for the body cover)


When this tester is out of use:
Attach the cover to the panel face for safekeeping.
When measuring:
Attach it either to the rear case side
or use it as a stand as shown below.

The specifications described in this manual are subject to


change without notice.
Application
This instrument is portable multitester designated for measurement of weak current circuits.

APPLICATION
Note: The definition of installation category,i.e.
CAT II : Local level, appliances, portable equipment etc.,
with smaller transient overvoltages than installation category III.
CAT III : Distribution level, fixed installation, with smaller
transient overvoltages than installation category IV.

2k/20k/200k/2M
(X1/X10/X100/X1K)
200M
(X100K)
10F
-10dB ~ +22dB
(for 10VAC)
~ +62dB

dB
C
1

159.5 X 129 X 41.5 mm/ approx. 320g


Instruction manual 1, Hand strap 1,

Dimensions and weight


Accessories

6kV AC (1min.) between input terminal and case

Withstand voltage

-10 ~ 50C 70%RH max. No condensation

Storage temperature/
Humidity

5 ~ 31C, 80%RH max.


31< ~ 40C, 80 ~ 50%RH (decreasing linearly)

Operating temperature
and humidity range

2.5m/25m/0.25
DCACV
50
10/50/250/750

ACV ~

F500mAH/250V 5.2 X 20mm Fast acting fuse

Internal fuse

R6 (IEC) or UM-3 1.5V X 2

Internal battery

250/1000

The circuit is protected by fuse even when voltage of up


to AC 230V is impressed on each range for 5 seconds.

Circuit protection

0.25/2.5/10/50

A taut-band structure is adopted in the meter part.


The meter part is designed to withstand shock.

Drop shock proof

DCV
(NULL)

5/25

DCVCV

0.1

Specification

General Specifications
(Temperature : 232C humidity 75% RH max. No condensation)

SCALE READING

02-0807 2040 2040

0 ~ 150mA at X1 range
0 ~ 15mA at X10 range
LI
0 ~ 1.5mA at X100 range Approximate Value
0 ~ 150 A at X1K range
0 ~ 1.5 A at X100K range
Use the optional probe
HV (DC
DC25kV
high volt)
hFE
1000 atX10 range

HV probe, HV-10T hFE probe, HFE-6T Test lead for repair, TL-6IT
Items

BODY COVER, TEST LEADS, HAND STRAP

*1 Not including the resistance of fuse.


*2 The maximum value when the pointer was moved by
charged current in the capacitor.
7
Function

Full scale value

Approximate Value
Approximate Value
5% of arc
3% of arc
3% against
full scale
3% against
full scale
4% against
full scale
Accuracy
5% against
full scale
3% against
full scale
3% against
full scale
5% against
full scale

HFE-6T probe
HV-10T probe
Current across test pins
Input impedance 9k1/V
*2
Center value 201
Max. value 2k1
Release voltage 3V
*1 Voltage drop 0.25V
*1 Voltage drop 0.1V
Input impedance 9k1/V
30Hz ~ 100kHz within
3% f.s. (AC10V range)
Input impedance 40k1/V
Input impedance 9k1/V
Input impedance 20k1/V
Remarks

Measurement Range and Accuracy

SPECIFICATIONS

Dempa Bldg., 4-4 Sotokanda 2-Chome


Chiyoda-Ku,Tokyo,Japan

Fig. 1

11. Never use the meter on the electric circuits that exceeds 3kVA.
12. Pay special attention when measuring the voltage of AC 33 Vrms
(46.7V peak) or DC 70V or more to avoid injury.
13. Never apply input signals exceeding the maximum rating input
value.
14. Never use the meter for measuring the line connected with equipment (i.e. motors) that generates induced or surge voltage since it
may exceed the maximum allowable voltage.
15. Never use the meter if the meter or test leads are damaged or broken.
16. Never use an uncased meter.

SANWA ELECTRIC
INSTRUMENT CO.,LTD.
INSTRUCTION MANUAL

MULTITESTER

YX360TRF

To ensure that the meter is used safely, follow all safety and operating instructions.

WARNING
Precautions for Safety Measurement
Symbols
The following cautionary signs appear on the multitester and
in this manual.
Disobedience to instructions with this sign may lead to trouble
with the tester and accidents such as electrical shock.
This sign cautions that high voltage is applied to parts marked
with it.

NAMES OF COMPONENTS
Thank you for purchasing a SANWA tester Model YX360TRF. You are
kindly requested to thoroughly read this manual before use for safety.
Especially, SAFETY INFORMATION and MEASURING PROCEDURE are important. Keep this manual together with the tester so as
no to lose it.

The following are precautions to prevent accidents such as electrical


shocks.
Be sure to read them before using the tester.

INTRODUCTION

17. Be sure to use a fuse of the specified rating or type. Never use a
substitute of the fuse or make a short circuit of the fuse.
18. Always keep your fingers behind the finger guards on the probe
when making measurements.
19. Be sure to disconnect the test pins from the circuit when changing
the function or range.
10. Before starting measurement, make sure that the function and
range are properly set in accordance with the measurement.
11. Never use the meter with wet hands or in a damp environment.
12. Never use test leads other than the specified type.
13. Never open the case except when replacing batteries or fuses. Do
not attempt any alteration of original specifications.
14. To ensure safety and maintain accuracy, calibrate and check the
meter at least once a year.
15. Indoor use only.

SAFETY INFORMATION
MEASURING PROCEDURE

WARNING
Confirm the range of use before measurement.
Preparation for Measurement
1 Adjustment of meter zero position:
Turn the zero position adjuster so that the pointer may align
right to the zero position.
2 Range selection:
Select a proper range for the item to be measured and set
the range selector knob accordingly.

NOTE
When determining a measuring range, select a one higher
voltage than the value to be measured as well as where the
pointer of a meter moves to a considerable extent.
However, select the maximum range and measure in case
the extent of value to be measured can not be predicted.
Measuring DCV
1 Set the range selector knob to
an appropriate DCV range.
2 Apply the black test pin to the
negative potential, and the red
test pin to the positive potential of the circuit.
3 Read the movement of the
pointer by V and A scale.

Measuring ACV
1 Turn the range selector knob to an appropriate ACV range.
2 Apply the test leads to the circuit to be measured.
3 Read the movement of the pointer by V and A scale. (Use
AC 10V scale for 10V range only.)
Since this instrument provides
the mean value system for its AC
voltage measurement circuit, AC
waveform other than sine wave
may cause an error.
Errors occur under such frequencies other than those in the speciFig. 4
fication table.

WARNING
Connect the meter in series with the load.

Measuring DCV (NULL)


1 Set the range selector knob to
an appropriate DCV (NULL)
range.
2 Turn the 01
adjuster so
that the pointer may align
exactly to 0 by DCV scale.
3 Apply the black test pin to the
negative potential side, and
the red test pin to the positive
potential side of the circuit.
4 Read the movement of the pointer by DCV scale.

1 Turn the range selector knob to an appropriate DCA range.


2 Take out the circuit to be measured and apply the black test
pin to the negative potential,
and the red test pin to the
positive potential of the circuit.
3 Read the movement of the
pointer by V and A scale.
Fig. 5

WARNING
Do not measure the resistance in a circuit where
a voltage is present.
1 Turn the range selector knob to an appropriate 1 range.
2 Short the red and black test pins and turn the 01 adjuster so
that the pointer may align exactly to 01. (If the pointer fails
to swing up to 01 even when the 01 adjuster is turned
clockwise fully, replace the internal battery with a fresh one.)
3 Apply the test pin to measur
resistance.
4 Read the movement of the pointer by 1 scale.
Note: The polarity of + and - is
reverse to that of the test
leads when measurement is
done in 1 range.

Measuring DCA

Fig. 3

Measuring 1

Cut off.

Fig. 6

Note: How to replace battery.


1 Loosen the screws fixing the rear
case and remove it.
2 Replace R6 (UM-3) to fresh dry
batterise.
3 Put back the rear case where it
was, and fix it with the screws.
Note: Be sure to use the same rated
Fig. 7
fuse. In case a fuse other than
the same rate (see SPECIFICATIONS) is used, an error in
indication occurs and/or circuit protection is made unable.

Measuring Capacity (C)


1 Set the range selector knob to C(F) .
2 Measure the capacitance by applying
the test pin to the capacitor to be measured after the 01 adjustment is made
in the same manner as in the resistFig. 8
ance measurement.
3 The pointer moves full scale by the charge current to the
capacitor. However, the pointer gradually starts returning to its
original position. Read the indicated maximum value on C(F)
scale.
Note: Be sure to short circuit the both ends of the capacitor for
discharge prior to the initial measurement or in such case
after the measurement is made.
Note: Pay due attention to the polarity (+ and -) of the capacitor.
(Connect + side of the capacitor to - side of the meter.)
Measuring AF Output (dB)
dB (decibel) is measured in the same way as the ACV
measurement, but by reading the dB scale instead.
For measurement on the 10V range, the dB scale (- 10dB ~
+22dB) is read directly, but, when measured on the 50V
range, 14dB is added. On the 250V range, 28dB is added to
the reading on the scale, and on the 1000V range, 40dB
added.
Thus, the maximum dB readable is 22 + 40 = 62 (dB) measured on the 1000V range.
Note: Cut direct current with a capacitor of 0.1F or more when
measuring such signal as having direct current.

10

11

MEASURING PROCEDURE

USAGE OF OPTIONAL PROBES

MAINTENANCE

AFTER - SALES SERVICE

Measuring of Iceo (Leak Current) for Transistor


1 Adjust 01 by setting the range selector
NPN
knob to a proper range from X1 ~ X1k.
2 For NPN transistor, apply a black test E
lead to the collector and a red one to
red
the emitter.
For PNP transistor, the red one to the colPNP
lector and the black one to the emitter.
3 Determine the leak current by ICEO
E
scale indicated on the scale plate. (Unit
black
in A, mA)

Usage of High Voltage Probe (HV-10T)


Up to DC 25kV of CRT anode voltage can be measured by
connecting optional HV-10T probe.

C
black
B

C
red

Fig. 9

Measuring of Diode (including LED)


1 Adjust 01 by setting the range selector knob to a proper
range from X1 (150mA) ~ X100k (1.5A).
2 Apply the black test lead to anode side and the red one to
cathode side when measuring IF (forward current). Apply
the black test lead to cathode side and the red one to anode
side when measuring IR (reverse current).
3 Read the indicated value by LI
black
red
I
scale. (The pointer moves to a
considerable extent for IF, and
A
K
little extent for IR)
I
4 The value indicated on LV scale
during the measurement is the
forward voltage of diode.
F

Fig. 10

12

WARNING
Keep the hand (finger) away from high voltage
power supply. Electric shock may occur due to discharge.
Measurement should be limited only to micro current circuits.
1 Turn the range selector knob and set it to HV PROBE (DC
2.5V range).
2 Connect the jack of the black lead of the probe to the black
test pin, and the jack of the red lead to the red test pin.
3 Apply the probe clip to the earth side and the measuring pin
to the point to be measured.
4 Read out measured value on 0 ~ 250 of V scale in kV unit
after multiplying it by 0.1.
Usage of hFE PROBE (HFE-6T)
1 Set the range selector knob to X10 range ( hFE PROBE).
2 Short circuit both the red and black test pins to adjust 01.
3 Connect the black test pin to the probe jack when a transistor to be measured is NPN, and the red pin to the probe
jack for PNP transistor.
4 Connect the black clip of the probe to the transistor base
and the red clip to the collector.
5 Connect the remaining test lead to the emitter and measure
hFE.
6 Read the indicated value of the meter on hFE scale.
13

How to Replace the Fuse


If voltage over 100V is applied
to DCA and 1 ranges, the fuse
may blow out to protect the
circuit.
1 Loosen the screws fixing the
rear case and remove it.
2 Pull the fuse out of holder on the
Fig. 11
circuit board and replace it.(Fig. 11)
3 Put back the rear case where it was and tighten the screws.
4 Check and see whether or not indications of respective
ranges are normal (check other parts for any failures).
Storage and Other Precautions
1 Avoid giving the meter any excessive shock or vibration by
loading it on the motorbike, for instance.
2 Keep off dust and moisture from the meter.
3 Do not leave the meter for a long time in places of a high
temperature (higher than 55C), a high humidity (higher than
80%), and dew condensation.
4 The meter cover is treated with antistatic coating. Do not
wipe it hard or clean it with volatile solvent. Use a soft
brush to remove dust.

14

Warranty and Provision


Under Sanwa's general warranty policy, each instrument is warranted to be free from defects in workmanship or material
under normal use for the period of one (1) year from the date of
purchase.
This warranty policy is valid within the country of purchase only,
and applied only to the product purchased from Sanwa authorized agent or distributor.
This warranty does not apply to fuses, disposables batteries, or
any product or parts, which have been subject to one of the following causes:
1. A failure due to improper handling or use that deviates
the instruction manual.
2. A failure due to inadequate repair or modification by people other than Sanwa service personnel.
3. A failure due to causes not attributable to this product
such as fire, flood and other natural disaster.
4. Non-operation due to a discharged battery.
5. A failure or damage due to transportation, relocation or
dropping after the purchase.
Repair
Please contact Sanwa authorized agent/distributor/service
provider, listed in our website, in your country with your
information.
SANWA web site
http://www.sanwa-meter.co.jp
E-mail: [email protected]

15