GEHC Service Manual - PROCARE B40 B20 Monitor PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 252

GE Healthcare

PROCARE* Monitor B40/B20


Users Reference Manual

PROCARE Monitor B40/B20


English
2050801-001 D (Paper)
2011 General Electric Company.
All Rights Reserved.

PROCARE Monitor B40/B20


User's Reference Manual
Related to software license VSP-A
Monitoring functions

0459

Conformity according to the Council Directive 93/42/EEC concerning Medical Devices.


All specifications are subject to change without notice.
Document no. 2050801-001 D
July 18, 2011

GE Medical Systems Information Technologies, Inc.


8200 West Tower Avenue
Milwaukee, WI USA
Zip: 53223
Tel: 1 414 355 5000 (outside US)
800 558 5102 (US only)
Fax: 1 414 355 3790
www.gehealthcare.com
Copyright 2011 General Electric Company. All rights reserved.

GE Healthcare
3F Building 1, GE Technology Park
1 Huatuo Road
Shanghai PRC 201203
Tel: +86 21 3877 7888
Fax: +86 21 3877 7451

Classifications
In accordance with IEC 60601-1

Class I and internally powered equipment - the type of protection against electric shock.

Type BF or CF equipment. The degree of protection against electric shock is indicated by


a symbol on each parameter module.

Equipment is not suitable for use in the presence of a flammable anesthetic mixture with
air or with oxygen or nitrous oxide.

Continuous operation according to the mode of operation.

Portable Monitor

In accordance with IEC 60529

IP21 - degree of protection against harmful ingress of water.

In accordance with EU Medical Device Directive

IIb.

In accordance with CISPR 11:

Group 1 Class A;

Group 1 contains all ISM (Industrial, scientific and medical) equipment in which there
is intentionally generated and/or used conductively coupled radio-frequency
energy which is necessary for the internal functioning of the equipment itself.

Class A equipment is equipment suitable for use in all establishments other than
domestic and those directly connected to a low-voltage power supply network
which supplies buildings used for domestic purposes.

Trademarks
Dash, PROCARE, DINAMAP, EK-Pro, Trim Knob, Unity Network, Datex, Ohmeda, S/5, D-fend, Dfend+, Mini D-fend, OxyTip+, EarSat, FingerSat, FlexSat are trademarks of GE Healthcare. All
other product and company names are property of their respective owners.

Table of Contents

Introduction
About this device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Intended purpose (Indications for use) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Intended audience. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Responsibility of the manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Product availability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Printed copies of this manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
CE marking information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Product Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Illustrations and names . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Conventions used in this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Related documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Safety message signal words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
ESD precautionary procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8

System description
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Principles of functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Functional options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
System introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Optional components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Rear panel connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
E-miniC module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Keyboards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Command Board keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Battery indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Conditioning a battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Abbreviations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14

i
Document no. 2050801-001

B40/B20 Patient Monitor

Technical specification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25


General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
Parameters specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27

Installation
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Unpacking instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Choosing location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Mounting the monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Connection to mains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Install the network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Install the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Inserting and removing the E-miniC module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
To insert module: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
To remove module: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Visual indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Installation checkout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Recommended tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Visual inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

Monitoring basic
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Starting and ending. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Starting monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Admitting a patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adding demographics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Loading previous data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contin. Previous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
During monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automatic saving of patient data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Discharging the patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automatic discharge of the patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ending monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup monitor before use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

ii
Document no. 2050801-001

4-1
4-1
4-4
4-4
4-4
4-4
4-4
4-5
4-5
4-6
4-6
4-6
4-6
4-6
4-7
4-7
4-8
4-8

Alarms
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Alarm indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Alarm categories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Alarm light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Alarm tones . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Alarm activation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Alarms Setup menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Adjusting alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Adjusting limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Choosing automatic limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Returning to default limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Changing alarm sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Silencing alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Silencing audible alarms temporarily . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Reactivating alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Silencing audible alarms permanently . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Reactivating alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Deactivating alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Automatic recording on alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Showing alarm history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Other adjustable features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Displaying limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Enabling or disabling alarm silencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Latching alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Reminder volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Changing the tone pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16

Monitor setup
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Configuration and passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Setting time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Battery setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Changing monitor installation settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Changing units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Changing alarm options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Changing printer settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Changing the monitor settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Changing the user modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Saving changes in user modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Changing the user mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Changing the startup mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Renaming a mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Loading modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Changing the Normal Screen layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Modifying waveform fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
iii
Document no. 2050801-001

B40/B20 Patient Monitor

Modifying digit fields. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9


Modifying split screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Modifying the minitrend length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Other adjustable screen features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Changing sweep speeds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Displaying pulse rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Changing parameter colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Changing the recorder and printer settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Recorder settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Printer settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Configuring trends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Configuring trend pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Setting trend length. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Setting trend scales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Setting the default trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Using Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14

Trends
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Trended parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Most common tasks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Minitrend view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Minitrend length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Removing minitrend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Graphical trend view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Graphical trend pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Changing trend length and resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Moving on graphical trend pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Recording and printing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Factory default parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
Numerical trend view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Numerical trend pages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Moving between numerical trend pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Recording and printing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Factory default parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Erasing trend data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11

Print and record


Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Direct function keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Recording waveforms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sample of waveform recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Selecting waveforms for recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

iv
Document no. 2050801-001

8-1
8-1
8-1
8-2
8-3
8-3

Changing the paper speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4


Controlling the recording time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Selecting the recording delay time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
Recording on alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Recording trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Recording numerical trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Tabular trend format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Selecting graphical trends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-8
Inserting recorder paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Printing currently displayed screen contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Printing all the information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Changing the printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Other adjustable features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10

Cleaning and care


Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preventive maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Daily and between the patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Regular checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Every six months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Every 12 months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power interruption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Changing fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Permitted detergents and disinfectants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

9-1
9-1
9-2
9-2
9-2
9-2
9-3
9-4
9-4
9-4
9-5
9-5
9-5
9-7

10 Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Power interruption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Other situations 8

11 ECG
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2
v
Document no. 2050801-001

B40/B20 Patient Monitor

Displaying ECG and heart rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3


Preparing the patient and placing the electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Preparing the patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Placing the electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
Patient connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Lead measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Color and letter coding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
ECG Setup menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
Selecting a lead. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-8
Selecting user leads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9
Viewing a cascaded ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9
Adjusting the ECG size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-9
Starting relearning manually . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-10
Setting heart rate alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-10
Setting PVC alarm limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-11
ST segment analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
Display of ST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
Monitoring the ST segment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-12
Setting the ST measurement points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-13
Setting ST alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-14
Description of the ST segment measurement algorithm . . . . . . . . . . . . . . . . . . . . . 11-15
Test results of ST segment measurement algorithm testing . . . . . . . . . . . . . . . . . . 11-15
Monitoring arrhythmia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-16
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-16
Adjusting arrhythmia alarm settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-17
Detecting the ECG arrhythmia alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-17
Selecting leads for the arrhythmia analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-18
Description of the arrhythmia algorithm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-18
Test results of arrhythmia algorithm testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-18
Monitoring pacemaker patients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-19
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-19
Other adjustable features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-20
ECG printout type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-20
ECG waveform sweep speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-20
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-20

12 Impedance respiration
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-2
Respiration detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-2
Respiration rate calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-3
Displaying impedance respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-3
Patient connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-4
Activating measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-5

vi
Document no. 2050801-001

Improving waveform readability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-6


Correcting the respiration number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-7
Measurement limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-8
Turning off the measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-8
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-9

13 Non-invasive blood pressure


Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-1
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-2
Direct function keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-3
Displaying non-invasive blood pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-3
Patient connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-4
Selecting a cuff and a cuff hose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-4
Connecting the cuff hose. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-5
NIBP Setup menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-5
Starting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-6
During measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-6
Autocycling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-7
Setting cycle time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-7
Setting custom mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-8
Starting and stopping autocycling: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-8
Starting and stopping a single manual measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-9
Starting and stopping a continuous measurement (STAT) . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-9
Principles of SuperSTAT Noninvasive Blood Pressure Determination . . . . . . . . . . . . . . . . 13-10
Systolic Search . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-11
Automatic NIBP double check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-12
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-12

14 Pulse oximetry
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-1
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-2
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-2
Display of pulse oximetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-3
Patient connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-4
Connecting the patient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-4
Pulse oximetry menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-5
During monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-6
Removing the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-6
Measurement limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-7
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14-7

vii
Document no. 2050801-001

B40/B20 Patient Monitor

15 Invasive blood pressure


Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-1
Direct function keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-2
Display of invasive blood pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-2
Patient connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-3
Starting with accurate values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-4
Invasive Pressures menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-5
IBPx Setup menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-5
Labeling channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-6
Cerebral perfusion pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-7
Adjusting alarm sources and limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-7
Smart InvBP and flushing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-8
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-8

16 Temperature
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-1
Displaying temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-2
Patient connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-2
Temp Setup menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-3
Changing temperature label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-3
Combining different temperatures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-4
Testing temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-4
Changing temperature units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-4
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16-4

17 Airway gas
Safety precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-1
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-1
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-1
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-2
Module description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-2
Display of gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-3
Patient connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-4
CO2 setup menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-5
Selecting alarm sources. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-5
Calibrating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-6
Interfering gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-8
Unit conversions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-8
Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17-9

viii
Document no. 2050801-001

Index
Appendix A:

Index-1
Installation and checkout form, B40/B20

A-1

ix
Document no. 2050801-001

B40/B20 Patient Monitor

x
Document no. 2050801-001

1 Introduction

Introduction

1 Introduction
About this device
Intended purpose (Indications for use)
This device is a portable multiparameter unit to be used for monitoring and recording of, and to
generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal
patients in a hospital environment and during intra-hospital transport. The B40/B20 Patient
Monitor is intended for use under the direct supervision of a licensed health care practitioner.
The B40/B20 Patient Monior is not intended for use during MRI.
The B40/B20 Patient Monior monitors and displays oscillometric non-invasive blood pressure
(systolic, diastolic and mean arterial pressure), invasive blood pressure, end-tidal carbon
dioxide, heart/pulse rate, respiration rate, ECG (including arrhythmia and ST segment analysis),
temperature with a reusable or disposable electronic thermometer for continual monitoring
Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/
Myocardial/Core/Surface temperature, and functional oxygen saturation (SpO2) and pulse rate
via continuous monitoring, including monitoring during conditions of clinical patient motion or
low perfusion.

Intended audience
This manual is intended for clinical professionals. Clinical professionals are expected to have a
working knowledge of medical procedures, practices, and terminology, as required for
monitoring critically ill patients.

Responsibility of the manufacturer


GE Medical Systems Information Technologies, Inc. is responsible for the effects on safety,
reliability and performance of the equipment only if:

Assembly operations, extensions, readjustments, modifications, or repairs are carried out


by persons authorized by GE.

The electrical installation of the relevant room complies with the requirements of the
appropriate regulations.

The equipment is used in accordance with the instructions for use.


The equipment is installed, maintained and serviced in accordance with the instructions
provided in the related technical manuals.

Product availability
Some of the products mentioned in this manual may not be available in all countries. Please,
consult your local representative for the availability.

1-1

B40/B20 Patient Monitor

Printed copies of this manual


A paper copy of this manual will be provided upon request. Contact your local GE
representative and request the part number on the first page of the manual.

CE marking information
CE compliance
The B40/B20 Patient Monior bears CE Mark CE-0459 indicating its conformity with the
provisions of the Council Directive 93/42/EEC concerning medical devices and fulfills the
essential requirements of Annex I of this directive. The country of manufacture can be found on
the equipment labeling.
The product complies with the requirements of standard IEC 60601-1-2 Medical Electrical
Equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic
compatibility - Requirements and tests. The product meets Class A EMC limits in accordance
with CISPR 11.

Product Compliance
The B40/B20 Patient Monitor is classified in the following categories for compliance:

1-2

This equipment is suitable for connection to public mains as defined in CISPR 11.

The invasive blood pressure parameter conforms to the IEC 60601-2-34

This Monitor conforms to general safety standard for medical devices to IEC 60601-1.
This Monitor conforms to EMC safety standard to IEC 60601-1-2.
This Monitor conforms to usability safety standard for medical devices to IEC 60601-1-6.
Software is developed in accordance with IEC 60601-1-4.
The application of risk management analysis to medical device conforms to ISO 14971.
The SpO2 Parameter conforms to ISO 9919.
The TEMP parameter conforms to EN 12470-4.
The CO2 parameter conforms to ISO 21647.
This Monitor conforms to particular safety standard for multifunction patient monitoring
equipment to IEC 60601-2-49
The ECG parameter conforms to IEC 60601-2-27, ANSI/AAMI EC13
The NIBP parameter conforms to IEC 60601-2-30, EN 1060-1, EN 1060-3, ANSI/AAMI SP10
The alarm systems of the Monitor conform to IEC 60601-1-8

Introduction

Warranty
This Product is sold by GE Healthcare (GE) under the warranty set forth in the following
paragraphs. Such warranty is extended only with respect to the purchase of this Product
directly from GE or GEs Authorized Dealers as new merchandise and is extended to the Buyer
thereof, other than for the purpose of resale.
For a period of twelve (12) months from the date of original delivery to Buyer, this Product,
other than expandable parts, is warranted against functional defects in materials and
workmanship and to conform to the description of the Product contained in this manual and
accompanying labels and/or inserts, provided that the same is properly operated under the
conditions of normal use, that regular periodic maintenance and service is performed and that
the replacements and repairs are made in accordance with the instructions provided, using
genuine parts and performed by a trained person. The foregoing warranty shall not apply if the
Product has been repaired by anyone other than GE or otherwise than in accordance with
written instructions provided by GE, or altered by anyone other than GE, or if the Product has
been subject to abuse, misuse, negligence, or accident.
GEs sole and exclusive obligation and Buyers sole and exclusive remedy under the above
warranty is limited to repairing or replacing, free of charge, at GEs option, a Product, which is
telephonically reported to the nearest GE office or GEs Authorized Dealers office and which, if
so advised by GE, is thereafter returned with a statement of observed deficiency, not later than
seven (7) days after the expiration date of the applicable warranty, to the GE office or GEs
Authorized Dealers office during normal business hours, transportation charges prepaid, and
which, upon GEs examination, is found not to conform to the above warranty. GE shall not be
otherwise liable for any damages including but not limited to incidental damages,
consequential damages, or special damages.
There are no express or implied warranties, which extend beyond the warranty hereinabove
set forth. GE makes no warranty of merchantability or fitness for particular purpose with
respect to the product or parts thereof.

1-3

B40/B20 Patient Monitor

About this manual


This Users Reference Manual describes the functions offered by the B40/B20 patient monitor
running the software license VSP-A. As the monitor setup may vary, some menus, displays and
functions described may not be available in the monitor you are using.
This manual is an integral part of the product and describes its intended use. Keep it always
close to the equipment. Observance of the manual is a prerequisite for proper product
performance and correct operation and ensures patient and operator safety.
NOTE: Before using your monitor, please read the Users Guide or this manual thoroughly.
This Users Reference Manual gives you more specific information about the clinical and
technical aspects. Pay special attention to WARNING and CAUTION statements.
The new user of the monitor should begin with sections "Safety precautions" "System
description" and "Monitoring basic." These sections describe the system and the basic
operation of the monitor.
The measurement sections describe the measurement technique, setup and how to adjust
displays and menus for patient monitoring and special views.
Section "Monitor setup" gives instructions about setting up the system and making changes in
the default settings. Section "Cleaning and care" describes cleaning and daily maintenance
procedures.

Illustrations and names


All illustrations in this manual are only examples, and may not necessarily reflect your system
settings or data displayed in your system. If a particular selection is not available in your
system, the selection is shown grayed in the menu.
All names used in examples and illustrations are fictitious.

Conventions used in this manual


To help you find and interpret information easily, the manual uses consistent text formats:

Hard keys

Names of the hard keys on the Command board are written in the
following way: Others.

Menu items

Software terms that identify window parts or menu items are written
in bold italic: Monitor Setup

Menu access

Menu access is described from top to bottom. For example, the


selection of the Monitor Setup hard key, the Screen Setup menu
item and the Waveform Fields menu item would be shown as
Monitor Setup - Screen Setup - Waveform Fields.

Messages

Messages (alarm messages, informative messages) displayed on the


screen are written inside single quotes: Please wait.

References

When referring to different sections in this manual or to other


manuals, manual names and section names are enclosed in double
quotes: See section "Cleaning and care." Please refer to "Technical
Reference Manual: Installation."

The following symbols are also used to distinguish procedures:

ECG

1-4

Press the menu key described.

Introduction

Turn the Trim Knob.

Push the Trim Knob.

Related documentation

Installation, technical solutions and servicing: PROCARE Monitor B40/B20 Technical


Reference Manual

Options and selections of the software: PROCARE Monitor B40/B20 Default Configuration
Worksheet

Compatible supplies and accessories: PROCARE Monitor B40/B20 Supplies and


Accessories

Other devices closely related to the monitor:

iCentral and iCentral Client User's Reference Manual


CIC Pro Clinical Information Center Operator's Manual

1-5

B40/B20 Patient Monitor

Safety precautions
The following list contains all the general warnings and cautions you should know before
starting to use the system. Warnings and cautions specific to parts of the system can be found
in the relevant section.

Safety message signal words


Safety message signal words designate the severity of a potential hazard.
DANGER: Indicates a hazardous situation that, if not avoided, will result in death or serious
injury. No danger messages apply to this system.
WARNING: Indicates a hazardous situation that, if not avoided, could result in death or serious
injury.
CAUTION: Indicates a hazardous situation that, if not avoided, could result in minor or
moderate injury.
NOTE: Indicates a hazardous situation not related to personal injury that, if not avoided, could
result in property damage.

Warnings
The following warning safety messages apply to this monitoring system.

1-6

Read all the safety information before using the monitor for the first time.

Single-use devices and accessories are not designed to be reused. Reuse may cause a
risk of contamination and affect the measurement accuracy.

There are hazards associated with the reuse of single-use sample tubing and adapters

Do not sterilize or immerse the sensor or cable in liquid. Do not spray or soak the
connectors.

Do not route cables in a way that presents tripping hazard

Always check that power cord and plug are intact and undamaged.

DISCONNECTION FROM MAINS - When disconnecting the system from the power line,
remove the plug from the wall outlet first. Then you may disconnect the power cord from

Equipment is intended for clinical professionals.


For continued safe use of this equipment, it is necessary that the listed instructions are
followed. However, instructions listed in this manual in no way supersede established
medical practices concerning patient care.

Inspect the sensor for signs of physical damage. Discard a damaged sensor immediately.
Never repair a damaged sensor; never use a sensor repaired by others.

Route all cables away from patients throat to avoid possible strangulation.
Do not touch the patient, table, instruments, modules or the monitor during defibrillation.
To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
Use only approved accessories, including mounts, and defibrillator-proof cables and
invasive pressure transducers. For a list of approved accessories, see the supplies and
accessories list delivered with the monitor. Other cables, transducers and accessories
may cause a safety hazard, damage the equipment or system, result in increased
emissions or decreased immunity of the equipment or system or interfere with the
measurement.

Introduction

the device. If you do not observe this sequence, there is a risk of coming into contact with
line voltage by inserting metal objects, such as the pins of leadwires, into the sockets of
the power cord by mistake.

If liquid has accidentally entered the system or its parts, disconnect the power cord from
the power supply and have the equipment serviced by authorized service personnel.

If service message appears, discontinue use as soon as possible and have the device
repaired.

PROTECTED LEADWIRES - Only use protected leadwires and patient cables with this
monitor

When applying devices intracardially, never contact electrically conductive parts


connected to the heart (pressure transducers, metal tube connections and stopcocks,
guide wires, etc.).

Do not use the monitor in high electromagnetic fields (for example, during MRI).
Do not tilt the monitor to avoid liquid entering.
Do not touch the electrical connector located within the extension rack housing

Cautions
The following caution safety messages apply to this monitoring system.

U.S. Federal law restricts this device to sale by or on the order of a physician.

Dispose of equipment in compliance with instructions and regulations

SUPERVISED USE - This equipment is intended for use under the direct supervision of a
licensed health care practitioner.
DATA LOSS - No Communication may occur at central station due to RF interference
Reset the monitor if loss of monitoring data occurs.
Leave space for circulation of air to prevent the monitor from overheating.
Before connecting power, check voltage and frequency ratings of equipment.
System time changes will result in time differences between stored and realtime data.

Notes
The following notice safety messages apply to this monitoring system.

NOTICE - The warranty does not cover damages resulting from the use of accessories
and consumables from other manufacturers.

If the device has been transported or stored outside operating temperature allow it to
stabilize back to operating temperature range before applying power.

Medical electrical equipment needs special precautions regarding electromagnetic


compatibility, EMC, and needs to be installed and put into service according to the EMC
information provided in the "Technical Reference Manual" by qualified personnel.

Portable and mobile RF communications equipment can affect the medical electrical
equipment.

The equipment is suitable for use in the presence of electrosurgery. Please notice the
possible limitations in the parameter sections and in "Technical specification" on page 225.

Service and repairs are allowed for authorized service personnel only.

1-7

B40/B20 Patient Monitor

ESD precautionary procedures

To avoid electrostatic charges to build up, it is recommended to store, maintain and use
the equipment at a relative humidity of 30% or greater. Floors should be covered by ESD
dissipative carpets or similar. Non-synthetic clothing should be used when working with
the component.

To prevent applying a possible electrostatic discharge to the ESD sensitive parts of the
equipment, one should touch the metallic frame of the component or a large metal object
located close to the equipment. When working with the equipment and specifically when
the ESD sensitive parts of the equipment may be touched, a grounded wrist strap
intended for use with ESD sensitive equipment should be worn. Refer documentation
provided with the wrist straps for details of proper use.

ESD precautionary procedure training


It is recommended that all potential users receive an explanation of the ESD warning
symbol and training in ESD precautionary procedures.

The minimum content of an ESD precautionary procedure training should include an


introduction to the physics of electrostatic charge, the voltage levels that can occur in
normal practice and the damage that can be done to electronic components if they are
touched by an operator who is electrostatically charged. Further, an explanation should
be given of methods to prevent build-up of electrostatic charge and how and why to
discharge ones body to earth or to the frame of the equipment or bond oneself by
means of a wrist strap to the equipment or the earth prior to making a connection.

At the end of lifetime, the product and its accessories must be disposed of in compliance
with the guidelines regulating the disposal of such products. If you have questions
concerning disposal of the product, please contact your local GE representatives.

Disposal

1-8

2 System description

System description

2 System description
Safety precautions
Warnings

All system devices must be connected to the same power supply circuit

Verify compatibility of all system components prior to installation.

Do not use with iCentral software V5.0.2 and earlier or Mobile Care Server software earlier
of V5.2.

EXCESSIVE LEAKAGE CURRENT - Do not use a multiple socket outlet or extension cord.
INTERFACING OTHER EQUIPMENT - Devices may only be interconnected with each other
or to parts of the system when it has been determined by qualified biomedical personnel
that there is no danger to the patient, the operator, or the environment as a result. In
those instances where there is any element of doubt concerning the safety of the
connected devices, the user must contact the manufacturers concerned (or other
informed experts) for proper use. In all cases, safe and proper operation should be verified
with the applicable manufacturers instructions for use, and system standards IEC 606011-1/EN 60601-1-1 must be complied with.
Do not use multiple modules with identical measurements in the same monitor.
The parameter modules are not able to withstand unpacked drops from a height of 1 m
without damaging the module latches. If the device is dropped, please service the device
before taking it back into use.

2-1

B40/B20 Patient Monitor

Principles of functions
The B40/B20 monitor is a modular multiparameter patient monitor. The monitor is especially
designed for monitoring in PACU, ED, Wards, Step down units, Intra-hospital transport, Private
sectors, Alternative care in mature markets; PACU, ED, Wards, Step down units (low/mid critical
care), Intra-hospital transport, ICU & CCU, OR in settings where anesthesia gas monitoring is
not required. It can also be used during transportation within the hospital.
The modular design makes the system flexible and easy to upgrade.

Functional options
Basic

Optional

3/5 lead ECG


NIBP
RESP
GE, Nellcor or Masimo SpO2
Two channel IBP
Two channel Tempertures
CO2
Recorder
Unity or S/5 Network
1 or 2 batteries

2-2

System description

System introduction
The B40/B20 monitor system may consist of the elements shown below.

NOTE: Your system may not include all these components. Consult your local
representative for the available components.

Figure 2-1

B40/B20 patient monitor system

1.

B40/B20 monitor frame

2.

E-miniC module: This module measure CO2 parameter

3.

Software: VSP-A

4.

Other monitors in the network


NOTE: You cannot view other monitors on the B40/B20 monitor.

5.

CIC Pro central station (for Unity network): The CARESCAPETM Network MC establishes
communication and allows patient data to be sent to an optional CIC Pro central station.
See the CIC Pro Clinical Information Center Operators Manual for operating instructions.

6.

iCentral station (for S/5 network): The S/5 network establishes communication and
allows patient data to be sent to an optional iCentral station. See the iCentral Users
Reference Manual for operating instructions.

7.

Laser printer: This device may be connected to the monitor, network, or to a central
station on the network. The laser printer can print waveforms, alarm waveforms, numeric
trends, and reports. See the "Print and record" chapter for more details.

2-3

B40/B20 Patient Monitor

Components
The basic components of the B40/B20 are the monitor frame with hemo module.

Figure 2-2
(1)

B40/B20 monitor front panel

Transportation handle

(2)

Alarm light

(3)

The Trim knob

(4)

Command board keys

(5)

Battery compartment

(6)

Guide rail for GCX mounting

(7)

Mains power and battery LEDs

(8)

On/standby key

(9)

Hemo connectors

(10) E-miniC module


(11) Recorder module

Optional components
Optional components are:

2-4

Extension rack
Interchangeable E-miniC module
Multi I/O connector

System description

Rear panel connections

Figure 2-3

Rear panel connections

(1)

Receptacle for power cord

(2)

Serial port

(3)

Defibrillator connector

(4)

Nurse call connector

(5)

Network connector

(6)

Equipotential connector

(7)

Multi I/O connector

NOTE: The Multi I/O with ports 2,3,4 are optional parts for customer.

E-miniC module

Figure 2-4

E-miniC

(1)

Water trap

(2)

Sample gas inlet

(3)

Gas outlet

2-5

B40/B20 Patient Monitor

Keyboards
You can control monitoring through the keys on the Command Board. For more information,
see section "Monitoring basic."

Command Board keys

Figure 2-5
(1)

Command Board keys

ON/standby key

(2)

Mains power ON (lit) or OFF (dark): indicates mains

(3)

For admitting or discharging a patient; for selecting user modes

(4)

For setting up the monitor and for activating the menu

(5)

For printing and recording different trends and waveforms

(6)

For activating parameter specific menus.


NOTE: All modules do not measure all of these parameters.

(7)

For starting or stopping the NIBP auto cycling

(8)

For starting or stopping local recording


NOTE: Functional with the recorder only.

(9)

For silencing the alarms

(10) For returning the Normal Screen view to the screen


(11) For zeroing the invasive pressure channels
NOTE: Functional with the IBP only.
(12) For starting or stopping the NIBP manual cycling
(13) For activating the Alarms Setup menu
(14) For viewing trends and alarm history

2-6

System description

Batteries
The monitor can be run either on mains power or batteries. Battery operation is initiated when
the power cord is disconnected or when the mains power is lost during monitoring.

NOTE: Before using the monitor for the first time, charge the batteries to their full
capacity. Charging time is two hours per battery pack.
The B40/B20 has two lithium-ion batteries at most, located in the battery compartment. They
can be charged separately, and screen symbols and monitor frame LEDs indicate their
charging level and possible failure, see below. You can also check the battery status through
Monitor Setup - Battery Setup. The internal battery capacity is up to 4.5 hours with fully
charged batteries.
Monitor
Setup

NOTE: When the monitor is battery powered, the green battery LED is on. When the
monitor is mains powered, the green mains LED is on. When the monitor is battery
charging, the orange battery LED is on.
If you wish to have the battery charge visible at all times, select it in one of the digit fields:
Monitor Setup - Screen Setup - Digit Fields - Battery. You can now see how much charging
time is left for each battery separately both in numbers and as symbols, and the total charging
time in numbers.

2-7

B40/B20 Patient Monitor

Battery indicators
The B40/B20 messages, screen symbols and front panel LED indicators tell the user about the
status of the batteries. For screen symbols, see page 2-14. For LED indicators, consult the table
below and for messages, see section Troubleshooting.

Table 2-1

Battery indicators

Screen symbol

Explanation

Front panel battery LED


indicators

Monitor is battery
powered. Batteries are fully
charged; the size of the
green bar indicates the
charging level.

Orange dark

Monitor is battery
powered. Battery A is
empty, battery B charge is
ok.

Orange dark

Monitor is battery
powered. Battery A failure,
battery B is full.

Orange flashing

Green lit

Green lit

Green lit

NOTE: If both batteries fail, the green battery LED is dark.

Monitor is mains powered.


Battery A is being charged
(white bar), battery B is
already charged.
Monitor is mains powered.
no screen symbol

2-8

Orange lit
Green dark

Orange dark
Green dark

System description

Checking the battery charge when the monitor is turned off


When the monitor is turned off, you can check the
battery charging level by pressing the test button on the
battery as indicated in the drawing on the left.
The charging indicator bar (1) lights up and the number
of lit segments indicates the charging level: the more lit
segments, the higher the charging level.
1

Figure 2-6

Charging indicator on the battery

Conditioning a battery
Batteries should be conditioned regularly to maintain their useful life. Condition a battery every
six months, when its run time becomes noticeably shorter, or when the message Condition
Battery x appears on the screen. Conditioning a battery is best done on an external charger.
Please, refer to instructions provided with the charger.
If you do not have an external charger, see section "Cleaning and care":"Conditioning the
batteries".

Network
The B40/B20 monitor has been verified to be able to work in CARESCAPETM Network and S/5
network environments. Other network infrastructures are not supported.
The monitor have the capability to communicate with GE CARESCAPE pro CIC version 4.0.8,
4.1.1 and 5.1.0; communication with iCentral version 5.0.3 and 5.1.1.
The monitor can talk to Aware Gateway Server V1.6 in Unity network and talk to Mobile Care
Server V5.2 in S/5 network. The B40/B20 monitor work with S/5 Collect V4.0.
The monitor can talk to at most 4 CIC Central stations, 1 Aware Gateway Server and 1000 other
devices simultaneously in one CARESCAPE MC network. The monitor cant act as the Time
Master in Unity network.
The B40/B20 monitor does not support Patient Data Server; B40/B20 realtime patient data
can't be viewed on other monitors (e.g. Dash 3000/4000/5000, Solar 8000, B850, B650) except
unit name, bed name and alarm message.
The monitors information transfer to central station as following:

All B40/B20 available parameters: 3/5 lead ECG, NIBP, IBP, SpO2, Resp, Temp, CO2
Real-time multi-parameter waveforms
Graphic/tabular trends
Real-time Alarm, including Arrhythmia alarms: ASYSTOLE, VFIB/VTAC, VTACH
Patient admit/discharge & patient name update
Multiple parameters alarm setting

Also in Unity network, B40/B20 support to set up parameters and waveforms on CIC remotely.

2-9

B40/B20 Patient Monitor

Symbols
On the rear panel this symbol indicates the following warnings and cautions:
-

Electric shock hazard. Do not open the cover or the back. Refer servicing to qualified
personnel.

For continued protection against fire hazard, replace the fuse only with one of the same type
and rating.

Disconnect from the power supply before servicing.

Do not touch the monitor during defibrillation.

Do not use the monitor without manufacturer approved mounting attached.

Electrostatic sensitive device. Connections should not be made to this device unless ESD
precautionary procedures are followed.
Type BF (IEC 60601-1) protection against electric shock. Isolated (floating) applied part suitable for
intentional external and internal application to the patient, excluding direct cardiac application.
Type BF (IEC 60601-1) defibrillator-proof protection against electric shock. Isolated (floating)
applied part suitable for intentional external and internal application to the patient, excluding
direct cardiac application.
Type CF (IEC 60601-1) protection against electric shock. Isolated (floating) applied part suitable for
intentional external and internal application to the patient, including direct cardiac application.
Type CF (IEC 60601-1) defibrillator-proof protection against electric shock. Isolated (floating)
applied part suitable for intentional external and internal application to the patient including
direct cardiac application.
When displayed in the upper left corner of the screen, indicates that the alarms are audio off.
When displayed in the menu or digit fields, indicates that the alarm source has been turned off or
alarm does not meet the alarm-specific activation criteria.
In the front panel: battery

Equipotentiality. Monitor can be connected to potential equalization conductor.

Alternating current
Bell cancel. Audio pause.

Home. Return to the main display.


Standby or power indicator.
Fuse. Replace the fuse only with one of the same type and rating

2-10

System description

Gas inlet.

Gas outlet.
IP21
SN,S/N

Degree of ingress protection.


Serial number
Date of manufacture. This symbol indicates the date of manufacture of this device. The four digits
identify the year.
Maunfacturer: This symbol indicates the name and the address of the manufacturer.

European authorized representative.


European Union Declaration of Conformity.

Rx Only U.S.

Prescriptive Device. USA only. For use by or on the order of a Physician, or persons licensed by
state law.
Fragile. Handle with care.

Keep dry. Protect from rain.

This way up.

Storage temperature

This symbol depicts the transportation and storage atmospheric pressure range of 700 to 1060
hPa

Recycled materials or may be recycled.

2-11

B40/B20 Patient Monitor

This symbol indicates that the waste of electrical and electronic equipment must not be disposed
as unsorted municipal waste and must be collected separately. Please, contact an authorized
representative of the manufacturer for information concerning the decommissioning of your
equipment.
The separate collection symbol is affixed to a battery, or its packaging, to advise you that the
battery must be recycled or disposed of in accordance with local or country laws. To minimize
potential effects on the environment and human health, it is important that all marked batteries
that you remove from the product are properly recycled or disposed. For information on how the
battery may be safely removed from the device, please consult the service manual or equipment
instructions. Information on the potential effects on the environment and human health of the
substances used in batteries is available at this url: http://www.gehealthcare.com/euen/weeerecycling/index.html
This product consists of devices that may contain mercury, which must be recycled or disposed
of in accordance with local, state, or country laws. (Within this system, the backlight lamps in the
monitor display contain

Battery operation and remaining capacity. The height of the green bar indicates the charging
level.
Battery (A) charging (white bar)

Battery (A) failure

Both batteries failed

Battery (A) missing


Submenu. Selecting a menu item with this symbol opens a new menu.
The monitor is connected to Network.
A blinking heart next to the heart rate or pulse rate value indicates the beats detected.
A lung next to the respiration rate value indicates that respiration rate is calculated from the
impedance respiration measurement.

Do not reuse.

Use by. Indicates the last use day.

2-12

System description

Date of manufacture

Do not immerse the sensor in liquids.

C
3ZG9

Medical Equipment
With respect to electrical shock, fire and mechanical hazards only in accordance with IEC 60601US 1, UL 60601-1; IEC 60601-2-27; IEC 60601-2-30; IEC 60601-2-34; IEC 60601-2-49; CAN/CSA C22.2
No. 601.1

2-13

B40/B20 Patient Monitor

Abbreviations

2-14

/min

beats per minute, breaths per minute

Celsius degree

Fahrenheit degree

microgram

arm (describing location)

alveolar

arterial

a/AO2

arterio-alveolar PO2 ratio

AaDO2

alveolo-arterial oxygen difference

AA

anesthetic agent

AAMI

Association for the Advancement of Medical Instrumentation

ABG

arterial blood gases

ABP

arterial pressure

ADU

Anesthesia Delivery Unit

AEP

auditory evoked potential

AirW

airway temperature

Alpha, Al

alpha frequency band

AM

Anesthesia Monitor

Amp

amplitude

Ant

anterior

APN

apnea

Arrh.

arrhythmia

Art

arterial pressure

ASY

asystole

ATMP

atmospheric pressure

ATPD

atmospheric/ambient temperature and pressure, dry gas

ATPS

ambient temperature and pressure, saturated gas

AV

atrioventricular

aVF

left foot augmented lead

avg

average

aVL

left arm augmented lead

aVR

right arm augmented lead

aw

airway

Axil

axillatory temperature

BAEP

brainstem auditory evoked potential

Bal

balance gas

bar

1 atmosphere

Beta, Be

beta frequency band

Bigem.

bigeminy

BIS

bispectral index

Blad

bladder temperature

Blood

blood temperature (C.O. measurement)

System description

Body

body temperature

BP

blood pressure

Brady

bradycardia

BSA

body surface area

BSR

burst suppression ratio

B-to-B

beat-to-beat

BTPS

body temperature and pressure, saturated gas

calculated/derived value

chest

C(a-v)O2

arteriovenous oxygen content difference

C.C.O.

continuous cardiac output

CFI

cardiac function index

C.I.

cardiac index

C.O.

cardiac output

cal.

calibration

Calc

calculated/derived value

Calcs

calculations

CAM

Compact Anesthesia Monitor

CaO2

arterial oxygen content

Casc.

cascaded (ECG)

cc

cubic centimeter

CCCM

Compact Critical Care Monitor

CCM

Critical Care Monitor

CcO2

capillary oxygen content

CCU

cardiac (coronary) care unit

CEL

Celsius degree

CFI

cardiac function index

CISPR

International Special Committee on Radio Interference

cmH2O

centimeter of water

CMRR

common mode rejection ratio

CO

carbon monoxide

CO2

carbon dioxide

COHb

carboxyhemoglobin

Compl

compliance

Cont.

continuous

Contrl

controlled ventilation

Core

core temperature

Count

count of responses

CPB

cardiopulmonary bypass

CPP

cerebral perfusion pressure

CSA

compressed spectral array

CT

computer tomography

CvO2

(mixed) venous oxygen content

2-15

B40/B20 Patient Monitor

2-16

CVP

central venous pressure

day

dB

decibel

DBS

double burst stimulation (NMT)

DEL

delete

Delta, De

delta frequency band

depr.

depression

Des

desflurane

Dia

diastolic pressure

Diagn

diagnostic (ECG filter)

DIFF

difference

DIS

S/5 Device Interfacing Solution

DO2

oxygen delivery

DO2I

oxygen delivery index

DSC

digital signal converter

dyn

dynamic

estimated

ECG

electrocardiogram

ECG1

first ECG waveform (top)

ECG1/r

real-time ECG

ECG2

second ECG waveform

ECG3

third ECG waveform

ED

emergency department

EDV

end-diastolic volume

EDVI

end-diastolic volume index

EE

energy expenditure (kcal/24h)

EEG

electroencephalogram

EEG1

first EEG waveform

EEG2

second EEG waveform

EEG3

third EEG waveform

EEG4

fourth EEG waveform

EEMG

evoked electromyogram

EEtot

total energy expenditure

elect

electrode

elev.

elevation

EMC

electromagnetic compatibility

EMG

electromyogram

Enf

enflurane

Entr

entropy

EP

evoked potential

ESD

electrostatic discharge

Eso

esophageal temperature

System description

ESV

end-systolic volume

ESVI

end-systolic volume index

ET, Et

end-tidal concentration

EtAA

end-tidal anesthetic agent

EtBal

end-tidal balance gas

EtCO2

end-tidal carbon dioxide

EtN2O

end-tidal nitrous oxide

EtO2

end-tidal oxygen

ET-tube, ETT

endotracheal tube

EVLW

extravascular lung water

EVLWI

extravascular lung water index

Exp

expiratory

foot (describing location)

FAH

Fahrenheit degree

FEMG

frontal electromyogram

FFT

fast Fourier transform

FI, Fi

fraction of inspired gas

FiAA

fraction of inspired anesthetic agent

Fib

fibrillation

FiBal

fraction of inspired balance gas

FiCO2

fraction of inspired carbon dioxide

FiN2

fraction of inspired N2

FiN2O

fraction of inspired nitrous oxide

FiO2

fraction of inspired oxygen

Flow

airway gas flow

Freq.

frequent

ft

foot, feet

FVloop

flow volume loop

Gauss

gram

GEDI

global enddiastolic volume index

GEDV

global enddiastolic volume

GEF

global ejection fraction

Graph.

graphical

hour

hand (describing location)

Hal

halothane

Hb

hemoglobin

Hbtot

total hemoglobin

HCO3-

bicarbonate

Hemo

hemodynamic

2-17

B40/B20 Patient Monitor

2-18

Hemo Calcs

hemodynamic calculations

HHb

reduced hemoglobin

HME

heat and moisture exchanger

HMEF

heat and moisture exchanger with filter

hPa

hectopascal

HR

heart rate

HRdiff

heart rate difference

ht

height

HW

hardware

Hz

hertz

IEC

International Electrotechnical Comission

I:E

inspiratory-expiratory ratio

IABP

intra-aortic balloon pump

IC

inspiratory capacity

ICP

intracranial pressure

ICU

intensive care unit

ID

identification

Imped.

impedance; impedance respiration

in

inch

Inf

inferior

Infl.

inflation (limit)

Insp

inspiratory

Inv.

invasive

Inv. BP

invasive blood pressure

Irreg.

irregular

Iso

isoflurane

ISO

International Standards Organisation

ISM

Industrial, Scientific and Medical

ITBV

intrathoracic blood volume

IVR

idioventricular rhythm

joule

kelvin

kcal

kilocalorie

kJ

kilojoule

kPa

kilopascal

leg (describing location)

left (describing location)

L, l

liter

l/min

liters/minute

Lab

laboratory

LAN

local area network

LAP

left atrial pressure

System description

Lat

lateral

lb

pound

LCD

liquid crystal display

LCW

left cardiac work

LED

light emitting diode

LVEDP

left ventricular end diastolic pressure

LVEDV

left ventricular end diastolic volume

LVSW

left ventricular stroke work

LVSWI

left ventricular stroke work index

MAC

minimum alveolar concentration

Max

maximum

mbar

millibar

mcg

microgram

Mean

mean blood pressure

mEq

milliequivalent

MetHb

methemoglobin

MF

median frequency

mg

milligram

min

minute

Min

minimum

ml

milliliter

MLAEP

middle-latency auditory evoked potential

mmHg

millimeters of mercury

mol

mole

Monit

monitoring (ECG filter)

MRI

magnetic resonance imaging

Mult.

multiple

Multif. PVCs

multifocal PVCs

MV

minute volume

MVexp

expired minute volume (l/min)

MVexp(BTPS)

expired minute volume in BTPS conditions

MVexp(STPD)

expired minute volume in STPD conditions

MVinsp

inspired minute volume (l/min)

MVspont

spontaneous minute volume

Myo

myocardiac temperature

neutral

N2

nitrogen

N2O

nitrous oxide

Na

sodium

Naso

nasopharyngeal temperature

neo

neonate

Net

network

2-19

B40/B20 Patient Monitor

2-20

NIBP

non-invasive blood pressure

Ni-Cd

nickel-cadmium

Ni-MH

nickel-metal hydride

NMT

neuromuscular transmission

NO

nitric oxide

NTPD

normal temperature and pressure, dry gas

Num.

numerical

O2

oxygen

O2ER

oxygen extraction ratio

O2Hb

oxygenated hemoglobin

OR

operation room

Oxy

oxygenation

Oxy Calcs

oxygenation calculations

partial pressure

pressure

P(BTPS)

pressure in BTPS conditions

P(g-a)CO2

difference between gastrointestinal carbon dioxide and arterial blood


carbon dioxide concentration

P(g-ET)CO2

difference between gastrointestinal carbon dioxide and end tidal


carbon dioxide concentration

P(STPD)

pressure in STPD conditions

P1, P2

invasive pressure channel identification on module

PA

pulmonary artery

Pa

Pascal (unit of pressure)

Paced

paced beats

PaCO2

partial pressure of carbon dioxide in the arteries

PAO2

partial pressure of oxygen in the alveoli

PaO2

partial pressure of oxygen in the arteries

PAOP

pulmonary artery occlusion pressure

PA

pulmonary arterial pressure

Paw

airway pressure

Pbaro

barometric pressure

PCWP

pulmonary capillary wedge pressure

PE

polyethylene

pedi

pediatric

PEEP

positive end-expiratory pressure

PEEPe

extrinsic positive end expiratory pressure

PEEPe+i

total positive end expiratory pressure (ICU)

PEEPe+PEEPi

total positive end expiratory pressure (ICU)

PEEPi

intrinsic positive end expiratory pressure

PEEPtot

total positive end expiratory pressure (anesthesia)

PgCO2

gastrointestinal carbon dioxide concentration

System description

pH

pH

pHa

arterial pH

pHi

intramucosal pH

pHv

(mixed) venous pH

PIC

patient interface cable

Pleth

plethysmographic pulse waveform

PM

pacemaker

PM non-capt.

pacemaker non-capturing

PM non-funct.

pacemaker non-functioning

Pmax

maximum pressure

Pmean

mean pressure

Pmin

minimum pressure

Ppeak

peak pressure

Pplat

plateau (pause) pressure

PR

pulse rate

Prev.

previous

psi

pounds per square per inch

pt

patient

PTC

post tetanic count (NMT)

pts

patients

PVC

polyvinylchloride

PVC

premature ventricular contraction

PVloop

pressure volume loop

PvO2

partial pressure of oxygen in (mixed) venous blood

PVR

pulmonary vascular resistance

PVRI

pulmonary vascular resistance index

Px

standard pressure label, x being 1, 2, 3, 4, 5, or 6

QRS

QRS complex

Qs/Qt

venous admixture

right (describing location)

RAP

right atrial pressure

Raw

airway resistance

RCW

right cardiac work

RCWI

right cardiac work index

RE

Response Entropy

Rect

rectal temperature

REF

right ventricular ejection fraction

ref.

reference

Resp

respiration rate (total) (set)

Resp Rate

respiration rate (total) (measured)

RF

radio frequency

RMS

average (root mean square) power

2-21

B40/B20 Patient Monitor

2-22

Room

room temperature

RQ

respiratory quotient

RR

respiration rate (total) (measured)

rtm

rhythm

RV

residual volume

RVEDV

right ventricular end-diastolic volume

RVESV

right ventricular end-systolic volume

RVP

right ventricular pressure

RVSW

right ventricular stroke work

RVSWI

right ventricular stroke work index

second

SA

sinoatrial

SaO2

arterial oxygen saturation

S.A.R.

specific absorption rate

SD

standard deviation

SE

State Entropy

SEF

spectral edge frequency

SEMG

spontaneous electromyogram

Sev

sevoflurane

SI

stroke index

Skin

skin temperature

SN, S/N

serial number

Spiro

patient spirometry

SpO2

oxygen saturation

Spont

spontaneous breathing

SQI

signal quality index

SR

suppression ratio

SR

sinus rhythm

SSEP

somatosensory evoked potentials

ST

single twitch (NMT)

ST

ST segment of electrocardiograph

stat

static

STAT

continuous NIBP cuff inflation for five minutes

STBY

standby

Stfilt

ST filter (ECG)

STPD

standard temperature and pressure, dry gas

Surf

surface temperature

SV

stroke volume

SVC

supraventricular contraction

SVI

stroke volume index

SvO2

(mixed) venous oxygen saturation

SVR

systemic vascular resistance

SVRI

systemic vascular resistance index

System description

SW

software

SVV

stroke volume variation

Sys

systolic pressure

time (min)

temperature

tesla

T(BTPS)

temperature in BTPS conditions

T1%

first stimulus as % of the reference value (NMT)

T1, T2

temperature channel identification on module

Tab.

tabular

Tachy

tachycardia

Tbl, Tblood

blood temperature

Tcorr

temperature correction

Temp

temperature

Theta, Th

theta frequency band

Tinj

injectate temperature

TOF

train of four (NMT)

TOF%

ratio of the 4th to the 1st response (NMT)

Trigem.

trigeminy

TV

tidal volume

TVexp

expired tidal volume (ml)

TVinsp

inspired tidal volume (ml)

Tx

temperature label, x being 1, 2, 3, or 4 or one of the other label choices

Tymp

tympanic temperature

venous

ventricular

volume

V/Q

ventilation/perfusion ratio

V0.5

volume expired during the first 0.5 seconds

V1.0

volume expired during the first second

VA

alveolar ventilation

VC

vital capacity

VCO2

carbon dioxide production

Vd

dead space

Vd/Vt

dead space ventilation

Vent Calcs

ventilation calculations

VFib

ventricular fibrillation

VO2

oxygen consumption

VO2Calc

calculated oxygen consumption*

VO2CalcI

calculated oxygen consumption index*

VO2I

oxygen consumption index

Vol

volume

2-23

B40/B20 Patient Monitor

V Run

ventricular run

V Tachy

ventricular tachycardia

WLAN

wireless local area network

wt

weight

extreme

yr

year

yrs

years

* with Fick equation

2-24

System description

Technical specification
WARNING

Operation of the monitor outside the specified values may cause inaccurate
results.
NOTE: Information in this section can be especially useful to clinicians.

General Specifications
Genenral specifications
Size
Monitor
Without extension
modules

3125 mm (H) * 3125 mm (W) * 1585 mm (D)

With extension modules

3125 mm (H) * 3525 mm (W) * 1785 mm (D)

Weight
With extension module,
recorder and CO2

6 kg

Environment
Operating temperature

Normal operation: +5 to +40C (41 to 104F)


Charging batteries: +5 to +35C (41 to 95F)

Storage and transport

-20 to +60C (-4 to 140F)

temperature
Operating humidity

20 to 90% noncondensing

Storage and transport


humidity

20 to 90% noncondensing

Operating atmospheric
pressure

700 to 1060 hPa

Storage and transport


atmospheric pressure

700 to 1060 hPa

(525 to 800 mmHg)


(525 to 800 mmHg)

Electrical
AC input voltage

100 to 240 V

AC input frequency

50/60 Hz

AC input power

150 VA

Power supply

Internal battery or AC power

Power cord type

cord connector IEC/EN 60320-1/C13

For USA, difference type of plugs should be used for connection to the alternate
voltage 13 A 125 V or 6 A 250 V.
Fuse

250 V, 2.5 Ah

Battery

Exchangeable lithium-ion, 2 pcs max.

2-25

B40/B20 Patient Monitor

Battery life

300 cycles minimum to 50% capacity

Battery information

model SM 201-6; 11.1 V, 3.52 Ah

Charging time

2 hours per battery pack

Operation time

Up to 4.5 hours

Recorder
Power comsumption

Standby: < 1.2 W


Printing: < 10 W

Recorder type

Thermal array

Resolution

Vertical

8 dots/mm (200 dots/inch)


in non-waveform mode

Horizontal 24 dots/mm (600 dots/inch)


mimimum in waveform mode
Paper width

50 mm, printing width 48 mm

Waveforms

Selectable 1, 2, or 3 waveforms

Print speed

1, 6.25, 12.5, 25 mm/s

Defibrillator synchronization connector


Synchronization pulse (Pin 5)
Pulse width:

10 ms positive pulse

Delay:

< 35 ms (R-wave peak to leading edge of pulse)

Amplitude:

CMOS compatible
3.5 V min. at 1 mA sourcing
0.5 V max. at 5 mA sinking

2-26

Output impedance:

50

Current limit:

10 mA

System description

Parameters specifications
ECG specifications
Leads available

3-lead configuration: I, II, III


5-lead configuration: I, II, III, aVR, aVL, aVF and VA

QRS detection range

0.5 to 5mV

QRS detection width (Q to


S)

40 to 120 ms

Defibrillation protection

5000 V, 360 J

Recovery time

<5 s

Input impedance

Common mode > 10 M @ 50/60 Hz


Differential > 2.5 M from 0.67 to 40 Hz

Common mode rejection

90 dB minimum at 50 Hz

Tall T wave rejection

>1.4 mV

ECG leads off detection

Active patient electrode: <30 nA


Reference electrode: <120 nA

Filter modes
60 Hz
Monitoring filter

0.5 to 40 Hz

ST filter

0.05 to 40 Hz

Dagnostic filter

0.05 to 150 Hz

50 Hz
Monitoring filter

0.5 to 35 Hz

ST filter

0.05 to 35 Hz

Dagnostic filter

0.05 to 150 Hz

Heart rate
Measurement range

30 to 300 bpm

Measurement accuracy

5 % or 5 bpm, whichever is greater

resolution

1 bpm

Heart rate response time (ANSI/AAMI EC13-2002 Section 4.1.2.1f)


Step increase from 80 to
120 bpm

average 6.9 s (6.5 to 7.5 s)

Step decrease from 80 to


40 bpm

average 8.2 s, (7.6 to 10.0 s)

The heart rate calculation operates with irregular rhythms of ANSI/AAMI EC132002 Section 4.1.2.1e, the heart rate after a 20 second stabilization period is:
Figure 3a

80 bpm

Figure 3b

59 bpm

Figure 3c

122 bpm

2-27

B40/B20 Patient Monitor

Figure 3d

117 bpm

Heart rate averaging computation (ANSI/AAMI EC13-2002 Section 4.1.2.1d):


Average of 10 second median values
The average time and time range ( ) to alarm (VFib or VTachy) for
tachycardia waveform are as follows (ANSI/AAMI EC13-2002 Section 4.1.2.1g)
Figure 4a halved
amplitude:

9.9 s (8.4 to 11.5 s)

Figure 4a

7.1 s (5.8 to 8.2 s)

normal amplitude:
Figure 4a

4.4 s (4.2 to 4.6 s)

doubled amplitude:
Figure 4b halved
amplitude:

7.0 s (6.1 to 7.5 s)

Figure 4b

5.8 s (4.5 to 7.4 s)

normal amplitude:
Figure 4b

6.1 s (5.1 to 7.0 s)

doubled amplitude:

ST
ST numeric range

-9 to 9 mm (-0.9 to 0.9 mV)

ST numeric accuracy

0.2 mm or 10%, whichever is greater (within the


range of -8 to 8 mm)

ST numeric resolution

0.1 mm

Pacemaker detection
Input voltage range

2 to 700 mV

Input pulse width

0.5 to 2 ms

Input overshoot

Specified for both Method A and Method B required


in ANSI/AAMI EC13-2002 Section 4.1.4.2

Pacer pulse rejection of


fast ECG signals

2.0 V/s (according to the test defined in ANSI/AAMI


EC13-2002 Section 4.1.4.3)

NOTE: Pacemaker detector may not operate correctly during the use of high-frequency (HF)
surgical equipment. The disturbances of HF surgical equipment typically cause false positive
pacer detection.

Direct cardiac application:


The display area reserved for the ECG measurement in the monitoring system screen may not
be adequate for displaying the complete ECG amplitude when measuring ECG direct from the
surface of the heart. Clipping of the signal can be reduced by adjusting the size of the signal on
the screen (for example, from the default 1.0 to 0.2) in the ECG menu.

2-28

System description

Impedance respiration specifications


Measurement range

4 to 120 resp/min

Measurement accuracy

5% or 5 resp/min, whichever is greater

Nomalized respiration
sensing current

<5.0 A

Impedance respiration
carrier frequency

31.25 kHz

GE SpO2 specifications
Measurement and display
range

0 to 100%

Calibrated against functional oxygen saturation.


Measurement accuracy

100 to 70%

2 digits

69 to 0%

unspecified

Display resolution

1 digit (1% of SpO2)

Display averaging

5 to 20 seconds

Wavelength of SpO2 probe Infrared LED


LEDs:
RedLED

660 nm

Maximum energy of SpO2


probe LEDs:

Infrared LED

42 J/pulse

Red LED

62 J/pulse

Artifact rejection

SpO2 values 3% and PR values 5%,

940 nm

when trapping with a finger for 10s


Pulse rate
Measurement and display
range

30 to 250 bpm

Display resolution

1 bpm

Measurement accuracy

5% or 5 bpm, whichever is greater

2-29

B40/B20 Patient Monitor

Nellcor SpO2 specifications


Measurement and display
range

1 to 100%

Measurement accuracy

Adult

100 to 70% 2 digits

Neo

100 to 70% 3 digits

Low perfusion 100 to 70% 2 digits


Display resolution

1% of SpO2

Display averaging

2 to 7 seconds

Pulse rate
Measurement and display
range

20 to 250 bpm

Display resolution

1 bpm

Measurement accuracy

3 digits

Sensor Light Source


Wavelength

Infrared: 890 nm (nominal)


Red: 660 nm (nominal)

Power Dissipation

Infrared: 22.5 mW (max)


Red: 30 mW (max)

Masimo SpO2 specifications


Measurement and display
range

1 to 100%

Measurement accuracy
Without motion

Adult/Pediatric 100 to 70% 2 digits


Neonate

100 to 70% 3 digits

With motion

Adult/Ped/Neo 100 to 70% 3 digits

Low perfusion

100 to 70% 2 digits


0~69% unspecified

Display resolution

1% of SpO2

Display averaging

2 to 16 seconds

Pulse rate

2-30

Measurement and display


range

25 to 240 bpm

Display resolution

1 bpm

Measurement accuracy

Without motion

3 bpm

With motion

5 bpm

System description

Sensor Light Source


Wavelength

Infrared: 905 nm (nominal)


Red: 660 nm (nominal)

Power Dissipation

Infrared: 22.5 mW (max)


Red: 27.5 mW (max)

NIBP
Measurement technique

Oscillometric with step deflation

Supported modes

Manual, automatic and stat

Measurement time

Adult/Pediatric inflate duration time less than 120 s


Neonate cycle time less than 85 s

Measurement ranges
Systolic

Adult/Pediatric: 30 to 290 mmHg


Neonate: 30 to 140 mmHg

MAP

Adult/Pediatric: 20 to 260 mmHg


Neonate: 20 to 125 mmHg

Diastolic

Adult/Pediatric: 10 to 220 mmHg


Neonate: 10 to 110 mmHg

Accuracy

According to AAMI SP10-2002 4.4.5.2 B, accuracy of


NIBP parameter was validated against the intraarterial method 1.

Default initial inflation


pressure

Adult/Pediatric: 135 15 mmHg

Over pressure allowed by


independent safety
controller

Adult/Pediatric: 300 to 330 mmHg

Neonate: 100 15 mmHg


Neonate: 150 to 165 mmHg

1. Blood pressure measurements determined with this device are equivalent to those
obtained by an intra-arterial blood pressure measurement device, within the limits
prescribed by the American National Standard, Manual, electronic, or automated
sphygmomanometers

2-31

B40/B20 Patient Monitor

Invasive blood pressure


Measurement range

-40 to 320 mmHg (-5.3 to 42.7 kPa)

Measurement accuracy

5% or 2 mmHg, whichever is greater

Frequency response

4 to 22 Hz

Transducer sensitivity

5 V/V/mmHg

Pulse rate
Range

30 to 250 bpm

Accuracy

5% or 5 bpm, whichever is greater

Display resolution

1 bpm

Zero adjustment range

150 mmHg

Temperature
Measurement units

Fahrenheit (F)
Celsius (C)

Measurement range

10 to 45C (50 to 113F)

Measurement accuracy

0.1C without temperature sensor

Display resolution

0.1C at 25 to 45 C with reusable probes

Probe types supported

Use only GE Healthcare recommend temperature


YSI probes.

Temperature self-check

At start-up and then every 10 minutes

Probe type time response


Reusable skin temperature 3 s
probe:

2-32

Reusable adult central


temperature probe:

6s

Reusable pediatric central


temperature probe:

4s

Disposable skin
temperature probe:

3 to 6 s

Disposable central
temperature probe, 12F:

5 to 8 s

Disposable central
temperature probe, 9F:

5 to 8 s

System description

Airway gases
Sampling rate

15025 ml/min (sampling line 2 to


3 m, normal conditions)

Maximum sampling

6m

line length:
Sampling delay

2.1 s typical with a 3-m sampling line

Total system response


time

2.4 seconds typical with a 3-m

Warm-up time

1 minute for operation

sampling line, including sampling delay and rise time


30 minutes for full specification

Autozeroing interval

4, 15, 30 and 60 minutes after start-up, then every


60 minutes

NOTE: The drift of measurement accuracy will be cleared after autozeroing.

Carbon dioxide (CO2)


Measurement range

0 to 20 vol%

Accuracy
0 to 15 vol%

(0.2 vol% + 2% of reading)

15 to 20 vol%

(0.7 vol% + 2% of reading)

Resolution

0.1%

Measurement rise time

< 300 ms with nominal flow

Valid for respiration rate < 40 breaths/min at I:E ratio of 1:1. (Relative error is
typically 10% for respiration rate 80 breaths/min at I:E ratio of 1:1.) The accuracy
is specified in simulated ventilation. With higher respiration rates and with
varying ventilation methods the specifications may not be met.

Respiration rate
Breath detection

1% change in CO2 level

Measurement range

4 to 80 breaths/min

Accuracy

1 breaths/min in the range 4 to 20


breaths/min 5% in the range 20 to 80 breaths/min

2-33

B40/B20 Patient Monitor

Resolution

1 breaths/min

Non-disturbing gases are


those with a maximum
effect on the CO2 reading
at 5.0 vol% < 0.2 vol%. The
effect is valid for specific
concentrations shown in
parentheses of the nondisturbing gas:

Ethanol C2H5OH (<0.3%)

Disturbing gases and their


effect on the CO2 reading
at 5.0 vol% CO2 are shown
below. Errors listed reflect
the effect of specific
concentrations (shown in
parentheses) of an
individual disturbing gas
and should be combined
when estimating the effect
of gas mixtures:

Acetone (<0.1%)
Methane CH4(<0.2%)
Nitrogen N2 (0 to 100%)
Water vapor (0 to 100%)
Dichlorofluoromethane (<1%)
Tetrafluoroethane (<1%)
Halothane (4%) increases < 0.3 vol%
Isoflurane(5%) increases < 0.4 vol%
Enflurane(5%) increases < 0.4 vol%
Desflurane(24%) increases < 1.2 vol%
Sevoflurane(6%) increases < 0.4 vol%
If O2 compensation is not activated:
O2 (40 to 95%) decreases < 0.3 vol%
If O2 compensation is activated:
O2 (40 to 95%) error < 0.15 vol%
If N2O compensation is not activated:
N2O (40%) increases < 0.4 vol%
If N2O compensation is activated:
N2O (40 to 80%) error < 0.3 vol%

NOTE: CO2 measurement is intended for patients weighing over 5 kg (11 lb).

2-34

3 Installation

Installation

3 Installation
Safety precautions
Warnings

The monitor or its components should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the monitor and its components
should be observed to verify normal operation in the configuration in which it will be
used.

After transferring or reinstalling the monitor, always check that it is properly connected
and all parts are securely attached. Pay special attention to this in case of stacked
mounting.

Do not use without manufacturer approved mounting.

The monitor display is fragile. Ensure that it is not placed near a heat source or exposed
to mechanical shocks, pressure, moisture or direct sunlight.

Never install equipment above the patient.

Cautions

Unpacking instructions
1.

Confirm that the packing box is undamaged. If the box is damaged, contact the shipper.

2.

Open the top of the box and carefully unpack all components.

3.

Confirm that all components are undamaged. If any of the components is damaged,
contact the shipper.

4.

Confirm that all components are included. If any of the components is missing, contact
your GE Healthcare distributor.

Choosing location
Consider the following aspects:

Lighting
Space
Connections
Electromagnetic and radio frequency interference. For details see ElectroMagnetic
Compatibility section in Users Guide
Environment

Mounting the monitor


Mounting of monitor to the Wall Mount, Rollstand, Wall Mount with standard arm or Counter
Top Mount is described in a separate instruction sheet delivered with each mount.

3-1

B40/B20 Patient Monitor

Connection to mains
Connect the power cord to the mains power inlet at the back of the monitor and to the wall
socket.
NOTE: Before taking the monitor into use for the first time, the batteries should be fully
charged. Keep the monitor connected to the mains until the Battery charging symbol
disappears, may take up to 5 hours if the batteries are fully discharged.

Battery charging symbol

WARNING

The power cord may only be connected to a three-wire, grounded, hospital


grade receptacle.

Install the network


NOTE: To install network, please ask the GE service personnel to support.

Install the batteries


1. Open the lid of the battery compartment by
the side of the monitor. Move the latch up or
down.
2. Put in the new battery. Make sure that the
charging indicator is facing to the back side of
the monitor, then push the battery in all the
way, move the latch and close the lid. Power
on the monitor, check the monitor indicators,
see above.

3-2

Installation

Inserting and removing the E-miniC module

To use the E-miniC module, your monitor need pre-configure the extension rack from
manufacture.

To insert module:
1.

Align the module with the insertion guides

2.

Push the module into the monitor frame until it clicks and stops.

3.

Pull the module outwards to insure the module is firmly seated.

To remove module:

WARNING

1.

Pressing the release latch, on the bottom of the module.

2.

Grasp the module firmly and pull out of the Frame. Make sure not to drop it when it
comes out.

When detaching modules, be careful not to drop them. Always support with
one hand while pulling out with the other.

Visual indicators
Function

Specification

Explanation

External power supply

Green LED

Indicates when monitor is powered from


mains

Battery operation

Green LED

Indicates when monitor is powered from


internal batteries

Battery condition

Orange LED

Indicates when monitor is charging


batteries (solid) or battery failure (flashing).

Alarm Light

Highly visible Red/


Yellow/Cyan light

Ease alarm detection from distance.

3-3

B40/B20 Patient Monitor

Installation checkout
It is recommended to be performed after monitor installation. Skip the tests that are not
applicable for the installed monitor.
These instructions include a Installation and checkout form, B40/B20 to be filled in when
performing the procedures.
An electrical safety check and a leakage current test should be performed prior to the monitor
installation. Please refer to Technical Reference Manual for details.

Recommended tools
NOTE: Use only properly maintained, calibrated and traceable measurement equipment for
the specified calibrations and adjustments to ensure accuracy.
NOTE: A functional tester cannot be used to assess the accuracy of pulse oximeter for monitor.

Table 3-1

Recommended accessories and tools


Accessories
A rigid cylinder or pipe
NIBP cuff
Adult NIBP cuff hose with cuff ID
Infant NIBP cuff hose with cuff ID
Tubing parts to connect a manometer and a pump to
the NIPB cuff and hose.
Dual invasive pressure adapter cable
ECG accessories, IEC or AHA
-

Multi-link 3-lead integrated cable and leadwire

Multi-link 5-leadwire set

Multi-link 3/5-lead ECG trunk cable

SpO2 finger probe


SpO2 Interconnect Cable
Temperature dual cable
CO2 Sampling line 3m/10 ft
Tool
A multiparameter patient simulator with IBP, Temp
adpter cables
NOTE: For details on recommended accessories see Supplies and Accessories catalog.

3-4

Installation

Visual inspection
Perform the following visual inspection to the installed monitoring system:

Carefully inspect the patient monitor if any damage.

Verify that the modules are properly connected and locked.

Verify that the patient monitor is properly mounted with specified mounting solutions.
Verify that the cables between the patient monitor and the connected peripheral devices
are intact and properly connected to the right connectors.
Verify that the battery door are properly locked.

The cleaning precautions, cleaning requirements, cleaning procedures, and recommended


cleaning solutions for the monitor are described in section "Cleaning and care". For details
about cleaning, disinfecting and sterilizing the accessories, see the instructions for use in the
accessory package.

Functional inspection
Start-up
1.

Turn on the patient monitor.

Verify that the monitor starts up normally:

The red, yellow and cyan alarm lights are lit momentarily.

Check and there are no error messages on the screen.

The speaker gives an audible beep.


Check that the GE logo screen is displayed, followed by the notes screen and the
normal monitoring screen appears.

NOTE: Refer to section "Conditioning the batteries" to see the procedure for battery
conditioning if you receive a a Condition Battery X message.
NOTE: Before taking the patient monitor into use for the first time, the battery should be fully
charged. Keep the monitor connected to the mains until the battery is fully charged.

Display
1.

Verify that all text is readable and all images are clear.

2.

Verify that the brightness is good. Adjust if necessary.

Frame unit
1.

Check that the clock on the screen shows correct time. Readjust the time and date, if
necessary.
NOTE: The B40/B20 cant be set as the TIME MASTER in network. You should adjust the
time and date from the central station.

Parameters measurements
Connect the accessories (no need connect simulator/patient), check the following
phenomenon will appeared.
3-5

B40/B20 Patient Monitor

ECG: After connecting ECG cable, leads off will display in the Waveform Field

SpO2: After connecting SpO2 cable and sensor, SpO2 sensor will be lit.

NIBP: After connecting NIBP hose to module, Adult/Pediatric or Neonatal will display in
NIBP Digital Field for several seconds
Temperature: After connecting Temp cable and sensor, Performing temp test: will
display in Temp Digital Field for several seconds.

IBP: After connecting IBP cable and transducer, InvBPs not Zeroed will display in
Message Field.

CO2: After installing the E-miniC module, Calibration gas sensor will display in CO2
waveform field for about 1 minutes

Recorder
1.

Press the Recorder Start/Stop key and check that the module starts recording the
selected waveforms. Press the Recorder Start/Stop key again to stop recording.

2.

Check that the quality of the recordings is acceptable.

Network connection
1.

Check that the CAT-5 cable connector is clean and intact, then connect it to the Network
connector on the backside of the monitor.
Check that the monitor connects to the network, i.e. the network symbol appears on the
upper right-hand corner of the screen.
NOTE: Pre-configure the network when install the monitor.

Conclusion

Power off the monitor

Perform final cleaning

Fill in all necessary documents

3-6

4 Monitoring basic

Monitoring basic

4 Monitoring basic
Warnings

Connect only one patient to the monitor at a time.


Always make sure that necessary alarm limits are active and set according to the
patient's clinical condition when you start monitoring.

Using menus
A menu is a list of functions or commands displayed on the monitor screen.
To display a menu, press one of the Command Board keys.

Figure 4-1

Example of a menu

(1)

Menu header

(2)

List of menu selections

(3)

Indicates the present selection

(4)

Adjustment window with other options

(5)

Short instructions

(6)

Entry indicator to submenus

4-1

B40/B20 Patient Monitor

Select items in the menus with the Trim Knob. For example, to change what is displayed in the
ECG display:
Select the desired
function by pressing
the menu key.

Turn the Trim Knob


to move the highlight
down in the menu to
the desired selection.

4-2

ECG

Monitoring basic

Push the Trim Knob to


enter an adjustment
window or a submenu

Turn the Trim Knob to


choose the desired
option or selection in
the window.
Push the Trim Knob to
confirm the selection.

Press the Normal Screen key to return to normal monitoring display.

Submenus are indicated by a symbol


contain less frequently used functions.

. They function just like the main menus and

4-3

B40/B20 Patient Monitor

Starting and ending


Warnings

Always make sure that necessary alarm limits are active and set according to the
patient's clinical condition when you start monitoring a patient.

Connect only one patient to the monitor at a time.

Overview
The monitor continuously saves patient data, such as trends. Saving is activated once the
patient is admitted. The monitor saves automatically.

In the monitor memory the most recent patient data up to 72 hours if the network is
not in use.

In the network the most recent patient data up to 72 hours from 2 to 90 days
depending on the configuration.

All the data can be printed through network and saved to the network for monitor memory.

Preparations
1.

Check that the monitor, accessories and monitor parts are clean and intact.

2.

Plug in the desired measurement module.

3.

Turn on the monitor from the ON/standby key. The monitor performs a self-test to
ensure correct functioning.

4.

If necessary, change the user mode:

Press the Admit/Discharge key and select Select Mode.

The mode defines what is displayed on screen and in the trends. Note that changing the mode
also changes settings such as alarm limits.
The monitor automatically reconfigures the display when modules are inserted.
Reconfiguration of the display may take up to 5 seconds.

Starting monitoring
1.

Prepare the patient connections according to the setup picture in the measurement
section.
Use only approved supplies and accessories, see the Supplies and Accessories catalog.
The alarms and parameter default settings become active.

2.

Check that you have the desired waveforms and digits in the fields. If necessary, adjust
the selections:

Press the Monitor Setup key.

Select Screen Setup.

Select Waveform Fields or Digit Fields.

3.

Zero invasive blood pressure lines.

4.

Check the alarm limits.

Press the Alarms Setup key.

The alarms are operative and the parameter default settings are active when the patient
is connected to the monitor.
5.

4-4

Start the measurement according to the instructions in the measurement section.

Monitoring basic

For example, zero the invasive pressure channels by pressing the IBP Zero All key, and
start NIBP measurement to get reference values.
6.

Enter or load patient data.

Admitting a patient
Monitoring of a new patient is started by admitting the patient. After the patient has been
admitted, you can start monitoring and trend gathering. To admit the patient:
1.

Press the Admit/Discharge key.

2.

Select Admit Patient and enter the patient data. Select letters and numbers by turning
and pushing the Trim Knob (max. 14 characters or numbers for each name and ID).

Admit/
Discharge

NOTE: Always observe the monitor and the patient carefully during start-up periods.

Adding demographics
1.

Press the Admit/Discharge key.

2.

Select Admit Patient - Demographics.

3.

Enter patient data: Height is adjustable from 15 to 250 cm (5 in to 8 ft 2 in), Weight is


adjustable from 1 to 250 kg (2 to 555 lb).
The body surface area, BSA, is calculated automatically by using du Bois formula. The
body surface area is used in calculating index values of some parameters.

4-5

B40/B20 Patient Monitor

Loading previous data


If the patient has already been admitted on the same monitor and this monitor has been
restarted, the Contin. Previous will be appeared.
Admit/
Discharge

Contin. Previous
Select this to load the most recent patient trends from the monitor memory when less than 15
minutes has elapsed from the turn-off.

NOTE: This selection is available if the patient is already admitted on this monitor.

During monitoring

If you need to avoid audible alarms, press the Silence Alarms key.
Empty the water trap container of the E-miniC module whenever it is more than half full.

Automatic saving of patient data


The monitor continuously and automatically collects and saves patient data such as trends.
Saving is activated once the patient is admitted or the monitor receives vital data.
If the monitor is not connected to the network, patient data is saved to the monitor memory.
If the monitor is connected to the network patient data is automatically saved to the network.
The network saves data from up to 2 to 90 days. The amount depends on the network
configuration.

Discharging the patient


When you end monitoring, discharge the patient. The monitor erases screen layout, trend data
and alarm and parameter settings that were active during monitoring, and returns to the
starting mode and its settings.

4-6

1.

Press the Admit/Discharge key.

2.

Select Discharge - Yes..

Monitoring basic

Automatic discharge of the patient


The monitor will discharge a patient automatically after 24 hours when vital signs for some
parameters (ECG, Art, NIBP, SpO2, Resp and CO2 (with E-miniC only)) are not available. When this
happens, all trend data will be cleared and alarm limits will be set to default values.

Ending monitoring
1.

Print necessary data.

2.

Press the Print/Record key.

Wait until the printing is finished. Then clear patient data and return settings, including
alarm limits, to their defaults by discharging the patient:

Press the Admit/Discharge key.

Select Discharge and Yes.

3.

Turn off the monitor from the ON/standby key if the monitor will not be used.

4.

Clean the monitor according to the instructions, see section "Cleaning and care."

4-7

B40/B20 Patient Monitor

Using modes
The B40/B20 monitor has seven user modes. These user modes are predefined combinations
of settings. They determine, for example, what is displayed on the screen and in trends and
what the alarm limits are. In other words, by choosing a specific mode you get suitable settings
on the screen without having to choose all features one by one.
Modes can be hospital specific. The monitor starts in start-up modes, which is one of the user
modes chosen during configuration. The default modes are STEP-DOWN, ED, PACU, CCU,
NEURO, PEDIATRIC and NEONATAL. Please refer to the Default Configuration Worksheet
delivered with the monitor for more information.
For more information about the installation settings and using modes, see section "Monitor
setup."

Setup monitor before use


For more information about the how to setup monitor, see section "Monitor setup."

4-8

5 Alarms

Alarms

5 Alarms
Overview
When an alarm for the monitored parameter becomes active:
1

Figure 5-1

View of alarms

(1)

Alarm messages appear in the message field in the order of priority.

(2)

The measurement value flashes. The color (red, yellow) indicates the alarm category (high
priority, medium priority). If the alarm is a low priority, the measurement value is not
flashing.

(3)

In some cases a message in the digit or waveform field gives more detailed information
using the color of the parameter.

An audible alarm is also triggered If enabled, also the alarm light flashes red (high priority),
yellow (medium priority), cyan (low priority) according to alarm levels. Refer to Default
Configuration Worksheet for more information about priorities and escalating.

5-1

B40/B20 Patient Monitor

Safety precautions
Warnings
Verify alarm processing is active and no arrhythmia occurred during power interruption.
Latched alarms are not retained through monitor reset if alarm condition is been
removed.

Do not rely on secondary alarm system for receipt of alarm signal.

The audible alarm signal may be paused temporarily from a central station.

Make sure alarm light brightness is adequate for care environment

Observe patient frequently while alarms or audio are paused or off.


No alarm sound and no remote alarms during Audio Pause
Alarms do not sound, alarm histories are not stored, alarm graphs do not print and
alarms are not sent to the central station during an alarm pause.
ALARMS - Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off during patient monitoring may result in a
hazard to the patient. Remember that the most reliable method of patient monitoring
combines close personal surveillance with correct operation of monitoring equipment.
After connecting the monitor to the central station and/or nurse-call system, verify the
function of the alarm system. The functions of the alarm system for monitoring of the
patient must be verified at regular intervals.
Make sure alarms are active and set according to patient condition
Only the highest priority alarms are sent on Unity Network.
ACCURACY - If the accuracy of any value displayed on the monitor, central station, or
printed on a graph strip is questionable, determine the patients vital signs by alternative
means. Verify that all equipment is working correctly.

Alarm indications

5-2

When the monitor is turned on, you will hear a beep: this tells you that the alarm audio
signal is working. Also the alarm light are lit up in red, yellow and cyan. You can also
check the functioning of the audio signal through Alarms Setup - Alarm Volume.

If alarms are turned off or a power interruption occurs when there is no battery backup,
for up to 15 minutes, check the alarm status before you start monitoring again.

If the monitor is connected to the network, the alarms can be heard and seen on the
Central as well. Please, consult the "iCentral User's Reference Manual: Alarms" or CIC Pro
Clinical Information Center Operator's Manual for details.

If the monitor is connected to the network, the alarms can also be silenced using the
Central if this feature has been enabled in Central configuration.

Alarms

Alarm categories
The alarms are classified into three categories according to the priority: HIGH PRIORITY/RED
ALARM, MEDIUM PRIORITY/YELLOW ALARM, LOW PRIORITY/CYAN ALARM, MESSAGE/WHITE
NOTE

NOTE: Asystole and ventricular fibrillation alarms are always high priority alarms.
The priority of an alarm depends primarily on the cause and the duration (generally minimum
20 seconds) of the alarm condition, the priority increasing with the duration and according to
the physiological significance. Thus, for example, brady advances rapidly to high priority,
whereas apnea is allowed a slightly longer duration.

Table 5-1

Alarm categories

Visual

Meaning

Alarm light (if Alarm in Message


enabled)
Field

Red

For life threatening situations:


HIGH PRIORITY ALARM

flashing red

Solid red background

flashing yellow

Thick yellow boundary,


blue-gray background

Solid Cyan

Thin Cyan boundary,


blue-gray background

No

blue-gray background
field

Require an immediate response.


Yellow

For serious but not life


threatening problems: MEDIUM
PRIORITY ALARM
Require a prompt response.

Cyan

Advisory: LOW PRIORITY ALARM


Require you to be aware of this
condition.

White

Message
Provide additional information.

Alarm light
In addition to the audible, the B40/B20 has an alarm light, located in the top of the monitor. The
alarm light flashes red, yellow and cyan according to the currently active highest priority
alarm. The brightness of the light is fixed.

5-3

B40/B20 Patient Monitor

Alarm tones
The monitor has four options for alarm tones and patterns: ISO, ISO2, General and IEC. The ISO2
and ISO patterns are very similar. The difference is the rising sound of the tone pattern. The
following frame colors and tones identify the alarm categories:

Table 5-2

Alarm tones

Visual

ISO pattern

ISO2 pattern

Gerenal pattern

IEC pattern

Red

11 beeps every 5
seconds or
continuous beep

10 beeps every 5 seconds


or continuous beep
--- -- --- -- (rising tone)

Continuous beep

10 beeps every 5
seconds or continuous
beep
--- -- --- -- (rising tone)

--- -- - ----

Yellow

Triple beep every 19 Triple beep every 19.5


seconds
seconds
--- 19 --- 19 --

Double beep every Triple beep every 19


5 seconds
seconds

Cyan

Single beep -

Single beep -

Single beep -

Single beep -

White

No

No

No

No

-- 5 -- 5 --

19

19

Alarm activation
To enable the alarms, connect patient cables. If the alarm source is selected, the alarms are
active also when the measurement is not displayed (except the impedance respiration alarms).
When an alarm becomes active, messages appear in order of priority. See default settings
presented in the "Default Configuration Worksheet."
Individual alarms have their own specific requirements before they become active, for
example:

5-4

Apnea requires five breaths to be activated.

Invasive pressures need to be within alarm limits for 20 seconds after zeroing.

Alarms

Alarms Setup menu


You can view and adjust patient alarm limits in the Alarms Setup menu.

Alarms
Setup

Adjust Limits

(1)

List of selections

(2)

Exit from the alarm limit adjustment area back to Alarms Setup menu

(3)

Parameter box with high and low limit values and a 10-minute trend showing the current
status

Adjusts individual measurement alarm limits. You can also access the adjustment menu
through each parameter menu.

NOTE: If the monitor is connected to the network, the alarm limits can also be changed
using the Central if this feature has been enabled in its configuration.
Auto Limits

Auto limits are calculated from the displayed patient reading at the time when auto limits are
selected.

Default Limits Sets the alarms to the default alarm limits.


NOTE: Default limits can be changed and saved to modes. For more information, see the
"Default Configuration Worksheet."
Cancel Changes Returns all the limits to the ones set before entering the Alarms Setup menu if you have not
exited the menu yet.
Arrh. Alarms

Opens a menu to select and adjust Severe arrhythmia analysis. Alarm priorities can be
configured in this menu.
Adjust = defines the alarm priorities (red, yellow, cyan, off).

NOTE: You can only set the priority for V Tachy.


NOTE: V Tachy alarm cannot be selected OFF.
NOTE: Alarm priorities can also be set using the Central, depending on its
configuration.
To redisplay the Alarms Setup menu, select Alarms Setup.

5-5

B40/B20 Patient Monitor

For more information on arrhythmia alarms, please see section "ECG."


Alarm Volume Adjusts the volume of the audio alarms. The range is from 1 (soft) to 10 (loud).

NOTE: Audible alarms cannot be totally silenced with the Alarm Volume function.
Alarm Light
This is a grey option menu. The brightness of the alarm light is fixed, cant be adjusted.
Audio ON/OFF Opens a menu to select Silence Apnea, Silence ECG, Silence Apn & ECG or Silence All.
NOTE: The silencing selections that is, all other selections than the Activate Alarms) are
available only if alarm silencing has been enabled by selecting Monitor Setup - Install/
Service - Installation - Alarm Options - Show Audio ON/OFF - Yes. By default, it is disabled. For
more details, see "Enabling or disabling alarm silencing" page 5-14.
Activate Alarms = Activates silenced alarms. This selection is always available for activating
alarms that have been permanently silenced using the Central.
Silence Apn = Silences apnea and disconnection alarms as well as CO2, respiration rate.
Silence ECG = Silences arrhythmia alarms and also HR limit alarms.
Silence Apn & ECG = Silences both of above.
Silence ALL = Silences permanently all alarms except FiO2<18%, Ppeak >70 cmH2O.
Remove Menu

5-6

Clears the menu selections from the display so that only 10 minute trends and limits are
displayed (push the Trim KnobTrim Knob to return the selections on the screen).

Alarms

Adjusting alarm limits


Adjusting limits
1.

Press the Alarms Setup key.

2.

Select Adjust Limits.

3.

Turn the Trim Knob to highlight the measurement. If the desired measurement is not
displayed in the window, select Next Page.

4.

Push the Trim Knob to open an adjustment window.

5.

Turn the Trim Knob to change the limits and accept them by pushing it. Move between
selections by turning the Trim Knob.

6.

To return to the Alarms Setup menu to select more measurements, push the Trim Knob
until the cursor is in the adjustment menu, then select Previous Menu or Alarms Setup.

7.

Press the Normal Screen key to return to normal monitoring view.

You can enter the alarm limit adjustment window also through the measurement menus.
When you are in an XX parameter menu, select XX Alarm or enter XX Setup and then select
XX Alarm.
NOTES:

ST high/low alarms will reach the cyan (low priority) level only.

If the monitor is connected to the network, the alarm limits can also be adjusted using the
Central if this feature has been enabled in its configuration. If alarm limits are adjusted
using the Central, the message 'Alarm settings changed from Central' is displayed in the
bedside monitor.

In NIBP measurement, the alarm limits change automatically according to the cuff hose
type used.

Choosing automatic limits


To activate automatic patient-specific alarm limits enabling close patient control, select Auto
Limits in the Alarms Setup menu. Limits are then calculated from the displayed patient
reading at the point of time when auto limits are selected.

Returning to default limits


Select Default Limits to set the alarms to the default alarm limits. The factory default alarm
limits are listed in the "Default Configuration Worksheet."

5-7

B40/B20 Patient Monitor

Changing alarm sources


For NIBP, IBP1 or IBP2, CO2 FI or ET (with Configuration), you can select which measured values
trigger the alarm. One or several alarm sources may be active at a time.

Parameters

Alarm triggered off by

Blood pressure
(NIBP, P1, P2)

Systolic, diastolic and/or mean pressure or off


(with off, there is no 10 minute trend on
display)

CO2

FI or ET

1.

Press the Alarms Setup key.

2.

Select Adjust Limits.

3.

Select the measurement. If the desired measurement is not displayed, select Next Page.

4.

Push the Trim Knob. An adjustment window is displayed.

5.

In the adjustment window, turn the Trim Knob to change the selection OFF or ON..

Or:
1.

Alarms
Setup

5-8

Select X Alarm in parameters menu and push the Trim Knob to change the selection OFF
or ON.

Alarms

Silencing alarms
Silencing audible alarms temporarily
NOTES:

The bedside alarms can also be silenced and acknowledged from the Central if this
feature has been enabled in the Central configuration. In this case, the message 'Alarms
silenced from Central' or 'Alarms acknowledged from Central' is displayed on the bedside
monitor display.

If the monitor is connected to the network and the network connection is lost, the
silenced alarms are reactivated and the volume level is automatically set to 7.

Pressing the Silence Alarms key once


To silence all alarms for two minutes, press the Silence Alarms key once.
Pressing the Silence Alarms key once silences the alarms that are currently active, and
presilences the upcoming alarms of other measurements. If the alarms are not active when
you press the Silence Alarms key, they are pre-silenced for 2.
The crossed bell symbol with a countdown timer is displayed in the upper left hand corner of
the screen to indicate that the alarms are silenced, and the message field is cleared of all the
previous alarm and note messages. The visual alarms in the digit and waveform fields remain
as long as they are valid. During silencing, all new alarms for the same reason and all alarms
for a different reason are indicated visually. Apnea alarms are activated after five breaths.

Pressing the Silence Alarms key twice


To silence an individual alarm that is currently active, press the Silence Alarms key
twice. This does not pre-silence the upcoming alarms.

The message Alarms acknowledged is displayed to indicate that the alarms that were
silenced by pressing the Silence Alarms key twice remain silent, whereas other upcoming
alarms will have an audible sound.

Reactivating alarms
You can reactivate the alarm sounds of new upcoming alarms by pressing the Silence
Alarms key once again during the silencing period. The alarms that were active when the key
was first pressed will not sound before the original silencing period is over.

5-9

B40/B20 Patient Monitor

Silencing audible alarms permanently


Silencing certain audible alarms entirely may be desirable for special cases.
NOTES:

The following silencing selections (that is, all other selections than Activate Alarms) are
available only if alarm silencing has been enabled (Monitor Setup - Install/Service Installation - Alarm Options - Show Audio ON/OFF - Yes). By default, it is disabled. You
need a password to enter the Install/Service menu. (For details, see "Changing alarm
options" page 6-4.)

If the monitor is connected to the network, the bedside alarms can also be silenced using
the Central if this feature has been enabled in its configuration.

Choosing silence selections:


1.
2.

Press the Alarms Setup key.


Select Audio ON/OFF menu.

Silence Apnea = Silences apnea and disconnection alarms as well as CO2,


respiration rate.

Silence ECG = Silences arrhythmia alarms and also HR limit alarms.

Silence Apn & ECG = Silences both of the above.

Silence ALL = Silences permanently all alarms.

Alarms
Setup

Selecting a choice displays a warning symbol.


If an active alarm is silenced, the monitor gives a reminder beep every two minutes. You can
adjust the volume of the reminder beep through Monitor Setup - Install/Service Installation - Alarm Options - Reminder Volume (a password is required for this operation).
For more information, see "Reminder volume" page 5-16.

5-10

Alarms

Reactivating alarms
1.

Press the Alarms Setup key.

2.

Select Audio ON/OFF menu.

3.

Select Activate Alarms to turn on audible alarms.

Deactivating alarms
You can set the alarm limits temporarily to OFF for the following parameters: HR, SpO2 and
respiration rate. This way, you can silence a parameter without having to adjust the alarm
limits.
1.

Press the Alarms Setup key.

2.

Select Adjust Limits.

3.

Turn the Trim Knob to highlight the measurement.

4.

Select and push the Trim Knob to change the selection Off.
The symbol

appears in the digit field.

When you select the alarm limit ON again, the previous alarm limits will be active.

5-11

B40/B20 Patient Monitor

Automatic recording on alarms


An automatic local recording is possible when Asystole, Tachy/Brady, Art High/Low, V Fib or V
Tachy alarms reach the red alarm level. When the red alarm level is reached, the recorder
prints the ECG + Art waveforms.
1.

Press the Print/Record key.

2.

Select Record Waveforms.

3.

Select Start on Alarms and Yes.

Print/
Record

Recording time is 30 seconds consisting of 12 seconds recording from the recorder memory
and 18 seconds real-time recording. The alarm source is always marked to the alarm
recordings. The following alarms start the recording:

5-12

Alarm

Recorded parameters

Asystole

ECG1 + Art waveforms, 25 mm/s

Tachy/Brady

ECG1 + Art waveforms, 25 mm/s

Art High/Low

ECG1 + Art waveforms, 25 mm/s

V Fib

ECG1 + Art waveforms, 25 mm/s

V Tachy

ECG1 + Art waveforms, 25 mm/s

Alarming level

Alarms which start the recording

Severe

Asystole, Tachy/Brady, Art High/Low, V Fib, V Tachy

Alarms

Showing alarm history

Press the Pt. Data & Trends key and select Alarm History. This displays a list of the
last 20 alarms that have reached the yellow, red and cyan alarm level. The time and type
of occurrence are displayed next to the alarm list items.

Other adjustable features


NOTE: This section describes the rest of the adjustable features regarding the alarms. You can
adjust each feature if you know the required password for entering the Install/Service menu. If
you wish to adjust the settings, we recommend that you contact the person responsible for the
entire configuration.

Displaying limits
You may select the alarm limits to be displayed next to the numerical parameter value.
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password (16-4-34).

3.

Select Installation - Alarm Options.

4.

Select Show Limits and YES or NO.

Monitor
Setup

The alarm limits become visible when patient monitoring begins and the monitor receives
patient data. The default setting will be NO.

5-13

B40/B20 Patient Monitor

Enabling or disabling alarm silencing


With this selection, you can determine whether the audible alarms can be turned off or not.
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Installation - Alarm Options.

4.

Select Show Audio ON/OFF and YES to enable alarm silencing or NO to disable it.

Monitor
Setup

The default setting will be NO.

5-14

Alarms

Latching alarms
If the Latching Alarms selection is active, the alarm messages stay on the screen even if the
initial alarm condition goes away. This enables unattended monitoring. You will also hear a
reminder beep every 10 seconds.
To clear the message field of the no-longer active alarm messages and to clear the beep, press
the Silence Alarms key once.
To clear the message field of all the alarms, enabling only new upcoming alarm messages,
press the Silence Alarms key twice.
To select latching alarms:
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Installation - Alarm Options.

4.

Select Latching Alarms - Yes.

Monitor
Setup

The default setting will be NO.

5-15

B40/B20 Patient Monitor

Reminder volume
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Installation - Alarm Options.

4.

Select Reminder Volume and adjust the alarm reminder volume with the Trim Knob.
Monitor
Setup

The default setting will be 5.

Changing the tone pattern


The monitor has four choices of alarming tone patterns: ISO, ISO2, General and IEC.
To change the tone pattern:
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Installation - Alarm Options.

4.

Select Alarm Tones and ISO, ISO2, IEC or General.

Monitor
Setup

5-16

6 Monitor setup

Monitor setup

6 Monitor setup
Overview
The B40/B20 monitor has numerous setup options for screen, parameters, alarms, etc. There
are two types of settings:

User mode settings: Some of the monitor setup options are preconfigured to be effective.
The preconfigured default settings, so called factory settings, form seven sets of user
modes. The monitor starts in the startup mode, which is one of the user modes. You can
change to another user mode, see "Changing the user mode" page 6-5.
The changes you make in the user mode settings are valid only temporarily until you
discharge the patient or change a mode, or until more than 15 minutes has elapsed from
the turn-off of the monitor. The changes need to be saved in the mode to become
permanent.

Monitor installation settings: the monitor installation settings are the same in all user
modes. The changes are permanent and preserved until changed again.

Before starting to use the monitor, check the monitor installation settings and what is
configured in the different user modes, and make necessary changes. You need passwords for
making the changes, see "Configuration and passwords" below.
This Users Reference Manual describes most of the configurable features. More detailed lists of
the options, and also of the factory settings, can be found in the "Default Configuration
Worksheet" delivered with each monitor.
NOTE: If you wish to make permanent changes, we recommend that you contact the person
responsible for the configuration, who is familiar with the configuration architecture. When
new settings are saved, they should be marked in the "Default Configuration Worksheet."

Configuration and passwords


NOTE: If you want to make changes that require a password, we recommend you contact the
system administrator.

The default password for entering the Install/Service menu is 16, 4, 34.

Press Monitor Setup and select the Install/Service menu, turn the Trim Knob in the opened
adjustment window until you hit the desired number, then push the Trim Knob to accept and
select the number. Continue until all three numbers are selected. After entering the third
number the Install/Service menu is displayed.
Most of the modifications are valid only temporarily unless you accept them in the Save Modes
menu, which is a submenu of the Install/Service menu. A password is also required for
entering the Save Modes menu.

The default password for entering the Save Modes menu is 13, 20, 31.

6-1

B40/B20 Patient Monitor

Setting time and date


The time is shown in the upper right corner of the screen. Turning off the monitor does not
affect the clock.
1.

Press Monitor Setup and select Time and date.

Monitor
Setup

2.

Turn and push the Trim Knob to set the time and date:

Hours, minutes and seconds.

Day, month and year.

NOTE: If the monitor is connected to the Central, the monitor follows the Centrals time settings
and the Time and date menu is not available.

Battery setup
Through this menu you can check the battery status:

6-2

1.

Press the Monitor Setup key.

2.

Select Battery Setup. Battery information is now available.

Monitor setup

Changing monitor installation settings


Monitor
Setup

Changing units
You can change units for height, weight and blood pressure. You can change temperature
units through Others - TempSetup and CO2 units through Airway Gas - CO2 Setup. The
changes are permanent. To change the units:
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Installation - Units.


Monitor
Setup

4.

Set the units for height, weight and blood pressure.

6-3

B40/B20 Patient Monitor

Changing alarm options


1.

Press the Monitor Setup key, select Install/Service and enter the password.

2.

Select Installation - Alarm Options.

Select Show limits and YES to show alarm limits in digit fields. NO is the default.
Select Show Audio ON/OFF and YES to enable alarm silencing. Selecting NO
(default) disables silencing options in the Audio ON/OFF menu in Alarms Setup.

NOTE: The Show Audio ON/OFF setting should be changed only by the system
administrator.

Select Latching Alarms and YES to keep alarm messages on screen until Silence
Alarms key is pressed. NO is the default.

Select Reminder Volume and adjust the volume of the audible alarm reminder tone.
Select Alarm Tones and IEC, ISO, ISO2 (default) or General.

Other alarm settings (alarm limits, and alarm volume) can be changed in the Alarms Setup
menu. To make the changes permanent, save them in user modes, see section "Saving
changes in user modes"

Changing printer settings


1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Installation - Printer.

Select ECG Printout Type and 2x6-25 mm/s (2 columns, 6 lines, 25 mm/s; default),
2x6-50 mm/s (2 columns, 6 lines, 50 mm/s) or 3x4-25 mm/s (3 columns, 4 lines, 25
mm/s).

Select Printer Connection and Net1Net16/None (default).


Select Paper Size and A4 (default) or Letter.

NOTE: Network printer only.


Other printer settings can be changed in the Record & Print menu. To make the changes
permanent, save them in user modes.

Changing the monitor settings


1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Installation - Monitor Settings.

Select Parameter Settings and select:

Humidity compensation type: CO2 Numbers - Dry (default), Wet.

6-4

Monitor setup

Changing the user modes


You can change the settings in each user mode to suit your specific needs. User modes are
predefined combinations of settings that include both general and measurement specific
settings. A user mode defines, for example, what is displayed on the screen and in the trends.
General settings can be changed in the Monitor Setup menu, other settings in the parameter
setup menus. You need a password to make changes through Monitor Setup - Install/Service
menus and to enter the Save Modes menu; for details, see "Configuration and passwords"
page 6-1.
This section describes the following changes that can be saved in user modes:
Changing the Normal Screen layout on page 6-7
Changing sweep speeds on page 6-11
Displaying pulse rate on page 6-11
Changing parameter colors on page 6-11
Changing the recorder and printer settings on page 6-12

Saving changes in user modes


The modifications are valid only temporarily unless you accept them by saving them in the
modes:
1.

Select Monitor Setup - Install/Service. Enter the password.

2.

Select Save Modes. Enter the password.

3.

Select the mode from the list and select Save.


Temporary modifications are valid until you discharge the patient or change a mode or
until more than 15 minutes has elapsed from the turn-off of the monitor.

Changing the user mode


The monitor always starts in startup mode. The startup mode is one of the user modes, and it is
chosen during configuration.
To change from the startup mode to another mode:
1.

Press Admit/Discharge and select Select Mode.

2.

Select one of the following modes:

STEP-DOWN

Mode for intermediate care

ED

Mode for emergency care

PACU

Mode for post-anesthesia care

CCU

Mode for ECG and ST care

NEURO

Mode for neurological monitoring

PEDIATRIC

Mode for pediatric ICU monitoring

NEONATAL

Mode for neonatal monitoring

NOTE: When you want to select NEONATAL mode. You should first:

Press the Admit/Discharge and select Patient Type to NEO

6-5

B40/B20 Patient Monitor

Admit/
Discharge

The selected mode is marked with a circle. You can return to the previous mode by selecting
Return to X.
During monitoring, you can make additional changes to the mode settings and, to make the
changes permanent, save them through the Save Modes menu.

Changing the startup mode


1.

Select Monitor Setup - Install/Service - Save Modes. Enter the password.

2.

Select Startup Mode - 1, 2, 3, 4, 5, 6 or 7.

Renaming a mode
1.

Select Monitor Setup - Install/Service - Save Modes. Enter the password.

2.

Select the mode, select Name and give a new name.

Loading modes
1.

Select Monitor Setup - Install/Service - Save Modes. Enter the password.

2.

Select Load Modes and one of the following:

From Network: Loads all modes from network and saves them in monitor's
permanent memory.

To Network: Copies all modes from monitor's permanent memory to network. This
selection is available only if saving modes to network is enabled in Central

NOTE: Make a discharge before loading modes.

6-6

Monitor setup

Changing the Normal Screen layout


At startup, the screen is arranged according to the startup mode definitions. Parameters that
are not used are not displayed and no space is reserved for them. You can decide which
waveforms and numerical information are displayed, and where on the screen they are
arranged. You can do this for the duration of monitoring or save the changes in the user mode.
To make the changes permanent, save them in the user mode through Monitor Setup Install/Service - Save Modes.

Figure 6-1

Display fields

To change the Normal Screen page:


1.

Press the Monitor Setup key.

2.

Select Screen Setup.

3.

Select Waveform Fields, Digit Fields,split Screen or Minitrend Length.

Monitor
Setup

NOTE: Choosing the same parameter in the waveform and digit field makes the
previously chosen field disappear.
6-7

B40/B20 Patient Monitor

Modifying waveform fields

1.

Press the Monitor Setup key.

2.

Select Screen Setup.

3.

Select Waveform Field.

Up to six waveforms can be displayed at a time.


Monitor
Setup

Screen Setup
Waveform Fields
Lower Field 1 ECG1
Lower Field 2 Resp

OFF

Lower Field 3 IBP1


Lower Field 4 IBP2

ECG2

Lower Field 5 Pleth


Lower Field 6 CO2

IBP1
IBP2
Pleth
CO2

Combine Pressures

NO

Previous Menu

Resp
OFF

Change waveform in field 2.


Select OFF to clear field.

When waveforms are configured to be displayed, they appear and disappear automatically
when module is connected or disconnected. The invasive pressure waveforms are displayed
only when the transducer is connected to the module.
Waveforms are always evenly spread to fill the entire waveform area. Whenever there are less
than 6 waveforms configured on the screen, the remaining waveforms are enlarged.
Changing the waveform to another also changes the numerical field to the right of the
waveform. It may also change the digit fields at the bottom: if you choose the same
measurement in the waveform field that is currently in the digit field, the digit field disappears.
Selecting Combine Pressures in the Waveform Fields menu displays invasive pressures in the
same waveform field with individual scales.
When you use 5-lead ECG measurement, up to three different ECG leads can be displayed
simultaneously in different fields.

6-8

Monitor setup

Modifying digit fields


Patient data may be displayed in up to four digit fields, located in the lower part of the screen.
The fields are numbered from left to right.

Figure 6-2

Digit fields

You may change the contents of each field, or turn them off individually. Before modifying the
digit fields remember to check that the desired parameter module is plugged in.
1.

Press the Monitor Setup key and select Screen Setup.

2.

Select Digit Fields and change the field contents.

Monitor
Setup

Screen Setup
Dight Fields
Lower Field 1

Resp

-More-

Lower Field 2

NIBP

CO2

Lower Field 3

T1+T2

Resp

Lower Field 4

CO2

T1+T2
Temp

Previous Menu

T1
T2
Battery
OFF
Change contents of lower digit
field 1. Select OFF to clear
field.

Changing the digit field may also change the waveform field setup. If you choose the same
measurement in the digit field that is currently in the waveform field, this measurement is
removed from the waveform field.
If a digit field is turned OFF, the remaining digit fields are enlarged to fill the space.

6-9

B40/B20 Patient Monitor

Modifying split screen


You can split Normal Screen so that one part continuously displays trend data.
To select a split screen view:
1.

Press the Monitor Setup key.

2.

Select Screen Setup.

3.

Select Split Screen and choose from the options: Trend or None.

Monitor
Setup

Figure 6-3

Split screen view

Modifying the minitrend length


You can choose to view minitrend data from the last 5-minute or 30-minute period next to the
parameters waveform field. The 5-minute minitrend is updated every 10 seconds, and the 30minute minitrend once every minute.
To modify the minitrend length:

6-10

1.

Press the Monitor Setup key.

2.

Select Screen Setup.

3.

Select Minitrend Length and choose 5 min or 30 min.

Monitor setup

Other adjustable screen features


Changing sweep speeds
You can change the speed of the waveforms on the screen. For CO2 and Resp parameter, the
selections are Fast (6.25 mm/s) and Slow (0.625 mm/s). For hemodynamic parameters, the
selections are 12.5, 25 and 50 mm/s. Slow waveforms have a sweep speed one tenth of
normal, for a full screen sweep. Slow waveforms show amplitude changes better than fast
waveforms.
1.

Press the Monitor Setup key.

2.

Select Sweep Speeds, select the parameter and adjust the value.

Displaying pulse rate


Combined heart rate and pulse rate can be displayed next to the ECG waveform. The current
HR source is displayed with bigger font and the heart rate symbol flashes next to it.

1.

Press the ECG key.

2.

Select ECG Setup.

3.

Select Display with HR and PR, PVC or None.

Changing parameter colors


You can select the color of each parameter to be yellow, white, green, red or blue. For some
parameters, also violet is possible. To change the color:
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Colors.

4.

Select the desired parameter and the color.

6-11

B40/B20 Patient Monitor

Changing the recorder and printer settings


Recorder settings
1.

Press Print/Record.

2.

Select Record Waveforms.

Select Waveform 1 and select a parameter or select OFF. Then select Waveform 2
and Waveform 3 and their parameters. You can record up to three waveforms
simultaneously. If you wish to record only one waveform, select the other waveform
fields OFF.

Select Start on Alarms - Yes to start automatic strip chart recording when the
following alarms reach the red alarm level (NO is the default):

3.

with VSP software license: bradycardia, tachycardia, asystole, Art high/low, V


Fib, V Tachy

Select Delay and OFF or 12 s. If the recording delay is OFF, the recording starts
when an event occurs and continues for 30 seconds or until manually stopped, or
until the recorder runs out of paper. If the delay is 12 seconds (default), the recording
starts when an event occurs and the 12 seconds prior to the event are recorded
from the recorder memory. The recording continues for 18 seconds if the length has
been set to 30 seconds or until the recorder runs out of paper.

Select Paper Speed and 1, 6.25, 12.5, or 25 mm/s (default).


Select Length and choose 30 s (default) or Cont.

Select Record Trends.

Select Trend Resolution and every 1 min, 5 min, 10 min (default) or 30 min. This
setting is for numerical trends.

Select Num Trend Type and Num. (default) or Tab. as the format of the numerical
trend recorded.

Select Graphic. Trend 1 and select the parameter and then do the same for Graphic
Trend 2. These settings define the graphical trends recorded in upper field and lower
field. You can record graphical trends of two parameters.

Printer settings

6-12

1.

Press the Print/Record.

2.

Select Print Graphical

3.

Select Page 1 (default), 2, 3, 4, the ones you want to print.

4.

Select Hours/Page for how many hours will be printed on one page: .1, 2 (default), 4, 6, 8,
10, 12, 24, 36, 48 or 72 hours on one page.

5.

Select End Time, printing starts from selected end time

6.

Select Trend Length and select hours to be printed: 1, 2, 4, 6, 8 (default), 10, 12, 24, 36, 48
or 72 hours.

Monitor setup

Configuring trends
Configuring trend pages
You can change the parameters on the trend fields:
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Trends and set up the Default Trend to Graph

4.

Select Graphical Trends.

5.

Select the trend page that you want to change.

6.

Select graphical parameters for each field.

The field numbers start from the top of the screen. Select one parameter for each field on the
trend page, or turn the field OFF. When all the fields are OFF, the page is displayed with empty
fields. The time scale and page number appear at the bottom of the page.
If several similar fields are selected on top of each other, they form one higher field. Equal fields
cannot be defined separate to each other.

NOTE: You cannot make changes in numerical trend page configuration.

Setting trend length


1.

Press Pt. Data & Trends.

2.

Select Trends - Graphical - Time Scale and select 20min, 1h, 2h (default), 4h, 6h, 8h, 10h,
12h, 24h, 36h, 48h or 72h.

Setting trend scales


You can change the scale for HR, ST, PVC, CPP, SpO2 and temperature trends.
1.

Press Pt. Data & Trends.

2.

Select Trends - Graphical - Trend Scales and adjust the scales.

Setting the default trend


You can select graphical or numerical trends to be displayed by default:
1.

Press the Monitor Setup key.

2.

Select Install/Service and enter the password.

3.

Select Trends.

4.

Select Default Trend and Graph or Num.

6-13

B40/B20 Patient Monitor

Using Network
Use the CAT-5 network cable to connect the monitor to the network.
1.

Make sure that the power is switched off.

2.

Connect one RJ-45 connector to the network port at the back of the monitor.

3.

Connect the other RJ-45 connector to the corresponding port on the wallbox.

4.

Switch on the power. Confirm that the network symbol and 'Connected to Network'
message are displayed in the upper part of the screen.

NOTE: If any problems or need advanced configuration, please ask the GE service personnel to
support.

6-14

7 Trends

Trends

7 Trends
Overview
The monitor displays two types of trend data: graphical and numerical. The monitor collects
graphical and numerical trend data automatically from trended variables. You can select the
trend time between 20 minutes and 72 hours.
You can view the trends through Pt.Data & Trends - Trends, or you can select graphical
minitrends to be displayed continuously next to the waveform fields (Monitor Setup Screen
Setup Split Screen).

Trended parameters

Electrocardiography (HR, ST)

Invasive pressures

Non-invasive blood pressure

Oxygen saturation (Pleth, SpO2)

Gases (CO2, Resp)

Impedance respiration (Resp)

Temperatures

Trend data is stored in the memory for 15 minutes after the power has been turned to standby.

7-1

B40/B20 Patient Monitor

Most common tasks


Displaying trends and
activating the Trends menu

Scrolling time with trend cursor

1. Turn the Trim Knob to move the cursor to the time


you want.
2. Numeric measurement values of that time are
displayed in the graphical trend next to the cursor.
3. Push the Trim Knob to return to the menu.

Scrolling pages to see more


parameters

Changing the time scale

1. Select Time Scale.


2. Select the trend time (20 min or 1, 2, 4, 6, 8, 10, 12, 24,
36, 48, 72 hours).

Selecting numerical or graphical


trends to the screen

Select Graphical or Numerical.

Changing trend scales

Select Trend Scales.

Printing trends

To print the currently viewed trend data, select


Graphical or Numerical and then Print Page.

To print all the graphical trend data, press the Print/


Record key and select Print Graphical select page

Press the Pt.Data & Trends key and select Trends.


The most recently displayed trend (graphical or
numerical) is displayed together with the Trends
menu.

Select Scroll Pages and turn the Trim Knob to move


from one page to the other.

- Print Graphs.
Recording trends

7-2

To record numerical trends, press the Print/Record


key and select Record Trends - Record Numerical.

To record graphical trends, press the Print/Record


key and select Record Trends - Record Graphical.

Erasing trend history

1. Press the Admit/Discharge key.


2. Select Discharge - Yes.

Selecting minitrend as split


screen option

1. Press the Monitor Setup key.


2. Select Screen Setup and Split Screen Trend.

Trends

Minitrend view

Figure 7-1

Minitrend view

You can split the Normal Screen page so that one fourth of the screen, on the left hand side,
continuously shows graphical minitrends beside waveforms. Note that the split screen option
is available only when the Normal Screen page shows waveforms.

7-3

B40/B20 Patient Monitor

To select a split screen view:


1.

Press the Monitor Setup key.

2.

Select Screen Setup.

3.

Select Split Screen and Trend.


Monitor
Setup

Minitrend length
You can choose to view trend data from the last five minutes or the last 30 minute period. The
five minute minitrend is updated every 10 seconds, the 30 minute minitrend is updated once
every minute.
To modify the split screen trend view:
1.

Press the Monitor Setup key.

2.

Select Screen Setup.

3.

Select Minitrend Length and choose 5 min or 30 min.

Removing minitrend
To remove the minitrend from the screen:

7-4

1.

Press the Monitor Setup key.

2.

Select Screen Setup.

3.

Select Split Screen and None.

Trends

Graphical trend view

Figure 7-2

Graphical trend page

(1)

Trends menu

(2)

Measurement trend field

(3)

Real time ECG

(4)

Numeric value of a measurement at the trend cursor point

(5)

Trend page number

(6)

Indication of the amount of data gathered and viewed

(7)

Time and marker field

7-5

B40/B20 Patient Monitor

Symbols
Trend bar, parameter scale to the left.
The gap shows the blood pressure mean value.

NIBP trend bar

Dotted vertical line across the trend field indicates change, such as
ST relearning or zeroing of an invasive blood pressure channel/
changing a label.

___________

A blue, white or red line above the marker field indicates the
following things:
Blue line indicates the amount of data on the screen: the left end of
the line shows the starting point of the trend data gathering. The
right end of the line shows the last moment the data has been
gathered.
White line indicates which proportion of the data you see on the
screen. If the line is on the left, there is more data to see after the
current view. If the line is on the right, there is more data to see
before the current view. If the line is in the middle, there is more
trend data to see towards the beginning and the end of the case.
Red line indicates the availability of trend data with 10 second
resolution (available for the last 30 minutes only).

7-6

Trends

Graphical trend pages


Graphical trends contain:

Four pages

Six fields on each page

Five fields are usually visible. The lowest (sixth) field is replaced by digit fields on the screen. All
six fields are printed.
Scale, label, unit and color of the parameter follow the real time waveform setting for each
parameter. For HR, ST, PVC, CPP, SpO2 and temperature you can select the scale through
Pt.Data & Trends - Trends - Trend Scales.

Changing trend length and resolution


1.

Press the Pt.Data & Trends key.

2.

Select Trends - Time Scale.

3.

Select the trend length.

Table 7-1

Trend length and resolution

Trend length on the


screen

Resolution

Trended time period

20 minutes

10 seconds

last 30 minutes

1 hour

1 minute

last 24 hours

2 hours

1 minute

last 24 hours

4 hours

2 minutes

last 24 hours

6 hours

3 minutes

last 24 hours

8 hours

4 minutes

last 24 hours

10 hours

5 minutes

last 24 hours

12 hours

6 minutes

last 24 hours

24 hours

12 minutes

last 24 hours

36 hours

18 minutes

last 36 hours

48 hours

24 minutes

last 48 hours

72 hours

36 minutes

last 72 hours

7-7

B40/B20 Patient Monitor

Moving on graphical trend pages


To see more parameters on other pages:

Select Scroll Pages in the Trends menu.

Numeric measurement values for trended parameters are displayed next to the cursor. The
cursor indicates the time when these values have been measured. To change the cursor
location:
1.

In the Trends menu, select Cursor.

2.

to move the cursor.

3.

to return to the menu.

To scroll the time, move the cursor past the right or left border of the trend.

Recording and printing


Recording
1.

Press the Print/Record key.

2.

Select Record Trends.

3.

Select Record Graphical.

The recording time of a trend corresponds to the time scale of the graphical trends. You can
choose the time scale (20 minutes to 72 hours) in the trends menu.
To select the parameters for graphical trend recording:
1.

Press the Print/Record key.

2.

Select Record Trends.

3.

Select Graphic Trend 1 and choose the parameter in the opened adjustment menu.

4.

Select Graphic Trend 2 and choose the parameter in the opened adjustment menu.

Printing
You can print all the graphical trends gathered:
1.

Press the Print/Record key.

2.

Select Print Graphical.

3.

Select the page and Print Graphs.

Factory default parameters


The default graphical trend setup varies according to mode types. For details refer to Default
Configuration Worksheet. To change these settings, see section "Monitor setup."

7-8

Trends

Numerical trend view

Figure 7-3

Numerical trend page

(1)

Trend menu

(2)

Page name and number

(3)

Real time ECG

(4)

Real time digit fields, if the Normal Screen shows waveforms

7-9

B40/B20 Patient Monitor

Numerical trend pages


Numerical trends contain:

four pages of maximum 72 hours trend information

real-time ECG on top of each page

Resolution is five minutes.

Moving between numerical trend pages

Use the Trim Knob to scroll the trend in vertical direction.

When the highlight reaches the top or the bottom of the view, next five minutes of information
appear.

Recording and printing


Recording
Parameters for recording are chosen during configuration. To change the parameters, see
section "Monitor setup."
1.

Press the Print/Record key.

2.

Select Record Trends.

3.

Select Record Numerical.

Printing
1.

Press the Pt.Data & Trends key and select Trends.

2.

To print all the numerical trend data, select Numerical - Print Page.

Factory default parameters


You cannot change the contents of numerical trend fields. The parameter units follow the real
time waveform settings of each parameter.
Page 1: Vital parameters
Mark

Time

HR

SpO2

NIBP
NIBP
sys/dia mean

Art
Art
sys/dia mean

CVP
mean

CO2
ET

IBP2
CVP

C.O.

REF

PCWP

Page 2: Hemodynamics
Mark

Time

IBP1
Art

Page 3: Gases
Mark

Time

CO2
ET/FI

RR

Page 4: Temperatures and saturations


Mark

7-10

Time

T1

T2

Tblood

SpO2

SvO2

Trends

Erasing trend data


Trends are erased when you discharge the patient.
1.

Press the Admit/Discharge key.

2.

Select Discharge.

3.

In the opened window, select Yes to erase the trends and to discharge the patient.

Admit/
Discharge

7-11

B40/B20 Patient Monitor

7-12

8 Print and record

Print and record

8 Print and record


Overview
You can manage recording and printing via the Print/Record menu. For recordings, you need a
extension rack with recorder, and for printouts you need connect to network and a laser
printer (PCL5 compatible, min. 2 MB memory). The monitor is connected to a laser printer via
network.

Recorder

Recording
The recorder:

Near real-time recording of up to three user-selectable waveforms simultaneously

Selectable paper speed

Prints up to 72 hours of graphical and numerical trends

Numerical printout with annotation when one or two waveforms are printed

Numerical trend resolution selectable

Configurable to create a 30-second recording automatically on alarm

NOTE: Recording on thermal paper may be destroyed when exposed to light, heat,
alcohol, and so on. Take a photocopy for your archives.

Direct function keys


There is a key for starting and stopping recorder on the Command Board:
Recorder
Start/Stop

For starting or stopping recording waveforms to local.


NOTE: Functional with the recorder only.

8-1

B40/B20 Patient Monitor

Recording waveforms
You can record three waveforms to a local recorder as you set up.
To start the recording, do one of the following:

Press the Recorder Start/Stop key, or


Press the Print/Record key and select Record Waveforms - Record to Local.

To stop the recording:

Press the Recorder Start/Stop key, or


Press the Print/Record key and select Record Waveforms - Stop Waveforms.

You can also use the network recorder by selecting Record to Net. The network recorder uses
the settings of the Central.
Print/
Record

8-2

Print and record

Sample of waveform recording

(1)

Recorder speed

(2)

Scale

(3)

Selected waveform

(4)

Date

(5)

Time

NOTE: Waveform scaling follows the displayed parameter scaling, when applicable.

Selecting waveforms for recording


You can select which waveforms and how many of them you want to record.
1.

Press the Print/Record key.

2.

Select Record Waveforms.

3.

Select Waveform 1 and a parameter for it, or select OFF.

4.

Select Waveform 2 and 3 and their parameters.

You can record simultaneously up to three waveforms. The following menu figure shows which
parameters have recordable waveforms.
Print/
Record

Record & Print


Record Waveforms
Record to Net
ECG1

Record to Local
Waveform 1

ECG1

ECG2

Waveform 2

OFF

ECG3

Waveform 3

OFF

IBP1

Start on Alarms
Delay
Paper Speed
Length

ON

IBP2

12 s

Pleth

25 mm/s

CO2

30 s

Resp

Previous Menu

OFF

Change waveform recorded


in upper field.
Select OFF to clear field.

8-3

B40/B20 Patient Monitor

Changing the paper speed


To see the waveforms more clearly or more generally, you can change the paper speed. The
recorder speed can be 1, 6.25, 12.5, or 25 mm/second.

Print/
Record

8-4

1.

Press the Print/Record key.

2.

Select Record Waveforms.

3.

Select Paper Speed.

Print and record

Controlling the recording time


You can adjust the recording time to be 30 seconds or continuous, which means that the
recording continues until the recorder runs out of paper.
1.

Press the Print/Record key.

2.

Select Record Waveforms.

3.

Select Length and choose 30 s or Cont.

The default setting for the recording time is 30 seconds.

NOTE: When recording is activated by alarms, the recording time is always 30


seconds.
Print/
Record

Selecting the recording delay time


If the recording delay time is set to OFF, the recording starts when an event occurs and
continues for 30 seconds or until it is manually stopped, or until the recorder runs out of paper.
If the delay time is set to 12 seconds (default), the recording starts when an event occurs and
the 12 seconds prior to the event are recorded from the recorder memory. The recording
continues for 18 seconds if the length has been set to 30 seconds or until the recorder runs out
of paper.
To change the delay:
1.

Press the Print/Record key.

2.

Select Record Waveforms.

3.

Select Delay and OFF or 12 s.

The recording can be started manually, or automatically when certain alarms occur. The
automatic alarm recording is explained in the following.

NOTE: When recording is activated by alarms, the delay is always 12 seconds.

8-5

B40/B20 Patient Monitor

Recording on alarms
An automatic strip chart recording is activated when the following alarms reach the red level:
Asystole, Tachy, Brady, Art High, Art Low, V Fib, and V Tachy. When recording is activated by
alarms, the recording time is always 30 seconds and the delay always 12 seconds.
1.

Press the Print/Record key.

2.

Select Record Waveforms.

3.

Select Start on Alarms and Yes.

Print/
Record

The following alarms start recording:

Alarm

Recorded parameters

Asystole
Tachy/Brady
Art High/Low
V Fib
V Tachy

ECG1 + Art waveforms, 25 mm/s


ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s

Arrhythmia analysis

Alarms which start the recording (if the alarm


is set on/level red)

Severe

Asystole, Tachy/Brady, Art High/Low, V Fib, V Tachy

All alarm recordings are marked with the alarm source.

8-6

Print and record

Recording trends
To record trends:
1.

Press the Print/Record key.

2.

Select Record Trends Record Numerical or Record Graphical.

3.

Stop recording by selecting Stop Numerical or Stop Graphical.

You can record numerical, graphical or tabular trends.

Recording numerical trends


Since the contents of the numerical trends are preconfigured, you cannot choose the
parameters or change their order.
The following parameters are printed in the numerical trend record:

Parameter

Printed values and units

HR and SpO2

bpm/SpO2

NIBP

Sys/dia or mean mmHg

IBP1 Art

Sys/dia or mean mmHg

IBP2 CVP

Sys/dia or mean mmHg

T1/T2

Celsius or Fahrenheit

Tblood

Not supported by B40/B20 monitor

CO2

ET/FI %, kPa or mmHg

Resp. Rate

Breaths per minute

Figure 8-1

Numerical trend printout

For pressures, either Sys/Dia or Mean are recorded depending on the digit format selected in
the pressure setups.

8-7

B40/B20 Patient Monitor

Selecting the format for the recorded numerical trends


You can select the format for the recorded numerical trend to be either Num. (vertical) or Tab.
(horizontal):
1.

Press the Print/Record key and select Record Trends.

2.

Select Num Trend Type and Num. or Tab.

Changing resolution
To select the resolution of a numerical or tabular trend record:
1.

Press the Print/Record key and select Record Trends.

2.

Select Trend Resolution. Choices are: every 1 minute, 5 minutes, 10 minutes, 30 minutes.

Tabular trend format


Tabular trend printout
Parameter

Printed values and units

HR

bpm

SpO2

NIBP or Art

sys/dia mmHg

CO2

Et%, mmHg or kPa

Selecting graphical trends


You can record graphical trends of two parameters.
To select the parameters for the graphical trends:
1.

Press the Print/Record key.

2.

Select Record Trends.

3.

Select Graphic Trend 1 or Graphic Trend 2.

The following figure shows which parameters have graphic trends that you can record.
Print/
Record

Record & Print


Record Trends
Record Numerical

-More-

Trend Resolution

5 min

HR

IBP1

Num. Trend Type

Num

ST

IBP2

IBP1

NIBP

IBP2

SpO2

NIBP

Pleth

SpO2

CO2

Record Graphical
Graphic Trend 1

HR

Graphic Trend 2

IBP1

Previous Menu

Change graphical trend


recorded in upper field.

8-8

Pleth
CO2

Resp

-More-

T1+T2

Print and record

Inserting recorder paper


To load paper in the reocrder:
1.

Open the recorder door by pressing the door latch.

2.

Remove the paper core.

3.

Place a new paper roll between the tabs of the paper holder. The paper should unroll from
underneath the paper roll.

4.

Begin to close the door while still holding onto and pulling about 6 to 8 cm of paper; then
close the door completely. You should hear a snap as the door latch is engaged.

5.

Attempt to print out a annotated strip by pressing the Recorder Start/Stop key, to
make sure you load paper in the correct direction.

Figure 8-2

Inserting recorder paper

8-9

B40/B20 Patient Monitor

Printing
The B40/B20 monitor support the laser printer in S/5 network. The user can directly print the
patient data from the laser printer when the monitor and printer are correctly configured on
iCentral Server.
The B40/B20 monitor dont support the laser printer in CARESCAPE Unity network. The user can
print the patient data from the CIC central station only.
You can print to a laser printer:

Graphical or numerical trends


You can print several pages or trend data through the Print/Record menu.

NOTE: Before you start printing, check that the printer is operational.

Printing currently displayed screen contents


To print a trend view
1.

Press the Pt.Data & Trends key.

2.

Select Trends - Graphical/Numerical.

3.

Select the desired trend page with Scroll Pages.

4.

Select Print Page.

Printing all the information


To print several pages of graphical trend data:
1.

Press the Print/Record key.

2.

Select Print Graphical.

Changing the printer


If you need to change the printer connection:
1.

Press the Print/Record key.

2.

Select Printer Connection.

3.

Select the printer from the list.

The default printer is set during configuration, see section "Monitor setup."

Other adjustable features


To adjust the ECG printout type, default printer connection or default paper size, please, see
section "Monitor setup."

8-10

9 Cleaning and care

Cleaning and care

9 Cleaning and care


Safety precautions
Warnings

Disconnect equipment from power line before cleaning

Regular preventive maintenance should be carried out annually.

Do not pour or spray any liquid directly on cables or leadwires or permit fluid to seep into
connections or openings.

Never use conductive solutions, solutions that contain chlorides, wax, or wax compounds
to clean devices, cables or leadwires.

Do not immerse any part of the device in liquids or allow liquid to enter the interior.

Clean the surface of the probe before and after each patient use.

If liquid has accidentally entered the system or its parts, disconnect the power cord from
the power supply and have the equipment serviced by authorized service personnel.
Do not use unspecified cleaners, materials or chemicals as they may damage device
surfaces, labels, or cause equipment failures.

The decision to sterilize must be made per your institutions requirements with an
awareness of the effect on the integrity of the cable or leadwire.
CABLE/SENSOR AFTER CARE

Do not immerse sensors or patient cables in water, solvents or cleaning solutions.

Do not reuse sensors intended for single patient use.

Do not sterilize sensors or patient cables by irradiation, steam, or ethylene oxide.

Clean the surface of the probe before and after each patient use.

Allow sensor and cable to dry completely after cleaning. Moisture and dirt on the
connector can affect the measurement accuracy.

If a probe is damaged in any way, discontinue use immediately.

Inaccurate SpO2 data can result if a sensor is past its useful life.

A damaged sensor or a sensor soaked in liquid may cause burns during


electrosurgery.

Since calibration gas contains anesthetic agents, always ensure sufficient ventilation of
the room during calibration.

Inaccurate readings due to:

use of unapproved accessories

reuse of single-use accessories

affect on pressure strong scavenging suction

The user may only perform maintenance procedures specifically described in this
manual.

On the E-MiniC module indicates that airway gases should be calibrated every six months
in normal use and every two months in continuous use.

9-1

B40/B20 Patient Monitor

Cautions

Dispose of packaging material by observing applicable waste control regulations

Overview
For safe and reliable function and operation of the monitor, regular care has to be carried out
according to the instructions in this manual and to the maintenance procedures described in
the "Technical Reference Manual."
If the monitor does not function as it should and troubleshooting cannot solve the problem,
contact your service or sales representative. Do not perform any other cleaning or
maintenance procedures than those described in the monitor manuals.

NOTE: Refer to the accessory package for detailed cleaning instructions.

Preventive maintenance
Daily and between the patients

9-2

Wipe the monitor and module surfaces.


Wipe the ECG trunk cable and leadwires, NIBP cuff and cables and SpO2 sensors. Avoid
excessive use of liquids.

Change or sterilize all airway and invasive patient accessories.

Check/note the following points regarding different parameters.

Clean, disinfect or sterilize reusable temperature probes.


Empty the D-fend water trap whenever half full.
Check that all accessories, cables and monitor parts are clean and intact.
Clean the device as described in the Cleaning part of this section.
When you start monitoring, check that the module is firmly in place, the accessories are
intact and properly connected, and that you have selected desired parameters to be
displayed in digit and waveform fields.

Cleaning and care

Regular checks
If the monitor does not work as described, see section "Troubleshooting." first.
When you start monitoring, check that:

The module is firmly in place.

Accessories are intact and properly connected.

You have selected desired parameters to be displayed in the digit and waveform fields.

ECG and impedance respiration


Check that the message 'Leads off' disappears and the waveforms are displayed when
the cable is connected to the patient.

Pulse oximetry
Check that the red light is lit in the sensor.
Check that the SpO2 value is displayed and the message 'SpO2 probe off' disappears
when the sensor is connected to the patient.

Temperature
Check that the temperature value is displayed when the probe is connected to the
patient.

InvBP
Check that the monitor recognizes cable connections (activates the display) for all the
pressure channels used and the pressure values are shown.

Make sure all transducers are zeroed correctly.

NIBP
Ensure that you are using correct cuff size and have selected correct inflation limits. For
neonatal/infant and when using hoses without identification, the inflation limit must be
set manually.

Check that the cuff hose detection works properly.


Check that the pressure values are displayed.

Airway gas (CO2)


Check that the water trap is empty.
Occlude the sampling line and check that the message Sample line blocked appears
within 30 seconds and gas waveforms are showing zero at the same time.

Functioning of the alarms


Turn on the patient monitor. Check that the red, yellow and cyan alarm lights are lit
momentarily.

Check that the speaker gives an audible beep.

Safety checks for software


The GE Healthcare software design controls include performance of a risk analysis using
methods consistent with ISO 14971 Medical devices - Application of risk management to
medical devices.
The monitor software employs watchdog timers, self-monitoring activities (memory,
communication and sensor checks and so on), and power-on self-diagnostics (for example,
memory checksums). For example, for SpO2 the software continuously monitors the SpO2
9-3

B40/B20 Patient Monitor

sensor and, if excessive sensor current is detected, the message Faulty probe is displayed in
the SpO2 number field and SpO2 Faulty in the monitor message field, and the old SpO2 data is
removed from the display.

Every six months


Gas calibration
Perform gas calibration for airway gas monitoring according to the instructions in the relevant
section. If gas measurement is in extensive use, calibration is recommended every two
months.

NOTE: Do not wash or disinfect calibration gas sampling lines.

Conditioning the batteries


Condition batteries regularly to maintain after their useful life. This is best done on an external
charger. Condition a battery every six months or when the message Replace Battery x
appears status. You can also check the status through Monitor Setup - Battery Setup.
If you do not have an external charger, proceed according to the following instructions.

NOTE: You cannot condition batteries during patient monitoring. Always disconnect
the modules first. Save the patient data and mode settings if necessary.
1.

Continue normal battery use until the green bar of a battery charge indicator in screen is
less than 3/4 of the full height. After this, remove the battery. Continue monitoring with
one battery until its charge is less than 3/4 of the full capacity.

2.

Insert both batteries and connect the monitor to the power supply. The monitor starts
charging both batteries, and the capacity indicators scroll accordingly. Keep charging the
batteries until both capacity indicators are full height.

3.

Continue charging for another two hours. After this, check that the orange battery LED in
the panel is no longer on. If it is, continue charging until it goes off.

4.

Disconnect the monitor from the power supply and leave it on until the batteries run out
and the monitor switches off. Wait for another 15 minutes.

5.

Reconnect the monitor to the power supply and turn it on. Continue charging the
batteries until both capacity indicators are full height and no longer scrolling.

6.

Keep charging for another two hours. After this, check that the orange battery LED in the
front panel is no longer on. If it is, continue charging until it goes off to indicate that the
battery conditioning is complete.

Every 12 months
Preventive maintenance check
The annual check according to detailed instructions of the "Technical Reference Manual"
requires trained service personnel and appropriate testing tools and equipment.

Calibration check of temperature, NIBP and invasive blood pressures


Calibration check of temperature, NIBP and invasive blood pressures should be performed at
least once a year by qualified service personnel as a part of the Maintenance, see "Technical
Reference Manual.

Power interruption
NOTE: If the monitor is turned off, trend data and the latest user-made settings remain
in the monitor memory for 15 minutes even if the mains power is interrupted. If not,

9-4

Cleaning and care

contact service personnel. After 15 minutes, trend data is lost and the monitor returns
to the user default settings.

Changing fuses
1.

Remove the power cord if used.

2.

Remove the fuse holder by pulling the holder out with screwdriver.

3.

If a fuse is blown, replace it with a fuse of the correct type and rating.

NOTE: Only a qualified medical personnel allow to operate.

Cleaning
The appropriate cleaning procedure depends on where and how the part or accessory is used
and on the patient's condition.

Permitted detergents and disinfectants

Tap water

Distilled water

Ethyl alcohol 99.5%

Ethyl alcohol 95 to 96%

Ethyl alcohol 90% + methyl alcohol 10%

Ethyl alcohol 80%

Isopropyl alcohol 60%

Phenol 2%

Purified benzine

Chloramine 5%

Glutaraldehyde 2%

Monitor casing
Wipe with mild hospital detergent solution. Make sure not to leave any liquid spills on any
metal part.

Let dry completely before connecting to power source.


Check ventilation holes and clean if necessary.

Modules
Wipe the front panel as any monitor casing.
Do not wash or immerse module in any liquid.
The internal sampling system of the airway module does no cleaning nor sterilization. The
water trap functions as a bacteria filter and there is no reverse flow back to the patient. If the
measuring chamber is suspected to be contaminated (for example, gas zero error), the airway
module should be serviced by authorized service personnel.

Display
Wipe all splashes immediately with a dry cloth.
Wipe the LCD display after use with a cloth moisturized with mild detergent solution.
ECG cables
Wipe the cables with mild detergent solution.
9-5

B40/B20 Patient Monitor

Avoid excessive use of liquids.


Disinfect when necessary.
Allow the product to dry completely after cleaning.

Pulse oximetry sensors


The GE Healthcare pulse oximetry sensors are latex-free. Take possible patient allergies into
account also when selecting the cleaning agent.
1.

Detach the sensor from the patient and the monitor.

2.

Wipe the sensor with mild detergent solution. Allow it to dry completely before use.

Sensors can be disinfected with chlorite compounds.


The sensors may be sterilized using an ethylene oxide mixture at 50 to 60C / 120 to 140F.
NOTE: After ethylene oxide sterilization, sensors must be well aerated in a ventilated place.

Temperature probes
Clean with mild detergent solution and rinse with water.
Disinfect or sterilize when necessary.
NIBP cuff hose
The cables and hose can be wiped with mild detergent solution.
Disinfect when necessary.
NIBP cuff
The NIBP cuffs listed in the Supplies and Accessories catalog are latex-free. Take possible
patient allergies into account also when selecting the cleaning agent.

Clean only when necessary. Wash the cuff in mild detergent solution. Do NOT use alcohol.

Invasive blood pressure cables


Wipe the cables with sterile alcohol-based detergent. After cleaning rinse surfaces by
wiping them with a cloth damped with sterile water. Dry with a dry cloth.

Airway adapter
Replace the single use adapter after each patient.
A reusable adapter can be disinfected with glutaraldehyde or alcohol.
If you want to clean the adapter before use: submerge the adapter in 70% alcohol solution for
30 seconds and rinse carefully with water.
Make sure that all traces of alcohol or detergent are rinsed away or dried before connecting to
the patient.

Sampling line
Do not reuse the sampling line. Reusing a cleaned sampling line may affect measurement
results.
NOTE: Do not wash or disinfect calibration gas sampling lines.

Water trap
The water trap is based on a hydrophobic membrane, which prevents water and secretions
from entering the measuring chamber. Condensed water and saliva are collected into a
washable container.
Replace the water trap every 24 hours or when the message Sample line blocked or Replace
D-fend persists.
9-6

Cleaning and care

The water trap container can be cleaned with disinfecting solutions or sterilized using cold
chemicals or ethylene oxide.
To lengthen the lifetime of the monitor and minimize downtime:

Empty the water trap container whenever it is more than half full.

Do not force air or oxygen through the water trap.

Do not open, wash or sterilize the water trap cartridge.


After washing or disinfecting the water trap container, make sure there is no alcohol nor
detergent left when used again. Traces of alcohol or other organic cleaning solutions
may affect measurement.
Do not allow smoke and dust to enter the water trap.
While administering nebulized medication, disconnect the gas sampling line from the
patient circuit for 30 minutes.

If the message Sample line blocked alarm occurs:

Replace the sampling line.


Empty the water trap container. It may be full.

Figure 9-1

Emptying the water trap container

To remove the water trap, push the trigger above the water trap. The water trap is springloaded. The message Check D-fend is displayed.

Detach the container from the water trap cartridge by pulling it carefully downward.
Empty and clean the container.
Attach the container tightly back into the cartridge.
Push the whole unit into its housing on the front panel until the latch is set.
Press the Normal Screen key to restart monitoring. Check that the message Check Dfend disappears.

Other accessories
See the accessory package for instructions for how to clean and check them. Do not reuse
single-use disposable accessories.

9-7

B40/B20 Patient Monitor

9-8

10 Troubleshooting

Troubleshooting

10 Troubleshooting
Overview
The troubleshooting section consists of two parts which together should help you resolve the
most common monitoring problems. The parts are "Checklist" and "Messages".

Checklist
Check the following things before monitoring to ensure that you have remembered to make all
essential preparations, and if any problems occur during monitoring. Parameter-specific
checklists can be found in the respective parameter sections.
Check that:

The monitor and its module do not have any visual defects such as cracks or loose parts.

Patient connection cables are attached to the module connectors.

Disposable accessories are not reused.

The batteries are inserted and charged.


The power cord is connected to an electrical wall outlet and to the monitor.
The module are pushed properly into the frame and the monitor recognizes the module
(parameter is displayed on the screen and menu selections are active).
Trends of the previous patient are erased.
Alarm limits are suitable for the patient.
The desired measurements have been selected for screen through Monitor Setup
Screen Setup - Waveform Fields or Digit Fields.
Sensors are not expired.

Power interruption
If the monitor is turned off, trend data and the latest user-made settings remain in the monitor
memory for 15 minutes even if the power (mains) is interrupted. After 15 minutes, trend data is
lost and the monitor returns to the user default settings (start-up mode).

10-1

B40/B20 Patient Monitor

Messages
Table 10-1

Messages

Message

Explanation

What to do

Alarms acknowledged from


Central

Silenced alarms remain silent. New


alarms will have an audible sound.
(Can be done using the Central).

If required, turn on the alarms


through Alarms Setup Audio
ON/OFF Activate Alarms.

Alarm setup changed from


Central

Alarm limits or arrhythmia alarm


priorities have been changed using
the Central.

Check the alarm limits and the


arrhythmia alarm priorities; see
"Alarms" and "Troubleshooting".

Alarms silenced from Central Alarms have been silenced using the
Central.

If required, turn on the alarms


through Alarms Setup Audio
ON/OFF Activate Alarms.

Apnea

Check the patient status.

Check the patient status.

Asystole

No breath detected for 20 seconds.

No QRS detected in ECG.

Check the ventilator and breathing


circuit.

Check the electrodes.

Battery low

About 20 minutes of battery operation


time left.

Replace the battery or connect the


monitor to power outlet.

Brady

HR is equal to or below the lower


alarm limit.

Check the patient status.

Calibration not protected

Unsuccessful calibration.

Perform a new calibration.

Call service: Error X

NIBP error

Contact authorized service


personnel.

Check D-fend

The water trap is not attached.

Check that the water trap is


properly attached to the module. If
the problem persists, contact
authorized service personnel.

Check NIBP

NIBP measurement affected by low


blood pressure and pulsation, or a
change in patient's condition.

Check the patient status.


Check the measurement setup.
Check the cuff.

Check sample gas out

Sample gas outlet is blocked.

Remove blockage from the sample


gas outlet.

Check SpO2 probe

SpO2: There is no detectable SpO2


signal, the sensor is faulty or it is
detached from the patient.

Check the sensor and connections.

Check network connectors

The virtual plug ID is changing

Check the virtual plug ID.

10-2

Troubleshooting

Message

Explanation

What to do

Condition Battery A,
Condition Battery B

The batteries need condition

Condition the battery according to


the instructions of the external
charger.

Cuff loose/NIBP cuff loose

Cuff is not attached to the patient or it


is too loose. The hose is not connected
to the module.

Check cuff and hose.

Cuff occlusion/NIBP cuff


occlusion

Tubes or hose are kinked.

Check tubes and hose.

Cuff overpressure

Cuff is squeezed during measurement


and pressure safety limits are
exceeded.

Check cuff, hose and tubes.

Check your power and reset the


settings.

The ECG cable is defective.

The ECG cable may be electrically


damaged or the connectors are wet.

Change the ECG cable and the lead


set.

Change the ECG module if the


module connector is wet. Keep the
connectors dry.

Avoid excessive use of liquids when


cleaning cables and connectors.

Reconnect the N-FREC or N-FCREC


module if you want to restart the
CO2 measurement.

Default settings returned


Faulty ECG cable

Gas measurements removed

Restart measurement.

InvBPs not zeroed

One or both InvBP channels have not


been zeroed.

Zero the channel indicated or zero


both channels.

Leads off

ECG trunk cable or any of the leads is


disconnected. Offset voltage between
two electrodes is too high.

ECG: Reconnect the disconnected


trunk cable, electrode or leadwire.

ECG: Change the trunk cable,


leadset and module.

NIBP measurement time too long.

NIBP: reduce patients motion.

Contact authorized service


personnel.

Long measurement time


Memory ERROR
Mode data reset

Error when loading settings from the


flash file.

Contact authorized service


personnel.

Network down: xxx

Network cable is not connected.

Check the network cable.

Check the NIBP setup and restart


autocycling.

Replace the batteries.

The Central is shut down.


NIBP manual
No battery backup

Autocycling mode is interrupted


because of an air leak or loose cuff.

Check the Central.

10-3

B40/B20 Patient Monitor

Message

Explanation

What to do

Noise

Unreliable HR calculation or distorted


waveform, possibly during
defibrillation or because of motion
artifacts.

ECG: Check the patient status.

Poor electrode quality or wrong


positioning.

Change the lead.

ECG: Check the electrodes.


Ensure that the patient is not
shivering.

Remove the ECG cable from the


connector and reinsert it.

No xx transducer

Transducer or channel x cable


disconnected.

Connect the transducer or the


cable.

No SpO2 probe

There is no SpO2 probe.

Check connection between the


SpO2 sensor and module.

Printer error

Printer is not working properly.

Check that the network printer is


operational.

Printing...

Printing network printer has started.

Please wait until the printing is


finished.

Printing ready

Monitor has completed sending


printing data to the printer.

Please wait until the printing is


finished.

Recorder module removed

There is no recorder module.

Reconnect the recorder module if


you need a recorder.

Recorder system error 29

Its a data transfer delay between


frame and recorder, the current
recording will stop also.

Press the Recorder Start/Stop key


again.

Replace Battery A
Replace Battery B

There is hardly any charge left in one


of the batteries. Also check the symbol
on screen.

Replace the battery as soon as


possible.

Replace D-fend

Water trap is partially blocked. This


decreases air flow.

Replace the water trap.

Sample line blocked

The sampling line inside or outside the


monitor is blocked, or the water trap is
occluded.

Change the sampling line and


water trap.

Select inflation limits

You are using a hose without an


automatic identification. When you try
to start the measurement, the monitor
goes automatically to the selection
NIBP Setup - Inflation Limits.

Select appropriate inflation limits.


NOTE: Auto option is not available
for these hoses.

SpO2 probe off

The finger or ear lobe may be too thin


or sensor halves are not aligned.

Check connection between sensor


and patient.

Replace the sensor.

10-4

Troubleshooting

Message

Explanation

What to do

Tachy

HR is equal to or above the higher

Check the patient status.

alarm limit.
Umbc Error

Umbc communication error

Contact authorized service


personnel.

Unable to measure Dia

Accurate diastolic pressure not


achieved because of artifacts, weak
pulsation etc.

Check the patient status.

Check the patient status.

Unable to measure Sys

Systolic pressure probably higher than


maximum inflation pressure, or
artifacts interfere in the systolic area.

Check the NIBP cuff placement.


Perform a new NIBP measurement.

Check the inflation limits.


Perform a new NIBP measurement.

Unstable zero pressure

Pressure is unstable at start of the


measurement.

Calm the patient and retry.

Weak pulsation

Weak or unstable oscillation signal


due to improper cuff position or
attachment, weak or abnormal blood
circulation, slow heart rate associated
with artifacts, moving or disturbed
patient during measurement, small air
leak.

Check the patient status.

Check that the cuff is not damaged.

x-Lead off

Check the leadwires and their


connections.

x high/low

Check the patient status.

Table 10-2

Adjust the alarm limits.

Other problems related to airway gas measurement

Symptom

Possible cause and solution

Airway gas values


are too low

Table 10-3

Check the NIBP cuff position and


attachments.

Check the sampling line and connectors for leakage.

Other problems related to arrhythmia measurement

Symptom

Possible cause and solution

Extra arrhythmia
alarms

The morphology of the ECG signal has changed.

Extra Ventricular
Fibrillations are
detected

Patients medical condition.

Start relearning manually through the ECG menu.


Check the patient status.

Low amplitude signal in some ECG leads.

Leads I and II: Select the one with the largest amplitude to ECG1.
After selecting the leads, start relearning manually.

10-5

B40/B20 Patient Monitor

Table 10-4

Other problems related to batteries

Symptom

Possible cause and solution

Battery operation
time is markedly
shortened

Table 10-5

Condition the batteries according to the instructions in this


manual.

Other problems related to ECG measurement

Symptom

Possible cause and solution

ECG signal is
noisy or no QRS is
detected

Ensure that the patient is not shivering.


Incorrect ECG filter.

Check the filter through ECG - ECG Setup - Filter.


Poor electrode quality or wrong positioning.

Check the electrodes and cables and their placement. See


section"ECG" for details.

Change the lead.

Check that:

The pacer markers have been selected ON.

The pacemaker has been adjusted correctly and not above R.

The pacemaker functions correctly: ECG cables, electrodes and


setups are correct.

Pacer markers are


not visible

Thick ECG
baseline

Remove the ECG cable from the connector and reinsert it.

ECG cable is looped.


Other electrical power cables are near the ECG leadwires.
Incorrect ECG filter.
Incorrect power frequency of the monitor.

Table 10-6

Symptom

Possible cause and solution

Respiration
measurement
fails

Check the electrode quality and positioning.

Other electrical devices may interfere with the measurement.

Table 10-7

10-6

Other problems related to impedance respiration measurement

Adjust the detection limits. During ventilator-supported breathing, the


respiration calculation may count only ventilator-produced
inspirations and expirations.

Other problems related to InvBP measurement

Symptom

Possible cause and solution

InvBP readings
seem unstable

Make sure there are no air bubbles in the transducer system. Flush
and zero.

Place the transducer on the patients mid-heart level and zero.

Troubleshooting

Table 10-8

Other problems related to NIBP measurement

Symptom

Possible cause and solution

NIBP
measurement
does not work or
values seem
unstable

Table 10-9

Check that cuff tubings are not bent, stretched, compressed or loose.
Prevent motion artifacts.
Use cuffs of correct size.

Other problems related to temperature measurement

Symptom

Possible cause and solution

Temperature
measurement
fails

Check that you are using a correct probe.


Try another probe.

Table 10-10 Other problems related to SpO2 measurement

Symptom

Possible cause and solution

SpO2 signal is
poor

Check the sensor and sensor positioning.


Change the averaging time from slow to normal.
Note that skin pigment causes differences.
Make sure that the patient is not shivering.

10-7

B40/B20 Patient Monitor

Other situations
The following table lists some other situations that may occur during monitoring and possible
explanations.

Table 10-11 Other operation problems

Symptom

Possible cause and solution

Printing is not
possible

Printer selection is None; change it through Print/Record - Printer


connection.

Printer is not connected to the network. Check printer cable.

The measured
values are not
displayed

Check that you have selected the desired parameter to a waveform


or digit field.

The monitor does


not start

Check that the power cord is properly connected.

Check that the measurement module is properly installed.

You cannot
perform a
measurement or
a function

10-8

Check that all cables are properly connected.


Check the fuses and replace them if necessary.

Check that all cables are properly connected.


Remove the module and reinstall it.

11 ECG

ECG

11 ECG
Safety precautions
Warnings

Make sure that the leadwire set clips or snaps do not touch any electrically conductive
material including earth.

Whenever patient defibrillation is a possibility, use non-polarizing (silver/silver chloride


construction) electrodes for ECG monitoring

The Display filter may alter the displayed ECG morphology. Do not make measurements
from the displayed or printed ECG when this filter is selected. Displayed ST values are
calculated before applying the Display filtering and may differ from values measured
from the displayed or printed ECG.

Single-use devices and accessories are not designed to be reused. Reuse may cause a
risk of contamination and affect the measurement accuracy.

Proper placement of defibrillator paddles in relation to the electrodes is required to


ensure successful defibrillation.

To ensure proper defibrillator protection, use only defibrillator proof transducers and
cables.

Set the pacemaker rate responsive mode off or turn off the impedance respiration
measurement on the monitor.

Do not use equipment for positioning (floating) temporary pacemaker leadwires,


performing pericardiocentesis, or other internal applications.

HEART RATE ALARM INTERFERENCE - Poor cable positioning or improper electrode


preparation may cause line isolation monitor transients to resemble actual cardiac
waveforms and thus inhibit heart rate alarms. To minimize this problem, follow proper
electrode placement and cable positioning guidelines provided with this product.

When using the electrosurgery unit, ensure proper contact of the ESU return electrode to
the patient to avoid possible burns at monitor measurement sites. Also ensure that the
ESU return electrode is near the operating area.

Patient's skin may become irritated after long contact with electrode gel or adhesive

Cautions

The impedance respiration measurement may cause rate changes in Minute Ventilation
Rate Responsive Pacemakers.

11-1

B40/B20 Patient Monitor

Overview
The electrocardiography, ECG, reflects the electrical activity generated by the heart muscle.
ECG monitoring is used for a heart rate measurement, for arrhythmia analysis and for
detecting pacemaker function and myocardial ischemia.
In Normal Screen, when measuring 5-lead ECG, you can simultaneously monitor the
waveforms of up to three different ECG leads. In 3-lead ECG, the monitor displays one ECG lead.
When monitoring the ECG, the monitor simultaneously analyzes ST segment changes.
.

ECG and
impedance
respiration
connector

Figure 11-1 Electrocardiographic (ECG) measurement connector

11-2

ECG

Displaying ECG and heart rate

Figure 11-2 Display of ECG and HR


(1)

ECG1 is displayed first

(2)

ECG2 is displayed below ECG1

(3)

ECG3 is displayed below ECG2

(4)

Selected lead label

(5)

ECG gain bar (1 mV reference)

(6)

Heart rate (HR) label

(7)

Heart rate/pulse rate source (ECG/Art/ABP/Pleth)

(8)

HR/PR value

(9)

Heart beat detector is flashing with every detected heart beat

(10) ST values are always displayed next to ECG2


(11) Fourth ST value, showing the largest absolute ST value
(12) Selected ST leads
(13) ST label
(14) Message field for parameter messages
NOTES:

You can change the number of ECG waveforms on the screen in the Monitor Setup menu
by selecting Screen Setup and Waveform Fields. The ECG leads can be chosen in the ECG
menu.

The input circuits are protected against the effects of electrosurgery and defibrillation.
However, the ECG waveform on the monitor screen may be disturbed during
electrosurgery.

11-3

B40/B20 Patient Monitor

Preparing the patient and placing the electrodes


Preparing the patient

Prepare the skin properly to ensure optimal signal quality.

Clean the skin using a mild soap and water solution.

Shave any hair from the electrode site. Gently rub the skin surface to increase capillary
blood flow and remove dead skin cells and oil.
Dry the skin completely before applying the electrodes.
Pre-gelled electrodes are recommended. Check that the electrodes are moist and have
not dried out during storage.

Placing the electrodes


When placing the electrodes, avoid bones close to the skin, obvious layers of fat and major
muscles.

Figure 11-3 Electrode positioning with 3-lead ECG

Figure 11-4 Electrode positioning with 5-lead ECG

11-4

ECG

Patient connection
NOTE: Keep the ECG cable, lead set and connectors dry. Avoid excessive use of liquids
when cleaning the cables and connectors.
L=YELLOW (IEC)
LA=BLACK (AAMI)

R=RED (IEC)
RA=WHITE (AAMI)

R=RED (IEC)
RA=WHITE (AAMI)

L=YELLOW (IEC)
LA=BLACK (AAMI)

LEAD I
L/ LA

LE
AD

LEAD III

R/RA

II

F/LL

C=WHITE (IEC)
V=BROWN (AAMI)
N=BLACK (IEC)
RL=GREEN (AAMI)

F=GREEN (IEC)
LL=RED (AAMI)

F=GREEN (IEC)
LL=RED (AAMI)

I I

R A L A

R L

R L

RA

L L

III

Figure 11-5 ECG setup


(1)

ECG connector

(2)

ECG trunk cable, or 3-lead ECG cable with integrated leadwires

(3)

3 or 5 leadwire set

ECG electrodes (pre-gelled electrodes are recommended). Check the expiration data.
NOTE: For a comprehensive list of accessories, see the Supplies and Accessories catalog.

Lead measurement
The following table lists the electrodes needed to measure different ECG leads:

Lead

Electrodes needed

R/RA, L/LA + F/LL or N/RL

II

R/RA, F/LL + L/LA or N/RL

III

LL/LA, F/LL + R/RA or N/RL

aVR

N/RL, R/RA, L/LA, F/LL

aVL

N/RL, R/RA, L/LA, F/LL

aVF

N/RL, R/RA, L/LA, F/LL

V5

N/RL, R/RA, L/LA, F/LL, C5

11-5

B40/B20 Patient Monitor

Color and letter coding


IEC standard
3-lead ECG

Position
on body
surface

5-lead ECG

Position
on body
surface

Position on
surface

R = red

right arm

R = red

right arm

right arm

L = yellow

left arm

L = yellow

left arm

left arm

F = green

left leg

F = green

left leg

left leg

N = black

right leg (neutral) right leg (neutral)

C = white

chest

4th intercostal
space at right
border of sternum

AAMI standard

11-6

3-lead ECG

Position
on body
surface

5-lead ECG

Position
on body
surface

Position on
surface

RA = white

right arm

RA = white

right arm

right arm

LA = black

left arm

LA = black

left arm

left arm

LL = red

left leg

LL = red

left leg

left leg

RL = green

right leg (neutral) right leg (neutral)

V = brown

chest

4th intercostal
space at right
border of sternum

ECG

ECG Setup menu


ECG

ECG

ECG Setup
Beat Sound Volume
HR Source

3
AUTO

Display with HR

None

Filter

Monit

Pacemaker

Hide

QRS width

Normal

Grid
5 - lead cable
V lead

OFF
5 elect
V5

Previous menu
Adjust volume of pulse / heart
beat sound.

Beat Sound Volume


Adjusts the pulse/heart beat sound volume (0 to 10). When the monitor has detected a QRS
complex or a pleth or pressure pulse, a beep tone is produced. You can raise, lower or turn off
the volume.
Adjusting beat sound volume affects also the beat sound volume of the HR source (ECG, Art or
Pleth).
When Pleth is monitored, the monitor provides a variable pulse beep, so that the tone of the
pulse beep rises with increasing oxygen saturation and falls with decreasing saturation. This
affects also the ECG tone.
HR Source

Selects the heart rate source (AUTO, ECG, Art, ABP or Pleth). When ECG is selected, HR is
always calculated from ECG. If the ECG signal is affected by too much noise for a reliable heart
rate calculation, pulse rate can be calculated from pressure (Art and ABP) or plethysmographic
pulse waveform. The selected heart rate source is displayed above the numerical display of the
heart rate. The color of the heart rate source indicator is the same as that of the source
parameter.
The AUTO selection priorities for heart rate calculation are: ECG, pressure (Art or ABP),
plethysmographic pulse waveform. The first heart rate source available is selected.

Display with HR Select PR, PVC or None to display combined heart rate and pulse rate or heart rate and PVC
rate next to the ECG waveform. The current HR source is displayed with bigger font and the
heart rate symbol flashes next to the reading.

Filter

Filters the ECG signal high frequency noise and slow respiratory artifacts.
Monit (monitor) filter effectively filters the artifacts caused by, for example, the electrosurgery
unit and respiration.
STfilt (ST filter) permits more accurate information of the ST segment. It filters the high
frequency artifacts caused by the electrosurgery unit but catches the slow changes in the ST
segment. The ST filter is more susceptible to baseline wander than the monitor filter.

11-7

B40/B20 Patient Monitor

Diagn (diagnostic) filter is used if more accurate information of the waveform is needed (for
example, of the P wave or AV block). The diagnostic filter is more susceptible to both high
frequencies and baseline wander than the monitor filter.
Pacemaker

Selects how to display the pacing spike of cardiac pacemaker. The selections are:
Hide = The pacing spike is filtered away from ECG data.
Show = The pacing spike is filtered away from ECG data but the spike is displayed as a
constant height marker.
Sensit = This selection uses a more sensitive pacemaker detection. Pacing spike is displayed on
ECG.

QRS width

Selects the QRS complex widths for ECG rhythms.


Normal option is suitable for adult/pediatric, between 70 to 120 ms.
Narrow option is suitable for neonatal, between 40 to 120 ms.

Grid

Selects the ECG gridlines to be displayed or not. If you select ON, you can view the ECG
waveforms over gridlines.

5-lead cable

Selects five or three electrodes.

V Lead

Selects the label for V lead according to the placement of the 5th electrode (see Placing the
electrodes on page 11-4).
NOTE: Selectable with the 5 lead trunk cable only.

Selecting a lead
The following lead selections are possible:

with 3 leadwire set: I, II or III

with 5 leadwire set: I, II, III, aVL, aVF, aVR or V

For channels ECG2 and ECG3, also a cascaded lead selection (Casc.) is available.

To select the ECG1 lead:

Press the ECG key and select ECG1 Lead.

You can select all leads (ECG1, ECG2 and ECG3) in the ECG menu.

ECG

The label of the lead is displayed in the ECG field.

11-8

ECG

Selecting user leads


1.

Press the ECG key.

2.

Select a lead for ECG1 -3 Lead.

With 3-lead ECG, you can select only one user lead (ECG1 Lead). With 5-lead ECG, you can
select three user leads.

Viewing a cascaded ECG


With a 3 leadwire set, ECG2 and ECG3 leads are automatically shown as cascaded. The same
ECG will be displayed in each waveform field. Thus more QRS complexes are displayed at the
same time.
With a 5 leadwire set either a cascaded ECG or different leads can be displayed. Select the
leads in the ECG menu.

To cascade a lead, press ECG and select ECG2/ECG3 Lead - Casc.

Adjusting the ECG size


Increasing or decreasing the ECG gain affects the size of the 1 mV bar at the left end of the ECG
waveform and the size of the ECG waveform accordingly.
ECG

11-9

B40/B20 Patient Monitor

Starting relearning manually


When the patients ECG pattern changes considerably, the monitor should start relearning a
new ECG pattern. The pattern changes, for example, when changing the patients position. To
start relearning manually:
1.

Press the ECG key.

2.

Select Relearn.

ECG

Setting heart rate alarm limits


To set the heart rate alarm limits:
1.

Press the ECG key.

2.

Select ECG Alarms.

3.

Select Adjust Limits.

You can also adjust the limits through Alarms Setup - Adjust Limits. For detailed instructions,
see section "Alarms."
ECG

11-10

ECG

Setting PVC alarm limits


To set the PVC alarm limits:
1.

Press the ECG key.

2.

Select ECG Alarms.

3.

Select PVC Alarm - ON.

4.

Select Adjust Limits.

You can also adjust the limits through Alarms Setup - Adjust Limits. For detailed instructions,
see section "Alarms."
ECG

11-11

B40/B20 Patient Monitor

ST segment analysis
Overview
The ST value, analyzed by the monitor, shows the difference of electrical activity between ISO
and ST points.
Myocardial ischemia appears in the ECG as an ST segment deviation from the isoelectric line
(ISO point). The ST segment generally rises above the PQ isoelectric line in the presence of
transmural ischemia and is pressed below the isoelectric line in the subendocardial ischemia.

NOTE: ST segment changes may also be affected by such factors as drugs or


metabolic and conduction disturbances.
NOTE: The significance of the ST segment changes needs to be determined by a
physician.

Display of ST

Figure 11-6 ST number field


(1)

Message field

(2)

ST values of ECG1, ECG2, and ECG3

(3)

Fourth ST value showing the largest absolute ST value

(4)

Lead label

NOTE: ST segment deviations are not displayed if the patient has a ventricular
pacemaker in use.

Monitoring the ST segment


The monitor analyzes ST for all measured leads and gives ST trends separately for each lead.
Numerical ST data is shown to the right of the second real-time ECG waveform field.
You can also select numerical ST data to the lower digit field. Press the Monitor Setup key
and select Screen Setup - Digit Fields. Select ST data to the field you prefer.
The ST analysis starts automatically after the leads have been connected and the QRS
detection has started. During a learning period of 32 accepted beats the median ST values are
displayed. Also, when the cable or the V lead is changed, or when an electrode is removed, or
relearning is started manually, the monitor starts to learn the ST segment.
NOTE: If the patient has an atrial pacemaker, ST calculations can be performed if the pacer
spike does not coincide with the ISO points adjustment range.

11-12

ECG

Setting the ST measurement points


Automatic setting of the J, ST and ISO points
The ST algorithm automatically searches for the J and ISO points. The distance between the ST
and J point is set according to the heart rate:

If the heart rate is less than 120 bpm, the ST point is set at J+80 ms.

If the heart rate is more than 120 bpm, the ST point is set at J+60 ms.

Setting the J, ST and ISO points manually


You can also set the J, ST and ISO points manually. If any of these points is manually set, the
other two are set at their current values.
Select Adjust ST and Set ISO point, Set J point or ST point.

Setting the J point manually


The J point is the point on the ECG trace where the S wave transitions to the ST segment.
Adjust the J point by turning the Trim Knob and confirm by pushing the Trim Knob. If the J
point setting is changed, the original point is shown as a dashed line. When you adjust
the J point, also the ST point is set according to the selected ST point setting (J+20, J+40,
J+60 or J+80).
The manually set J point remains until the monitor is turned off or reset. After a turn-off or
reset, the J point is set automatically.

Setting the ST point manually


ST segment is the component of the ECG trace between the end of the QRS complex and
the T wave.
Manual ST point adjustment sets the distance between ST and J points in ms. Selection
values are J+20, J+40, J+60, J+80.
The manually set ST point remains until the monitor is turned off or reset. After a turn-off
or reset, the ST point is set automatically according to the heart rate.

Setting the ISO point manually


The ISO point is on the isoelectric line. Adjust the ISO point by turning the Trim Knob and
confirm by pushing the Trim Knob. If the ISO point setting is changed, the original point is
shown as a dashed line.
The manually set ISO point remains until the monitor is turned off or reset. After a turn-off
or reset, the ISO point is set automatically.

11-13

B40/B20 Patient Monitor

Setting ST alarm limits


To set the ST alarm limits:
1.

Press the ECG key.

2.

Select ECG Alarms.

3.

Select ST Alarms.

4.

Select Adjust Limits.

You can also adjust the limits through Alarms Setup - Adjust Limits.
ECG

11-14

ECG

Description of the ST segment measurement algorithm


The ST segment begins at the point where the QRS ends (J point). Diagnostic criteria of ST
segment changes are measured at 60 ms after the J point. For monitoring purposes it is
important to keep the measurement point fixed during monitoring to notice the ST changes on
the respective trends.
The sophisticated algorithms of B40/B20 monitor search the J and isoelectric (ISO) points. The
system learns the ECG and stores the reference QRST complex. The algorithm sets the ISO and
J points. Due to the large variation with QRST complexes the user has possibility to adjust the
ST measurement points manually. The QRS analysis classifies each beat by using several
criteria and rejects distorted complexes from the ST calculation.
NOTE: The ST algorithm has been tested for accuracy of the ST segment data. The significance
of the ST segment changes needs to be determined by a physician.

Test results of ST segment measurement algorithm testing


The algorithm testing has been performed by using The European Society of Cardiology ST-T
Database (ESC DB).
Average results from ischemic ST detection:
Episode sensitivity

84%

Episode positive predictive accuracy

70%

Duration sensitivity

64%

Duration positive predictive accuracy

74%

11-15

B40/B20 Patient Monitor

Monitoring arrhythmia
Warnings

LOSS OR DETERIORATION OF ARRHYTHMIA DETECTION -Automated arrhythmia analysis


programs may incorrectly identify the presence or absence of an arrhythmia. A
physician must therefore interpret the arrhythmia information in conjunction with other
clinical findings. Please take special note of the following ECG waveform conditions:

Noisy waveforms. Noisy portions of ECG waveforms are typically excluded from
analysis. The exclusions are necessary to reduce the occurrence of inaccurate beat
interpretations and/or rhythm alarms. If the excluded noisy portions of the ECG
waveform contain true arrhythmia events, those events may remain undetected by
the system.

Beat amplitude and duration. Accurate detection and interpretation of beats


becomes increasingly difficult as the amplitude and/or duration of those beats
approach the design limits of the analysis program. Thus, as beats become
extremely wide or narrow, or especially as beats become small, arrhythmia
interpretation performance may degrade.

Other morphology considerations. Automated arrhythmia detection algorithms are


designed fundamentally to detect significant changes in QRS morphology. If an
arrhythmia event is present and does not exhibit a significant change from the
patients predominant morphology, it is possible for those events remain
undetected by the system.

SUSPENDED ANALYSIS - Certain conditions suspend arrhythmia analysis. When


suspended, arrhythmia conditions are not detected and alarms associated with
arrhythmias do not occur. The messages which alert you to the conditions causing
suspended arrhythmia analysis are: LEADS OFF and DISCHARGED.

Always monitor ECG for arrhythmia detection purposes. HR calculated from pulsatile
SpO2 waveform may differ significantly from ECG HR measured values. Users should be
aware that the SpO2 probe off and No SpO2 pulse technical alarms escalate no higher
than a Medium priority.

VFIB/VTAC should not be considered a substitute for the V TACH arrhythmia call.

The Severe arrhythmia analysis mode used by the B40/B20 detects asystole, bradycardia,
tachycardia, ventricular fibrillation and ventricular tachycardia.

11-16

ECG

Adjusting arrhythmia alarm settings


To open the adjustment menu:
1.

Press ECG.

2.

Select ECG Alarms - Arrh. Alarms.

In the adjustment menu, turn and push the Trim Knob to select the priority for all alarms except
asystole and ventricular fibrillation, which are always red (high priority), and ventricular
tachycardia, which cannot be selected OFF.

NOTE: Alarm priorities can also be set using the Central, depending on its
configuration.
ECG

Detecting the ECG arrhythmia alarms


Alarm

Criteria

Asystole

Cardiac arrest, no QRS complexes for five seconds.

Brady

HR below the HR alarm limit.

Tachy

HR over the HR alarm limit.

V Fib

Fibrillatory waveform caused by ventricular fibrillation.

V Tachy

Five or more consecutive PVCs and rate of successive beats


over 100 bpm.

11-17

B40/B20 Patient Monitor

Selecting leads for the arrhythmia analysis


When measuring 5-lead ECG, you can affect the selection of the two ECG leads used for
detecting beats and ventricular fibrillation. The selection of user leads (ECG1, ECG2, ECG3) on
the monitor affects the leads used for detection. The first lead used for detection is lead I or II.
The algorithm uses the lead appearing first in user leads. The second lead used for detection is
one of the precordial leads (V1-V6). The algorithm uses the precordial lead appearing first in the
user leads.
To change the user lead:
1.

Press the ECG key.

2.

Select a lead for ECG1 Lead, ECG2 Lead, ECG3 Lead.


The monitor starts relearning the new ECG pattern automatically.

NOTE: With the 3 leadwire trunk cable, the algorithm uses the only one available lead
ECG1 Lead, which is I, II or III, depending on the selected user lead.

Description of the arrhythmia algorithm


The bedside arrhythmia algorithm is based on template matching. A template is a group of
beats matching the same morphology. The algorithm detects QRS complexes, generates QRS
templates and performs beat labeling.
Parallel to this process there is an algorithm for detection of ventricular fibrillation. Detection of
ventricular fibrillation is based on waveform analysis.

NOTE: A physician must analyze the arrhythmia information in conjunction with other
clinical findings.

Test results of arrhythmia algorithm testing


The algorithm testing has been performed by using the following databases:

AHA (The American Heart Association Database for Evaluation of Ventricular Arrhythmia
Detectors)

MIT-BIH (The Massachusetts Institute of Technology- Beth Israel Hospital Arrhythmia


Database)

NST (The Noise Stress Test Database)

CU (The Creighton University Sustained Ventricular Arrhythmia Database)

Gross results for beat-by-beat detection (AHA)

11-18

Test

Gross

QRS Sensitivity

99.84%

QRS positive preductive


accuracy

99.88%

VEB sensitivity

95.24%

VEB positive predictive


accuracy

97.24%

VEB false positive rate

0.274%

ECG

Gross results for sample-by-sample detection of ventricular


fibrillation (AHA)
Test

Gross

VF Sensitivity (duration)

99%

VF positive preductive accuracy 100%


(episode)
The gross results are calculated as overall results of all records.
Abbreviations:
VEB

Ventricular ectopic beat

QRS

The waveform presented in an ECG during ventricular depolarization

VF

Ventricular fibrillation or ventricular flutter

ST

Segment of the ECG between the end of the QRS complex and the
start of the T-wave.

Monitoring pacemaker patients


Warnings

Do not diagnostically interpret pacemaker spike size and shape.

The shape of QRS complex may be changed so much because of the pacemaker that
QRS detection may be affected.

The monitoring of pacemaker patients can only occur with the pace program activated.

A pacemaker pulse can be counted as a QRS during asystole in either pace mode. Keep
pacemaker patients under close observation.

PACEMAKER PATIENTS - Rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter
alarms. Keep pacemaker patients under close surveillance. See "ECG specifications"for
the pacemaker pulse rejection capability of this instrument.

The monitor detects and rejects pacemaker pulses (see selection Pacemaker in the ECG Setup
menu). Sometimes this may lead to unnecessary asystole alarms.

NOTE: The shape of QRS complex may be changed because of the pacemaker so
much that QRS detection may be affected.
NOTE: If the patient has an atrial pacer, ST calculations can be performed if the pacer
does not coincide with the ISO points adjustment range.
NOTE: Pacemaker detector may not operate correctly during the use of highfrequency (HF) surgical equipment. The disturbances of HF surgical equipment
typically cause false positive pacer detection.

11-19

B40/B20 Patient Monitor

Other adjustable features


NOTE: This section describes the rest of the adjustable features related to ECG
measurement. To adjust the following features, you need a password. If you wish to
adjust the settings of the features, we recommend that you contact the person
responsible for the entire configuration.

ECG printout type


See "Changing printer settings" on page 6-4.

ECG waveform sweep speed


To change the waveform sweep speed, Monitor Setup - Sweep Speeds. Select
Hemodynamics and adjust the value.

Checklist
Check that:

11-20

Electrode gel is moist.


Electrodes have good skin contact.
Electrodes are positioned correctly.
Correct leadwire set is selected.
The trunk cable is connected properly.
Leadwire set is properly connected to the trunk cable.
Correct leadwire type is selected in the ECG Setup menu.
Pacemaker selection in the ECG Setup menu is Show when a pacemaker is used.
ECG is selected for screen through Monitor Setup - Screen Setup.

12 Impedance
respiration

Impedance respiration

12 Impedance respiration
Safety precautions
Warnings

Make sure that the leadwire set clips or snaps do not touch any electrically conductive
material including earth.

The monitor may not detect all episodes of inadequate breathing, nor does it distinguish
between central, obstructive and mixed apnea events.

ELECTRODE CONFIGURATION - Impedance respiration monitoring is not reliable when


ECG electrodes are placed on the limbs.

This device is not an apnea monitor system intended to alarm primarily upon the
cessation of breathing. In central apnea it indicates an alarm after a pre-determined time
since the last breath detection. Do not attempt to use it for detecting obstructive or mixed
apneas, since respiration movements and impedance variations may continue in these
cases.

The impedance respiration measurement is inherently very sensitive as it measures very


small physiologic signals (changes of impedance of the patient's chest area). Conducted
RF current above 1 Vrms may cause erroneous measurements at various frequencies, for
example interference with the signal/waveform leads to respiration rate readings
inconsistent with the patient's true respiration rate. If you notice this, use another form of
respiration monitoring. For further information, see the "Technical Reference Manual.

Electrical interference - Electrical devices, such as electrosurgery units and infrared


heaters, that emit electromagnetic disturbance may cause artifacts or disable the
respiration measurement completely.

Movement artifacts - Changing position, moving the head, moving the arms or shivering
may result in movement artifacts. Also the heart may cause noticeable movement and
sometimes this may interfere with the respiration measurement.

Intermittent mechanical ventilation - During spontaneous breathing the ventilator may at


times support the patients ventilation with an extra inspiration. If these ventilator
inspirations are substantially larger than the spontaneous breaths, the respiration
calculation may mistakenly count only the inspirations and expirations produced by the
ventilator.

The impedance respiration measurement may cause rate changes in Minute Ventilation
Rate Responsive Pacemakers. Set the pacemaker rate responsive mode off or turn off the
impedance respiration measurement on the monitor.

When using the electrosurgery unit, ensure proper contact of the ESU return electrode to
the patient to avoid burns at measurement sites. Also ensure that the ESU return
electrode is near the operating area.

Cautions

12-1

B40/B20 Patient Monitor

Overview
Impedance respiration is measured across the thorax. When the patient is breathing or is
ventilated, the volume of air changes in the lungs, resulting in impedance changes between the
electrodes Respiration rate is calculated from these impedance changes, and a respiration
waveform is displayed on the monitor screen.

ECG and
impedance
respiration
connector

Figure 12-1 Impedance respiration measurement connector

Respiration detection
The respiration rate is the sum of the respirations that have exceeded the detection limit.
The dotted lines present the zero line and the detection limit. The signal strength produced by a
respiration should thus exceed this minimum limit to be included in the respiration rate
calculation. Peaks within the grids are not calculated.

If the detection mode is AUTO, the grid lines present the minimum limits. The limits in use may
be larger. The RR value could include fewer respirations than indicated by the gridline.

12-2

Impedance respiration

Respiration rate calculation


Respiration rate is calculated automatically when ECG or CO2 is measured unless the
respiration measurement is turned off. When CO2 is measured, the respiration rate is
automatically calculated from CO2.
Respiration rate calculation switches back to impedance respiration if you press the Silence
Alarms key during an Apnea alarm.
The respiration rate source is displayed above the respiration numeric value.

Figure 12-2 Respiration rate source indication for impedance and CO2
The impedance respiration waveform is displayed next to the RR value also when the RR value
is calculated from CO2.
To manually select impedance respiration as respiration rate source:
1.

Select respiration in a waveform or a digit field.

2.

Press the Others key.

3.

Select Resp Setup.

4.

Select Resp Rate Source and select AUTO, CO2 or Imped.

Others

Displaying impedance respiration

Figure 12-3 Impedance respiration waveform and numeric respiration rate value.

12-3

B40/B20 Patient Monitor

Patient connections
The setup is the same as for the ECG measurement. For more information, see section ECG.
L=YELLOW (IEC)
LA=BLACK (AAMI)

R=RED (IEC)
RA=WHITE (AAMI)

R=RED (IEC)
RA=WHITE (AAMI)

L=YELLOW (IEC)
LA=BLACK (AAMI)

LEAD I
L/ LA

LE
AD

LEAD III

R/RA

II

F/LL

C=WHITE (IEC)
V=BROWN (AAMI)
N=BLACK (IEC)
RL=GREEN (AAMI)

F=GREEN (IEC)
LL=RED (AAMI)

F=GREEN (IEC)
LL=RED (AAMI)

I I

R A L A

R L

R L

RA

L L

III

Figure 12-4 Impedance respiration setup


(1)

ECG connector

(2)

Multi-link 5 lead standard cable

(3)

Multi-link leadwire set (3 or 5 leads)


ECG electrodes (pre-gelled electrodes are recommended). Check the expiration date.

NOTE: For a comprehensive list of accessories, see the "Supplies and Accessories"
catalog.

12-4

Impedance respiration

Activating measurement
Select respiration in a waveform or a digit field, otherwise the respiration information is not
included in the trends and the alarms are not operative.
Monitor
Setup

Screen Setup

Screen Setup

Waveform Fields

Dight Fields

Lower Field 1 ECG1

Lower Field 1

Resp

-More-

Lower Field 2 Resp

OFF

Lower Field 2

NIBP

CO2

Lower Field 3 IBP1


Lower Field 4 IBP2

ECG2

Lower Field 3

T1+T2

Resp

IBP1

Lower Field 4

CO2

Lower Field 5 Pleth


Lower Field 6 CO2

IBP2

Previous Menu

T1+T2
Temp
T1

Combine Pressures

NO

Previous Menu

Pleth
CO2

T2

Resp

Battery

OFF

OFF

Change waveform in field 2.


Select OFF to clear field.

or:

Change contents of lower digit


field 1. Select OFF to clear
field.

Check that the measurement is on:


1.

Press the Others key.

2.

Select Resp Setup.

3.

Select Measurement and ON.

Others

12-5

B40/B20 Patient Monitor

Improving waveform readability


To improve readability, increase the waveform size.
1.

Press the Others key.

2.

Select Resp Setup.

3.

Select Size and adjust the waveform size.


Others

The bar on the left side of the waveform always indicates a 1 reference.

12-6

Impedance respiration

Correcting the respiration number


Normally, we recommend the use of the AUTO detection limit. However, in some specific cases
you may wish to adjust the limits manually.

When the respirations are weak, you can manually adjust the detection limits
(measurement sensitivity) closer to each other to ensure that all respirations are included
in the RR value. In this case, the dotted line represents the absolute detection limits.

When there are lots of artifacts, the grids can be adjusted further apart to separate
smaller artifacts from larger, true respiration peaks. The small peaks fall within the grids
and are not calculated, while the bigger peaks cross the grids and are calculated as true
respirations.

To ensure the correct respiration number, adjust the limits closer to each other:
1.

Press Others.

2.

Select Resp Setup - Detection Limit and adjust the limit.


Others

The percentage is the ratio to the 1 reference bar which is 100%.

12-7

B40/B20 Patient Monitor

Measurement limitations
Movement artifacts
Changing position, moving the head, moving the arms or shivering may result in movement
artifacts. Also the heart may cause noticeable movement and sometimes this may interfere
with the respiration measurement.

Electrical interference
Electrical devices, such as electrosurgery units and infrared heaters, that emit electromagnetic
disturbance may cause artifacts or disable the respiration measurement completely.

Intermittent mechanical ventilation


During spontaneous breathing the ventilator may at times support the patients ventilation
with an extra inspiration. If these ventilator inspirations are substantially larger than the
spontaneous breaths, the respiration calculation may mistakenly count only the inspirations
and expirations produced by the ventilator. You can correct this by adjusting the detection
limits manually:
1.

Press the Others key.

2.

Select Resp Setup.

3.

Select Detection limit and adjust the limits.

Turning off the measurement


If the impedance respiration measurement signal interferes with other measurements, such as
ECG, you can turn it off.
1.

Press the Others key.

2.

Select Resp Setup.

3.

Select Measurement OFF.


Others

12-8

Impedance respiration

Checklist
Check that:

Electrode gel is moist.


Electrodes have good skin contact.
Electrodes are positioned correctly.
Correct leadwire set is selected.
The trunk cable is connected properly.
Leadwire set is properly connected to the trunk cable.
Correct leadwire type is selected in the ECG Setup menu.
ECG + Resp are selected for screen through Monitor Setup - Screen Setup.

12-9

B40/B20 Patient Monitor

12-10

13 Non-invasive
blood pressure

Non-invasive blood pressure

13 Non-invasive blood pressure


Safety precautions
Warnings

The NIBP parameter will not measure blood pressure effectively on patients who are
experiencing seizures or tremors.

Arrhythmias will increase the time required by the NIBP parameter to determine a blood
pressure and may extend the time beyond the capabilities of the parameter.

Do not apply external pressure against the cuff while monitoring. Doing so may cause
inaccurate blood pressure values.

Do not place the cuff on a limb being used for A-V fistulas, intravenous infusion or on any
area where circulation is compromised or has the potential to be compromised.

Accuracy of NIBP measurement depends on using a cuff of the proper size. It is essential
to measure the circumference of the limb and choose the proper size cuff.

The cuff size selected in the NIBP menu and the cuff size used must be correct to obtain
reliable NIBP data and to prevent overpressure in neonatal or pediatric use.

For SuperSTATTM NIBP Only - It takes one to three minutes for the NIBP parameter to
identify an irregular rhythm after ECG is connected. For patients with irregular rhythms,
simultaneous monitoring of ECG will enhance NIBP performance. Wait three minutes
after ECG has been connected and ECG heart rate is present on the monitor screen
before performing an NIBP determination.

NIBP READINGS MAY TIME OUT WHEN USING IABP An IABP balloon pump creates nonphysiological arterial waveforms. These waveforms create an oscillometric signal that
may not be interpreted by the NIBP algorithm, causing NIBP to time out. The patient
blood pressure can be monitored from the balloon pump device.

GE Healthcare monitors are designed for use with dual-hose cuffs and tubing. The use of
single-hose cuffs with dual hose tubing can result in unreliable and inaccurate NIBP data.

If a patients beat-to-beat pulse amplitude varies significantly (e.g., because of pulsus


alternans, atrial fibrillation, or the use of a rapid-cycling artificial ventilator), blood
pressure and pulse rate readings can be erratic, and an alternate measuring method
should be used for confirmation.

If Luer lock connectors are used in the construction of tubing, there is a possibility that
they might be inadvertently connectd to intravascular fluid systems, allowing air to be
pumped in to a blood vessel.

Devices that exert pressure on tissue have been associated with purpura, skin avulsion,
compartmental syndrome, ischemia, and/or neuropathy. To minimize these potential
problems, especially when monitoring at frequent intervals or over extended periods of
time, make sure the cuff is applied appropriately and examine the cuff site and the limb
distal to the cuff regularly for signs of impeded blood flow.

Cautions

13-1

B40/B20 Patient Monitor

Periodically check patient limb circulation distal to the cuff. Check frequently when using
Auto NIBP in one and two minute intervals. The one and two minute intervals are not
recommended for extended periods of time.

The monitor sets the inflation pressure automatically according to the previous
measurement. Reset the case or discharge the patient to reset the inflation limits before
measuring NIBP on a new patient.

Overview
The non-invasive blood pressure (NIBP) measurement uses the oscillometric measuring
principle. The cuff is inflated with a pressure slightly higher than the presumed systolic
pressure, then slowly deflated at a speed based on the patients heart rate, collecting data
from the oscillations produced by the pulsating artery. Based on this data, the unit calculates
values for systolic, mean and diastolic pressures.
Blood pressure measurements determined with this device are equivalent to those obtained by
an intra-arterial blood pressure measurement device. Multiple intra-arterial sites were used.
You can set the NIBP module into an automatic cycling mode to make measurements at
desired time intervals. You can also measure NIBP continuously for five minutes in STAT mode
or take separate single measurements.

NOTE: Intervals below 10 minutes and STAT measurements are not recommended for
extended periods of time.

NIBP
connector

Figure 13-1 Non-invasive blood pressure measurement connector

13-2

Non-invasive blood pressure

Direct function keys


There are two keys for starting NIBP on the Command Board:
NIBP
Auto ON/OFF

Starts and stops autocycling measurements.

NIBP
Start/Cancel

Starts a single measurement, and cancels any measurement.

Displaying non-invasive blood pressure


NIBP can be displayed in the digit field:

Figure 13-2 NIBP digit field display


(1)

Systolic and diastolic pressure value of non-invasive blood pressure

(2)

Label

(3)

Mean pressure value of NIBP

(4)

Time since the last autocycle measurement

(5)

NIBP autocycle time indicator

NOTE: When 60 minutes has passed from the latest NIBP measurement, the NIBP
numeric value digits turn gray. When 245 minutes has passed from the latest NIBP
measurement, the gray numeric value digits are replaced by a dashed line.

13-3

B40/B20 Patient Monitor

Patient connections

1
2

Figure 13-3 NIBP setup


(1)

NIBP connector in monitor

(2)

Cuff hose

(3)

Cuff of correct size

Place the arrow (4) over the brachial artery. Check that the index line (5) falls within the range
markings on the cuff, and wrap the cuff around the upper arm.
For a comprehensive list of accessories, see the Supplies and Accessories catalog. The listed
NIBP cuffs are latex-free.

Selecting a cuff and a cuff hose


Two different cuff hoses with different cuff connections are available:

BLACK hose for adults and pediatric (corresponding inflation limits A/P including cuff
identification.

LIGHT BLUE hose for neotatal (corresponding inflation limit NEO without cuff
identification.

To set up the inflation limits:


1.

Press the NIBP key and select NIBP Setup - Inflation Limits.

2.

Select the limit according to the hose with the Trim Knob.

NOTE: The monitor automatically identifies the setting of the inflation limits. When
using hoses without identification, the monitor goes to this selection automatically
when you try to start the NIBP measurement. You must set the inflation limits
manually. With these hoses, AUTO option is not available.
NOTES:

You cannot select A/P inflation limits when using an infant cuff.

The NIBP system incorporates a safety circuit to prevent overpressure or prolonged


inflation of the cuff.

The alarm limits change automatically according to the cuff hose type used.

To determine the correct cuff size, check that the index line on the outer edge of the cuff falls
between the range lines. If not, use a larger or smaller cuff.

13-4

Non-invasive blood pressure

Connecting the cuff hose


1.

Connect the cuff hose to the NIBP cuff by placing the opposite connectors in contact and
locking them together.

2.

Plug the NIBP cuff hose to the module.

NIBP Setup menu


NIBP

Ready Prompt Ready prompt gives an audible tone when the NIBP measurement is ready. Adjust the volume
of the beep tone from 1 (soft) to 10 (loud), or to 0 (OFF.)
Inflation Limits When this selection is Auto, the monitor automatically identifies the cuff hose and selects the
right inflation pressure and alarm limits for different cuff size.
This selection allows you to override the automatic safety limit feature for the hose/cuff being
used.
The selections are: Auto, A/P and NEO. The selection AUTO is not available when using hoses
without identification.

NOTE: You cannot select adult limits with an infant cuff hose.
NOTE: When using very large adult cuffs, use Adult limits to prevent Cuff loose
message from displaying.
Calibration Check
Enables the calibration check with an external manometer:
Remove cuff from the module connector before entering the menu. When the menu is
displayed, attach an external manometer with pump to the connector. Pump approximately
200 mmHg and compare the readings of the manometer and screen. If the difference is more
than 3 mmHg, calibration by authorized service personnel may be necessary.
Previous Menu Returns to the NIBP menu level.

13-5

B40/B20 Patient Monitor

Starting
Note that the measurement unit may be mmHg or kPa. The unit is selected during
configuration through Monitor Setup - Install/Service - Installation - Units.
You can start the NIBP measurement using either the direct function keys, or from the NIBP
menu by pressing the NIBP key.
NIBP

In the beginning of the measurement, sys and dia labels are replaced by the inflation limit
indication (Adult/Pediatric, Neonatal) for five seconds. The cuff pressure is displayed in the
mean pressure value field.
If motion artifacts are detected, the monitor automatically holds deflation until the motion
stops (maximum of 30 seconds). If the artifacts prevent proper measurement, a new
measurement starts automatically.
When the measurement is ready, you can hear a short beep and see the result numbers
flashing.

During measurement

13-6

Observe the cuffed limb frequently. Measurement may impair blood circulation. Intervals
below 10 minutes and STAT measurements are not recommended for extended periods
of time.

Make sure that the tubes are not bent, pressed or stretched. Measurement may be
impaired.

Blood pressure values may be affected by a change in the patient's position.


The presence of some arrhythmias during NIBP measurement may increase the time
required for the measurement.

Non-invasive blood pressure

Autocycling
The NIBP Auto On/Off key sets automatic NIBP measurement at selected intervals on and
off. To automatically measure NIBP at set time intervals, you must first set the cycle time
before setting the automatic measurements.
You also can configure a custom auto mode to meet the need of your clinical situation.
Autocycling is synchronized to real time, for example, if the first measurement was at 12.02,
the next measurement is at 12.05 and again at 12.10 (5 min. interval.)

Setting cycle time


The possible intervals for autocycling are 1, 2, 3, 4, 5, 10, 15, 20, 30, 60, 90, 120 minutes, and the
custom option.
To set the cycle time:

Press the NIBP key on the Command Board and select Cycle Time.
Select the alternative with the Trim Knob.

NIBP

NOTE: When you select the custom option, the NIBP measurements will follow custom
mode. You need configure or verify the custom modes set up before start the
automatic measurements.

13-7

B40/B20 Patient Monitor

Setting custom mode


You can set up to 4 separate steps for auto NIBP determinations by setting the time interval
and how many times this interval is repeated.
To set the custom mode:

Press the NIBP key.


Select Custom Setup.
Set up time interval and repeat times for every series.
The cycle time can be set to 1 to 120 min or OFF.
The repeat option can be set to 1 to 25 times or OFF or cont.

NIBP

Starting and stopping autocycling:


To start the autocycling, do one of the following:

Press the NIBP Auto On/Off key, or


press the NIBP key and select Start Cycling.

The bar at the bottom of the NIBP field shows the time remaining to the next
measurement.
To stop autocycling:

13-8

Press the NIBP Auto On/Off or


press the NIBP key on the Command Board and select Stop Cycling.

Non-invasive blood pressure

Starting and stopping a single manual measurement


To start the measurement, do one of the following:

Press the NIBP Start/Cancel key, or


press the NIBP key and select Start Manual.

To cancel any NIBP measurement, do one of the following:

Press the NIBP Start/Cancel key, or


press the NIBP key and select Cancel.

Starting and stopping a continuous measurement (STAT)


The STAT mode initiates continuous measurement for five minutes. A new NIBP measurement
starts immediately after the previous one.
After five minutes the monitor automatically returns to the previously selected cycling interval
or to manual mode.
To start and stop the measurement, do one of the following:

From the keyboard

To stop the measurement, press the NIBP Start/Cancel button again.

From NIBP menu

Press the NIBP key and select Start STAT.

To stop the measurement, select the Stop STAT.

13-9

B40/B20 Patient Monitor

Principles of SuperSTAT Noninvasive Blood Pressure


Determination
The oscillometric method of determining NIBP is accomplished by a sensitive transducer which
measures cuff pressure and pressure oscillations within the cuff. For the first determination
taken on a patient, the algorithm stores the pattern of the patient's oscillation size as a function
of the pressure steps. For subsequent manual (determined as such when the previous
determination is less than 16 minutes old), auto or stat determinations taken of the same
patient,
as few as four pressure steps may be necessary to complete the determination process. When
employing fewer pressure steps, the system uses the stored information from the previous
blood pressure determination to decide the best pressure steps to take. The algorithm
measures the consistency of pulse size to tell if the oscillations taken at a step are good and if
more steps are needed.
The first determination settles at an initial target pressure of 135 mmHg (adult mode) and 100
mmHg (neonate mode), depending on initial target pressure preset. To allow for rapid settling
of cuff pressure, the monitor will momentarily inflate to a higher pressure then immediately
deflate to the target pressure. After inflating the cuff, the NIBP parameter begins to deflate. The
oscillations versus cuff pressure are measured to determine the mean pressure and calculate
the systolic and diastolic pressures.
During an NIBP determination, the parameter deflates the cuff one step each time it detects
two pulsations of relatively equal amplitude. The time between deflation steps depends on the
frequency of these matched pulses (pulse rate of the patient). However, if the monitor is unable
to find any pulse within several seconds, it will deflate to the next step. The process of finding
two matched pulses at each step provides artifact rejection due to patient movement and
greatly enhances the accuracy of the monitor. In stat mode, some steps may require only one
pulse.

Figure 13-4 Full NIBP Determination Sequence for Adult


At each step the microprocessor stores cuff pressure, the matched pulse amplitude, and the
time between successive pulses. The stepped deflation and matched pulse detection continues
until diastolic pressure is determined or total cuff pressure falls below 8 mmHg. The parameter
then deflates the cuff (to zero detected pressure), analyzes the stored data, and updates the
screen.

13-10

Non-invasive blood pressure

The operating cycle is composed of four parts: inflation time, deflation time, evaluation time,
and wait time. Wait time, which varies from mode to mode, is affected by the cycle time (auto
mode) or operator intervention (manual mode). The figure shows the basic operating cycle for
an NIBP determination.

Figure 13-5 NIBP - Auto Mode

Systolic Search
If systolic pressure is not found, the NIBP parameter can search at cuff pressures higher than
the initial target pressure. The parameter will inflate the cuff above the initial target pressure to
get better data in the systolic region. The maximum pressure allowed in systolic search is
limited by the normal range for cuff pressures.
In any operating mode, if a patient's systolic pressure exceeds the inflation pressure, the
parameter will begin normal deflation sequence, detect the absence of a systolic value, stop
deflation, reinflate to a higher (than initial) inflation pressure (290 mmHg maximum in adult
mode; 145 mmHg in neonatal mode), and resume the normal deflation sequence.
If it has been 16 minutes or less since the last determination and the current blood pressure is
similar to the previous reading, the monitor will try to make an accelerated determination of
blood pressure. During irregular rhythms, only pulses from the current determination are used
in calculating the blood pressure values. In order to ensure adequate artifact rejection
capability and optimal SuperSTAT NIBP performance, several criteria used to match and
qualify the oscillometric pulses at each pressure step are relaxed while supplementing the
criteria with additional information from ECG.

NOTE: (Adult/Pediatric only) When ECG is monitored, SuperSTAT NIBP is able to


determine blood pressure in the presence of irregular heart rhythms. At the beginning
of a SuperSTAT NIBP determination, the coefficient of variation from the previous 120
ECG R-R intervals is used to determine if an irregular rhythm is present.
Accuracy of the NIBP parameter was validated against the intra-arterial method. Do not use
the auscultatory method to verify the accuracy of the NIBP parameter. Auscultatory method
(using cuff and stethoscope) calculates the mean pressure value from audible sounds at
systolic and diastolic, but the NIBP parameter method detects all three values.

13-11

B40/B20 Patient Monitor

Automatic NIBP double check


If the NIBP value exceeds the alarm limits, a low level alarm will be given and the auto
measurement takes place, if the alarm situation persists, the alarm will be raise to yellow level.

Checklist
Check that:

13-12

The hose is correct: grey NIBP hose for adults and pediatric and light blue for neonatal.
The O-ring on the hose connector is intact.
The connector is firmly pushed inside the cuff tube.
The NIBP hose is properly connected to the module and will not detach if pulled.
The NIBP cuff is correct for the patient size.
There are no holes or cracks in the cuff bladder or cuff tube.
The symbol indicating the center of the bladder is over the artery.
All residual air is squeezed out of the cuff before wrapping it around the arm.
The cuff is not loose.
The cuff is at heart level.
The cuff tubes or NIBP tube are not kinked or squeezed together.
Non-invasive blood pressure is selected to be displayed through Monitor Setup Screen Setup.

14 Pulse oximetry

Pulse oximetry

14 Pulse oximetry
Safety precautions
Warnings

Allow sensor and cable to dry completely after cleaning. Moisture and dirt on the
connector can affect the measurement accuracy.

To prevent erroneous readings, do not use physically damaged sensors, cables or


modules. Discard a damaged sensor or cable immediately. Never repair a damaged
sensor or cable; never use a sensor or cable repaired by others. A damaged sensor or a
sensor soaked in liquid may cause burns during electrosurgery.

Inaccurate SpO2 data can result if a sensor is past its useful life. Therefore, re-evaluate the
measurement periodically by performing additional assessment of the patient and
equipment, including consideration of use of alternate monitoring methods such as direct
measurement of arterial oxyhemoglobin saturation (SaO2).

Conditions that may cause inaccurate readings and impact alarms include interfering
substances, excessive ambient light, electrical interference, ventricular septal defects
(VSD), excessive motion, low perfusion, low signal strength, incorrect sensor placement,
poor sensor fit, and/or movement of the sensor on the patient.

Physiological characteristics of the patient can affect the SpO2 signal and readings.
NEONATAL - The display of inaccurate pulse oximetry (SpO2) values has been linked to the
presence of poor signal strength or artifact due to patient motion during signal analysis.
This condition is most likely to be encountered when the monitor is used on neonates or
infants. These same conditions in adults do not impact the SpO2 values to the same
extent.
We recommend the application of the following criteria when using the pulse oximetry
function on neonates and infants:

The peripheral pulse rate (PPR) as determined by the SpO2 function must be within
10% of the heart rate, and

The SpO2 signal strength should be adequate. This is indicated by the display of two
or three asterisks or the absence of a Low Signal Quality message.

Procedures or devices previously applied in your facility for SpO2 monitoring should be
used in the event the SpO2 value from the monitor cannot be validated by the above
criteria.

If the accuracy of any measurement does not seem reasonable, first check the patients
vital signs, then check for conditions that may cause inaccurate SpO2 readings. If the
problem is still not resolved, check the SpO2 module or sensor for proper functioning.

Oximetry performance may be impaired when patient perfusion is low or signal


attenuation is high.

The operator is responsible for checking the compatibility of the pulse oximetry monitor,
sensor, and patient cable prior to use. Incompatible components can result in degraded
performance and/or device malfunction.

SpO2 readings may be inaccurate for a short time after defibrillation.

14-1

B40/B20 Patient Monitor

Change the sensor site immediately if there is evidence of blistering, skin erosion, or
ischemic skin necrosis (such as skin discoloration or reddening). Otherwise, change the
site every four hours.

A pulse oximeter should not be used as an apnea monitor.

Interfering substances can affect the SpO2 reading.

Prolonged monitoring or incorrect sensor application can cause skin irritation or impaired
circulation. Recommend checking probe site every four hours (more frequently for poor
perfusion or neonate). Refer to instructions supplied with sensor.

A pulse oximeter should be considered an early warning device. As a trend toward


patient deoxygenation is indicated, blood samples should be analyzed by a laboratory
CO-oximeter to completely understand the patient's condition. Check that the pulse
oximetry waveform is physiological in shape to ensure waveform quality and minimize
noise spikes caused by motion conditions. (Not applicable when monitoring SpO2 with
Masimo SET technology).
Improper sensor placement can affect the SpO2 signal and readings.
Do not allow tape to block the probe light detector.
During electrosurgery the SpO2 measurement results may be incorrect.

Cautions

Overview
Oxygen saturation SpO2 is the percentage of saturated hemoglobin compared to total
hemoglobin measured by a two wavelength pulse oximeter (also called functional or In Vivo
oxygen saturation.
The SpO2 value is measured by light absorption technique: Red and infrared light is emitted
from the emitter side of the sensor. The light is partly absorbed when it passes through the
monitored tissue. The amount of transmitted light is detected in the detector side of the sensor.
When the pulsative part of the light signal is examined, the amount of light absorbed by
arterial hemoglobin is discovered and the saturation level can be calculated.
The plethysmographic pulse wave is derived from variations of the intensity of the transmitted
light and reflects the blood pulsation at the measuring site. Thus the amplitude of the
waveform reflects the perfusion.
The B40/B20 monitor have three options for SpO2 configuration: GE, Masimo and Nellcor. The
set up will be preconfigured by the manufacturer according to your choice. Different setup
may have different performance, please refer to "Technical specification"for more details.
For Masimo SpO2, possession or purchase of this device does not convey any express or
implied license to use the device with unauthorized sensors or cables which would, alone, or in
combination with this device, fall within the scope of one or more of the patents relating to this
device.
Masimo SpO2 is covered under one or more of the following U.S.A. patents: 5,758,644,
6,011,986, 6,699,194, 7,215,986, 7,254,433, 7,530,955 and other applicable patentslisted at:
www.masimo.com/patents.htm

14-2

Pulse oximetry

Pulse
oximetry
connector

Figure 14-1 Pulse oximetry measurement connector

Display of pulse oximetry


1

2
3

Figure 14-2 Display of SpO2 value and pleth waveform


(1)

Scale of plethysmogram

(2)

Label

(3)

Oxygen saturation (SpO2) value

(4)

Pulse oximetry message field

14-3

B40/B20 Patient Monitor

Patient connections
OXY-E-UN

OXY-SE-3

OXY-F-UN

OXY-W-UN

OXY-AP

OXY-AF

Figure 14-3 Pulse oximetry setup


(1)

Compatible SpO2 measurement capability

(2)

Interconnect cable

(3)

Reusable sensors

(4)

Disposable sensors

NOTE: The listed sensors are latex-free.


NOTE: For a comprehensive list of accessories, see the "Supplies and Accessories"
catalog delivered with the monitor.
NOTE: For each SpO2 accessory, refer to the instructions for use in the accessory
package for patient weight limits.

Connecting the patient


NOTE: Use dry and clean sensors only.
1.

Connect the sensor to the blue connector in the module.

2.

Clean the application site: Remove nail polish, artificial fingernails, earrings etc., clip long
fingernails.

3.

Position the sensor correctly. For proper sensor positioning, see the Instructions for use
accompanying each sensor.

4.

Attach the sensor to the patient.

5.

Attach the sensor cable to the wrist or bed clothes to prevent the cable and sensor from
moving.

The message Pulse Search is display in the message field. After the pulse search is completed.
the plethysmographic pulse waveform and the SpO2 reading are displayed on the screen.

14-4

Pulse oximetry

Pulse oximetry menu


SpO2

Pleth Scale (GE SpO2 configuration only)


If the pleth scale selection is AUTO, the scale of the plethysmographic waveform display is
automatically set during Pulse Search. The scale is then kept constant throughout the case to
enable easy detection of changes in the patient's perfusion. To adjust the pleth scale further,
select one of the following: 2, 5, 10, 20, 50, Auto.
With automatic AUTO scaling mode, the scale changes automatically if the amplitude of the
pleth waveform either exceeds the current scale or falls below the maximum value of the next
lower scale by 10% for 30 seconds or more. When the scale changes, the message Scale
changed is displayed.
The scale indicator number is displayed on the left side of the waveform.
SpO2 Response (GE SpO2 configuration only)
Selects the SpO2 averaging time. The selections are Normal: 12 seconds (default setting), Fast:
3 seconds.
Beat Sound Volume
Adjusting the SpO2 beat volume also affects the beat volume of ECG.
When SpO2 is monitored, the monitor provides a variable pulse beep so that the tone of the
pulse beep rises with increasing oxygen saturation and falls as saturation decreases. You can
select the volume from 1 (soft) to 10 (loud). With selection 0 there is no audible sound.
HR Source

Selects the heart rate source. If the ECG signal is affected by too much noise for a reliable heart
rate calculation, heart rate can also be calculated from invasive pressure (Art/ABP) or
plethysmographic pulse waveform Pleth. The selected heart rate source is displayed above the
numerical display of the heart rate. The color of the heart rate is the same as that of the source
parameter.
The AUTO selection priorities for heart rate calculation are: ECG (the lead with highest R-wave),
pressure (Art/ABP) and plethysmographic pulse waveform.

SpO2 Alarm

Alarms can be adjusted in this menu or in the Alarms Setup menu which opens by pressing the

Alarms Setup key.

14-5

B40/B20 Patient Monitor

During monitoring
Patient condition or prolonged use may require changing the sensor site periodically. Check
skin integrity, circulatory status and correct alignment and change sensor site at least every
four hours.
For patients with poor peripheral blood circulation or sensitive skin, change the site at intervals
of 30 minutes to one hour. To confirm the circulatory status, observe the size of the
plethysmographic waveform with a fixed pleth scale. Take special care of this when monitoring
small children.
If possible, do not attach the SpO2 sensor on a limb that is used for NMT measurement or for
administrating cold infusions.
NIBP measurement and arterial blood pressure measurement
To avoid erroneous readings, do not use a blood pressure cuff or arterial blood pressure
measurement device on the same limb as the sensor.
Plethysmographic pulse wave
To get an optimal pulse wave use smaller scale indicators when using measuring sites with
poor perfusion. A small pulse wave may be a sign of impaired circulation that may require
increased attention.
Higher scale indicators together with a well defined pulse wave indicate strong circulation and
a relaxed patient.

Removing the sensor

14-6

Open the sensor. Do not pull the sensor from its cable.

Pulse oximetry

Measurement limitations

The B40/B20 monitors are designed to minimize the interference of electrosurgery. Under
some circumstances electrosurgery may cause noise on the screen. Therefore, be careful
in interpreting the results, especially the plethysmographic pulse waveform, during
electrosurgery.

The saturation values may be somewhat higher for smokers. Special care should be
taken with patients who have burns or carbon monoxide (CO) intoxication. When carbon
monoxide intoxication is suspected, always confirm the pulse oximetry reading with a
blood sample measurement.

Intravascular dyes may cause erroneous readings. For example, methylene blue, indigo
carmine, indocyanine green or any substances that contain dyes, interfere with the SpO2
measurement.

Vasoconstrictive drugs, such as phenylephrine hydrochloride and dopamine, may affect


the accuracy of the measurement.

The pulse oximeter cannot distinguish between oxyhemoglobin and dyshemoglobins, for
example, met- or carboxyhemoglobins.

Poor perfusion may affect the accuracy of measurement when using the ear probe.

For Masimo SpO2, loss of pulse signal can occur when:

To avoid erroneous readings, do not use a blood pressure cuff or arterial blood pressure
measurement device, or NMT sensor in the same limb as the SpO2 sensor.

Sensor is too tight

The patient has hypotension, severe vasoconstriction, severe anemia or


hypothermia.

There is arterial occlusion proximal to the sensor.

Checklist
Check that

Only one module with pulse oximetry measurement is inserted.


Correct SpO2 sensor is selected for each patient size.
The sensor is completely dry after cleaning.
Sensor or interconnect cable is plugged to the blue connector in the module.
Sensor is properly connected to the interconnection cable if used.
Sensor is positioned correctly to the patient.
SpO2 is selected for screen through Monitor Setup - Screen Setup.

14-7

B40/B20 Patient Monitor

14-8

15 Invasive blood
pressure

Invasive blood pressure

15 Invasive blood pressure


Safety precautions
Warnings

All invasive procedures involve risks to the patient. Use aseptic technique. Follow catheter
manufacturer's instructions.

Make sure that no part of the patient connections touches any electrically conductive
material including earth.

Mechanical shock to the invasive blood pressure transducer may cause severe shifts in
zero balance and calibration, and cause erroneous readings.

When initializing the IBP parameter, Invasive blood pressure alarm activation criteria may
result in inactive limit alarms.

Overview
You can measure and monitor two invasive blood pressures at the same time using a dual
invasive blood pressure cable. To measure invasive blood pressure you need the IBP
configuration in monitor.
During the invasive blood pressure measurement, the transducer converts pressure variations
into electrical signals. The electrical signals are amplified and displayed as numeric pressure
values and waveforms.

Invasive pressures
connector

Figure 15-1 Invasive pressure measurement connector

15-1

B40/B20 Patient Monitor

Direct function keys


There is a key for zeroing IBP on the Command Board:
IBP
Zero All

Zeros all pressure transducers.

NOTE: Selecting Zero ALL does not zero ICP. Zero it separately.

Display of invasive blood pressure


2

Figure 15-2 InvBP display


(1)

Invasive blood pressure waveforms with zero and reference lines

(2)

Selected pressure label

(3)

Selected pressure scale

(4)

Field for messages and alarm limit settings

(5)

Systolic, diastolic and mean pressure values of invasive blood pressures

You can have a combined display of all those waveforms that are selected on the screen. This
combined display uses the whole waveform field area and the same zero line for all
waveforms.
To select:

15-2

1.

Press the Monitor Setup key and select Screen Setup.

2.

Select Waveform Fields.

3.

To combine all waveforms in one field, select Combine Pressures and Yes.

Invasive blood pressure

Patient connections
1.

Connect the pressure transducer to the transducer adapter cable. Connect the cable to
the red connector in the module, or to the dual invasive blood pressure adapter cable.

NOTE: Invasive pressures need to be zeroed after reconnecting the pressure


transducer or cable, and whenever the patients position is changed. The transducer is
always leveled to the mid right atrium.
2.

Prepare the transducer kit according to the manufacturers instructions. Mount the kit
with the transducer zeroing port at mid-heart level.

3.

Ensure that there is no air in the line. Refer to transducer manufacturers instructions on
how to remove trapped air from the transducer.

4.

Connect the patient catheter to the pressure line.

5.

Open the dome stopcock to room air.

6.

Zero the transducer. See "Starting with accurate values" on page 15-4.

7.

Open the dome stopcock to pressure catheter and check the quality of the waveform.

4
5

7
1
6

Figure 15-3 Invasive blood pressure setup


(1)

Compatible InvBP measurement capability

(2)

Heparinized fluid bag with pressure infusor

(3)

Flushing set

(4)

Disposable catheter

(5)

Transducer

(6)

Adapter cable for the InvBP transducer

(7)

Adapter cable for dual InvBP measurement

NOTE: Patient connections made according to the picture above using approved accessories
are defibrillator-proof.

15-3

B40/B20 Patient Monitor

Starting with accurate values


Pressure transducers generally produce a small signal even when no pressure is applied to
them. It is necessary to zero the monitor with the transducer to establish an accurate electrical
zero point.
Also, the position of the transducer effects the accuracy of the measurement. An error of 10
mmHg of static pressure is introduced for every 13.6 cm (5.4 inches) difference in height
between the mid-heart and the transducer.
To zero the transducer, open the transducer to air and do one of the following:

Press the IBP Zero All key. This starts zeroing immediately.

Or:

Press the IBP key on the Command Board and select Zero Pressures.

IBP

During the zeroing process, the message Zeroing is displayed. After the transducer is zeroed,
the message Zeroed is displayed in the digit field. After each channel is zeroed, the time of
zeroing is displayed in the menu.

NOTE: Selecting Zero ALL does not zero ICP. Zero it separately.
NOTE: Check zero level after power interruptions.
NOTE: Invasive pressures need to be zeroed after reconnecting the pressure
transducer or cable, and whenever the patients position is changed. If all channels
have not been zeroed, the message InvBP not zeroed appears. However, the alarms
advance to yellow and red levels regardless of zeroing.

15-4

Invasive blood pressure

Invasive Pressures menu


IBP

Zero Pressures Opens a menu to zero both pressures or one of them.


Ventilation Mode Respiration causes artifacts in invasive pressures. At the end of expiration the artifact is at its
smallest. Select Spont for spontaneous respiration and Contr for controlled ventilation.

IBPx Setup menu


Label

Start-up labels are Art and CVP. Other labels are IBP1, IBP2, PA, RAP, RVP, LAP, ICP, ABP, UAC
and UVC. Assigning the appropriate label automatically changes other pressure settings
accordingly.

Scale

Scales are assigned by the monitor when the channel is labeled. Scales can also be individually
adjusted between 10 and 300 mmHg in steps of 10.

Digit Format

With the numeric display format you can choose either the Systolic/Diastolic numbers (S/D) or
the Mean pressure value (Mean) in large size to the screen. You can also choose all values (S/D/
M) to the screen. If the label is ICP, also the CPP selection is available (see also page 15-7).

Response

Use the Trim Knob to change the invasive blood pressure averaging time. The available values
are Normal and B-TO-B (beat-to-beat). With selection Normal, normal averaging is used.
Depending on the label, the values are updated approximately every five seconds. With
selection B-TO-B, no averaging is used and the values of the last detected pulse are displayed.
These values can change up to three times per second. This feature is useful when it is
necessary to detect fast pressure changes. NOTE: This setting affects only the displayed
values, not the averaging of invasive pressure trends.

Filter FrequencyMeasured signal is filtered to remove noise and artifacts. Use the Trim Knob to adjust the filter
between 4 and 22 Hz.
HR Source

IBPX Alarm

If the ECG signal is affected by too much noise for a reliable heart rate calculation, heart rate
can be calculated mechanically from the pressure (Art) or plethysmographic pulse waveform.
The selected heart rate source is displayed above the numerical display of the heart rate. The
color of the heart rate is the same as that of the source parameter. Auto selection prioritizes
the heart rate calculation in a specified order: ECG (the lead with highest R-wave), pressure
(Art), and plethysmographic pulse waveform.
Alarms can be adjusted in this menu or in the Alarms Setup menu which opens by pressing the

Alarms Setup key.

15-5

B40/B20 Patient Monitor

Determining pressure values visually


By moving the horizontal cursor across the pressure waveform, you can get accurate pressure
values at selected points. This may be useful, for example, if the patient's breathing pattern is
irregular. The cursor is not available for pressures shown with a combined scale.
1.

Press the IBP key.

2.

Select Px Setup - Px Cursor.

3.

Move the cursor up or down by turning the Trim Knob. Every time the cursor is moved, the
time (hours and minutes) and pressure values are displayed on the screen. This way, you
can keep track of the changes made.

4.

You can remove the cursor by selecting Remove cursor. Note that if the cursor is not
removed, it remains visible in the Normal Screen.

Labeling channels
The label of the pressure channel sets its display scale, color, filter, alarm source and alarm
limits. The label descriptions are preconfigured.
To change the label:
1.

Press the IBP key.

2.

Select IBP1 Setup.

3.

Select Label.

IBP

The labels are the following:


IBP1, IBP2

15-6

Standard labels

Art

Arterial pressure

CVP

Central venous pressure

PA

Pulmonary arterial pressure

RAP

Right atrial pressure

RVP

Right ventricular pressure

LAP

Left atrial pressure

ICP

First intracranial pressure

ABP

Arterial blood pressure

UAC

Umbilical artery catheters

UVC

umbilical venous catheters

Invasive blood pressure

Both Art and ABP labels are available for situations when two arterial lines are desired but you
want to use different settings or alarm labels.

Table 15-1

LABEL

Invasive blood pressure labels and descriptions

P1, Art, P2, CVP


ABP

Scale mmHg/kPa

RAP,
LAP

ICP

PA

RVP

UAC

UVC

200/30

20/4

20/4

20/4

60/8

60/8

100/14

10/4

Color

Red

Blue

White

White

Yellow

White

Red

White

Alarm source

Sys

Off

Off

Off

Off

Off

Sys, Dia,
Mean

Mean

Digit format

S/D

Mean

Mean

CPP

S/D

S/D

S/D

Mean

Filter (Hz)

22

14

14

Response

normal

normal

normal

normal

normal

normal

normal

normal

Cerebral perfusion pressure


Cerebral Perfusion pressure CPP is calculated by subtracting ICP mean pressure from Art mean
pressure.
CPP mean is displayed next to the ICP value.

NOTE: Set the Label of the channel to ICP, and the Digit Format to CPP.

Adjusting alarm sources and limits


You can adjust or turn off pressure alarm limits in the Alarms Setup/Adjust Limits menu. Enter
the menu by pressing the Alarms Setup key.
For each pressure channel you can choose as a source Sys (Systolic), Dia (Diastolic), Mean or
Off. Note that you can choose one or both sources to be active at a time.

15-7

B40/B20 Patient Monitor

Alarms
Setup

For more information about alarms and adjusting, see section "Alarms."

Smart InvBP and flushing


Flushing is performed to keep the lines open. It prevents blood from clotting and occluding the
lines and measurements. Infusion that is used for flushing goes through the dome into the
patients artery.
Two types of flushing are used simultaneously: flushing with continuous infusion and manual
flushing. Flushing with continuous infusion uses a higher pressure than the patients blood
pressure and contains Heparin to prevent blood from clotting. Manual flushing is always used
after having taken a blood sample, and every now and then to ensure that the lines remain
open.
The monitor detects the pressure used for infusion to flush the invasive blood pressure line.
There is a 40 second time-out for performing the flushing. The digits change to dashes, ---/--- ,
and there are no alarms. After 40 seconds, or when flushing is completed and patients pulse
detected, the pressure digits are displayed and alarms become active. If pressure values are
outside the alarm limits, an alarm is given.
The monitor functions the same way when blood samples are taken.

Checklist
Check that:

15-8

Invasive blood pressure transducer cable is plugged to the adapter cable, and this is
connected to the red connector in the module.

Pressure transducer is connected to the cable.

Pressure transducer is zeroed.

Patient catheter is connected to the pressure line.


There is no air in the transducer dome or catheter line and transducer is at mid heart
level.
Invasive blood pressure is selected for screen through Monitor Setup - Screen Setup.

16 Temperature

Temperature

16 Temperature
Safety precautions
Warnings

Temperature measurement response time is affected by use of esophogeal stethoscope


with certain temperature sensors.

Overview
You can simultaneously measure and monitor temperature of two sites with TEMP
configuration.
As a measuring probe use only GE Healthcare temperature probes or defibrillator-proof
YSI 400 series probes. You can measure, for example, esophageal, nasopharyngeal, rectal, and
skin temperature.

NOTE: Monitoring of perioperative body temperature is recommended when inducing


hypothermia or if unexpected temperature changes occur.

Temperature
connector

Figure 16-1 Temperature measurement connector

16-1

B40/B20 Patient Monitor

Displaying temperature

Figure 16-2 T1+T2 digit display


(1)

Labels

(2)

Temperature measurement value

(3)

Calculated T2-T1 difference

The other options are individual T1 and T2 readings in a digit field.

Patient connections
3

3
2
1
5
4

Figure 16-3 Temperature measurement setup


(1)

Compatible temperature measurement capability

(2)

Adapter cable for temperature probes

(3)

Reusable temperature probe

(4)

Adapter cable for disposable temperature probe

(5)

Disposable temperature probe

To connect the patient:


1.

Attach the temperature probe to the patient.

2.

Connect the adapter cable to the module connector.

NOTE: For a comprehensive list of accessories, see the "Supplies and Accessories"
catalog.

16-2

Temperature

Temp Setup menu


Others

T1, 2 Label

Allows you to label each temperature measurement site.

Unit

Allows you to select the units to be either degrees Celsius or degrees Fahrenheit.
Allows you to adjust temperature alarm limits for two measurement sites, measured by T1 or
T2. You can choose one or both sources to be active at a time.

Temp Alarm

Alarms can be adjusted in this menu or in the Alarms Setup menu which opens by pressing the

Alarms Setup key.


Previous Menu Returns to the previous menu level.

Changing temperature label


1.

Press the Others key.

2.

Temp Setup - T1 Label or T2 Label.

3.

Select the label with the Trim Knob. Labels are T1 and T2, or:
Eso (Esophageal temperature)
Naso (Nasopharyngeal temperature)
Tymp (Tympanic temperature)
Rect (Rectal temperature)
Blad (Bladder temperature)
Axil (Axillary temperature)
Skin (Skin temperature)
AirW (Airway temperature)
Room (Room temperature)
Myo (Myocardial temperature)
Core (Core temperature)
Surf (Surface temperature)

16-3

B40/B20 Patient Monitor

Combining different temperatures


The monitor displays the difference between different temperatures if they are displayed in the
same digit field. differences (T2-T1,) are displayed in the temperature digit field if you choose
them in the same digit field. For example, to display T1+T2:
1.

Press the Monitor Setup key and select Screen Setup.

2.

Select Digit Fields.

3.

Select T1+T2 to one of the lower fields.

Testing temperature
The temperature measurement functioning is automatically tested periodically. During the
test, the message Performing temp test is displayed. If the test fails, the monitor displays the
message Temperature error.

Changing temperature units


You can select the temperature unit to be either degrees Centigrade or degrees Fahrenheit:
1.

Press the Others key.

2.

Temp Setup - Unit.

3.

Select the unit (C or F) with the Trim Knob.

Checklist
Check that:

16-4

Temperature adapter cable is properly inserted into the connector in the module, and the
probe is inserted into the adapter cable.

Temperature probe is positioned correctly.


Temperature is selected for screen through Monitor Setup - Screen Setup.

17 Airway gas

Airway gas (CO2)

17 Airway gas (CO2)


Safety precautions
Warnings

Always check the airway adapter for a tight connection and proper operation before
attaching it to the patient.

Remove the airway sampling line from the patients airway while nebulized medications
are being delivered.

Keep the monitor horizontal when the E-miniC module is used. Tilting the monitor may
cause erroneous results in the E-miniC module's readings and damage the module.

Leak in breathing circuit (water trap or sampling line) may cause inaccurate readings.

Strong scavenging suction may change the operating pressure of the module and cause
inaccurate readings or excessive sample gas flow.

Do not use gas or CO2 sidestream sampling modules on patients who may be adversely
affected by the specified withdrawal rates (e.g. a neonate with low tidal volume).

To avoid the spread of infectious disease, do not allow the exhaust to discharge in the
direction of the patient or user.

Handle the water trap and its contents as you would any body fluid. Infectious hazard
may be present.

CO2 measurement is intended for patients weighing over 5 kg (11 lb).

Do not apply pressurized air to any outlet or tubing connected to the monitor, pressure
may destroy sensitive elements.

Malfunction may cause gases to vent.

You can connect and disconnect airway modules like any other modules.

Blocked gas exhaust may cause inaccurate readings.


EtCO2 values may differ from blood gas readings.
Be sure CO2 measurement is off before removing water trap.
Never connect any tubing to reference gas inlet connector. The inlet must be open at all
times.

Since sample gas contains anesthetic agents, make sure that it is not released in the
room.

Cautions

Check the water trap every 48 hours at least.

Notes
For a comprehensive list of accessories, see the "Supplies and Accessories" catalog.
In the monitoring system, use only one module for measuring airway gases.

17-1

B40/B20 Patient Monitor

The message Sampling line blocked may result if you attach the sampling line to the
water trap after turning the monitor on.

When the warning symbol is displayed beside the O2 value FiO2, low alarm limit is set
below 21%.

Overview
With the E-miniC module, you can measure and monitor the gases being delivered to the
patient and exhaled by the patient.
The modules are equipped with a water separation system and they measure only CO2.
Respiration rate is the frequency of peak (end tidal) CO2 measurements per minute. A breath is
defined as a change in the CO2 signal which exceeds 1% (8 mmHg). All concentrations are
measured and displayed breath by breath.

Module description
The airway gas measurement is housed in the E-miniC module.
You can connect and disconnect airway modules like any other modules.

E-miniC module
for CO2 measurement

Figure 17-1 E-miniC module for CO2 measurement

17-2

Airway gas (CO2)

Display of gases

Figure 17-2 Airway gas waveform display


(1)

30 minute trend for CO2 (selected in the Monitor Setup menu)

(2)

Gas waveforms

(3)

Message field for gases

(4)

Gas label

(5)

Digit field for ET and FI gas values

(6)

Respiration rate

(7)

Scale

NOTE: When the measured CO2 value is outside the specified measurement range, the
numeric value is gray.

Figure 17-3 Lower digit field for gases

17-3

B40/B20 Patient Monitor

Patient connections
1.

Insert the E-miniC module in the monitor and push until it clicks.

2.

Check visually that the airway adapter connections are tight and that the adapter is
operating properly, then attach it to the patient.

3.

Make sure that the water trap container is empty and properly attached. The water trap
should be changed between patients and emptied whenever it is more than half full.

4.

Connect the gas sampling line to the sampling line connector on the water trap.

5.

Turn on the monitor. The monitor performs a self-check.

6.

Before connecting the patient, wait until the message Calibrating gas sensor disappears.
Then make the patient connections as described below and connect the sampling line to
the airway adapter.

7.

Position the adapter with sampling port upwards. This prevents any condensed water
from entering the sampling line.

If E-miniC module is used with O2 and/or N2O contents higher than 40%, make sure that FiO2
Level and N2O Level are set accordingly through Airway Gas - CO2 Setup to enable O2 and/
or N2O compensation.

Take the gas sample as close to the patients airway as possible, as shown in the illustrations,
and connect the sampling line to the patients airway adapter.
Position the adapters sampling port upwards to prevent any condensed water from entering
the sampling line.

NOTE: The message Sampling line blocked may result if you attach the sampling line
to the water trap after turning on the monitor.

17-4

Airway gas (CO2)

CO2 setup menu


Airway
Gas

Scale

Allows you to select the following scales:


scale options for %

scale options for kPa

scale options for mmHg

0-6%
0-10%
0-15%
0-20%

0 - 6 kPa
0 - 10 kPa
0 - 15 kPa
0 - 20 kPa

0 - 50 mmHg
0 - 80 mmHg
0 - 100 mmHg
0 - 160 mmHg

FiO2 Level

Selects the FiO2 level. The FiO2 level is used in gas compensations of the CO2 measurement to
increase the measurement accuracy. Make sure the level is set if the FiO2 level is higher than
40%.

N2O Level

Selects the N2O level. The presence of N2O causes the CO2 value to appear higher than the
actual value. Use this option to compensate for the presence of N2O.

Unit

Selects the CO2 display unit: %, kPa or mmHg.

Resp Rate Source


You can select the rate to be calculated from the ECG leads (impedance measurement) or the
CO2 measurement. If you select AUTO, the rate is automatically calculated from the measured
CO2. If impedance measurement is present and the source is AUTO, the respiration rate
calculation switches back to impedance respiration if you press the Silence Alarms key
during an Apnea alarm.
CO2 Alarm

Opens the CO2 alarm adjustment menu to change the CO2 alarm limits.

Resp Rate Alarm


Opens the Resp Rate alarm adjustment menu to change the respiration rate alarm limits.
Previous Menu Returns to the previous menu.

Selecting alarm sources


To select alarm sources, go to the CO2 Setup menus submenu CO2 Alarm. The selections are
FI or ET as the high and low alarm limit.

17-5

B40/B20 Patient Monitor

Calibrating
NOTE: Ensure that the calibration gas and regulator are functioning properly before
calibration. Perform annual maintenance of the regulator as required.
NOTE: Do not wash or disinfect calibration gas sampling lines.
The airway module should be calibrated once every six months or whenever there are
indications of errors in the gas readings.
Calibrate the gas measurement with the GE Healthcare calibration gas. Do not use any other
calibration gases.

Use the regulator 755534 or equivalent.

Use the gas 755580 only and set the CO2 concentration to 20%.

If you do not use the recommended calibration gases, the calibration does not succeed.
During gas calibration, % units are always used for CO2 regardless of selected measuring units.

Figure 17-4 Attaching regulator to calibration can


1.

Attach the regulator to the gas container.

2.

Attach a new sampling line to the water trap. Connect the loose end of the sampling line
to the regulator on the gas container.

Figure 17-5 Connecting sampling line to the gas valve and feeding gas

17-6

Airway gas (CO2)

3.

Turn on the power. For maximum accuracy, let the monitor warm up for 30 minutes. The
menu item Gas calibration remains gray as long as the message Calibrating gas sensor
is displayed.

4.

Press the Airway Gas key and select Gas calibration.

5.

Wait until Zero ok and then Feed gas messages are displayed after each gas on the
screen.

6.

Open the regulator and feed calibration gas until the message Adjust is displayed, then
close the valve. If you use an older brass regulator, the feeding pressure should be
adjusted between 5 and 7 psi.

7.

Check that the displayed gas values match the values on the calibration gas container.

NOTE: Set the CO2 level according to the gas, for example with 755580, set the CO2
level to 20%.
NOTE: If an error occurs during calibration or if no gas is fed, the highlight goes
automatically over the item Recalibrate and the text Calibr. error is displayed after
each gas. Push the Trim Knob to perform a new calibration.
If adjustments are required:

Turn the Trim Knob to highlight the first gas to be adjusted and then push the Trim Knob.
Turn the Trim Knob until the displayed value matches the desired value in the gas bottle
and push it again.

Repeat these two steps for each gas.

The use of an old regulator with the new aerosol cylinders requires an adapter available from
GE Healthcare. For ordering details, see the "Supplies and Accessories" catalog.
If the message Zero error is displayed, press the Normal Screen key and repeat the
calibration procedure. If the problem persists, contact authorized service personnel.
The monitor performs autozeroing after start up. The autozeroing intervals are: 4 min, 15 min,
30 min and 60 minutes after start-up, thereafter every 60 minutes.

NOTE: If you are using separate gas cylinders, calibrate each gas separately.
The time of the last calibration is shown at the bottom of the menu page.

17-7

B40/B20 Patient Monitor

Interfering gases
Non-disturbing gases are those with a maximum effect on the CO2 reading < 0.2 vol%. The
effect is valid for specific concentrations shown in parentheses of the non-disturbing gas:
Ethanol C2H5OH (<0.3%)
Acetone (<0.1%)
Methane CH4 (<0.2%)
Nitrogen N2
water vapor
Dichlorofluoromethane (<1%)
Tetrafluoroethane (<1%)
Disturbing gases and their effect on the CO2 reading at 5.0 vol-% CO2 are shown below. Errors
listed reflect the effect of specific concentrations (shown in parentheses) of an individual
disturbing gas and should be combined when estimating the effect of gas mixtures:
Halothane (4%) increases < 0.3 vol%
Isoflurane (5%) increases < 0.4 vol%
Enflurane (5%) increases < 0.4 vol%
Desflurane (24%) increases < 1.2 vol%
Sevoflurane (6%) increases < 0.4 vol%
Helium (50%) decreases < 0.3 vol%
If O2 compensation is not activated: O2 (40 to 95%) decreases < 0.3 vol%
If O2 compensation is activated: O2 (40 to 95%) error < 0.15 vol%
If N2O compensation is not activated: N2O (40%) increases < 0.4 vol%, N2O (40 to 80%)
increases < 0.8 vol%
If N2O compensation is activated: N2O (40 to 80%) error < 0.3 vol%

Unit conversions
Relationship between gas concentration and its partial pressure:
Reading in mmHg (dry gas) =
(ambient pressure in mmHg) x (gas concentration in%)
100
Reading in mmHg (water vapor saturated gas) =
(ambient pressure in mmHg - 47 mmHg) x gas concentration in%
100
Reading in kPa (dry gas) =
(ambient pressure in mmHg) x (gas concentration in%)
750
Reading in kPa (water vapor saturated gas) =
((ambient pressure in mmHg - 47mmHg) x (gas concentration in%)
750

NOTE: 47 mmHg is the partial pressure of the saturated water vapor at 37C.

17-8

Airway gas (CO2)

Checklist
Check that:

Water trap is locked into the module.


Water trap container is empty.
A new sampling line is used after each patient.
Sampling line is connected to the water trap.
Monitor is turned on and self-check is performed with the sampling line attached.
Sampling line is connected to the airway adapter.
The humidification and/or bacteria filter are in correct place.
Breathing circuit or accessories have no residuals of alcohol based disinfectants.
Desired gas parameter is selected for screen through Monitor Setup - Screen Setup.

17-9

B40/B20 Patient Monitor

17-10

Index

Index

B
Batteries
capacity 2-7
conditioning 2-9
indicators 2-8
Bladder temperature 16-3

A
Abbreviations 2-14
Adjusting
alarm light brightness 5-6
Admitting 4-5
Adults
cuff hose detection 9-3
inflation limit 13-5
Airway adapter
cleaning 9-6
Airway gases
calibrating 17-6
checklist 17-9
CO2 setup menu 17-5
display 17-3
sampling line 9-6
Airway sampling line 9-6
Airway temperature 16-3
Alarm
activation 5-4
AUDIO OFF 6-4
categories 5-3
changing the tone pattern 5-16
changing tone 6-4
deactivating 5-11
displaying limits 5-13
enabling and disabling silencing 5-14
latching alarms 5-15
PVC alarm limits 11-11
reactivating temporarily silenced 5-9
recording 5-12
reminder volume 5-16
Setup menu 5-5
silencing permanently 5-10
ST limit 11-14
Alarm tones 5-4
Arrhythmia
alarm settings 11-17
arrhythmia algorithm 11-18
detecting 11-17
selecting leads 11-18
Axillary temperature 16-3

C
Calibration
airway gas 17-6
NIBP 13-5
Canceling
alarm limit changes 5-5
Capacity
batteries 2-7
Central
setting time 6-2
Changing the water trap 9-7
Checklist
temperature 16-4
children 13-4
Cleaning
airway adapters 9-6
ECG cables 9-6
NIBP cuff and hose 9-6
parameter cables 9-6
temperature probes 9-6
color 13-4
Colors
changing parameter colors
Command Board keys 2-6
Configuration
digit fields 6-9
installation settings 6-3
split screen 6-10
trends 6-13
units 6-3
user modes 6-5
waveform fields 6-8
Connecting patients 4-4
Core temperature 16-3
Cuff
cuff hose 13-4

6-11

D
Date setup 6-2
Digit field
modifying 6-9
Display

Index
airway gases 17-3
ECG 11-3
heart rate 11-3
invasive blood pressure
SpO2 14-3
ST 11-12
temperature 16-2

I
15-2

E
ECG
3-lead 11-4
5-lead 11-4
adjusting size 11-9
cascaded waveforms 11-9
changing waveform sweep speed 6-11
cleaning of cables 9-6
filter 11-7
monitoring the ST segment 11-12
setup 11-7
starting relearning manually 11-10
user leads 11-9
Electrodes
ECG 11-4
Emptying the water trap 9-7
E-PSM 16-2
E-PSMP 16-2
Esophageal temperature 16-3

F
Filter
ECG 11-7
Functioning of the alarms

L
Label
ECG lead label 11-8
temperature 16-3
V lead label 11-8

M
Maintenance
water trap 9-6
Menus 4-1
Messages
system 10-2
Monitor
installation settings 6-1
Monitoring
starting 4-4
Myocardial temperature 16-3

N
9-3

G
Gas calibrating 17-6
Graphical trends 7-7

H
Half-annual maintenance 9-4
Heart rate
display 11-3
HR Source 11-7
setting limits 11-10
Hose
adult 13-4
children 13-4
color 13-4
infant 13-4
HR 11-7
2

Impedance respiration setup 12-4


infant 13-4
Installation checkout 3-4
Invasive blood pressure
patient connections 15-3
setup menu 15-5
Invasive blood pressure setup 15-3

Nasopharyngeal temperature 16-3


NIBP
cuff and hose cleaning 9-6
patient connections 13-4
setup 13-5
NIBP setup 13-4
Normal Screen
digit fields 6-9
minitrend length 6-7
waveform fields 6-8
Numerics 6-9

P
Parameter
colors 6-11
trended parameters 7-1
Patient connections
invasive blood pressure 15-3
NIBP 13-4
Preparations 4-4

Index
Printing
numerical trends 7-10
Printout
tabular trend printout 8-8
Problems
messages 10-2
other situations 10-8
Pulse oximetry setup 14-4
Pulse rate
display 11-7

monitoring ST segment 11-12


setting alarm limit 11-14
Surface temperature 16-3
Sweep speed
changing waveform sweep speed
Symbols
trend 7-6
System
configuration 6-1
messages 10-2

Recording
delay time 8-5
numerical trend format 8-8
numerical trends 7-10
selecting graphical trends 8-8
selecting recorded waveforms 6-12
Rectal temperature 16-3
Removing
water trap 9-7
Room temperature 16-3

Temperature
changing label 16-3
changing units 16-4
checklist 16-4
cleaning the probes 9-6
combining temperatures 16-4
display 16-2
label 16-3
testing 16-4
Temperature measurement setup 16-2
Time
setup 6-2
Trend
configuring 6-13
graphical pages 7-7
graphical trend view 7-5
printing numerical 7-10
recording numerical 7-10
symbols 7-6
trended parameters 7-1
Tympanic temperature 16-3

S
Sampling line 9-6
Scale
pleth 14-5
trend time scale 6-13
Screen setup
digit fields 6-9
parameter colors 6-11
waveform fields 6-8
Setup
changing units 6-3
digit field 6-9
ECG 11-7
NIBP 13-5
parameter colors 6-11
SpO2 14-5
temperature 16-3
temperature units 16-4
time and date 6-2
trends 6-13
waveforms 6-8
Size
ECG 11-9
impedance respiration waveform
Skin temperature 16-3
ST
display 11-12

6-11

U
Unit
changing for different parameters
temperature 16-4
User leads 11-9
User mode
renaming 6-6

6-3

12-6

Water trap 9-6


care 9-6
Waveform 6-8
modifying the field

6-8

Index

Installation and checkout form, B40/B20

APPENDIX A: Installation and checkout form,


B40/B20
Customer
Service
Service engineer

Date

OK = Test OK

N.A. = Test not applicable

Fail = Test failed

NOTE: Refer to the Technical reference manual for Electrical Safety tests procedure, Chapter
Maintenance.
Electrical Safety Tests

OK

N.A.

Fail

2.4. Power Outlet Test


2.5. Power cord and plug
2.6. Ground (Earth) Integrity
Ground continuity test
Impedance of protective earth connection

< 0.1 ohms/


< 0.2 ohms
< 0.1 ohms/
< 0.2 ohms

2.7. Ground (earth) wire leakage current tests


Normal Condition (NC)
Single Fault Condition (SFC)

< 500 A/
< 300 A
< 1 mA

2.8. Enclosure (Touch) leakage current test


Normal Condition (NC)
Single Fault Condition (SFC)

< 100 A/
< 500 A/
< 300 A

2.9. Patient (source) leakage current test


Normal Condition (NC)

< 10 A/

2-1
Document no. 2050802-001

B30 Patient Monitor

Electrical Safety Tests


Single Fault Condition (SFC)

OK

N.A.

Fail

OK

N.A.

Fail

OK

N.A.

Fail

< 50 A/

2.10. Patient (sink) leakage current test (mains voltage


< 50 A/
on the applied part)
Notes
Visual Inspection
. Visual inspection

Functional Inspection
. Start-up
. Display
. Frame unit
. Parameters measurements
. Recorder
. Network connection
. Conclusion
Notes

Signature

2-2
Document no. 2050802-001

Asian Headquarters
GE Medical Systems
Information Technologies, Inc.
8200 West Tower Avenue
Milwaukee, WI 53223 USA
Tel:+ 1 414 355 5000
1 800 558 5120 (US only)
Fax:+ 1 414 355 3790

GE Medical Systems
Information Technologies GmbH
Munzingerstrasse 5
79111 Freiburg
Germany
Tel: + 49 761 45 43 - 0
Fax: + 49 761 45 43 - 233

GE Medical Systems
Information Technologies Asia
1 Huatuo Road
Zhangjiang Hi-tech Park Pudong
Shanghai, P.R. China, 201203
Tel: + 86 21 3877 7888
Fax: + 86 21 3877 7451

GE Medical Systems Information Technologies, a General Electric Company, going to market as


GE Healthcare
www.gehealthcare.com

0459

You might also like