0% found this document useful (0 votes)
44 views5 pages

Absorbance of Iron Solutions of Varying Concentrations at 508nm

This document presents the results of an experiment measuring the absorbance of iron solutions at 508nm over a range of concentrations. A linear regression analysis of the absorbance data versus concentration yielded a very high R2 value of 0.9996, indicating absorbance increases linearly with iron concentration. The regression equation was then used to calculate the concentrations of unknown iron solutions based on their measured absorbances.

Uploaded by

JordanOGrady
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
44 views5 pages

Absorbance of Iron Solutions of Varying Concentrations at 508nm

This document presents the results of an experiment measuring the absorbance of iron solutions at 508nm over a range of concentrations. A linear regression analysis of the absorbance data versus concentration yielded a very high R2 value of 0.9996, indicating absorbance increases linearly with iron concentration. The regression equation was then used to calculate the concentrations of unknown iron solutions based on their measured absorbances.

Uploaded by

JordanOGrady
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
You are on page 1/ 5

Absorbance of Iron Solutions of Varying Concentrations at 508nm

0.6

y = 10523x + 0.0001
R = 0.9996

0.5

Absorbance at 508nm

0.4

0.3

Series1
Linear (Series1)
Linear (Series1)

0.2

0.1

0
-0.00001

-0.1

0.00001

0.00002

0.00003

Iron Concentration(moles/L)

0.00004

0.00005

0.00006

Conc.(M)

Sample
0 Blank
0 Blank
0 Blank
0.00000536
0.00000536
0.00000536
0.0000134
0.0000134
0.0000134
0.0000268
0.0000268
0.0000268
0.0000402
0.0000402
0.0000402
0.0000536
0.0000536
0.0000536

1
1
1
2
2
2
3
3
3
4
4
4
5
5
5

Absorbance at 508nm
-0.002
-0.001
-0.002
0.054
0.054
0.054
0.142
0.142
0.142
0.286
0.285
0.287
0.428
0.43
0.43
0.558
0.558
0.557

slope

10522.9348

Intercept

0.00014286

SD regress.

0.00449195 =

m
10522.9348 0.00014286
Sm
55.562255 0.00166926
r^2
0.99955413 0.00449195
F
35868.5674
16
ssreg
0.7237436 0.00032284
Calculated Conc. Of Unkown(M)
2.25E-05
M
3
SD of estimated results for unkown
2.66E-07

Unknown Fe sol. Absorbance


0.236
0.237
0.237
Average
0.236666667

b
Sb
Sy
DOF
sum of squares
Calculated Conc. Of Standard(M)
1.84857E-05
M
3
SD of estimated results for unkown
2.66231E-07

Standard Fe sol. Absorbance


0.195
0.195
0.194
Average
0.194666667

You might also like