ProSim 3 and 2

Vital Signs Simulators


Dont need a comprehensive patient monitor
tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim
2 Vital Signs Simulators are clear choices for
biomedical engineers and field service techni-
cians that need a quality, feature-rich device
with high portability. Choose one of these
modern vital signs simulators for preventive
maintenance, troubleshooting and repair.
The ProSim 3 and 2 feature the perfect
amount of features for testing in the field. We
like to call it the Just Right feature set, and it
includes:
ECG
pacemaker
arrhythmia and performance testing
respiration
invasive blood pressure
temperature
cardiac output (ProSim 3 only)
fetal/maternal (ProSim 3 only)
Technical Data
Portable, for evaluating the performance of patient monitors in the eld
20 % lighter and 25 % smaller than preceding technology
Just Right feature set includes: ECG, pacemaker, arrhythmia and performance testing,
respiration, invasive blood pressure, temperature, cardiac output (ProSim 3 only), fetal/
maternal (ProSim 3 only)
43 high-quality waveforms
With four IBP channels, ProSim 3 tests even the highest acuity scenarios
Stay-connected ECG posts for secure lead connections
Improved user interface and online Advantage Training demos
Upgraded DIN connectors ensure consistency with the ProSim family; minimize cable
compatibility issues
Field upgradeable, and easily paired with other devices for comprehensive testing
ProSim 3 and 2 are 510(k) approved products
Key features
2 Fluke Biomedical ProSim 3 and 2 Vital Signs Simulators
Specications
General specications
Temperature Operating 10 C to 40 C (50 F to +104 F)
Storage -25 C to +50 C
(-13 F to +122 F)
Humidity 10 % to 80 % non-condensing
Altitude 2,000 meters (6,562 ft)
Dimensions (LxWxH) 14.0 cm x 20.6 cm x 4.5 cm (5.5 in x 8.2 in x 1.8 in)
Display LCD greyscale display
Communication USB device upstream port
Power Two 9 V alkaline batteries
Battery life 8 hours continuous operation
Weight 0.47 kg (1 lb, 4 oz)
Safety standards IEC 61010-1, Pollution degree 2
Certifications CE, CSA, C-TICK N 10140, RoHS
Electromagnetic compatibility
(EMC)
IEC 61326-1; 2006
Detailed specications
Normal-sinus-rhythm waveform
ECG Reference The ECG amplitudes specified are for Lead II (calibration), from the
baseline to the peak of the R wave. All other leads are proportional.
Normal sinus rhythm 12-lead configuration with independent outputs referenced to right
leg (RL). Output to 10 universal ECG Jacks, color-coded to AHA and
IEC Standards
Amplitude 0.05 mV to 0.45 mV (0.05 mV steps); 0.5 mV to 5.5 mV (0.5 mV steps)
Amplitude accuracy 2 % of setting Lead II. All other leads 5 %
ECG rate 30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260,
280 and 300 BPM
Rate accuracy 1 % of setting
ECG waveform selection Adult (80 ms) or pediatric (40 ms) QRS duration
ST-segment elevation Adult mode only: -0.8 mV to +0.8 mV (0.1 mV steps)
Additional steps: +0.05 mV and -0.05 mV
Power-on default 80 BPM, 1.0 mV, adult QRS, ST-segment elevation of 0 mV, and a P-R
interval of 0.16 seconds
Pacemaker waveform
Pacer pulse Amplitude 0 (off), 1, 2, 5, 10 mV 10 % for
lead II (reference lead) with other
leads proportional as for perfor-
mance waves.
Accuracy Reference lead II: (5 % setting
+ 0.2 mV)
Pacer pulse width 0.1, 0.5, 1.0, 1.5, 2.0 ms 5%
3 Fluke Biomedical ProSim 3 and 2 Vital Signs Simulators
Paced arrhythmias Atrial 75 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-ventricular sequential
Noncapture (one time)
Nonfunction
Power-on default Off
Arrhythmia
Baseline NSR 80 BPM
PVC focus Left focus, standard timing (except where specified)
Supraventricular arrhythmia Atrial fibrillation (coarse or fine); atrial flutter sinus arrhythmia;
missed beat (one time); atrial tachycardia; paroxysmal atrial tachy-
cardia; nodal rhythm; and supraventricular tachycardia
Premature arrhythmia (All one-time events) Premature atrial contraction (PAC); premature
nodal contraction (PNC); PVC1 left ventricular; PVC1 left ventricular,
early; PVC1 left ventricular, R on T; PVC2 right ventricularl PVC2
right ventricular, early; PVC2 right ventricular, R on T; and multifocal
PVCs
Ventricular arrhythmia PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy;
trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); ventricu-
lar tachycardia; ventricular fibrillation (coarse or fine); and asystole
Conduction defect First-, second-, or third-degree AV block; and right- or left-bundle-
branch block
Power-on default None (off)
ECG performance testing
Amplitude 0.05 to 0.45 mV (0.05 mV steps), 0.5 to 5.5 mV (0.5 mV steps)
Pulse wave 30, 60 BPM, with 60 ms pulse width
Square wave 2.0, 0.125 Hz
Triangle wave 2.0, 2.5 Hz
Sine wave 0.5, 5, 10, 40, 50, 60, 100 Hz
R-wave detection waveform Haver-Triangle
R-wave rate 30, 60, 80, 120, 200, and 250 BPM
R-wave width 20 to 200 ms (10 ms steps) Additional Steps: 8, 10, and 12 ms
Rate accuracy 1 %
Amplitude accuracy 2 %, Lead II (Exception: 5 % for R waves 20 ms)
Power-on default None (off)
Fetal/Maternal ECG (ProSim 3 only)
Fetal heart rate (Fixed) 60, 90, 120, 140, 150, 210 and 240 BPM
Fetal heart rate (IUP) 140 BPM at beginning, then varies with pressure
Intrauterine-pressure waveforms Early deceleration, late deceleration, and uniform acceleration
Wave duration 90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg and
returning to 0
IUP period 2, 3, or 5 minutes; and manual
Power-on default FHR 120 BPM, early deceleration, manual
4 Fluke Biomedical ProSim 3 and 2 Vital Signs Simulators
Invasive blood pressure
Channels 4, each independently settable with identical parameters and are
individually electronically isolated from other signals
Input/output impedance 300 10 %
Exciter input range 2.0 to 16.0 V rms
Exciter-input frequency range DC to 5000 Hz
Transducer sensitivity 5 or 40 V/V/mmHg
Pressure accuracy 2 % of setting + 2 mmHg (valid for dc excitation only)
Static Levels, Channel 1 -10, 0, 80, 160, 240, 320, 400 mmHg
Static Levels, Channel 2 -10, 0, 50, 100, 150, 200, 240 mmHg
Static Levels, Channel 3
(ProSim 3 only)
-5, 0, 20, 40, 60, 80, 100 mmHg
Static Levels, Channel 4
(ProSim 3 only)
-5, 0, 20, 40, 60, 80, 100 mmHg
Dynamic waveforms, Channel 1 Arterial: 120/80, Radial Artery: 120/80, Left ventricle: 120/00,
Right ventricle: 25/00
Dynamic waveforms, Channel 2 Arterial: 120/80, Radial artery: 120/80, Left ventricle: 120/00,
Right atrium (central venous or CVP): 15/10, Right ventricle: 25/00,
Pulmonary artery: 25/10, Pulmonary-artery wedge: 10/2,
Left atrium: 14/4
Dynamic waveforms, Channel 3 Arterial: 120/80, Radial artery: 120/80, Left ventricle: 120/00,
Right atrium (central venous or CVP): 15/10, Right ventricle: 25/00,
Pulmonary artery: 25/10, Pulmonary-artery wedge: 10/2,
Left atrium: 14/4
Dynamic waveforms, Channel 4 Swan-Ganz sequence, Right atrium (CVP), Right ventricle (RV),
Pulmonary artery (PA), Pulmonary-artery wedge (PAW)
Respiration artifact BP delta changes from 3 to 16 mmHg
Output connector DIN 5-Pin
Power-on default 0 mmHg
Respiration
Rate 0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM
Waves Normal or ventilated
Ratio (inspiration: expiration) Normal 1:1, 1:2, 1:3, 1:4, 1:5
Ventilated 1:1
Impedance variations ( ) 0.2, 0.5, 1 or 3 peak-to-peak variation of lead impedance
Accuracy delta 10 %
Baseline 500, 1000, 1500, 2000 , Leads I, II, III
Accuracy baseline 5 %
Respiration lead LA or LL
Apnea selection OFF, 12, 22 or 32 seconds (one-time events), or continuous (Apnea
ON = respiration OFF)
Power-on default 20 BrPM, delta 1.0 , 1000 baseline
5 Fluke Biomedical ProSim 4 Vital Signs Simulator
Temperature
Temperature 0 C (32 F), 24 C (75.2 F), 37 F (98.6 C), and 40 C (104 F)
Accuracy 1 C
Compatibility Yellow Springs, Inc. (YSI) Series 400 and 700
Output connector Circular DIN 4-pin
Power-on default 0 C (42 F)
Cardiac output
Catheter type Baxter Edwards, 93a-131-7f
Calibration coefficient 0.542 (0 C) injectate), 0.595 (24 C injectate)
Blood temperature 37 C (98.6 F) 2 %
Injectate volume 10 cc
Injectate temperature 0 C or 24 C 2 % value
Cardiac output 2.5, 5, 10 liters per minute 5 %
Faulty-injectate curve Waveform for simulation available
Left-to-right shunt curve Waveform for simulation available
Calibrated pulse 1.5
Output connector DIN 7-PIN
Power-on default 2.5 liters perminute, 0 C injectate
6 Fluke Biomedical ProSim 3 and 2 Vital Signs Simulators
Fluke Biomedical.
Better products. More choices. One company.
Fluke Biomedical
6045 Cochran Road
Cleveland, OH 44139-3303 U.S.A.
Fluke Biomedical Europe
Science Park Eindhoven 5110
5692EC Son, The Netherlands
For more information, contact us:
In the U.S.A. (800) 850-4608 or
Fax (440) 349-2307
In Europe/M-East/Africa +31 40 267 5435 or
Fax +31 40 267 5436
From other countries +1 (440) 248-9300 or
Fax +1 (440) 349-2307
Email: [email protected]
Web access: www.ukebiomedical.com
2013 Fluke Biomedical. Specications subject
to change without notice. Printed in U.S.A.
3/2013 4263771A_EN
Modification of this document is not permitted
without written permission from Fluke Corporation.
About Fluke Biomedical
Fluke Biomedical is the worlds leading manufacturer of quality biomedical test
and simulation products. In addition, Fluke Biomedical provides the latest medical
imaging and oncology quality-assurance solutions for regulatory compliance.
Highly credentialed and equipped with a NVLAP Lab Code 200566-0 accredited
laboratory, Fluke Biomedical also offers the best in quality and customer service
for all your equipment calibration needs.
Today, biomedical personnel must meet the increasing regulatory pressures,
higher quality standards, and rapid technological growth, while performing their
work faster and more effciently than ever. Fluke Biomedical provides a diverse
range of software and hardware tools to meet todays challenges.
Fluke Biomedical Regulatory Commitment
As a medical test device manufacturer, we recognize and follow certain quality
standards and certifcations when developing our products. We are ISO 9001 and
ISO 13485 medical device certifed and our products are:
CE Certifed, where required
NIST Traceable and Calibrated
UL, CSA, ETL Certifed, where required
NRC Compliant, where required
Ordering information
Models/descriptions
ProSim 3 ProSim Vital Signs Simulator
ProSim 2 ProSim Vital Signs Simulator
Standard accessories
ProSim 2/3 Instruction Sheet (multi-language)
4253822 ProSim 2/3 Users Manual CD
614487 Two 9-volt alkaline batteries (minimum 8
hours continuous use)
2392173 IBP Cable, unterminated
2392199 3010-0289FG, CI-3 Cable Assembly
(Cardiac Output Box)
1671807 USB cable
2248623 ProSim 2/3 Carrying Case
AC power cords
4219453 AC/DC Power Supply
769422 AC Power Cord (Schuko)
284174 AC Power Cord (USA)
769455 AC Power Cord (UK)
658641 AC Power Cord (Australia)
2200218 AC Power Cord (Denmark)
2200229 AC Power Cord (India)
2200241 AC Power Cord (Israel)
2198785 AC Power Cord (Italy)
769448 AC Power Cord (Switzerland)
Optional accessories
2523334 YSI 400 Series (UT-4)
2199019 YSI 700 Series (UT-2)
4022300 Cardiac output switch for GE