0% found this document useful (0 votes)
51 views

Prob. 7.75 Data

This document contains calibration data for a rotameter, including the inner diameter and volume of the rotameter, fluid density, atmospheric pressure, temperature, and flow rate data for both air and water. The air flow data includes the rotameter scale reading and corresponding volumetric flow rate in cubic feet per second. The water flow data provides the scale reading, weight of water collected, and collection time.

Uploaded by

jarot
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLS, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
51 views

Prob. 7.75 Data

This document contains calibration data for a rotameter, including the inner diameter and volume of the rotameter, fluid density, atmospheric pressure, temperature, and flow rate data for both air and water. The air flow data includes the rotameter scale reading and corresponding volumetric flow rate in cubic feet per second. The water flow data provides the scale reading, weight of water collected, and collection time.

Uploaded by

jarot
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLS, PDF, TXT or read online on Scribd
You are on page 1/ 1

Data for Problem 7.

75: Calibration of a Rotameter

d, in. V, in.^3 ρf, slug/ft^3 Hatm, in. Hg T, deg F


1.40 1.50 15.1 29.05 78

Air Flow Data

SR Q, ft^3/s
14.6 0.23
21.5 0.32
28.1 0.41
33.6 0.49
39.2 0.56
44.8 0.64
50.2 0.71
55.9 0.8
63.1 0.89
68.6 0.97
73.5 1.05
76.2 1.08

Water Flow Data

SR W, lb t, s
13.1 6.52 19.9
18.5 8.01 17.7
24.2 7.02 10.4
28.2 7.81 10.1
37.1 8.20 8.4
45.7 9.21 7.5
52.6 8.19 5.7