0% found this document useful (0 votes)
53 views1 page

History Background Document: IP 15-8-1 Process Stream Analyzers

This document provides background history for corresponding paragraphs in IP 15-8-1 (Rev. 0) on process stream analyzers. It notes requirements for analyzers containing toxic materials to ensure safe handling. It also specifies requirements for analyzers using radioactive sources to ensure containment and certification for operating in electrically classified areas. The document is meant to convey minimum performance standards and safety information for installing and operating different types of process stream analyzers.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
53 views1 page

History Background Document: IP 15-8-1 Process Stream Analyzers

This document provides background history for corresponding paragraphs in IP 15-8-1 (Rev. 0) on process stream analyzers. It notes requirements for analyzers containing toxic materials to ensure safe handling. It also specifies requirements for analyzers using radioactive sources to ensure containment and certification for operating in electrically classified areas. The document is meant to convey minimum performance standards and safety information for installing and operating different types of process stream analyzers.
Copyright
© Attribution Non-Commercial (BY-NC)
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

INTERNATIONAL PRACTICE

HISTORY BACKGROUND DOCUMENT


Process Stream Analyzers

IP 15-8-1
Page 1 of 1 Rev. 0 June 1997

BACKGROUND HISTORY DOCUMENTATION FOR CORRESPONDING PARAGRAPHS OF IP 15-8-1 (REV. 0)

S S

[H] [H]

4.2.i 4.4

Required in case analyzer contains toxic materials, e.g., Beryllium windows in x-ray analyzers. Required for safe handling and operation of radioactive/ionizing sources which are used in, e.g., x-ray absorption analyzers. Ensures containment of process sample. Ensures analyzer is correctly certified and safe to operate in electrically classified areas. Specifies minimum performance requirements for process stream analyzers. Performance specifications are not available in Industry Standards. Ensures flame type detectors in gas chromatographs cannot be a source of ignition for external flammable vapors. Avoids any potential hydrocarbon or toxic gas release on start-up of the analyzer. Ensures minimum information required to safely install and operate the analyzer is displayed on the analyzer.

S S O

[H] [H] [H]

5.4 5.5 5.13

[H]

6.6

S S

[H] [H]

8.9 9.1

Notes: (1) (2) Each paragraph to identify the Purpose Code(s) of the corresponding paragraph number of the applicable International Practice (IP). The symbol [H] is used to denote all History Background Paragraphs which are for INFORMATION ONLY.

THIS INFORMATION FOR AUTHORIZED COMPANY USE ONLY EXXON RESEARCH AND ENGINEERING COMPANY FLORHAM PARK, N.J.