Agilent 6611C-07A Random Error 113
Agilent 6611C-07A Random Error 113
Agilent 6611C-07A Random Error 113
6611C-07A
S E R V I C E
N O T E
Supersedes: 6611C-07
On faster computers, when programming over GPIB, a random error will occur.
To Be Performed By: Agilent-Qualified Personnel Parts Required: P/N
Description
Qty.
0699-3058 5063-4841
1 1
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY: LOCATION CATEGORY: AVAILABILITY: AUTHOR: MC [[]] IMMEDIATELY x ON SPECIFIED FAILURE [[]] AGREEABLE TIME [[]] CUSTOMER INSTALLABLE [[]] ON-SITE x SERVICE CENTER PRODUCTS SUPPORT LIFE PRODUCT LINE: 33 STANDARDS: LABOR: 1.0 Hours [[]] RETURN USED [[]] RETURN x SCRAP PARTS: x SCRAP [[]] SEE TEXT [[]] SEE TEXT AGILENT RESPONSIBLE UNTIL: obsolescence SERVICE INVENTORY:
ADDITIONAL INFORMATION:
November 2, 2005
Page 1 of 2
Page 2 of 2
6611C-07A
Situation: When programming a unit over GPIB, a random error -113 (Undefined header) will appear on the unit. This error can be read beck via the front panel or over the GPIB bus.
Solution/Action: There are two potential solutions. Solution 1 Change resistor R1246 (see picture for location) from 511 Ohms to 100 Ohms (part number 06993058). This requires un-soldering and soldering a surface mount component and can be very difficult. This does not require re-calibration.
Solution 2 Replace the entire GPIB assembly (part number 5063-4841) with a newer one that has the 100 Ohm resistor. This will require the unit to be re-calibrated.