Agilent 6611C-07A Random Error 113

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

MODIIFIICATIION RECOMMENDED MOD F CAT ON RECOMMENDED CORRECTS MANUFACTURIING OR DESIIGN DEFECTS CORRECTS MANUFACTUR NG OR DES GN DEFECTS

6611C-07A

S E R V I C E

N O T E
Supersedes: 6611C-07

Agilent 6611C DC Power Supply


Serial Numbers: MY43000488/MY43000759

On faster computers, when programming over GPIB, a random error will occur.
To Be Performed By: Agilent-Qualified Personnel Parts Required: P/N

Description

Qty.

0699-3058 5063-4841

100 Ohm resistor GPIB Assembly, tested

1 1

ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:

MODIFICATION RECOMMENDED
ACTION CATEGORY: LOCATION CATEGORY: AVAILABILITY: AUTHOR: MC [[]] IMMEDIATELY x ON SPECIFIED FAILURE [[]] AGREEABLE TIME [[]] CUSTOMER INSTALLABLE [[]] ON-SITE x SERVICE CENTER PRODUCTS SUPPORT LIFE PRODUCT LINE: 33 STANDARDS: LABOR: 1.0 Hours [[]] RETURN USED [[]] RETURN x SCRAP PARTS: x SCRAP [[]] SEE TEXT [[]] SEE TEXT AGILENT RESPONSIBLE UNTIL: obsolescence SERVICE INVENTORY:

ADDITIONAL INFORMATION:

AGILENT TECHNOLOGIES, INC. 2005 PRINTED IN U.S.A.

November 2, 2005
Page 1 of 2

Page 2 of 2

6611C-07A

Situation: When programming a unit over GPIB, a random error -113 (Undefined header) will appear on the unit. This error can be read beck via the front panel or over the GPIB bus.

Solution/Action: There are two potential solutions. Solution 1 Change resistor R1246 (see picture for location) from 511 Ohms to 100 Ohms (part number 06993058). This requires un-soldering and soldering a surface mount component and can be very difficult. This does not require re-calibration.

Solution 2 Replace the entire GPIB assembly (part number 5063-4841) with a newer one that has the 100 Ohm resistor. This will require the unit to be re-calibrated.