Flow & Level Calibration Notes: Differential Pressure Transmitter Calibration
Flow & Level Calibration Notes: Differential Pressure Transmitter Calibration
Adjust the zero first, since span error is corrected only after an accurate zero is established. Zero is properly set when a 10% input produces a 10% output. Adjust the span at 90%. Since zero and span frequently interact, after one of these errors has been corrected, the other may require readjustment.
Five-Point Check
To begin the calibration of a magnetic flowmeter, calculate the input signal value. The input signal is equal to the upper range multiplied by the calibration factor and by the phase band factor. These values are indicated on the instrument's data plate. Input Signal = Upper Range x Calibration Factor x Phase Band Factor Record the output values at each test point, and from this data determine if the instrument is within manufacturer's specifications. The following formula tells if the range of error is within manufacturer's specifications: Accuracy = (Deviation / Span ) * 100 Deviation = Expected Value - Actual Value Adjust zero at the lowest point in the instrument's range by turning the zero adjust screw until the output reading is correct. Then adjust span, and , since zero and span often interact, verify both until no further adjustment is necessary. To conclude the calibration, recheck the upscale and downscale readings to verify that the instrument is properly calibrated.
Before making the connections, determine the vortex shedding frequency. The vortex shedding frequency is usually provided by the manufacturer, but if it is not listed in the manufacturer's literature, calculate the frequencies using this formula: Vortex Shedding Frequency = RF x CF x ( URV/TIME) where: The vortex shedding frequency is represented in pulses per second or PPS. PPS = represents the alternate shedding of vortices on either side of the bluff body RF = stands for reference factor which can be found on the transmitter's data plate and is usually represented in pulses per US gallon CF = is the conversion factor, and is a number found in the manufacturer's conversion table, the CF converts the RF to actual volume or mass flow rate units URV = is the upper range value in US gallons per minute TIME = is related to the increment in which the flow is measured Are span jumpers required to calibrate this transmitter? Refer to the manufacturer's instructions for the appropriate setting for the coarse span jumpers. Once the PPS has been determined and span jumpers set in their proper positions, the frequency generator can be connected to the input terminals of the transmitter. The output side of the calibration loop is connected in series. After all the connections are made, set the fine span.
Correcting Errors
To establish the zero and span values, the flow tube must be full of process liquid with no flow. Because actual process fluid is used, this procedure is typically done on an installed transmitter. Press the "format" key followed by the "proceed" key. Then press the "autozero" key. The autozero automatically establishes zero to the properties of the process fluid. The resulting display will indicate that the mass flowmeter is properly calibrated.
Five-Point Check
Determine the instrument's range and test points for calibration. For the lower range value measured in inches of water, divide the minimum height of the liquid in inches by the liquid's specific gravity. The upper range value is the maximum height of the liquid in inches of water divided by its specific gravity. The span then, is the difference between these values. Perform the five-point upscale and downscale check. Correct the zero at 10% of input span, adjusting zero until the output produced is 10% of the output span. Next, correct the span error, applying 90% input and adjusting the span until 90% output is produced.
Calibrate the transmitter span first before compensating zero for transmitter height location.
Calibration
When the chamber is empty, the corresponding output should be 4 mA. If the milliammeter displays a value that is greater than or less than 4 mA, adjust the zero. To correct span, fill the chamber to the upper range value, and turn the span adjustment until 20 mA is produced. Linearity is not always adjustable on this type of transmitter, check to manufacturers specifications. Adjust both zero and span until transmitter performs within specifications.
Liquid-Liquid Interface
This same sensing principle used to measure liquid vapor interface, can be used to locate the interface between two liquids. The heavier of the two liquids exerts more buoyant force, so as the lower phase rises or falls, the displacer travels with it. To create calibration conditions that replace the process, use liquids that have the same specific gravity as the process fluids. Fill the chamber with the lighter phase to check and adjust the zero. Check zero by filling the chamber to 100% with the lighter fluid. Check span by filling the chamber 100% full of the heavier phase. Adjust span until 20 mA output is produced. Check mid-range output and recheck zero and span.