Whitepaper Precision Data for the Cement Industry Vitriox Electric (2)
Whitepaper Precision Data for the Cement Industry Vitriox Electric (2)
Whitepaper Precision Data for the Cement Industry Vitriox Electric (2)
The data show consecu�ve prepara�ons in the same fusion machine done
on the same day, except of where it is men�oned differently.
RV
10 times 1 day Unit 1 2 3 4 5 6 7 8 9 10 Mean STD Diff
result
Cement 04.08.'15 04.08.'15 04.08.'15 04.08.'15 04.08.'15 04.08.'15 04.08.'15 04.08.'15 04.08.'15 04.08.'15
Al2O3 % 3,628 3,627 3,626 3,612 3,622 3,615 3,641 3,629 3,628 3,642 3,627 0,010 3,640 -0,013
CaO % 64,189 64,222 64,215 64,083 64,233 64,165 64,152 64,074 64,150 64,204 64,169 0,055 64,100 0,069
Cr2O3 % 0,015 0,016 0,015 0,015 0,016 0,017 0,017 0,015 0,017 0,017 0,016 0,001 0,019 -0,003
Fe2O3 % 4,958 4,971 4,983 4,967 4,976 4,964 4,979 4,982 4,976 4,967 4,972 0,008 5,030 -0,058
K2O total % 0,687 0,684 0,681 0,685 0,688 0,686 0,686 0,688 0,685 0,683 0,685 0,002 0,665 0,020
Loss on ignition % 0,980 0,980 0,990 0,990 0,980 0,980 0,990 0,990 1,000 1,000 0,988 0,008 1,200 -0,212
MgO % 0,834 0,824 0,824 0,816 0,830 0,817 0,825 0,823 0,816 0,818 0,823 0,006 0,815 0,008
Mn2O3 % 0,096 0,095 0,095 0,093 0,095 0,092 0,093 0,094 0,095 0,093 0,094 0,001 0,100 -0,006
Na2O % 0,190 0,202 0,200 0,205 0,203 0,210 0,200 0,202 0,212 0,201 0,202 0,006 0,200 0,002
P2O5 % 0,387 0,387 0,388 0,385 0,390 0,387 0,390 0,385 0,390 0,386 0,387 0,002 0,380 0,007
SiO2 % 20,598 20,577 20,662 20,633 20,653 20,625 20,630 20,598 20,643 20,647 20,626 0,027 20,610 0,016
SO3 % 2,937 2,941 2,963 2,953 2,953 2,962 2,940 2,952 2,963 2,983 2,955 0,014 2,940 0,015
SrO % 0,043 0,040 0,038 0,038 0,034 0,038 0,040 0,037 0,036 0,037 0,038 0,002 0,040 -0,002
TiO2 % 0,226 0,224 0,228 0,226 0,227 0,225 0,224 0,225 0,229 0,228 0,226 0,002 0,230 -0,004
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark
FXBA-1017-01 2/8
RV
10 days 1 time Unit 1 2 3 4 5 6 7 8 9 10 Mean STD Diff
result
Cement 30.07.'15 31.07.'15 04.08.'15 05.08.'15 06.08.'15 07.08.'15 10.08.'15 11.08.'15 12.08.'15 13.08.'15
Al2O3 % 3,621 3,635 3,650 3,631 3,630 3,645 3,643 3,625 3,641 3,637 3,636 0,009 3,640 -0,004
CaO % 64,196 64,259 64,288 64,085 64,273 64,125 64,203 64,234 64,158 64,143 64,196 0,068 64,100 0,096
Cr2O3 % 0,016 0,016 0,017 0,016 0,017 0,016 0,017 0,016 0,014 0,015 0,016 0,001 0,019 -0,003
Fe2O3 % 4,976 4,979 4,976 4,959 4,971 4,973 4,980 4,982 4,983 4,975 4,975 0,007 5,030 -0,055
K2O total % 0,688 0,679 0,682 0,680 0,681 0,676 0,679 0,678 0,682 0,676 0,680 0,004 0,665 0,015
Loss on ignition % 1,020 1,040 1,000 1,050 1,010 1,080 0,980 0,980 1,070 1,030 1,026 0,035 1,200 -0,174
MgO % 0,800 0,819 0,805 0,811 0,811 0,834 0,820 0,822 0,825 0,827 0,817 0,011 0,815 0,002
Mn2O3 % 0,094 0,096 0,097 0,095 0,094 0,096 0,096 0,094 0,095 0,093 0,095 0,001 0,100 -0,005
Na2O % 0,203 0,223 0,207 0,203 0,195 0,205 0,183 0,201 0,198 0,190 0,201 0,011 0,200 0,001
P2O5 % 0,385 0,389 0,393 0,385 0,386 0,385 0,386 0,386 0,384 0,389 0,387 0,003 0,380 0,007
SiO2 % 20,620 20,614 20,596 20,624 20,590 20,592 20,604 20,580 20,611 20,592 20,602 0,015 20,610 -0,008
SO3 % 2,955 2,967 2,957 2,962 2,937 2,948 2,947 2,952 2,950 2,964 2,954 0,009 2,940 0,014
SrO % 0,042 0,035 0,041 0,039 0,042 0,035 0,039 0,040 0,037 0,040 0,039 0,003 0,040 -0,001
TiO2 % 0,229 0,228 0,227 0,226 0,226 0,226 0,227 0,228 0,229 0,230 0,228 0,001 0,230 -0,002
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark
FXBA-1017-01 3/8
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark
FXBA-1017-01 4/8
CEM II/A-V
Unit 1 2 3 4 5 Mean STD
52,5N
Al2O3 % 8,086 8,081 8,063 8,077 8,100 8,081 0,013
CaO % 55,217 55,184 55,199 55,196 55,225 55,204 0,017
Cr203 % 0,011 0,012 0,010 0,011 0,011 0,011 0,001
Fe2O3 % 3,427 3,431 3,424 3,433 3,429 3,429 0,003
K2O % 1,154 1,153 1,153 1,149 1,158 1,153 0,003
Loss on ignition % 1,810 1,910 1,980 1,880 1,840 1,884 0,066
MgO % 0,968 0,965 0,954 0,958 0,968 0,962 0,006
Mn2O3 % 0,161 0,163 0,159 0,162 0,160 0,161 0,001
Na2O % 0,249 0,255 0,262 0,256 0,251 0,255 0,005
P2O5 % 0,410 0,407 0,413 0,410 0,412 0,410 0,002
SiO2 % 25,502 25,474 25,535 25,471 25,563 25,509 0,040
SO3 % 2,438 2,452 2,448 2,429 2,468 2,447 0,015
SrO % 0,048 0,045 0,046 0,045 0,049 0,047 0,002
TiO2 % 0,382 0,384 0,381 0,384 0,387 0,384 0,002
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark
FXBA-1017-01 5/8
Loss on ignition
as received
% 4,770 4,650 4,890 4,700 4,960 4,794 0,129
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark
FXBA-1017-01 6/8
Loss on ignition
as received
% 1,170 1,140 1,160 1,190 1,170 1,166 0,018
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark
FXBA-1017-01 7/8
Loss on ignition
as received
% 42,860 42,810 42,880 42,880 42,860 42,858 0,029
Loss on ignition
as received
% 34,470 34,490 34,450 34,510 34,530 34,490 0,032
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark
FXBA-1017-01 8/8
Natural RV
Unit 1 2 3 4 5 Mean STD Diff
Gypsum result
Al2O3 % 2,606 2,612 2,622 2,597 2,598 2,607 0,011 2,630 -0,023
CaO % 25,038 25,042 25,056 25,052 25,036 25,045 0,009 25,090 -0,045
Cr2O3 % 0,004 0,003 0,004 0,003 0,005 0,004 0,001 0,050 -0,046
Fe2O3 % 1,041 1,042 1,026 1,037 1,032 1,035 0,007 1,040 -0,005
K2O % 0,619 0,623 0,622 0,622 0,619 0,621 0,002 0,610 0,011
Loss on ignition
950 °C
expressed in % 19,990 20,010 20,020 20,050 20,030 20,020 0,022 19,990 0,030
after drying at
45 °C
MgO % 5,670 5,682 5,688 5,681 5,683 5,681 0,007 5,630 0,051
Mn2O3 % 0,014 0,015 0,015 0,014 0,015 0,015 0,000 0,020 -0,005
Na2O % 0,002 0,004 0,013 0,001 -0,001 0,004 0,006 0,070 -0,066
P2O5 % 0,039 0,038 0,038 0,039 0,037 0,038 0,001 0,045 -0,007
SiO2 % 10,705 10,726 10,704 10,703 10,747 10,717 0,019 10,810 -0,093
SO3 % 33,323 33,480 33,377 33,381 33,485 33,409 0,071 33,740 -0,331
SrO % 0,340 0,344 0,339 0,341 0,347 0,342 0,003 0,355 -0,013
TiO2 % 0,150 0,147 0,146 0,151 0,151 0,149 0,002 0,135 0,014
RAW V03
(BAS 683- Unit 1 2 3 4 5 Mean STD Certificate Diff
1)
Al2O3 % 2,466 2,470 2,457 2,468 2,459 2,464 0,006 2,440 0,024
CaO % 7,924 7,932 7,895 7,905 7,927 7,917 0,016 7,890 0,027
Cr2O3 % 0,033 0,032 0,034 0,032 0,033 0,033 0,001 0,023 0,010
Fe2O3 % 79,784 79,748 79,679 79,646 79,749 79,721 0,057 79,400 0,321
K2O % 0,158 0,161 0,157 0,158 0,157 0,158 0,002 0,176 -0,018
Loss on ignition
as received
% -0,870 -0,860 -0,860 -0,850 -0,800 -0,848 0,028
MgO % 1,833 1,849 1,852 1,856 1,879 1,854 0,017 1,700 0,154
Mn2O3 % 0,689 0,690 0,693 0,685 0,680 0,687 0,005 0,657 0,030
Na2O % 0,057 0,049 0,053 0,058 0,050 0,053 0,004 0,060 -0,007
P2O5 % 0,354 0,335 0,338 0,340 0,339 0,341 0,007 0,336 0,005
SiO2 % 7,267 7,263 7,254 7,243 7,239 7,253 0,012 7,160 0,093
SrO % 0,012 0,010 0,011 0,011 0,010 0,011 0,001
TiO2 % 0,151 0,152 0,153 0,152 0,153 0,152 0,001 0,160 -0,008
Tel 49(0)2821 997320/ Fax 49(0)2821 9973229, e-mail: [email protected], web: www.fluxana.com, FLUXANA® GmbH & Co.KG, Borschelstr.3, 47551 Bedburg-Hau Germany,
GM Dr. R. Schramm, reg.in Kleve HRA 2935, personal liability FLUXANA Verwaltungs-GmbH Bedburg-Hau, reg.in Kleve HRB 8211, FLUXANA®registered trademark