Guia Incerteza 2020
Guia Incerteza 2020
Guia Incerteza 2020
a) ܻ = 5 ∗ ܺଵ ∗ ܺଶ + 3 ∗ ܺଷ
y con pdf normal), y los siguientes datos medidos, calcular el resultado de Y=y±u(y)
Temp (ºC) Tensión (V) Temp (ºC) Tensión (V) Temp (ºC) Tensión (V) Temp (ºC) Tensión (V)
-20 -0.939 0 0.059 20 1.060 40 2.061
-19 -0.880 1 0.123 21 1.121 41 2.122
-18 -0.821 2 0.178 22 1.178 42 2.181
-17 -0.772 3 0.228 23 1.230 43 2.229
-16 -0.728 4 0.273 24 1.271 44 2.272
-15 -0.688 5 0.308 25 1.309 45 2.311
-14 -0.651 6 0.349 26 1.350 46 2.349
-13 -0.608 7 0.390 27 1.391 47 2.389
-12 -0.560 8 0.442 28 1.443 48 2.440
-11 -0.502 9 0.497 29 1.498 49 2.499
-10 -0.439 10 0.559 30 1.561 50 2.560
-9 -0.376 11 0.620 31 1.620 51 2.620
-8 -0.320 12 0.680 32 1.680 52 2.679
-7 -0.270 13 0.731 33 1.731 53 2.730
-6 -0.230 14 0.772 34 1.770 54 2.771
-5 -0.189 15 0.809 35 1.811 55 2.808
-4 -0.153 16 0.848 36 1.849 56 2.848
-3 -0.110 17 0.892 37 1.892 57 2.891
-2 -0.061 18 0.940 38 1.940 58 2.940
-1 -0.002 19 0.997 39 2.000 59 3.000
5) Dados los siguientes intervalos de confianza (IC) y suponiendo el nivel de confianza
(NC) del intervalo y función densidad de probabilidad indicados, proponga valores para
μ(X) y σ(X) de las pdf (Lea la seccion 4 de la nota técnica 1297 del NIST –
NIST_tn1297s.pdf – como ayuda).
a) IC=[7,8; 9,1], NC=90%, pdf: normal
b) IC=[1,4; 1,8], NC=95%, pdf: normal
c) IC=[-7; -5], NC=100%, pdf: rectangular
d) IC=[-1,5; 1,5], NC=100%, pdf: triangular
e) IC=[4; 4,5], NC=100%, pdf: rectangular
8) Se tiene una PT100 que cumple la norma IEC-751 Clase A (100Ω ± 0.06% @ 0°C, TCR =
0.00385Ω/Ω°C, modelo del sensor R(T) = R0 * (1+TCR*(T-T0)) ), a la cual se le hace
circular una corriente y se mide la tensión que aparece sobre ella, ambos valores
reportados con un factor de cobertura de 1. Dada la siguiente información, hallar el
valor de la temperatura medida con un 99% de confianza. Aclare las suposiciones que
realiza.
I= 1±0.01mA V= 122±0.2mV
1
=ܩ
1 1
ܣ௩ + ቌ
ቍ
ܴ݂
1 + ܴ1