Title | Calibration of the X-ray Tomograph |
Author(s) | Avramidou, R ; Berbiers, Julien ; Boudineau, C ; Drakoulakos, D G ; Fabjan, Christian Wolfgang ; Grau, S ; Gschwendtner, E ; Krumshtein, Z ; Maugain, J M ; Rangod, Stephane ; Rieder, H ; Rohrbach, F ; Sbrissa, E ; Schuh, S ; Sedykh, E ; Sedykh, Yu ; Smirnov, Y ; Vertogradov, L S ; Woudstra, M ; Vichou, I |
Publication | 2000 |
Imprint | 01 Mar 2000 |
Number of pages | 14 |
Note | revised version number 1 submitted on 2001-01-31 16:08:44 |
Subject category | Detectors and Experimental Techniques |
Accelerator/Facility, Experiment | CERN LHC ; ATLAS |
Free keywords | ATLAS ; Calibration ; Tomograph ; X-ray ; accuracy ; assurance ; chambers ; measurement ; muon ; quality |
Abstract | An X-ray Tomograph has been constructed to measure the mechanical accuracy of the ATLAS Muon chambers. Here we report on the choice of the measurement scheme, the global calibration procedure and practical aspects of the calibration. We reached for the measurement accuracy a value of 2 /mum systematic and 2 /mum statistical uncertainties in the horizontal and vertical directions. The measurement area is up to 220 cm (horizontal) by 60 cm (vertical). |