CMS Note
Report number CMS-CR-2017-281
Title High precision, low disturbance calibration of the High Voltage system of the CMS Barrel Electromagnetic Calorimeter
Author(s) Marzocchi, Badder (INFN, Rome ; Rome U.)
Publication 2017
Collaboration CMS Collaboration
Imprint 22 Sep 2017
Number of pages 4
Presented at New Developments in Photodetection 2017, Tours, France, 3 - 7 Jul 2017
Subject category Detectors and Experimental Techniques
Accelerator/Facility, Experiment CERN LHC ; CMS
Keywords ECAL
Abstract The CMS Electromagnetic Calorimeter is made of scintillating lead tungstate crystals, using avalanche photodiodes (APD) as photo-detectors in the barrel part. The high voltage system, consisting of 1224 channels, biases groups of 50 APD pairs, each at a voltage of about 380 V. The APD gain dependence on the voltage is 3pct/V. A stability of better than 60 mV is needed to have negligible impact on the calorimeter energy resolution. Until 2015 manual calibrations were performed yearly. A new calibration system was deployed recently, which satisfies the requirement of low disturbance and high precision. The system is discussed in detail and first operational experience is presented.
Copyright/License Preprint: (License: CC-BY-4.0)

 