მთავარი > Recent exploits of the ISOLTRAP mass spectrometer |
Article | |
Title | Recent exploits of the ISOLTRAP mass spectrometer |
Author(s) | Kreim, S (CERN ; Heidelberg, Max Planck Inst.) ; Kowalska, M (CERN) ; Naimi, S (Heidelberg, Max Planck Inst.) ; Blaum, K (Heidelberg, Max Planck Inst.) ; Breitenfeldt, M (Leuven U.) ; Rossel, R E (Mainz U., Inst. Phys. ; Fachhochschule Wiesbaden) ; Fink, D (Heidelberg, Max Planck Inst. ; Heidelberg U.) ; Stanja, J (Dresden, Tech. U.) ; Atanasov, D (Heidelberg, Max Planck Inst.) ; Borgmann, Ch (Heidelberg, Max Planck Inst.) ; Cocolios, T E (CERN ; U. Manchester (main)) ; Zuber, K (Dresden, Tech. U.) ; Wolf, R N (Greifswald U.) ; George, S (Heidelberg, Max Planck Inst.) ; Neidherr, D (Darmstadt, GSI) ; Nicol, T (ENSICAEN) ; Rosenbusch, M (Greifswald U.) ; Lunney, D (CSNSM, Orsay) ; Boehm, Ch (Heidelberg, Max Planck Inst.) ; Manea, V (CSNSM, Orsay) ; Herlert, A (FAIR, Darmstadt) ; Koester, U (Laue-Langevin Inst.) ; Beck, D (Darmstadt, GSI) ; Wienholtz, F (Greifswald U.) ; Kellerbauer, A (Heidelberg, Max Planck Inst.) ; Ramirez, E Minaya (Darmstadt, GSI ; Helmholtz Inst., Mainz) ; Schweikhard, L (Greifswald U.) |
Publication | 2013 |
Note | 16th International Conference on ElectroMagnetic Isotope Separators and Techniques Related to their Applications (EMIS), RIKEN Nishina Ctr Accelerator Based Sci, Matsue, JAPAN, DEC 02-07, 2012 |
In: | Nucl. Instrum. Methods Phys. Res., B 317 (2013) 492-500 |
In: | 16th International Conference on Electromagnetic Isotope Separators and Techniques Related to their Applications, Matsue, Japan, 2 - 7 Dec 2012, pp.492-500 |
DOI | 10.1016/j.nimb.2013.07.072 |
Subject category | Detectors and Experimental Techniques ; Nuclear Physics - Experiment |
Accelerator/Facility, Experiment | CERN ISOLDE |
Abstract | The Penning-trap mass spectrometer ISOLTRAP, located at the isotope-separator facility ISOLDE (CERN), is presented in its current form taking into account technical developments since 2007. Three areas of developments are presented. The reference ion sources have been modified to guarantee a sufficient supply of reference ions for mass measurements and systematic studies. Different excitation schemes have been investigated for manipulation of the ion motion in the Penning trap, to enhance either the purification or measurement process. A multi-reflection time-of-flight mass separator has been implemented and can now be routinely used for purification and as a versatile tool for beam analysis. (C) 2013 Elsevier B.V. All rights reserved. |