Abstract 


Background

Delivery of cardiac resynchronization therapy (CRT) requires left ventricular myocardial capture to achieve clinical benefits.

Objective

We sought to determine whether ineffective pacing affects survival.

Methods

Ineffective ventricular pacing (VP) was defined as the difference between the percentage of delivered CRT (%VP) and the percentage of effective CRT in CRT devices. Using the Optum de-identified electronic health record data set and Medtronic CareLink data warehouse, we identified patients implanted with applicable devices with at least 30 days of follow-up. Kaplan-Meier and Cox proportional hazards models assessed the impact of %VP and % ineffective VP on survival.

Results

Among 7987 patients with 2.1 ± 1.0 years of follow-up, increasing ineffective VP was associated with decreasing survival: the highest observed survival was in the quartile with <0.08% ineffective VP and the lowest survival was in the quartile with >1.47% ineffective VP (85.1% vs 75.7% at 3 years; P < .001). As expected, patients with more than the median %VP of 97.7% had better survival than did patients with <97.7% VP (84.2% vs 77.8%; P < .001). However, patients who had >97.7% VP but >2% ineffective VP had similar survival to patients with <97.7% VP but ≤2% ineffective VP (81.6% vs 79.4%; P = .54). A multivariable Cox proportional hazards model demonstrated that <97.7% VP (adjusted hazard ratio 1.29; 95% confidence interval 1.14-1.46; P < .001) and >2% ineffective VP (hazard ratio 1.35; 95% confidence interval 1.18-1.54; P < .001) were both significantly associated with decreased survival.

Conclusion

Ineffective VP is associated with decreased survival. In addition to maximizing the percentage of delivered CRT pacing, every effort should be made to minimize ineffective VP.

Citations & impact 


This article has not been cited yet.

Impact metrics

Alternative metrics

Altmetric item for https://www.altmetric.com/details/165925033
Altmetric
Discover the attention surrounding your research
https://www.altmetric.com/details/165925033

Similar Articles 


To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.


Funding 


Funders who supported this work.

Medtronic

    Medtronic Inc