Pàgina inicial > CERN Experiments > LHC Experiments > ATLAS > ATLAS Preprints > A Gas Leak Rate Measurement System for the ATLAS MUON BIS-Monitored Drift Tubes |
Internal Note and Article | |
Report number | SN-ATLAS-2002-016 ; ATL-COM-MUON-2001-024 |
Title | A Gas Leak Rate Measurement System for the ATLAS MUON BIS-Monitored Drift Tubes |
Author(s) | Alexopoulos, T (Natl. Tech. U., Athens) ; Avramidou, R ; Dris, M ; Filippas-Tassos, A ; Gazis, E N ; Katsoufis, E C ; Maltezos, S ; Savva, Panagiota S ; Stavropoulos, G D ; Tsipolitis, G |
Publication | 2004 |
Imprint | 20 Oct 2001 |
Number of pages | 18 |
Note | Accepted as Scientific Note |
In: | Nucl. Instrum. Methods Phys. Res., A 521 (2004) 332-342 |
DOI | 10.1016/j.nima.2003.10.095 |
Subject category | Detectors and Experimental Techniques |
Accelerator/Facility, Experiment | CERN LHC ; ATLAS |
Free keywords | Gas leak rate measurement ; Muon drift detectors ; QA_QC |
Abstract | A low-cost, reliable and precise system developed for the gas leak rate measurement of the BIS-Monitored Drift Tubes (MDTs) for the ATLAS Muon Spectrometer is presented. In order to meet the BIS-MDT mass production rate, a total number of 100 tubes are tested simultaneously in this setup. The pressure drop of each one of the MDT is measured, within a typical time interval of 48 hours, via a differential manometer comparing with the pressure of a gas tight reference tube. The precision of the method implemented is based on the system temperature homogeneity, with accuracy of ÄT = 0.3 oC. For this reason, two thermally isolated boxes are used testing 50 tubes each of them, to achieve high degree of temperature uniformity and stability. After measuring several thousands of the MDTs, the developed system is confirmed to be appropriate within the specifications for testing the MDTs during the mass production. |
Copyright/License | Preprint: (License: CC-BY-4.0) |