Report number | CERN-PH-EP-2006-014 |
Title | A segmented Hybrid Photon Detector with integrated auto-triggering front-end electronics for a PET scanner |
Author(s) | Chesi, Enrico Guido ; Braem, André ; Joram, C ; Mathot, S ; Séguinot, Jacques ; Weilhammer, P ; Ciocia, F ; De Leo, R ; Nappi, E ; Vilardi, I ; Argentieri, A ; Corsi, F ; Dragone, A ; Pasqua, D |
Affiliation | (CERN) |
Publication | 2006 |
Imprint | 08 May 2006 |
Number of pages | 26 |
In: | Nucl. Instrum. Methods Phys. Res., A 564 (2006) 352-363 |
DOI | 10.1016/j.nima.2006.05.084 |
Subject category | Detectors and Experimental Techniques ; Health Physics and Radiation Effects |
Abstract | We describe the design, fabrication and test results of a segmented Hybrid Photon Detector with integrated auto-triggering front-end electronics. Both the photodetector and its VLSI readout electronics are custom designed and have been tailored to the requirements of a recently proposed novel geometrical concept of a Positron Emission Tomograph. Emphasis is put on the PET specific features of the device. The detector has been fabricated in the photocathode facility at CERN. |